Accuracy improvement of quantitative analysis in laser-induced breakdown spectroscopy using modified wavelet transform X. H. Zou,1,2 L. B. Guo,1,2 M. Shen,1 X. Y. Li,1,* Z. Q. Hao,1 Q. D. Zeng,1 Y. F. Lu,1 Z. M. Wang,1 and X. Y. Zeng1 1

Wuhan National Laboratory for Optoelectronics (WNLO), Huazhong University of Science and Technology, Wuhan, Hubei 430074, China 2 These authors contributed equally to this work * [email protected]

Abstract: A modified algorithm of background removal based on wavelet transform was developed for spectrum correction in laser-induced breakdown spectroscopy (LIBS). The optimal type of wavelet function, decomposition level and scaling factor γ were determined by the root-meansquare error of calibration (RMSEC) of the univariate regression model of the analysis element, which is considered as the optimization criteria. After background removal by this modified algorithm with RMSEC, the rootmean-square error of cross-validation (RMSECV) and the average relative error (ARE) criteria, the accuracy of quantitative analysis on chromium (Cr), vanadium (V), cuprum (Cu), and manganese (Mn) in the low alloy steel was all improved significantly. The results demonstrated that the algorithm developed is an effective pretreatment method in LIBS to significantly improve the accuracy in the quantitative analysis. ©2014 Optical Society of America OCIS codes: (100.7410) Wavelets; (300.6365) Spectroscopy, laser induced breakdown; (160.3900) Metals.

References and links 1.

W. B. Lee, J. Y. Wu, Y. I. Lee, and J. Sneddon, “Recent applications of laser-induced breakdown spectrometry: a review of material approaches,” Appl. Spectrosc. Rev. 39(1), 27–97 (2004). 2. Z. Wang, T. B. Yuan, Z. Y. Hou, W. D. Zhou, J. D. Lu, H. B. Ding, and X. Y. Zeng, “Laser-induced breakdown spectroscopy in China,” Front. Phys. (2014). 3. L. B. Guo, Z. Q. Hao, M. Shen, W. Xiong, X. N. He, Z. Q. Xie, M. Gao, X. Y. Li, X. Y. Zeng, and Y. F. Lu, “Accuracy improvement of quantitative analysis by spatial confinement in laser-induced breakdown spectroscopy,” Opt. Express 21(15), 18188–18195 (2013). 4. P. Fichet, P. Mauchien, J. F. Wagner, and C. Moulin, “Quantitative elemental determination in water and oil by laser induced breakdown spectroscopy,” Anal. Chim. Acta 429(2), 269–278 (2001). 5. L. Dudragne, P. Adam, and J. Amouroux, “Time-resolved laser-induced breakdown spectroscopy: application for qualitative and quantitative detection of fluorine, chlorine, sulfur, and carbon in air,” Appl. Spectrosc. 52(10), 1321–1327 (1998). 6. J. P. Singh and S. N. Thakur, Laser-Induced Breakdown Spectroscopy (Elsevier Science, 2007). 7. Z. Wang, Z. Y. Hou, S. L. Lui, D. Jiang, J. M. Liu, and Z. Li, “Utilization of moderate cylindrical confinement for precision improvement of laser-induced breakdown spectroscopy signal,” Opt. Express 20(S6), A1011– A1018 (2012). 8. T. Fujimoto, Plasma Spectroscopy (Clarendon, 2004). 9. X. G. Shao, A. K. M. Leung, and F. T. Chau, “Wavelet: a new trend in chemistry,” Acc. Chem. Res. 36(4), 276– 283 (2003). 10. C. M. Galloway, E. C. Le Ru, and P. G. Etchegoin, “An iterative algorithm for background removal in spectroscopy by wavelet transforms,” Appl. Spectrosc. 63(12), 1370–1376 (2009). 11. X. G. Ma and Z. X. Zhang, “Application of wavelet transform to background correction in inductively coupled plasma atomic emission spectrometry,” Anal. Chim. Acta 485(2), 233–239 (2003). 12. C. X. Ma and X. G. Shao, “Continuous wavelet transform applied to removing the fluctuating background in near-infrared spectra,” J. Chem. Inf. Comput. Sci. 44(3), 907–911 (2004).

#207574 - $15.00 USD (C) 2014 OSA

Received 3 Mar 2014; revised 11 Apr 2014; accepted 14 Apr 2014; published 21 Apr 2014 5 May 2014 | Vol. 22, No. 9 | DOI:10.1364/OE.22.010233 | OPTICS EXPRESS 10233

13. J. Schlenke, L. Hildebrand, J. Moros, and J. J. Laserna, “Adaptive approach for variable noise suppression on laser-induced breakdown spectroscopy responses using stationary wavelet transform,” Anal. Chim. Acta 754, 8– 19 (2012). 14. B. Zhang, L. X. Sun, H. B. Yu, Y. Xin, and Z. B. Cong, “Wavelet denoising method for laser-induced breakdown spectroscopy,” J. Anal. At. Spectrom. 28(12), 1884–1893 (2013). 15. T. B. Yuan, Z. Wang, Z. Li, W. D. Ni, and J. M. Liu, “A partial least squares and wavelet-transform hybrid model to analyze carbon content in coal using Laser-induced breakdown spectroscopy,” Anal. Chim. Acta 807, 29–35 (2014). 16. X. G. Shao, W. S. Cai, and Z. X. Pan, “Wavelet transform and its applications in high performance liquid chromatography (HPLC) analysis,” Chemom. Intell. Lab. Syst. 45(1–2), 249–256 (1999). 17. S. G. Mallat, “A theory of multiresolution signal decomposition: the wavelet representation,” IEEE Trans. Pattern Anal. 11(7), 674–693 (1989). 18. S. G. Mallat, A Wavelet Tour of Signal Processing: The Sparse Way (Academic, 2008). 19. X. G. Shao, L. M. Shao, and G. W. Zhao, “Extraction of extended X-ray absorption fine structure information from the experimental data using the wavelet transform,” Anal. Commun. 35(4), 135–137 (1998). 20. D. Chen, X. G. Shao, B. Hu, and Q. D. Su, “A background and noise elimination method for quantitative calibration of near infrared spectra,” Anal. Chim. Acta 511(1), 37–45 (2004). 21. I. Daubechies, Ten Lectures on Wavelets (Society for Industrial and Applied Mathematics, 1992). 22. A. E. Kramida Y. Ralchenko, J. Reader, and NIST ASD Team, “NIST Atomic Spectra Database (version 5.1)” (National Institute of Standards and Technology, 2013), http://physics.nist.gov/asd. 23. L. X. Sun and H. B. Yu, “Automatic estimation of varying continuum background emission in laser-induced breakdown spectroscopy,” Spectrochim. Acta, B At. Spectrosc. 64(3), 278–287 (2009).

1. Introduction As a promising analytical technique, laser-induced breakdown spectroscopy (LIBS) based on laser-material interactions has a wide range of applications, including geological analysis, pharmaceutical industry, and environmental monitoring [1–5], owing to its advantages of rapid detection, no samples pretreatment, real-time analysis and remote measurement [6, 7]. In LIBS, elemental concentration is determined by the peak intensities in plasma emission spectra. However, as collected LIBS spectra are often overwhelmed by intense continuum background, which interferes the true intensity of spectra and compromises spectral clarity, and hence reduces the accuracy of quantitative analysis. In LIBS, continuum background is generated by the bremsstrahlung radiation, recombination radiation, as well as stray lights [8]. As the delay time increases, isolated characteristic lines go down along with background. The signal-to-background ratio (SBR) of the peak intensities increases initially and then decreases. To obtain a maximum SBR, a time-resolved detector is indispensable. However, for a nongated detector for applications requiring portability and cost effectiveness, background subtraction is particularly important. Therefore, finding an effective compensation algorithm for accurate estimation and subtraction of background to reduce background interference and enhance spectral SBR, is essential for improving the accuracy of quantitative analysis. Wavelet transform (WT) is a new signal processing technique which has undergone rapid development in the past two decades. Compared with Fourier transform, WT has the characteristics of time-frequency localization and multi-resolution analyses of signals [9]. Given the superiority for the analysis of non-stationary signals, WT is an extremely powerful tool for correction of spectra, which has been extensively applied in many fields, including Raman spectroscopy [10], inductively coupled plasma atomic emission spectroscopy [11], and near-infrared spectroscopy [12]. Recently, a few researches have explored the wavelet denoising in LIBS. Schlenke et al. adopted stationary WT via adaptive variable threshold for noise reduction, which successfully achieved noise suppression and signal preservation [13]. Zhang et al. established a new double threshold optimization model of wavelet denoising, and demonstrated the performance of the method for both synthetic and actual LIBS signals [14]. However, its application in the background subtraction of LIBS has not been systematically studied. Yuan et al. implemented WT to subtract background for LIBS application by utilizing the wavelet coefficients instead of the spectral information after reconstruction for the first time, and analyzed quantitatively the high-concentration carbon element in coal [15]. Shao et al. proposed a concept of factor of the low-frequency part of the highest level of

#207574 - $15.00 USD (C) 2014 OSA

Received 3 Mar 2014; revised 11 Apr 2014; accepted 14 Apr 2014; published 21 Apr 2014 5 May 2014 | Vol. 22, No. 9 | DOI:10.1364/OE.22.010233 | OPTICS EXPRESS 10234

wavelet decomposition to amend the estimated background in high performance liquid chromatography (HPLC) [9, 16]. The value of the factor was just estimated by the operator according to the position of simulated background and the results were judged only by observation of the chromatogram. This method was not suitable for batch processing of spectra data and unable to evaluate the accuracy of the factor. In this study, a modified algorithm based on WT for automatically subtracting the continuum background was proposed for the purpose of low concentration elements analysis. Through considering the root-mean-square error of calibration (RMSEC) of the univariate regression model as the criterion, the three most important parameters, wavelet function, decomposition level and the scaling factor γ were optimized and discussed in details. To verify the validity, seven low alloy steel samples were used to compare the results with and without using the algorithm, which confirmed the effectiveness of the algorithm to improve the quantitative analyses. 2. Algorithm description WT is a signal decomposition method. According to the multiresolution signal decomposition algorithm proposed by Mallat [17], a given signal f(t) can be expanded to the linear combination of a series of wavelet bases [18]. Generally, the spectrum contains three main types of frequency information: high-frequency noise, low-frequency background, and medium-frequency peaks. The traditional background subtraction method [11, 19, 20] based on discrete WT considered that the low-frequency part of the highest level of wavelet decomposition only contained background information, and it was completely removed before signal reconstruction to achieve the purpose of background correction. However, for the real spectrum, the low-frequency part of wavelet decomposition often consists of the continuum background and part of the spectral peaks. Considering that the background occupies a certain proportion in the low-frequency component, the corresponding wavelet approximation coefficients should be multiplied by a scaling factor γ which varies from 0 to 1. The expansion equation of the reconstructed signal can be written as f ( t ) =  d j , kψ j , k ( t ) + (1-γ )  cJ , k φJ , k ( t ) (1) j ≤ J k ∈Z

j − 2

k ∈Z

j − 2

where ψ j , k (t ) = 2 ψ ( 2− j t − k ) , and φ j , k (t ) = 2 φ ( 2− j t − k ) . ψ (t ) and φ (t ) are the mother and father wavelets (scaling function). j denotes the scale corresponding to the frequency. k is related to the translation corresponding to the spectral wavelength. J represents the scale of the highest decomposition level. dj,k and cJ,k are the wavelet detail coefficients and wavelet approximation coefficients. Both terms in Eq. (1) represent the detail part (high-frequency component) and approximation part (low-frequency component) of the signal, respectively. Obviously, the result of background correction depends on the wavelet function, decomposition level and scale factor γ. The Daubechies wavelet family was adopted [15, 19– 21]. It should be given priority to the selection of the wavelet function and decomposition level, and then determine γ, which can effectively extract background information from the low-frequency component of the optimal wavelet decomposition and then further optimize the result of background estimation. The background-corrected spectra are used to establish the univariate regression model. The RMSEC and the determination coefficient (R2 c) of the calibration curve, the root-mean-square error of cross-validation (RMSECV) and the determination coefficient (R2 v) of the leave-one-out cross-validation curve, as well as the average relative error (ARE) are used as the evaluation parameters to test the performance of the proposed method. To improve the accuracy of the regression model after background correction, the RMSEC was used as the optimization criterion to obtain the optimal wavelet function, decomposition level and γ. The program used for background subtraction of spectra was written in Matlab R2010b. The algorithm includes the following steps: #207574 - $15.00 USD (C) 2014 OSA

Received 3 Mar 2014; revised 11 Apr 2014; accepted 14 Apr 2014; published 21 Apr 2014 5 May 2014 | Vol. 22, No. 9 | DOI:10.1364/OE.22.010233 | OPTICS EXPRESS 10235

(1) Select the wavelet function which varies from db1 to db10, and choose the decomposition level which ranges between 4 and 13. (2) Set the approximation coefficients cJ,k to zero, which is equivalent to γini = 1. Then reconstruct the spectra. Extract the intensity information of analysis element to build the calibration regression model, and calculate the RMSEC. (3) Determine the optimum values of the wavelet function (W) and decomposition level (L) which make the RMSEC achieve the minimum. (4) Perform wavelet decomposition again using the wavelet function (W) and decomposition level (L) for the spectra. Then solve the optimum γopt by the least squares method, and the RMSEC is still considered as the optimization goal. That is, γopt satisfies Eq. (2) RMSEC (W , L, γ opt ) = RMSEC min (W , L, γ ) (2) (5) Multiply cJ,k by (1-γopt) for the spectra and then reconstruct the spectral signal according to Eq. (1). Finally, the background-corrected spectra are obtained. 3. Experimental

Fig. 1. Schematic diagram of the experimental setup. Table 1. Composition of Cr, V, Cu, Mn and Fe elements from the low alloy steel samples N o. 1 2 3 4

Cr 0.164 0.322 0.092 0.409

Concentration (wt.%) V Cu Mn 0.010 0.154 0.151 0.023 0.201 1.100 0.010 0.105 0.571 0.040 0.054 1.500

Fe 97.883 97.462 96.851 96.372

No. 5 6 7

Cr 0.601 0.062 0.036

Concentration (wt.%) V Cu Mn 0.053 0.021 2.000 0.007 0.288 0.110 0.244 0.346 0.043

Fe 95.614 98.845 97.605

The LIBS experimental apparatus used in this study is schematically shown in Fig. 1. A Qswitched Nd:YAG laser (wavelength: 532 nm, repetition rate: 3 Hz, pulse width: 8 ns, pulse energy: 80 mJ) was used to obtain spectral data under air environment. The laser beam was reflected and then focused onto the surface of the low alloy steel sample by a lens with a focal length of 15 cm. The plasma emission was coupled into an optical fiber by a light collector, then collected by an echelle spectrometer (Andor Tech., Mechelle 5000) with a wavelength range of 200–950 nm and a spectral resolution of λ/Δλ = 5000 . A digital delay generator was adopted to trigger the laser pulses and control the gate delays and widths of the ICCD (Andor Tech., iStar DH-334T). The detector was operated in the gated mode whose delay time and gate width were set to 2 and 10 μs, respectively, to obtain high spectral intensity and SBR. To reduce the effect of the laser energy fluctuation on the spectral intensity, each spectrum accumulated 30 pulses and six spectra were taken for each sample and averaged. Seven low alloy steel samples with main elements of Cr, V, Cu, and Mn in the matrix element Fe were used in this study. Table 1 lists the certified concentrations of these elements.

#207574 - $15.00 USD (C) 2014 OSA

Received 3 Mar 2014; revised 11 Apr 2014; accepted 14 Apr 2014; published 21 Apr 2014 5 May 2014 | Vol. 22, No. 9 | DOI:10.1364/OE.22.010233 | OPTICS EXPRESS 10236

4. Results and discussions To verify the accuracy of the background correction algorithm, the quantitative analysis was performed for the four elements (namely, Cr, V, Cu, and Mn) in the seven low alloy steel standard samples. According to the Atomic Spectra Database [22], together with the observation of actual spectra, the analyte lines of Cr I 425.43 nm, V II 311.07 nm, Cu I 327.40 nm and Mn I 476.64 nm were chosen. Cr was used as an example element to demonstrate how the wavelet function, decomposition level and scaling factor γ influence the RMSEC. 4.1 Optimization of the wavelet function and decomposition level As described above, the wavelet function and decomposition level should be firstly determined. Theoretically, to obtain the best background estimation, the structure of the wavelet function selected should be similar to a processed spectrum as much as possible. However, it is too difficult to achieve. Practically, the most suitable filter can be quickly determined by the specific optimization goal of the spectral analysis. In this study, RMSEC was used as the evaluation criterion for the optimal wavelet and level. As presented in Fig. 2, the minimal value of RMSEC could be obtained when the wavelet function was db8 and decomposition level was 10. The background information was certainly stored in the approximation coefficients of the highest level of the wavelet decomposition.

Fig. 2. RMSEC values at different wavelet functions and decomposition levels.

4.2 Optimization of the scaling factor γ

Fig. 3. Influence of the scaling factor γ. (a) The effect of γ on the RMSEC with optimal decomposition level for each wavelet function. (b) Comparison of one original and its corrected spectrum without and with γ at wavelet function was db8 and decomposition level was 10.

As described in Section 4.1, compared to the wavelet function, the decomposition level is an important parameter. We analyzed the effect of the scaling factor γ on the RMSEC under the condition of optimal decomposition level corresponding to each specific wavelet function, as depicted in Fig. 3(a). Based on the above analysis and performing procedure 4 in the algorithm, the optimum value of the scaling factor γ was acquired. Figure 3(a) shows that all the RMSEC could become smaller by taking the scaling factor of the low-frequency #207574 - $15.00 USD (C) 2014 OSA

Received 3 Mar 2014; revised 11 Apr 2014; accepted 14 Apr 2014; published 21 Apr 2014 5 May 2014 | Vol. 22, No. 9 | DOI:10.1364/OE.22.010233 | OPTICS EXPRESS 10237

component into consideration. For db8 and 10 decomposition levels, γ = 0.89. Take a spectrum of the sample No. 5 for example, the original and background-corrected spectrum can be seen from Fig. 3(b). It was clear that the background of the spectrum reduced after applying the algorithm and avoided overestimation compared with the traditional method without introducing γ. The traditional method means fully discarding the lowest frequency part of wavelet coefficients. 4.3 Comparison of quantitative analysis To further verify the effectiveness of the background subtraction, the similar procedure was implemented for V, Cu, and Mn as well. For comparison, the polynomial fitting method proposed by Sun et al. [23] was also adopted for background subtraction. The detailed data are summarized in Table 2. In comparison with other methods, the proposed algorithm can further optimize the results. The improvement for Cr, V, Cu and Mn was obtained with lower RMSEC(V), higher R2 and lower ARE. The above analysis revealed that the proposed method can effectively improve the accuracy of the regression model. Table 2. Influence of different background subtraction methods on the univariate regression model Calibration regression curve Element

Cr I 425.43

V II 311.07

Cu I 327.40

Mn I 476.64

Method Original Polynomial WT (without γ) WT (with γ) Original Polynomial WT (without γ) WT (with γ) Original Polynomial WT (without γ) WT (with γ) Original Polynomial WT (without γ) WT (with γ)

RMSEC (wt.%) 0.0360 0.0174 0.0101 0.0091 0.0088 0.0072 0.0024 0.0024 0.0279 0.0254 0.0147 0.0142 0.1554 0.0739 0.0855 0.0855

R2 c 0.9674 0.9925 0.9974 0.9981 0.9907 0.9921 0.9992 0.9991 0.9475 0.9649 0.9875 0.9867 0.9686 0.9908 0.9903 0.9903

ARE (%) 41.02 14.41 10.64 8.54 33.57 48.69 13.28 13.21 16.50 45.86 8.51 7.66 84.92 31.37 22.83 22.83

Leave-one-out cross-validation regression curve RMSECV R2 v ARE (%) (wt.%) 0.0506 0.9418 56.35 0.0207 0.9896 16.75 0.0147 0.9948 13.97 0.0133 0.9963 11.28 0.0219 0.9732 39.01 0.0148 0.9822 60.83 0.0041 0.9987 17.51 0.0035 0.9987 17.01 0.0347 0.9237 24.14 0.0344 0.9398 59.19 0.0164 0.9827 11.65 0.0160 0.9820 11.06 0.2036 0.9463 119.30 0.1067 0.9803 41.58 0.1017 0.9856 30.99 0.1017 0.9856 30.99

5. Conclusions A modified algorithm based on wavelet transform for background subtraction was developed in this study for quantitative analysis in LIBS spectra of the low alloy steel samples. The method developed requires no prior knowledge of the selection of the proper background points and the mathematical assumption of the background distribution. The results obtained show that for the four elements of Cr, V, Cu, and Mn, their RMSEC(V), R2 and ARE are all significantly improved, successfully proving the validity of the background correction algorithm. This method can effectively improve the quality of the signals and the accuracy of the regression model. Accordingly, the proposed method can be used as a competitive preprocessing tool for LIBS and other spectral analysis. Acknowledgments This research has been financially supported by the National Special Fund for the Development of Major Research Equipment and Instruments (No. 2011YQ160017), by the National Natural Science Foundation of China (No. 51128501), and by the Fundamental Research Funds for the Central Universities of China (No. CXY13Q022).

#207574 - $15.00 USD (C) 2014 OSA

Received 3 Mar 2014; revised 11 Apr 2014; accepted 14 Apr 2014; published 21 Apr 2014 5 May 2014 | Vol. 22, No. 9 | DOI:10.1364/OE.22.010233 | OPTICS EXPRESS 10238

Accuracy improvement of quantitative analysis in laser-induced breakdown spectroscopy using modified wavelet transform.

A modified algorithm of background removal based on wavelet transform was developed for spectrum correction in laser-induced breakdown spectroscopy (L...