Edwin Ronald
E. Frederick, W. Hahn,
BS
PhD2 #{149} Michael #{149} Gerald Entine,
R. Squillante, PhD
Accurate Automatic for Mammography:
terms: #{149} Physics
Breast
radiography,
#{149} Radiography,
much
1991;
is one of the techniques for
ic range.
safer.
Problems
still
must
subjectively
degree
technology,
exist,
how-
anticipate
of exposure
based
breast compression ly adjust a “darkness”
technology,
The need frequently
178:393-396
in the
and
x-ray
detectors
manualknob.
used,
is
software studied.
control Improvements
I
Hunt
From
Radiation
St,
Watertown,
Monitoring MA
Devices,
02172.
Received
Inc.
Current
c RSNA,
address:
1991
Eaton
Corp.
Beverly,
Mass.
existing
units
and
The
system
ride
al-
detectors.
design
mithms
are
uses
prototype to be
mammography
to be built
tector
into
four The
and
new
ones.
cadmium details
tellu-
of the
software
discussed
MATERIALS
44
April
26, 1990; revision requested June 20; revision received September 7; accepted September 1 1. Supported in part by National Cancer Institute contract N44-CM-67807. Address reprint requests to MRS. 2
has
added
onto
laboratory the potential
de-
algo-
the
diagnostic
0.1
mm
energy
dynam-
range
is less
telluride.
than
Cadmium
telluride
detectors
(Radiation
that
Monitoring
Mass)
were
were
Devices,
fabricated
Wa-
on crystal-
were
etched
polished
and chemically
in
to
provide smooth, clean electrodes were applied uum deposition. Because the detectors
surfaces. Metallic by means of vac-
small
have
and
breasts
are
may
variations
in density,
age breast
attenuation
relatively significant
estimating
the
requires
aver-
that multi-
ple detectors be used to monitor a representative area of the image. If an insufficient area is monitored, the system could calculate an incorrect exposure time for most mined
of the breast by examination
of radiographs
image.
that even
tissue
adequate
for
with
It was of a large
performance
highly
the
deternumber
textured
signals
from
three
colinear detectors. The signal from the detector receiving the middle-value flux level was used to determine the exposure.
controller.
wide
detector
1 cm wafers
breast
METHODS
and
x-ray
sensitivity to lowwide dynamic range, and a Photovoltaic cadmium telmeet these requirements.
line wafers of cadmium telluride diameter and 2 mm thick. The
ed
sensitivity
for the high
in cadmium
tertown,
Such
high
system.
tellunide is a semiconductor with a band gap of 1.47 eV, which makes it nearly ideal for use at room temperature. This cornbination of properties provides excellent sensitivity to x rays and little noise. The
Research into the new AEC system was divided into two areas: (a) detector design and (b) development of data analysis and control algorithms. Development of the detector required modifying the design of an existing x-ray detector that providboth
calithe
Design
The requirements were that it have
is obtained
below.
AND
empirical
to optimize
of the
Detector
used
had to be made both in the design of the x-ray detector and in software for data analysis and control. The result is a stand-alone system that
and
measurements
performance
in-
of fundamental
relationships
bration
cadmium
harden-
ing of the x-ray beam, and failure of the photographic reciprocity law (1,2). To address these problems, a new AEC detector system complete with the appropriate gomithms was
physical
of algorithms
combination
Cadmium telluride provides a high x-ray attenuation because it has an average atomic number of 50 and a density of 6.2 g/cm3 (0.62 kg/rn3). At 30 keV, for example, the mean free path of x rays within
the
for such subjective input attributed to limitations
the
energy x rays, fast response. luride detectors
on the
then control
#{149}
Development
volved
even, in obtaining adequate control of the exposure times to prevent the need for performing the procedure again. Many designs for automatic exposure controllers (AECs) have been tested, but most do not work well over the typical mange of breast size and density encountered. With most AECs, an experienced technolo-
00.11
Radiology
AMMOGRAPHY most reliable
MS
Controller Performance’
early diagnosis of breast cancer. Over the past few years, advances in the design of mammography machines and in film have significantly meduced the dose of x nays received by a patient and have made the procedure
gist Index 00.11
J Cirignano,
#{149} Leonard
Exposure Design and
M
An automatic exposure controller has been designed that controls the optical film density for film, screen, and radiographic techniques typically used in mammography to within 0.05 over a range of 1.3-6.7cm thickness of Lucite. This degree of accuracy is better than that reported for presently available controllers. The detector system consists of four cadmium telluride detectors and involves the use of a control algorithm to read the detectors and turn off the mammography unit at the correct time. This algorithm is implemented by a microprocessor, which also provides the means for a convenient calibration.
PhD
if one
a design
proved
of the
detectors
tenuating
smaller Abbreviation:
region,
to work
was
as could
well
even
outside
the
happen
for
at-
breasts. AEC
=
automatic
exposure
393
Figure
1.
detector lunide
Detector configuration
housing showing for four cadmium tel-
array
photovoltaic
detectors.
The
center-to-
center spacing of detectors A, B, and C is 2 cm. The spacing between C and D is 4 cm. These values were selected as appropriate based on the examination of 12 randomly chosen mammograms.
a.
b. 3.0
2.5 2.0
0
4
8
12
Lucite
16
Lucite
Figure
Signal current versus Lucite layer thickness for two photovoltaic cadmium telluride devices at increasing tube voltages. The detectors were similar in size and con2.
struction
and
varied
only
in
the
To ensure
satisfactory
the
in-
size, it is necessary
to
selected;
for a large
triplet A, C, and D is used. The is whether detector D is exposed direct x rays.
Algorithm
with
software
was developed
data
control
and
the
to analyze time.
attenuation of a monoenergetic beam is given by the well-known
where unit
N
area
absorbed N,
i
is the
per
time
nor scattered,
is the
linear
of
that N0
material, through
For polyenergetic
x-ray
394
emitted
Radiology
#{149}
and x is the the material.
beams,
by radiography
energy.
How-
radiation without bias), have a nearly
on the detector, photocurrent
follows
I
such machines,
as
=
an
the “2’
I0e,
,‘
where
I is the
is a function
photogenerated of both
detector
and
response,
current, the
and
do
a
cadmium
x is the
I
used:
Ortho-M
where
-a,
which
is seen
to be a
of the kilovolt peak. To develop an algorithm for controlling film exposure, it was also necessary to determine the film and screen response as a function of Lucite thickness.
for
different
opti-
Two mammographic
different
film
(Eastman
speeds
were
Kodak,
Roches-
ter, NY) and SP15O (Du Pont, Wilmington, Del). Each film type was used with a Min-R screen and cassette (Eastman Kodak). Exposures were made with a Mammomat-B mammographic unit (Siemens Medical Systems, Iselin, NJ). By means of a 15-step Lucite phantom, each exposure provided 16 data points. A plot of net optica! film density versus Lucite layers for both films are shown in Figure 3 for a family of exposures at two kilovolt-peak settings. Figure 4 was then generated with the values obtained from the data in Figure 3
of each
line is
of data
phantom. with
by ty.
thickness.
sets
films
Figure 2 shows the detector response as a function of attenuation by Lucite (Du Pont Diagnostic Imaging Division, Billerica, Mass). The slope
attenuator
this,
wedge
the cad-
function
coefficient
of the attenuating propagation length those
per
are neither is the incident
To
cal film densities were obtained by exposing the film and screen at various exposures between 5 and 320 mAs with a step
of x rays incident
The
(1)
attenuation
vary
the
x-ray equa-
photons
not
function
a range a matea beam
mium telluride equation
N0CCX,
number
unit
does
linear response to x-ray flux over of six orders of magnitude. When rial with a low Z value attenuates
telluride =
with
mode
tion N
varies
(ie, detecting externally applied
breast,
research,
exposure
flux
exponential
provide a good approximation (3). Cadmium telluride detectors, when optimized and operated in the photovoltaic
criterion to the
the detector
photon
as a simple
ever, for materials with low atomic numbers (Z) such as tissue, Equation (1) does
Development
In parallel
transmitted
exactly
of x because
the spacing between the three detectors, depending on whether the attenuation is by a large or small breast. This variance was achieved with four detectors spaced as shown in Figure 1 . For a small breast, the detector triplet A, B, and C is vary
automatically
d.
Figure 3. Optical film densities of two mammographic films for each region of the step wedge phantom obtained at varying milliampere-second settings: (a) Kodak film at 28 kVp, (b) KOdak film at 35 kVp, (c) Du Pont film at 28 kVp, (d) Du Pont film at 35 kVp. The numbers adjacent to each curve are the milliampere-second values used. These curves are similar to a “reversed” Hurter and Dniffield curve, plotted as a function of phantom thickness rather than as a function of x-ray intensity.
preparation
performance
of breast
Layers
Lucite
Layers
c
process of the semiconductor surface. Detector 3 (Det #3) was used in all of the subsequent measurements.
dependent
8
8
Layers
means From
of a constant Figure 4 we
It the
=
J is the x-ray exposure time.
characteristic calibration (3) can
combined
both
of the constants. be solved
optical film densiobtain the relation
Jot tube current is a constant
film,
and
and t is that is
to get Jt as a function
to are
Jo and
Equations for e’ and
(2)
and
then of I:
February
1991
lows 2BkVp
35kVp
A
on
KODAK 4
DUPONT
0’
. 0
16
Lucite
.
4
a.
.
8
Lucite
Layers
12
16
(3).
4.
Data
obtained from J is the x-ray voltage 28 kVp;
exposing film with a density of 1.0 through Lucite layers of tube current and t is the exposure time as given in Equation (b) tube voltage 35 kVp.
thickness.
(a) Tube
kilovolt-peak
mammography
selection
unit.
The
sor.
In both calibration modes, time is determined with the settings on the mammography I 750
U
1 500
1 500
any E
1 250
.
1 000
number
I 250
thickness
I 000
until
of times,
and
the
0 750
0 500 2
2
4
Lucite
Thickness
3
4
Lucite
[cmj
5
6
Thickness
7
[cm]
Figure 5. Film telluride-based points
are
for
density
results
AEC
to control
single
obtained
at 25 kVp
exposure
times
(a) and
30 kVp
for the Lorad
(b)
by using
mammography
the
Data
measurements.
(4)
gle constant
be combined
c, and t
into
a sin-
we have (5)
=
where K(J) = JoCJoto/J is a constant for a given current J. Thus the control circuit must determine the correct exposure time and turn off the mammography system
by means
of Equation
and c need calibration
(5). The constants
to be determined measurements.
by means
phy
machine.
uation.
Once
The
circuit
then
selection
calculation
data
of time
by an Intel 8052AH-BASIC
Design
sor that
basic
has a built-in
the
pulse accordof the selected
time t described
pro-
only
are saved and logged by the
once
exceptions
are
interpreter
values
off, calibrate A (thin Lucite), calibrate B (thick Lucite), and operate. Once calibrated, the AEC operates without the need for any interven-
two AEC electronic
compo-
nents, and (d) the radiography machine. The principle behind the design of the electronics is as follows. The signal current from each detector is the input to a current-to-voltage
en circuit. is used
converter
The
output
to select
the
and
voltage
amplifi-
of this
appropriate
stage
triplet at the beginning of the exposure based on the relative amplitude of the voltage for detector D (see Detector Design above). The three voltage outputs of the selected triplet are then sampled and digitized by means of a multiplex analogto-digital converter. Each digitized signal is then
Volume
used
178
to calculate
Number
#{149}
the
2
optimal
expo-
and
a real-
Operation
tion
and
a relatively
thus
can
be
one panel switch:
successfully
inexperienced
control.
used
by
radiography
technician.
Calibrate
detector
functions
The AEC has only This is a four-position
AEC
panel
the
are
operation
of the
requires that the two constants K and c in Equation (5) be determined for each detector and each kilovolt peak. To achieve this, a pair of flat sheets of Lucite for each kilovolt-peak setting are used as attenuation standards. The AEC detersystem
mines and stores The measurement
calibra-
possible
techthe noaging
retained
in the
in the
event
operate
on the
switch
that
would
cause
setting
also
on
position.
The
mammography
are set to a milliampere-second This
sys-
of a power
control
to the
controls
a slight serves
unit
setting overexposure.
as a fail-safe
timer
is mithe AEC
terminates the exposure. The optical film density of the mammogram at the location of the controlling cadmium telluride
detector mined
should be the density during the calibration
ticular
kilovolt
peak
as deterfor that par-
to within
the required is accomplished
constants. as fol-
AND
the
accura-
CONCLUSIONS
The AEC system LR-1 mammography
was
tested with machine
(Lomad Medical Systems, Danbury, Conn). With this unit, exposure times of 0.2-5 seconds and x-ray tube anode values of 22-30 kVp can be selected
from
the
R intensifying
control
panel.
screen
and
A Mm-
Ortho-M
film were used (Eastman Kodak). system was calibrated at 25 and
kVp mode-The
is set
RESULTS
tween
these
the
the
AEC
Lucite
the x-ray operator;
microproces-
BASIC
with time
math
if the
cy of the AEC.
t is done
components of the complete system are (a) the cassette that contains the x-ray film, the intensifying screen, and the four cadmium telluride photovoltaic detectors, (b) the AEC electronics (analog and digital), (c) the interface be-
intrinsic clock.
is achieved.
procedure
override for the AEC. An exposure tiated in the normal manner, and
selects
is made,
the turn-off
ing to the calculated detector. The above and
of this voltage the mammogra-
of the exposure time, the middle value atten-
the
AEC generates
cessing
System
A logic
the middle value which represents K
of
sure time t, according to Equation (5), for the attenuation at each detector location. Because the correct exposure time is a function of the x-ray tube voltage, an analog voltage representative is input to the AEC from
density
tion sheets niques are
calibration
Jt = can
fl/a
Lucite
setting
trial-and-error
tern memory even failure. Operate mode-The Finally,
the
and replacement of the x-ray tube. At most, this procedure would need to be carried out only two times per year. The
cadmium
system.
film
this
is required
table
b.
changing
milliampere-second
desired
However, 0 750
the exposure panel control unit itself,
not with the AEC. If the proper film exposure is not achieved during either calibration step, those steps can be repeated
3OkVp
Vp
255 1 750
correct
exposure time (or milliampere second) is selected for a thin Lucite calibration phantom. The AEC control knob is switched to calibrate A, and an exposure is made; the AEC measures current I from the detectors and the length of time t that the mammography unit is running an exposure. The procedure is repeated with the AEC switched to calibrate B and with the thick Lucite calibration phantom in the beam. From the values obtained, the unknowns K and c from Equation (5) are calculated and stored by the microproces-
Layers
b.
Figure varying
for a given the
with
target
film
densities
The 30
set at
1.2.
The
design
allowed
a simple
wire interface between the Lorad system. The sign required that one from the Lomad system
dicate in use,
which one
anode lead
kilovolt
to the
three-
the AEC and interface deelectrical lead be used to in-
peak
Lonad
Radiology
is
system
395
#{149}
transmits ground
The
the stop pulse, and a lead connects the two units.
AEC
from the tom itself
the
received Lonad directly
its start system sensing
x-ray beam. Tests to evaluate
ty of the
AEC
the
and
the
signal the detecthe onset
of
reproducibili-
performance
ned out. First, several were carried out with attenuator,
by
ment controlled exposure as a function of attenuator thickness, we made exposures over a wide range of Lucite thickness and at two x-ray tube
were
exposures a given signal
from
car-
a sin-
gle detector was used to terminate the exposure. Variations in exposures that were terminated by signals from each of the four detectors were also evaluated. The standard deviation in resulting film density from run to run (for a given detector) or from detectom to detector was less than ±0.03. To determine how well the instnu-
Radiology
#{149}
quality
the
need
and
1.
1.3-4.5
cm
at 25 kVp,
exposures
densities for which
within 0.05 the system
resulting
the
AEC
These results cleanly demonstrate that the cadmium tellumide detectors combined with a suitable algorithm developed for determining the connect exposure time overcome the dif-
without
manual
adjust-
U
References
in film
of the density was calibrated.
mammograms
for “expert”
ment.
Iklason
LT,
Barnes
GT,
Rubin
E.
raphy phototimer technique gy 1985; 157:539-540. 2.
LaFrance
R, Gelskey
cuit modification graphic phototimer gy 1988; 3.
Schleien
Health diological Nucleon
DE, that
Mammog-
chart. Barnes
RadioloGT.
improves performance.
A cir-
mammoRadiolo-
166:773-776. B, Terpilak
Physics.
MS.
In: The
eds.
health
Medical
physics
health
handbook.
Olney,
Lectern
Associates,
1984;
raMd:
chap
10.
ficulties encountered with other methods of exposure control. The resulting film density is close to the set density and is independent of attenuatom thickness over the range of in-
terest
396
congood-
voltages. The results are shown in Figure 5. The data indicate that, over the manges of 2.25-6.75 cm at 30 kVp produced
Lucite
tion in the clinical setting, this trollem should reliably provide
for mammography.
In opera-
February
1991