Vol. 10, No. 3

JOURNAL OF CLINICAL MICROBIOLOGY, Sept. 1979, p. 346-350 0095-1137/79/09-0346/05$02.00/0

Application of Fluoroimmunoassay to Cerebrospinal Fluid Immunoglobulin G and Albumin ROY W. STEVENS,* DANIEL ELMENDORF, MICHAELYN GOURLAY, ERNST STROEBEL, AND HASSAN A. GAAFAR Division of Laboratories and Research, New York State Department of Health, Albany, New York 12201 Received for publication 22 June 1979

Special solid-phase fluoroimmunoassay protocols were used to measure the amount of immunoglobulin G (IgG) and albumin in 1,511 samples of cerebrospinal fluid. The fluoroimmunoassay is an inhibition test conducted on the surface of a plastic probe, whose fluorescence is inversely related to the concentration of spinal fluid IgG or albumin. A microcomputer interfaced with the fluorometer calculates the sample IgG or albumin from a calibration curve based on standard cerebrospinal fluid values. The test and interpretation take less than 90 miin. Correlation coefficients for over 100 cerebrospinal fluid samples tested by both fluoroimmunoassay and radial immunodiffusion were: IgG, 0.90 (slope, 1.04); albumin, 0.95 (slope, 0.99). The within-run precision (coefficient of variation) was: IgG, 4.4%; albumin, 6.3%. Run-to-run precision on a midrange sample was: IgG, 7.7%, albumin, 11.7%. These findings establish the simplicity, speed, and precision of the modified fluoroimmunoassay system for specific cerebrospinal fluid proteins. Abnormal concentrations of specific proteins, especially immunoglobulin G (IgG), in cerebrospinal fluid (CSF) are important indicators in the differential diagnosis and treatment of a number of neurological disorders (11). Chief among these are encephalomelitis, neurosyphilis, subacute sclerosing panencephalitis, multiple sclerosis, and systemic lupus erythematosis affecting the central nervous system. These abnormal concentrations of specific CSF proteins can occur in fluids with normal total protein. Three immunological procedures, diffusion (8, 12), electrophoresis (9, 18), and nephelometry (5, 14-16; R. W. Stevens, H. Lamonda, N. Trornbley, and H. A. Gaafar, Abstr. Annu. Meet. Am. Soc. Microbiol. 1973, M34, p. 79), are commonly used in clinical laboratory determinations of CSF IgG and albumin. Tables of normal values have been compiled, and abnormal IgG and albumin concentrations and IgG/albumin ratios in neurological disorders have been described (3, 6, 10, 13). At our suggestion special protocols were developed by the manufacturer to adapt the existing (4) solid-phase serum-protein fluoroimmunoassay (FIA) (FIAX system, International Diagnostic Technology, Santa Clara, Calif.) for analysis of IgG and albumin in CSF. To accommodate the ultra-low-level IgG and albumin in CSF, the fluorescent-antibody and specimen volumes and ratios of the FIAX serum protein

system were adjusted, and unique microcomputer programs were used. We have investigated the precision of this modified system and its agreement with the

radial immunodiffusion (RID) techniques, using reagents supplied by the manufacturer for investigational use. (This report was presented in part at the American Society for Microbiology Second Conference on Diagnostic Immunology, June 1978, Atlanta, Ga.). MATERIALS AND METHODS FIA. The FIAX equipment includes a horizontal shaker (200 oscillations per min, 2.5-cm stroke) with a tube holder, a filter fluorometer, and a microcomputer (optional). The special protocols for determinations of IgG and albumin in CSF are identical, so only the procedure for IgG is described here. For the test, 5 ,ul of CSF in 250 pl of phosphatebuffered saline, pH 7.4, is mixed in a 12- by 75-mm tube with 5 ul of fluorescein isothiocyanate-labeled, monospecific, goat anti-human-IgG serum. The mixture is incubated on the horizontal shaker for two 10min periods. During the first period a plastic probe with a shallow well surface, which has a high capacity to adsorb protein, is shaken in another tube containing 500 pl of human IgG. During the second period this probe is shaken in a rinse tube containing 500 Al of phosphate-buffered saline. The probe is then inserted in the CSF-antiserum mixture, which is shaken for another 20 min. A final 10-min phosphate-buffered

450 angle

346

SPINAL FLUID IgG AND ALBUMIN ASSAY

VOL. 10, 1979

saline rinse removes all CSF and antiserum proteins not coupled to the probe. To measure the amount of antiserum complexed to IgG on the probe, the probe is inserted onto the fluorometer stage. Its fluorescence is converted to digital signal units, which are displayed on the instrument panel and simultaneously transmitted to the microcomputer. Probe fluorescence is inversely related to the IgG concentration in the CSF. Four fluids of known protein concentration are used to establish the calibration curves. The calibrator fluids used in this study contained 0.6, 1.5, 3.0, and 9.2 mg of IgG per dl (or 3.6, 10.2, 17.5, and 52.5 mg of albumin per dl). A calibration curve plotted from their fluorescence signals (Fig. 1) is used to determine IgG concentrations from the specimen signals. The microcomputer is used to construct the best-fit curve, compare the curve with an ideal curve for the CSF program, and report any deviation. If the calibration curve is acceptable, the specimen concentrations are computed and printed out in milligrams per deciliter. Specimens with fluorescence signals above the upper (or below the lower) reference value are reported as too high (or too low) for computation. RID. Agar gel plates (Immunoplate III, Hyland Laboratories, Costa Mesa, Calif.) containing antiserum specific for human IgG (or albumin) and precut sample wells were used exactly as prescribed by the manufacturer for limit diffusion ("precision") measurements. The immunodiffusion standards contained (per deciliter) 1.9, 4.1, and 7.9 mg of IgG or 12.3, 25.2, and 50.0 mg of albumin. After 48 h of incubation at 37°C the diameters of the RID precipitation zones formed by the standards and the CSF specimens were measured to the nearest 0.1 mm. The standard measurements were squared and plotted on linear graph paper against the known concentrations. Specimen concentrations were then determined from the calibration curve. Total protein assay. Total protein was determined by applying Folin reagent to a trichloroacetic acid precipitate of the CSF specimen (17). This test is relatively insensitive to the CSF albumin/globulin ra180 Cf)

Z~ 0150

\

z 0

120

z

CoC) 90 0 -J

IMMUNOGLOBULIN G (mg/dl)

FIG. 1. IgG FIA calibration curve.

347

tio and e,liminates interference by nonprotein substances. Specimens. The 1,511 samples studied were from lumbar CSF specimens submitted to our reference service for diagnostic tests over 6 months. A total of 1,840 specimens were accessioned, but 329 showed evidence of plasma or microbial contamination or were of insufficient volume for all laboratory tests. These were excluded from the study. To determine the correlation of FIA results with those for the established RID assay, 108 fluids were tested undiluted by each procedure in lots of 54 samples. After all 108 had been examined, those with IgG (or albumin) concentrations greater than the maximum reference value were diluted 1:4 in normal saline and reexamined. To determine normal IgG (and albumin) FIA values, we selected 50 specimens with normal results in all other laboratory tests (fluids, clear and colorless; qualitative tests for hemoglobin, negative; cardiolipin complement fixation tests for syphilis, nonreactive; colloidal gold curves, normal type A; total protein values, 15 to 45 mg/dl). These specimens were from an equal number of male and female individuals 18 to 60 years of age; no diagnosis or other information on disease was given. Accuracy and precision. For FIA within-run precision measurements, 19 samples from one pool were tested in a single test run. For run-to-run precision, one pool of fluids was examined on each of 25 test days. In addition, World Health Organization IgG International Reference Preparation 67/95 was rehydrated with 1 ml of distilled water and diluted 1:200 in saline to contain 3.83 mg of IgG per dl (7). This dilution was tested in quadruplicate by FIA and RID.

RESULTS Comparison of FIA and RID. Agreement between the FIA and RID results was excellent after samples outside the calibration range were excluded (Table 1). The specific protein concentrations of nearly all 108 specimens compared fell within the FIA calibration range of 0.6 to 9.2 mg/dl for IgG (3.6 to 52.5 mg/dl for albumin). No sample had less IgG (or albumin) than the miniimum FIA calibrator; two contained extraordinarily high albumin concentrations. With RID, however, the relatively high minimum calibrator values (1.9 mg/dl for IgG, 12.3 mg/dl for albumin) precluded measurement of one or both of these proteins in 12 specimens. One contained too little of either IgG or albumin to be measured, six contained too little IgG only, and five others contained too little albumin. The maximum RID calibrator value of 7.9 mg/dl for IgG was exceeded in five specimens; the maximum of 50.0 mg/dl for albumin was exceeded in four of the same specimens. Accuracy and precision of FIA. Withinrun precision for the FIA system for IgG was

348

J. CLIN. MICROBIOL.

STEVENS ET AL.

In the group of 564 specimens with total probetter than for RID, and run-to-run precision how- tein values outside the normal range, only 4 had 45 mg/dl, the values of 3.56, 3.60, 3.87, and 3.90 mg/dl). The IgG levels of 52 and the albumin levels of 73 RID mean for the same dilution was 4.45 mg/dl, were greater than the maximum calibrator valand the coefficient of variation was 7.9% (for ues. Virtually all of these off-scale high samples were fluids with total protein concentrations >65 values of 4.1, 4.1, 4.8, and 4.8 mg/dl). Normal values. Normal CSF values (in mil- mg/dl (IgG, 41 of 52 or 90%; albumin, 72 of 73 or ligrams per deciliter) for IgG and albumin, found 99%). by FIA in 50 normal samples, were (mean ± DISCUSSION standard deviation): IgG, 2.33 ± 0.53 (range, ± 4.30 10.6(range, 1.14-3.90); albumin, 19.46 Our results with the special FIAX protocols 31.3). IgG as a percentage of total protein was for CSF determinations showed excellent agree7.42 ± 1.88 (range, 3.9-12.8). ment with findings obtained after 48 h with the Analysis of specimens. Of the 1,511 fluids limit-diffusion RID procedure. A recent study of tested, 947 had normal total protein concentra- the standard FLAX test for serum immunoglobtions (15 to 45 mg/dl). Most of these (81.6%) ulins showed results for IgG in close agreement contained IgG, as a percentage of total protein, with those found by timed-diffusion (16-h incuwithin 2 standard deviations of the mean normal bation) RID (4). value (Table 3). However, 167 (17.6%) were :2 The special FIA system for CSF analysis has standard deviations above normal, and 67 (7.1%) greater sensitivity and relatively longer calibrawere -4 standard deviations above normal. One tion curves than RID. The FIA calibration curve sample was >2 standard deviations below nor- range from 0.6 to 9.2 mg of IgG per dl (3.6 to mal. Six others contained too little IgG to meas- 52.5 mg of albumin per dl) accommodated all ure; three of these six also had too little albumin. 108 specimens in the comparison study. In the None of the 947 samples had IgG or albumin in tests of all 1,511 diagnostic specimens only 10 excess of the maximum reference values. contained too little IgG to be measured, and 6 of

was nearly as good (Table 2). For albumin, ever, the RID system was more precise.

TABLE 1. Comparison of FIA and RID for determination of IgG and albumin in CSF Determination

IgG FIA RID

se.imn s

menst

108 108

Albumin 108 FIA 108 RID aDiluted 1:4 and retested.

Values outside calibration range Too Too low high0

Results compared (no.)

Mean concn

3.75 4.18

0 7

1 5

101 101

0 6

2 4

102 102

(mg/dl)

24.4 24.2

Snepreg

Correlatson

lnterceptline

coefficient

0.274

1.039

0.903

-0.179

0.999

0.946

TABLE 2. Precision of FIA and RID for deterrnination of IgG and albumin in CSF Run to Runb

Within run'

Determination

Mean

(mg/dl)

Range Rage

SD

CV

Mean

SD(mg/dl)

Range Rag

S

SD

CV %

0.31 0.25

7.7 6.3

3.31 1.41

11.7 6.1

IgG 3.6-4.5 3.99 4.4 0.13 2.99 2.83-3.23 FIA 3.80-4.30 3.98 7.3 0.29 3.85-4.50 RID 4.02 Albumin 28.1 21.3-33.6 1.53 6.3 24.1 22.1-27.0 FIA 20.5-24.5 3.2 23.0 0.74 22.5-24.0 RID 23.4 a One sample, n = 19. SD, Standard deviation; CV, coefficient of variation. 'One pool for FIA, n = 25; one pool for RID, n = 8.

SPINAL FLUID IgG AND ALBUMIN ASSAY

VOL. 10, 1979

TABLE 3. IgG distribution in 947 fluids with normal total protein (15 to 45 mg/dl) % of total pro-

>26.2 22.5-26.2 18.8-22.4

Deviation from normal

(SD)a >10

Specimens No. 3 11

8-10 6-8 19 15.0-18.7 4-6 34 11.2-14.9 2-4 100 3.7-11.1 ±2 773 2 1 >2 Undetectable 6 aMean of 50 normal samples was 7.42% SD, Standard deviation.

% 0.3 1.2 2.0 3.6 10.6 81.6 0.1 0.6 (see text).

the 10 contained too little albumin. Thus, specific protein levels of only about 0.7% of the specimens could not be determined by FIA. An estimate of the protein values might have been obtained by testing a greater sample volume in the same reaction mixture. At the other extreme, 52 specimens contained too much IgG to be measured by FIA of an undiluted specimen; 73 contained too much albumin. All of these had total protein > 65 mg/ dl. Specific protein concentrations of such specimens might be determined if the sample is diluted 1:4 in saline. The IgG values in normal adult CSF established with the special FIA protocol are comparable to those reported with other immunological assays. Published means range from 0.9 to 5.0 mg/dl (5, 6, 8) or from 6.4 to 11.4% of total protein (14). The accuracy of our FIA values of 2.3 mg and 7.4% is supported by the number of samples studied, by the method used to select samples, and by the accuracy of the total protein assay. In infants and young children normal total protein and, thus, IgG concentrations are presumed to be lower (1). Total protein concentration alone is an inadequate measure of CSF composition, since a significant number (7.1%) of fluids with normal total protein contained a markedly high concentration of IgG. A smaller percentage (0.7%) of fluids had a markedly low concentration of IgG, which may occur in autoimmune disorders with central nervous involvement (10). Thus, IgG measurement is an important adjunct to total protein findings in differentiation of neurological disorders. The CSF FIA tests were completed very quickly: from materials setup to report printout for 40 specimens took about 90 min. The fluorescence of each sample probe was read, and the protein concentration printed in about 5 s. This efficiency of solid-phase immunoassay

349

was largely due to the separation of the specific antigen-antibody reactions (2). The dipstick probe was removed from the specimen reaction mixture and washed; neither reagent nor specimen blanks were needed. Specimen filtration or clarification was also unnecessary, since neither materials in suspension or solution (protein aggregates, drugs, contrast media, etc.) nor particulates (cells, tissue debris, etc.) would be carried to the fluorometer. Furthermore, the solid-phase assay was unaffected by an excess of antigen within the limits of the reference curve, since the immunofluorescence, unlike nephelometry, is independent of the character of the antigenantibody complexes formed. The test precision meets the special requirements of clinical laboratory determinations, and the results obtained with the World Health Organization reference standard indicate the accuracy of FIA for CSF specimens. In conclusion, the advantages of simplicity, speed, and precision of the FIA system recommend its use for determinations of CSF IgG and albumin. ACKNOWLEDGMENT Gerald I. Kaufman kindly provided statistical analysis for correlation of FIA and RID results.

LITERATURE CITED 1. Abramowicz, M. 1969. Normal values for cerebrospinal fluid protein concentration in children. Clin. Fed. 8: 300-301. 2. Baier, R. E. 1975. Applied chemistry at protein interfaces. Adv. Chem. Ser. 145. American Chemical Society, Washington, D.C. 3. Berner, J. J., V. A. Ciemins, and E. F. Schroeder. 1972. Radial immunodiffusion of spinal fluid: diagnostic value in multiple sclerosis. Am. J. Clin. Pathol. 58:145154. 4. Blanchard, G. C., and R. Gardner. 1978. Two immunofluorescent methods compared with a radial immunodiffusion method of measurement of serum immunoglobulins. Clin. Chem. 24:808-814. 5. Deaton, C. D., K. W. Maxwell, R. S. Smith, and R. L Creveling. Use of laser nephelometry in the measurement of serum proteins. Clin. Chem. 22:1465-1471. 6. Hartley, T. F., D. A. Merrill, and Ho N. Claman. 1966. Quantitation of immunoglobulins in crebrospinal fluid. Arch. Neurol. 15:472479. 7. Humphrey, J. H., and I. Batty. 1974. International Reference Preparation for human serum IgG, IgA, IgM. J. Immunol. 113:428. 8. Laffin, RK J. 1970. A simple immunologic method for the quantitation of immunoglobulin G in human cerebrospinal fluid. J. Lab. Clin. Med. 76:816-823. 9. Laurell, C. 1966. Quantitative estimation of proteins by electrophoresis in agarose gel containing antibodies. Anal. Biochem. 15:45-52. 10. Levin, A. S., H. H. Fudenberg, L. D. Petz, and G. C. Sharp. 1972. IgG levels in cerebrospinal fluid of patients with central nervous system manifestations of systemic lupus erythematosus. Clin. Immunol. Immunopathol. 1: 1-5.

350

STEVENS ET AL.

11. Lumsden, C. E. 1972. The proteins of cerebrospinal fluid-the importance of methods, p. 333-367. In D. McAlpine, C. E. Lumsden, and E. D. Acheson (ed.), Multiple sclerosis; a reappraisal, 2nd ed. The Williams & Wilkins Co., Baltimore. 12. Mancini, G. A., A. 0. Carbonara, and J. F. Heremans. 1965. Immunochemical quantitation of antigens by single radial immunodiffusion. Immunochemistry 2:235254. 13. Riddoch, D., and R. A. Thompson. 1970. Immunoglobulin levels in the cerebrospinal fluid. Br. Med. J. 1:396399. 14. Ritchie, R. F., C. A. Alper, J. Graves, N. Pearson, and C. Larson. 1973. Automated quantitation of proteins in serum and other biological fluids. Am. J. Clin. Pathol. 59:151-159. 15. Savory, J., M. G. Heintges, L. M. Killingsworth, and

J. CLIN. MICROBIOL. J. AL Potter. 1972. Manual and automated determination of inmmunoglobulin in unconcentrated cerebrospinal fluid. Clin. Chem. 18:37-42. 16. Sternberg, J. C. 1977. A rate nephelometer for measuring specific proteins by immunoprecipitin reactions. Clin. Chem. 23:1456-1464. 17. Stevens, R. W., J. T. Utter, and E. Stroebel. 1966. Folin phenol reagent for determination of cerebrospinal fluid total protein, p. 125. In Annual report of the Division of Laboratories and Research. New York State Department of Health, Albany. 18. Tourtellotte, W. W., B. Travolato, J. A. Parker, and P. Comiso. 1971. Cerebrospinal fluid electroimnmunodiffusion. An easy, rapid, sensitive, reliable, and valid method for the simultaneous determination of immunoglobulin G and albumin. Arch. Neurol. 25:345-350.

Application of fluoroimmunoassay to cerebrospinal fluid immunoglobulin G and albumin.

Vol. 10, No. 3 JOURNAL OF CLINICAL MICROBIOLOGY, Sept. 1979, p. 346-350 0095-1137/79/09-0346/05$02.00/0 Application of Fluoroimmunoassay to Cerebros...
675KB Sizes 0 Downloads 0 Views