Lone
R. Pelc,
Gary
H. Gloyer,
PhD #{149} Norbert PhD #{149} Robert
Arterial Noninvasive
J. Pelc, ScD J Herfkens,
Venous
and
Index
terms:
Blood,
961.91,
966.91,
flow dynamics, 984.91 #{149} Blood,
CCURATE
I
From
assessment
904.91,
MR.
the
=
185:809-812
Departments
of Radiology
gery (5CR., L.J.C., D.C.M.), Richard aging, Stanford University School of April 2, 1992; revision requested June part by grants to the Department of Phil N. Allen Trust, and GE Medical ‘ RSNA, 1992
(L.R.P.,
of blood
flow
the major vessels is an essential aspect of cardiovascular evaluation. Current noninvasive techniques have limitations such as the use of ionizing radiation, contrast agents, or both or limited vascular access. Many techniques have been reported for measurement of blood flow by means of magnetic resonance (MR) imaging (1-5). However, the in vivo accuracy of these MR measurements has not been firmly established by comparison with simultaneous measurement of volume flow performed with a technique proved to be accurate. Phase-contrast, gradient-echo MR imaging methods are well suited for measurement of blood flow because of their speed, sensitivity to blood flow across a broad range of flow rates, and inherent quantitative foundation. To measure blood flow in one direction, two complete sets of raw MR data are obtained with pulse sequences that have a different gradient first moment in the direction of interest. For blood flow that is uniform throughout a voxel and constant throughout the image, the change in the phase of the transverse magnetization in a voxel is proportional to the flow velocity in the encoded direction. Phase-contrast data can be used to produce an image in which the pixel intensity depends on the magnitude of the transverse magnetization in each voxel (magnitude images) and a velocity image proportional to the phase shift. The measured phase shift 4 is converted to velocity v by the following equation: V
1992;
Flow: with MR
through
904.1299, 957.1299, 961.1299, 966.1299, 984.1299 #{149} Blood, US, 904.12989, 957.12989, 961.12989, 966.12989, 984.12989 #{149}Magnetic resonance (MR), phase imaging
Radiology
Blood
Quantltatlon
Quantitative measurements of arterial and venous blood flow were obtamed with phase-contrast cine magnetic resonance (MR) imaging and compared with such measurements obtained by means of implanted ultrasound (US) blood flow probes in anesthetized dogs. The US flowmeter was enabled during a portion of each MR imaging sequence to allow virtually simultaneous data acquisition with the two techniques. MR imaging data were gated by means of electrocardiography and divided into 16 phases per cardiac cycle. The rates of portal venous blood flow measured with MR imaging and averaged across the cardiac cycle (710 mL/ mm ± 230 [standard deviation]) correlated well with those measured with the flowmeter and averaged in like fashion (751 mL/min ± 238) (r = .995, slope 1.053). The correspondence in arterial blood flow was almost as good. No statistically significant difference existed between the paired measurements of blood flow obtained with MR imaging and the implanted probe. It is concluded that, as a noninvasive means of accurate quantification of blood flow, phase-contrast MR imaging may be especially useful in deep blood yessels in humans.
957.91,
J.
Stephen C. Rayhill, MD #{149}Luis Castro, MD MD #{149}D. Craig Miller, MD #{149}R. Brooke Jeffrey,
#{149}
N.J.?.,
G.H.G.,
R.J.H.,
R.B.J.)
-yz.M1’
and
Cardiothoracic
(1)
MD
Imaging’ where ment
Mu1 is the change in the two sequences
in first and
moy is
gyromagnetic ratio. The flow-encoding strength is sometimes parameterized by Venc, the velocity that produces a phase shift of 180#{176}. For quantitative flow measurements, flow encoding is generally performed in the section-select direction, a region of interest (ROI) that includes the vessel of interest is defined, and the product of the average velocity within the vessel and the vessel area the
yields
the
flow
rate
in milliliters
per
minute. To measure pulsatile blood flow, this technique is combined with cardiac gating (3) or cine imaging (5). Velocity images have noise that depends on the strength of the flow encoding and on the signal, S. and noise, if, in the magnitude image (5). The standard deviation (SD) in the velocity, (Tv, is calculated as follows: Venc
Thus,
noise
0
(2)
ffv=
ITS
in flow
measurements
is
minimized whenever velocity encoding (Venc) 5 as small as possible. However, because phase shifts can be uniquely measured only over a range of 360#{176}, it is best to ensure that no yelocity of interest exceeds Venc (flow aliasing). Because velocity images are produced by the difference in acquisitions
that
quantitative insensitive geneities
are
almost
identical,
characteristic to magnetic and
many
other
their
is relatively field inhomoimperfec-
tions in the system. However, differential eddy current effects secondary to the gradient changes that produce flow encoding can cause undesired
Sur-
M. Lucas Center for Magnetic Resonance Spectroscopy and ImMedicine, 300 Pasteur Dr. Stanford, CA 94305-5488. Received 3; revision received July 13; accepted July 27. Supported in Radiology from the Society of Computed Body Tomography, Systems. Address reprint requests to L.R.P.
Abbreviations:
RF = radio frequency, ROI of interest, SD = standard deviation, = standard error of the mean, TE = echo time, TR = repetition time, v,. = velocity encoding.
=
region SEE
809
phase
shifts.
These
effects
appear
with
as
additive velocity errors (ie, static structures appear to have nonzero velocity) that vary only slowly across the image. Accurate measurements of blood flow must account for the possible presence of such errors. Phase-contrast measurements of blood flow have been shown to have acceptably
low
interand intraob(less than 5% and
server variability 10% of the measured blood flow, respectively [6]). The accuracy of the method has been repeatedly validated in vitro (4,6,7). Some in vivo validation has also been reported. Generally
good
agreement
has
has
Doppler
ultrasound
(US)
are
complicated
comparisons
also
been
blood flow. We studied vessels: six portal veins, two renal veins, two renal arteries, one carotid artery, and three common iliac ar-
teries.
Two to three centimeters
allowed
to recover
spontaneous
sufficiently
to
respiration.
The
Hoffman-La by Roche
Roche,
Rico) (0.25 mg/kg)
hydrochloride
(2.5-5.0
The
US flowmeter
screened
room,
(approximately
and ketamine
was
inside
2 m)
of cable
flecting surement
brackets within
minimum
were the
production
phase-
of flow rate, whereas Dopis not. Thus, the purpose of study was to validate measure-
Sunnyvale,
MR sequence
the sampling
of the
our
synchronized,
of blood flow obtained in vivo with phase-contrast cine MR imaging by means of comparison with simultaneous measurements obtained with an implanted blood flow probe.
ter samples
dogs
received
an
Because
uniformly
METHODS
was
injection
of sodium
the dogs were was maintained 1%-2%
halothane.
(model
T1OID;
NY)
were
sure
perivascular
of blood 810
One
Transonic
implanted
blood
or two
US
probes
Systems, flow.
invalid
because
of the
of blanking,
throughout
samples
were
missed
the cycle.
dog
The
Ithaca, to mea-
accuracy
obtained
half
the MR
of the last three
car-
periods was generated (9) and stored in the computer (Fig 2). All measurements obtained during flowmeter blanking were
excluded. Flowmeter measurements obtamed during all cardiac cycles throughout the MR acquisition
16 or 32 bins equally
were
spaced
grouped
into
in the cycle.
cardiac
The
trigger
read
(Flowmeter
just before until portion of each
just after sequence.
axis
nulled.
the radiofrequency
(dotted
(Gr,d)
was
The US flowmeBlanking) from the a
=
data flip
acquisition angle of
(RF) pulse.
Rou
__J1_iI_fl_JLJLfl_RJL.fl_.fl_Jl_JLJl.JLJJ_JLJI_I1._J1
Binaking
___
-
“*S’
SR
Figure 2. Schematic representation of cardiac-phase signal generation. Time elapsed since the beginning of the cycle (r) divided by the running average of three preceding RR intervals (running average RR) was en-
tered blood
into a file and flow
used
to partition
measurements
flowmeter
sample
obtained
when
the
was enabled.
The bin to which a sample was assigned was selected with use of the cardiac-phase signal. The average flowmeter measure-
ment
for all acceptable
bin and
the average
ment
during
puted.
Thus,
meter
with
with
each
waves).
first-order-moment ter was blanked
amination.
For comparisons
with
sine
imaging samples
diac of
with different section-select direction gradient waveforms (Gci,ct., dotted and solid lines) were rapidly interleaved throughout the acquisition. Phase encodings (Gptase) were in-
Phase-contrast cine MR produced 16 frames, equally spaced throughout the cardiac cycle, that comprise a composite cardiac cycle averaged throughout the ex-
cycle to the average
and anesthesia of inhalation
in each
flow measurements
Radiology
#{149}
pentothal
intubated, by means
were
(20 mg/kg),
(0.4 mg/kg).
with
blood
of the MR pulse sequence (and flowmeter blanking) were not some of the stored flowme-
intravenous
injec-
maleate
induced
and disabled shortly before acquisition por(Fig 1). Since flow probe and
data, the flowmeter samples were similarly grouped into “frames.” For this purpose, a “cardiac-phase” signal equal to the ratio of the time since the beginning of the
intramuscular
tion of acepromazine Anesthesia
(RF) intercircuit was
external blanking. To enable identification of these samples, the blanking signal was also digitized and stored in the computer.
Arterial and venous blood flow was measured in eight anesthetized mongrel dogs of either sex that weighed 20-30 kg. All
Calif).
tion of every execution therefore
ments
mea-
fed through the screen-room penetration panel to a personal computer (Macintosh Ilci; Apple Computer, Cupertino, Calif) with data acquisition hardware and software (Workbench; Strawberry Tree,
added to the US flowmeter flowmeter operation from until shortly after the data
imagmea-
to enable
was used to monitor the dog and provide synchronization for the acquisition of MR images. The flowmeter signal was sampled every 20 msec and
surements pler US
AND
the
magnetic field with of artifact. A lead II
To prevent radio-frequency ference, an external blanking
MATERIALS
7 ft
between
used
Blanking
Figure 1. Pulse sequence used for collection of phase-contrast cine MR data for measurement of blood flow. Two sequence types
cremented
the
approximately
many
contrast MR velocity measurements can be expected to be consistently lower than Doppler measurements. Conversely, phase-contrast MR ing is well suited for volumetric
NJ:
Manati,
with
iiii2Hc”:
G Rowmu
flow probe and the meter. Specially designed blood flow probes with brass re-
and While
Thus,
0
mg/kg).
electrocardiogram
periods.
Nutley,
Products,
some MR techniques measure the distribution of velocities within a voxel and can therefore estimate peak yelocity, the most commonly used method yields a value that represents the average velocity within a voxel. Furthermore, all MR measurements represent either averages over the entire imaging time or at least over cardiac
of each
acoustic gel (HR Lubricating Jelly; CarterWallace, New York) was used to fill any air space within the probe bracket. Flowprobe cables were secured and exteriorized, the incision was closed, and the dog
Puerto
the
G
vessel was isolated, a probe of appropriate size was selected and implanted, and
(Valium;
(7). These
peak velocity resolution.
10% of actual
a total of 14 blood
distributed
fact that phase-contrast MR imaging and Doppler US measure different quantities. Doppler US is generally used to measure has high temporal
by the plus or mi-
dog was then transported to the MR imaging facility, and for the duration of the imaging protocol it was sedated with intravenous bolus injections of diazepam
with by
probes is guaranteed to be better than
maintain
been
compared
nus
was
obtained between phase-contrast measurements of ventricular outflow and MR measurements based on planimetry (6,8). Phase-contrast MR imaging
these
manufacturer
the cardiac this
data enabled
in each measure-
cycle were
processing
of the
comparison
comflow-
with
data and also compensated lost during blanking.
All MR imaging a 1.5-T
Systems,
were
readings flowmeter
system
Milwaukee).
used
to select
data
were
(Signa;
obtained GE Medical
Localizer
a plane
MR
for
images
perpendicular
to the direction of blood flow in each yessel of interest, as close as possible to the position of the blood flow probe. In the immediate vicinity of the probe, RF and static field perturbations caused signal loss within the vessel. Typical imaging locations were within 5 mm of the probe. At these locations, the vessel lumen was relatively bright. Any residual phase shifts due to the presence of the probe were
December
1992
b.
a.
Figure
3.
MR images
of canine
256; section thickness, 5 mm; (dotted lines). (b) Magnitude tual data used for calculation tively, and gray as 0 cm/sec.
iliac artery
with
and field of view, 24 cm. image shows ROI (arrow) of blood flow. Gray scale
the following
1000
parameters:
30#{176}; two signals averaged; matrix, 256 x with ROIs used as mask for correction of additive phase errors of blood flow. (c) Magnitude-weighted velocity image contains accm/sec) as black or white in superior or inferior directions, respec-
(a) Magnitude image used for calculation portrays z’ (±100
Because
1200
B
c.
obtained
the
noise
in the
50/12;
measured
flip angle,
veloc-
(n
ity is high in regions with tensity (Eq [2J) and because outside the object do not
low signal inpixels in air exhibit phase-
shift were signal
that excluded weighted by to suppress
air
selected
of
1800
#{149} “noun 600
ani
0
errors, static regions selected. The fit was magnitude squared
noise
from
low
inadvertently
signal
then
intensity.
subtracted
The
from
all
the phase-contrast 200
400
600
Blood
Figure
4.
Mean
flow measured the flowmeter cates the best plus venous); dence intervals. coefficient).
800
flow(rnl/min)
arterial
iou
MIII
and
venous
blood
with MR imaging (x axis) and (y axis). The solid line mdilinear fit to all data (arterial the dashed lines, 95% confir2 = r (sample correlation
can
be sensitive
Care ders
was taken consistently
studies. ROIs tude images to the
fore
ROI
and
window
to the
two
phase-contrast
surements
and
were
subtracted.
contrast a 256
cine MR x 256 matrix
data were and two
ages. Respiratory compensation
compensation in the readout
The
parameters
following
of view,
mm;
18-24
cm;
flip angle,
msec/echo and
collected signal were
time
(TE) per
used:
=
field
velocity
24-50/9.5-12;
cycle.
out
causing
flow
gions tures
(5). A correction of these effects that were
contained identified
error image, the measured regions and function
Volume
of space,
185
in the
flow
means
both
3c).
Be-
level
set
signal Whenever
and
regression. blood
flow
with
both
arterial
modalities
was
images errors magnetic for was
are suscaused field
cycle) plot-
a cor-
RESULTS The
the possible used. Re-
sured
Number
#{149}
C11 + C2y
3
(5).
(3)
mean
venous
blood
flow
for each
is
flow
vessel.
Results
flows
combined.
The
mean
dif-
ference between measurements obtamed with MR imaging and those obtained with the flowmeter in all vessels was -11 mL/min ± 11 (standard error of the mean). Average blood flow measured with both methods was calculated in each vessel and compared by means of lin-
regression
.995
(Fig 4). The
of correlation,
(standard
[SEE]
=
mea-
with MR imaging was 710 mL/ mm ± 230 (SD) versus mean venous blood flow of 751 mL/min ± 238 measured with the flowmeter (ii = 8). The corresponding mean arterial blood flow was 137 mL/min ± 6 (measured with MR imaging) and 131 mL/min ± 6 (measured with the flowmeter)
of the
error
3.59
technique
following
r, equal to estimate
mL/min). The intercept zero and the slope was 1.0, indicating that
is accurate
without
lyzed tions
fit to the
data:
flowmeter (where
x
MR
imaging),
=
mL/min;
flowmeter r = .980,
SEE In addition, cardiac cycle
flow
blood r
for =
+
arterial 0.88x
flow
+ 0.95x 5.71
blood mL/
measured
.963, SEE = blood flow,
=
venous
73.6 =
for 5.87
=
with
blood
venous ana-
separately, the following equafor the lines described the best
mm 1.66
the
need
for a calibration curve. When and arterial blood flows were
flow,
no vascular struc(Fig 3a). A velocity
t’() +
blood
linear
as follows: =
measurements
in blood
of the paired Student t test were not significant for differences between portal or arterial blood flows or all
was nearly approximately
blood
over
representative cardiac cycle to assess relation of pulsatile blood flow.
difference
coefficient MR
data
measurement
paired
equation for the line described the best linear fit to all data (venous plus arterial): flowmeter = -7.54 + 1.053x mL/min (where x = blood flow measured with MR imaging), with the
from
(ie,
of
ear
average
cardiac
range
intensity sub-
In addition,
a representative
the
of the large
flows measured, a more relecomparison of the techniques
to half
calculated.
flow calculated with flowmeter
resolved
during
cycle
were
of linear
ted for each
e(x, y), was calculated with velocity data in these static a least squares fit to a linear
e(x, y)
(10). bor-
aliasing.
Phase-contrast velocity ceptible to additive velocity by eddy current-induced changes presence
3b,
window
were
blood compared
measured
Velocity
encoding (v) was varied from 20 cm/sec for venous flow to 100 cm/sec for arterial flow. The was chosen in each case to be near the minimum value possible with-
(Fig
intraluminal image.
16 frames
Average data was
flow
data
display
width
temporally
(TR)
vascular and within
drawn on the magnicine set and then ap-
selection,
of blood
by
3-5
time
msec
cardiac
rate
of ROI
vessel motion during the cycle was separate ROIs were selected for each The rate of blood flow during each
of the
flow used.
thickness,
30#{176}; repetition
16 frames
with aver-
and
were
section
stantial seen, frame.
meaPhase-
of
of blood
to estimate between
the maximum in the magnitude common
images
to choice
were of the
was
set.
measurements
flow
plied
image
velocity
cine
Phase-contrast
1200
areas
error
6). Because
=
blood vant
mL/min,
mL/min.
plots of the combined were made for arterial
measured
with
both
meth-
ods. flow
A representative plot of blood in the iliac artery is shown in Figure 5. Excellent agreement is seen.
Radiology
811
#{149}
Blood
DISCUSSION This
study confirms of phase-contrast
the
in vivo cine MR
curacy aging for the measurement of blood flow. A large range of mean blood flows were included in these measurements
(Fig
4). This
was
acim-
purposely
accomplished by inclusion of different blood vessels to represent the physiologic range of blood flows that might be seen clinically and to avoid artificially inflated correlation coefficients that can be observed over a narrow range (11). In addition, a large variation in blood flow exists because of pulsatility within a representative cardiac cycle, which can be appreciated by
examination
Blood neously
flowmeter
of Figure
flow
measured
with
MR
imaging
5.
simultaand
the
US
did
not vary significantly across a large range of average blood flows. The accuracy of the US flowmeter technique across the range of blood flows included in this study is within plus or minus 10% (12,13) of the measured value. The observed temporal and quantitative correlation of pulsatile blood flow throughout the cardiac cycle is excellent between the two methods. In cases of high heart rates, long TR, or both, MR measurements have limited temporal resolution. In this study, heart rates were 112-164 beats per minute, a range that produced a minimum of seven data points during each cardiac cycle before interpolation by the cine algorithm (with a worst-case TR of 50 msec). The number of sequences per heart cycle controls the temporal resolution of the data. If this number is too small, pulsatile flow waveforms will underestimate cyclical variations. In our data, underestimation of peak flow was less than 15%. Furthermore, limited temporal resolution should not produce errors in the measured mean flow rate. Our data support this position.
Our study examined blood large veins and medium-size
812
Radiology
#{149}
flow in arteries.
flow
studied.
in large
arteries
However,
the
was
principal
not limi-
tations in quantitation of blood flow with phase-contrast MR imaging are associated with the ratio of flow yelocity to Venc and to the number of pixels across the vascular diameter. Slow blood flow is more difficult to measure than fast blood flow because longer TE is required. Thus, the results are expected to be representative of what should be achieved in studies of large arteries if the pixel size and Venc are appropriately scaled. A limiting factor to the use of this technique is the requirement for a regular sinus rhythm. Our implementation is capable of discarding data from approximately 20 ectopic beats for the duration of the acquisition. Once this number is exceeded, the acquisition continues as if in a regular rhythm, and gating may be maccurate. When this occurs, the pulsatile nature of the flow waveform may be lost,
but
the
measured
mean
rate
References
2.
B -*-S
Frami
Figure
measured
4.
Nayler CL, Firmin DN, Longmore DB. Blood flow imaging by cine magnetic resonance. J Comput Assist Tomogr 1986; 101: 715-722. Spritzer CE, Pelc NJ, Lee JN, Evans AJ, Sostman HD, Riederer SJ. Rapid MR im-
aging
of blood
flow
(in milliliters
per
with
MR
imaging
and
the
flow-
5.
6.
frame
Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR. Phase contrast cine magnetic resonance imaging. Magn Reson Q 1991; 7:229-254. Kondo C, Caputo GR, Semelka R, Foster E, Shimakawa A, Higgins CB. Right and left ventricular stroke volume measurements with velocity-encoded cine MR imaging: in vitro and in vivo validation. AJR 1991; 157: 9-16.
7.
Caputo
8.
Right and left lung perfusion: in vitro and in vivo validation with oblique-angle, yelocity-encoded cine MR imaging. Radiology 1991; 180:693-698. Firmin DN, Nayler GL, Klipstein SR. Un-
GR, Kondo
derwood
T, et al.
SR. Rees RSO, Longmore
vivo validation Comput Assist
9.
C, Masui
of MR velocity Tomogr 1987;
Pelc NJ, Shimakawa
A.
DB.
In
imaging. 11:751-756.
Reduced
pulsatil-
artifacts in magnetic resonance angiography (abstr). In: Book of abstracts: Society of Magnetic Resonance in Medicine 1991. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1991; 821. Pelc NJ, Sommer FG, Enzmann DR. Pelc ity
LR, Glover
11.
12.
13.
GH.
Accuracy
and precision
of phase-contrast MR flow measurements (abstr). Radiology 1991; 181(P):189. Armitage P. Statistical methods in medical research. Oxford, England: Blackwell Scientific, 1971; 186-213. Burton RG, Gorewit RC. Ultrasonic flowmeter uses wide-beam transit-time technique. Transonics Systems 1984; 15:68-73. Hartman JC, Koerner J, Lancaster L, Gorczynski R. In vivo calibration of a transit
time ultrasound cending macologist
system
for measuring
aorta volume flow 1985; 27:217.
(abstr).
asPhar-
flow with a phase-sensitive,
limited-flip-angle,
sequence: ogy 1990;
Blood
meter. Numbers on the x axis indicate number of composite cardiac cycle.
556.
3.
5.
minute) versus time during a composite cardiac cycle of blood flow in the iliac artery
10.
SingerJR, Crooks LE. Nuclear magnetic resonance blood flow measurements in the human brain. Science 1983; 221:654-656. Edelman RR, Mattle HP, Kleefield J, Silver MS. Quantffication ofblood flow with dynamic MR imaging and presaturation bolus tracking. Radiology 1989; 171:551-
M
-a--
of
blood flow should remain accurate. We have demonstrated that phasecontrast cine MR imaging can produce accurate estimates of arterial and venous blood flow in vivo over a large physiologic range. This technique is already being applied in a variety of clinical and experimental situations and represents a major advance in noninvasive measurement of blood flow. #{149}
1.
B
gradient
preliminary
recalled
experience.
pulse
Radiol-
176:255-262.
December
1992