Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article. Authors’ Disclosures or Potential Conflicts of Interest: Upon manuscript submission, all authors completed the author disclosure form. Disclosures and/or potential conflicts of interest: Employment or Leadership: None declared. Consultant or Advisory Role: G.H. White, CLSI; Andrea R. Horvath, CLSI. Stock Ownership: None declared. Honoraria: None declared. Research Funding: None declared. Expert Testimony: None declared. Patents: None declared. Other Remuneration: G.H. White, CLSI; Andrea R. Horvath, CLSI.

Endocrinology Prince of Wales Hospital Sydney, Australia 4 School of Medical Sciences University of New South Wales Sydney, Australia. *

Address correspondence to this author at: Chemical Pathology Directorate SA Pathology Flinders Medical Centre Bedford Park, SA 5042, Australia Fax ⫹61-8-8374-0139 E-mail [email protected]. Previously published online at DOI: 10.1373/clinchem.2014.227645

Biological Variation of Hemoglobin A1c: Consequences for Diagnosing Diabetes Mellitus To the Editor:

Graham H. White1,2* Craig A. Campbell3,4 Andrea R. Horvath3,4

For optimal monitoring and diagnosing of patients with diabetes by use of glycated hemoglobin (Hb A1c)1 measurements, the analytical CV (CVa) of the Hb A1c assay and the within-person biological variation (CVwp) are of relevance. CVwp refers to an inherent biological variation around an individual patient’s set point of a biological parameter. Several studies have been published assessing the CVwp of Hb A1c (1–3 ). However, none of those studies used ⬎1 Hb A1c method to determine the CVwp. The aim of this study was to apply 4 different IFCC and National Glycohemoglobin Standardization

Chemical Pathology Directorate SA Pathology Flinders Medical Centre Bedford Park, Australia 2 Medical Biochemistry Flinders University Bedford Park, Australia 3 SEALS North Department of Clinical Chemistry and

© 2014 American Association for Clinical Chemistry 1 Nonstandard abbreviations: Hb A1c, glycated hemoglobin; CVa, analytical CV; CVwp, withinperson biological variation; NGSP, National Glycohemoglobin Standardization Program; SRMP, secondary reference measurement procedure; DCCT, Diabetes Control and Complications Trial; PRMP, primary reference measurement procedure; CVbp, between-person biological variation.

References 1. Lundberg GD. When to panic over abnormal values. Med Lab Obs 1970;4:47–54. 2. Lundberg GD. Critical (panic) value notification: an established laboratory policy (parameter). JAMA 1990;263:709. 3. Hortin GL, Csako G. Critical values, panic values, or alert values? J Clin Path 1998;109: 496 –7. 4. Piva E, Plebani M. From “panic” to “critical” values: which path toward harmonization? Clin Chem Lab Med 2013;51:2069 –71. 5. Campbell CA, Horvath AR. Harmonization of critical result management in laboratory medicine. Clin Chim Acta 2014;432:135– 47.

1

1570 Clinical Chemistry 60:12 (2014)

Program (NGSP) certified secondary reference measurement procedures (SRMPs) using different assay principles and calibrated in SI units (mmol/mol) and Diabetes Control and Complications Trial (DCCT) units (% Hb A1c) to see whether there were differences in the CVwp obtained. In addition, we addressed the consequences for diagnosing diabetes mellitus of the CVwp found. We recruited 21 presumed healthy hospital employees to participate in the study (11 men and 10 women). Five K2EDTA-treated whole blood samples were collected from each individual every 2 weeks for 2 months. On collection, aliquots of each sample were immediately stored at ⫺80 °C. Full analysis was performed at the end of the 2-month collection period. The samples were analyzed in a single run in duplicate using the following 4 SRMPs: • Tina-quant Gen.2 HbA1c on Integra 800, immunoassay, IFCC and NGSP certified (Roche Diagnostics); • Premier Hb9210, boronate affinity chromatography HPLC, at the time not yet officially certified (Trinity Biotech); • Tosoh G8, Cation-Exchange HPLC, IFCC certified (Tosoh Bioscience); and • Ultra2, boronate affinity chromatography HPLC, IFCC and NGSP certified (Trinity Biotech). All 4 SRMPs have documented good results in the IFCC and NGSP monitoring programs [CV ⬍3.0% in SI units, ⬍2.0% in DCCT units, no bias or a very small bias (⫾1 mmol/mol) compared to the IFCC primary reference measurement procedure (PRMP)] and were calibrated using the IFCC secondary reference material with assigned IFCC and derived DCCT values. The data were analyzed using a 2-level nested ANOVA model.

Letters to the Editor

Table 1. CVa, CVwp, and CVbp of Hb A1c in healthy individuals.a Grand mean, %

Total CV, %

CVa, %

CVwp, %

CVbp, %

mmol/mol

34.2 (33.8–34.6)

7.6 (5.9–11.0)

1.3 (1.2–1.5)

0.8 (0.4–1.2)b

7.4 (5.6–10.9)

%

5.28 (5.24–5.32)

4.5 (3.5–6.5)

0.8 (0.7–0.9)

0.5 (0.3–0.7)c

4.4 (3.3–6.4)

2

Ultra

Premier mmol/mol

32.5 (32.1–32.9)

7.6 (5.9–11.0)

0.7 (0.6–0.8)

1.3 (1.1–1.6)

7.5 (5.7–10.9)

%

5.13 (5.09–5.16)

4.5 (3.5–6.5)

0.4 (0.4–0.5)

0.8 (0.6–0.9)

4.3 (3.3–6.3)

mmol/mol

33.4 (32.9–33.8)

8.2 (6.3–11.9)

0.7 (0.6–0.8)

1.7 (1.4–2.0)b

8.0 (6.1–11.7)

%

5.20 (5.17–5.24)

4.8 (3.7–7.0)

0.4 (0.4–0.5)

1.0 (0.8–1.2)c

4.7 (3.6–6.9)

mmol/mol

32.6 (32.3–33.0)

7.1 (5.6–10.1)

1.9 (1.7–2.2)

1.4 (0.9–1.8)

6.7 (5.1–9.8)

%

5.14 (5.10–5.17)

4.1 (3.2–5.9)

1.1 (1.0–1.3)

0.8 (0.5–1.1)

3.9 (2.9–5.7)

Tosoh G8

Tina-quant

a b

Data are SI units (mmol/mol) and DCCT units (%) (95% CI). and c are significantly different from each other.

This analysis model assumes homogeneity, normality, and independence when estimating CIs. The homogeneity of analytical imprecision was confirmed using Cochran test, and within-person variation was confirmed using Bartlett test. Homogeneity was achieved by removing 2 individuals from the data set (P ⬎ 0.05/ 8 ⫽ 0.006, P value adjusted for multiple testing). One individual was removed from the data set because of a clear linear increase in Hb A1c values during the 2-month testing period (i.e., dependence of data), and the other was removed from the data set because of increased Hb A1c values (previously undiagnosed diabetes). Normality was checked using Shapiro–Wilk test (P ⬎ 0.006) for both within-person and analytical components for the 4 methods. The index of individuality was calculated as CVwp/between-person biological variation (CVbp). The mean CVwp of 19 healthy individuals using 4 SRMP Hb A1c methods was 1.3% in SI units (range 0.8% to 1.7%) and 0.8% in DCCT units (range 0.5%

to 1.0%), depending on the Hb A1c method used (Table 1). The CVwp measured with the Ultra2 was significantly different (95% CI CVwp in SI units: 0.4 –1.2) from the Tosoh G8 (1.4 –2.0) but not from Roche Tina-quant Gen.2 (0.9 –1.8) or the Premier Hb9210 (1.1–1.6). All CVs calculated in DCCT units were much lower compared with CVs calculated in SI units (Table 1) owing to the lower specificity of the NGSP PRMP. The master equation for converting SI units to DCCT units is: DCCT ⫽ (0.0915 * IFCC) ⫹ 2.15, where the positive y-intercept value reflects the nonspecificity of the NGSP PRMP method. When Hb A1c measured with the IFCC PRMP is zero, the NGSP PRMP method still measures “something” (low specificity), and this nonspecificity is added to the results, making the results in DCCT units higher. Because the SD will be the same in the 2 situations, the CV will be lower when calculated in DCCT units. The NGSP PRMP measures Hb A1c on the interval scale because of the positive y-intercept, whereas the IFCC PRMP measures on the ratio

scale (4 ). In principle, CVs should not be calculated for constituents measured on the interval scale. In previous studies, CVwp in DCCT and SI units have been used interchangeably and compared with each other (2, 3, 5 ). The Premier Hb9210 and the Ultra2 use the same assay principle measuring the same measurand, and therefore the difference in CVwp measured with the Tosoh G8 and the Ultra2 cannot be explained by measuring different measurand, since the CVwp measured with the Tosoh G8 and the Premier Hb9210 did not significantly differ from each other. As shown in Table 1, the CVbp was large (Table 1) and Hb A1c had a marked individuality. The index of individuality varied between 0.11 and 0.22. The mean Hb A1c of the healthy individuals ranged from 28 mmol/mol (4.7% DCCT) to 39 mmol/mol (5.7% DCCT). With the publication of the WHO and American Diabetes Association guidelines advocating the use of Hb A1c for the diagnosis of diabetes, there have been major shifts in the role of Hb A1c testing. The low inClinical Chemistry 60:12 (2014) 1571

Letters to the Editor dex of individuality found in this study confirms the conclusion of Braga et al. (2 ) that the individuality in Hb A1c is limiting the use of outcome-based or populationbased reference limits for test interpretation and will therefore also limit the use of a fixed cutoff value of 48 mmol/mol (6.5% DCCT) for the diagnosis of diabetes for all individuals/patients. The reason for this is that a small change in Hb A1c may indicate a clinically significant change, e.g., development of diabetes in a patient, even if the difference between the 2 results does not cross the threshold. The CVwp of stable diabetic patients should especially be used for monitoring purposes. However, as shown by Carlsen et al. (3 ), the CVwp of stable diabetic patients does not differ significantly from that of healthy individuals, and therefore it is probable that the findings in this study will also refer to stable diabetes patients. In conclusion, the withinperson biological variation of Hb A1c in healthy individuals is very low compared with the betweenperson biological variation of Hb A1c, affirming the absolute individuality of Hb A1c. The data also suggest that population-based reference limits and fixed cutoff values should be more closely examined for the diagnosis of diabetes. CVs in DCCT units are lower than CVs in SI units and therefore cannot be directly compared.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article. Authors’ Disclosures or Potential Conflicts of Interest: No authors declared any potential conflicts of interest.

1572 Clinical Chemistry 60:12 (2014)

References 1. Rohlfing C, Wiedmeyer HM, Little R, Grotz VL, Tennill A, England J, et al. Biological variation of glycohemoglobin. Clin Chem 2002;48:1116 – 8. 2. Braga F, Dolci A, Montagnana M, Pagani F, Paleari R, Guidi GC, et al. Revaluation of biological variation of glycated hemoglobin (HbA1c) using an accurately designed protocol and an assay traceable to the IFCC reference system. Clin Chim Acta 2011;412:1412– 6. 3. Carlsen S, Petersen PH, Skeie S, Skadberg Ø, Sandberg S. Within-subject biological variation of glucose and HbA1c in healthy persons and in type 1 diabetes patients. Clin Chem Lab Med 2011;49:1501–7. 4. Weykamp CW, Mosca A, Gillery P, Panteghini M. The analytical goals for hemoglobin A1c measurement in IFCC units and National Glycohemoglobin Standardization Program units are different. Clin Chem 2011;57:1204 – 6. 5. Braga F, Dolci A, Mosca A, Panteghini M. Biological variability of glycated hemoglobin. Clin Chim Acta 2010;411:1606 –10.

Erna Lenters-Westra2,3* Thomas Røraas4 Roger K. Schindhelm5,6 Robbert J. Slingerland2,3 Sverre Sandberg4,7,8 2

Department of Clinical Chemistry Isala Zwolle, the Netherlands 3 European Reference Laboratory for Glycohemoglobin Isala Zwolle, the Netherlands 4 Norwegian Quality Improvement of Primary Care Laboratories (NOKLUS) Haraldsplass Hospital Bergen, Norway 5 Department of Clinical Chemistry Hematology & Immunology Medical Centre Alkmaar Alkmaar, the Netherlands 6 Department of Clinical Chemistry and Hematology Gemini Hospital Den Helder the Netherlands 7 Department of Global Health and Primary Health Care Faculty of Medicine and Dentistry University of Bergen, Norway 8 Laboratory of Clinical Biochemistry Haukeland University Hospital Bergen, Norway *Address correspondence to this author at: Isala

Department of Clinical Chemistry Dr. Van Heesweg 2 8025 AB, Zwolle, the Netherlands Fax ⫹31-38-4242-676 E-mail [email protected] Previously published online at DOI: 10.1373/clinchem.2014.227983

Carcinoembryonic Antigen Monitoring to Detect Recurrence of Colorectal Cancer: How Should We Interpret the Test Results? To the Editor: It is routine clinical practice, supported by national guidelines in both North America and Europe, to measure blood carcinoembryonic antigen (CEA)1 to detect recurrence of colorectal cancer during followup after primary treatment. Blood CEA is usually measured every 3– 6 months, and patients with a CEA concentration above an absolute threshold (5 ␮g/L according to American Society of Clinical Oncology guidelines) are investigated further by radiological imaging. However, the evidence underpinning both guidelines and routine practice is weak. We recently reported the interim results of the Follow-up After Colorectal Surgery (FACS) trial, a clinical trial comparing different types of posttreatment follow-up in 1200 patients with colorectal cancer (1 ). This trial confirmed that measuring CEA is an effective way of detecting recur-

© 2014 American Association for Clinical Chemistry 1 Nonstandard abbreviations: CEA, carcinoembryonic antigen; FACS, Follow-up After Colorectal Surgery; AUC, area under the curve. Disclaimer: The views and opinions expressed therein are those of the authors and do not necessarily reflect those of the Health Technology Assessment program, the National Institute for Health Research, the National Health Service, or the Department of Health.

Biological variation of hemoglobin A1c: consequences for diagnosing diabetes mellitus.

Biological variation of hemoglobin A1c: consequences for diagnosing diabetes mellitus. - PDF Download Free
71KB Sizes 0 Downloads 9 Views