Werner Moshage, Kurt Bachmann,

MD MD

Stephan Achenbach Peter Wegener, MS

Biomagnetic Arrhythmias’

complexes

(PVCs),

three

with ventricular tachycardia four healthy subjects with induced paced beats were recorded for 2-15 minutes. After correction for superimposed repolanization activity, the site of origin of the arrhythmias was localized from the magnetic field and

distribution

at the

onset

of the

B

ec-

topic beats. The localization results of paced beats showed an error of a few millimeters in relation to the position of the catheter tip. The results of spontaneous PVC and VT were confirmed with endocardial mapping or associated with ischemic lesions. The authors conclude that multichannel magnetocardiographic studies enable the completely noninvasive localization of ventricular arrhythmias.

every electric a magnetic

causes

trical

excitation

specific

of the

magnetic

current field, the

heart

field

These

patterns can be measured without contact over the thorax to obtain the magnetocardiogram (MCG). These magnetic fields, unlike the common electrocardiogram, remain almost uninfluenced by the conductive properof the

tissues

cal sources system.

and

between

the

the

SUBJECTS Study

measurement

Therefore,

the

field

sequentially

over

many

points of a measuring grid to record sufficiently large field distribution. Localization could be done only in

a

the Wolff-Parkinson-White

1991;

180:685-692

in

syndrome

The recent development channel, superconducting interference-device

of multiquantum

systems

opened

I

From

Poliklinik, Oestliche

the

Medical

Clinic

II (Cardiology)

and

University of Erlangen-Nuremberg, Stadtmauerstrasse 29, 8520 Erlangen, Germany (W.M., S.A., KG., A.W., KB.); and the Medical Engineering Group, Siemens, Erlangen, Germany (P.W., S.S., W.H.). From the 1990 RSNA scientific assembly. Received November 29, 1990; revision requested February 11, 1991; revision received May 2; accepted May 9. Address reprint requests to W.M. C RSNA, 1991

as 37 sensors,

these

systems

can

record the magnetic field distribution over a sufficiently large area without repositioning

of the

system

or the

and

four

healthy

of these associated

10 patients, the extrasystoles with ischemic myocardial

were dam-

age documented by means of coronary angiography, cardiography of the left yentide, and MR imaging. The other six patients

presented

Three patient,

duced dial

parasystole.

with

patients had recurrent VT. In one several episodes of VT were inby means of programmed intracar-

stimulation

during

registration

MCG. Some of the episodes minated by burst stimulation; nonsustained. Two patients incessant

of the

had to be tersome were showed spon-

VT.

In four healthy subjects, paced beats were induced at a known site within the heart by use of the specially designed nonmagnetic pacing catheter (described subsequently) biomagnetic

to verify localization.

Registration

and

the accuracy

of

of the

Evaluation

MCG

new fields of clinical applications for the biomagnetic method (5). With as many

arrhythmias

i3

with paced beats (Table). All patients and healthy subjects had given informed consent after the nature of the study had been fully explained. Ten patients showed spontaneous, pnedominantly monomorphic PVCs. In four

taneous

repeated of accesdelta wave

a group of i7 individuals: spontaneous or induced

volunteers

distributions measured over the body surface permit the quantification and the three-dimensional localization of sources within the heart. They also permit the evaluation of the propagation of physiologic and pathologic heart excitation. Even though the magnetocardiographic method was described as early as 1963 by Baule and McFee (1), its clinical applicability was limited for a long time because measurements had to be obtained with systems composed of only one to seven channels. These systems had to be adjusted

METHODS

Group

ventricular

magnetic

AND

We studied patients with

electri-

(2-4). Radiology

MD H#{228}rer,MS

graphic localization in patients with premature ventricular contraction and ventricular tachycardia (VT) and the verification of the localization accuracy of the method.

elec-

creates

patterns.

disorders with periodically signals, such as localization sony pathways from the

Index terms: Biomagnetism #{149} Heart, experimental studies, 51.1299 #{149}Heart, function, 515.1429,524.1429 S Heart, rhythm

#{149} Andreas Weikl, PhD #{149} Wolfgang

of Ventricular

ECAUSE

ties

patients (VT),

G#{246}hl,MD Schneider,

#{149}

Localization

The magnetic fields caused by electrical activity of the human heart can be coherently measured with a highly sensitive, multichannel, superconducting quantum interference-device system and can enable noninvasive localization of the underlying electrical activity. The magnetocardiograms (MCGs) of 10 patients with spontaneous premature ventricular

#{149} Konrad Siegfried

#{149} #{149}

The MCG was registered over the thorax without physical contact by use of a planar

biomagnetic

(Krenikon; (5).

pa-

lying

The

Siemens, patients

on a wooden

and

multichannel

system

Erlangen,

Germany)

healthy

couch

subjects

inside

were

a magnet-

tient. Besides drastic shortening of the recording times, coherent registration of the data for the first time permits study of spontaneous on sporadic events.

The cal

goal

application

of this

study

was

of magnetocardio-

the clini-

Abbreviations: ECG = electrocardiogram, MCC = magnetocardiogram, PVC = premature ventricular complex, S/N = signal-to-noise ratio, VT = ventricular tachycardia.

685

ically

shielded

room.

recordings

The

ranged

addition channels,

duration

from

of the

2 to 15 minutes.

In

Examinations

ration

(ECC)

channel

were

leads

and

recorded

Subject No.

simulta-

To localize the magnetic

the electrical activity from field distribution, the PVC,

beats

during

tachycardia,

were

averaged

or paced

to reduce

beats

background

and thus improve localization (6). Only data sampled during

tion were averaged ence of respiratory localization

to minimize movements

results

(7).

to 50 expiratory tamed

in each

In most the

the onset by

two

be ob-

fields

of the

Digital the

of the PVC was

magnetic

repolarization

separate

study,

could

caused

previous

subtraction

fields

tracted

from

the

combination

to

(8):

was done

To compensate for any the electronic components recording

periods

channel

separately,

signal

during

tracted

15

For

mean

recorded

intervals

was

the measured

these

steps

of electrical

three

distributions

single finite

localization

imaging

(Figs

Paced

NA

Paced

65/M

NA

Paced

Note.-+ = performed, Echo = echocardiography, resonance (MR) imaging, tar tachycardia.

= not performed, EPS = electrophysical NA = not applicable,

matched

the

Nonmagnetic, Imaging-compatible Catheter

in

field

To

of a

MCG,

dipole in the inspace (9).

with

visible

marks

in MR

to the to the

sponding

plane

positioning support

the same

ECG

to

and

was

with

system fixed

position

cardiography netocardiography,

artifacts

and images

MR the

both imaging. support

corre-

of a special

(5): A T-shaped to the patient’s

and

gat-

were

of the

the help

during

by

respiration

of the MCG MR

plastic chest in

magnetoFor magcarried

four current-driven wire coils, which were replaced by agaroseor copper sulphatefilled cubes during MR imaging. The position of the wire coils in magnetocardiography was determined by measuring the magnetic

field

rent

run

686

#{149}

through

Radiology

caused

+

-

-

+

+

-

+

+

+

-

+

+

+

-

+

+

+

-

+

+

+

-

+

+

+

-

+

+

+

+

4-

+

+

+

+

+

+

+

+

+

+

-

-

+

+

-

-

+

+

-

-

+

+

-

-

+

+

im-

the

Pacing

localization

a bipolar

pacing

in a way that exactly known

accuracy

catheter

of the

was

permitted site within

de-

stimulation the heart The of tip,

two platinum electrodes apart. The copper lead-in

are fixed 10 mm wires are

twisted so that the from the electrodes

current not cause

turbing net the lead-in

electric does

by

the coils

the

and

known

could

cur-

be

to the

magnetic field. Additionally, wires detach easily from

radio-frequency

MR imaging visibility is filled

at MR imaging, with lyophilized

tetate

dimeglumine,

the

To imthe cathegadopen-

echo sequences planes. Several

during

an MR imaging

contrast agent. Preliminary examinations with the catheter placed in a tank full of saline solution confirmed the compatibility of the catheter with biomagnetic measurements.

In the four magnetic

healthy

pacing

subjects,

catheter

was

the

non-

stably

posi-

tioned at the apex of the right ventricle via the brachial or subclavian vein. During registration of the MCG, paced beats were induced scribed

according hereafter.

to the

protocol

After magnetocardiography, wires were removed, with remaining in place, under guidance. the catheter MR imaging

fields

can be prevented.

prove ter tip

Figure 1. Schematic representation of digital subtraction performed in a PVC emerging from the repolarization of a normal beat. Averaged normal beats were subtracted from the combination of a normal beat followed by an extrasystole to obtain the isolated extrasystole. This procedure was carried out in each channel separately.

to and a dis-

due

of movement

coordinates

+

MR

during registration of the MCC (10). bipolar catheter (5 F) is made entirely nonmagnetic material. At the catheter

of MR

frontal,

verify

signed at an

MR

assigned

application

in axial,

with

and

triggering ing. The transferred

-

= angiography, CHD = coronary heart disease, study, MI = myocardialinfarction, MR.! = magnetic PVC = premature ventricular complex, VT = ventricu-

electrodes and can be removed with the catheter remaining in its intracardial position. In this way, heating of the electrodes

free

+

-

AC

-

sagittal planes. MR imaging was chosen because soft tissues can be pictured clearly and because cross-sections of the thorax can be obtained in every desired plane made

+

+

beats

the

model

were

by

2-7)

+

+

aging.

localized

the

results

structures

NA

50/M

17

at each

Assignment

anatomic

-

beats

each

processing,

was

current half

Topographic Imaging The

of data

applying

equivalent homogeneous

42,/M

16

sep-

from the magnetic

by

+

sub-

signal

activity

dimensions

+

beats

instant.

After

51/M 67/M 76/NI 1WM 41/M 65/M 41/F

MRI

during

activity.

the

these

from

performed

zero

41/1s4

PVC PVC PVC PVC PVC PVC PVC PVC PVC PVC VT VT VT Paced

Echo

additional

was

of known

7

Parasystole Parasystole Parasystole Parasystole Parasystole Parasystole CHD, MI CHD, MI CHD, MI CHD Idiopathic CHD, MI Cardiomyopathy NA

AG

offset caused by of the measur-

systems,

correction

42/F 45/M 22/F 47/M 50/M 67/M

EPS

of a normal

line (Fig 1). This procedure arately in all channels.

and

I 2 3 4 5 6

PVC, VT, or Paced Beats

A

beat and the following PVC to obtain isolated PVC activity arising from zero base-

baseline

(y)/Sex

Multichannel

beats

subtraction template was obtained by averaging normal beats. After careful time alignment, the template was digitally sub-

site

Underwent

nor-

was done

superimposed

Age

8 9 10 11 12 13 14

subject.

cases,

mal beat.

accuexpira-

the influon the

In this

averages

superimposed

ing

Who

Abnormality of the Heart

a respi-

neously.

by

in 17 Subjects

to 32 of 37 available magnetic the signals of 12 standard elec-

trocardiographic

noise racy

Performed

de-

exclude within raphy,

Subsequently, tip could be by application in three measures

a movement the heart: the catheter

the lead-in the catheter fluoroscopic the position documented of gradientorthogonal were taken

orthogonal the

registration

of the views

to

of the catheter Before magnetocardiogwas firmly anchored

the trabecular system of the right cle, and the position of the catheter was documented with orthogonal graphs and markers on the body Fluoroscopy

of with

catheter was

of the

position

repeated MCC,

in

ventritip radiosurface. just

before

in before and

after MR imaging, and especially during removal of the lead-in wires. In no case was a movement or change of position

September

of

1991

a. Figure

b.

C.

2. Biomagnetic localization results in a patient with a parasystolic focus high in the interventricular septum, verified with an endocardial catheter mapping procedure. The magnetic field distribution recorded at the onset of the PVC and the magnetocardiographic localization result (arrow) projected on the corresponding cross section of an MR image (a) show that the ectopic focus (0 ) lies high in the interventricular septum. The localization results during the first milliseconds of the ectopic beat form a path that indicates the spreading excitation moving distally along the interventricular septum in frontal view (b) and in axial view (c).

cases

of programmed

ulation msec

the after

ration voltage

ECG-triggered

impulse

was

sensing

stim-

applied

of the

300-350

R wave.

The

du-

of the impulse was I msec, with twice the stimulation threshold

a

(0.8-1.2 V). The stimulation pulse was followed by a low-voltage compensation current

of 12-23

msec

to prevent

effects at the electrodes Induction of VT-VT

polarizing

(Ii).

was induced during registration of the MCG by means of the nonmagnetic catheter. The stimulation protocol included one to three extrastimuli at three basic cycle lengths (600, 500,

and 400 msec) in the right ventricular apex. If VT was sustained, it was terminated by ventricular burst stimulation. Endocardial mapping-To determine 4-

Figures

3, 4.

(3) Biomagnetic localization result of monomorphic PVC in a patient with akinesia of the left ventricular apex. The results of magnetocardiographic localization, obtained during the onset of the PVC transferred to the corresponding MR section, depict the excitation spreading from the margin of the postinfarction scar. (4) Biomagnetic localization result of the origin of monomorphic PVC in a patient with a large aneurysm of the left ventricle. The circle indicates the position of the arrhythmogenic focus, determined with the MCG, at the margin of the lesion. postinfarction

the

exit point of ventricular arrhythmias, endocardial pacing was performed at different sites in the right ventricle, left yentricle, or both. The site of stimulation was systematically

lead

changed

ECG,

paced

the QRS

beats

or PVC

until,

and those

were

very

in the

complexes

12-

of the

of spontaneous

VT

similar.

RESULTS the catheter served.

tip within

During

MR imaging,

of the catheter tal, axial, time

tip were

and

interval

the first and respiratory

the ventricle

sagittal

cross-sections,

of about

45 minutes

last sequences.

weighs

inspiration

Furthermore,

in fron-

QRS

between

physiologic

diastolic

of the right

ventricular

tamed

with

a trigger

msec,

according

during

ratio

an exact

delay

could

be ob-

exceeding

coupling

300

intervals

the nonmagnetic,

that

Volume

pacing rendered

180

presented

topographic possible

Number

#{149}

MR imag-

catheter

intracardial

and,

with programmed registration of the

the

such

if necessary, stimulation MCG.

ing

catheter

raphy:

during

In all four

the

3

referverifica-

beats by

were

ulation

was

during technique

intervals. 100

the site of stimula-

ventricular

induced

extrastimulus

coupling

magnetocardiog-

cases,

The stimuli

apex. sinus

Paced

rhythm

at different frequency of stimper

Localization

of Spontaneous

minute;

in

PVCs

At the onset of every singular extrasystole, bipolar magnetic field patterns were observed that permitted

the three-dimensional an equivalent electrical within the heart. The

Studies

In the course of this study, three different electrophysiologic procedures were performed: Induction of paced beats-Paced beats were induced with the nonmagnetic pac-

hon was the right

ing-compatible ence

to

stimulation.

In this way,

arrhythmias

be induced

Electrophysiologic

cross-sections

apex

to the

terminated during

accuracy.

expiratory

in magnetocardioexpiration out-

in the

complex,

a

localization

complex

as VT could

with

of 2:1. Because MR imaging was conducted with ECG gating and triggered the

of biomagnetic

To exclude only

were evaluated In MR imaging,

lion

coordinates

identical

movements,

events graphy.

ob-

magnetic fields and the localized dipole

Transfer ity,

obtained

of the

site

from

localization of current dipole strength of the the strength increased

of electrical the

magnetic

of quickly.

activfield

distributions during the first milliseconds of the PVC, to the MR image permitted the localization of the ongin of the ectopic beat with respect to morphology.

The

orientations

localized dipoles roughly the direction of propagation

of the

agreed with of the

Radiology

#{149} 687

b.

a.

c.

Figure 5. MCG and electrophysiologic images from a patient with cardiomyopathy and incessant VT. (a, b) Results of biomagnetic localization. The dipole localizations during the beginning of the VT were transferred to the frontal (a) and axial (b) MR images and show the exit point of the VT (0) in the proximal left interventricular septum. (c, d) Results of the electrophysiologic pace-mapping examination. The electrodes of the pacing catheter are placed in

the left ventricle tion that

at the interventricular

at this site creates close of paced beats (d).

excitation.

With

increasingly cardium, represent

and

can

fined

septum

similarity

between

the depolarization

of

large parts of the the localized dipoles a summation of the

no longer

anatomic

In only

myotend to activity

be related

to con-

structures.

two

cases

in this

ectopic beat by repolariza-

tion activity, and simple baseline connection before the onset of the ectopy was sufficient for localization of the equivalent current source. In the other eight cases, the onset of PVC was superimposed by the repolarization activity of the previous beat. Evaluation of these patients showed that correct posed fields

ization.

separation is crucial

Whenever

physiologic

performed known,

had

of superimfor correct

invasive

mapping

and the

had

the

been

focus

was

localization

of up to 80 mm

superimposition

local-

electro-

ectopic

biomagnetic

an error

if the

of magnetic

fields

was not eliminated by means of digital subtraction. Only after the superimposed fields were separated with digital subtraction, as previously descnibed,

subsequent

baseline

con-

rection was performed the onset of the PVC

directly to eliminate

before sys-

tem

and

offset

localized With

could

the

correctly. this method,

ectopic

focus

anatomically

be

and

pathophysiologically meaningful localization was obtained in nine of iO patients. In two patients, the noninvasive biomagnetic localization was con-

firmed by means mapping. In one 688

Radiology

#{149}

below ECC

of endocardial pacepatient, we still had

the aortic

valve

morphology

(circle,

c). Stimula-

of spontaneous

VT and

doubts about the accuracy of the magnetocardiographic localization. Figure 2 consists of images from a patient

with panasystole. magnetic field set

of PVC

study, the onset of the was not superimposed

right the

of the

Figure distribution

extrasystole

2a shows at the projected

d.

the ononto

the MR image of the patient. The position of the MR section corresponds with the depth of the biomagnetically localized equivalent current dipole, which is depicted by the arrow and represents the site of origin of the ectopic

beat.

Invasive

electrophys-

iologic studies performed have proved the localization ectopic

focus

high

tricular septum agreement with

in the

in this case of the interven-

in good topographic the biomagnetic

findings. The propagation of the electrical activity during the first milliseconds of the ectopic beat is shown in

Figure

2b and

clearly

follows

2c. The the

excitation

contours

of the

interventricular septum in a distal direction. Figure 3 shows the result of magnetocardiographic localization in a patient

with

PVC

after

farction.

Coronary

showed

a distal

myocardial

in-

occlusion

of the

left

ventricle was documented at candiography of the left ventricle, echocardiography, and MR imaging. At magnetocandiography, the onset of the ventricular extrasystoles was localized at the margin of the damaged myo-

cardium (Fig 3). In another patient disease,

an

6. The magnetic field distribution and biomagnetic localization (arrow), measured during intracardial stimulation with the amagnetic pacing catheter, projected onto the corresponding MR cross section. The gadolinium-filled catheter tip causes a circumscript dark area while the moving blood is depicted in light gray in gradientecho sequences.

angiography

anterior descending artery. An akinetic area at the apex of the left

heart

Figure

with

extensive

coronary infarction

of the

anterior

wall

had

led

to a large

aneurysm documented with cardiography of the left ventricle, echocardiography, and MR imaging. Two years

after

infarction

sodes tion,

of VT and the patient

On

the

MCG

and

the

after

ventricular presented

ectopic

localized at the margin rysm toward the distal lar septum (Fig 4).

several

epi-

fibrillawith PVC.

focus

was

of the aneuinterventricu-

September

1991

b.

c.

Figure 7. Results of biomagnetic localization during (ta), and sagittal (c) sections in MR images. The circular electrical stimulus (circle) was localized within 4 mm (cross) was localized 9 mm from the site of stimulation.

Localization

ent

field

of VT

For the multichannel

first

time, system

registration

the biomagnetic permitted coher-

of the

magnetic

fields

during times

VT (Fig 8) with recording shorter than 1 minute, which

were

well

tolerated

by the

With the nonmagnetic ten, VT could repeatedly

measurement. The result tion

patients.

pacing cathebe induced

in its intracandial interfering with

of biomagnetic

in all three

agreement

with

the

trophysiologic

result

In all cases,

netic

field

the

at the

the

a stable

showed

only

magnetic

bipolar

and

of

field

pattern.

The

the equivalent and orientation

slight

changes

maximum of the R wave (Fig 9a). After the R-wave orientation of the current quickly changed polarization, the

onset

(Fig field

di-

until

the

was reached peak, the dipole

9b). During minimum

reand

maximum were interchanged compared with those at depolarization; otherwise, the field pattern most identical and remained until the end of nepolarization

9c). Toward the

localized

the

end

was alstable (Fig

of repolarization,

equivalent

current

di-

pole approached the site of the onset of VT. The stability of the magnetic Volume

180

Number

#{149}

3

and toward facilitated of VT.

may

be the

signal-

onset

of de-

the end of localization

reason

why

ously and the repolarization last beats could thus serve template.

It was

the origin

accuracy

simple

of the as a subalso

of VT with

from

that was stimulation

high

unaveraged

MCG recording. agreement with sive

confirmed

tricular

point

part

by

prothe

pace-mapping

the of the

proximal septum

proce-

left yenas the

exit-

of VT (Fig 5c, 5d).

Localization During

of Paced intracardial

localization

with

magnetocardiog-

The stimuli applied after complete repolarization of the previous beat could be localized in three dimensions

after

baseline

correction:

stimulus

registered

in the

mitted the localization lus. The average error and z axes was 8 mm, respectively,

pacing,

prema-

the

MCG

per-

of the stimualong the x, y, 5 mm, and with

an

average

spatial error of 11 mm in all three dimensions. Figure 6 shows the magnetic field distribution during intracardial stimulation and the biomagnetic

localization

indicated

by

the

yellow arrow. Both are projected onto the MR section corresponding to the depth of the localized dipole. The cmdank

mark

is caused

by

the

cath-

eten tip. After the termination of the compensation current (13-24 msec aften the stimulation), the magnetic field pattern caused by the paced beats

was

localization

registered

and

of the

ten tip, the circle netic localization

tune paced beats were induced at the catheter tip. Because the position of the catheter tip could be exactly documented at MR imaging, the exact location of both the electrical stimulus and the following myocardial depolarization was known and thus permitted verification of the accuracy of

In all

four healthy subjects, the bipolar magnetic field distribution during

site

depolarization, shown planes in Figure 7. The responds to the position

Beats

(a), axial tip. The paced beat

raphy.

cular

In good topographic this result, an inva-

endocardial

dune

induced during

catheter transferred to frontal represents the site of the catheter after stimulation, the onset of the

2 mm,

possible

data. Averaging of about 50 beats in VT only slightly improves the localization result. Figure 5a and 5b shows the result of magnetocardiographic localization in a patient with spontaneous VT grammed

characteristic

9). Immediately

magnetic field pole localization

mag-

of VT showed

remarkable

depolarization,

elec-

a high

at the

baseline correction at the very onset of the R wave yielded the same localization result as digital subtraction. The latter could be performed if the VT episodes terminated spontane-

spatial

proce-

registered

distributions

a typical,

showed

of the

(S/N)

polarization repolarization of the origin

to localize

in good

pace-mapping

dune.

(Fig

was

and

ratio

traction

localiza-

patients

distributions

to-noise

This

and terminated during registration of the MCG after the previously mentioned stimulation protocol. The catheter could remain position without

intracardial stimulation with the nonmagnetic dark mark at the apex of the right ventricle from the catheter tip. Thirteen milliseconds

the

cross

depicts of the

represents

the

permitted

of myocardial in all three dark mark conof the cathe-

the biomagstimulus, and site

of electri-

cal activity caused by myocardial vation localized from the MCG msec

after

stimulation.

The

acti13

difference

of about 10 mm from the localization of the stimulus is in good agreement with the known conduction velocity (approximately 1 mm/msec) in the myocardium. Radiology

#{149} 689

In two programmed premature stimulation to apply stimuli during

examinations, was performed the repolar-

ization

beat

of the

previous

I’

--‘

V

y

,1V

y

V

v

y

V

:V

,V

to simu-

late the pathologic activity being superimposed by background activity. The localization result of premature stimuli showed inaccuracies of several centimeters in all directions, with an overall error of 100 mm and 73 mm in the two studies in comparison with stimuli applied at zero baseline if the superimposed repolarization was not correctly eliminated. After digital subtraction of the repolarization template, the localization error of the premature stimuli could be reduced to 7 mm and 9 mm, respectively.

,

1.-’

.#{248}#{149}’ i---

\

t.



.

Jr

lLxj,e

,.,

.‘

vi

‘i

r

ii



-

-

.

.

,.

‘j

:

y



‘x#{149},--

,,\,

.s

‘#{149}‘

4#{149}

.-

2j41Ut .., I’,

#{149}; .-

.-

.

..

\

...

----

-

I

/-,,-.--

-

DISCUSSION Previous reports have indicated the possibility to biomagnetically localize the arrhythmogenic tissue in patients with ventricular extrasystoles and VT (12,13).

However,

these

#{149} Radiology

1

..

1.51S

3’C1’?A.P*T2)b4.#{248}1

e.500

2.SSS

.EXT2.

3.SSS

3.51S

4.SSS

4.511

(SECI

S

measure-

ments were obtained with singlechannel systems. The magnetic fields could not be registered coherently, and the measurements required long recording times. To obtain a sufficient number of expiratory averages, the recording time had to be prolonged even further. Arrythmias cannot be tolerated for a long time by the patient, especially if they are to be hemodynamically effective and thus of clinical relevance. Therefore, such arrhythmias can be only insufficiently studied with single-channel systems. Coherent registration of the magnetic field with biomagnetic multichannel systems in connection with adequate imaging, such as MR imaging, is the technical prerequisite for precise localization of transient events like singular PVC (14,15) or hemodynamically effective arrhythmias such as VT that can only be tolerated for a limited time. In this way, recording times of less than 15 minutes for PVC and less than 30 seconds for VT are sufficient to obtain an acceptable number of expiratory averages. A high S/N provides for exact localization of the ectopic activity. The following factors are among the most important contributors to localization inaccuracies with the biomagnetic method: patient movements, inaccuracy of the transformation of localization results to the MR images, and insufficient source and torso modeling. The greatest localization inaccuracies were found in the examination of deep sources, which confirmed earlier observations in phantom studies with a dipolar 690

..

Figure 8. Magnetocardiographic registration of nonsustained VT with the biomagnetic multichannel system. The magnetic signal recorded during 4.6 seconds is shown in eight of the 37 MCG channels (top). Three of the 12 simultaneously recorded ECG leads are shown beneath this signal. The bottom curve shows the respiration channel.

source filled

submerged with

in a plastic

saline

tank

solution.

The development of a nonmagnetic pacing catheter compatible with the exact three-dimensional imaging technique of MR imaging permitted verification of the localization accuracy of this method. The results of our pacing studies indicate a high precision of magnetocardiographic localization, even if data evaluation and localization are done with simplified modeling rounding

of the volume

future, contribute

the study

more

sources and conductors.

of paced

to the

events

development

sophisticated

Experiments

the surIn the

models. with

the

nonmagnetic

stimulation catheter permitted study and solution of a major lem in magnetocardiography, perimposition

may of

of magnetic

the probthe su-

fields:

If

several sources are active within the heart at the same time, the registered magnetic field always represents a summation of several components. In this study, superimposed fields were recorded in patients with ventricular extrasystoles arising from the repolarization activity of the previous beat. Our results show that the problem can be solved by applying digital subtraction if the disturbing background fields are known (eg, if normal beats not followed by a PVC are registered

and can be used as a subtraction ternplate). The study of VTs revealed some characteristic

properties:

The

localiza-

tion of the equivalent current dipole remained stable and showed a high S/N during the complete upstroke of the R wave. For this reason, the site of origin of the VT could be localized without

cases

digital

even

During

subtraction,

from

in some

unaveraged

repolarization,

the

data. magnetic

field map had reverse polarity and, again, remarkable stability. This observation might correspond to specific patterns and mechanisms of excitation during VT. In one patient with VT, interpretation of the data with simplified models

was

not

possible,

because

the

mag-

netic field did not show a typical bipolar pattern. The high density of isofleld lines in the cramo-left-lateral direction might have been caused by a border effect indicating insufficiency of the model of the infinite half space, which is usually applied to describe the thorax as a volume conductor. Interpretation

of this

with the commonly model of a single yielded centimeters

tours. The

a localization outside

correct

field

used current

distribution

source dipole

result several the heart con-

localization

of the September

site 1991

I

.

! -

.

.

..

00

.‘#{149}#{149}-

iPj

r1 .,

Figure 9. The magnetic field distribution registered coherently with the biomagnetic multichannel system during VT. The field maps stroke ization

during

stable . .

-

.

.

._.i_

--

.

during

;Lt:

.

.

7;;#{149}ir’

.,

#{149}

I

..

w’

L.

..

. ..

no

n

a

electrical

.

Volume

180

#{149} Number

3

Toward

and

dipole

of the

the

remain R wave.

orientation

at

(b) During the

dipole

ap-

end

of repo-

the field map shows a similar disinterchanged polarity with that of the onset of VT (c).

with

activity

of heart

arrhythmias

can be completely and noninvasively examined. This method not only enables localization of the site of origin of arrhythmias, but it also offers new insights in the mechanisms of ectopic

/\

w,

its onset.

repolar-

charactenis-

of origin of VT in this case could only be obtained by the application of a new localization algorithm that is based on the reconstruction of the current density distribution (“lead fields”) (16). A very high number of evenly distributed current dipoles is assumed, and the strength and direction of the dipoles are chosen to optimally describe the measured field distribution. The number of sources is then systematically reduced until a minimal number of dipolar sources still describing the measured field distribution within the S/N is reached. With this algorithm (which, in this study, was applied to clinical magnetocardiographic data for the first time, to our knowledge), a localization result in exact agreement with invasive pace-mapping was obtained. In eight other cases in this study, in which the localization of a single current dipole had yielded verified or physiologically meaningful results, the site of highest current density was found in the same area. These observations may suggest that localization could be improved in some cases if improved source models were used. Adequate evaluation of data recorded with a biomagnetic multichannel system combined with adequate imaging, such as MR imaging, is the first and only method with which electrical

.

changes

the equivalent

compared

1

the upstroke

quickly

repolarization,

tribution

:.ci t

current

of the R wave

larization,

:

down-

field distribution

the maximum proaches

:::

(a) and

wave and during a remarkable

equivalent

The dipole .

:[yT.

:

R

tic: The magnetic localized

..

the upstroke

(b) of the (c) display

:

heart

activity.

Because the annual mortality rate due to cardiac rhythm disorders is about 1,500 per 1 million inhabitants in industrialized nations, a simple and noninvasive method to investigate these disorders is of outstanding importance.

Epidemiologic

studies

with

broad application of this noninvasive method could lead to better underRadiology

691

#{149}

standing and classification of this inhomogeneous group of diseases. The results may, in turn, help to improve methods of treatment in the future. Biornagnetisrn

might

be

a key

5.

al. study

6.

7.

References 1.

2.

3.

4.

692

Baule C, McFee R. Detection of the magnetic field of the heart. Am Heart J 1%3; 66:95-96. Ern#{233} SN. High resolution magnetocardiography: modeling and sources localizalion. Med BiolEng Comput 1985; 23(suppl): 1447-1450. Fenici RR, Melillo C, Maselli M, Capeffi A. Magnetocardiographic three-dimensional localization of Kent bundles. In: Atsumi K, Kotani M, Ueno S, et al, eds. Biomagnetism ‘87. Tokyo: Tokyo Denki University Press, 1987; 140-141. Katila T, Montoneu L, Maekijaervi M, Nenonen J, Raivic M, Siltanen P. Localization of the accessory cardiac conduction pathway. In: AtSUnIi K, Kotani M, Ueno S, et al, eds. Biomagnetism ‘87. Tokyo: Tokyo Denki University Press, 1987; 430-433.

#{149} Radiology

S, Hoenig

Multichannel

E, Rechenberger

biomagnetic

system

H, et for and

of electrical activity in the brain Radiology 1990; 176:825-830. Abraham-Fuchs K, Schneider 5, Reichenberger H. MCG inverse solution: influence of coil size, grid size, number of coils, and SNR. IEEE Trans Biomed Eng 1988; 573-576. Abraham-Fuchs K, Weikl A, Schneider S. et al. Application of a biomagnetic multichannel system to the comparative localization of accessory conduction pathways in patients with WPW syndrome. In: Williamson SJ, Hoke M, Stroink C, Kotani M, eds. Advances in biomagnetism. New York: Plenum, 1989; 369-372. Achenbach 5, Moshage W, Weild A, et al. Elimination of electronic offset and physiological background activity in magnetocardiographic localization. Biomed Tech (Berlin) 1990; 35(suppl 3):160-161. Williamson SJ, Romani CL, Kaufman L, Modena J. Biomagnetism: an interdisciplinary approach. New York: Plenum, 1982. Moshage W, Achenbach 5, BoLz A, et al. A non-magnetic, MR-compatible pacing catheter for clinical application in magnetocardiography. Biomed Tech (Berlin) 1990; 35(suppl 3):162-163. Schaldach M, Boheim C, Edelh#{228}user R. Em beitrag zur erkennung evozierter po-

8.

9.

10.

11.

Biomed Tech (Berlin) 1988; 2):343-344. Fenici RR, Melillo C, Capelli A, Deluca C, Maseffi M. Magnetocardiographic localization of a pacing catheter. In: Williamson SJ, Hoke M, Stroink G, Kotani M, eds. Advances in biomagnetism. New York: Plenum, 1989; 361-364. Fenici RR, Melillo C, Capelli A, Deluca C, Maselli M. Atrial and ventricular tachycardias: invasive validation and reproducibility of magnetocardiographic imaging. In: Williamson SJ, Hoke M, Stroink C, Kotani M, eds. Advances in biomagnetism. New York: Plenum, 1989; 441-444. Moshage W, Achenbach 5, Weikl A, et al. Magnetocardiography (MCG): Noninvasive localization of ventricular extrasystoles with a biomagnetic multichannel system. Presented at Update in Clinical Cardiology, Herzlya, Israel, May 20-22, 1990. Moshage W, Weild A, Abraham-Fuchs K, et al. Magnetokardiographie: erste klinische erfahrungen mit einem biomagnetischen vielkanalsystem. HerzfKreisl 1990; 22:245249. Joannides AA, Bolton JPR, Clarke CJS. Continuous probabilistic solutions to the biomagnetic inverse problem. inverse Problems 1990; 6:523-542. tentiale.

33(suppl 12.

heart.

to solv-

ing a problem that is of preeminent importance because of the high preyalence of life-threatening arrhythmias. #{149}

Schneider

13.

14.

15.

16.

September

1991

Biomagnetic localization of ventricular arrhythmias.

The magnetic fields caused by electrical activity of the human heart can be coherently measured with a highly sensitive, multichannel, superconducting...
2MB Sizes 0 Downloads 0 Views