Neuroradiology Brigitte

Poncelet,

P.

Brain with

(MR)

measured

tion

imaging, the

of brain

sitive,

J.

Van

#{149}

intrinsic

pulsatile

parenchyma.

mo-

Phase-sen-

two-dimensional

planar

cine

throughout a spin-echo,

MR

pulse

images

were

the cardiac cycle blipped echo-

sequence.

Trans-

verse and coronal planes were obtamed in 14 healthy volunteers. Corrections were made for gross head

motion.

Brain

motion

consisted

of a

rapid displacement in systole, with slow diastolic recovery. The motion occurred chiefly in the cephalocaudal

a

and lateral directions; the anteroposterior motions were relatively small. Cephalocaudal velocities increase with

proximity

num.

The

to the

lateral

foramen

motion

mag-

is mainly

compressive motion of the Brain parenchymal velocities high as 2 mm/sec caudally

a

thalami. were in the

as

brain stem and 1.5 mm/sec medially in the thalami. Net parenchymal excursions were at most 0.5 mm. Phasebased echo-planar velocity measurements agreed well with echo-planar Fourier velocity zeugmatography measurements and were consistent with reported values. Velocity mapping with echo-planar imaging offers

a rapid

and

flexible

ing the pulsation man brain.

method velocities

of several techniques, including phase contrast (5,6), Fourier velocity zeugmatography (FVZ) (7,8), and image mnterferography (9). We present a method to map CNS velocity components by using spin-echo echo-planar

MR

imaging,

which

of assess-

the

fluid, MR. (MR), cine (MR), echo planar

representation of the (3D) dynamics of

AND

METHODS

2D sensitivity

echo-planar

image.

is created

by

incorpo-

1992; 185:645-651

From the MGH-NMR Center, Department of Radiology, 13th St. Charlestown, MA 02129. Received December 27, 1992; final reyision received June 30; accepted July 6. Supported GE Medical Systems, and Advanced NMR Systems. Address ‘ RSNA, 1992 See also the article by Enzmann and PeIc (pp 653-660) and this issue.

S. Cohen,

PhD

rating ing

a pair pulse

of unipolar

gradients

velocity-sensitiz-

along

a chosen

direc-

lion into an MR spin-echo pulse sequence (11). This spin-echo sequence allows us to measure slow motion by using long, velocity-sensitizing, gradient pulses without extensive T2* decay, a potential source of limitation in gradient-echo techniques.

Data

Acquisition

All imaging was performed with a 1.5-T Signa imager (GE Medical Systems, Muwaukee) retrofitted with Instascan technology (Advanced NMR Systems, Wilmington, Mass). The imaging pulse sequence was a blipped echo-planar acquisition (12,13) augmented by a pair of unipolar velocity-sensitizing gradient pulses placed before and after the 180#{176} radio-frequency pulse (Fig 1). The yeloc-

ity-sensitizing gradient either the section-select

was applied

along

axis (ie, throughplane motion) or the phase-encoding axis (in-plane motion); amplitudes were 0.11.0 C/cm (0.1-1.0 x 10 T/cm); gradient pulse duration, 8, was 20-40 msec; and the time interval between gradient pulses, , was 23-46 msec (total motion-encoding duration, + 8 = 43-86 msec). These pulses provided velocity sensitivities of 0.7-5.0 radians . mm sec. The velocity sensitivity, K, was made as large as possible, consistent with the re-

quirement not wrap;

that the measured that

is, p

=

K

phases,

V must

p,

satisfy

I’ I

< rr radians at all points. Thus K < radians/V,,), where Vma. 5 the maximum velocity to be imaged. The full expression of ip shows (see Appendix) that the measured phase corresponds to an average measurement of velocity over the duration of the total motion encoding + b. Acquisition parameters were as follows: echo time of 80-120 msec, acquisilion matrix of 64 x 128, in-plane resolution of 3 x 3 mm, and section thickness of 5-10 mm. The sequence was electrocardiography (ECC) gated; an image was collected every one to four R-R intervals with (IT

Our basic technique for brain motion measurement is to analyze a velocity-dependent phase map obtained from a yeVelocity

149,

to

CNS.

locity-sensitized

Radiology

can be used

acquire a two-dimensional (2D) image in less than 0.1 second (10). These rapid rates of acquiring echo-planar imaging data facilitate acquisition of more extensive data sets, offering a

of the

Mark

#{149}

Measurement Imaging’

the rigid cranium, blood, cerebrospinal fluid (CSF), and brain tissue engage in a complex and imperfectly understood three-dimensional (3D) motion. Although different proposed models represent brain motion as the “circulation pump” of the CSF (1,2), direct validations of these models in the human brain have been limited. Because derangements of central nervous system (CNS) dynamics may play an important role in a variety of diseases (3,4), a better understanding of the dynamics of normal CNS is required. As several recent reports have mdicated, magnetic resonance (MR) imaging allows detection and measurement of motion throughout the CNS precisely and noninvasively by means

MATERIALS Index terms: Cerebrospinal 10.1214 #{149} Magnetic resonance study #{149} Magnetic resonance

PhD

ITHIN

more complete three-dimensional

hu-

M. Weisskoff,

#{149} Robert

Motion: MR

W

magnetic resothe authors

electrocardiography-gated,

acquired by using

MD

Wedeen,

Parenchyma Cine Echo-Planar

With echo-planar nance

MS

Massachusetts General Hospital, Bldg 1991; revision requested February 12, in part by NIH grant no. 5RO1HC3937, reprint requests to B.P.P. editorial

by Feinberg

(pp

630-632)

in

Abbreviations:

CNS

=

central

nervous

system,

CSF = cerebrospinal fluid, ECG = electrocardiography, FVZ = Fourier velocity zeugmatography, 3D = three-dimensional, 2D = two-dimensional.

645

1. Blipped echo-planar Instascan (Advanced NMR Systems) spin-echo pulse sequence. The readout gradient (G-Readout) is a resonant gradient producing a train of 64 echoes, one every 500 sec, applied in combination with the “blipped” gradient in the phaseencode direction, for a 2D single-shot acquisition (12,13). An auxilFigure

iary

motion-encoding

gradient,

of duration

8, is applied

along

either

the section-select direction (as shown) or one of the in-plane direclions. The total motion-encoding duration is + b. This velocityencoding gradient is applied either in a single polarity for each image (solid boldface line) or with alternating polarities for pairs of images (solid boldface and dashed lines). RF = radio frequency.

G-Readout

RJ.Jii A

a cardiac

delay

by 50 msec

progressively

per

image,

G-22LflJLfL

incremented

for 25 images.

Typi-

cally, the delay range spanned two full cardiac cycles. Depending on technique (see below), one or two image series were obtained, with either a single-polarity velocity encoding (Fig 1, solid boldface line) or an interleaved acquisition with alternating positive and negative polarities (Fig 1, solid boldface and dashed lines). The total data acquisition limes were from 20 seconds to 3 minutes. For studies of cephalo-

caudal

and anteroposterior

velocity,

the

subject was positioned supine in the head coil, with restraint to reduce head motion.

Due to system

limitations,

repositioned

subjects

in a 90#{176} lateral

position

for studies

were

Motion

observed

phase

in each

Cp1[x,yJ=

where

image:

pj[X,y]

-

p[x,yJ/N,

(1)

is the

ith phase image after correction, Pi[X,Y] S the ith phase image, and i = 1, . . N, where N is the total number of phase data collected over the car[X,y]

.,

diac cycle. ages 1[X,y]

Each of the

corrected

phase

im-

is purged of all phase terms that are constant over the image series (5,16). The simple methodology embodied

decubitus

of the mediolateral

velocity.

Equation (1) is applicable provided that the image phases do not wrap; more cornplex methods are required if phase wrap is in

Phase

Corrections

The phase

of the reconstructed

image

data reflects the velocity of the brain tissue at each pixel, but also contains additive phase errors of several types that must be

removed to isolate the velocity-dependent phase shifts. Two methods were investigated to remove such phase errors. (We shall abuse notation and refer to phase and velocity interchangeably, recaffing that phase p = K V for a constant of sensilivity K in each experiment [14].) .

Method

1 : phase

subtraction-Subtracting

the phase of pairs of images collected with velocity-encoding gradient pulses of opposite polarities cancels constant phase errors and isolates the motion-specific phase terms (15): (j = (p1f ‘P1’ where and p denote the velocity encodings of

_

the respective signs for the ith phase image and (j S the final phase-subtracted image. The phase-subtraction technique assumes that the brain motion is periodic, so that the same sequence of velocity dis-

tributions is played out in each heartbeat. It also assumes that the velocity-sensitizing gradient substantial Method

Unlike has

pulse does not itself cause phase errors. 2: cycle mean phase correction.-

blood

and CSF, solid brain

no net translation:

over one cardiac

cycle

Its average

must

tissue velocity

be zero.

This

principle allows us to correct any timeconstant phase errors, such as those of magnetic field variation or radio-frequency phase dispersion. At each pixel, compute the mean phase over the cardiac cycle and subtract this mean from the

we 646

Radiology

#{149}

present.

Data Processing: Compensation To identify

the motion

five to the skull, skull

Head

Motion of the brain

the rigid

is measured

and

motion

rela-

of the

subtracted

from

the

parenchyma velocity maps. A vector component of the motion of a rigid body

brain

(translation

plus

function L1[x,y]:

of the L1[x,yJ

where

the rates

x,y

rotation) image

=Ax

must be a linear coordinates,

+ By

of translation

+ V0,

Encoding

Gradients

of the ith lime point. This procedure removes the instantaneous contribution of rigid head motion on a frame-by-frame basis. This is not a substitute for immobilization of the head, since internal brain motions (“sloshing”) may be induced by rigid accelerations, particularly rotation, of the head. This procedure also corrects echo miscentering in k space, accomplished in other cases with the aid of stationary reference phantoms. The effect of this head motion compensation on observed parenchymal motion is illustrated

for

two

different

regions

(Fig

2).

Cephalocaudal velocity curves are shown for a frontal pole region (Fig 2a) and for a deep gray region (Fig 2b). The curves represent the computed, rigid motion at the selected locations, as well as the brain yelocity before and after rigid motion correclion. Cerebral periphery velocities (Fig 2a) are seen to be correlated strongly to the head motion and are reduced substanlially after correction. By contrast, the yelocities of the central structures appear less strongly correlated with the rigid head motion.

(2)

RESULTS

and rotation

determine the constant V0 and the rotalion rate determines the constants A and B. The values of V0, A, and B could be fixed by determining the value of V at three locations, but it is preferable from a signalto-noise standpoint to measure V, at a

This

study

volunteers years).

examined

of both Informed

14 healthy

sexes

(aged

consent

was

20-40 ob-

large set of points and to solve for the coefficientrmn Equation (2) by a linear least-

tamed according to the Subcommittee on Human Studies guidelines at our institution. Single or multiple axial and/or coronal images were obtained.

squares

In total,

fit. In practice,

we have typically 150 points in the scalp, which we assume moves with the skull and which furnishes a readily identifled MR signal. For adequate skin signal, we used four R-R intervals for repetition

measured

V at

and

about

cording

L1[x,y],

locity

at each

where and

point:

V[x,y]

=

V[x,y]

V[x,y] is the corrected yeV1[x,y] is the observed velocity

series

to section

were

collected

in Table

1 ac-

orientation,

section

acquisition mode, and velocity ponent measured (cephalocaudal, anteropostenor, or mediolateral).

time and 1.0 cm for section thickness. To find the rigid motion corrected velocity for each lime frame, a linear-fit function is subtracted from the observed velocity

84 cine

are summarized

Brain -

Motion

com-

Observations

Figure 3 shows a typical, throughplane motion study. By using the phase-subtraction

technique,

a cine

December

1992

Cephalic

Cephalic

I

At 0

(C

a) C’,

a)

E E

E E

>‘

>‘

U)

0

(C

0 a,

0

a)

>

-0.5

>

-

\.

.

Caudal

‘I

#{149}

0

.

Ii

“1’

-1

100

200

300 Delay

.

from

400 A-wave

500

600

Caudal

700

0

100

200

(msec)

300

400

500

Delay from R-wave

a.

600

700

(msec)

b.

Figure

2. Head motion (b) in a deep gray region.

(El) head

motion

correction For each

correction,

and

at the region,

the local

periphery and center. Time three velocity-time curves

head

motion

(U) derived

curves for correspond

from

cephalocaudal to the measured

the linear

velocities measured (a) in the parenchymal velocity, with

fit, computed

at these

I

frontal brain and (0) and without

locations.

Figure

3.

Phase

subtraction

cine study

level man.

of the basal A magnitude

vals/90

ganglia image

[repetition

time

indicate

ECG

gate

at the

in a 37-year-old (four R-R intermsec/echo

time])

at top left, followed by 14 images caudal velocity covering a cardiac times

of ye-

head-motion-corrected, cephalocaudal locity imaged in a transverse plane

delays

is

of cephalocycle. from

the

The R

wave. The velocities are shown in gray-scale representation, ranging over ±2 mm/sec. Darker shades correspond to caudal velocities,

and

zero

velocity

lighter

shades,

is medium

to cephalic

velocities;

gray.

image frames

at top left; the following 15 are the velocity maps for one full cardiac cycle, with velocity rendered in gray scale. Figure 4 shows the cephalocaudal velocity in a coronal plane through the fourth ventricle and brain stem (in-plane motion study). The magnitude image is at left, and two frames of the cine images are shown measured at gate delays of 0 and 75 msec after the ECG R wave. Figure 5 shows a study of a healthy data set of 25 images was obtained for cephalocaudal velocities in a transverse plane at the level of basal gan-

performed cine data des, with

glia,

50 msec.

Vnliime

and

rigid ic

motion

#{149} Number

correction

I

was

as described set covered a progressive Figure

3 shows

two

above. The cardiac cygate delay of a magnitude

volunteer

in which

the

nal velocity components consecutively in transverse The anteroposterior and

three

orthogo-

were imaged planes. cephalocau-

Radin1nv

#{149} 1s47

dal

components

were

measured

in an

Head

identical transverse plane 1.5 cm above the orbitomeatal line; the mediolateral component was imaged at a similar but not identical location coyenng the thalami. Figure 5a shows magnitude and velocity images at ECG

gate

delays

of 50 and

100

tion

Motion

occurs

biphasic

Scalp motion, assumed to be directly equal to head motion, was evaluated in a healthy volunteer and was found to be highly reproducible. Qualitatively, most of this head mo-

in systole cephalocaudal

and

consists

of a

displacement

followed by quiescence during diastole (see Fig 2a). To evaluate intercycle variability, studies of multiple heartbeats were same individual.

compared In a typical

in the case,

six

msec,

in all three motional axes. Figure 5b shows the three corresponding, complete velocity-time curves for a region of interest in the thalamus. In summary, brain motion appears to consist of a single displacement in systole

the

followed

initial

by

a slow

configuration

This displacement of the midbrain

and

return

to

in diastole. includes brain

a descent stem to-

ward the foramen magnum, with yelocities increasing with proximity to the foramen ( 2 mm/sec) and medial compression of the thalami on the third ventricle ( 1.5 mm/sec). Table 2 summarizes our measurements of maximum cephalocaudal, mediolateral, and anteropostenor velocities for six subjects transverse

on selected plane through

regions the

Figure 4. Phase-subtraction cine coronal study of head-motion-corrected cephalocaudal yelocities (coronal plane) through the quadrigeminal plate cistern and the brain stem in a 25year-old woman. At left is the T2-weighted magnitude image (four R-R intervals/90), and at right are two images of the cephalocaudal velocity, corresponding to delays of 0 and 100 msec after the ECG R wave. These data were acquired with motion encoding in the phase-encode direction. Velocity gray scale ranges over ±2 mm/sec. darker shades being caudal and lighter

of a lateral

and third ventricles (Fig 6). Cornputed net pulsatile displacements of the brain parenchyma are not greater than 0.5 mm. The axial motion of the midbrain is subtly asymmetric in the anteroposterior direction, with iniliation with respect to systole occurring more posteriorly. The motion of the cerebral periphery is relatively small. Anteroposterior motions tend also to be small in the forebrain, no more than 25% of the observed cephalocaudal and lateral motion, but may be of larger magnitude in the brain stem.

Cephalic Anterior Right

shades

being

foramen

cephalic.

Note

the

increase

in caudal

yelocities

(darker)

with

proximity

to the

magnum.

-

Caudal Posterior Left

-

0

100

200

-

300 Delay

400 From

-..-

Cephalo-Caudal

-

Antero-Posterior

._;;_

Left-Right

---

A-wave

500

600

700

800

(msec)

b. a. Figure 5. (a) Phase-subtraction cine study (four R-R intervals/90) of head motion-corrected cephalocaudal, anteroposterior, and lateral yelocities (transverse plane) through the third ventricle and the thalami in a 33-year-old man. Anatomic images (center) and velocity maps (right) corresponding to 50 and 100 msec after the R wave are shown for cephalocaudal motion (top), anteroposterior motion (middle), and mediolateral motion (bottom). The mediolateral measurement required 90#{176} rotation of the head, approximating the prior section position. Note the strong negalive cephalocaudal velocities, indicating caudal displacement of the midbrain, as well as the strong opposite mediolateral yelocities in the thaIami, associated with a medial compression of the thalami on the third ventricle. Ant = anterior, caud = caudal, cepli = cephalic, post = posterior. (b) Three corresponding complete velocity-time curves for a region of interest in the left thalamus indicated on the magnitude images. MR

Radinlnv

#{149}

December

1992

cine

series

were

of cephalocaudal

collected

and

ations of series were computed; to 0.3 mm/sec.

and

velocity

the standard

devi-

of equal-delay images they varied from 0.1 depending on time

tion cine ure 3.

location.

with

Comparison To verify native ment, FVZ

one

=

our

velocity

data,

shown

sional velocity gies may omit CNS

results

with

dimension)

in Fig-

DISCUSSION Observations Implications

technique,

as ([7] with

dal velocities were measured at six different cardiac phases (0-500-msec cardiac delays from the R wave) in the same volunteer and at the same section as that in Figure 3. Figure 7b shows multiple velocity profiles across the section, extracted from our 3D Fourier-transformed data set collected during early systole (see Fig 7

This

prior

study

of Brain Motion for Imaging confirms

MR imaging

several observed

points:

and

and

observations First,

the

on

Phase

6.

T2-weighted

MR

image

(four R-R for Ta-

area gray callo-

of

are observed

for CSF

venous

difference

flow

flow

is confirmed

and

in the

more

arterioneck

(6).

high of brain If this generally,

MR imaging not being for accurate measurement persons. Our echo-planar

finding

is at variance

preliminary findings Mark (7); it suggests

with

us to identify the of brain motion.

subtraction

in

the

of Feinberg and that one-dimen-

niques lional

and

represent

cycle-mean-

alternative

tech-

for the removal of nonmophase terms. Two contrasts

between

these

techniques

are

of basic

artifacts. The phase alternation method should remain accurate in these conditions. Thus, the cycle-

measurement, however, may be valuable in the study of transient, experimentally induced perturbation of brain velocity by mechanical or pharmaceutical means. The present study indicates that at least two vector components contribute substantially to the total motion, the cephalocaudal and the lateral. This

should

importance. First, the cycle-meansubtraction method is inaccurate where the temporal-mean velocity is nonzero, for it removes nonzero mean velocities along with the phase

it will validate the previous and future use of conventional ECG-gated multishot velocity MR imaging technique in the CNS and is suggestive of

Figure

time,

will help source”

subtraction

maximum

The study also demonstrates beat-to-beat reproducibility motion in healthy subjects.

intervals/80) shows regions selected ble 2. A = frontal lobe, B = anterior corpus callosum, Cl and C2 = deep matter, D = posterior area of corpus sum, E = occipital lobe.

measurement-encoding

ternthe

Technique

extends

(7). Second, the observed CNS motion is temporally simple, being essentially the velocities of a single monophasic displacement of the brain (in various directions) that occurs chiefly during cardiac systole, confirming previous results (7,9). Similar temporal patterns

single-shot essential healthy

of the

motion.

sampling “initiating

parenchymal velocities are on the order of 2-3 mm/sec. which is in agreement with Feinberg and Mark

finding

methodolofacets

allow for a finer sampling of the sequence of movements within the thalami, brain stem, and spinal cord. Finer

an alter-

others have done in conventional with 2D FVZ) or echo-planar ([8] 3D FVZ) imaging mode. Cephalocau-

imaging substantial

Future improvements in the poral resolution, by shortening

3D FVZ

method of velocity measurewe applied a 2D FVZ (3D two spatial dimensions and

velocity

legend for full technical description). Note the good correlation of this result with Figure 7c, representing the corresponding profiles derived from the first frame of the phase-subtrac-

mean-subtraction method would not detect the small, steady velocity of CSF drift. Conversely, phase alternalion requires an identical motion pattern in successive heartbeats (true periodicity), subtraction

whereas method

mean subtraction ful if the cardiac although a large may ased

the cycle-meandoes not. Thus,

should rhythm number

remain useis irregular, of images

be needed to guarantee sample. A mean phase

an may

unbialso

be measured from ungated data, vided the cardiac cycle is covered formly

and

densely

sense). Comparison

(in

of our

the

prouni-

Nyquist

2D phase-con-

trast rapid imaging method with FVZ suggests a few general points. FVZ suffers important limitations corn-

pared with presented.

the 2D instant FVZ is a more

suming procedure, measurements.

The

requiring velocity

techniques time-conmultiple spectrum

obtained with such a method provides an orderly way to organize the velocity distribution within a voxel and so to eliminate “partial voluming”

effects

between

multiple

velocity

components. (Similarly, it will spread velocity information among multiple levels in the case of cardiac variability or nonperiodic motion.) Unfortunately for brain this offers little

have pointed in the brain FVZ velocity results

from

attenuating with high Volume

185

Number

#{149}

3

parenchyma motion, interest. As others

out (8), proton diffusion introduces a blurring into spectra. Such blurring diffusion

progressively

the signal from images encoding gradients. ReferRadiology

#{149} 649

present methodology. For example, covering the brain with 20 sections, 10 gate delays, and three velocity cornponents requires 1,200 images. At 150

ring to the FVZ experiment described in Figure 7, velocity line broadening due to brain water apparent diffusion (10 cm2/sec) would be 0.4 mm/sec (kernel full width at half maximum). Deconvolulion methods to reduce blurring have been proposed (8).

msec only

versus

Total

Motion

tudes

components.

of these

effects

are

The

magni-

not

known,

however, and may influence producibility and interpretation brain motion data. The methodology of linear scribed

here

the

computed

this

possibility,

may

introduce

images. the

the

facts in MR signals from

memory

only

by

space

of

imaging single

to disperse voxels will

fit deinto

signal

loss

affect

measurement

fusion

due

to brain

constant.

knowledge

To investigate

lion

noninvasively.

Further

imaging

studies of CNS kinematics with brain motion-sensitive MR imaging will be needed to evaluate its ability to define normal and abnormal states, such as normal pressure hydrocephalus

the result

motion

of the model

dif-

increased

features

motion,

adequate

can

of the water Despite

parenchymal

fit model

parameters were evaluated by means of linear regression. With the optimal acquisition parameters indicated in Materials and Methods, the standard deviation, a, of the difference between the data and the linear fit was mm/sec. The standard error of the model varies with position, being lowest near the centroid of the sample points. By assuming the data deviations are distributed normally, the standard error of the linear regression would be approximately a/ N near the centroid of the sample points (central brain) and range up to a / 2/N peripherally (17), where N is the least-

of brain

we

with

still

which

lack

an

to detera.

a -T------

0.11

squares points.

fit total By using

standard

error

number

sec

throughout

smaller

than

variability

I 3

the



-3 ‘t 3 2 1

of data

typical



-4 C

-1

--

1____ -2 -3

is 0.01-0.02 mm/ the brain, much

of 0.1-0.3

b L____-

N = 150 points, the of the linear model in

these experiments

motion

of appar-

in false signal attenuation and quantitative errors in, for example, relaxation time measurements. In addition,

reof

noise

linear

image

prototype instrument. We believe that small head and brain motions are probable contaminants of most MR imaging procedures. Such motions add to the incoherent noise present on all images and, in some cases, may be a limiting factor in applications where critical evaluation of signal intensities is required. The tendency for motion arli-

stead, to be influenced strongly by the mechanics of the neck. As noted, mathematical correction of rigid movement does not remove all the effects of an external impulse on the brain, which may include both rigid

plastic

set requires lime. Such

is precluded

limited

such

ent diffusion coefficient. This will be the object of further investigations for modeling and simulation. High-speed MR imaging has a unique capacity to image brain mo-

our

We corrected for the rigid head motion because we assume this motion to reveal little about the CNS, but in-

and

this data of imaging

an acquisition the

Relative

per image, 3 minutes

mine quantitatively how will affect the measurement

beat-to-beat

mm/sec.

CONCLUSION The present investigation prompted by our belief that standing of CNS kinematics

advanced

by an approach

was an underwould be

that

in-

creases data rates, allowing large sets of the full position-lime-velocity

sub-

matrix

effiIn

ciently

the image

of the brain to be sampled and presented coherently. present framework, a complete of CNS

motion

would

vector with one spatial coordinates.

temporal While

such

data

are

herein,

their

acquisition

650

#{149} Radiology

presented

is feasible

with

Figure

7.

Comparison

(a) T2-weighted pared in b and 3D echo-planar

the

a

sensitized repetition

of cephalocaudal

2D cine

MR image (four R-R intervals/80) c. Lateral to the head are static FVZ

tion of multiple

specify

3D velocity and three

not

b.

method.

2D time

to motion time was

data,

Velocity

profiles

stepped

through

along the section-select four R-R intervals, and

were collected by using a 3-msec cardiac well with the phase-subtraction profiles by using velocities

the 50-msec as positive

frame from and caudal

the data velocities

phase-subtraction shows the location

phantoms. are

obtained

32 gradient

direction. echo time

delay shown

(b) Velocity by means

steps

(±1.0

profiles

obtained

of Fourier

G/cm

2D FVZ. profiles corn-

with

a

transforma-

1±0.0001

T/cm]),

Velocity resolution was 0.5 mm/sec. was 170 msec. Velocity profiles shown the R wave. These FVZ profiles compare

after in c. The

set shown as negative

studies with of velocity

phase-sensitive

in Figure 2. Both (in millimeters

profiles

were

b and c show per second).

derived

cephalic

December

1992

(18,19)

and

nal cord

diseases

involving

the

plexuses as the generator of the force for flow of the fluid and ventricular enlargement. J Neurosurg 1962; 19:405-413.

spi-

(20,21). 2.

APPENDIX In the presence with

zero

of any gradient

net moment,

I: where

G(t)

=

that

G(t)dt

0,

pulse

4.

KE, et al. Marked cerebrospinal fluid void: indicator of successful shunt in patients with suspected normal-pressure hydrocephalus. Radiology 1991; 178:459-466. Hennig J, Ott D, Adam T, Friedburg H.

+ -1 fT[ft

v(t’)dt’}}G(t)dt

lion (Al),

That

and

=

--YfoT

[ft

=

-1

w(t)v(t)dt.

IT

Jo

is, the phase

weighted dient-encoding

by using

Equa-

G(t’)dt’]v(t)dt

7. on the

velocity

during

period.

R, Greitz of pulsatile

imaging

discussions

David shy,

sion

with

Mikulis,

Gilberto

Gonzalez,

MD. We thank

Robert

the gra-

8.

Enzmann surement trast cine

grateful

for the support

MD,

9.

and

McKin-

an initial especially ofThomas J. Brady,

ver-

and

Bering

EA.

Circulation

of the cerebrospi-

nal fluid: demonstration

185

#{149} Number

of the choroid

3

using

DE, Ross MR. Pelc NJ. Meaof CSF flow using a phase conMR pulse sequence (abstr). In:

fluid

DA, Jakab studied

line scan, and echo-planar

Magn

Reson

Med

1990;

(abstr).

In: Book

Weisskoff

Reson

10.

Poncelet

Cohen

BP, Wedeen

MS, Brady motion

with

VJ, Weisskoff

TJ.

14.

the human heart using a new, whole-body MR imaging system. AJR 1987; 149:245-250. Moran P. A flow velocity zeugmatographic interlace for NMR imaging in humans. Magn Reson Imaging 1982; 1:197-

IL.

15.

203. O’Donnell M. using multiecho,

NMR blood phase

flow imaging

contrast

sequence.

Med Phys 1985; 12:59-64. Wedeen VJ, Crawley AP, Weisskoff Holmvang

C, Cohen

MS.

RM,

Real-time

MR

imaging of structured fluid flow (abstr). In: Book of abstracts: ninth annual meeting of the Society of Magnetic Resonance in Medicine, 1990. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1990; 164.

18.

20.

Bevington

P.

Data

reduction

and

error

analysis for the physical sciences. New York: McGraw-Hill, 1969. Bradley WG, Atkinson DJ, Nitz WR, et al. Quantitative cerebrospinal fluid velocity imaging: comparison of normals and patients with shunt-responsive normal pressure hydrocephalus. JMRI 1992; 2(P):59. HennigJ, Wahkloo AK, Jungling FD. MR interferographic examination of brain motion. JMRI 1992; 2(P):108. Levy LM, Di Chiro C, McCullough DC,

Dwyer AJ, Johnson DL, Yang SSL. Fixed spinal cord: diagnosis with MR imaging. Radiology 1988; 169:773-778. 21.

Levy

LM,

Wang

H, Carson

B, Schellinger

D, Di Chiro G, Bryan RN. Dynamic MR imaging for Chiari malformations (abstr). Radiology

1991;

181(P):172.

RM,

Measurement

EPI (abstr).

RR, Pykett

Ultra-fast im1991; 9:1-37. Instant images of

Rzedzian

imag-

tenth

RM.

Imaging

13.

16:280-293.

of abstracts:

855.

MS,

19.

annual meeting of the Society of Magnetic Resonance in Medicine, 1991. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1991; 44.

1991;

Magn

circulation

PD. Tissue perfusion by Fourier velocity dis-

in Medicine,

aging.

17.

imaging.

Soci-

Cohen

16.

HennigJ, Wahkloo AK, Koch D, Laubenberger J. The examination of ECG-dependent brain motion with MR-interferogra-

brain

Volume

motion

of the

12.

(abstr).

cerebrospinal

Feinberg in humans

phy MD.

References 1.

Mea-

Book of abstracts: tenth annual meeting of the Society of Magnetic Resonance in Medicine, 1991. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1991; 155. Feinberg DA, Mark AS. Human brain mo-

ing.

helpful

for his help in providing of a pulse sequence. We are MD,

brain

M.

meeting

Resonance in Medicine, Calif: Society of Magnetic

Stejskal EO. Use of spin echoes in a pulsed magnetic-field gradient to study anisotropic, restricted diffusion and flow. Chem Phys 1965; 43:3597-3603.

an inter-

demonstrated with MR velocity Radiology 1987; 163:793-799.

N

We appreciate

Resonance

C, St#{228}h-

F, Stubgaard

MR phase

tribution, Acknowledgments:

using

D, Thomsen

surement

tion

depends

average

flow

annual

11.

In: Book of abstracts: ninth annual meeting of the Society of Magnetic Resonance in Medicine, 1990. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1990; 1107.

v(t’)dt’]G(t)dt.

by parts,

Wirestam

AR, Kortman

MR technique based on the imaging sequence. Magn Re1990; 8:543-536.

lebrg F, Nordell

6.

integrating

Whittermore

of CSF

ferographic RARE-FAST son Imaging

,yfTx#{216}G(t)dt

=

WG,

Measurement

5.

[ft

+

{0

Bradley

amplitude,

position, v(t) = velocity, and _)‘ = gyromagnetic constant, the full expression of the phase shift of a moving spin is given by

fT

3.

(Al)

xo = initial

‘p = -y

Bostick on pulfluid 13:497-

tenth

ety of Magnetic 1991. Berkeley,

523.

is,

=

gradient

pulse

Du Boulay G, O’ConnellJ, CurrieJ, T, Verity P. Further investigations satile movements in the cerebrospinal pathways. Acta Radiol Diagn 1972;

abstracts:

In: Book

of of

Radiology

#{149} 651

Brain parenchyma motion: measurement with cine echo-planar MR imaging.

With echo-planar magnetic resonance (MR) imaging, the authors measured the intrinsic pulsatile motion of brain parenchyma. Phase-sensitive, electrocar...
1MB Sizes 0 Downloads 0 Views