Controversies and Complications in the Perioperative Management of Transcatheter Aortic Valve Replacement Andrew A. Klein, MBBS, FRCA,* Nikolas J. Skubas, MD, FASE,† and Joerg Ender, MD, PhD‡ Transcatheter aortic valve replacement (TAVR) is performed with increasing frequency in the United States since Food and Drug Administration approval in 2011. The procedure involves the replacement of a severely stenosed native or bioprosthetic aortic valve with a specially constructed valvular prosthesis that is mounted onto a stent, without the use of cardiopulmonary bypass and the complications of a major open surgical procedure. TAVR has been performed mostly in elderly patients with multiple comorbidities or who have undergone previous cardiac surgery. The most commonly used access routes are the femoral artery (transfemoral) or the cardiac apex (transapical), but the transaortic and transubclavian approaches are also used with varying frequency. Conscious sedation may be used in patients undergoing transfemoral TAVR, but the use of general anesthesia has not been shown to carry greater risk and permits the use of transesophageal echocardiography to assist in valve positioning and diagnose complications. Cardiovascular instability during TAVR is relatively common, necessitating invasive monitoring and frequent use of vasoactive medications. Complications of the procedure are still relatively common and the most frequent is vascular injury to the access sites or the aorta. Cardiovascular collapse may be the result of major hemorrhage pericardial effusion with tamponade or coronary occlusion due to incorrect valve placement. Persistent hypotension, myocardial stunning, or injury requiring open surgical intervention may necessitate the use of cardiopulmonary bypass, the facilities for which should always be immediately available. Ongoing and planned trials comparing conventional surgery with TAVR in lower risk and younger patients should determine the place of TAVR in the medium- to long-term future.  (Anesth Analg 2014;119:784–98)

S

ince the first transcatheter aortic valve replacement (TAVR) was performed in 2001, there has been an explosion of interest in the procedure, and it has captured the attention of the medical profession and the public alike.1 After initial introduction into mainstream practice in Europe, TAVR is performed with increasing frequency in the United States since Food and Drug Administration approval was granted in 2011. The procedure involves the replacement of a severely stenosed native or bioprosthetic aortic valve (AV) with a catheter-inserted, specially constructed valvular prosthesis that is mounted onto a stent. A TAVR is performed without the use of cardiopulmonary bypass (CPB) and the complications of a major open surgical procedure. TAVR is not currently indicated for aortic regurgitation because the valve stent is not sutured as in surgical AV replacement (AVR), but anchored inside the calcified native aortic annulus or the in situ prosthetic stent From the *Department of Anaesthesia, Papworth Hospital, Cambridge, United Kingdom; †Department of Anesthesiology, Weill Cornell Medical College, New York, New York; and ‡Department of Anesthesiology and Intensive Care Medicine, Herzzentrum, Universität Leipzig, Leipzig, Germany. Accepted for publication June 19, 2014. Funding: None. The authors declare no conflicts of interest. Supplemental digital content is available for this article. Direct URL citations appear in the printed text and are provided in the HTML and PDF versions of this article on the journal’s website (www.anesthesia-analgesia.org). Reprints will not be available from the authors. Address correspondence to Andrew A. Klein, MBBS, FRCA, Department of Anaesthesia, Papworth Hospital, Papworth Everard, Cambridge, CB 23 3RE, United Kingdom. Address e-mail to [email protected]. Copyright © 2014 International Anesthesia Research Society DOI: 10.1213/ANE.0000000000000400

784    www.anesthesia-analgesia.org

(the latter is referred to as a “valve-in-valve” procedure). However, a number of centers have reported undertaking TAVR off-license for aortic regurgitation when no other treatment modalities were thought to be possible.2 The majority of published experience is with 2 valves: the balloon-expandable Edwards SAPIEN (Edwards Lifesciences, Irvine, CA) and the self-expanding Medtronic CoreValve (Medtronic CoreValve Revalving System, Minneapolis, MI). A number of other valves are currently being trialled.3 TAVR is most often performed jointly by cardiologists and cardiac surgeons, although in Europe, cardiologists may undertake it alone. The first multicenter, prospectively randomized trial (Placement of AoRTic TranNscathetER Valve [PARTNER]) showed a marked and significant improvement in terms of mortality associated with TAVR, compared with medical treatment alone, in high-risk aortic stenosis patients judged not to be suitable candidates for surgical AVR.4 A subsequent randomized controlled study showed that TAVR and conventional surgical AVR were associated with a similar mortality at 1 and 2 years in high-risk surgical candidates.5 However, important differences in morbidity were observed; surgical AVR was associated with increased bleeding and related complications, and stroke was more common after TAVR.5 Similarly, 30-day and 1-year all-cause mortality were similar among intermediate-risk surgical patients (Society of Thoracic Surgeons [STS] risk score between 3% and 8%) with symptomatic severe aortic stenosis, who were propensity scorematched between surgical and TAVR treatment.6 In a recently published trial in the United States, high-risk patients randomized to TAVR, using the CoreValve, had a significantly higher rate of survival after 1 year (absolute reduction in risk 4.9%, October 2014 • Volume 119 • Number 4

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

P < 0.001 for noninferiority and P = 0.04 for superiority compared to surgical AVR).7 However, TAVR is associated with a relatively high incidence of controversies and major complications. Indeed, these may be more frequent than those associated with conventional surgical AVR performed via sternotomy, and particular care with regards to anesthetic care and perioperative management is required.8,9 Despite a relative lack of evidence of superiority of TAVR over surgical AVR, a greater number of patients and their clinicians are keen to explore or perform the less-invasive TAVR procedure, hence the importance of understanding the particular problems and challenges associated with it.10

Development of TAVR

The first percutaneous procedure performed on the AV was balloon aortic valvuloplasty, which was developed in the 1980s. This technique, however, did not improve valve area greatly, though modest but short-lived hemodynamic and symptomatic improvement was reported.11 Thereafter, the concept of balloon-expandable and self-expandable transcatheter valve implantation was introduced, and in 2001, Cribier et al.12 implanted the first transcatheter AV in Paris, France, in a patient with acute cardiogenic shock as a treatment of last resort. The TAVR procedure was designed to offer a less invasive alternative for patients who were deemed to be too high-risk for conventional surgical AVR. Over the last 12 years, improvements in the design of the valve and stent in which it is mounted, along with advances in valve delivery technology, have made TAVR much easier and even less invasive to perform. Recently, the delivery systems’ sizes have been greatly reduced; initially they were 22-F to 25-F (9 to 10 mm diameter), which meant the incidence of arterial injury was relatively high. Newer low-profile systems are compatible with 16-F to 19-F sheaths (outer diameter 6 or 7 mm), enabling their use in more patients.

Overview of the Procedure

The 2 TAVR systems in mainstream clinical use differ in construction. The Edwards SAPIEN has a cobalt chromium alloy tubular frame with leaflets from bovine pericardium, whereas the CoreValve has a nitinol frame with leaflets from porcine pericardium. They also differ in method of deployment. The Edwards SAPIEN valve is balloon-expanded, whereas the CoreValve is self-expanding. Both valves are most commonly inserted in a retrograde transarterial fashion

via the femoral artery (transfemoral [TF] approach). Severe peripheral arterial disease, including extreme tortuosity, marked calcification, narrowing (diameter less than 5.5 mm with current delivery sheaths), or aortic disease including aneurysm and thrombus, may prohibit the TF approach. In such cases, another suitable vessel, such as subclavian, axillary, or ascending aorta are used (the latter via a mini-sternotomy) for retrograde insertion or an antegrade transapical (TA) approach may be considered (via a left anterior intercostal mini-thoracotomy).8 In the TA-TAVR, the left ventricular (LV) cavity is punctured with a needle lateral to the apex and the ventriculotomy is then dilated serially to accommodate the valve deployment sheath. After the procedure has been completed, the LV apex is repaired with preinserted pledget sutures. A number of other access routes may also be used (Table 1) depending on operator experience. Balloon aortic valvuloplasty is usually performed before valve insertion to facilitate passage of the prosthesis through the stenotic native AV. In the case of a prosthetic valve stenosis (valve-in-valve procedure), this is not required. With the balloon-expandable Edwards SAPIEN valve, rapid ventricular pacing is used to decrease arterial blood pressure and transvalvular flow, reducing the risk of movement of the valve into the aorta during deployment. The longer profile of the CoreValve, which extends from the aortic annulus to the supracoronary aorta, allows for the gradual release of the prosthesis without the need for pacing.13

Patient Selection

The use of TAVR is geared toward patients with severe aortic stenosis, who are considered to be at substantially increased risk for conventional surgical AVR. These include patients who require reoperative sternotomy with potential damage to patent coronary arterial grafts, and those with multiple comorbidities, such as respiratory or vascular disease. These patients are deemed to be high risk based on a STS risk score >10% or logistic EuroSCORE >20%. Of note, these 2 risk stratification schemes are not equivalent, with the STS risk score considered more representative of highrisk AV surgery.14 As a result, it is not straightforward to compare the patient cohorts among the various TAVR registries. The most recent American College of Cardiology/ American Heart Association Valvular Heart Disease Guidelines recommend surgical AVR if the predicted risk of mortality is below 8%, and TAVR is reserved for those elder, frail patients with significant morbidity who are considered

Table 1.  Transcatheter Aortic Valve Replacement—Approach Prosthesis Access Anesthetic type Echocardiography Postoperative disposition

Retrograde Transaortic/subclavian Sapien or CoreValve Aorta (mini-sternotomy) or subclavian artery Sedation (MAC) or general General anesthetic anesthetic Transthoracic Transesophageal Variable, institution Intensive care unit dependent: step-down, (usually) monitored (telemetry) setting

Transfemoral Sapien or CoreValve Femoral artery

Antegrade Transapical Transcaval68 Sapien or CoreValve Sapien only Femoral vein into IVC then Cardiac apex abdominal aorta (mini-thoracotomy) General anesthetic General anesthetic (less often: thoracic epidural) Transesophageal Transesophageal Intensive care unit intensive care unit (usually)

MAC = monitored anesthesia care; IVC = inferior vena cava.

October 2014 • Volume 119 • Number 4

www.anesthesia-analgesia.org

785

E Review Article Table 2.  Summary of Evidence Comparing General Anesthesia and Monitored Anesthesia Care Author, year Guarracino, 201127

Access, prosthesis TAx, CV, or ES

N 22

Durand, 201224

TF, ES

151

Dehedin, 201126

TF, CV, or ES

125

GA (propofol or sevoflurane/ isoflurane/desflurane + opioid) = 91

Fassl, 200954

TA, ES

100

Covello, 201067

TF, Tscl, ES, or CV

69

GA (sevoflurane + remifentanil) GA (sevoflurane + remifentanil) = 27

Ree, 200868

TF, ES

40

Grube, 200769

TF, Tscl, CV

86

GA

Guinot, 201070

TF = 62, TA = 28, ES or CV

90

Covello, 200971

TF, ES

18

GA (propofol or sevoflurane/ isoflurane/desflurane + opioid) GA (sevoflurane + remifentanil) = 15

Mukherjee, 200930

TA, ES

1

Motloch, 201225

TF, ES

74

Bergmann, 201123

TF, ES, or CV

Petridis, 201229

TA, ES

Yamamoto, 201328

TF, ES, or CV

174

Oguri, 201431

TF, ES, or CV

2326

100

General anesthetic GA (propofol or sevoflurane + remifentanil) = 8

Other anesthetic Local infiltration (ropivacaine) + MAC (remifentanil) = 14 Local infiltration (lidocaine) + MAC (midazolam + nalbuphine) Local infiltration + block (iliohypogastric or ilio-inguinal with ropivacaine) + MAC (remifentanil)

Comments

Fluoroscopy instead of TEE 3 patients (larger sheath) converted to GA Similar outcome, TEE in GA, mini-TEE probe in some MAC CPB (1st–10th), off CPB (11th–100th)

Lidocaine + block (iliohypogastric or ilio-inguinal or superficial cervical plexus with ropivacaine) + sedation (remifentanil) = 42 (3 converted to GA) MAC = 4 (1st–4th)

GA = 36 (5th–40th)

Complications prompted change of technique Prospective, observational, comparing 18F and 21F

MAC in at least 9/36 patients with 18F sheath

GA (desflurane/isoflurane or propofol/etomidated + remifentanil) = 33 GA = 17 (converted from MAC due to complications)

1 GA (propofol + remifentanil) = 44 GA 1377 patients (technique not specified)

MAC (remifentanil + lidocaine) = 3 Thoracic epidural at T2-3 with ropivacaine+fentanyl MAC (midazolam + piritramide) + local (lidocaine) = 41 MAC (midazolam + remifentanil) + local (lidocaine) Thoracic epidural at T2-3 with bupivacaine+fentanyl MAC (remifentanil+propofol) + local (lidocaine) = 130 Local anesthetic +/- sedation (MAC) 949 patients

Inserted the day before Higher risk patients received MAC Conversion to GA due to procedural complications Inserted the day before 6 conversions to GA due to procedural complications Incidence of postprocedural aortic regurgitation≥mild was significantly lower in GA than in MAC (15.0% vs 19.1%; P = 0.015). No difference in 30-day survival.

CV = Medtronic CoreValve Revalving System (Medtronic, Minneapolis, MN); ES = Edwards Sapien (Edwards Lifesciences, Irvine, CA); TA = transapical; Tax = transaxillary; T-scl = transsubclavian; TF = transfemoral; GA = general anesthesia; MAC = monitored anesthesia care.

at high risk for surgical AVR (class of recommendation I and level of evidence B for prohibitive risk, class of recommendation IIa and level of evidence B for high surgical risk).15 Multidisciplinary patient consideration is considered to be ideal,16 and the involvement of anesthesiologists early in the process of patient selection is imperative. Successful practice requires careful preoperative evaluation by a surgeon, cardiologist, and anesthesiologist to allow for the correct patient selection and delivery of tailored, patient-focused care. Concurrently, there should also be consideration as to whether sternotomy should be performed if a catastrophic event occurs; this will depend on the underlying morbidities and whether the patient has undergone previous surgery.17 Agreeing on an emergency plan in advance is very helpful in the event of a major complication, and this should ideally be documented in the medical notes.

786    www.anesthesia-analgesia.org

Planning for TAVR

Even when a percutaneous approach to the femoral artery is planned, the potential for serious complications that may result in the rapid deterioration of the patient must be respected. A surgical team, including a perfusionist and a “dry” CPB setup must be on standby. Additional available equipment includes intra-aortic balloon pump, cell salvage,16 external defibrillator pads, a warming blanket, and pumps for infusion of IV medication. Communication among team members during the procedure is vital so that complications may be avoided or recognized promptly if they do occur. An interdisciplinary approach to TAVR by means of a heart team may facilitate alternative procedures with preplanned protocols and lead to acceptable outcomes, despite severe intraprocedural complications.18 For these reasons, a heart team approach is

anesthesia & analgesia

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

Table 3.  Anesthetic Techniques Approach Transfemoral

Anesthetic management General anesthesia • Propofol + remifentanil infusion26 • Propofol or volatile + opioid bolus or remifentanil infusion27,63

Transapical

General anesthesia • Sevoflurane + remifentanil infusion50 • Propofol or volatile + opioid bolus or remifentanil infusion27,65 General anesthesia • Propofol or sevoflurane + remifentanil infusion28

Transaxillary

Conscious sedation • Propofol + remifentanil infusion27 • Remifentanil infusion + midazolam bolus29 • Bolus midazolam + bolus nalbuphine23 Thoracic epidural • Bupivacaine + fentanyl infusion26 • Ropivacaine infusion29 Conscious sedation • Remifentanil infusion28

Table 4.  Echocardiographic Imaging During Transcatheter Aortic Valve Replacement Preprocedure (TTE more often than TEE)  Verify severity of aortic stenosis  Estimate number of aortic valve cusps, aortic annulus size, anatomy of aortic root  Evaluate thoracic aorta for atheromatous disease Intraoperatively (2D and 3D TEE)  Reassess the aortic annular diameter:   2D: annular anterior-posterior diameter in a long-axis view   3D: annular anterior-posterior diameter in a long-axis view and a short-axis view (multiplanar reconstruction mode)   All measurements performed in systole, at the hinge point of the leaflets.  Aortic annulus to coronary ostia distance:   2D: aortic annulus to right coronary ostium in the long-axis view   3D: aortic annulus to the right and left coronary ostia using mulitplanar reconstruction   (Both distances should be ≥11mm to avoid coronary occlusion during implantation)  Imaging of access site (transapical approach): slightly off the cardiac apex (ME 2 chamber view)  Evaluate the proper position of guidewires, balloon (for aortic valvuloplasty), and prosthetic valve stent  Imaging of prosthetic valve deployment and proper positioning  Evaluation of severity and location of trans- and paraprosthetic aortic regurgitation  Assessment of valvular hemodynamics:   Peak and mean gradients, valve area, Dimensionless velocity index  Assess baseline global and regional LV function and mitral regurgitation  Monitor for complications:   New regional wall motion abnormalities   Severity of mitral regurgitation   Pericardial effusion/tamponade   Aortic dissection   Coronary occlusion Postprocedure (TTE more often than TEE)  Assess function of the aortic prosthesis  Assess ventricular function and mitral regurgitation TEE = transesophageal echocardiography; TTE = transthoracic echocardiography; LV = left ventricular.

mandated by societal guidelines19 and, at least in the United States, is linked to reimbursement.20

General Anesthesia Versus Sedation

There are no randomized, prospective studies comparing general anesthesia (GA) to local anesthesia with infiltration of the access site and/or monitored anesthesia care (MAC) techniques (Table 2). The choice of the particular anesthetic technique is usually influenced by the patient’s comorbidities and institutional practices. Based on recent surveys, GA appears to be the preferred technique for TAVR irrespective of cannulation access in either European or North American institutions.21 Many authors published their experience with the use of local anesthetic for infiltration of the access site in TF-TAVR22–26 or even transaxillary TAVR,27 whereas some have performed an iliohypogastric and ilioinguinal block with local anesthetic,28 supplemented by IV sedation. One center reported the use of local infiltration and IV sedation by the operator, without the presence of an anesthesiologist.24 There are case reports of thoracic epidural techniques for

October 2014 • Volume 119 • Number 4

the TA approach29,30 (Table 3). A MAC technique may have some theoretical benefits, such as avoidance of the hemodynamic side effects of the anesthetic drugs, and therefore, decreased use of vasoactive drugs, along with the ability to monitor the central nervous system for embolic events, decreased cost, and faster patient recovery. On the other hand, the patient under GA is not at risk for airway compromise, which may require emergent intervention in the midst of an unfavorable hemodynamic environment during critical parts of the procedure. The patient remains immobile to facilitate the surgical procedural steps, and transesophageal echocardiography (TEE), which is of particular importance for the diagnosis of paraprosthetic leaks, can be performed. There is no evidence for a difference in outcome between sedation versus GA. A recent large observational study of the French Aortic National Registry in 2326 patients showed smilar device success and cumulative 30-day survival in GA compared with MAC (97.6% vs 97.0%; P = 0.41 and 91.6% vs 91.3%; P = 0.69, respectively), whereas the incidence of postprocedural aortic regurgitation ≥ mild was

www.anesthesia-analgesia.org

787

E Review Article

Figure 1. Native aortic valve with (A) aortic stenosis and (B) no aortic regurgitation (TEE, ME aortic valve LAX). Ao = aorta; LA = left atrium; LVOT = left ventricular outflow tract; TEE = transesophageal echocardiography; ME = midesophageal; LAX = long-axis.

another, and for inhaled or IV maintenance of anesthesia. Short-acting anesthetic drugs are preferred if tracheal extubation is planned at the end of the procedure, unless there is ongoing cardiovascular compromise or bleeding.

Transesophageal Echocardiography

Figure 2. Unfavorable septal anatomy for transcatheter aortic valve replacement. The basal septal hypertrophy (*) is abutting to, and extending at least 7 mm below the aortic annulus, making the seating of the transcatheter valve impossible (TEE, ME aortic valve LAX view). Ao = aorta; LA = left atrium; LVOT = left ventricular outflow tract; TEE = transesophageal echocardiography; ME = midesophageal; LAX, long axis.

significantly lower in GA than in MAC (15.0% vs 19.1%; P = 0.015).32 This may be as a result of greatly increased use of TEE in patients who underwent GA compared with MAC (76.3% vs 16.9%; P < 0.001). In TF-TAVR, sedation is considered a safe option,21 but emergency conversion to GA is often required (in 1 study, the conversion rate was 17 of 100 patients), mostly due to hemorrhage after vascular complications23 or procedure-related hemodynamic alterations due to cardiac tamponade, cardiac arrest, myocardial infarction, or stroke.25 Thus, it is reasonable to carefully consider the perioperative risks associated with the TAVR procedure when planning for the anesthetic, and if sedation is used, the presence of an anesthesiologist in the operating room is vital to administer suitable drugs and monitor the patient. In terms of anesthetic drugs for a GA technique, again high-quality evidence is lacking, and individual or institutional preferences are most frequently used (Table  3). There is no clear benefit in terms of one induction drug over

788    www.anesthesia-analgesia.org

For successful TAVR, high-quality imaging modalities are needed, and fluoroscopy and TEE have complementary roles. Fluoroscopy is better to assess the location of guidewires and catheters as well as the proper position of the valve stent, but at the expense of radiation exposure to the patient and personnel. TEE is considered essential during TAVR by many operators; however, some do not use it routinely (Table  4). Use of TEE necessitates GA, but some reported their experience with mini-TEE probes on patients under MAC.31 TEE is considered the most accurate (and rapid) tool32 for: (1) Confirming the preoperative diagnosis (Fig. 1). (2) Excluding unfavorable anatomy that may not allow for proper seating of the prosthesis across the aortic annulus (Fig. 2). (3) Verifying the choice of prosthesis size (23, 26, or 29 mm for SAPIEN; 23, 26, 29, or 31 mm for CoreValve), by remeasuring the aortic annulus diameter (Fig. 3). (4) Establishing the baseline right and LV function and preexisting valvular disease (especially the severity of mitral regurgitation), pericardial effusion, or AV/ aorta atheromas. (5) Guiding the introduction and positioning of guidewires (Figs. 4 and 5), balloon (if aortic valvuloplasty is performed), and valve stents (Fig. 6). (6) Verifying deployment of the prosthesis (Fig. 7). (7) Examining the success of the procedure (Fig. 8) and assessing the effective orifice area of the newly implanted aortic prosthesis. (8) Diagnosing complications, such as aortic regurgitation (Fig. 9), trauma to the aorta (Fig. 10) and neighboring structures,33 and pericardial effusion/ tamponade.

anesthesia & analgesia

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

Figure 3. A, Aortic annulus measurement in the ME aortic valve LAX view. B, Orthogonal plane measurements of the aortic annulus with (A) 2-dimensional TEE and (B) 3-dimensional multi-plane reconstruction. LA = left atrium; LV = left ventricle; LVOT = left ventricular outflow tract; Ao = aorta; TEE = transesophageal echocardiography; ME = midesophageal; LAX = long-axis.

Figure 4. Introduction of a guidewire (arrowheads) through the stenosed aortic valve (ME aortic valve LAX): notice the arrhythmia introduced. LA = left atrium; LVOT = left ventricular outflow tract; RV = right ventricle; Ao = aorta; ME = midesophageal; LAX = long-axis.

(9) Reassessing the hemodynamic environment, for example, evaluation of the ventricular function at baseline, after ventricular pacing and valve deployment.32 (10) Measuring the distance from the annulus to the coronary ostia, which may be useful to predict coronary occlusion from the TAVR stent34 (Video 1, Supplemental Digital Content 1, http://links.lww. com/AA/A983). The most recent European Association of Echocardiog­ raphy/American Society of Echocardiography guidelines35

October 2014 • Volume 119 • Number 4

recommend the use of TEE whenever there is doubt about the anatomy of the aortic root, aortic annulus size or number of AV cusps. There is considerable debate about which imaging modality is the “gold standard” for the evaluation of the aortic annulus and root at this time. Transthoracic echocardiography alone has been shown to underestimate the native annulus size by up to 15% or 1.36 mm (95% confidence interval, 1.75–4.48 mm).36 There is some evidence that 3-dimensional (3D) computed tomography (CT) as well as real-time 3D echocardiography may be more accurate.37 Three-dimensional TEE is considered more accurate in measuring the true aortic annulus diameter38 (via proper alignment of the short and long axes of the AV), and the distance of the coronary ostia from the aortic annulus (to avoid occlusion from the valve stent)34 and for the quick evaluation of the degree of (any) paraprosthetic aortic regurgitation after deployment. Adherence to sizing variables defined by cross-sectional 3D-TEE was associated with a lower incidence of paraprosthetic aortic regurgitation than conventional 2-dimensional TEE cutoffs and is advocated whenever good cross-sectional CT data are not available39 Multidetector CT measurements of the aortic annulus derive aortic annular diameter that is larger than the TEE-derived diameter by 1.5 ± 1.6 mm and may become the ideal technique for TAVR sizing.38 Aortic root assessment with cardiac magnetic resonance imaging may be a valid imaging alternative in patients unsuitable for CT.40 Transthoracic echocardiography is an alternative if sedation is chosen, but this may not allow such rapid diagnosis

www.anesthesia-analgesia.org

789

E Review Article

Figure 5. Entrapment of a guidewire (transapical approach for insertion of an Edward Sapien aortic valve). An echogenic guidewire (arrow) is entangled with the subvalvular mitral apparatus (A, B; transgastric long-axis views) resulting in severe mitral regurgitation (C, mid esophageal long-axis view). Withdrawal of the guidewire (B) results in decrease of mitral regurgitation (D, midesophageal long-axis view). LA = left atrium; LV = left ventricle.

because of difficulties with acquiring good quality echocardiographic windows during the procedure and with the patient in the supine position. Intracardiac echocardiography31 may be an option, but experience with this technique is limited.

Hemodynamic Monitoring and Management

Figure 6. Introduction of an Edwards Sapien valve (arrowheads) across the aortic valve via a transfemoral approach (TEE, mid esophageal aortic valve LAX view). LVOT = left ventricular outflow tract; Ao = aorta; LA = left atrium; TEE = transesophageal echocardiography; LAX = long axis.

Figure 7. Deployment of a balloon-expandable Edwards Sapien transcatheter aortic valve during rapid pacing (ME aortic valve LAX view). Ao, aorta; LA, left atrium; LVOT, left ventricular outflow tract; ME, midesophageal; LAX, long axis

790    www.anesthesia-analgesia.org

Invasive monitoring should be considered standard because cardiovascular compromise may be rapid and unpredictable. This should include invasive arterial and central venous pressures. The use of pulmonary artery catheters in this setting may be considered controversial by some, but may be particularly useful in patients with moderate or severe pulmonary hypertension.41 Many of the TAVR patients have baseline pulmonary hypertension. Despite the sustained decrease of pulmonary artery systolic pressure after surgical AVR or TAVR,42 the presence of pulmonary hypertension was found to be an independent risk factor for mortality in TAVR patients, irrespective of TF or another approach.43 In a small cohort, severe pulmonary hypertension (pulmonary artery systolic pressure > 60) had an odds ratio of 7.56 and right ventricular dysfunction had an odds ratio of 3.55 for all-cause mortality at 6 months after TAVR.44 Some have used normothermic femoro-femoral CPB electively in such patients and reported favorable outcomes.45 When performing balloon aortic valvuloplasty or during deployment of a balloon-expandable valve (Edwards Sapien), rapid ventricular pacing is used to arrest cardiac ejection. The heart may take some time to recover from this insult, especially if its function was significantly impaired before TAVR, or if rapid pacing was prolonged because of adjustments to balloon or valve position before inflation. This may necessitate bolus administration of phenylephrine or meteraminol, depending on institutional preference. In an emergent situation, epinephrine (10–20 mcg) may be injected into the aortic root via the pigtail catheter normally used for contrast administration to enable the drug to work more quickly, especially if the temporarily dysfunctional heart is hardly contracting. Other drugs that may be

anesthesia & analgesia

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

Figure 8. A, Successfully deployed Edwards Sapien transcatheter aortic valve, end-diastolic frames in ME aortic valve SAX views. B, Central, trace aortic regurgitation due to a guidewire and no paraprosthetic aortic regurgitation. LA = left atrium; RV = right ventricle; ME = midesophageal; SAX = short axis.

Figure 9. A, Paraprosthetic and central (wire-related) aortic regurgitation following implantation of Edwards Sapien valve (TEE, ME aortic valve LAX view). B, Three-dimensional echocardiography with multiplanar reconstruction showing a paraprosthetic aortic regurgitation following deployment of an Edward Sapien valve. Notice that the aortic regurgitation is not seen in the equivalent ME aortic valve LAX view (top right). LA = left atrium; RA = right atrium; LVOT = left ventricular outflow tract; TEE = transesophageal echocardiography; ME = midesophageal; LAX = long axis.

October 2014 • Volume 119 • Number 4

www.anesthesia-analgesia.org

791

E Review Article Vascular Injury

Figure 10. Posterior rupture of aortic root (pseudoaneurysm) following a transcatheter aortic valve deployment (TEE, ME aortic valve LAX view). Ao = aorta; LA = left atrium; TEE = transesophageal echocardiography; ME = midesophageal; LAX = long axis.

Video 1. Midesophageal 4-chamber view showing normal left ventricular (LV) function preimplantation; then long-axis aortic view showing calcified plaque in the left ventricular outflow tract and ascending aorta only a short distance from the coronary ostia. Following transcatheter aortic valve replacement (TAVR) implantation, regional wall abnormality (RWMA) can be seen affecting the left coronary (LCA) territory. In the long-axis view, the LCA can be seen to be occluded by the TAVR.

Assessing the minimum diameter of the arterial access vessel is essential for proper patient selection. A relatively compliant nondiseased artery can accommodate a vascular sheath that is slightly larger (1–2 mm) than its internal diameter. By general consensus, the TF-TAVR is the most popular approach and the femoral vessels are usually percutaneously accessed. In relatively small, calcified, or tortuous vessels, the risk of injury is significant; major vascular complications include dissection, rupture, or pseudoaneurysm. In patients deemed to be at increased risk of vascular injury due to anatomical considerations, surgical exposure may be used to facilitate surgical control and repair if required. The incidence of major vascular complications in the PARTNER A randomized controlled trial comparing TAVR and surgery was 11% in the TAVR group, compared with 3.2% in the surgical group (P < 0.001),5 but without a statistically significant difference in mortality (3.4% in the TAVR group and 6.5% in the surgical group, P = 0.07). In the first few years after the introduction of TAVR, the incidence of vascular complications was reported to be as high as 27%; vascular dissection, perforation, and hematoma at the access site were encountered in up to 15.3% of the patients and associated with significantly higher 30-day rates of complications, and 30-day and 1-year mortality.46–48 However, the size of the delivery systems has been reduced over the last 2 years to less than 5.5 mm outer diameter (depending on manufacturer and size of valve to be implanted) with a concurrently decreased incidence of access-related vascular damage to as low as 3.1% in a more recent European registry.22 Hayashida et al.50 reported a sheath:femoral artery ratio of >1.05 as predictive for major vascular complication and 30-day mortality in 102 patients undergoing TF-TAVR with either the CoreValve or the Edwards Sapien valve.49

Pericardial Hemorrhage

considered in this situation include phenylephrine, norepinephrine, or meteraminol. If the heart does not recover quickly (within 30–60 seconds), external cardiac massage should be initiated, and this may have the effect of pumping the inotrope and restoring cardiac function. However, if this is not successful, urgent consideration should be given to the institution of CPB. In most instances, this can be performed quickly by placement of peripheral venous and arterial cannulas over guidewires already in situ in the femoral vessels. In some cases, sternotomy may be required; preoperative discussion of the appropriateness of such an invasive intervention in an emergency situation is therefore essential.

Intrapericardial bleeding is possible at any point during the procedure due to injury to the heart or to the aorta, either as a result of wire perforation or annular rupture. Early detection of pericardial effusion is easier and earlier detected with echocardiography (Video 2, Supplemental Digital Content 2, http://links.lww.com/AA/A984; Video 3, Supplemental Digital Content 3, http://links.lww.com/AA/A985; Video 4, Supplemental Digital Content 4, http://links.lww. com/AA/A986). Life-threatening bleeding into the pericardium that may occur in up to 13% of patients during or immediately after TAVR is more common with a TA approach (odds risk 3.7, 95% confidence interval 1.73%–7.9%, P = 0.01) and is associated with significantly more frequent 1-year mortality (odds risk 2.54, 95% confidence interval 1.3%–4.9%, P = 0.002).50 Arterial bleeding from annular rupture will most likely require emergent sternotomy and surgical control. On the other hand, venous bleeding, which may occur as a result of injury from the pacing wire, may be managed by percutaneous pericardial drainage and close observation.

Complications

Problems with Conduction and Arrhythmias

The incidence of procedural complications depends on the access chosen for TAVR and on the type of implanted valve.

792    www.anesthesia-analgesia.org

The conduction system passes superficially through the interventricular septum immediately below the AV. After

anesthesia & analgesia

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

Video 2. Sweep of the probe from transgastric position showing posterior pericardial effusion after CoreValve transcatheter aortic valve replacement implantation to the midesophageal position showing compromised right atrium. LA, left atrium, LV, left ventricle.

implantation of a pacemaker. In the PARTNER trial cohorts (Edwards SAPIEN), the reported incidence of permanent pacemaker implantation in TAVR was only slightly higher than surgical AVR (5.7% vs 5%, P = 0.68)5 and lower than medical therapy (4.5% vs 7.8%, P = 0.27).4 Conductance disturbances leading to permanent pacemaker implantation are more common for the CoreValve than for the Edwards Sapien valve. The incidence for permanent pacemaker implantation ranges from 4.9%–6% for the Edwards Sapien valve to 23.4%–39% for the CoreValve.51–53 The CoreValve stent is longer and extends inside the LV outflow tract (LVOT) where it can compress the conduction system. A higher implantation technique for the CoreValve seems to decrease the risk for conductance disturbances.51 Left bundle-branch block with left axis deviation, an interventricular septal dimension >17 mm at end-diastole, or increased baseline thickness of the native noncoronary cusp AV (>8 mm) are considered predictors for the need of a new cardiac pacemaker.52 Conduction disturbances can occur up to 72 hours after implantation; therefore, patients should be closely monitored for up to 3 days after TAVR. However, based on 1-year follow-up from a single center, the periprocedural permanent pacemaker insertion does not seem to affect clinical outcomes adversely.53 Although the incidence of ongoing ventricular fibrillation after rapid ventricular pacing is very rare, external defibrillator paddles should be attached to every patient before the procedure so that defibrillation may be performed without delay after rapid ventricular pacing if the rhythm fails to normalize.54

Valve Malpositioning

Video 3. Discrete paravalvular leak due to annulus rupture after CoreValve transcatheter aortic valve replacement implantation.

Video 4. View from the left ventricular outflow tract to the aortic annulus. Three-dimensional color showing annulus rupture after CoreValve implantation.

conventional AVR, mechanical trauma, tissue edema, and local inflammation at the level of the aortic annulus and the neighboring atrioventricular node and bundle of His may induce conduction disturbances and increase the need for

October 2014 • Volume 119 • Number 4

During TAVR, there is a risk that the valve is incorrectly implanted into the aorta and the sealing cuff of the TAVR valve is not apposed to the aortic annular tissue; most commonly, the valve is placed somewhat too low (in the LVOT) or too high (in the aortic root) in the aorta. Exact positioning of the valve is important: if too low in the LVOT, it may impinge on the anterior mitral leaflet; if it is too high in the aorta, a coronary artery annulus may be blocked leading to myocardial ischemia and potential cardiovascular collapse (Video 1, Supplemental Digital Content 1, http://links.lww. com/AA/A983; Video 5, Supplemental Digital Content 5, http://links.lww.com/AA/A987; Video 6, Supplemental Digital Content 6, http://links.lww.com/AA/A988). Coronary ostial occlusion occurred in 44 of 6668 patients (0.66%) in a recently published large registry.55 If the latter occurs, coronary stenting may be attempted and can be lifesaving. Three-dimensional TEE may be the most accurate method of measuring the distance between the aortic annulus and the left main coronary ostium before implantation.56 The registry data reported that percutaneous coronary intervention was attempted in 75% of the cases and was successful in 82%; 30-day mortality was 41%. After a median follow-up of 12 (2 to 18) months, the cumulative mortality rate was 45.5%, and there were no cases of stent thrombosis or reintervention. Some may preemptively insert a stent inside the coronary ostia to avoid inadvertent obstruction from the valve prosthesis,57 or alternatively a wire can be placed before TAVR deployment to allow rapid stent placement in an emergent situation.

www.anesthesia-analgesia.org

793

E Review Article

Video 5. Preoperative midesophageal long-axis view showing plaque near to the right coronary ostium. LA, left atrium; LV, left ventricle; Ao, aorta.

Video 7. Angiography showing malposition of the valve due to a short period of ejection because of pacemaker malfunction during rapid pacing.

Video 6. Regional wall motion abnormality (RWMA) of the inferior wall due to obstruction of the right coronary ostium after valve deployment.

Video 8. Biplane view showing extensive calcium on the septum leading to aortic regurgitation after implantation (see also Video 9).

Incorrect positioning along with undersizing, incomplete expansion (Video 7, Supplemental Digital Content 7, http://links.lww.com/AA/A989), or asymmetric calcification (Video 8, Supplemental Digital Content 8, http:// links.lww.com/AA/A990; Video 9, Supplemental Digital Content 9, http://links.lww.com/AA/A991; Video 10, Supplemental Digital Content 10, http://links.lww.com/ AA/A992) may result in paraprosthetic aortic regurgitation. Intraoperative TEE echocardiography may identify calcification of the commissure between the right coronary and noncoronary cusps and the area cover index as independent predictors of significant paravalvular aortic regurgitation after TAVR.58 If significant aortic regurgitation is detected after valve implantation, either by fluoroscopy/aortography or TEE (Fig. 9, Video 11, Supplemental Digital Content 11, http://links.lww.com/AA/A993; Video 12, Supplemental Digital Content 12, http://links. lww.com/AA/A994), consideration should be given to a second balloon dilation of the prosthetic valve (only possible in the case of the Edwards Sapien valve). Such a procedure may cause further complications because it may lead to valve leaflet disruption, but if carefully performed, it may improve valve function and reduce paravalvular regurgitation.

794    www.anesthesia-analgesia.org

Video 9. Paravalvular aortic regurgitation at the site of preoperatively detected extensive calcium.

Moderate or severe residual aortic regurgitation is much more common than after surgical AVR, occurs in 11.7% (95% confidence interval 9.6–14.1) of the patients,59 and is an independent predictor of mid- to long-term mortality (hazard ratio: 1.68 [95% confidence interval 1.21–1.44, P < 0.01] for the Edwards Sapien valve,60 or

anesthesia & analgesia

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

Video 10. Midesophageal aortic short-axis view showing asymmetric calcification with a mobile noncoronary cusp.

Video 11. Deep transgastric view showing severe paravalvular aortic regurgitation after valve implantation.

Video 13. Midesophageal aortic long-axis view showing a primary good implantation result. LA, left atrium; LV, left ventricle.

Video 14. Midesophageal aortic long-axis view showing valve embolization into the left ventricle minutes after primary correct deployment of the prosthesis. LA, left atrium; LV, left ventricle.

http://links.lww.com/AA/A995; Video 14, Supplemental Digital Content 14, http://links.lww.com/AA/A996) into the aorta. Prosthetic valve malpositioning may occur during deployment if the heart continues to eject during rapid ventricular pacing (Video 9, Supplemental Digital Content 9, http://links.lww.com/AA/A991). Therefore, correct function of the temporary pacemaker should be ensured and the invasive pressure tracing checked to exclude continued ventricular ejection before implantation of the valve. In addition, during rapid ventricular pacing, the temporary pacemaker should be used in a fixed mode with no sensing and maximum output to minimize the risk of ventricular ejection.

Stroke

Video 12. Three-dimensional color of the left ventricular outflow tract showing severe paravalvular aortic regurgitation.

4.89, [95% confidence interval 3.95–15.81, P < 0.001] for the CoreValve).61 Inappropriate sizing of the implanted valve is also a risk for valve malpositioning, and may, in the worst-case scenario, lead to retrograde embolization into the LV or antegrade embolization (Video 13, Supplemental Digital Content 13,

October 2014 • Volume 119 • Number 4

The incidence of stroke after TAVR has declined since the inception of the procedure, from 7.8% to 2.1%–2.8%.62 When compared to conventional surgical AVR or medical treatment, TAVR is associated with a higher incidence of neurological events. Neither the approach, TF or TA, nor device are associated with the risk of neurological events, despite the different rate of high intensity transient signals recorded with transcranial Doppler.63 The etiology of stroke in TAVR patients is multifactorial: atherotic material from the ascending aorta or arch; calcific material from the native AV; thromboembolism

www.anesthesia-analgesia.org

795

E Review Article from the catheters used in the procedure; and air embolism during LV cannulation, prolonged hypotension, or dissection of brachiocephalic vessels. In the PARTNER trial, the incidence of 30-day major stroke was not statistically different between TAVR and surgical AVR (3.8% vs 2.1%, P = 0.20), but the central nervous system-associated morbidity was higher in TAVR patients (5.5% vs 2.4%, P = 0.04).4 Similarly, Panchal et al.64 reported a higher incidence of neurological morbidity at 66 weeks among 1426 patients from randomized registries, while surgical AVR was “protective” because it was associated with a risk reduction of 0.56 (95% confidence interval 0.35–0.88, P = 0.01). The miniaturization of the newer version of prostheses may lead to a decrease in the incidence of neurologic complications and a number of debris/calcium-capturing devices, currently under investigation, may further reduce the incidence of stroke.

Renal Function

Renal morbidity in TAVR patients is frequent and may require hemodialysis for treatment. A retrospective series (N = 213) reported an incidence of 11.7% of acute kidney injury, an independent predictor of postoperative mortality, and 1.4% need for hemodialysis. However, among patients with chronic kidney failure, acute kidney injury and need for hemodialysis were lower in TAVR than in surgical AVR.67 TAVR was not associated with more frequent renal injury in the PARTNER cohorts.4,5

CONCLUSIONS

There is no doubt that demand for TAVR is increasing rapidly, and that the procedure poses unique demands for the cardiovascular anesthesiologist. Ideally, the anesthesiologist should be involved in patient selection as well as management during and immediately after the procedure because he/she has a unique and vital role on the heart team. For the TF approach, the decision between GA and conscious sedation is not straightforward. The intraoperative use of TEE however may be advantageous in terms of valve positioning and detection of complications, and would necessitate GA. Randomized controlled trials in progress or planned should determine whether the use of TAVR in younger patients with fewer comorbidities should be permitted, and indeed whether TAVR becomes more prevalent than surgical AVR in the medium to long-term future. E DISCLOSURES

Name: Andrew A. Klein, MBBS, FRCA. Contribution: This author helped prepare the manuscript. Attestation: Andrew A. Klein approved the final manuscript. Name: Nikolas J. Skubas, MD, FASE. Contribution: This author helped prepare the manuscript. Attestation: Nikolas J. Skubas approved the final manuscript. Name: Joerg Ender, MD, PhD. Contribution: This author helped prepare the manuscript. Attestation: Joerg Ender approved the final manuscript. This manuscript was handled by: Martin J. London, MD. REFERENCES 1. Webb JG, Wood DA. Current status of transcatheter aortic valve replacement. J Am Coll Cardiol 2012;60:483–92 2. Huczek Z, Kochman J, Scisło P, Rymuza B, Kowalik R, Opolski G. Transcatheter aortic valve implantation (TAVI) in a patient

796    www.anesthesia-analgesia.org

with severe aortic insufficiency of aortic valve homograft. Kardiol Pol 2013;71:1325 3. Treede H, Mohr FW, Baldus S, Rastan A, Ensminger S, Arnold M, Kempfert J, Figulla HR. Transapical transcatheter aortic valve implantation using the JenaValve™ system: acute and 30-day results of the multicentre CE-mark study. Eur J Cardiothorac Surg 2012;41:e131–8 4. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Brown DL, Block PC, Guyton RA, Pichard AD, Bavaria JE, Herrmann HC, Douglas PS, Petersen JL, Akin JJ, Anderson WN, Wang D, Pocock S; PARTNER Trial Investigators. Transcatheter aorticvalve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 2010;363:1597–607 5. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Williams M, Dewey T, Kapadia S, Babaliaros V, Thourani VH, Corso P, Pichard AD, Bavaria JE, Herrmann HC, Akin JJ, Anderson WN, Wang D, Pocock SJ; PARTNER Trial Investigators. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med 2011;364:2187–98 6. Piazza N, Kalesan B, van Mieghem N, Head S, Wenaweser P, Carrel TP, Bleiziffer S, de Jaegere PP, Gahl B, Anderson RH, Kappetein AP, Lange R, Serruys PW, Windecker S, Jüni P. A 3-center comparison of 1-year mortality outcomes between transcatheter aortic valve implantation and surgical aortic valve replacement on the basis of propensity score matching among intermediate-risk surgical patients. JACC Cardiovasc Interv 2013;6:443–51 7. Adams DH, Popma JJ, Reardon MJ, Yakubov SJ, Coselli JS, Deeb GM, Gleason TG, Buchbinder M, Hermiller J Jr, Kleiman NS, Chetcuti S, Heiser J, Merhi W, Zorn G, Tadros P, Robinson N, Petrossian G, Hughes GC, Harrison JK, Conte J, Maini B, Mumtaz M, Chenoweth S, Oh JK; U.S. CoreValve Clinical Investigators. Transcatheter aortic-valve replacement with a self-expanding prosthesis. N Engl J Med 2014;370:1790–8 8. Klein AA, Webb ST, Tsui S, Sudarshan C, Shapiro L, Densem C. Transcatheter aortic valve insertion: anaesthetic implications of emerging new technology. Br J Anaesth 2009;103:792–9 9. Billings FT 4th, Kodali SK, Shanewise JS. Transcatheter aortic valve implantation: anesthetic considerations. Anesth Analg 2009;108:1453–62 10. Mukherjee C, Ender J, Shanewise JS, Cheung AT. Transcatheter aortic valve replacement: an encounter with innovation in the field of cardiothoracic anesthesiology. Anesth Analg 2013;116:517–9 11. Block PC, Palacios IF. Clinical and hemodynamic follow-up after percutaneous aortic valvuloplasty in the elderly. Am J Cardiol 1988;62:760–3 12. Cribier A, Eltchaninoff H, Bash A, Borenstein N, Tron C, Bauer F, Derumeaux G, Anselme F, Laborde F, Leon MB. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation 2002;106:3006–8 13. Popma JJ, Adams DH, Reardon MJ, Yakubov SJ, Kleiman NS, Heimansohn D, Hermiller J Jr, Hughes GC, Harrison JK, Coselli J, Diez J, Kafi A, Schreiber T, Gleason TG, Conte J, Buchbinder M, Deeb GM, Carabello B, Serruys PW, Chenoweth S, Oh JK; CoreValve United States Clinical Investigators. Transcatheter aortic valve replacement using a self-expanding bioprosthesis in patients with severe aortic stenosis at extreme risk for surgery. J Am Coll Cardiol 2014;63:1972–81 14. Wendt D, Osswald BR, Kayser K, Thielmann M, Tossios P, Massoudy P, Kamler M, Jakob H. Society of Thoracic Surgeons score is superior to the EuroSCORE determining mortality in high risk patients undergoing isolated aortic valve replacement. Ann Thorac Surg 2009;88:468–74 15. Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP, Guyton RA, O’Gara PT, Ruiz CE, Skubas NJ, Sorajja P, Sundt TM, Thomas JD. 2014 AHA/ACC Guideline for the management of patients with valvular heart disease: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol 2014;10:e57–185 16. Ashworth A, Klein AA. Cell salvage as part of a blood conservation strategy in anaesthesia. Br J Anaesth 2010;105:401–16

anesthesia & analgesia

Controversies in Anesthesia for Transcatheter Aortic Valve Replacement

17. Vallabhajosyula P, Bavaria JE. Transcatheter aortic valve implantation: complications and management. J Heart Valve Dis 2011;20:499–509 18. Huffmyer J, Tashjian J, Raphael J, Jaeger JM. Management of the patient for transcatheter aortic valve implantation in the perioperative period. Semin Cardiothorac Vasc Anesth 2012;16:25–40 19. Seiffert M, Conradi L, Baldus S, Schirmer J, Blankenberg S, Reichenspurner H, Diemert P, Treede H. Severe intraprocedural complications after transcatheter aortic valve implantation: calling for a heart team approach. Eur J Cardiothorac Surg 2013;44:478–84 20. Tommaso CL, Bolman RM 3rd, Feldman T, Bavaria J, Acker MA, Aldea G, Cameron DE, Dean LS, Fullerton D, Hijazi ZM, Horlick E, Miller DC, Moon MR, Ringel R, Ruiz CE, Trento A, Weiner BH, Zahn EM. Multisociety (AATS, ACCF, SCAI, and STS) expert consensus statement: operator and institutional requirements for transcatheter valve repair and replacement, part 1: transcatheter aortic valve replacement. J Am Coll Cardiol 2012;59:2028–42 21. Bufton KA, Augoustides JG, Cobey FC. Anesthesia for transfemoral aortic valve replacement in North America and Europe. J Cardiothorac Vasc Anesth 2013;27:46–9 22. Di Mario C, Eltchaninoff H, Moat N, Goicolea J, Ussia GP, Kala P, Wenaweser P, Zembala M, Nickenig G, Alegria Barrero E, Snow T, Iung B, Zamorano P, Schuler G, Corti R, Alfieri O, Prendergast B, Ludman P, Windecker S, Sabate M, Gilard M, Witowski A, Danenberg H, Schroeder E, Romeo F, Macaya C, Derumeaux G, Maggioni A, Tavazzi L; Transcatheter Valve Treatment Sentinel Registry (TCVT) Investigators of the EURObservational Research Programme (EORP) of the European Society of Cardiology. The 2011-12 pilot European Sentinel Registry of Transcatheter Aortic Valve Implantation: in-hospital results in 4,571 patients. EuroIntervention 2013;8:1362–71 23. Bergmann L, Kahlert P, Eggebrecht H, Frey U, Peters J, Kottenberg E. Transfemoral aortic valve implantation under sedation and monitored anaesthetic care–a feasibility study. Anaesthesia 2011;66:977–82 24. Durand E, Borz B, Godin M, Tron C, Litzler PY, Bessou JP, Bejar K, Fraccaro C, Sanchez-Giron C, Dacher JN, Bauer F, Cribier A, Eltchaninoff H. Transfemoral aortic valve replacement with the Edwards SAPIEN and Edwards SAPIEN XT prosthesis using exclusively local anesthesia and fluoroscopic guidance: feasibility and 30-day outcomes. JACC Cardiovasc Interv 2012;5:461–7 25. Motloch LJ, Rottlaender D, Reda S, Larbig R, Bruns M, MüllerEhmsen J, Strauch J, Madershahian N, Erdmann E, Wahlers T, Hoppe UC. Local versus general anesthesia for transfemoral aortic valve implantation. Clin Res Cardiol 2012;101:45–53 26. Dehédin B, Guinot PG, Ibrahim H, Allou N, Provenchère S, Dilly MP, Vahanian A, Himbert D, Brochet E, Radu C, Nataf P, Montravers P, Longrois D, Depoix JP. Anesthesia and perioperative management of patients who undergo transfemoral transcatheter aortic valve implantation: an observational study of general versus local/regional anesthesia in 125 consecutive patients. J Cardiothorac Vasc Anesth 2011;25:1036–43 27. Guarracino F, Covello RD, Landoni G, Baldassarri R, Stefani M, Cariello C, Ruggeri L, Franco A, Gerli C, Pappalardo F, Zangrillo A. Anesthetic management of transcatheter aortic valve implantation with transaxillary approach. J Cardiothorac Vasc Anesth 2011;25:437–43 28. Yamamoto M, Meguro K, Mouillet G, Bergoend E, Monin JL, Lim P, Dubois-Rande JL, Teiger E. Effect of local anesthetic management with conscious sedation in patients undergoing transcatheter aortic valve implantation. Am J Cardiol 2013;111:94–9 29. Petridis FD, Savini C, Castelli A, Di Bartolomeo R. Awake transapical aortic valve implantation. Interact Cardiovasc Thorac Surg 2012;14:673–4 30. Mukherjee C, Walther T, Borger MA, Kempfert J, Schuler G, Mohr FW, Ender J. Awake transapical aortic valve implantation using thoracic epidural anesthesia. Ann Thorac Surg 2009;88:992–4 31. Oguri A, Yamamoto M, Mouillet G, Gilard M, Laskar M, Eltchaninoff H, Fajadet J, Iung B, Donzeau-Gouge P, Leprince P, Leguerrier A, Prat A, Lievre M, Chevreul K, MD; DuboisRande J-L, Chopard J, Van Belle E, Otsuka T, Teiger E, on behalf of FRANCE 2 Registry Investigators. Clinical Outcomes and Safety of Transfemoral Aortic Valve Implantation Under General Versus Local Anesthesia. Subanalysis of the French

October 2014 • Volume 119 • Number 4

Aortic National CoreValve and Edwards 2 Registry. Circ Cardiovasc Interv. 2014 Jul 8. [Epub ahead of print]. 32. Moss RR, Ivens E, Pasupati S, Humphries K, Thompson CR, Munt B, Sinhal A, Webb JG. Role of echocardiography in percutaneous aortic valve implantation. JACC Cardiovasc Imaging 2008;1:15–24 33. Généreux P, Reiss GR, Kodali SK, Williams MR, Hahn RT. Periaortic hematoma after transcatheter aortic valve replacement: description of a new complication. Catheter Cardiovasc Interv 2012;79:766–76 34. Mukherjee C, Hein F, Holzhey D, Lukas L, Mende M, Kaisers UX, Ender J. Is real time 3D transesophageal echocardiography a feasible approach to detect coronary ostium during transapical aortic valve implantation? J Cardiothorac Vasc Anesth 2013;27:654–9 35. Zamorano JL, Badano LP, Bruce C, Chan KL, Gonçalves A, Hahn RT, Keane MG, La Canna G, Monaghan MJ, Nihoyannopoulos P, Silvestry FE, Vanoverschelde JL, Gillam LD. EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. J Am Soc Echocardiogr 2011;24:937–65 36. Holmes DR Jr, Mack MJ, Kaul S, Agnihotri A, Alexander KP, Bailey SR, Calhoon JH, Carabello BA, Desai MY, Edwards FH, Francis GS, Gardner TJ, Kappetein AP, Linderbaum JA, Mukherjee C, Mukherjee D, Otto CM, Ruiz CE, Sacco RL, Smith D, Thomas JD. 2012 ACCF/AATS/SCAI/STS expert consensus document on transcatheter aortic valve replacement. J Am Coll Cardiol 2012;59:1200–54 37. Smith LA, Dworakowski R, Bhan A, Delithanasis I, Hancock J, Maccarthy PA, Wendler O, Thomas MR, Monaghan MJ. Realtime three-dimensional transesophageal echocardiography adds value to transcatheter aortic valve implantation. J Am Soc Echocardiogr 2013;26:359–69 38. Jilaihawi H, Doctor N, Kashif M, Chakravarty T, Rafique A, Makar M, Furugen A, Nakamura M, Mirocha J, Gheorghiu M, Stegic J, Okuyama K, Sullivan DJ, Siegel R, Min JK, Gurudevan SV, Fontana GP, Cheng W, Friede G, Shiota T, Makkar RR. Aortic annular sizing for transcatheter aortic valve replacement using cross-sectional 3-dimensional transesophageal echocardiography. J Am Coll Cardiol 2013;61:908–16 39. Tamborini G, Fusini L, Gripari P, Muratori M, Cefalù C, Maffessanti F, Alamanni F, Bartorelli A, Pontone G, Andreini D, Bertella E, Fiorentini C, Pepi M. Feasibility and accuracy of 3DTEE versus CT for the evaluation of aortic valve annulus to left main ostium distance before transcatheter aortic valve implantation. JACC Cardiovasc Imaging 2012;5:579–88 40. Pontone G, Andreini D, Bartorelli AL, Bertella E, Mushtaq S, Gripari P, Loguercio M, Cortinovis S, Baggiano A, Conte E, Beltrama V, Annoni A, Formenti A, Tamborini G, Muratori M, Guaricci A, Alamanni F, Ballerini G, Pepi M. Comparison of accuracy of aortic root annulus assessment with cardiac magnetic resonance versus echocardiography and multidetector computed tomography in patients referred for transcatheter aortic valve implantation. Am J Cardiol 2013;112:1790–9 41. Bergmann L, Großwendt T, Kahlert P, Konorza T, Wendt D, Thielmann M, Heusch G, Peters J, Kottenberg E. Arrhythmogenic risk of pulmonary artery catheterisation in patients with severe aortic stenosis undergoing transcatheter aortic valve implantation. Anaesthesia 2013;68:46–51 42. Ben-Dor I, Goldstein SA, Pichard AD, Satler LF, Maluenda G, Li Y, Syed AI, Gonzalez MA, Gaglia MA Jr, Wakabayashi K, Delhaye C, Belle L, Wang Z, Collins SD, Torguson R, Okubagzi P, Aderotoye A, Xue Z, Suddath WO, Kent KM, Epstein SE, Lindsay J, Waksman R. Clinical profile, prognostic implication, and response to treatment of pulmonary hypertension in patients with severe aortic stenosis. Am J Cardiol 2011;107:1046–51 43. Seiffert M, Sinning JM, Meyer A, Wilde S, Conradi L, VasaNicotera M, Ghanem A, Kempfert J, Hammerstingl C, Ojeda FM, Kim WK, Koschyk DH, Schirmer J, Baldus S, Grube E, Möllmann H, Reichenspurner H, Nickenig G, Blankenberg S, Diemert P, Treede H, Walther T, Werner N, Schnabel RB. Development of a risk score for outcome after transcatheter aortic valve implantation. Clin Res Cardiol 2014;103:631–40 44. Auffret V, Boulmier D, Oger E, Bedossa M, Donal E, Laurent M, Sost G, Beneux X, Harmouche M, Verhoye JP, Le Breton H. Predictors of 6-month poor clinical outcomes after transcatheter aortic valve implantation. Arch Cardiovasc Dis 2014;107:10–20

www.anesthesia-analgesia.org

797

E Review Article 45. Drews T, Pasic M, Buz S, Dreysse S, Klein C, Kukucka M, Mladenow A, Hetzer R, Unbehaun A. Elective use of femorofemoral cardiopulmonary bypass during transcatheter aortic valve implantation. Eur J Cardiothorac Surg 2014. [Epub ahead of print] PMID: 24627439 46. Kappetein AP, Head SJ, Généreux P, Piazza N, van Mieghem NM, Blackstone EH, Brott TG, Cohen DJ, Cutlip DE, van Es GA, Hahn RT, Kirtane AJ, Krucoff MW, Kodali S, Mack MJ, Mehran R, Rodés-Cabau J, Vranckx P, Webb JG, Windecker S, Serruys PW, Leon MB; Valve Academic Research Consortium-2. Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document. J Thorac Cardiovasc Surg 2013;145:6–23 47. Leon MB, Piazza N, Nikolsky E, Blackstone EH, Cutlip DE, Kappetein AP, Krucoff MW, Mack M, Mehran R, Miller C, Morel MA, Petersen J, Popma JJ, Takkenberg JJ, Vahanian A, van Es GA, Vranckx P, Webb JG, Windecker S, Serruys PW. Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research Consortium. Eur Heart J 2011;32:205–17 48. Généreux P, Webb JG, Svensson LG, Kodali SK, Satler LF, Fearon WF, Davidson CJ, Eisenhauer AC, Makkar RR, Bergman GW, Babaliaros V, Bavaria JE, Velazquez OC, Williams MR, Hueter I, Xu K, Leon MB; PARTNER Trial Investigators. Vascular complications after transcatheter aortic valve replacement: insights from the PARTNER (Placement of AoRTic TraNscathetER Valve) trial. J Am Coll Cardiol 2012;60:1043–52 49. Hayashida K, Lefèvre T, Chevalier B, Hovasse T, Romano M, Garot P, Mylotte D, Uribe J, Farge A, Donzeau-Gouge P, Bouvier E, Cormier B, Morice MC. Transfemoral aortic valve implantation new criteria to predict vascular complications. JACC Cardiovasc Interv 2011;4:851–8 50. Borz B, Durand E, Godin M, Tron C, Canville A, Litzler PY, Bessou JP, Cribier A, Eltchaninoff H. Incidence, predictors and impact of bleeding after transcatheter aortic valve implantation using the balloon-expandable Edwards prosthesis. Heart 2013;99:860–5 51. Fraccaro C, Buja G, Tarantini G, Gasparetto V, Leoni L, Razzolini R, Corrado D, Bonato R, Basso C, Thiene G, Gerosa G, Isabella G, Iliceto S, Napodano M. Incidence, predictors, and outcome of conduction disorders after transcatheter self-expandable aortic valve implantation. Am J Cardiol 2011;107:747–54 52. Jilaihawi H, Chin D, Vasa-Nicotera M, Jeilan M, Spyt T, Ng GA, Bence J, Logtens E, Kovac J. Predictors for permanent pacemaker requirement after transcatheter aortic valve implantation with the CoreValve bioprosthesis. Am Heart J 2009;157:860–6 53. Buellesfeld L, Stortecky S, Heg D, Hausen S, Mueller R, Wenaweser P, Pilgrim T, Gloekler S, Khattab AA, Huber C, Carrel T, Eberle B, Meier B, Boekstegers P, Jüni P, Gerckens U, Grube E, Windecker S. Impact of permanent pacemaker implantation on clinical outcome among patients undergoing transcatheter aortic valve implantation. J Am Coll Cardiol 2012;60:493–501 54. Fassl J, Walther T, Groesdonk HV, Kempfert J, Borger MA, Scholz M, Mukherjee C, Linke A, Schuler G, Mohr FW, Ender J. Anesthesia management for transapical transcatheter aortic valve implantation: a case series. J Cardiothorac Vasc Anesth 2009;23:286–91 55. Ribeiro HB, Webb JG, Makkar RR, Cohen MG, Kapadia SR, Kodali S, Tamburino C, Barbanti M, Chakravarty T, Jilaihawi H, Paradis JM, de Brito FS Jr, Cánovas SJ, Cheema AN, de Jaegere PP, del Valle R, Chiam PT, Moreno R, Pradas G, Ruel M, SalgadoFernández J, Sarmento-Leite R, Toeg HD, Velianou JL, Zajarias A, Babaliaros V, Cura F, Dager AE, Manoharan G, Lerakis S, Pichard AD, Radhakrishnan S, Perin MA, Dumont E, Larose E, Pasian SG, Nombela-Franco L, Urena M, Tuzcu EM, Leon MB, Amat-Santos IJ, Leipsic J, Rodés-Cabau J. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. J Am Coll Cardiol 2013;62:1552–62 56. Conte AH, Gurudevan SV, Lubin L, Shiota T, Labounty T. Transesophageal echocardiographic evaluation of left main coronary artery stent protection during transcatheter aortic valve implantation. Anesth Analg 2013;116:49–52 57. Gripari P, Ewe SH, Fusini L, Muratori M, Ng AC, Cefalù C, Delgado V, Schalij MJ, Bax JJ, Marsan NA, Tamborini G, Pepi M. Intraoperative 2D and 3D transoesophageal echocardiographic

798    www.anesthesia-analgesia.org

predictors of aortic regurgitation after transcatheter aortic valve implantation. Heart 2012;98:1229–36 58. Athappan G, Patvardhan E, Tuzcu EM, Svensson LG, Lemos PA, Fraccaro C, Tarantini G, Sinning JM, Nickenig G, Capodanno D, Tamburino C, Latib A, Colombo A, Kapadia SR. Incidence, predictors, and outcomes of aortic regurgitation after transcatheter aortic valve replacement: meta-analysis and systematic review of literature. J Am Coll Cardiol 2013;61:1585–95 59. Hayashida K, Lefèvre T, Chevalier B, Hovasse T, Romano M, Garot P, Bouvier E, Farge A, Donzeau-Gouge P, Cormier B, Morice MC. Impact of post-procedural aortic regurgitation on mortality after transcatheter aortic valve implantation. JACC Cardiovasc Interv 2012;5:1247–56 60. Gotzmann M, Korten M, Bojara W, Lindstaedt M, Rahlmann P, Mügge A, Ewers A. Long-term outcome of patients with moderate and severe prosthetic aortic valve regurgitation after transcatheter aortic valve implantation. Am J Cardiol 2012;110:1500–6 61. Zahn R, Gerckens U, Grube E, Linke A, Sievert H, Eggebrecht H, Hambrecht R, Sack S, Hauptmann KE, Richardt G, Figulla HR, Senges J; German Transcatheter Aortic Valve Interventions-Registry Investigators. Transcatheter aortic valve implantation: first results from a multi-centre real-world registry. Eur Heart J 2011;32:198–204 62. Kahlert P, Al-Rashid F, Plicht B, Hildebrandt H, Patsalis P, Chilali KE, Wendt D, Thielmann M, Bergmann L, Kottenberg E, Schlamann M, Eggebrecht H, Jakob H, Heusch G, Konorza T, Erbel R. Incidence, predictors, origin and prevention of early and late neurological events after transcatheter aortic valve implantation (TAVI): a comprehensive review of current data. J Thromb Thrombolysis 2013;35:436–49 63. Kahlert P, Al-Rashid F, Döttger P, Mori K, Plicht B, Wendt D, Bergmann L, Kottenberg E, Schlamann M, Mummel P, Holle D, Thielmann M, Jakob HG, Konorza T, Heusch G, Erbel R, Eggebrecht H. Cerebral embolization during transcatheter aortic valve implantation: a transcranial Doppler study. Circulation 2012;126:1245–55 64. Panchal HB, Ladia V, Desai S, Shah T, Ramu V. A meta-analysis of mortality and major adverse cardiovascular and cerebrovascular events following transcatheter aortic valve implantation versus surgical aortic valve replacement for severe aortic stenosis. Am J Cardiol 2013;112:850–60 65. Bagur R, Webb JG, Nietlispach F, Dumont E, De Larochellière R, Doyle D, Masson JB, Gutiérrez MJ, Clavel MA, Bertrand OF, Pibarot P, Rodés-Cabau J. Acute kidney injury following transcatheter aortic valve implantation: predictive factors, prognostic value, and comparison with surgical aortic valve replacement. Eur Heart J 2010;31:865–74 66. Greenbaum AB, O’Neill WW, Paone G, Guerrero ME, Wyman JF, Cooper RL, Lederman RJ. Caval-aortic access to allow transcatheter aortic valve replacement in patients otherwise ineligible: initial human experience. J Am Coll Cardiol 2014;63:2795–804 67. Covello RD, Ruggeri L, Landoni G, Guarracino F, Bignami E, Gonfalini M, Virzo I, Michev I, Colombo A, Zangrillo A. Transcatheter implantation of an aortic valve: anesthesiological management. Minerva Anestesiol 2010;76:100–8 68. Ree RM, Bowering JB, Schwarz SK. Case series: anesthesia for retrograde percutaneous aortic valve replacement–experience with the first 40 patients. Can J Anaesth 2008;55:761–8 69. Grube E, Schuler G, Buellesfeld L, Gerckens U, Linke A, Wenaweser P, Sauren B, Mohr FW, Walther T, Zickmann B, Iversen S, Felderhoff T, Cartier R, Bonan R. Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation selfexpanding CoreValve prosthesis: device success and 30-day clinical outcome. J Am Coll Cardiol 2007;50:69–76 70. Guinot PG, Depoix JP, Etchegoyen L, Benbara A, Provenchère S, Dilly MP, Philip I, Enguerand D, Ibrahim H, Vahanian A, Himbert D, Al-Attar N, Nataf P, Desmonts JM, Montravers P, Longrois D. Anesthesia and perioperative management of patients undergoing transcatheter aortic valve implantation: analysis of 90 consecutive patients with focus on perioperative complications. J Cardiothorac Vasc Anesth 2010;24:752–61 71. Covello RD, Maj G, Landoni G, Maisano F, Michev I, Guarracino F, Alfieri O, Colombo A, Zangrillo A. Anesthetic management of percutaneous aortic valve implantation: focus on challenges encountered and proposed solutions. J Cardiothorac Vasc Anesth 2009;23:280–5

anesthesia & analgesia

Controversies and complications in the perioperative management of transcatheter aortic valve replacement.

Transcatheter aortic valve replacement (TAVR) is performed with increasing frequency in the United States since Food and Drug Administration approval ...