Guy
H. Sebag,
MD
#{149} Sheila
G. Moore,
MD
Effect of Trabecular Bone on the Appearance of Marrow in Gradient-Echo Imaging of the Appendicular Skeleton’
‘
This prospective study evaluated the effect of trabecular bone on the appearance of marrow in gradientecho (GRE) images of the appendicular skeleton in vivo at high magnetic field strength. Magnetic resonance (MR) imaging of 10 normal extremities in five patients was performed with spin-echo (SE) and GRE sequences. The latter were obtamed with gradient recalled acquisition in a steady state. SE and GRE sequences had identical spacing and planes of imaging. Cortical bone appeared as a signal void regardless of the pulse sequences and parameters. Marrow in contact with trabecular bone exhibited a shortened effective transverse relaxation time (T2*) and resultant signal loss because of local field inhomogeneities where mineralized matrix interfaced with it. This T2* effect was increased in regions with more trabecular bone (epiphysis) than regions with little trabecular bone (diaphysis). A low signal intensity on GRE images may represent fatty marrow with a high content of trabecular bone and should not be interpreted only as hematopoietic marrow. Index
terms: Bones, MR studies, 40.1214 #{149} Extremities, MR studies, 40.1214 #{149} Magnetic resonance (MR), tissue characterization #{149} Magnetic (MR), pulse
resonance
Radiology
1
ogy, 052,
From the Stanford Stanford,
Department of Diagnostic Radio!University Medical Center, S. CA 94305-5105 (G.H.S., 5GM.),
ceived
August
Enfants
of Pediatric Malades,
7, 1989;
ber 1 1; revision received November 15. Supported from the Georges Lurcy York. 0
Address reprint RSNA, 1990
revision
Radiology, Paris
(G.H.S.).
requested
Re-
Octo-
October 31; accepted in part by a grant Foundation, New requests to 5GM.
(GRE) sequences used for mus-
RADIENT-ECHO
are
increasingly
culoske!etal
low fast imaging
imaging
ation
magnetic with the
because
they
al-
(MR) of threedimensional acquisition. GRE sequences have several contrast characteristics that are distinct from those of spin-echo (SE) sequences (1). The signal intensity of tissue is dependent on effective transverse relaxtime
(T2*)
resonance advantage
as opposed
to the
T2
dependence of SE sequences. To our knowledge, the consequence of this T2* dependence on bone marrow contrast has not been previously studied in vivo in the appendicular skeleton. A 1986 study (2) has shown that fat and water in contact with trabecular bone in vitro exhibits shortened T2* relaxation times. To determine if this T2* effect extends to the appendicular skeleton in vivo at high field strength, a prospective study using GRE was undertaken. PATIENTS MR imaging
AND
METHODS
of 10 normal
extremities
in five patients (aged 5-40 years; three females and two males) was performed on a 1.5-T superconducting MR system with both SE and GRE pulse sequences. SE sequences were used with a repetition time (TR) of 600-800 msec and an echo time (TE) of 20 msec for Ti-weighted images and a TR of 2,000 msec and TE of 80 msec for T2-weighted images. GRE images
were
1990; 174:855-859
and the Department H#{244}pital des
sequences
G
obtained
with
gradient
recalled
ac-
quisition in a steady state (GRASS) (3). A TR of 60 msec, flip angle of 30#{176}, and TE of 10, 12, 14, 16, 20, 24, 30, and 40 msec were used. An additional set of images was obtamed with a TR of 250 msec, flip angle of 300, and TE of 10, 12, 14, 16, 20, 24, 30, and 40 msec. Spacing and plane of imaging were the same for GRE and SE sequences. Five-millimeter-thick corona! and transverse sections of the femur, tibia, or both, including proximal and distal epiphyses, were obtained with a 24-48cm field of view, six to eight repetitions, and a 256 X 256 matrix. At the intermediate flip angle (300) and short TR (60 msec) used for this study, the
signal intensity of the tissue is dependent on the ratio of T2* and Ti relaxation times. Tissues with a long T2*/T1 give a high signal intensity. T2* dependence on both short TR (60 msec) and long TR (250
msec) GRASS sequences is increased by lengthening the TE (3). Furthermore, because of its different spectral components, fat signal exhibits two chemical shift-induced modulations on GRASS images at 1.5 T. The
first
modulation has a period 4 msec and the secof approximately 20 msec
(p) of approximately ond
a period
(4). For each
modulation,
the spectra!
components are in phase when the TE is equal to integer multiples of the period - p. They are out of phase when the TE is equal to [(2n + 1)12] . p (ie, for both modulations, the spectral components are in phase at approximately 20 and 40 msec
n
and out of phase at approximately 10 and 30 msec). We therefore used variable echo delays to take sideration.
this
phenomenon
into
con-
RESULTS Cortical bone appeared as a signal void regardless of the pulse sequences and parameters. On SE images, fatty (yellow) marrow was seen as increased signal intensity on Tiweighted images (Fig 1) and decreased signal intensity on T2weighted images commensurate with the decrease in signal intensity of subcutaneous fat. Signal intensity of hematopoietic (red) marrow was intermediate (similar to that of muscle) on SE sequences. On the GRASS images the signal intensity of marrow was the same regardless
msec
of which
TR
or 250 msec).
was
used
(60
intensity of diaphyseal fatty marrow was similar to that of subcutaneous fat regardless of TE. Signal intensity of the epiphyseal fatty marrow, however, was moderately less than that of subcuta-
Abbreviations:
Signal
GRASS
in a steady state, echo, SE = spin echo, TE repetition time.
gradient
quisition
GRE =
=
recalled
ac-
gradient
echo time, TR
=
855
Figure
1.
Transverse Ti-weighted (b) the metaphysis, and
epiphysis,
neous
fat
on
the
(TR
800 msec,
=
(c) the
lO-msec-TE
diaphysis.
TE
20 msec)
Fatty
marrow
SE images
of the distal
is isointense
to subcutaneous
of transverse
GRASS
left femur fat
in a 40-year-old in the
three
man
segments
show of the
(a) the femur.
image
(Fig 2a). On the images obtained with increasing TE and constant TR and flip angle, the signal intensity of the epiphyseal fatty marrow decreased as compared with the signal intensity
of subcutaneous
fat
(Fig
2c). Epiphyseal signal remained creased regardless of the phase subcutaneous thus displaying phase/out-of-phase
fat
relative
2b,
deof
to muscle,
no significant effect
in(Fig
3).
The effect on metaphysea! marrow was intermediate to the effect on diaphyseal or epiphyseal marrow. The signal
intensity
marrow
of the
was
metaphyseal
moderately
less
than
that of subcutaneous msec-TE image.
fat However,
on
the iOthe signal
intensity did not
metaphyseal to the same
marrow degree
of the decrease
as that
of the
epiphysea!
marrow
with increasing TE. The presence of hematopoietic versus fatty marrow in the metaphysis did not affect the decrease in metaphyseal marrow signal
2a.
Figure illustrates
2.
Matrix the
interaction
between
umn distal metaphysis, right and 14 msec (bottom row) (Fig
images
marrow
column
site
of the same (left
patient
column
middiaphysis)
=
and
as in Figure
1. This series middle colTE. (a) TE of 12 msec (top row) distal
epiphysis,
2 continues).
intensity.
DISCUSSION GRE based
gle than
is a fast on
the
excitation
imaging technique use of a partial flip
pulse
90#{176}) followed
(typically by
an-
less
a gradient
re-
versa! to refocus the echo and generate the signal (1). Factors that influence GRE imaging contrast are the acquisition technique and tissue characteristics. It is reported that images acquired with an appropriate
TR, TE, and
flip
contrast similar (i,3). However,
appearance
angle
can
GRE
provide
a
to that of SE imaging we describe marrow
on GRE
images
that
dif-
fers from marrow appearance on SE images in that fatty marrow signal is not always isointense to subcutane-
856
Radiology
#{149}
ous fat but varies according to the anatomic location of the marrow and the TE selected on the GRE sequence. The amount of trabecular bone in addition to marrow composition (red versus yellow) is reflected on GRE images. On SE images, hydrogen nuclei associated with aliphatic carbon chains (fatty marrow) and water protons (hematopoietic marrow) are the major contributors to the signal pattern (5), as opposed to trabecular bone, which has a minor or absent contribution. Regardless of the SE sequence, bone marrow signal is homogeneous taneous skeleton
and isointense to the subcufat in those regions of the containing exclusively yel-
low marrow (ie, epiphyses and adult distal appendicular skeleton). Several contrast characteristics of GRE images are distinct from those of SE images and have a significant influence on signal intensity. Most notably, signal intensity on GRE images is dependent on the T2*, as opposed to the T2 dependence on SE sequences. On GRE images, there is no rephasing of local magnetic field inhomogeneities by a i800 pulse, and thus T2* dephasing is not recovered. Therefore GRE imaging is very sensitive to field inhomogeneities arising either extrinsically (nonuniformity in the applied main static field) or intrinsically
(heterogeneity
of the
March
body
1990
quence. The hydrogen nuclei in the fatty acid chains exhibited a signal decrease in regions containing more trabecular metaphysis)
bone (the epiphysis than regions with
and little
or no trabecular bone (the diaphysis and subcutaneous fat) (Fig 4). The progressive increase in signal loss with increasing TE indicated that fatty marrow
bone
in contact
exhibits
T2*
and
with
trabecular
a markedly
shortened
dephasing
secondary
a spin
to local field inhomogeneities. The effect on marrow signal seen in our study is not due to dephasing by proton diffusion across local field gradients during the interecho time of a
iso0
SE pulse.
This
phenomenon,
af-
fecting only water, cannot be recovered by a i80#{176} pulse and results in T2 shortening. Therefore the subsequent
signal
seen
decrease
should
be also
on SE sequences.
Adipose
tissue
spectral finic
contains
components, protons
several
including
ole-
of unsaturated
fatty
acid and central long-chain fatty of approximately 64 MHz resonance
methy!ene of the acid. Chemical shift 4.i ppm (260 Hz at frequency) be-
tween
and
ene
the
vinyl
protons
quency
central
results
methyl-
in a high-fre-
modulation
with
a period
of
approximately 4 msec. In addition, there is a second, low-frequency modulation with a period of approximately 20 msec related to the chemical shift difference between the central CH2 and the -CH2-Oprotons
of the
Hence,
fatty
acid
fat signal
shift-induced
chains
other,
a). Epiphyseal signal seal signal intensity dle
column,
loss is increased is intermediate
with increasing TE (right column, b and c). Metaphybetween the diaphyseal and epiphyseal intensities (mid-
a-c).
itself) (6). On SE images TR shortening increases the degree of Ti weighting, but with TR shortening (