Arkadiusz
Polacin,
Evaluation and Image
PhD
Wihi
#{149}
A. Kalender,
PhD
Guy
Marchal,
#{149}
MD
of Section
Noise
Spiral computed tomography (CT) offers continuous volume scanning of complete organs or body sections within a single breath hold. Almost all image quality characteristics of spiral CT are identical to those of conventional section-by-section CT; however, there is a change in pixel noise values and degradation in the shape of the section sensitivity profiles (SSPs). Computer simulations, phantom measurements, and clinical studies were used in evaluating the SSP and noise characteristics of two new section-interpolation algorithms. The results were compared with standard CT and spiral CT data processed with the commonly employed linear section-interpolation algorithm. Degradation of SSP quality was insignificant for a table feed distance per 360#{176} revolution equal to the section thickness when the new algorithms were applied; noise values, however, increased. SSP width increased for table feed distances greater than the section width, the effect being less pronounced with the new algorithms. The value of these algorithms is primanly seen in the improved quality of multiplanar reformations and cine and three-dimensional displays.
Index terms: Computed tomography (CT), image processing #{149} Computed tomography (CT), image quality #{149} Computed tomography (CT), physics ‘ Computed tomography (CT), spiral technology
Sensitivity in Spiral CT’
S
PIRAL
was lice
computed introduced
in 1989
(1-5)
Profiles
tomography into clinical and
has
since
section
may
cause
omission
of struc-
tures or lesions and degradation the quality of multiplanar and dimensional It has been
displays. demonstrated
section
CT with
respect
185:29-35
in three-
that
image quality of spiral scanning equivalent to that of standard to most
the
is singlepa-
rameters; only image noise and the shape of the section sensitivity profiles (SSPs) are changed (5,23-25). The magnitude and direction of these effects depend on the scanning parameters employed and the reconstruction algorithms used. The shape of the resulting SSPs is a particular point of concern. We investigated different sectioninterpolation algorithms by means of computer simulations, phantom ex-
periments, termine and the
and
clinical
studies
their effects on SSPs, quality of multiplanar
termine whether pitch of greater 1992;
been
investigated and applied in a large number of studies (6-22). Spiral or continuous-volume scanning involves moving the patient through the gantry while CT data are simultaneously acquired with multiple 360#{176} acquisilions (5). This scanning mode is of particular clinical interest in contrast media studies, in which complete organs or a large volume need to be measured during the short time of maximum enhancement, or in examinations of organs subject to motion, in which misregistration from section to
mations and three-dimensional plays. In particular, we sought
Radiology
(CT) prac-
acceptable
image
to denoise, refordis-
to de-
spiral CT with a than 1 can produce
quality.
(Pitch
is de-
1 From Siemens Medical Systems, Henkestrasse 127, 8520 Erlangen, Germany (A.P., W.A.K.); and the Department of Radiology, University Hospitals, Louvain, Belgium (G.M.). From the 1991 RSNA scientific assembly. Received February 27, 1992; revision requested April 22; revision received May 18; accepted May 28. Address reprint requests to W.A.K.
C RSNA,
1992
fined
as the
table
360#{176} rotation section
feed
divided
distance
per
by the
nominal
thickness.)
AND
MATERIALS
METHODS
Section-Interpolation
Algorithms
Direct reconstruction of images from data obtained over any 360#{176} segment of a spiral CT acquisition will result in motion artifacts due to patient transport. To avoid such been must
artifacts, the data that would have obtained in the planar geometry be synthesized from the spiral data
set. This has to be done point by point for a complete data set (ie, for all projections over the 360#{176} angular range of rotation and for all data within each projection) (5). The new data set, which can be calculated for any arbitrary table position z within
the scanned volume, is subsequently subjected to the same image reconstruction process as any data set obtained with conventional single-section scanning. There are a number of alternatives for synthesizing “planar” data sets from spiral data. In the simplest case, this is done by means of linear interpolation (LI) between the two neighboring data points obtained at the same angle of rotation, z’ and z’ + d, with
d denoting
the table
feed
distance
per 360#{176} rotation (Fig la). Thereby, data from a range corresponding to 2 x 360#{176} are used, with z’ varied from z d to z. -
We refer LI.
to this
type
of processing
as 360#{176}
To limit the SSP broadening that results from using this large range, we exploit data for interpolation that were obtained with 180#{176} opposite views. This is equivalent to calculating a second spiral from the original that is offset by a distance d/2, as depicted in Figure lb. Thereby, data are available for interpolation that were obtained a total
closer range
being
used.
to the desired z axis position, of only 2 x (180#{176}+ fan angle)
We use the term
180#{176} LI when
Abbreviations: FWHM = full width at half maximum, FWTA = full width at tenth area, FWTM = full width at tenth maximum, HI =
higher-order lation,
SSP
interpolation, =
section
sensitivity
LI
=
linear
interpo-
proffle.
29
a.
b.
Figure
1.
would views
have been obtained
Data processing one to limit
allows
points
U between
in spiral in planar
CT. (a) Linear
geometry. (b) Interpolation the scan range used per image.
z” and z’
+ d
is em-
Rel. 1.2
ployed. A better
estimate
be obtained terpolation two
data
of the true
In our
z’ and z’ + d is most commonly data points and data points
between
points
between
measured
used to estimate data that derived from 180#{176} opposite
value I
I
may
value
by means of higher-order in(HI) with use of more than points.
interpolation
1.1
implementation,
we employed a modified cubic spline interpolation between points z z”, z’ + d, and z” + d (Fig lb). We use the term 180#{176} HI for this procedure.
1.0
-
0.9
-
0.8
-
0.7
Noise
Analysis
0.6
It has been shown that noise is reduced by a factor of V’2/3 in spiral CT with 360#{176} U processing when compared with the in standard single-section CT with the same nominal section thickness and radiation dose (5,24,25). In 180#{176} LI, noise
to increase
the central
noise
by a factor
ray in comparison
in 360#{176} LI, as is derived
is
of V2 for with
the
in the Ap-
pendix. Compared with the noise in standard CT at the same section thickness and
dose,
this means
or 1.15.
a noise
The increase
increase
of \/4/3
is less for off-axis region of interest 4 cm in
rays; for a central diameter, the average noise increase is expected to be 13%. For 180#{176} HI, no general prediction can be made, since the algorithm will adapt to the object. Simulation
of SSPs
It has been shown that SSPs in spiral CT with 360#{176} LI result as a convolution of the original SSP with the table motion function, which is given by a triangle of 2d base width and unity area; results of simulations and measurements were found to be in excellent agreement (25). The same agreement was found for 180#{176} LI with this simulation method; the triangular table motion function at the center of rotation has a base width of only d and unity area. For 180#{176} HI, a different approach had to be created.
We synthesized
CT projection
data of an infinitely
thin disk scanned
spiral
reconstructed
geometry
and
as a function of table position, directly yielding the SSP. This method was also applied to the other section-interpolation 30
Radiology
#{149}
in
images
SSP
area.
0.4
noise
expected
of 0.5
0.3
FWTA
0.2
I
0.1
FWTM I
0.0 -123
-10.0
-5.0
-7.5
-2.5
I
II
0.0
2.5
5.0
7.5
Figure
2.
algorithms, with those method. Figures
Width
parameters
[mm]
for SSPs. (See text for explanation.)
providing results consistent of the earlier convolution
of Merit
for
SSPs
The full width at half maximum (FWHM) is the most widely used descriptor of SSPs; this value is typically used to describe the nominal section width. The FWHM does not give any indication of whether the profile approaches the ideal form of a rectangle or deviates from this form substantially. In standard CT, section proffles approximate the ideal rectangular form well, in particular when postpatient collimators are used. In spiral CT, the scan is obtained
with
tion width; owing tive section width, broadened proffle, proffle is degraded
a selected
12.5
10.0
Z-positlon
nominal
sec-
to the motion, an effecthe FWHM of the will result. Even if the substantially, this will
not necessarily result in a large increase in the FWHM of the profile (5,25). We therefore employed additional descriptors, as illustrated in Figure 2. We determined the full width at one-tenth of the maximum (FWTM) and the full width at the points of the profiles where 90% of the area
is covered
and
10% of the area
is ex-
duded (full width at tenth area [FWTAJ). We consider FWTA the most adequate measure
of SSP quality.
Measurements All measurements Somatom Medical
Plus-S Systems,
were CT scanner Erlangen,
obtained
with
a
(Siemens Germany),
which acquires a full 360#{176} data set in 1 second and offers a spiral CT option with 32 seconds of continuous scanning. Section widths of 1-10 mm can be chosen; table feed speeds of 1-10 mm/sec are available. October
1992
Rd.
Rd.
value
I.
I
-
Processing LO
,.‘I
algorithm
none
-
value
I
(d-0)
-
360L1
\#{149}\
:/
:..:.:
0.8
Table 0.6
0.4
feed
L
d=5rnm/’
I
I
-
-
0.2
0.c
.12.5
.10.0
.7#{149}5 -5.0
.2.5
0.0
2.5
5.0
7.5 10.0 1 -5 Z-positlon [mm]
b.
a. rd.
rd.
FWHM
FWTM
5.
....G... -#{149}-
-.e.-
.
360
LX
180 180
LI HI
0.
1.
2.
0.
3.
1.
Pitch
3.
2. Pitch
d. rd.
FWTA
I Figure
5.
3.
SSP5
in spiral
CT for
different
section-interpolation
algo-
rithms at 5-mm nominal section thickness. (a, b) Profile plots for table feed speeds of 5 mm/sec (a) and 10 mm/sec (b). (c-c) Plots of FWHM FWTM (d), and FWTA (e), relative to the value for the original profile (d = 0), as a function of pitch.
4.
(c),
3.
Table 1 Image Noise
2.
in Spiral
Algorithm
Predicted
360#{176} LI
180#{176}LI 180#{176} HI
CT
0.82 1.13 Not available
Measured 0.83 1.12 1.29
0. 0.
1.
2.
3.
e.
We employed an experimental software package that allowed table feed speeds up to 20 mm/sec
and
and 180#{176} HI algorithms. Noise was measured
included
the
in a water
of
180#{176} LI
phan-
torn 20 cm in diameter. To evaluate spatial resolution in the longitudinal direction or
Volume
185
Number
#{149}
1
Note-Values given are relative standard CT at the same nominal thickness and dose. in
Pitch
z axis, we scanned spheres in a water phantom. We employed acrylic spheres
and 10 mm in diameter
and a wooden
sphere 10 mm in diameter. SSPs were measured with a phantom containing a 0.5-mm-thick disk oriented exactly perpendicular to the z axis. Phantom mea-
4
to the
noise
section
surements were obtained with both standard CT and spiral scanning for direct comparison.
Raw data from spiral CT acquisitions were available for a large number of cmical studies, obtained mA over 32 seconds
at 137 kVp and 145 with varying section
Radiology
#{149} 31
and table feed values.
thickness
We arbi-
trarily selected 20 studies of the abdomen and processed them with the three different algorithms for direct comparison.
RESULTS Noise
was
measured
for
standard
CT and spiral CT with 120 kVp, 165 mA, 5-mm nominal section thickness, and 5-mm/sec table feed speed. Re-
sults
are given
in Table
1; they
were
in excellent agreement with the predictions in the “Noise Analysis” seclion. Simulated SSPs for the different data processing algorithms are plotted in Figure 3a and 3b. It is apparent that SSPs are broadened in spiral CT for all algorithms. However, this effect is much
reduced
with
the
180#{176} algo-
rithms. In particular, even the profiles for 180#{176} HI at a pitch of 2 (ie, a table feed distance per 360#{176} revolution equal to twice thickness) are
those
the still
nominal narrower
confirmed parameters the width Figure
lower
of 1. This is
3c-3e.
These
results
apply
water confirmed
given
above
(Fig
the
re-
(Table
noise
FWHM FWTA FWHM FWfM FWTA
spheres
are
more
While
the
and
shape
pronounced
dard section-by-section however.
addi-
of the in stan-
CT scanning, in noise
levels
is apparent in the transaxial CT images, the effects of the SSPs on image quality were relatively hard to assess in clinical studies (Fig 5a-5c). Since SSP5
determine
spatial
resolution
The 180#{176} algorithms
proved resolution substantially these displays. These findings consistently Results
ues 32
obtained obtained
greater Radiology
#{149}
than
the
for Standard
CT
()* (.)*
(mm)t (pj)t
(j)t
case
were
quality.
is highest
for
180#{176} HI pro-
CT and Spiral
CT at 5-mm
CT
=
5 mm/sec.
=
10 mm/sec.
360#{176} LI
180#{176} LI
180#{176} HI
5.0
6.3
5.0
5.0
6.1
11.1
8.0
7.4
5.0
8.3
5.9
5.4
5.0
10.8
6.5
6.4
6.1
19.8
11.3
10.3
5.0
14.1
8.4
7.7
for
imin were
for all studies. at table feed val-
nominal
section
limited is particularly
scans
5-mm
lion and
in imthe
slightly
This
obtained
tion thicknesses. in Figure 6 were
nominal
with
large
occurs particularly in lung nodule screening (4,6,9,11,15) or in examination of patients who are not able to cooperate. Some studies show deci-
sec-
The images shown obtained with a
section
table
feed
acquisition;
thickness speed
over
effective
sive trast
and a
180#{176} HI,
respectively
These differences can ated in the multiplanar (Fig 6d-6i).
(Table
advantages medium
for spiral applications.
CT in conDupuy et
al (13) and Costello et al (18) reported improved image quality for spiral CT
sec-
thicknesses (FWHM) of 10.8, 6.5, 6.4 mm resulted for 360#{176} LI, 180#{176} 2).
best be apprecireformations
even
when
using
amounts
been
demonstrated
Simulations
and
Spiral CT is a promising CT technique. Its particular advantages result
ments yield consistent results. In transaxial CT reconstructions,
from terest
the fact that are scanned
the volumes continuously,
of inpro-
SSP degradation ticed in most
viding
a seamless
or overlapping
set
of images in every case. of particular importance anatomic
or involuntary
regions
subject
patient
This feature in scanning to voluntary
motion.
This
feed
is
of
CT on SSPs is well understood
(25) and has detail above.
DISCUSSION
reduced
a contrast medium bolus. With respect to image quality, the only concerns brought forward so far refer to the shape of SSPs (5,23-25). The influence of patient transport in
spiral
in
the longitudinal direction, the effect of data processing can best be demonstrated in multiplanar reformations
(Fig 5d-5i).
age
LI, and
difference
and
1). Multipla-
elongation
in the
Descriptors)
(SSP
Table feed speed t Table feed speed
10-mm/sec 32-second
this
1800 algorithms
Thickness
*
in spiral
CT;
Results Section
Descriptor
thickness
distortion
by the
Spiral
nar displays were generated for standard CT images obtained at 5-mm increments and spiral CT images calculated in 1-mm increments (Fig 4b). It is evident that the spheres are displayed elongated to different degrees tional
HI).
Standard
to
phantom
right = 180#{176} is increased
comparison of standard and spiral CT images. left = standard CT, upper right = 360#{176} LI, Noise is lowest for 360#{176} LI processing resolution
2
Simulation Nominal
thicknesses provided original profile is the same. There was good agreement between simulations and measurements; on average, differences between measured and calculated width parameters were about 0.1 mm. Spiral scans of the spheres in the
20-cm-diameter
(b) coronal
= 180#{176} LI, lower longitudinal direction
section of the
4a) visually
CT images of spheres; reformations (upper
Multiplanar
left
in the cessing.
quantitatively by the width given in Table 2. Plots of parameters are given in
arbitrary the shape
sults
4.
(a) CT images,
Table
section than
for 360#{176} LI at a pitch
Figure
the
distance
nominal
ticular details
structures
cases
can hardly be nowhen the table
is less
section
in experi-
than
or equal
thickness,
to
in par-
for thin sections. Attention to like small calcifications or bone
reveals
this
influence, October
how1992
t
r.
:.
--
3-
:
I
‘
t
,
,I
-
k
I
.
JiL d.
-
JL
g
‘
#{216}r, e.
h.___
.?t
V
,.
, I.
Figure 5. Spiral CT studies (32-second acquisition, 5-mm section thickness, 5-mm/sec table feed speed) of upper abdomen. (a-c) CT images at level of upper pole of the left kidney. (d-f) Coronal reformations through the spine. (g-i) Sagittal reformations through the left kidney. (a, d, g = 360#{176} LI; b, e, h = 180#{176}LI; c, f, i = 180#{176}HI.) The continuous acquisition in spiral CT provides excellent-quality multiplanar reformations. Note the absence of contour irregularities, which are commonly seen when reformations are obtained from consecutive breath-hold CT scans acquired at 5-mm section thickness. Anatomic detail is best displayed in reformatted images obtained with 180#{176} HI (note clear delineation of the diaphragm).
ever. The effect may become pronounced when thick sections are used in combination with high table feed values. Particular advantages of spiral scanfling and of the different data processing algorithms become apparent in multiplanar reformations. It is understood
that
spiral
scanning
elongate structures nal direction. This c
#{149} Number
tends
to
in the longitudieffect can be re1
duced substantially with the new algorithms, especially with 180#{176} HI. However, it is more important to realize that the same detrimental effect occurs in standard CT (Fig 4b). In standard CT, the choice of thickness and spacing of sections is decisive. Resolution in the longitudinal direction is much inferior to resolution in the x, y, or image plane in any case. Spiral
CT offers
considerable
im-
provement
in this
us to reconstruct table positions
ing. This provides tours
in multiplanar
Steps
commonly
respect,
images at arbitrarily
as it allows for
arbitrary fine spac-
smoothness
of con-
reformations.
done
in contiguous
section-by-section scanning, in particular for thick sections, are strongly reduced. Discontinuities due to incon-
sistent
breathing
are practically
levels eliminated.
between
scans
In addition, Radiology
#{149} 33
-.-..v
-.
L.
\
\
‘
‘
-
#{149}t:
.
1_ a.
,
b.
Figure
6.
quisition,
Spiral 5-mm
CT studies section
(32-second
thickness,
10-mm/sec
table feed speed) of patient with carcinoma of the right kidney. (a-c) Bolus-enhanced images show the hypervascular and partially
necrotic
carcinoma
C-
ac-
in the
CT
posterior
I
as-
pect of the upper pole of the right kidney. Note the differences in blurring between the lower edge of the liver, the anterior surface of the right kidney, the posterior surface of the left kidney, and the aorta. Also note the improved image contrast for the 180#{176} algonthms (arrows in c). (d-f) Coronal reformations. (g-i) Sagittal reformations. (a, d, g = 360#{176} LI; b, e, h
spiral lively fully
=
c, f, i
180#{176}LI;
=
180#{176}HI.)
scanning allows us to retrospecselect the table position that yields the anatomic or patho-
e.
h.
logic detail of interest in the section. This is particularly effective in interaclive cine displays (12,16,19). Efforts to scan the largest possible volume with spiral scanning in a given lime (ie, by choosing table feed values greater than the nominal seclion thickness) have to be seen and weighed
in the
context
of 5SF
degra-
dation (Fig 3). The selection of parameters will depend on the clinical demand and the diagnostic task. However, even for a pitch of greater than 1, scanning is continuous and in no case do gaps result or will lesions be overlooked. For lesions smaller than the section thickness, a reduclion in contrast may be observed, owing to the partial volume effect. Lung scanning-with the inherent high contrast of most lesions-will hardly be limited by this effect and therefore may become a primary field of application. In addition to producing increased scan volumes, pitch values of greater
than
1 bring
about
in mean organ dose, posure is distributed ume. Scanning
than 1 offers but it requires
with
a reduction
as the given over a larger
exvol-
1.
choices of section thickness, table feed value, and section interpolation algorithm still need to be determined for the various clinical applications.
P2(O)
=
O/’rrP+i(O
of greater
potential advantages, 180#{176} data processing to keep
adequate
This demands additional tional efforts and causes creased noise. The most
34
#{149} Radiology
-
.P1(0)for0
a pitch
SSPs.
computaslightly inappropriate
of image
ir)
+ (1
-
P(0)
+
(2
-
-
TI)
0/u)P1#{247}1(0)]dO
+
uT
[(2
+
(0/ui
-
-
0/ur)P+i(0
1)P1+1(0
We can rearrange
-
rr)
-
2ui)]dO.
the above
(A3)
by chang-
in the second
inte-
C [0,u-r)
0/ui)P1+1(0
P1+i(O
(1
-
2r
‘180LI =
(A2)
in
ing integration limits gral, which leads to
0/u)
P(0)d0,
J0 [(O/’rr)P+1(O
I180’LI
pixel
or
.
algorithms
derivation
noise for the 180#{176} LI algorithm, let us concentrate on the center of rotation and, correspondingly, on data measured with the central detector. For each section position z and angular position 0, the synthesized projection value P2(0) is calculated as
f2i
‘180’LI
resulting
APPENDIX To simplify
value (5,25). To get the image pixel value, we have to integrate over all projections:
-
-
ri)
+ (0/u
-
1)
2rr) for 0 C [rr, 2u),
where j is the number of the last scan reaching angle 0 before reaching position z and P(0) is a corresponding projection
f +
2[(O/’rr)P3+1(O
(1
-
-
ur)
0/ur)P1+1(0)]dO.
(A4)
(Al) Denoting variance of projection data as o.2(0) and taking into account statistical independence of the corresponding cornponents, we can express the variance of
October
1992
image
pixel noise
f
80’LI +
We can &(0)
3.
by
(I
-
4.
0/u)2o(8)]d0.
assume
further
(A5)
that a(0
ur)
-
(ie, the noise level is the same for all projections). This allows us to write cTso.u
4
=
J2
f
[(0/rr)2
+ (I
(A6)
Evaluation
of the integral
(A7) With
use of Equation
(A2)
Thus,
(ie,
8.
=
2w
4
(A8)
(A9
‘
that for image
References Kalender WA, Seissler W, Vock P. Singlebreath-hold spiral volumetric CT by continuous patient translation and scanner rotalion (abstr). Radiology 1989; 173(P):414. Vock
P,Jung
breathhold iary system
11.
H, Kalender
WA.
Single-
12.
13.
14.
15.
185
#{149} Number
spiral
M, Kalender volumetric CT with sin-
technique.
Radiology
16.
1990;
Kalender WA, Seissler W, Klotz E, Vock Spiral volumetric CT with single-breath-
P. 17.
hold technique, continuous transport, and continuous scanner rotation. Radiology 1990; 176:181-183. Costello P, Anderson W, Blume D. Spiral volumetric CT in the search for a pulmonary nodule (abstr). Radiology 1990; 177(P):295. Zimmerman PA, Gusnard DA. Craniocervical spiral CT (abstr). Radiology 1990; 177(P):12l. Bautz W, Skalej M, Kalender WA. Diagnosis of pancreatic disease: clinical impact of spiral CT (abstr). Radiology 1990; 177(P):
18.
19.
20.
1
Midorikawa Suzuki K
S, Hashimoto N, Katakura T, New approach to lung cancer with helical volume CT (abstr). Radiology 1990; 177(P):174. Kaneko M, Eguchi K, Ono R, et al Threedisplay
of bronchial
images
21.
by
scanning CT (abstr). Radiology 1990; 177(P):174. Marchal C, Rigauts H, Baert AL Improved contrast enhancement of the pancreas by spiral scanning (abstr). Eur Radiol 1991; l(S):22l. Rigauts H, Marchal C, Baert AL Cine mode display of images reconstructed out of spiral scans (abstr). Eur Radiol 1991; i(S): 192. Dupuy DE, Costello P, Ecker CP. Comparative study of spiral CT and conventional CT of the pancreas (abstr). Radiology 1991; i81(P):350. Costello P, Ecker CP, Tello RJ, Hartnell CC. Spiral CT assessment of coronary artery bypass grafts (abstr). Radiology 1991; 181(P):i50. Bautz W, Strotzer M, Lenz Kalender WA. Preoperative
M, Dittler evaluation
Fishman EK, Ney DR. KuhlmanJE, Cameron JL, Savader BL. Multidimensional display of spiral CT data: evaluation of the pancreas (abstr). Radiology 1991; 181(P): 259. Remy-Jardin M, Deffontaines C, RemyJ. Chest evaluation with use of spiral volumetric CT with the single-breath-hold technique (abstr). Radiology 1991; 181(P):273. Costello P, Dupuy DE, Ecker CP. Spiral CT of the thorax with small volumes of contrast material: a comparative study (abstr). Radiology 1991; 181(P):274. Kuhhnan JE, Ney DR, Fishman EK. Twoand three-dimensional imaging of the in vivo lung: combining spiral CT with multiplanar and volumetric rendering techniques (abstr). Radiology 1991; 181(P):274. Schafer M, Prokop M, Dohring W, Schafers F, Galanski M. Spiral CT of the tracheobronchial system: optimized technique and dinical applications (abstr). Radiology 1991; 181(P):274. Prokop M, Schafer CM, Dohring W, LaasJ, Nischelsky JE, Galanski M. Spiral CT for three-dimensional cular anatomy
helical
volumetric CT of the hepatobil(abstr). Radiology 1989; 173(P):
377.
Volume
Lung:
dimensional
reconstruction
without any interpolation or standard CT with the same patient dose. #{149}
2.
9.
for 180#{176} LI is
center of rotation is expected to increase by a factor of ‘/4/3 ( 1.15) when corn-
1.
WA.
M, Daepp
screening
Thus, the standard deviation of the pixel noise for 180#{176} LI interpolation in the
with
7.
results
u=lcrcr.
pared
P, Soucek
154.
f21 a(O)d0
the variance
6.
for recon-
struction without any interpolation P[0J = P1[0]), image pixel variance as in conventional CT: cTrr
5.
gives
finally
the vessels of the upper abdomen with spiral CT: comparison with conventional CT and arterial DSA (abstr). Radiology 1991;
CT of hilar
176:864-867.
0/ir)2]dO.
-
WA.
18i(P):261.
452. Vock
gle-breath-hold
=
cr
=
M, Kalender
adenopathy with 1-second and subsecond scan times (abstr). Radiology 1989; 173(P):
rr)
[(0/u)22(0
Oudkerk
imaging
of complex 1991;
vas-
(abstr). Radiology
181(P):293. 22.
23.
24.
25.
H, Kondo T, Katada T, Ida Y, Tsujioka K, Koga S. Three-dimensional coronary anglo-imaging with helical scanning CT (abstr). Radiology 1991; 181(P):336. Rigauts H, Marchal C, Baert AL, Hupke R. Spiral scanning and the influence of the reconstruction algorithm on image quality. J Comput Assist Tomogr 1990; 14:675-682. Crawford CR, King K. Computed tomography scanning with simultaneous patient translation. Med Phys 1990; i7:%7-982. Kalender WA, Polacin A. Physical performance characteristics of spiral CT scanning. Med Phys 1991; 18:910-915. Anno
HJ, of
Radiology
#{149} 35