Downloaded from www.ajronline.org by 58.42.232.6 on 11/04/15 from IP address 58.42.232.6. Copyright ARRS. For personal use only; all rights reserved

359

Improved Imaging Scan Equalization

J. C. Wandtke1 D. B. Plewes1’2

of Bone with Radiography

Use of scan equalization radiography (SER) for imaging bone in the head, face, neck, and shoulder was evaluated in a clinical comparison study with conventional radiographs of randomly selected patients. Two hundred nine pairs of normal and abnormal SER images and conventional radiographs were compared by four radiologists in a side-byside viewing situation. The radiologists compared the visibility of specific anatomic features and rated the SER images as better than, equal to, or worse than the conventional

radiographs.

much

preferred

detail

and

images

In the the

marked

SER

images

in detail

the radiologists

the conventional

of the images

of the

of the

63% of the time because

improvement

of the shoulder,

preferred

evaluation images

radiograph

preferred

soft

spine,

the

radiologists

of slight improvement

tissues.

In the

the SER image

only 5% of the time

(p

in bone

evaluation

58%

< .05).

of the

of the time and In the

evaluation

the radiologists preferred the SER images 62% of the time and the conventional images 4% of the time (p < .05). The superior image quality with the SER technique was recognized by all radiologists in the study and was the overwhelmingly preferred way of imaging the shoulder, neck, head, and face. AJR

of the skull, face,

of the

cervical

157:359-364,

August

and mandible,

1991

Conventional plain film radiography portions of the body that exhibit large

areas

of underexposure

is often variations

or overexposure

suboptimal when in tissue thickness.

on the radiograph.

used to image This results in

Nowhere

problem more evident than in conventional chest and bone radiography. radiography, the thickest part is optimally exposed but the thinner

overexposed. equalization of radiation

In response

to these

limitations

of conventional

is this

With bone areas are

radiography,

scan

radiography (SEA) was developed. SEA generates a nonuniform field on the patient that is modulated in response to transmission variations

of the particular positioning of the patient during film exposure [i -31. The application of SEA to chest radiography has been studied, and the results

Received August 9, 1 990; sion March 26, 1991.

This work was supported

accepted

after

revi-

by grants from the

National Cancer Institute (CA-27823) and the Eastman Kodak Co. 1 Department of Diagnostic Radiology, University

of Rochester Medical Center, 601 Elmwood Ave., Rochester, NY 14642. Address reprint requests to J.C. Wandtke.

are a significant improvement in image quality [4, 5] and diagnostic accuracy [6, 7]. Similar studies with another equalization radiography unit (AMBER) have shown improvement in chest radiography [8]. The problem of large variations in transmission through different body tissues is a common problem associated with

bone radiography also, especially radiography of the head, neck, shoulder, and spine. Most attempts to overcome these limitations have been studies of chest radiography with the use of specialized X-ray filters [9, i 0]. Other approaches include the use of image processing of digitized radiographs [i i , i 2], photostimulable storage phosphor digital radiographic systems [i 3, 14], and direct digital scanning

detector

North York, Ontario, Canada M4N 3M5.

Subjects

and Methods

0361-803X/91/1572-0359

In this investigation, the application mandible, and shoulder was studied.

2

Present address:

physics,

©

University

American

Department

of Toronto,

Roentgen

2075

Ray Society

of Medical BioBayview

Ave.,

systems

[1 5, 16].

of SEA to imaging This was evaluated

of the cervical spine, head, face, by a subjective assessment of

WANDTKE

360

AND

Downloaded from www.ajronline.org by 58.42.232.6 on 11/04/15 from IP address 58.42.232.6. Copyright ARRS. For personal use only; all rights reserved

image quality by a panel of radiologists who reviewed matched conventional and SEA images. The details of the configuration and the operation of the SEA system used in this study have been previously reported [i -31. This SEA system uses a 2.5- by 4.0-cm beam of radiation scanned in a

PLEWES

ventional

AJR:157,

and SEA radiographs.

Nonetheless,

August

the radiologists

1991

in that

study said that they preferred the side-by-side comparison. A sideby-side comparison permits easier selection of a preferred image when the differences are small.

series of 39 vertical lines. The X-ray output varies up to 2000 times

The radiologists were asked to compare the visibility of all anatomic features and overall image quality and then indicate on a five-point

during

scale whether

a 2-sec scan, and this is achieved

by computer-controlled was designed to be but was adjusted to

than

provide artifact-free

kVp with a maximum

with the patient’s

equalized

of 600 mA. The conventional

images at 80-90 radiographs

were exposed

radiation by using a standard phototimed technique, the total scanning time ranged posteroanterior,

lateral,

or oblique

views

with 75-kVp

system. For the SEA from i .2 to 4.5 sec for

depending

on the thickness screens

of the body part. Kodak T Mat-G film and Lanex regular were used for the conventional

radiograph.

This is the screen-film

system most often used in our department. Because this prototype SEA system is limited by tube output, a faster combination of Kodak T Mat-G film and Lanex fast screens was used for the SEA images.

The source-to-image distance for the SEA images was fixed at 180 cm, but the conventional bone radiographs were obtained at i 22 cm. The conventional radiographs were obtained with a wall-mounted Bucky grid assembly (grid ratio, 8:1). No grid was used for the SEA images because scatter rejection was achieved by a 3.0-cm slit mounted in the aft collimator assembly. All the comparison examinations were done in the upright position

better

the SEA images were much better than (+2), slightly

modulation of pulse width. The SEA system operated at 1 20 kVp for chest radiography,

than (+i), (-2)

the

equal to (0), slightly conventional

images.

worse The

than (-1), radiologists

age and sex but not the patient’s

or much worse were

provided

clinical history.

There were no time constraints for viewing the radiographs. Before the study, the comparative features of the SEA and conventional radiographs were illustrated in a 30-mm training session with four demonstration cases. A brief technical explanation of the equalization technique was also given. The training sessions consisted of individual meetings with verbal instructions and the opportunity for discussion.

Results The 175 patient examinations resulted in 209 pairs of comparison images on which each of four radiologists made 501 observations. A summary of the radiologists’ opinions of the neck, shoulder, and head shown in Table i The radiologists .

radiologic examinations is had a marked preference

because this was a design limitation of the SEA prototype. For the head and neck radiograph obtained with the SEA system, the patient was positioned as close to the film cassette as possible, but, because of the size of the aft collimator assembly, it was not feasible to place the film cassette in direct contact with the exit side of the body part examined; however, only about 5% magnification resulted. During some examinations of the cervical spine and shoulder, there were

for the SEA images. The average improvement in score (on the scale of -2 to +2) of the 2004 SEA images compared with conventional bone radiographs was +0.75. In 690 comparisons, no difference was seen. In i 233 comparisons, the SEA image was better, whereas in 8i comparisons the

slightly different degrees of magnification in positioning of the patients. Positioning

was a significant

SEA system aminations

as with the conventional were

examination. nologist

in a different

system. room

The comparison

5-1 5 mm after

In about half of the cases, a different

positioned

tional radiographs film density

done

owing to slight differences was as convenient with the

the patient

for the SEA

were obtained

and on-line

monitoring

examination.

The

ex-

the initial

radiologic

tech-

conven-

in the usual fashion with the usual of film quality

by a radiologist.

The

SEA system was set by using a phantom to provide an excellent image and was not readjusted during the study. Phantom studies have shown that for SEA head and shoulder examinations, radiation

exposure is less than for conventional elimination

of overexposure

around

studies. This is due to the

the thin edges

of these

anatomic

structures

and the film-screen used for SEA, which is twice as fast. Comparison SEA and conventional radiographs were collected from 175 volunteers (38-76 years old) during a 3-month period after informed consent was obtained. Three or four volunteer patients were recruited on each of 3 days a week. Patients were recruited without bias as to type or severity of bone disease. Most of the patients asked to volunteer did participate. A variety of abnormal

conditions suIted

shoulder

in

were included

as well as normal examinations.

209 pairs of comparison

radiographs

(65), and head (63). The comparison

within 1 5 mm of each other. The images were reviewed

This

re-

of the neck (81),

images were obtained

of the radiologists

Cervical

lateral,

by four

radiologists

about

the relative

quality

image was better. By chi-square (p < .05) improvement.

analysis,

SER

Spine

Cervical

spine

of 47 patients resulted in 47 and i 4 anteropostenior comparison ex-

examinations

20 oblique,

aminations. For each of these groups the radiologists were asked to comment on the clarity of the anatomy at the CiC2 level, at the C3-C7 level, and of the soft tissues. Table i

shows that for cervical spine examinations there was a preference for the SEA images, with 209 (22%) regarded as much better, 395 (41 %) as slightly better, and 333 (34%) as equivalent. Only 35 (3%) of the SER images were classified as inferior to the conventional images. Comparison of an SEA and conventional cervical spine examination is shown in Figure i , where all four radiologists preferred SEA. A markedly improved soft-tissue detail in the upper pharynx and larynx is usually seen. Most of the cases included in the study had normal soft tissue. The bones were frequently abnormal. The improved contrast in the bony structunes and better visibility of C7 was due to the adjustments in exposure

independently

possible

with

SEA.

in

viewing sessions that comprised approximately 50 comparison pairs each. Side-by-side comparison was used because the radiologist could not be blinded since it was obvious which radiograph was the conventional and which was the SEA. A previous side-by-side and independent comparison study [4] showed no significant difference in the opinions

conventional

of con-

Shoulders

For the 65 comparison shoulder examinations, the radiologists were asked to compare the visibility of all anatomic detail of the humerus, scapula, and clavicle. The SER images

AJR:157,

August

SCAN

1991

TABLE

1:

Ratings

of Scan

EQUALIZATION

Equalization

Radiography

BONE

RADIOGRAPHY

(SER)

and

36i

Conventional

Radiography

by

Four Radiologists Rating

Imaged

Region

+2

spine Anteroposterior

+i

0

-2

-i

Downloaded from www.ajronline.org by 58.42.232.6 on 11/04/15 from IP address 58.42.232.6. Copyright ARRS. For personal use only; all rights reserved

Cervical

(n

=

C1-C2 C3-C7

23 7

28 19

5 26

0 4

0 0

19

33

4

0

0

17 0 20

37 20 30

25 51 29

1 9 1

0 0 0

C1-C2 C3-C7

31 25

75 63

79 83

2 16

1 i

Soft tissues

67

90

3i

0

0

209

395

333

33

2

20 23 47

i07 131 1 27

117 95 73

i6 ii 12

0 0 1

90

365

285

39

1

Soft tissues Oblique (n = 20) Ci-C2 C3-C7 Soft tissues Lateral (n = 47)

Subtotal Shoulder (n Humerus

=

65)

Scapula

Clavicle

Subtotal Head

14)

(n = 63) Skull (n = i 1) Face (n = 22) Mandible (n = 30) Subtotal Total

9

19

i4

2

0

24

42

20

2

0

20

60

38

2

0

72

6

0

690

78

3

53 (n

=

209)

352

i2i 881

Note-A rating of +2 means SEA is much better than conventional radiography, +1 means SEA is slightly better than conventional, 0 means SEA is equal to conventional, -1 means SEA is slightly worse than conventional, and -2 means SEA is much worse than conventional.

Fig. 1.-A, of cervical

Conventional spine

vere degenerative B,

in a patient

radiograph with se-

arthritis.

Scan equalization radiograph shows better detail, especially of soft tissue, base of skull, and C7.

WANDTKE

Downloaded from www.ajronline.org by 58.42.232.6 on 11/04/15 from IP address 58.42.232.6. Copyright ARRS. For personal use only; all rights reserved

362

AND

were regarded as being either much better or slightly better than the conventional images 58% of the time. The SER images were equivalent 37% of the time, and 5% were found to be inferior. Examinations of the shoulder showed marked differences between the SEA and conventional radiographs. With the SER technique, the soft tissues were seen uniformly. In many cases, the bony structures were seen optimally without the use of a bright light. An example in which all four radiologists thought the antomic detail was better with SEA is shown in Figure 2. Head

Examinations

1).

of the head resulted

in improved

visibility

of

AJR:i57,

August 1991

anatomic detail when examinations were performed not only of the skull but also of the face and mandible. The uniformity of exposure resulted in improved contrast and visibility, especially around the edges of the calvaria. The underexposure commonly seen in the thicker parts of the face and skull also was reduced. This is seen well in an example of the anteroposterior examination of the face (Fig. 3). Table 2 is a summary of Table 1 , grouped so that it is easier to compare the ratings of visibility in terms of percentages. The preference of the individual radiologists for either the SER or conventional study is listed in Table 3; the categories of much better and slightly better and of much worse and slightly

The 63 radiographic examinations of the head area included 1 1 radiographs of the skull, 22 of the face, and 30 of the mandible. The SEA images were judged to be superior in i 74 (69%) of the images, equivalent in 72 (29%), and inferior in six (2%) (Table

PLEWES

worse

were

grouped.

Statistical

analysis

with the chi-

square method showed a significant difference (p < .05) for all categories and radiologists except one. Table 4 shows a summary of the data in Table 3, listing each radiologist’s preferences for the images of the head, neck, and shoulder. The five categories were grouped into three because not all the radiologists found it easy to differ-

Fig. 2.-A, Conventional radiograph of shoulder in patient with fracture of humeral head shows overexposure of acromioclavicular joint. B, Scan equalization radiograph shows better detail of soft tissue, ac-

romioclavicular humeral head.

joint, and lateral part of

Fig. 3.-A, Conventional radiograph of facein patientwith bilateral maxillary sinusitis. B, Scan equalization radiograph shows more uniform exposure and improved contrast of all bony and softtissue structures.

Downloaded from www.ajronline.org by 58.42.232.6 on 11/04/15 from IP address 58.42.232.6. Copyright ARRS. For personal use only; all rights reserved

AJA:157,

August

SCAN

1991

EQUALIZATION

BONE

entiate between slightly better and much better. Nonetheless, the difference between individual radiologists is still noticeable. The general trend was a preference for the SER images, ranging from 42% (radiologist 1) to 88% (radiologist 4) being judged as superior to the conventional images. Even though radiologists 1 and 4 had a different frequency of preference for the SER images, there was less of a difference in the opinion that the SEA was rarely worse (2-i 1 %) than the conventional square, p

image.

There

was

a significant

preference

7]. Although radiographs

TABLE

Equal Worse Mandible

deemed

so because

of slight

underexposure

TABLE Quality

that

is characteristic

2: Ratings of Visibility of Anatomic on Scan Equalization Radiography

Conventional

(SER)

Ratings Images

for Scan

Radiography

of all SEA

Worse Facial bones Better

Equal

Worse

or

Better

Equal Worse Scapula Better Equal Worse Clavicle/acromion

with

4 7 0

10 i 0

4 6 ia

iO 0 i

15 7

18 4

12

8

1

0

0

2

0

14 15

23 7

16 4

27 2

1

0

0

1

14 51 0

39 26 0

26 25 14

48 15 2

23 41 1

39 26 0

40 18 7

52 10 3

64 1 0

Better Equal

34

47

61

70

47

32

19

10

Worse

0

2

1

1

17 60 4

22 55 4

38 25 18

61 6 4

48 33 0

56 25 0

74 6 1

78 3 0

C3-C7 Equal

Worse

63

34

3

Better

Shoulder Skull

58 64 74 66

37 32 24 32

5 4 2 2

Equal Worse

Soft tissue

Note-Better

=

equal to conventional radiography. a Difference

4

Detail and Image Radiographs

SER better than conventional radiography; equal = SEA radiography; worse = SEA worse than conventional

not statistically

Quality

significant.

by Individual

No. (%) of Images by Radiologis

Radiologists

t

No.

SEA Rating Better

than

conventional

radiography

Equal to conventional radiography Worse than conventional radiography

21

27 ii

Better

TABLE 4: Ratings of Visibility of Anatomic on Scan Equalization (SER) vs Conventional

4

27

Cervical spine

34

3

24 2

Worse

62

2

39

Equal

Average

No.

23 0

Better

Mandible

Radiologists

42

Co nventional Study ( % of Images)

Face

consist-

Equal Worse Ci-C2

Better

[4,

by Region

Region

the

Humerus

Detail and Image (SER) vs

Quality o f SEA Compared

on conventional light,

Equalization

by Individual

Better Equal

differences in the positioning of the patients. Some of those SEA examinations might have been repeated, but the protocol did not permit it. Some of the comparison conventional radiographic examinations were repeated. The frequency of repeat examinations was not recorded. We think that the primary value of the SEA image is the consistency of image quality uniformity

a bright

Radiologist

The capability of SEA to provide a more uniform exposure over the entire radiograph, regardless of variations in body tissue X-ray transmission, results in improved visibility of normal anatomic detail as judged by this subjective analysis by four radiologists. Despite the faster film-screen combination and higher-kilovoltage X-ray spectrum used for the SEA images, the SEA image quality was regarded as equivalent (34%) or superior (62%), with only a very few cases (4%) regarded as inferior. Most of the 4% of SER cases that were

and exposure

with

significance. Skull Better

were

that is overexposed

be visualized

3: Preference

Radiographic

Discussion

inferior

anatomy can

(chi

Improved imaging of bone with scan equalization radiography.

Use of scan equalization radiography (SER) for imaging bone in the head, face, neck, and shoulder was evaluated in a clinical comparison study with co...
909KB Sizes 0 Downloads 0 Views