European Heart Journal Advance Access published March 5, 2015

REVIEW

European Heart Journal doi:10.1093/eurheartj/ehv060

Infected cardiac-implantable electronic devices: prevention, diagnosis, and treatment Jens Cosedis Nielsen 1*, Jens Christian Gerdes 1, and Niraj Varma 2 1 Department of Cardiology, Aarhus University Hospital, Skejby, Palle Juul-Jensens Boulevard 99, Aarhus N 8200, Denmark; and 2Department of Cardiovascular Medicine, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH, USA

Received 15 December 2014; revised 18 February 2015; accepted 23 February 2015

Cardiac implantable electronic device (CIED) infection, according to current trends, appears to be an increasing problem. It can be indolent and its diagnosis challenging. Cardiac implantable electronic device infections are potentially lethal, and timely diagnosis and early initiation of correct treatment are of highest importance for patient prognosis. For reducing CIED infections, careful patient selection, preventative measures, and appropriate choice of device are key. The current review presents available data and consensus opinion within the field of CIED infection and identifies important current practice points and aspects for future development. Strategies for reducing CIED infection should be tested in sufficiently powered and well-designed multicentre randomized controlled trials.

----------------------------------------------------------------------------------------------------------------------------------------------------------Keywords

Cardiac implantable electronic device † Infection † Prevention † Diagnosis † Treatment

Cardiac implantable electronic device (CIED) therapy is safe and effective. Complications are infrequent, but among them, infection poses a major hazard, causing significant morbidity, increasing mortality, and generating considerable healthcare costs.1 – 4 Its incidence is growing, only partly explained by increasing numbers of CIED implants resulting from widening indications and increasing numbers of generator replacements. The absolute rate of device infection itself increased recently.2,5,6 Causes for this trend are uncertain, but likely relate to increasing proportion of implantable cardioverter defibrillator (ICD) and cardiac resynchronization therapy (CRT) devices implanted,2,7 coupled to implantations in ‘higher risk’ candidates, that is, patients with heart failure, diabetes, and renal failure.2 Current knowledge is mainly based upon observational data. While no international guidelines exists for handling infected CIED’s, an expert consensus statement on lead extractions,8 an American Heart Association scientific statement9 and a recent British guideline paper10 are useful for guiding decisions. This review aims to present available data and consensus opinion within the field of CIED infection and identify important current practice points and aspects for future development.

Incidence of cardiac implantable electronic device infection Reported CIED infection rates vary according to definitions and follow-up duration. In a large cohort of pacemaker recipients,

estimated incidence was 1.82/1000 device-years and higher within the first 12 months.11 Infection risk after pacemaker implant is 0.5 – 1% within the first 6–12 months.11 – 13 The more complex CIED system implanted the higher infection risk. Risk is higher with ICDs,14 recently estimated 1.7% within 6 months among 200 000 ICD recipients,3 even higher in CRT-recipients, 9.5% over 2 years15 and among these highest with defibrillators (CRT-D).16 Infection risk is higher for devices implanted at thoracotomy and may be 2 –4× greater after device replacement and upgrades compared with primary implants.3,11 – 13,17

Pathogenesis and presentation Breach of the skin barrier introduces bacteria into the pocket. These may colonize and not always cause clinically relevant infection. Pathogenesis of CIED infections is associated with biofilm formation on device surfaces, described as ‘a structured community of bacterial cells enclosed in a self-produced polymeric matrix and adherent to an inert or living surface’.18 Bacteria in biofilm are protected from killing by host defences and antimicrobial agents. Therefore, entire device removal is needed to treat infection. Most infections are caused by bacteria, which are part of the patient’s normal skin flora. Gram-positive species are most frequent, usually Staphylococcus species, less commonly Gram-negative bacteria. Fungi, anaerobes, or polymicrobial infections are rare.14,19 Cultures are negative in 15%. Pocket infection is the most common presentation of CIED infection.15,19,20 This should not be assumed localized since infection commonly tracks along leads and/or causes secondary

* Corresponding author. Tel: +45 78452039, Email: [email protected]; [email protected] Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2015. For permissions please email: [email protected].

Downloaded from by guest on May 31, 2015

Introduction

Page 2 of 7 blood-stream infection and endocarditis. Less commonly, the mechanism of blood-stream infection is haematogenous spread of bacteria from another infection site with secondary involvement of intravascular CIED parts.

Prevention of cardiac implantable electronic device infection Cardiac implantable electronic device infection is serious for the patient, potentially fatal, and virtually always necessitates complete CIED system removal. The most important issue within this field therefore is to prevent CIED infections.

Before implantation

Table 1 Patient-, device-, and intervention-related factors associated with increased cardiac implantable electronic device infection risk Early CIED re-intervention3,12

2.7–15

Pocket haematoma3,30 CIED replacement11

4.0–6.7 2.7

More complex CIED3,11,20

1.3–5.4

Temporary pacing12,23 Prior device procedure11

2.5–5.0 2.7–8.7

Prior CIED infection23

11.3

Fever/systemic infection12 Renal failure (no dialysis)3,23

5.8 1.3

Haemodialysis19

8.6

Chronic skin disease23 Corticosteroid treatment20

10.6 13.9

COPD3,23

1.2–9.8

Diabetes7 Higher Charlson comorbidity index23

2.3 2.7–3.0

Male gender11

1.5

Younger age11 Low body-mass index

1.4–4.5 NA

Heart failure

NA

Indwelling lines Anticoagulation/antithrombotic drugs

NA NA

Longer procedure time

NA

Low operator experience

NA

Peri-operatively General recommendations for reducing surgical site infections should be applied. It is recommended to use appropriate antiseptic agents for skin preparation pre-operatively to minimize burden of microorganisms in normal skin flora.26 Comparative efficacy of chlorhexidine-alcohol vs. povidone-iodine for clean-contaminated surgery has yielded mixed results.7,27 Chlorhexidine-alcohol reduces surgical site infections following clean surgery compared with alcohol-based povidone-iodine.22,28 Use of transparent films or diathermia or substances preventing bleeding have not been proven beneficial.10 Pre-operative antibiotics are mandatory for CIED procedures. Observational data indicate reduction in CIED infections to less than half with one single dose of pre-operatively intravenously administered antibiotics,11,12,20 and are supported by one meta-analysis (OR 0.26 with antibiotics).29 One randomized controlled trial confirmed these findings—risk of CIED infection within 6 months was reduced to 0.64% using single-dose cefalozin vs. 3.28% with placebo (P ¼ 0.016).30 Topical antibiotics are often used additionally, but lacks supporting evidence. Antibiotic regime differs widely with respect to type of antibiotics, treatment duration, and topical antibiotics. There is no documentation that one specific regime is superior. Most implanters use one dose of antibiotics 60–90 min pre-operatively. Antibiotic choice is influenced by local burden of methicillin-resistant Staphylococci that is increasing worldwide. An ongoing randomized cluster

Downloaded from by guest on May 31, 2015

Physicians should prescribe CIED treatment carefully. A thorough evaluation of its indication is mandatory for each patient with a balanced assessment of risks and benefits, accounting for individual patient characteristics and comorbidities.21 Lower infection risk with simpler CIED systems should be weighed against potential benefits of more complex systems. Use of smaller devices and shorter leads in asthenic patients reducing pocket bulk, or sub-pectoral implants to prevent erosions should be considered, and attention given to battery longevity to reduce anticipated generator replacements. Need for device upgrade should be evaluated extremely carefully, given the high infection risk.22 Meticulous attention to pre-operative preparation is necessary (Table 1). Fever ,24 h before implant is associated with 5.8 times higher infection risk.12 It is therefore important to diagnose and treat ongoing infections before CIED implant. Patients presenting with need for acute cardiac pacing are challenging. It is unsettled whether the best management strategy is temporary transvenous pacing and treatment of infection, thus postponing permanent CIED implantation, or proceeding to primary permanent CIED implantation with concomitant treatment, particularly with ‘low-risk’ conditions such as urinary tract colonization. Most commonly, the first of these strategies is chosen to reduce potential infection of the permanent device. However, temporary pacing itself is associated with increased risk of CIED infection (odds ratio, OR 2.512,23) and should be performed under sterile conditions. Its duration should be minimized since longer hospital stay awaiting permanent CIED implantation is associated with significant complication risk.23,24 Chronic skin conditions increase infection risk23 and should be appropriately treated to avoid skin defects especially around implantation site. Telemetry electrodes should not be mounting in an area of anticipated implantation to prevent skin reactions and defects. Indwelling lines, for example, central venous catheters and chest tubes should be removed in timely fashion before CIED implantation. Ideal timeframe is unknown, most operators prefer .24 h. Patients who are nasal carriers of Staphylococcus aureus (20–30% of population) are at increased risk of health care associated infections with this organism. Recently, a randomized controlled trial documented that screening and decolonizing of nasal carriers of these bacteria reduces number of surgical site infections (relative risk 0.42) among patients admitted to hospital for surgery.25 Whether such a strategy reduces CIED infection is unknown. No data support screening and treatment of health personnel.

J.C. Nielsen et al.

Infected cardiac implantable electronic devices

The high-risk patient Several patient-related factors are associated with heightened risk of CIED infection, such as gender, age, body-mass index, renal failure and dialysis, heart failure, diabetes (especially with hyperglycaemia), respiratory failure, corticosteroid treatment, chronic skin conditions, indwelling lines, and higher comorbidity index (Table 1). Since these conditions are largely irreversible/unavoidable, attention should focus on optimization before CIED implantation. Conceivably, this patient group may benefit most from innovations as leadless pacemakers and subcutaneous ICD systems. Leadless pacemakers37 are likely to bear lower risk of device infection than systems used today. However, long-term efficacy and safety that is at least noninferior to traditional pacemakers needs confirmation in randomized controlled trials before advocating for their widespread utilization. In event of their infection, extraction risk is unknown. Subcutaneous

ICD systems using a subcutaneous shock electrode was introduced.38 It is unlikely that pocket infection rate is much lower than for traditional pacemakers. However, this system does not bear the risk of blood-stream infection or device endocarditis and thus could be attractive for patients not needing bradycardia or antitachycardia pacing and at particularly high risk of CIED infection, for example, renal failure patients treated with haemodialysis. Again, documentation of at least non-inferiority with respect to efficacy and safety as compared with transvenous ICD systems is needed before expanding this treatment to larger patient groups. Extraction risk is however negligible.

Diagnosis of cardiac implantable electronic device infection There are intra- and extra-cardiac components. Diagnosis may be difficult and a high index of clinical suspicion is warranted. Presentation demonstrates a wide spectrum from subtle complaints of mild pain without pocket abnormalities, device migration occurring years after implant, erosion, to a hot, red pocket, and septic shock. Unidentified or partially treated low-grade chronic sepsis can result in multisystem disease processes. Cardiac implantable electronic device infection commonly presents as local infection around the device. Such pocket infection usually occurs within 12 months post-implant, but may occur later. Blood-stream infection is less common, may appear with or without concomitant pocket infection, and with or without valve endocarditis.39 Symptoms and blood tests may be misleadingly normal. Bacteraemia may occur during the entire CIED lifetime. Blood-stream infection is particularly common among patients with end-stage renal failure and haemodialysis.19,40 Any symptom from the device pocket should raise suspicion of CIED pocket infection, and the patient referred to a CIED specialist for evaluation (Figure 1). When using remote monitoring for device follow-up, educating patients about this topic is especially important. In particular, this form of follow-up, if initiated at implant, does not obviate need for standard 2-week wound check and 3-month in person evaluation post-implant.41 Typical signs are local erythema, warmth, pain, and swelling, adherence of skin to device, and incipient or overt erosion of skin with a draining sinus.42 Erosion is de facto infected. Early post-operatively it may be difficult to differentiate between CIED pocket infection and superficial wound infection. Diagnostic percutaneous puncture with pocket fluid aspiration should be avoided. In CIED patients developing systemic infection without any obvious focus, blood-stream infection affecting the CIED should be suspected, and repeated blood cultures obtained. At least two sets of blood cultures should be obtained before starting empiric antibiotic treatment. Growth of bacteria typical for CIED infections, Staphylococcal species, Corynebacteria or Propionobacteria, support the diagnosis. If suspecting CIED infection, echocardiography is indicated. Transoesophageal imaging more accurately visualizes leads, valvular involvement, and vegetation. However, echocardiography cannot discriminate infective vegetation from non-infected thrombotic or fibrous masses, often prevalent on chronic indwelling leads.43 Absence of CIED lead vegetation does not rule out CIED blood-stream infection. Occult Gram-positive

Downloaded from by guest on May 31, 2015

crossover trial enrolling 10 800 patients compares single-dose pre-operative antibiotics (cefazolin or vancomycin) with an incremental antibiotic strategy adding intra-operative wound pocket wash (bacitracin) and 2 days post-operative cephalexin/cephadroxil/ clindamycin.31 The TYRXTM Envelope (Medtronic, Monmouth Junction, NJ, USA) is an antibacterial envelope made from polyprolene mesh releasing the antibiotics minocycline and rifampin in the generator pocket after CIED implantation. This envelope eliminates Staphylococcal species and prevents biofilm formation on implanted pacing devices in an animal model,32 and reduces CIED infections in high-risk patients in observational studies.33 – 35 Recently a bio-absorbable version, disappearing within 9 weeks after implantation was introduced. This will be evaluated in WRAP-IT (worldwide randomized antibiotic envelope infection prevention trial, NCT02277990, commenced 2015) aiming to enrol 7.764 patients scheduled for CIED replacement or upgrade procedure, primary CRT-D implantation or late lead or pocket revision. Primary endpoint is major CIED infection within 12 months. Meticulous surgical technique to prevent tissue damage, assure haemostasis, and ensure secure subcutaneous cover for the device should be standard. Shorter procedure times was associated with lower infection risk.16,30 Cephalic or axillary vein access is recommended for lead implantations, to better preserve long-term lead integrity and avoid re-operations. During device replacement procedures, some authors support capsulectomy. No controlled studies have tested this strategy. Benefit of removing avascular tissue must be weighed against risk of haematoma, associated with higher infection rates.16,22,30 Clopidogrel and aspirin increases risk threefold. Among anticoagulated patients, continued warfarin reduced risk of significant haematoma after CIED implantation markedly (OR 0.19) as compared with heparin bridging.36 How to handle patients treated with new oral anticoagulants is less clear. When hematoma occurs, management must be considered carefully. Early re-intervention probably is the strongest risk factor for later CIED infection.3,12,16,17,23 Pocket re-entry should be avoided unless haematoma is particularly tense or painful. Large haematomas gradually soften and resorb over few weeks. Appropriate education and a reasonable implant volume for each operator should be assured to decrease risk of lead dislodgements and haematomas requiring early re-interventions.13

Page 3 of 7

Page 4 of 7

J.C. Nielsen et al.

bacteraemia is indicative for CIED blood-stream infection and Class 1 indication for explantation.8 Systemic symptoms may warrant pulmonary or brain imaging for septic emboli. Recently, fluorine-18 marked fluorodesoxyglucose (18F-FDG) positron emission tomography and computed tomography was found useful for differentiating patients with CIED infection from those without.44 Yet experience is limited, and further validation of this method is needed.

Treatment of cardiac implantable electronic device infection Confirmation of CIED infection—systemic or localized to pocket— mandates removal of all hardware (generator and leads), antibiotic therapy, and a strategy for re-implant. Only exception to this rule is for minor incisional/suture abscesses few days after implantation, not communicating with the pocket, which may be treated with antibiotics and careful follow-up. Timely, correct and complete treatment are of highest importance for patient prognosis. Delay further increases morbidity and mortality. Partial procedures, for example,

generator removal and capping of leads are associated with almost invariable relapse,20,42,45,46 even if presentation is pocket infection. Percutaneous extraction is favoured though associated with a small risk of major complications, particularly vascular lacerations causing haemothorax and death. Mortality rates are 0.3 –0.8%.47 Given this, risk –benefit ratio favours extraction, which should be performed in centres with procedural volume sufficient to maintain operator skills and capacity for immediate surgical backup. Primary surgical extraction may be considered with failed/incomplete percutaneous extraction, epicardial leads, and valvular endocarditis demanding surgical repair. A hybrid approach is sometime necessary.48

Pre-operative evaluation Adequate inter-disciplinary preparation is mandatory. Infection disease specialists may direct antibiotic therapy. This may be commenced before system explant after blood and incisional drainage cultures. Septic shock requires immediate explant. There are few data regarding course of antibiotic therapy. Generally, courses of 10 –14 days after pocket infection, 14 days for bacteraemia, and 4–6 weeks for endocarditis are used.9 When long-term intravenous antibiotics are needed, venous access sites not interfering with future

Downloaded from by guest on May 31, 2015

Figure 1 Indolent swelling of pocket (A), adherence between skin and device (B), imminent perforation (C), and overt erosion (D), all representing pocket infection.

Page 5 of 7

Infected cardiac implantable electronic devices

implantations should be considered. Temporary pacing needs careful planning. Following extraction, we use a ‘permanent’ active fixation endocardial right ventricular pacing electrode. Externally, this is anchored to sterilized skin using an anchoring sleeve, covered, and connected to a generator, which may be affixed externally to the skin of the shoulder or chest at a convenient location. Heart failure teams should be involved when anticipating haemodynamic support after CRT device removal. All hardware, including abandoned leads, even if contralateral to infected system, must be removed. Risks of explant increase with older leads and superior vena cava coils. A chest X-ray is important. Imaging may discover unreported leads, leads jailed by stents, or unusual trajectories suggestive of extravascular or even arterial courses.49 This may require contrast CT-scan for visualization of, for example, perforated tips of endocardial leads, epicardial leads, and loculated pericardial collections, which may affect choice of approach. Transoesophageal echocardiography is an important part of evaluation. Large vegetation particularly when adherent to the tricuspid valve must be assumed infected, and may indicate surgical explant (Figure 2). Traditionally, transvenous extraction is avoided for vegetations .2 cm.8 Most are friable, and even at open surgical explant tend to disintegrate on handling and cannot be removed en masse. These

disintegrate during transvenous extraction and seldom cause pulmonary artery obstruction. They may be visualized with intra-cardiac echo (Figure 3). However, dispersion may cause sepsis syndrome and hypotension. A thorough evaluation of the individual patient’s condition, risks, and benefits must be taken into account. For example, in someone very fragile or with poor prognosis due to overwhelming comorbidities presenting with localized pocket infection of a 10-year-old dual coil ICD system, best management may be pocket debridement and commitment to chronic antibiotic therapy, or tailored salvage therapy.50

Operative procedure Extraction methodology is beyond this article’s scope. However, surgical principles need to be followed with debridement of necrotic-infected tissue and drainage of purulent abscesses. This includes pocket capsulectomy with removal of avascular tissues. Lead tip and pocket culture are important.19,46

Re-implantation Firstly, the original indication needs re-evaluation. Some arrhythmia problems may resolve,19 for example, post-operative atrioventricular Downloaded from by guest on May 31, 2015

Figure 2 Transoesophageal echocardiography and operative photograph showing densely adherent tissue to lead and tricuspid valve requiring surgical debridement.

Figure 3 Large vegetations visualized with intra-cardiac echocardiography.

Page 6 of 7 block or sinus node disease, as confirmed by 0% pacing on explanted device diagnostics. Alternatively, indication for a more complex CIED may have developed, for example, need for CRT in a patient with previous single-lead device. Other comorbidities may have overtaken the risk from cardiovascular disease, for example, malignancy in patients with prophylactic ICD, or significant renal disease.21 In such cases, there may be merit in not re-implanting at all. When re-implantation is necessary, this should be done on opposite side of the chest, or via transiliac approach. There are no randomized trials guiding appropriate timing of re-implantation. Carefully attention to an adequate period of antibiotic therapy and indication for CIED is needed. Decisions must be individualized. Epicardial placement of pacemaker leads may be considered for those at high risk of re-infection or with limited vascular access, for example, dialysis patients. A continued dialogue, commencing pre-operatively, among electrophysiologist, surgeon, and infection disease specialist, is critical to ensuring individual management plans.

Conclusion

J.C. Nielsen et al.

9.

10.

11.

12.

17.

Conflict of interest: none declared.

18.

References

19.

1. Sohail MR, Henrikson CA, Braid-Forbes MJ, Forbes KF, Lerner DJ. Mortality and cost associated with cardiovascular implantable electronic device infections. Arch Intern Med 2011;171:1821 –1828. 2. Greenspon AJ, Patel JD, Lau E, Ochoa JA, Frisch DR, Ho RT, Pavri BB, Kurtz SM. 16-year trends in the infection burden for pacemakers and implantable cardioverterdefibrillators in the United States 1993 to 2008. J Am Coll Cardiol 2011;58: 1001 –1006. 3. Prutkin JM, Reynolds MR, Bao H, Curtis JP, Al-Khatib SM, Aggarwal S, Uslan DZ. Rates of and factors associated with infection in 200 909 Medicare implantable cardioverter-defibrillator implants: results from the national cardiovascular data registry. Circulation 2014;130:1037 –1043. 4. Ferguson TB Jr, Ferguson CL, Crites K, Crimmins-Reda P. The additional hospital costs generated in the management of complications of pacemaker and defibrillator implantations. J Thorac Cardiovasc Surg 1996;111:742 –751; discussion 751 –742. 5. Cabell CH, Heidenreich PA, Chu VH, Moore CM, Stryjewski ME, Corey GR, Fowler VG Jr. Increasing rates of cardiac device infections among Medicare beneficiaries: 1990 –1999. Am Heart J 2004;147:582 –586. 6. Voigt A, Shalaby A, Saba S. Rising rates of cardiac rhythm management device infections in the United States: 1996 through 2003. J Am Coll Cardiol 2006;48:590 –591. 7. Qintar M, Zardkoohi O, Hammadah M, Hsu A, Wazni O, Wilkoff BL, Tarakji KG. The Impact of Changing Antiseptic Skin Preparation Agent used for Cardiac Implantable Electronic Device (CIED) Procedures on the Risk of Infection. Pacing Clin Electrophysiol 2015;38:240 –246. 8. Wilkoff BL, Love CJ, Byrd CL, Bongiorni MG, Carrillo RG, Crossley GH III, Epstein LM, Friedman RA, Kennergren CE, Mitkowski P, Schaerf RH, Wazni OM,

13.

14.

15.

16.

20.

21. 22.

23.

24. 25.

26.

Downloaded from by guest on May 31, 2015

Cardiac implantable electronic device infection, according to current trends, appears to be an increasing problem. It can be indolent and its diagnosis challenging, and potentially lethal if left untreated. Its treatment—explantation and eradicative antibiotic therapy—may be challenging and not without risk. These include risks associated with device removal and re-implantation, adverse reactions to antibiotics and generation of resistant bacteria, and risks of long-term vascular access. Given this backdrop, careful patient selection, preventative measures, and appropriate choice of device, are key. Future strategies for reducing CIED infection should be tested in sufficiently powered and well-designed multicentre randomized controlled trials.

Heart Rhythm S, American Heart A. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm 2009;6:1085 –1104. Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, Masoudi FA, Okum EJ, Wilson WR, Beerman LB, Bolger AF, Estes NA 3rd, Gewitz M, Newburger JW, Schron EB, Taubert KA, American Heart Association Rheumatic Fever E, Kawasaki Disease C, Council on Cardiovascular Disease in Y, Council on Cardiovascular S, Anesthesia, Council on Cardiovascular N, Council on Clinical C, Interdisciplinary Council on Quality of C, American Heart A. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation 2010; 121:458 – 477. Sandoe JA, Barlow G, Chambers JB, Gammage M, Guleri A, Howard P, Olson E, Perry JD, Prendergast BD, Spry MJ, Steeds RP, Tayebjee MH, Watkin R. Guidelines for the diagnosis, prevention and management of implantable cardiac electronic device infection. Report of a joint Working Party project on behalf of the British Society for Antimicrobial Chemotherapy (BSAC, host organization), British Heart Rhythm Society (BHRS), British Cardiovascular Society (BCS), British Heart Valve Society (BHVS) and British Society for Echocardiography (BSE). J Antimicrob Chemother 2015;70:325 –359. Johansen JB, Jorgensen OD, Moller M, Arnsbo P, Mortensen PT, Nielsen JC. Infection after pacemaker implantation: infection rates and risk factors associated with infection in a population-based cohort study of 46299 consecutive patients. Eur Heart J 2011;32:991 –998. Klug D, Balde M, Pavin D, Hidden-Lucet F, Clementy J, Sadoul N, Rey JL, Lande G, Lazarus A, Victor J, Barnay C, Grandbastien B, Kacet S, Group PS. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation 2007;116:1349 –1355. Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J 2014;35:1186 –1194. Uslan DZ, Sohail MR, St Sauver JL, Friedman PA, Hayes DL, Stoner SM, Wilson WR, Steckelberg JM, Baddour LM. Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study. Arch Intern Med 2007;167:669 –675. Palmisano P, Accogli M, Zaccaria M, Luzzi G, Nacci F, Anaclerio M, Favale S. Rate, causes, and impact on patient outcome of implantable device complications requiring surgical revision: large population survey from two centres in Italy. Europace 2013; 15:531 – 540. Romeyer-Bouchard C, Da Costa A, Dauphinot V, Messier M, Bisch L, Samuel B, Lafond P, Ricci P, Isaaz K. Prevalence and risk factors related to infections of cardiac resynchronization therapy devices. Eur Heart J 2010;31:203 –210. Nery PB, Fernandes R, Nair GM, Sumner GL, Ribas CS, Menon SM, Wang X, Krahn AD, Morillo CA, Connolly SJ, Healey JS. Device-related infection among patients with pacemakers and implantable defibrillators: incidence, risk factors, and consequences. J Cardiovasc Electrophysiol 2010;21:786 – 790. del Pozo JL, Patel R. The challenge of treating biofilm-associated bacterial infections. Clin Pharmacol Ther 2007;82:204–209. Tarakji KG, Chan EJ, Cantillon DJ, Doonan AL, Hu T, Schmitt S, Fraser TG, Kim A, Gordon SM, Wilkoff BL. Cardiac implantable electronic device infections: presentation, management, and patient outcomes. Heart Rhythm 2010;7:1043 –1047. Sohail MR, Uslan DZ, Khan AH, Friedman PA, Hayes DL, Wilson WR, Steckelberg JM, Stoner SM, Baddour LM. Risk factor analysis of permanent pacemaker infection. Clin Infect Dis 2007;45:166 –173. Patton KK, Poole JE. Defibrillator therapy in chronic kidney disease: is it worth it? Circ Arrhythm Electrophysiol 2014;7:774 – 776. Uslan DZ, Gleva MJ, Warren DK, Mela T, Chung MK, Gottipaty V, Borge R, Dan D, Shinn T, Mitchell K, Holcomb RG, Poole JE. Cardiovascular implantable electronic device replacement infections and prevention: results from the REPLACE Registry. Pacing Clin Electrophysiol 2012;35:81– 87. Sohail MR, Hussain S, Le KY, Dib C, Lohse CM, Friedman PA, Hayes DL, Uslan DZ, Wilson WR, Steckelberg JM, Baddour LM, Mayo Cardiovascular Infections Study G. Risk factors associated with early- versus late-onset implantable cardioverterdefibrillator infections. J Interv Card Electrophysiol 2011;31:171 – 183. Risgaard B, Elming H, Jensen GV, Johansen JB, Toft JC. Waiting for a pacemaker: is it dangerous? Europace 2012;14:975 –980. Bode LG, Kluytmans JA, Wertheim HF, Bogaers D, Vandenbroucke-Grauls CM, Roosendaal R, Troelstra A, Box AT, Voss A, van der Tweel I, van Belkum A, Verbrugh HA, Vos MC. Preventing surgical-site infections in nasal carriers of Staphylococcus aureus. N Engl J Med 2010;362:9 –17. Mangram AJ, Horan TC, Pearson ML, Silver LC, Jarvis WR. Guideline for Prevention of Surgical Site Infection, 1999. Centers for Disease Control and Prevention (CDC) Hospital Infection Control Practices Advisory Committee. Am J Infect Control 1999; 27:97 –132; quiz 133 –134; discussion 196.

Infected cardiac implantable electronic devices

39. Murdoch DR, Corey GR, Hoen B, Miro JM, Fowler VG Jr, Bayer AS, Karchmer AW, Olaison L, Pappas PA, Moreillon P, Chambers ST, Chu VH, Falco V, Holland DJ, Jones P, Klein JL, Raymond NJ, Read KM, Tripodi MF, Utili R, Wang A, Woods CW, Cabell CH, International Collaboration on Endocarditis-Prospective Cohort Study I. Clinical presentation, etiology, and outcome of infective endocarditis in the 21st century: the International Collaboration on Endocarditis-Prospective Cohort Study. Arch Intern Med 2009;169:463 – 473. 40. Hickson LJ, Gooden JY, Le KY, Baddour LM, Friedman PA, Hayes DL, Wilson WR, Steckelberg JM, Sohail MR. Clinical presentation and outcomes of cardiovascular implantable electronic device infections in hemodialysis patients. Am J Kidney Dis 2014;64:104 –110. 41. Varma N, Michalski J, Epstein AE, Schweikert R. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circ Arrhythm Electrophysiol 2010;3:428 –436. 42. Chua JD, Wilkoff BL, Lee I, Juratli N, Longworth DL, Gordon SM. Diagnosis and management of infections involving implantable electrophysiologic cardiac devices. Ann Intern Med 2000;133:604 – 608. 43. Downey BC, Juselius WE, Pandian NG, Estes NA 3rd, Link MS. Incidence and significance of pacemaker and implantable cardioverter-defibrillator lead masses discovered during transesophageal echocardiography. Pacing Clin Electrophysiol 2011;34: 679 –683. 44. Sarrazin JF, Philippon F, Tessier M, Guimond J, Molin F, Champagne J, Nault I, Blier L, Nadeau M, Charbonneau L, Trottier M, O’Hara G. Usefulness of fluorine-18 positron emission tomography/computed tomography for identification of cardiovascular implantable electronic device infections. J Am Coll Cardiol 2012;59:1616 –1625. 45. Pichlmaier M, Knigina L, Kutschka I, Bara C, Oswald H, Klein G, Bisdas T, Haverich A. Complete removal as a routine treatment for any cardiovascular implantable electronic device-associated infection. J Thorac Cardiovasc Surg 2011;142:1482 –1490. 46. Klug D, Wallet F, Lacroix D, Marquie C, Kouakam C, Kacet S, Courcol R. Local symptoms at the site of pacemaker implantation indicate latent systemic infection. Heart 2004;90:882 –886. 47. Wazni O, Epstein LM, Carrillo RG, Love C, Adler SW, Riggio DW, Karim SS, Bashir J, Greenspon AJ, DiMarco JP, Cooper JM, Onufer JR, Ellenbogen KA, Kutalek SP, Dentry-Mabry S, Ervin CM, Wilkoff BL. Lead extraction in the contemporary setting: the LExICon study: an observational retrospective study of consecutive laser lead extractions. J Am Coll Cardiol 2010;55:579 –586. 48. Varma NJ, Sellke FW, Epstein LM. Chronic atrial lead explantation using a staged percutaneous laser and open surgical approach. Pacing Clin Electrophysiol 1998;21: 1483 –1485. 49. Baranowski B, Wazni O, Chung R, Martin DO, Rickard J, Tanaka-Esposito C, Bassiouny M, Wilkoff BL. Percutaneous extraction of stented device leads. Heart Rhythm 2012;9:723 – 727. 50. Lopez JA. Conservative management of infected pacemaker and implantable defibrillator sites with a closed antimicrobial irrigation system. Europace 2013;15: 541 –545.

Downloaded from by guest on May 31, 2015

27. Darouiche RO, Wall MJ Jr, Itani KM, Otterson MF, Webb AL, Carrick MM, Miller HJ, Awad SS, Crosby CT, Mosier MC, Alsharif A, Berger DH. Chlorhexidine-Alcohol versus Povidone-Iodine for Surgical-Site Antisepsis. N Engl J Med 2010;362:18– 26. 28. Dumville JC, McFarlane E, Edwards P, Lipp A, Holmes A. Preoperative skin antiseptics for preventing surgical wound infections after clean surgery. Cochrane Database Syst Rev 2013;3:CD003949. 29. Da Costa A, Kirkorian G, Cucherat M, Delahaye F, Chevalier P, Cerisier A, Isaaz K, Touboul P. Antibiotic prophylaxis for permanent pacemaker implantation: a meta-analysis. Circulation 1998;97:1796 – 1801. 30. de Oliveira JC, Martinelli M, Nishioka SA, Varejao T, Uipe D, Pedrosa AA, Costa R, D’Avila A, Danik SB. Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial. Circ Arrhythm Electrophysiol 2009; 2:29–34. 31. Connolly SJ, Philippon F, Longtin Y, Casanova A, Birnie DH, Exner DV, Dorian P, Prakash R, Alings M, Krahn AD. Randomized cluster crossover trials for reliable, efficient, comparative effectiveness testing: design of the Prevention of Arrhythmia Device Infection Trial (PADIT). Can J Cardiol 2013;29:652 – 658. 32. Hansen LK, Berg K, Johnson D, Sanders M, Citron M. Efficacy of local rifampin/minocycline delivery (AIGIS(RX)(R)) to eliminate biofilm formation on implanted pacing devices in a rabbit model. Int J Artif Organs 2010;33:627 –635. 33. Bloom HL, Constantin L, Dan D, De Lurgio DB, El-Chami M, Ganz LI, Gleed KJ, Hackett FK, Kanuru NK, Lerner DJ, Rasekh A, Simons GR, Sogade FO, Sohail MR, Investigators COMsMaCAD. Implantation success and infection in cardiovascular implantable electronic device procedures utilizing an antibacterial envelope. Pacing Clin Electrophysiol 2011;34:133 –142. 34. Kolek MJ, Dresen WF, Wells QS, Ellis CR. Use of an antibacterial envelope is associated with reduced cardiac implantable electronic device infections in high-risk patients. Pacing Clin Electrophysiol 2013;36:354 –361. 35. Mittal S, Shaw RE, Michel K, Palekar R, Arshad A, Musat D, Preminger M, Sichrovsky T, Steinberg JS. Cardiac implantable electronic device infections: incidence, risk factors, and the effect of the AigisRx antibacterial envelope. Heart Rhythm 2014;11:595–601. 36. Birnie DH, Healey JS, Wells GA, Verma A, Tang AS, Krahn AD, Simpson CS, Ayala-Paredes F, Coutu B, Leiria TL, Essebag V, Investigators BC. Pacemaker or defibrillator surgery without interruption of anticoagulation. N Engl J Med 2013;368: 2084– 2093. 37. Reddy VY, Knops RE, Sperzel J, Miller MA, Petru J, Simon J, Sediva L, de Groot JR, Tjong FV, Jacobson P, Ostrosff A, Dukkipati SR, Koruth JS, Wilde AA, Kautzner J, Neuzil P. Permanent leadless cardiac pacing: results of the LEADLESS trial. Circulation 2014;129:1466 –1471. 38. Lambiase PD, Barr C, Theuns DA, Knops R, Neuzil P, Johansen JB, Hood M, Pedersen S, Kaab S, Murgatroyd F, Reeve HL, Carter N, Boersma L, Investigators E. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. Eur Heart J 2014;35: 1657– 1665.