Joseph

W. Carlson,

PhD

#{149} David

MR Relaxometry Work

relaxation

rates

in extremities

heads

of healthy volunteers. The of the measurement is sufficient to obtain a distinctive relaxation rate dispersion behavior for different tissues. sensitivity

Index terms: MR. 452.1214

technology

Brain, MR, 10.1214 #{149} Magnetic resonance #{149} Magnetic resonance

Extremities, (MR), (MR), tissue

#{149}

characterization Radiology

Antonio

Brito,

#{149}

BS

1992;

184:635-639

resonance (MR) imaging is a useful diagnostic tool in clinical imaging because it provides a wide range of image contrasts. It is well known that image contrast in MR imaging is determined mainly on the basis of the longitudinal and transverse relaxation times Ti and T2, the density of hydrogen nuclei haying a substantial but comparably smaller variability in the body. It is also well known that Ti of tissue varies considerably with different magnetic field strengths. These differences are ultimately due to interactions of ‘

lvi

AAGNETIC

the hydrogen cal environment. Most of the

strength rates

nucleus observable

dependence in the

with

range

its chemifield

of relaxation of field

the Radiologic of California, Dr, South

Imaging Laboratory, San Francisco, 400 San Francisco, CA 94080

(J.W.C., A.B., L.K.); and Tosiba America MRI Inc. South San Francisco, Calif (D.M.G.). Received 13, 1992; revision requested March 16; revision received March 30; accepted April 8. Address reprint requests to J.W.C. C RSNA, 1992

January

PhD

trometers

(2-4)

by

MR

imaging

sequence.

MATERIALS

high

field

strengths.

In

the intermediate regime-a range of field strengths currently being used with clinical imagers-we should see the transition behavior between the limiting cases of very high and very low field strengths. Techniques for measuring the field strength dependence of Ti have been devised for use with laboratory spec-

cy-

article,

Apparatus

Measurement performed with

of relaxation an existing

system for

A 0.064-T

system

ica

South

MRI,

modified

backs. well

magnet,

for

this

of large can

be

variations

a pulsed aperture human resting patients

(20-cm gap). investigation on their sides on their

with

is centered

A pulsed netization value; no reception

magnet

the

relatively

strength.

small

In addition,

magnetic

on

gion of the relaxation

is

because

in relaxation

will see that the baseline strength

was

electro-

imager

changes

explored

Amer-

Calif)

application

of field

field

strength.

whole-body

The low-field-strength suited

elec-

Toshiba

Francisco,

by including

regimen rates

field

(ACCESS; San

was MR

a pulsed of a cycle

magnetic

permanent

imaging

rates clinical

by adding application

the baseline

of a clini-

METHODS

Experimental

magnet with a small This small gap limits to heads for patients or to extremities for

at very

In this

AND

corresponding

of 1

of field

we describe the implementation field-cycling relaxometry with cal, whole-body MR imager.

onto

to a frequency

means

cling. During part of the pulse sequence, an electromagnet is turned on and magnetization is allowed to approach equilibrium in that field. The electromagnet is then turned off and the magnetization measured quickly by means of a conventional

relevant to imaging systems is a result of hydrogen interactions with biologic macromolecules, particularly proteins (1). The typical rotational correlation time for a protein molecule is 1 psec,

stant

From University Grandview

Kaufman,

imaging tromagnet

strengths

MHz. Longitudinal relaxation of a hydrogen spin occurs more rapidly when the Larmor frequency is much lower than the characteristic frequency for protein reorientation than when it is much higher. In general, Ti will approach distinct limiting values at very high and very low field strengths in proteinaceous tissues. At Larmor frequencies much less than i MI-Iz (corresponding to a field strength of 200 G), Ti will decrease to a value independent of field strength. At frequencies much higher than 1 MHz, Ti will approach a higher value and again become con-

I

Leon

#{149}

in Progress’

of relaxation rate disperfrom magnetic resonance images was demonstrated on a clinical, whole-body imaging system. Study of the behavior of relaxation rates over a range of field strengths probes the structural environment of imaged hydrogen protons and reveals information about the composition of tissue. The authors deterand

PhD

Imaging

Acquisition sion curves

mined

M. Goldhaber,

an

we

field

interesting

re-

rate spectrum. is used

to approach radio-frequency is performed

to allow

mag-

an

equilibrium transmission while the pulsed

electromagnet

is on, nor

magnetization

undergoing

is any

or

transverse

free preces-

sion. Therefore, the homogeneity of the electromagnet is relatively unimportant. For example, although the homogeneity of the static magnetic ppm, the homogeneity

tromagnet typical

field is typically of the pulsed

is on the order field

Abbreviations: repetition time.

of 30%

50 elec-

across

a

of view.

ROl

=

region

of interest,

TR

=

635

A small electromagnet duce power requirements ing.

The

of two

was used to refor fast switch-

electromagnet

circular

we

bundles

cross-sectional

consisted

(ROIs) in particular tissues. A numerical best fit of the data to a monoexponential approach to an equilibrium value was

with

formed

used

of wires

dimension

7 x 7

of

a

cm

an average radius of 10 cm. A current 60 A provided a maximum variation 450

G with

these

tromagnet

was

parable fiers.

measurements.

driven

to those Voltage

supplies

as gradient

limitations

resulted

The

com-

ampli-

on the current

in a rise

time

to full

cycle

field strength of 30 msec. The fall time of the electromagnet was reduced substantially to approximately 8 msec with the use of a flying capacitor dump circuit. The rate of change of the magnetic field strength in these measurements was 6 T/sec, which is sufficiently high to induce retinal magnetophosphenes (5) but below levels for motor nerve stimulation reported in echo-planar imaging (6). In the imaging pulse sequence, we applied an initial 90 pulse to flip magnetization

from

the

static

field

direction

into

the

transverse plane. The pulsed electromagnet was then turned on for a time r, ranging from 50 to 350 msec. Transverse magnetization is quickly and irretrievably dephased in the highly nonuniform field. The cycle field was then turned off, and

the field was allowed

to settle.

shielding of the pole piece counterwound “pancake”

nated

eddy

dard

imaging

formed

currents

without

contamination

baseline field strength. To obtain an accurate essential

that

a stan-

the

of the

through

it is the

elec-

tromagnet be turned off completely. This requires an introduction of additional delay time for the current to fully settle. The time between turning off the amplifier pulse and start of a pulse sequence was 25 msec.

After

the current

two-dimensional

echo sequence

settled, Fourier

a standard transform

was performed

spin-

nal from an asymmetric echo was collected, with an echo time of 30 msec, a section thickness of 10 mm, a 1.7 x 1.4-mm in-plane resolution, and a 128 x 256 ma-

trix. Field

cycle time was the same

for all

phase-encoded acquisitions, then it was increased and the imaging sequence repeated. The time of cycle field application ranged from 50 to 350 msec in steps of 75 msec, for a total of five different values. With the 25-msec delay time for current settling and 150 msec for performing ra-

dio-frequency

pulses,

data

acquisition,

and unused sequence time, the repetition time (TR) was 225-525 msec, for a total imaging time of 4 minutes per field strength value. Current investigations with this apparatus have included extremities and heads of healthy volunteers and a rat model.

Data

Analysis

Mean signal from manually

636

Radiology

#{149}

intensities located

were regions

echo

recorded of interest

be

Two

in-

into

ac-

taken

sequence

results

in some

ph

zation,

magnetization

90#{176}-I80#{176} measurement following form: m

+ {[m(f)

-

exp_b/TI(fo)J}

‘1’ ‘

The signal

intensity

an

of the collected magnetization

exponential

attenuation

data is

multiplied from

T2

decay. The first term in braces in Equation (1) contains the two instrumental details described earlier; this term is a constant independent of the cycle field strength. The second term in braces is linearly dependent on cycle field strength by means of the equilibrium magnetization value mO(f). The derivation of Equation (1) is

explained The

in the Appendix. most

reliable

estimates

of TI are fit to all data from each of the selected ROIs for all cycle field strength values. Thirty different cycle field strength values were used to obtain a typical set of images in these preliminary studies. As a result, data analysis required a 33-parameter fit to the data: one parameter for the constant term in Equation (I), two for the linear function in Equation (1), and 30 for the Tls at the different field strengths. In analyzing the data, we did not assume any relationship between the Tls at different field strengths. The computational effort required for this analysis is modest. Before performing a global fit of the data, only the data from a single field strength value are fit to a three-parameter model. These results are then used as a starting point in a simplex search for a global best fit. The entire data analysis for each ROI at all field strength values required approximately 2 minutes with use of a minicomputer. Errors were estimated for the individual data points by evaluating the variance of the data points from the best-fit curve, that is, the reduced 2 was set equal to 1. Upper and lower confidence limits of Tl

by performing

a global

the

inhomogeneous

cycle

field.

The

expT/I(f)J.

-

by

The regrowth occurs for a time r, at an increased or decreased field strength. The electromagnet is turned off, and, after allowing the current to completely shut off, the magnetization is measured with a conventional spin-echo two-dimensional Fourier transform sequence. rf = radio frequency.

mz]exp_b/TI(fo)}

-

-----

dephased

(m) before the sequence has the

+ m,(f0)[l

mexpbhhl(fo)

=

times.

-

Figure 1. Diagram of pulses for the relaxation rate measurements. An initial 90#{176} pulse flips longitudinal magnetization into the transverse plane, where it is quickly

relaxation

and the cycle and delay

longitudinal

d

r

to the baseline field strength magnetization at the TI of that field. The assumed model of signal intensity depends on the TI at the cycle field strength, Tl(f), the TI at the baseline field strength, Tl(f,), the equilibrium magnetization values at the cycle and baseline field strengths, m0(f) and m,(f0), respectively, the residual longitudinal magneti-

achieved

(Fig 1). Sig-

density. must

per-

relaxation

count for this data analysis to be successful. First, transmitter coil nonuniformity or misadjustment will result in nonzero residual magnetization, m,, along the static field after application of the 90#{176} prepulse. Second, the delay time 6 between turning off the current and the start of the spin-

by

measurement,

current

details

the longitudinal

to be per-

the tissue

magnetization

x [1

with use of a winding elimi-

sequence

and

Active

and allowed

pulse

rate

strumental

elec-

by amplifiers

used

and

of of

to calculate

f,Ld

I

were established by performing a numeric search for the values of TI to find where x2 increased by 1 (7). The confidence limits describe only random errors in the data and do not help estimate effects from systematic errors in the instrumentation. Cycle field strength over the entire field of view varies by approximately 30%. Therefore, we must be careful to calibrate the local field strength when presenting data of relaxation rates. The regions studied herein were of the electromagnet,

always

ations were much been calibrated.

close so that

less

than

to the center the field van-

30%

and

had

RESULTS The relaxation process for hydrogen nuclei is the result of a combination of many effects at the molecular level. Individual effects, if they act independently, contribute additively to the overall relaxation rate (the inverse

of Ti).

pendent come

Contributions

relaxation more

plotted

from

processes

apparent

when

as relaxation

rates.

Phantom

inde-

will data

beare

Studies

Archetypal behavior of the relaxation rates as determined with phantom studies are shown in Figure 2. Relatively constant relaxation rate behavior over the field strengths measured

was

seen

in phantoms

contain-

ing simple solutions (Fig 2a), such as a solution of a paramagnetic relaxation agent, nickel chloride, and water or an emulsion of mineral oil and water (approximately i :1 mixture of mineral oil and

water

oleate and monooleate Phantoms romolecules,

with

sorbitan

mono-

polyoxyethylene-sorbitan added as emulsifiers). containing biologic macsuch as hen egg albuSeptember

1992

6

T

300 400

2 1000

400

Field

600

Strength

800

000

ues were returned

400

decreased

:

600

200

300 :

6

4

S

)

2OO

are probably artifacts of the order of data acquisition. Data were initially acquired at the baseline field strength of the magnet; the field strength val-

150

150

1200

0

200

400

(guss)

F1d

600

Strength

800

1000

1.00

( g.ss)

b. Figure

2. rates

ation emulsion strength phantom strengths associated phantom

Relaxation rate dispersion curves for in vitro samples. (a) Curves show the relaxfor phantoms containing a solution of nickel chloride and water (solid line) and an of mineral oil and water (dotted line). Relaxation rates were insensitive to field over the ranges investigated and showed no apparent structure. (b) Curves from a containing hen egg albumen show a comparably larger change over the field studied. The denatured egg phantom (dotted line) exhibited peaks at 500 and 650 G with nitrogen cross-relaxation that were not present in the native egg albumen (solid line).

higher than the native value. Also, structure appeared in the relaxation rate curve, with the appearance of two peaks at approximately 500 and 650 C. These peaks were ascribed to cross-relaxation effects of proton interactions with nitrogen nuclei in cross-linked protein molecules (2). These peaks represent not only the chemical composition of the phantom but also the structure of the composition. These peaks in relaxation rates

are a result

of the

short-range

the tissue caused by heating phantom (8). The field strength values

peaks

Figure

3.

Representative

MR image

ob-

tamed in the knee of a healthy volunteer (TR = 525 msec, echo time = 30 msec [525/ 30]). This is one of the five images used to determine relaxation rates. The cycle field used in this image was set to give a total field strength of 1,200 G.

men (Fig 2b), exhibited substantial changes in the relaxation rates for these same field strengths. The lower curve in Figure 2b exhibits a decrease

in the

relaxation

rate

from

3 sec-1

(Ti,

333 msec) at 200 G to 2 sec (Ti, 500 msec) at 1,100 C. Substantial changes occur in egg albumen with heat denaturation. After a phantom containing albumen was heated at approximately 60#{176}C for 10 minutes, the overall relaxation rate increased, as did the dispersion in the rates. At 200 C, the rate had nearly doubled, to roughly 5 sec (Ti, 200 msec), and decreased to 2.5 sect (Ti, 400 msec) at 1,100 C, a rate 25%

Volume

184

Number

#{149}

3

are

a result

of the

order

of these

quadrupole

transition frequencies of the nitrogen nuclei and are independent of the chemical composition of the material. Winter and Kimmich (2) calculated three field strength values for these peaks: i95, 490, and 685 C, with the peak at lowest field seen only in highly ordered biologic tissues. Koenig and Brown (i) reported slightly different values for the higher two peaks, of 500 and 655 C.

Human

Volunteer

marrow,

and

Study

fat; data

The

were

macromolecular

tissue

with

substan-

tially less structure. Data obtained the region of the pons exhibited stantial structure in the relaxation behavior (Fig 5c).

in subrate

Study

The measurements obtained in human volunteers are intended only to help assess the sensitivity of the technique in the differentiation of normal tissue. To determine both the feasibility and potential of the technique in the evaluation of pathologic tissue, we obtained measurements in a rat with an implanted mammary adenocarcinoma (9). A representative transaxial image of two rats is shown in Figure 6. The tumor had a discernable necrotic center surrounded by a viable region. Relaxation rates for the necrotic center, viable region, and normal rat muscle tissue are shown in Figure 7. The overall

relaxation

rate

of the

necrotic

center

of the slightly

tumor (Fig compared

7b) decreased with that

viable

region

7a). A small,

statistically seen in the Conversely,

field tially This

(Fig

of the

but

significant, structure was necrotic center at 650 C. the normal muscle (Fig

strengths, as well larger cross-relaxation enables a qualitative

tative differentiation and muscle in this

higher at lower

re-

as a substanstructure. and quanti-

between particular

tumor model.

DISCUSSION

pro-

cessed as described earlier. The basic behavior of the three regions is shown in Figure 4. The fat (ROI volume, 1.2 cm3) and bone marrow regions (ROI volume, 2.5 cm3) showed relatively flat relaxation rate behavior, with little structure apparent. The small structures seen at 640 C, the baseline field strength of the magnet,

then muscle

region, however (ROI volume, 1.5 cm3), exhibited an overall behavior similar to that seen in the cross-linked protein phantom of the previous experiments. This same sequence was repeated with transaxial sections of the brain of the healthy volunteer. Relaxation rate dispersion in white (Fig 5a) or gray (Fig 5b) matter is typical of that of

7c) showed a substantially laxation rate, particularly

A set of measurements was obtamed from sagittal images acquired in the knee of a healthy 34-year-old man. A representative image of the knee is shown in Figure 3. ROIs were located in different regions of muscle,

bone

incrementally, value, and

incrementally.

Rat Model

in

the

increased to baseline

These

intended

preliminary

only

of relaxation rate a clinical whole-body

imaging

time

experiments

to show

the

are

feasibility

determination imager.

of 4 minutes

with With an

per

field

point, the total data collection time for the 32 field points used in the human studies was approximately 2 hours.

Patient

acceptance

with

the

imaging

Radiology

637

#{149}

I00

100

100

8

r ISO

ISO

S

6

‘, 3)

20(1

400

600

Field

800

StI’ngth

a. Figure

1000

200

200

300

300

400

400

1300

600

1000

1000

200

I

200

400

(glJ)

FIeld

600

800

1000

200

300 400

2 600

200

rates for fat (a), bone marrow indicative of the short-range

400

(g.us)

Strength

600

800

Strength

Field

b.

4. Relaxation cross-relaxation peaks

1000

1200

000

12

0

(gauss)

C.

(I,), and muscle (c) in a healthy order in this tissue.

volunteer.

The long

muscles

of the extremity

showed

large

200

200

300

300

400

400

600

600

i S

10011

CO

200

400

600

800

1000

1000

120(1

I Fiald

200

F,td

a.

600

400

(gauss)

Strangth

600

1000

1200

0

sion

5. Relaxation components with

ROIs

were

times

3.0 cm3, I .9 cm3, and

used

would

rate dispersion little apparent

in these

be poor,

of field

Our

data

determine structure

show

matter,

in the of the

brain pons

of a healthy volunteer. (c) shows a much larger

gray

matter,

and

pons,

shorten reducing

by

will

points that

than were We can

allow

can

used reduce

us to acquire

in this the

Figure

images

positioning). is needed

Furto deter-

minutes.

The clinical applicability of this procedure may lie in observations of changes in nitrogen peaks associated with structural changes in the underPathologic

such

structural

as demyelination

conditions

changes

in

of brain

6.

Considerations

of total

imaging

aging.

In both

ence

is that

dimensions

hence, tional

spectroscopy

spectroscopy of spatial

total imaging to the product

ings.

However,

relax-

from along differ-

requires

MR images

(525/30)

disperof the

of two rats with

im-

phase

total

encoding;

imaging

time

is

independent of the number of frequency points of spectroscopy information. In relaxation rate dispersion measurements, the total imaging time is dependent on the product of only one

of the

spatial

dimensions

and

strength

of the

magnet,

we

the

field

were

to probe the cross-relaxation peak structure and very low field relaxation rate dispersion with relatively small changes in field strength. Attempts to perform this procedure with a mediumor high-field-strength

MR imager

two

time is proporof spatial encod-

peaks.

Radiology

remiim-

and

number of field points sampled. Because of the low baseline

#{149}

Transaxial

resolution are in spectroscopic

or scar tissue in muscle, may be mdicated by changes in cross-relaxation

638

show large The volumes

planted mammary adenocarcinomas. A small necrotic region is visible in the center of a large tumor in the leg of the rat on the right side of the figure. As in Figure 3, only one of five images obtained at 1,200 G is shown.

of

mine if we can reliably calculate Ti with less than five different time delays; if so, we could reduce examination time further, to approximately 15

tissue,

(b) matter peak effect.

clearly

ometry, additional information each image voxel is acquired with the image. The primary

with

gray

peak fewer

including patient ther investigation

tissue.

White (a) and cross-relaxation

measured. we

time and spatial niscent of those

associated

1200

respectively.

the head in 64 phase-encoded steps. These reductions will result in a total examination time of 30 minutes (not

lying

1000

the the

number of field points to 15 and still have sufficient data to characterize both cross-relaxation peaks and the overall trend in relaxation rates. A sacrifice in image resolution to 1.4 x 3 mm

800

( aa.s)

we anticipate

cross-relaxation with substantially

data points investigation.

600 Strngth

C.

determined A region

2.0 cm3, for white

400 F,1d

experiments

but

that we can substantially overall imaging time

number

curves structure.

200

S trngth

b.

Figure

1000

able

tions

This culties

larger field technical

with

changes,

experience Future limited

require baseline more

associated

strength

field

will

from the will cause

and

larger the

variastrength. diffi-

field patient

will

larger rates of flux change. work with use of uniform

cycle

coils

would

to extremity

probably

and

head

be

imag-

ing because of the drawbacks associated with larger-aperture coils. However, a nonuniform “surface” field coil could be used to investigate relax-

ation rates nonuniformity

of the heart muscle. The of the cycle field September

1992

.

ISO 6

150

ISO

6 200_

200

4

200

I

4 300

300

400

400

400

600

600

2

_

1000

200

400 Fild

600

800

1000

1000

1200

200

400

(gauss)

St,angth

800 (gauss)

600

Strsngth

Field

a.

1000

Figure

7.

region

(a) and

would niques

Relaxation

rate

necrotic

dispersion

region

curves

determined

in the center

of a tumor

require data processing techto calibrate the field of the coil,

_O

this

is not

a formidable

problem.

for

a rat

(b) decreased,

of this

next

[m2exp_ThhI(f)

x expStTI(fo)

APPENDIX understand

the derivation

+ m,(f)[l

m2exp’T1(I)

-

with

an

with

implanted

little

mammary

variation

magnetization period

+ m,(f)(l

adenocarcinoma.

compared

at the

The cycle field is then turned off and declines to within I % of the baseline field strength within 8 msec. An additional 17 time

delay

is necessary

for

the

re-

maining cycle field to completely decay. We can approximate that the field strength for the entire 25-msec period is the baseline field strength. The evolution

Volume

184

Number

#{149}

3

with

2.

-

+ m0(f0)(I

expT1f)1 -

3.

exp8/Tl(1o)).

dent on the TI at the baseline field strength and the TR. However, because this term is small and can usually be neglected in comparison to the regrowth contribution, this effect is not important. U

Winter

600

800

Strength

1000

1200

(gauss)

Koenig SH, Brown RD III. Relaxometry tissues. In: Gupta RK, ed. NMR spectroscopy of cells and organisms. Vol 2. Boca ton, Fla:CRC, 1987; 75-114.

4.

Relaxation

muscle

F, Kimmich

R.

6.

8.

9.

viable

Y, Muller

RN.

Field-cycling

relaxometry: medical applications. Radiology 1988; 168:843-849. Marg E. Magnetostimulation of vision: direct noninvasive stimulation of the retina and the visual brain. Optom Vision Sci 1991; 68:427-440. Cohen MS. Weisskopff RM, Rzedzian RR,

Kantor varying 7.

for

(c).

NMR field-cycling of bovine serum

spectroscopy

Haverbeke

5.

rates

of normal

albumin, muscle tissue, micrococcus lufeus and yeast: ‘4N’H-quadrupole dips. Biochim Biophys Acta 1982; 719:292-298. Koenig SH, Brown RD III, Adams D, Emerson D, Harrison CG. Magnetic field dependence of 1/T1 in tissue. Invest Radiol 1984; 19:76-81. Rinck PA, Fisher HW, Vander Elst L, Van

KL. Sensory stimulation by timemagnetic fields. Magn Reson Med

1989; 14:409-414. Bevington PR. Data reduction and error analysis for the physical sciences. New York: McGraw-Hill, 1969; 242-243. Beaulieu CF. ClarkJl, Brown RD. Spiller M,

Koenig

References 1.

that

relaxation

is

of

exp’T1(f)J.

400

Field

After a rearrangement of terms, this expression is Equation (I). Equation (1), start with the assumption Of the two contributions to the first that before the application of the initial 90#{176} term in braces in Equation (1), the residual pulse, magnetization is purely longitudilongitudinal magnetization factor is much nal and aligned with the static baseline smaller than the term from regrowth durmagnetic field. Application of the initial, ing the period 8. For example, if the typinominal 90#{176} pulse flips magnetization into cal Ti at the baseline field strength is 300 the transverse plane, with some possible msec, the regrowth in 25 msec is 8% of m,; residual component in the z direction, m2. if the ffip angle is misset by 5#{176}, the residual The cycle field is applied for a time ‘r. The magnetization is only 0.4%. magnetization will exponentially apIf the magnetization has not fully come proach the equilibrium value m0(f) with to equilibrium at the baseline field the characteristic TI constant of the cycle strength by the end of the measurement field. The longitudinal magnetization after phase of the pulse sequence, the residual the time i is magnetization would be slightly depenTo better

200

C.

of the longitudinal end

1000

1200

b.

msec

300

2 600

but

0

SH.

Relaxometry

of calf lens ho-

mogenates, including cross-relaxation by crystallin NH groups. Magn Reson Med 1988; 8:45-57. Davis PL, Sheldon P. Kaufman L, et al. Nuclear magnetic resonance imaging of mammary adenocarcinomas in the rat. Cancer 1983; 51:433-439.

of Ra-

Radiology

639

#{149}

MR relaxometry imaging. Work in progress.

Acquisition of relaxation rate dispersion curves from magnetic resonance images was demonstrated on a clinical, whole-body imaging system. Study of th...
952KB Sizes 0 Downloads 0 Views