Nonlinear light-sheet fluorescence microscopy by photobleaching imprinting Liang Gao, Liren Zhu, Chiye Li and Lihong V. Wang J. R. Soc. Interface 2014 11, 20130851, published 29 January 2014

Supplementary data

"Data Supplement" http://rsif.royalsocietypublishing.org/content/suppl/2014/01/24/rsif.2013.0851.DC1.htm l

References

This article cites 12 articles, 2 of which can be accessed free

Subject collections

Articles on similar topics can be found in the following collections

http://rsif.royalsocietypublishing.org/content/11/93/20130851.full.html#ref-list-1

bioengineering (176 articles)

Email alerting service

Receive free email alerts when new articles cite this article - sign up in the box at the top right-hand corner of the article or click here

To subscribe to J. R. Soc. Interface go to: http://rsif.royalsocietypublishing.org/subscriptions

Downloaded from rsif.royalsocietypublishing.org on May 11, 2014

Nonlinear light-sheet fluorescence microscopy by photobleaching imprinting Liang Gao†, Liren Zhu†, Chiye Li and Lihong V. Wang

rsif.royalsocietypublishing.org

Research Cite this article: Gao L, Zhu L, Li C, Wang LV. 2014 Nonlinear light-sheet fluorescence microscopy by photobleaching imprinting. J. R. Soc. Interface 11: 20130851. http://dx.doi.org/10.1098/rsif.2013.0851

Optical Imaging Laboratory, Department of Biomedical Engineering, Washington University, St Louis, MO 63130, USA We present a nonlinear light-sheet fluorescence microscopy (LSFM) scheme based on photobleaching imprinting. By measuring photobleaching-induced fluorescence decay, our method simultaneously achieves a large imaging field of view and a thin optical section. Furthermore, the scattered-lightinduced background is significantly reduced, considerably improving image contrast. Our method is expected to expand the application field of LSFM into the optical quasi-ballistic regime, enabling studies on non-transparent biological samples.

1. Introduction Received: 19 September 2013 Accepted: 7 January 2014

Subject Areas: bioengineering Keywords: light-sheet fluorescence microscopy, nonlinear microscopy, photobleaching

Author for correspondence: Lihong V. Wang e-mail: [email protected]

Light-sheet fluorescence microscopy (LSFM) is an emerging technology that combines optical sectioning with wide-field detection to image biological tissues and living organisms [1–3]. Despite widespread application, LSFM faces a fundamental trade-off between imaging field of view (FOV) and sectioning thickness [4]. If a high numerical aperture (NA) objective is used for excitation, then thin optical sectioning will be achieved at the expense of decreasing the excitation beam’s depth of focus, reducing the FOV (figure 1a). Conversely, if a low NA is chosen to generate a large imaging FOV, then the excitation beam’s waist will be compromised, thickening the optical section (figure 1b). In addition, if the sample is turbid, then LSFM suffers from the scattering of photons through the specimen [5–8], a fact that restricts LSFM’s application at depths. To provide a simple solution to these long-standing problems in LSFM, here we present a nonlinear light-sheet fluorescence (LSF) imaging scheme based on photobleaching memory effect. The resulting method, light-sheet fluorescence photobleaching imprinting microscopy (LSF–PIM), mitigates the trade-off between imaging FOV and sectioning thickness by extracting high-order fluence-dependent signals from photobleaching-induced fluorescence decay. Moreover, we demonstrate that LSF –PIM can effectively suppress the background fluorescence excited by scattered photons—akin to two-photon microscopy, thereby considerably increasing image contrast.

2. Theory and simulation The operating principle of LSF –PIM is shown in figure 1c. Upon light sheet excitation, the light intensity measured by a wide-field microscope is an integration of the fluorescence emitted from all depths: ð Iðx; yÞ ¼ C {ma ðx; y; zÞðFb ðx; y; zÞ þ Fs ðx; y; zÞÞ}  PSFz ðx; yÞdz; ð2:1Þ



These authors contributed equally to this study. Electronic supplementary material is available at http://dx.doi.org/10.1098/rsif.2013.0851 or via http://rsif.royalsocietypublishing.org.

where C is a constant, ma is the absorption coefficient of the fluorophore, Fb represents the ballistic excitation (i.e. unscattered photon) fluence in J m22, Fs represents the scattered excitation (i.e. scattered photon) fluence, PSFz is the point-spread-function (PSF) of the fluorescence imaging system at depth z, and the operator asterisk denotes two-dimensional convolution in the x 2 y plane. The imaging FOV is determined by the excitation beam’s depth of focus, d, which is approximated by d¼

1:8l NA2

:

& 2014 The Author(s) Published by the Royal Society. All rights reserved.

ð2:2Þ

Downloaded from rsif.royalsocietypublishing.org on May 11, 2014

(b)

(a) FOV

low NA objective

rsif.royalsocietypublishing.org

high NA objective

2 FOV

ST

ST z x imaging objective

imaging objective

t1

z

J. R. Soc. Interface 11: 20130851

(c)

ti

y x (e)

(d)

Dz

z

Dz¢

y x

In (n ≥ 1)

I0

Figure 1. The operating principle of LSF – PIM. (a,b) The trade-off between imaging FOV and optical sectioning thickness (ST) in conventional LSFM. (c) LSF – PIM is conducted by imaging time-lapse photobleaching-induced fluorescence decay. The fluorophores at the focal plane experience higher excitation fluence, and thereby much faster photobleaching than those at out-of-focus planes. By fitting the measured fluorescence time course to a polynomial function, the PIM components that have a high-order dependence on excitation fluence are extracted. (d,e) The effective optical section associated with a zero-order (i.e. conventional LSFM) and a high-order LSF– PIM component, respectively. (Online version in colour.)

Here, l is the excitation laser wavelength. Within this FOV, the ballistic excitation fluence is assumed to be uniform along the x- and y-directions. Equation (2.1) can be simplified to ð Iðx; yÞ ¼ C ma ðx; y; zÞðFb ðzÞ þ Fs ðx; y; zÞÞ  PSFz ðx; yÞdz: ð2:3Þ

time t, we obtain

In fluorescence microscopy, photobleaching occurs when excited electrons are trapped in a relatively long-lived triplet state. Compared with the singlet – singlet transition, the forbidden triplet –singlet transition provides a fluorophore a much longer time to undergo irreversible chemical reactions with the environment [9]. The photobleaching reduces the local absorption coefficient, resulting in an exponential decay with time:

Substituting equation (2.6) into (2.3) gives

ma ðtÞ ¼ ma0 expðktÞ:

ð2:5Þ

where B is a constant. By substituting equation (2.5) into (2.4), followed by Taylor expansion with respect to

ð2:6Þ

9 ð ðBÞn nþ1 > > Iðx; y; tÞ ¼ C t ma0 ðx; y; zÞðFb ðzÞ þ Fs ðx; y; zÞÞ > > > n! > > n¼0 > = PSFðx; yÞdz > > > 1 > X > > n > In ðx; yÞ  t ; ¼ > ; 1 X

n

n¼0

ð2:7Þ

ð2:4Þ

Here, t is time, ma0 is the initial absorption coefficient of the fluorophore and k is the photobleaching rate. The photobleaching rate, k, is a function of the total excitation fluence [10]. For one-photon excitation, the relation is described by k ¼ BðFb þ Fs Þ;

ma ðtÞ ¼ ma0 expðBðFb þ Fs ÞtÞ 1 X ðBÞn ma0 ¼ ðFb þ Fs Þn tn : n! n¼0

where ð In ðx; yÞ ¼ Dn ma0 ðx; y; zÞðFb ðzÞ þ Fs ðx; y; zÞÞnþ1  PSFðx; yÞdz; ð2:8Þ and Dn ¼ CðBÞn /n!. Here, In(x,y) is the coefficient associated with tn and can be derived from the polynomial fitting of I(x, y, t).

Downloaded from rsif.royalsocietypublishing.org on May 11, 2014

(d) 1.0

LSFM (NA = 0.1)

(a)

3 LSFM (NA = 0.1)

normalized intensity

z x LSM–PIM (I1 μ F2)

(b)

LSM–PIM (I4 μ F5) 0.6 0.4 0.2

LSM–PIM (I4 μ F5)

(c)

0

5

10 distance (mm)

(e)

20 mm 0

15

20

LSFM (NA = 0.25)

20 mm 1

normalized intensity

Figure 2. Simulated effective excitation volume in (a) LSFM with an excitation objective of NA ¼ 0.1, (b) the corresponding first-order LSF – PIM and (c) fourthorder LSF– PIM. (d) Comparison of signal intensities along the dashed line in (a – c). (e) Simulated effective excitation volume in LSFM with an excitation objective of NA ¼ 0.25. The optical sectioning thickness in (c) is comparable to that in (e), yet with a six times larger (in diameter) imaging FOV. (Online version in colour.) Binomial expansion of the term ðFb ðzÞ þ Fs ðx,y,zÞÞn in equation (2.8) gives ð nþ1   X n nþ1k In ðx;yÞ ¼ Dn ma0 ðx;y;zÞ F ðzÞFks ðx;y;zÞ  PSFðx;yÞdz: k b k¼0

ð2:9Þ Because Fb ðzÞ  0 outside the optical section (z , z0 or z . z0 , z0 ¼ 0:61l/NA) and Fb ðzÞ  Fs ðx,y,zÞ within the optical section (z0  z  z0 ), the cross terms in equation (2.9) are negligible. Equation (2.9) becomes ð z0 9 > > In ðx; yÞ ¼ Dn ma0 ðx; y; zÞFnþ1 ðx; y; zÞ  PSFðx; yÞdz > s > > 1 > > ð z0 > = nþ1 þ Dn ma0 ðx; y; zÞFb ðzÞ  PSFðx; yÞdz > z0 > > > ð1 > > > nþ1 ; þ Dn m ðx; y; zÞF ðx; y; zÞ  PSFðx; yÞdz: > z0

a0

s

ð2:10Þ The ballistic fluence Fb ðzÞ reaches a maximum at the depth z ¼ 0. Rearranging equation (2.10) yields (ð   z0 Fs ðx;y;zÞ nþ1 nþ1 In ðx;yÞ ¼ Dn Fb ð0Þ ma0 ðx;y;zÞ  PSFðx;yÞdz Fb ð0Þ 1   ð z0 Fb ðzÞ nþ1 þ ma0 ðx;y;zÞ  PSFðx;yÞdz Fb ð0Þ z0 )   ð1 Fs ðx;y;zÞ nþ1  PSFðx;yÞdz : þ ma0 ðx;y;zÞ Fb ð0Þ z0 ð2:11Þ In the case that Fb ð0Þ . Fs ðx,y,zÞ at all depths, the first and third terms in equation (2.11) approach zero when n  1. Therefore, for a high-order PIM component, the

background signals associated with scattered excitation fluence (the first and third terms in equation (2.11)) are suppressed. Additionally, the optical section associated with ballistic excitation fluence (the second term in equation pffiffiffiffiffiffiffiffiffiffiffi (2.11)) is also reduced by a factor of n þ 1 in thickness (figure 1d,e). To show that LSF –PIM can decrease optical sectioning thickness while maintaining a relatively large FOV, we computed the effective excitation volume by tracking the wavefront using the Fourier optical transformation [11]. With an objective of NA ¼ 0.1 for light sheet excitation (l ¼ 550 nm), the derived cross-sectional (x 2 z) images associated with conventional LSFM, LSF–PIM (I1) and LSF–PIM (I4) are shown in figure 2a–c, respectively. The signal intensities along the dashed line in figure 2a–c are shown in figure 2d. The full width at half maximum (FWHM) of the corresponding axial responses is 2.4, 1.7 and 1.1 mm, respectively. As expected, a higher-order LSF–PIM signal results in a thinner optical section. For comparison, we also simulated the effective excitation volume with an objective of NA ¼ 0.25 for light sheet excitation (figure 2e). In this case, the light sheet’s thickness (1.1 mm) approaches that in LSF–PIM (I4); however, the drawback lies in a much shorter sheet extent along the x-axis. Consequently, provided that the sectioning thickness is comparable, LSM–PIM is superior to conventional LSFM in terms of imaging FOV.

3. Results and discussion To demonstrate LSF–PIM experimentally, we first imaged a phantom sample consisting of red fluorescent beads (diameter, 0.9 mm; maximum emission wavelength, 600 nm) uniformly mixed and sealed in gelatin. The sample was excited by a 532 nm continuous wave (CW) laser and imaged on

J. R. Soc. Interface 11: 20130851

0

rsif.royalsocietypublishing.org

LSM–PIM (I1 μ F2)

0.8

Downloaded from rsif.royalsocietypublishing.org on May 11, 2014

(c)

(b)

(a)

4 LSF–PIM (I4)

rsif.royalsocietypublishing.org

LSF–PIM (I2)

LSFM

normalized intensity

1

y 2 mm

0

x (d) normalized intensity

1

LSFM LSF–PIM (I2) LSF–PIM (I4)

0

2

6 4 distance (mm)

8

10

Figure 3. Light-sheet fluorescence imaging of red fluorescent beads (0.9 mm in diameter) mixed in gelatin. (a) Conventional LSFM image. (b) LSF – PIM (I2) image. (c) LSF– PIM (I4) image. (d ) Comparison of signal intensities across the dashed line in (a – c). (Online version in colour.) a custom-built LSF microscope (Material and methods). With an objective of NA ¼ 0.1 for light-sheet excitation, a conventional LSFM image was captured and is shown in figure 3a. The imaging depth is approximately 500 mm below the sample surface. Owing to the relatively thick optical section and light scattering, most in-focus fluorescent beads are obscured by out-of-focus light. By contrast, after measuring 100 timelapse fluorescence images (0.1 s frame integration time), we calculated the LSF–PIM (I2, I4) images, shown in figure 3b,c, respectively. Three corresponding areas from figure 3a–c are zoomed in for further comparison. The fluorescent beads are much more distinguishable in figure 3b,c because of the much thinner effective optical section. Additionally, the signal intensities across the dashed line in figure 3a–c are compared. The results (figure 3d) indicate that the higher-order PIM component leads to higher image contrast. Next, we imaged a zebrafish embryo 6 days postfertilization (dpf ) in vivo with LSF – PIM. In the transgenic embryo, the myelin basic protein (mbp) promoter drives expression of mcherry throughout the nervous system [12,13], and the spinal cord was imaged. The light sheet’s incident direction (green arrow) was perpendicular to the

spinal cord axis (figure 4a). Imaging a 6 dpf or older fish in vivo is generally considered as a challenge for conventional LSFM owing to light scattering [5]. Here, for that reason, the conventional LSFM image (figure 4b) suffers from severe blurring. By contrast, in the corresponding LSF – PIM image (figure 4c), because most out-of-focus light is rejected, the sample’s spinal cord is clearly seen. To provide a gold standard, the same embryo was also imaged by a confocal fluorescence microscope (Material and methods). The captured cross-sectional image of the spinal cord is shown in figure 4d. The signal intensities across the dashed lines in figure 4b – d were compared (figure 4e). The result implies that while the conventional LSFM fails to provide the contrast, the LSM – PIM accurately resembles the gold-standard image. The image contrast was improved approximately six times by using LSM – PIM. In addition, we also acquired a three-dimensional image of the embryo’s spinal cord by scanning the sample along the z-axis (figure 4f ). The recovery of the spinal cord in three dimensions from a strong fluorescence background substantiates LSF – PIM’s superior imaging capability through a scattering medium.

J. R. Soc. Interface 11: 20130851

15 mm

Downloaded from rsif.royalsocietypublishing.org on May 11, 2014

(a)

(b)

x

(c)

(d)

5

y

y

300 mm

20 mm 0

1 normalized intensity

(e)

(f)

CM LSFM LSF–PIM 0.5

1000 y (mm)

normalized intensity

1.0

500

0 0

0

40

80 120 distance (mm)

160

1000

500 z (mm)

0

0

500 1000 m) x (m

Figure 4. LSF – PIM imaging of a 6 dpf tg(mbp:mcherry-caax) zebrafish embryo. (a) Bright-field transmission image. In the transgenic embryo, the myelin basic protein (mbp) promoter drives expression of mcherry throughout the nervous system [12,13], and the spinal cord is imaged. The light sheet’s incident direction (indicated by the green arrow) is along the y-axis. (b) Conventional LSFM image (0.1 s frame integration time). (c) LSF – PIM image (I10). (d ) Gold-standard confocal image. (e) Comparison of signal intensities along the dashed line in (b – d). CM, confocal microscopy. (f ) Three-dimensional image of the spinal cord acquired by LSF –PIM (I10). (Online version in colour.) The contrast enhancement by LSF–PIM is accomplished at the expense of an increased acquisition time. The accuracy of fitting time-lapse fluorescence decay into a polynomial function is highly dependent on the number of temporal samplings and the signal-to-noise ratio (SNR). Thus, to extract the higher-order PIM components, more temporal samplings are needed and more photons must be collected at each sampling point. Depending on the sample and application, one can determine an optimal experimental setting to achieve the desired image contrast and temporal resolution. Here, to extract the PIM component I10 (figure 4c), a total of 200 time-lapse images was measured with a 0.1 s frame integration time.

implementation and superior imaging capability, LSF–PIM is expected to open new areas of investigation in studies involving relatively large and non-transparent living organisms.

5. Material and methods 5.1. Preparation of the phantom sample Gelatin was dissolved in warm deionized water to form a 10% solution, then uniformly mixed with red fluorescent beads (R900, Fisher Scientific; average diameter, 0.9 mm). The mixture was cooled to 48C to form a gel.

5.2. Zebrafish husbandry

4. Conclusion In summary, we presented a nonlinear LSF imaging approach based on PIM. Compared with conventional LSFM, our method features two advantages: on the one hand, it simultaneously achieves a large imaging FOV and a thin optical section; on the other hand, it effectively removes the scattering-induced background, thereby allowing much deeper penetration into a thick medium. We demonstrated the deep imaging capability of LSF–PIM by imaging a scattering phantom and a 6 dpf fish embryo in vivo—while conventional LSFM showed almost no contrast between the in-focus signals and background, LSF–PIM clearly resolved the sample features and closely resembled the gold standard. Owing to its easy

Zebrafish embryos were raised at 28.58C and staged according to published methods [14]. All animal work was performed in compliance with Washington University’s institutional animal protocols. tg(mbp:mcherry-caax) zebrafish were a kind gift of Dr David Lyons (University of Edinburgh) [12,13]. Larvae were raised in 0.0045% phenylthiourea in embryo medium to prevent the formation of melanin pigments. For visualization, larvae were anaesthetized with tricaine methanesulfonate and immobilized in low melt agarose.

5.3. Light-sheet fluorescence microscopy A light-sheet microscope was built for fluorescence imaging (see electronic supplementary material, figure S1). A CW laser (diameter, 2 mm; wavelength, 532 nm) is used for fluorescence excitation. The excitation strength is adjusted by a combination

J. R. Soc. Interface 11: 20130851

x

rsif.royalsocietypublishing.org

z

Downloaded from rsif.royalsocietypublishing.org on May 11, 2014

5.4. Photobleaching imprinting microscopy and image processing

Acknowledgements. We thank James Ballard for careful reading of the

The PIM is conducted by measuring time-lapse fluorescence decay, followed by polynomial fitting and extraction of high-order, fluence-dependent coefficients. To make the PIM image reconstruction algorithm easily accessible to the biological research community, we implemented it as an open-source plugin for the

5.5. Confocal fluorescence microscopy The gold-standard image shown in figure 4d was captured on a commercial confocal laser scanning fluorescence microscope (FV1000, Olympus) with 1 s acquisition time. We used a 40 microscope objective (Plan Fluorite, NA ¼ 0.75) to excite the fluorophore and collect fluorescence, and an RFP filter set (Olympus) to separate fluorescence from excitation light. An optimal pinhole size was automatically chosen by the microscope to balance the in-focus SNR and optical sectioning thickness. To increase the image SNR, a Kalman filter was applied during data acquisition.

manuscript. We thank David Lyons (University of Edinburgh) for the transgenic zebrafish. We also thank Sarah DeGenova and Kelly Monk for helping prepare the zebrafish embryo sample. This work was sponsored in part by National Institutes of Health (NIH) grants DP1 EB016986 (NIH Director’s Pioneer Award), R01 CA134539 and R01 CA159959. L. Wang has a financial interest in Microphotoacoustics, Inc. and Endra, Inc., which, however, did not support this work.

References 1.

2.

3.

4.

5.

Santi PA. 2011 Light sheet fluorescence microscopy: a review. J. Histochem. Cytochem. 59, 129–138. (doi:10.1369/0022155410394857) Reynaud EG, Krzic U, Greger K, Stelzer EHK. 2008 Light sheet-based fluorescence microscopy: more dimensions, more photons, and less photodamage. Hfsp J. 2, 266–275. (doi:10.2976/1.2974980) Voie AH, Burns DH, Spelman FA. 1993 Orthogonalplane fluorescence optical sectioning: threedimensional imaging of macroscopic biological specimens. J. Microsc. 170, 229–236. (doi:10.1111/ j.1365-2818.1993.tb03346.x) Fahrbach FO, Gurchenkov V, Alessandri K, Nassoy P, Rohrbach A. 2013 Self-reconstructing sectioned Bessel beams offer submicron optical sectioning for large fields of view in light-sheet microscopy. Opt. Express 21, 11 425 –11 440. (doi:10.1364/OE.21. 011425) Keller PJ, Schmidt AD, Santella A, Khairy K, Bao ZR, Wittbrodt J, Stelzer EHK. 2010 Fast, high-contrast

6.

7.

8.

9.

imaging of animal development with scanned light sheet-based structured-illumination microscopy. Nat. Methods 7, U637 –U655. (doi:10.1038/nmeth. 1476) Mertz J, Kim J. 2010 Scanning light-sheet microscopy in the whole mouse brain with HiLo background rejection. J. Biomed. Opt. 15, 016027. (doi:10.1117/1.3324890) Cella Zanacchi F, Lavagnino Z, Faretta M, Furia L, Diaspro A. 2013 Light-sheet confined superresolution using two-photon photoactivation. PLoS ONE 8, e67667. (doi:10.1371/journal.pone.0067667) Keller PJ, Dodt HU. 2012 Light sheet microscopy of living or cleared specimens. Curr. Opin. Neurobiol. 22, 138– 143. (doi:10.1016/j.conb.2011.08.003) Tsien RY, Ernst L, Waggoner A. 2006 Fluorophores for confocal microscopy: photophysics and photochemistry. In Handbook of biological confocal microscopy (ed. JB Pawley), pp. 338– 348. Berlin, Germany: Springer.

10. Patterson GH, Piston DW. 2000 Photobleaching in two-photon excitation microscopy. Biophys. J. 78, 2159– 2162. (doi:10.1016/S0006-3495(00)76762-2) 11. Barrett HH, Myers KJ. 2004 Foundations of image science. Wiley Series in Pure and Applied Optics. Hoboken, NJ: Wiley-Interscience. 12. Czopka T, Ffrench-Constant C, Lyons DA. 2013 Individual oligodendrocytes have only a few hours in which to generate new myelin sheaths in vivo. Dev. Cell 25, 599 –609. (doi:10.1016/j.devcel.2013. 05.013) 13. Almeida RG, Czopka T, Ffrench-Constant C, Lyons DA. 2011 Individual axons regulate the myelinating potential of single oligodendrocytes in vivo. Development 138, 4443 –4450. (doi:10.1242/ dev.071001) 14. Kimmel CB, Ballard WW, Kimmel SR, Ullmann B, Schilling TF. 1995 Stages of embryonic development of the zebrafish. Dev. Dyn. 203, 253 –310. (doi:10. 1002/aja.1002030302)

6

J. R. Soc. Interface 11: 20130851

widely used IMAGEJ software (http://rsb.info.nih.gov/ij/). The LSF–PIM IMAGEJ plugin is freely available at http://code.google. com/p/lsf-pim.

rsif.royalsocietypublishing.org

of a 1/2 l waveplate (AHWP05M-600, Thorlabs) and a polarizing beam splitter (PBS101, Thorlabs). The excitation laser is focused into a line by a cylindrical lens of 50 mm focal length (LJ1695RM-A, Thorlabs). Then, the laser line image is relayed to the back aperture of a four times microscope objective (Plan Fluorite, Nikon) with NA ¼ 0.1 through a 4-f telecentric imaging system consisting of a lens of 25 mm focal length (LA1951-A, Thorlabs) and a lens of 100 mm focal length (LA1509-A, Thorlabs). The fluorescence is collected by a 10 microscope objective (Plan Fluorite, Nikon) with NA ¼ 0.3, filtered by an emission filter (central wavelength, 559 nm; bandwidth, 34 nm), and imaged by a CCD camera (ORCA-R2, Hamamatsu). The FWHM of the point-spread-function on the fluorescence imaging side is 12.2 mm, and is sampled by 2  2 camera pixels ( pixel size, 6.45 mm). To achieve volumetric imaging, the sample is scanned along the z-axis with a step size of 50 mm.

Nonlinear light-sheet fluorescence microscopy by photobleaching imprinting.

We present a nonlinear light-sheet fluorescence microscopy (LSFM) scheme based on photobleaching imprinting. By measuring photobleaching-induced fluor...