Physico-mechanical properties of degradablepolymersused in medical applications:a comparativestudy Israel Engelberg and Joachim Kohn Department of Chemistry, Rutgers, The State University of New Jersey, New Brunswick NJ, USA (Received 9 January 1990; accepted 11 July 1990)
The physico-mechanical properties of degradable polymers used for medical applications have been characterized. The following polymers were included in this study: three samples of poly(ortho esters) derived from 3,g_bis(ethylidene 2,4,8,10-tetraoxaspiro[5,5]undecane) and various ratios of 1,8-hexanediol and traos-cyclohexane dimethanol, poly(glycolic acid), six samples of poly(t-lactic acid) and poly(~,~lactic acid) with mol wt from 21 000 to 550 000, poly(&-caprolactone), poly(j3-hydroxybutyrate) and three copolymers of /I-hydroxybutyric acid and various amounts of hydroxyvaleric acid, one sample each of two different types of poly(anhydrides), poly(trimethylene carbonate) and two different poly(iminocarbonates). For each polymer, the thermal properties (glass transition temperature, crystallization, melting and decomposition points) were determined by differential scanning calorimetry and by thermogravimetric analysis. The tensile properties (Young’s modulus, tensile strength and elongation at yield and break) were determined by tensile testing on an lnstron stress-strain tester. The flexural storage modulus as a function of temperature was determined by dynamic mechanical analysis. Keywords: Polymers, biodegradation, mechanical properties, thermal analysis
last 20 years, many degradable
Overthe suggested
as possible
application delivery range
of these
polymers Currently
devicele5. of new
important luminal
applications
trend
degradable
in
grafts,
coronary
the reblocking
implants
sophisticated
polymers
with
Particularly
important
strength
and
transition (storage
detail.
modulus),
the
A further
engineering
and
the
Correspondence to Dr J. Kohn.
selected
20
the mechanical
strength,
determined
(glass transition,
require
crystallization
range
of
temperature
analysis thermal
(glass
measured
point,
anical analysis (DMA).
important in
data
on the
is related
to the
facilitate
Biomaterials
199 1. Vol 12 April
(Young’s
The thermal
and melting scanning
(TGA)
stability
as a function
preparation
polymers
and conducted
was and
a
properties. modulus, were
properties
temperatures)
calorimetry
(DSC).
used to determine the
decomposition
( Td). Finally, the flexural storage modulus was
properties
properties
results,
at yield and break)
testing.
by differential
Thermogravimetric the
properties.
properties
and elongation
by mechanical
measured
as biomaterials
of temperature
of all test samples
the intrinsic
by dynamic
mech-
It stands to reason that the consistent by the same laboratory
variability
the direct comparison
may
of the test results and may of the properties
of different
polymers. Unfortunately
it is unavoidable
suffer from several limitations.
that such a study will
First, the number
polymers
is too large to allow a comprehensive
available
materials.
Second,
polymers
0 1991 Butterworth-Heinemann 292
degradable
Specifically, tensile
being investigated
or
reliable
polymers
therefore
(tensile
these
published
polymers.
We
decrease that
is in turn a function
study of some of their engineering
skin
not been investigated
in locating
of new
of different
were
melting
which
are prepared,
even if available, often cannot be used for a direct comparison
properties
viscoelastic
revealed
have often
difficulty
properties
thermal
of the sample
of the way the test samples
comparative
orthopaedic
material
on the morphology
since the mechanical
of polymers are strongly dependent
and
and tan 6).
review
properties
defined
of the test results;
properties
currently
or staples’-“. usually
variability
into
include
artificial
high-strength
(T,), softening
temperature)
literature
engineering
grafts,
are the mechanical
and loss moduli
A
vascular
intrinsic
and rheological
as intra-
after successful
applications
narrowly
temperature
degradation
example,
the collapse
of blood vessels
and degradable,
is an
are implanted
to prevent
such as bone nails, screws,
Such
For
investigated
that
drug
of a wide
‘* 7. Other advanced applications of temporary
for burn victims
studied
polymers
research.
devices
(restenosis)
the development
for degradable
in an attempt
balloon angioplasty
have been
widely
identification
are now being
stent-like
arteries
One
is as an implantable the
biomaterials
polymers
polymers
biomaterials.
of available study of all
of the same
Ltd. 0142-9612/91/030292-13
family
Physico-mechanical
have often tions
been prepared
that
differ
Although our
it has
molecular Finally,
not
within
as uniform
mechanical carefully
techniques
controlling
mechanical
polymers
USED
Poly(ortho Poly(ortho
reported
that
the
polymeric
IN THIS
are
more
hydrophobic
poly(lactic
investigated.
device
thinner
with
time
over
a period
of
acid)
names
were
the
intensively
acid and lactic acid have
sutures
Vicryl’”
of PGA with
(PLA)
of glycolic
as alternative
the trade
of PGA to a wider
copolymers
and are being
and Polyglactin
marketed
9 1 OTMzo,“.
number
devices
for drug
There Originally,
two
Chronome?’
Since
to
drugs
process,
esters)
esters)
Heller
of
were
use
prepared
Upon
products
upon hydrolysis
poly(ortho
The
rates
(t-CDM) with
type
with
2,4,8,1
with
various
hydrolysis soluble
the
by-
autocatalytically
included
in this
1,6-hexanediol
different
(1,6-HD).
ratios
(Structures
study
were
with trans-cyclohexane of t-CDM
Polymer
morphological
hydration
degrade
0,x;
Since distinct from
.+CHz~CHz+
+I
1.6.HI)
DETOSU 1 a. DETOSU 1b 1 c DETOSLJ
t-CDM
t-CDM : 1.6.HD = 100 : 35 : 65 t-CDM : 1.6-H,, = IO0 : 70 : 30 KDM 1.6.HD = IO0 : 90 : 10
in practice,
more
L-PLA
is an
amorphous
acid)
Poly(glycolic polyester
acid)
(PGA)
(Structure
melting was
used
and poly(lactic is the
2). Since
a monophasic
point in
and low the
simplest
solubility
development
orthopaedic linear
aliphatic
crystalline,
in organic of
and
the
it has
solvents.
first
totally
PLA, intensively applications. prehensively
PGA
stereocan
is not often
be
used
here. active
but
D and L
the
optically
Generally,
L-PLA
is
the hydrolysis
is the
naturally
of D,L-PLA
ramifications. it
is
delivery,
are
two
meso-PLA
crystallinity
Since
and
D,L-PLA
usually
considered
where
it is important
of the active
On the other
species
for to
within
hand, the semicrystalline where
required,
high e.g.,
mechanical sutures
and
’ their
copolymers
for a large
research
reviewed
In two
obtained
which
in applications
and
to
acid.
dispersion
investigated This
acid,
polymer,
the
D-PLA, since
of lactic
toughness
It exists
the optically
amorphous.
in the
devices*-’
tend
polymer
materials,
than
as drug
acid;
further
practical
matrix.
is preferred
strength
acid)
PGA is highly
such
are
meso-PLA
from
L(+)-lactic
differences
acid.
in the rates
copolymers
molecule,
L-lactic
is always
have important
the
crystallinity
acid and lactic
L-PLA
Since
frequently
yields
crystalline,
is the racemic
derived
L-PLA
L-PLA
and
and
stereoisomer
and
corresponding
give nse to four morphologically
are semicrystalllne D,L-PLA
relationship acid
the
Thus,
it is not considered
have a homogeneous
Structure 1: Polyforthoesters)
Poly(glycolic
D-
o,L-lactide.
The polymers
applications
DETOSIJ
D-PLA of
from
PLA is more
lead to an increase
is a chiral
D,L-PLA
a mixture
obtained
of thin
backbone
PGA or PLA20,21,23.
which
polymers.
occurring
X
acid
forms
polymers;
The !I+
lactic
than
of
lactic
of
hydrolysis.
rapidly
stereoisomeric
employed
AH2
and
more
to
of glycolic
changes
uptake
rate
is no linear
acid
PGA is highly
of ZZ
there
lost in copolymers
inactive
+,+
that
in
PGA.
of glycolic
Whereas
group
than PGA.
In addition,
than
copolymers. These
the
to PGA”.
solvents
ratio
methyl
the water
reduces
properties
monomers
Ctl,
and
physico-mechanical
to 1,6-HD
la-c).
of an extra
3) is more hydrophobic
as compared in organic
di-
acidic
2%”
It is noteworthy
of O-
presence
of PLA limits
about
is rapidly
of poly(ortho
do not release
of DETOSU
and
three
characterized
polymers
increase
to
regular
esters)
copolymers
specimens
a new
esters)
names
to the
hydrophobicity
between
rates.
poly(ortho
dimethanol
that
and do not exhibit
degradation
condensation
these
(DETOSU)
These
con-
the degradation
of 3,9_bis(ethylidene
tetraoxaspiro[5,5]undecane)
increasing
the trade
hydrolysis,
et a/.18 synthesized
the
and a dialcohol”.
that autocatalyse
based on the reaction
by
Owing
films
esters).
Jr, Poly(lactic acid)
lactic acid, PLA (Structure The
of poly(ortho
poly(ortho
under
in degradation
alcohols.
release
L Structure 3:
(PW
rate, poly(ortho
for controlled the
are known
by-products
resulting
release
that there are a significant
types
or Alzamer”.
acidic
the
crumbling.
of 2,2_diethoxytetrahydrofuran
poly(ortho
e.g.,
and will tend to
applications’4-‘6.
major
poly(ortho
densation
of
than
describing
delivery
are
erosion,
at a constant
It is not surprising
of publications
degradable
tend
useful
Poly(glycollc acid)
can be formulated
surface
to be particularly
drug delivery13.
PGA
to the
tend to lose their
(PGA)
for a number
esters)
rather
the polymer
within
synthetic
only at its surface
embedded
appear
of
undergoes
degrades slab-like
a high
Copolymers
been developed under
Owing
sutures
properties
the data on
the data collected
of PGA in Dexon
typically
materials
applications,
limitations
development
made of poly(ortho
device
become
were
the
of possible
are representative,
PGA
the trade
STUDY
a family
surface-eroding,
esters)
To adapt range
Structure 2:
esters) Devices
release
here
interpreting
that have been under
esters)
parameters,
Kohn
implantation”.
by
improved
under
of 46-52%“.
rapidly,
and J
Cyanamidlg.
The crystallinity
of PGA, Dexona
strength
wk after
the
1970.
I. Engelberg
available
in the range
nature
esters)
polymers
esters)
Since
by American
commercially
since
mechanical 2-4
suture
been
is typically
hydrophilic of
polymers:
l-3.
POLYMERS
years”.
be
best possiblevalues.These when
name Dexona sutures
of polymers.
polymer.
can
the processing
in mind
in
we have not optimized
for each test of
strength
be borne
in Tables
time.
family
cover
for the preparation
as possible,
properties
but not necessarilythe
Such
a given
to
have
all
samples
possible
of degradable
absorbable
sutures
properties.
representative been
synthetic
composi-
material
in order to keep our procedures
the processing
so
of molecular
respective
always
compositions
test specimen
must
in a variety
their
we have tried to include
study,
the
in
properties
by
effort
are
number has
Lewis24.
B~omatenal.5
also
recently PLA,
being
of drug-delivery PGA
199 7. Vol
been
com-
and
their
12 Awl
293
Physic@mechanical
copolymers degradable
are currently polymers
In view research,
properties of degradable polymers: I. Engelberg and J. Kohn
of the
the most widely
in human importance
six representative
used synthetic,
medicine.
samples
of PLA
and copolymers
with
in biomaterials
of PLAwere
included
our study. Table 1
Poly(j? -hydroxybutyrate) hydroxyvaleric acid
in
Poly(P-hydroxybutyrate)
(PHB)
recently been identified
(Structure
as a promising
#a)
biomaterialz5.
has
only
PHB is
Biodegradable polymers used in this study
MVdb
SOWX
Polymera
Poly(ortho esters) [ 1] t-CDM: 1,6-HD = 35:65 t-CDM: 1,6-HD = JO:30 t-CDM: 1,6-HD = 90: 10 Poly(glycolic acid) [Z] Poly(lactic actds) (31 L-PLA L-PLA L-PLA D,L-PLA D.L-PLA D.L-PLA Poly(B-hydroxybutyrate) [4] Homopolymer (0 mol% HV) Copolymer (7 mol% HV) Copolymer (11 mol% HV) Copolymer (22 mol% HV) Poly(e-caprolactone) [5] Polyanhydrides [6,7] Poly(CPP-SA-IS0 anhydride) Poly(SA-HDA anhydride) Poly(trtmethylene carbonate) [8] Polytminocarbonates [ 10,l 1] Poly(BPA-iminocarbonate) Poly(DTH-iminocarbonate)d
Sample characterizationC
WV
Ml?
[rll
19600 43 200 150 000 13400 66 300 163 500
0.6 1 1.65 3.08 0.25 0.64 2.01
SRI lnternattonal SRI International SRI lnternatlonal Polysciences Inc.
99 700 101000 131000 50 000
Polysctences Inc. Polysciences Inc. Polysctences Inc. Polysciences Inc. Stolle R&D Corp. Stolle R&D Corp.
50 100 300 21 107 550
000 000 000 000 000 000
64 800 139 600 375 100 16500 98 500 410 500
ICI ICI ICI ICI Aldrich Chemical Co.
370 450 529 227 44
000 000 000 000 000
212400 173 300 531 500 285 700 72 500
3 1 24 2 42
Nova Pharmaceuttcal Co. Nova Pharmaceutical Co. Amgen inc.
31000 142 000 48 000
147 900
55 100
0.42 1.13
Prepared in our lab Prepared rn our lab
N/A N/A
105 000 101000
62 000 70 000
0.40 0.31
1.49 1.04 1.97 1.30 0.79
400 400 900 800 500
aNumbers in square brackets refer to the structures provided in the text. bAbsolute weight average molecular weights as prowded by the suppliers of the polymer samples. ‘Data obtained by GPC in chloroform relative to polystyrene standards without further correction. The intrinsic wscostty [q] was determtned in chloroform at 30°C in an lJbbelohdetypeviscometer.The intrinsicviscosity [q] is indL/g. Owing tothe insolubilityof PGAin common organtc solventsand owingtothe limited availability of the poly(ortho esters) these polymer samples were not independently characterized. The polyanhydrides were too unstable in solutton to give reliable results. dDTH: Dat-Tyr-Hex, see Structure Il.
Table 2
Thermal properties of the biodegradable polymers
Polymera
M
T,(C)
W
Poly(ottho esters) t-CDM: 1,6-HD = 35:65 [ 1a] t-CDM: 1.6-HD = 70:30 [ 1 b] t-CDM:1,6-HD = 9O:lO [lc] Poly(glycolic acid) (21 Poly(lactlc acids) [3] L-PLA L-PLA L-PLA D,L-PLA D,L-PLA D,L-PLA Poly(fi-hydroxybutyrate) [4] Homopolymer (0 mol% HV) [4a] Copolymer (7 mol% HV) [4b] Copolymer (1 1 mol% HV) [4c] Copolymer (22 mot% HV) [4d] Poly(e-caprolactone) [5] Polyanhydrides Poly(CPP-SA-IS0 anhydride) [6] Poly(SA-HDA anhydride) [J] Poly(trimethylene carbonate) (81 Polyiminocarbonates Poly(BPA-iminocarbonate) [lo] Poly(DTH-tmmocarbonate) [l 11’
99 700 101000 131000 50 000
55 a4 95 35
50 000 100 000 300 000 21000 107 000 550 000
54 58 59 50 51 53
370 450 529 227 44
000 000 000 000 000
31 000 142 000 48 000 105 000 101000
-1 2 -5 ~62
-15 69 55
Jd(OC)
H,(Jg-’
210
358 362 338 254
Amorphous Amorphous Amorphous 71
170 159 178
242 235 255 255 254 255
41 20 39 Amorphous Amorphous Amorphous
171 160 145 137 57
252 243 235 251 350
51 32 12 7 34
46 49
297 292 261
1.2 2.5 Amorphous
135 138
Amorphous Amorphous
aNumbers in square brackets refer to the structures prowded in the text. bX, was calculated from H,. based on a calibration value of 72.3 J/g determined for PGA with a crystallinity of 52% (Reference 55) ‘DTH: Dat-Tyr-Hex.
294
Biomaterials
199 1, Vol 12 April
)
T,,,(“C)
XJWb
52 30 15 29
Physico-mechanical
Table 3
Mechanical
properties
of btodegradable
Polymer”
of degradable
polymers.
I Engelberg
and
J. Kohn
polymers
strength
modulus
Flexural modulus”
(MPa)
(MPa)
(MPa)
20 19 27 N/A
820 800 1150 N/A
28 50 48 N/A 29 35
Mb%
Poly(ortho esters) t-CDM:l.&HD = 35:65 [la] t-CDM:1,6-HD = 70:30 [1 b] t-CDM: 1.6.HD = 90: 10 [ 1c] Poly(glycollc actd) [2] Poly(lactlc acids) [3] L-PLA L-PLA L-PLA D,r-PLA D.L-PLA D.L-PLA Poly(/&-hydroxybutyrate) /4] Homopolymer (0 mol% HV) [4a] Copolymer (7 mol% HV) [4b] Copolymer (1 1 mol% HV) [4c] Copolymer (22 mol% HV) [4d] Poly(E-caprolactone) [5] Polyanhydndes PolyjCPP-SA-IS0 anhydride) 161 Poly(SA-HDA anhydrlde) [7] Poly(tnmethylene carbonate) 18) Polylmwxarbonates PolyjBPA-lmlnocarbonate) [lo] PolyjDTH-lmvwcarbonate) [ 1 11’
propertn?s
TellSlIe
Elongation Yteld (‘%,j
Break
1250 N/A
41 41 34 NA
220 180 7.0 N/A
1200 2700 3000 N/A 1900 2400
1400 3000 3250 N/A 1950 2350
37 26 18 N ‘A 40 35
6.0 3.3 20 N/A 6.0 5.0
36 24 20 16 16
2500 1400 1100 620 400
2850 1600 1300 750 500
142 000 48 000
N/A 4 05
N/A 45
N/A N/A N/A
105 000 101 000
50 40
2150 1630
2400 N/A
99 700 101000 131 700 50 000 50
000
100
000
300
000
21000
107 000 550 000 370
000
450
000
529
000
227
000
44
000
31
000
950 1000
wi
22 2.3 55 8.5 70 N
2.5 2.8 17 36 80
‘A
N/A 85 160
14 20 3.5 35
4.0 7.0
“Numbers KI square brackets refer to the structures provided an the text. bFlexural storage modulus as measured by DMA at room temperatures (23°C) ‘DTH- Dat-Tyr-Hex.
a degradable,
biocompatible,
by microorganisms”. whose
function
energy=?.
thermoplastic
is to
provide
constituent
made
polymer,
of carbon
by soil bacteria28,2g.
to o-3-hydroxybutyric
of human
storage
a reserve
PHI3 can be degraded
PHB degrades
polyester
It is an intracellular
acid which
blood. The low toxicity
PHB
In vivo,
is a normal
of PHB may at
Poly(&-caprolactone) polyester
that
can
degraded
be
:
4 a 4 b 4 c
d
4
homopolymer
of HB.
IHVI
Y = 0
copolymer of HB and HV.
: :
acid
(Structure
been
biomateria13’.
Initially,
by
microorganisms,
of PCL as a biodegradable
certain
circumstances,
enzymatically,
erosion33.
Low-molecular-weight
reportedly
taken
to
The
degradation
available
= 2’2
design
under the trade
PHB homopolymer the copolymers crystalline, potential
more flexible such
skin,
paramedical
as
as well
disposables.
can
be
surface
of
PCL
and degraded
are
intra-
of
long-term,
implantable
slower
most suitable drug-delivery
than
for the systems
contraceptive
device3’.
and
for the manuof sugars
by the structure
PHB and its copolymers
5: Poly(E-caprolactone)
acid are now commercially Biopol’M26.
is very crystalline
Since PCL is a semicrystalline and brittle whilst
about
acid (HV) are less
temperature.
and more readily processiblez6.
of these
by a
acid
process
of PH B with hydroxyvaleric
applications
applications artificial
name
PCL
of PCL IS significantly
Y (mole%i)
with up to 30% of 3-hydroxyvaleric
the
cellularly34.
of HB and HV.
eutrophus.
to
materia13’.
enzymatic
fragments
up by macrophages
a
conditions3*.
cross-linked
leading
copolymer
Alcaligenes
leading
physiological
that of PGA or PLA. PCL IS therefore
of PHB, based on the fermentation
bacterium
as
under
= 11
a biosynthetic
aliphatic that PCL
packaging
Y (mole%)
ICI developed
is an
that PCL can also be degraded
mechanism
such as CapronorTM, a 1 yr implantable
facture
4).
investigated
it was recognized
of HB and HV.
with hydroxyvaleric
5)
intensively
copolymer
copolymers
of 7,
(Structure
Later, it was discovered
Y [molen/o) = 7
Structure 4: Poly(f3-hydroxybutyrate)
HV contents
hydrolytic degraded
hydroxyvaleric
(PCL)
has
potential
Under . (HB)
with
were characterized
Poly(E-caprolactone)
evaluation
acid
copolymers
respectively,
and
least in part be due to this fact.
hydroxvbutyric
and three
1 1 and 22%,
polymers
controlled as industrial
drug
include release,
applications
The
biomedical sutures, such
-6O”C,
PCL is always Among
this is an unusual
polymer
the more common
property,
which
to the very high permeability
with a low Ts of
in a rubbery
state
allphatic
undoubtedly
at room
polyesters, contributes
of PCL for many therapeutic
drugs35.
as
Another form
interesting
compatible
blends
property with
of PCL is its propensity a
B/omatenals
wide
range
199 I, Vol
of
12 Apr//
to
other
295
Physico-mechanical
properties
polymers36. merized
In
with
oxide,
addition,
noteworthy
other
THF,
methyl
part of the evaluation non-toxic
be
(e.g.,
of Capronor”.
and lactic acid
38. extensively
and tissue-compatible
Drug Administration
materials.
has been
regarded
(FDA)-approved
as
US Food and
phase
I and phase
II
It is interesting has so far been
to note that despite
predominantly
drug-delivery
used as a biodegradable
reviewed
its versatility,
considered
applications. staple
blends and copolymers applications
aliphatic under
PCL
carbonate)
polycarbonate,
physiological
PCL is being
and it is likely that PCL or
with PCL will find additional
in the future. The current
addition,
a few reports
medical
merizations
(carbonate-ester)
and
found
were
to
applications.
carbonate
be
their
0
synthesis
hydrolytically potentially
limitation
who
of the
polyanhydrides
has
most
reactive
and The
Because
without
of their
degrade
erosion
and a
high rate of by a surface
the need to incorporate
into the formulation.
of polyanhydrides
surface
use of
used as biomaterials.
polyanhydrides
mechanism
the
A study
rate is both a blessing
of polyanhydrides.
catalysts or excipients
industrial
as a potential
suggested
the
polymers
when
various
This is a potential
over poly(ortho
only
esters),
various
which
additives
amines
have been shown
during
high-temperature
reactivity of the polymer
must therefore
be considered
controlled
release
reactive
Despite
amino
these
been approved clinical
controlled
trials
when
processing43.
The
using
of peptides,
aliphatic
carbonate)
research,
of aliphatic
use in the
however,
therefore
polycarbonates
industry.
points
have pre-
In biomaterials
of 40-60°C
are not necessarily PTMC
extremely
points and the low
polycarbonates
plastics
softening
strength included
become
The low softening
strength
their
mechanical
and low
disadvantages.
nucleophiles
polyanhydrides
proteins
The commercially derived
from
very stable
strength
used polycarbonate
Bisphenol
polymer
physiological carbonate)
A (BPA).
that is virtually
conditions.
Important
are its excellent
derived
is an aromatic non-degradable properties
processibility,
and its exceptional
shatter
group
modification
decreases
without
significantly
of poly(BPA-
its high mechanical
materia15’.
forms hydrolytically
degradable
in appearance
and
oxygen
stability of the polymer
the mechanical
Consequently,
is derived
the carbonyl
9 and 10). This backbone
the hydrolytic
affecting
properties
of malignant
anhydride),
bone structures,
films and fibres that are very
mechanical
strength
to those
made of poly(BPA-carbonate).
Biomaterials
characterized
Poly(CPP-SA-IS0
199 1, Vol
for agent)
the in
brain tumours44. anhydride)
two polyanhydrides
6:
in phase
devices
(a chemotherapeutic
poly(CPP-SA-IS0
were
Stmctu,e
delivery
12 April
anhydride)
CH3
Jn
and poly-
with different
(Structures
L
back-
6 and 7).
of
poly(BPA-iminocarbonate)
and sebacic acid have
are being tested
is a under
resistance47-4s.
by replacing
(Structures
by an imino
polymer
Poly(BPA-carbonate)
In a formal sense, poly(BPA-iminocarbonate)
the
or drugs
polyanhydrides
propane
of BCNU
We
in our study.
Polyiminocarbonates
similar
limitations,
polyanhydrides
In our study,
296
poly-
matrix toward
as implantable
release
the treatment (SA-HDA
n
by the FDA for clinical trials in a remarkably
short time. These III
to react with
groups.
from bis-p-(carboxyphenoxy)
in several
I-
Poly(trlmethylene
from poly(BPA-carbonate)
anhydrides
containing
mechanical vented
are
incorporated. Several
in copolyoxide. These
and Langer42.
among
high degradation many
for
biomaterials.
by Domb
are
unstable
degradation,
Generally,
soft at about 40-60°C.
1958
was later recognized
been published
8:
stability
high-molecular-weight
Polyanhydrides
In the
by Hill and possible
as degradable
of
in detail
for
et al.4’
by Langer
polyanhydrides
slow
describing
have been evaluated
of PCL was used in
in
major
extremely
as a monomer
an
degrade
communication).
lactide or ethylene
copolymers
Their low hydrolytic
considered fibres4’.
This feature
advantage
for the
albeit with personal
8).
to
status of PCL has been
sample
first investigated
as textile
potential
found
applications45.46.
Structure were
applications
show
(Structure
recently
II
Carothers3’
advantage
(PTMC)
(CH&-0-C-O
Polyanhydrides
erosion
anhydride)
have been published
with glycolide,
drug delivery
Polyanhydrides
limitation
was
conditions,
use of trimethylene
for controlled-
In Europe,
by Pitt3’. A commercial
recently
Poly(SA-HDA
carbonate)
this study.
was
7:
rates (C.G. Pitt and K.J. Zhu,
clinical trials3’.
release
Poly(trimethylene Poly(trimethylene
Consequently,
undergoing
Structure
as
Based on a large number
and PCL are currently
system
copolyethylene
Particularly
of c-caprolactone
extensively37,
and J. Kohn
4-vinylanisole,
vinylacetate).
of PCL has been studied
of tests, c-caprolactone the Caprono?’
can
S-valerolactone,
are copolymers
The toxicology
I. Engelberg
monomers
methacrylate,
that have been studied
polymers:
c-caprolactone
numerous
chloroprene,
styrene,
of degradable
Structure 9: Poly(Bispheno1 A carbonate)
Structure 10: Poly(Bispheno1 A iminocarbonate)
Physico-mechanical
Owing tothe and
mechanical
find some
combination strength,
applications
of degradability,
as a disposable
view of the growing
concern
degradable
plastics
in the environment,
degradable
materials
The
suggested5’,
physically
entrapped
pression moulded
within
poly( BPA-iminocarbonate) upon subcutaneous
biocompatibility
of various
dyes
films
com-
52. Although
in mice and rabbits5’,
the
of poly(BPA-iminocarbonate)
has
so far not been established. To reduce carbonate), tyrosine formed
the
we
potential
have
dipeptide5’,53. between
tyrosine
These
iminocarbonate-amide also be regarded
with
derivatives
iminocarbonate hydroxyl groups
polymers
(Structure
as pseudo-poly(amino
was
present at the
tyrosine-derived
copolymers
of
bond
are
7 1) that can
acids)54.
Films were prepared
0
by solvent casting and/or
depending
cast from
10%
temperature
glass
plates.
vacuum
on the specific
(w/v)
room
between
After
mm
drying,
were
For compression
thermostated
Polymers
were
placed
0.2
tonnes
were
were
A=0
1n
0
Structure
These
data,
Tab/es
l-3.
of
dynamic
In general,
pseudo-poly(amino
acids)
when amino acids are linked together
more
favourable
conventional The
poly(amino
tyrosine-derived resulting Based
properties
monomers
these
investigations,
(Structure
1 7)
promising
polymer
in the formulation devices
The carbonate),
of the
the
has
properties
been
of
investigated53.
poly(Dat-Tyr-Hex been that
currently
of high-strength
as
a
being
orthopaedic
polyiminocarbonates were 100
included and
synthesized
in
this
study,
poly(Dat-Tyr-Hex-imino-
in our laboratory
and had mol
000.
in this study
1. Polymers
were
and their used
to
sources
as received.
are The
standards. were
without
polymer
correction
in chloroform,
the
was used to determine
the degree a/,55 for polymers
were
thermal
determined
were
per
with indium
from the Hf
on a DuPont
The 951
Cohn et Td of all TGA
under an atmosphere
storage
made
of PGA and PLA.
of 52%.
as detected
heating materials
as determined
at a
of nitrogen.
onset
points
for
by TGA. The measure-
modulus
(E’) as a function
using a DuPont
mode (1 Hz) at a heating
AND
four
specimens
was calculated Jg-’
were
and the heat of fusion
in Tab/e 2 represent
of the flexural
fixed frequency
(X,)
a crystallinity
decomposition
temperature
(T,)
value of 72.3
reported
values
of at least
For crystalline
as well. For samples
rate of 20”C’min
The values
average
tensile
determined
the Tg and T,. The standard
temperature
with
were
9 10 DSC calibrated
of crystallinity
PGA
temperature
strength,
break
was 1 O”C/min.
(H,) were determined
were made using an at room
from four separate
A DuPont
and
relative
for the hydro-
In all cases, tensile
arithmetic
obtained
rate for all polymers
and
by GPC
Number
[q] of all polymers
Tensile
at yield
in
distri-
using an Ubbelohde
1122
D882-83.
conditions.
from sample.
983
DMA
of
in the
rate of 5”C/min.
DISCUSSION
characterization
The suppliers absolute
for
listed
calculated
viscosity
model
ASTM
measurements
Sample included
tester
are
weight
in our laboratory
averages
The intrinsic
elongation
Materials in Table
cut
of their samples.
the molecular
were analysed
standards
tensile
calculated
ments
were
suppliers,
Tensile measurements
heating
mm
weights
by the
at 30°C
RESULTS
shown
for
were asked to provide the
type viscometer.
EXPERIMENTAL
The polymers
0.3-0.6
molecular
weight
using a calibration
imino-
identified is
the
polymers
measurements.
relative to polystyrene
the crystallization
of several
such as bone pins and screws.
poly(BPA-iminocarbonate) wts of about
many
structure
has
particularly evaluated
the
and
carbonate)
fixation
than
acids)54. between
polyiminocarbonates on
acids) tend to retain
of the amino acids whilst exhibiting
engineering
relationship
bonds. lt
polymers
to crystalline
In order to obtain
under ambient
obtained
polymers).
was determined
modulus,
by non-amide
was found that such pseudo-poly(amino the good biocompatibility
are
(for crystalline
samples
as provided
volume.
according iminocarbonate)
and
or 10°C
to amorphous
about
average
molecular
lnstron
11: poly[Dat-Tyr-Hex
polymers)
(T,)
applied
of polymer
to polystyrene
CH,
used.
mould
characterization weight
weight
Icy5
steel
applied
tests and DMA
in chloroform
I
a stainless
press
was
a thickness
bution, all polymers
NH
laboratory
platens
with
absolute
O-&O_
a Carver
10 min. The mould then was rapidly cooled. Samples
All suppliers
u
about
tests.
about
Materials
0
under
and
length,
above Tg (for amorphous
of 2 tonnes
stored
40 mm
above the melting temperature Loads
were
heated
into
at
d). Strips of
moulding,
with
heated to 30°C
films
chloride
chloride-treated
was reached (2-3
cut for mechanical
equipped
Films were
in methylene
the
weight
compression
polymer.
trimethylsilyl
thickness,
mm width
mechanical
CH2-CH&H-CH-CH2
solutions
until constant
0.1 5-0.20
and
f
was
preparation
moulding,
5-7
of poly(BPA-imino-
BPA
The
the phenolic
side chains.
toxicity
replaced
by Li and
Trimethylsilyl
from Aldrich.
Sample
was found to be tissue compatible
implantation
grade.
of has
and
of HPLC
reported
(used to coat glass moulds for solvent casting)
obtained
were
was
chloride
of non-
solvents
and J. Kohn
In
material.
discs has been investigated5’,
of polyiminocarbonates
I fngelberg
All
as biomaterials
solvent-cast
polymers:
Kohn5’.
important.
5’ and the release
of degradable
may
the development
increasingly
use of polyiminocarbonates
also been
ultimate
plastic
about the accumulation
becomes
synthesis
low toxicity
poly(BPA-iminocarbonate)
properties
of polymer
weight
average
samples
were asked to provide the
molecular
Biomaterials
weight
of their samples.
199 1. Vol
12 April
297
Physico-mechanical
propwties
of degradable
polymers:
I. Engelberg
We assumed
these values to be correct and included
Tables
Since
7-3.
polymers
depend
on the
molecular
polymer
sample
the
physico-mechanical
by GPC, relative
The relative molecular
weights
in Tab/e 1. These
listed
hydrodynamic
volume
significantly by the
distribution,
were
from the absolute
suppliers.
calculatethe
Our
polydispersity
results weight were
with
polymer
samples
low polydispersity.
exceptions
e.g., the
and its copolymers more
detail
There
with
below.
we
(limiting
viscosity
samples.
The intrinsic viscosity
data are discussed
in
Since
the
(SSOC),
below
body
(1,6-HD)
esters)
having
dimethanol
were obtained
(t-CDM)
polymer
physico-mechanical
were
expected
observation unit
to
of t-CDM
can possibly the
increase
be explained
backbone.
to permit
The onset
sharply
the
by the restriction
thermal
molecular
dialcohols stability
with
similar
of the
stability of the poly(orthoesters)
polyesters The 1,6-HD.
thermal
mechanical were
The
strength,
should,
cleavage,
techniques
at the same
when designing
applications. readily
processible
moulding
with as
by
and formed between
decomposition suitable
be readily
such
and
70:30
the
indicates precautions
processible
injection
by melt
moulding
or
extrusion. We
noticed
that
over drying
poly(ortho
weight.
our observation
storage
at ambient
had
completely
esters)
molecular
after
agent
Although
for about
10
temperature, degraded
month all three
to
we did not investigate
may indicate during
that some prolonged
a
poly(ortho storage
is rigorously
low
further, esters)
at ambient
excluded.
PGA above are
reactivities, was
not
Since we were
unable to obtain high-molecular-weight
for our study,
we
Polysciences
had to use a commercial
Inc.) with
a nominal
sample
PGA (from
mol wt of only 50 000.
that
was significantly other
aliphatic
failed
to 9O:lO.
significantly
strength
of t-CDM.
of poly(ortho
degree
of copolymers
esters)
to 1,6-HD
small the A
copolymer
tensile
was
modulus
42%
than the
was that the 90: 10
elongation,
indicating
a
It thus seems
I
is the price for the increased containing
in the ratio of large
of the corresponding
Figure of
12 April
1
Determination
temperature
molecular B: ratio HD
199 1. Vol
60
100
80
120
,Temperature ("Cl
that in this type
will result in relatively
properties
40
a high proportion
small increases
above 70:30
in the mechanical
The between
the t-CDM/1,6-HD
our results indicate
relatively
of
not significant.
of brittleness.
copolymers.
Biomaterials
when
lower
in brittleness
Furthermore,
were
difference
at a much
increased
that an increase
elongation.
This
higher
Another
three
ratios
and modulus
only occurred
copolymer.
the
in terms of their tensile
copolymers
increased
of
t-CDM/1,6-HD
and
strength
and had a 44%
copolymer
changes
polymers
hydrolytic
even if moisture
esters)
(Tab/e 3)
having
modulus
change
ratio was
t-CDM
the
related to the ratio of t-CDM/
were very similar
in tensile
significant
tensile
than
films. The large safety margin
temperature,
and t-CDM
of most
properties
and the 35:65
70:30
stability
copolymers
tensile
differences
stronger
Con-
degraded
were
copolymers
ratio. It is noteworthy
not linearly
and 70:30
70:30
implantation.
after
may lose its stiffness
for practical
Ts
initial
drop
may be important
copolymers
low
(e.g. PLA) in this study.
copolymers 35:65
tested
tends to
service in viva.
circumstances
implantation
stability
was well
poly(ortho
thethermal
the
after
have limited
chemical
on the t-CDM/1,6-HD
than
of
motion.
decomposition
dependent better
faster
of water
must be heated to a higher
all three copolymers. Since 1,6-HD
300”Cfor
both aliphatic
certain
(37°C)
of the cyclic t-
Consequently,
extensive
of thermal
an
Ts
drug-delivery
has a relatively
copolymer
copolymers
these
of the
with
and the
into the polymer
during
even if both copolymers
fabrication
(Tab/e 2). This
by the presence
a larger ratio of t-CDM
temperature
298
of
function
in the polymer
motion caused in
containing
the
properties
temperature
exceeded
imbibition
Ts and the onset point of thermal against
Since t-CDM
is a linear flexible
to be a sensitive
found
portion
the molecular
temperature
rate. These considerations
sealed
Ts was
The
copolymer
the 65:35
copolymer,
that
1,6-hexanediol
and
1,6-HD
of trans-
composition.
increasing
CDM
whilst
the
copolymers
ratios
from SRI International.
is a rigid cyclic molecule, molecule,
different
and had a
copolymers
of drugs
its Tg may under
clear transparent
poly(ortho
the
in the formulated
either solvent casting or compression
esters)
cyclohexane
As
the temperature
the possible
65:35
much
All
Three
when
to the incorporation
value
specific
Poly(ortho
MPa).
for all three
reduce the Ts of the polymers
rigidity
studies.
E’ (1250 declined
After implantation,
polymer value
values of E’ (950
may be lower than the Tg of the pure bulk polymer.
[a] for most
used in our
similar and the
whilst the 90: 10 copolymer
matrix, the Ts of the polymer
sequently,
samples
yielded
copolymer
I).
intrinsic
[q] is an unambiguous
higher
sharply
the
the polymer
the 35:65
had nearly identical
E’ values
dropped
further
noteworthy
determined
number)
characterizes
In most
of poly(hydroxybutyrate)
HV. These
Finally,
viscosity
that uniquely
however,
rose, the
device
materials
of the E’ by DMA
MPa, respectively),
Owing
to
samples
distributions.
measurement
copolymer
1000
provided
it possible
well-defined
were,
ICI samples
70:30
(Figure
differed
of the polymer
and to assess their molecular
The
results. At room temperature,
significantly
are
for the
weights
made
(M,/M,)
cases,
the
standards.
and therefore
molecular
GPC
each
in our laboratory not corrected
of the polymer
of
but also
analysed
to polystyrene
obtained
values
weight
we
J. Kohn
them in
properties
not only on the molecular weight
and
by
compositions. of
= 35:65.
t-CDM
of the flexural
DMA
to
for Curve
1,6-HD
three A: rat:0 = 70:30;
storage
modulus
poly(ortho
estersJ
of t-CDM Curve
(E’J as a function having
to 1,6-HD C: ratio
of
different
= 90: 10; Curve t-CDM
to
1.6
Phywxwnechanical
Owing to the insolubility
of PGA in common
we did not obtain the GPC and viscosity The insolubility impossible
of PGA in chlorinated
to prepare
procedure.
strength.
specimens,
Owing
value
and
hydrocarbons
to the
highest
m.p.
close to the value of 72.3 PGA of similar
lack
with previously
made it
reliably for
of suitable
test
properties
(210°C)
results55,
among
all tested
Jg-‘, which is very
was 7 1
Jg-’
molecular
reported
PGA had the highest H,
to be semicrystalline.
the
that had been
reported
for r
weight55.
2(#
PIA difference
and the stereoregular whilst
between
L-PLA
of L-PLA could
illustrated
for a sample
(Figure
2).
Upon
heating,
observed
be readily
different
Thermograms
was quenched
resulting
in the
solid.
second
heating,
Upon
crystallization
exotherm.
be observed resulted polymer.
molecular melting
initial
transition,
test
B). Slow
178”C,
respectively;
000
and 300
had the lowest
inconsistent
is provided
samples
of
Likewise,
behaviour
had
distributions
significantly (f 1 O’C)
temperature
in the range of 50-59°C. In addition,
molecular
for any given
L-PLA was somewhat changes,
however,
irrespective versus
or the
were
unable
evaluation
= 21 000).
to obtain of
improved
weight.
D,L-PLA
mechanically
Compression
moulded testing.
samples
formed
with
Thus,
no
was allowed
run).
to slowly
crystallization.
the
thermogram
was
its strength strength
and modulus
degradable L-PLA.
tensile
increased
with
value
3000
of
the Tg of the
rather
insignificant.
A
to be 55°C
or
strength,
suitable
increasing and
storage
for the
of this polymer
by solvent strips.
too brittle for reliable
quantitative
data
for
the
surpassed
(+26%). increased
polymer
for high-
implants
(e.g.,
weight
Also
and reached No
in terms is
of tensile
one
available.
a
other
of tensile
in terms
L-PLA
of
moduli
of
the
Among
the
in our study, only the polyiminocarbonates strength
of L-PLA. Devices
moulding
or solvent
made of
casting
were
on the degree of crystalllnlty
during
processing.
Our results
of good mechanical
properties
of either D,L-PLA or L-PLA requires a minimum
mol wt of about
100
A limitation weight
000.
of all samples
Although
there
(irrespective
of
was their pronounced
all specimens
behaviour
L-PLA showed stress-strain
of PLA
and stereoregularity)
low elongations,
tended
were
subtle
break
at
differences
to
in
of L-PLA versus D,L-PLA. Whilst
a purely elastic deformation
for most of the
curve, D,L-PLA was more ductile and exhibited
a significantly
larger
3). These
can be attributed stereoregular
storage
respectively.
currently
to turbid, depending
from samples
(Figure
the
increase
to D.L-PLA. Based on
L-PLA
modulus.
by the specimen
relatively
and
significant
flexural
MPa,
indicate that the achievement
brittleness.
D,L-PLA
and
molecular
3250
polymers
the tensile
transparent
weight samples
module
by compression
molecular
casting Increasing
as orthopaedic
high-molecular-weight
of the polymer of the
as compared
a
that
by solvent
and modulus
such
in our study
obtained
was
polymer
bone nails). In our series of three test samples
the
L-PLA
000
moulding.
of
molecular
of L-PLA very significantly
applications
modulus
for Tg
107
these results, L-PLA is clearly the preferred
polymer
of the polymer.
of
processible
in a further
(+20%)
D,L-PLA are properties
increasing
wt
films
resulted
strength
included
weight,
strong
000
polymers
weight
mol
discs by compression
mol wt to 550 in tensile
a
with
yet readily
transparent,
transparent
approached
molecular
were
1 OO”C/min.
a spontaneous
the mechanical
dramatically
strong,
(Tab/e 1). Overall,
low-molecular-weight
Films prepared
at
showmg
m spontaneous
(third
= 50 000) A: First run of
is semicrystalline.
melt
of low-molecular-weight
D,L-PLA
is a representative
casting were sticky and could not be cut into uniform mechanical
properties
the stress-strain
mechanical
re-analysed
in Tab/e 3. However,
medical
of the stereoregularity
L-PLA)
the
C: The polymer
resulting
(M,
run.
strongest
value for the Tg of PLA appeared
representative
mechanical
for the
higher than that of D,L-
were
from
amorphous.
Curve
5”C/min,
was
first
was
at 89°C.
at
material
to the
mol wt
molecular
There was a small tendency molecular
specimen. We
000
of D,L-PLA and L-PLA were
to increase with increasing
(D,L-PLA
100
000 of the
of L-PLA.
The Tg of all tested samples
(+5”C),
the 100
different
and polydispersities that 1 70°C
value for the melting
PLA. These
a mol wt of
of crystallinity
of the
and
by our GPC results, which indicate that all
L-PLA
our results indicate
stereoregular
and their
L-PLA samples
of L-PLA (Tab/e 2). An explanation
three specimens apparently
the
had m.p. of 170
degree
melt
acid)
by DSC. Curve
Inc. The material
quenching
the material
The crystallinity
between
of L-PLA with
had a lower m.p. at 159°C.
mol wt sample
of the
to the first
specimens
000
the sample
cooling
state
found
points or their degree of crystallinity.
mol wt of 50 000
the
rapid
included
only (Curve C). was
of the three
the could
during cooling and in is identical
endotherm
the
Polysciences,
after
exotherm
from
of poly(L-lactic
as determmed
be
molten
endotherm
semicrystalline
no correlation
weights
could the
of an amorphous
(Curve
from
run
to quenching
When
histories
semicrystalline.
glass
crystallization
the melting
For L-PLA,
weight
the
Thus, the third thermogram
and shows
sample
formation
as obtained 6: Second
identical
by rapid cooling at a rate of
in the thermogram of the
cool
by DSC, as
From
and the melting
in spontaneous
the regeneration
A).
Curve
of samples
thermal
c~stallization
a mol wt of 50 000 endotherm
(Curve
100”C/min,
(M,
2
having
Owing
The crystallization
was
melting
D,L-PLA
is amorphous,
observed
the polymer
only the
in the thermogram
state, the polymer
100
material.
of L-PLA with
As obtained,
first
the racemic
is that D,L-PLA
L-PLA is a semicrystalline
behaviour
with
Figure
polymer
The most important
J. Kahn
solvent-casting
for the mechanical
The H, of our sample
materials.
and
in Table 3.
included
In correspondence PGAwasfound
I Engelberg
of PGA yielded compression
noquantitativedata
of PGA were
polymers:
organic solvents,
discs that were too brittle to be tested
mechanical
of degradable
data for this sample.
films by our standard
The available sample
moulded
properties
proportion
differences
of
plastic
deformation
in the stress-strain
to the presence
of a crystalline
behaviour phase in the
L-PLA.
B~omater/als
199 1, Vol
12 Apnl
299
Physico-mechanical
propetiies of degradable polymers: I. Engelberg and J. Kohn
Molecular
60 1
Weight 120
475
(weight average) 17 7.6
1
3.8
D ‘..._,
10
0
2
1
3
4
6
5
Elongation
(%I
Figure 3 Stress-strain cwves for various samples of PLA, Curves A and 6 were obtained from samples ofpoly(L-lactic acid) with mol wt of (A) 300 000 and (R) 100 000. The cwves indicate mostly elastic deformation. Curves C and D were obtained from pOiy(D,L-lactic acid) with mol wt of(C) 550 000 and(D) 107 000. These curves show both elastic and plastic deformation.
PBH and copolymers GPC analysis
with
(up to 35% (Figure
fraction
by weight) to
be
molecular-weight calculated from curve were
consisted
weights
the
provided
molecular
fractions. the entire
significantly
The true molecular values
The
low
viscosity
7) was
significant test
intrinsic
a further
amount
samples.
found for all ICI samples biosynthetic
origin
significant different
lots
limitations, about among
average
of the
several
increasing
content
copolymers
(Table 2). systems.
decreased, This
A qualitatively
copolymer
composition
had been
glycolic
reported
and
modulus
HV
elongation
was
of 2.8%,
made
highest
ductility.
all other
and
solvent-
polymers.
The
with HV decreased
with
with
optimum polymers
of HV
would
of the
defined
many
relationship
for copolymers
as
well
cantly
copolymer
between and
the
melting
of lactic acid and
DMA.
with
the
onset 252”C,
point
of the
as determined
199 1, Vol 12 April
when
thermal
decom-
by TGA, some
HV content
that
molecular
polymer
weight
PH B was kept in the
similar
about
had the brought
reduction
of
Our results indicate
in the
as the
DMA
polymers
for a wide range properties
in polydispersity,
samples
with
more
of one
narrowly
would exhibit signifi-
properties. results
were
E’, as a function
E’ was significantly
temperature.
of 22%
tensile
distributions
mechanical of the
the
to variations
improved
seen
occurred
reached
and toughness
Since
Qualitatively
gradually
at a low
improvement
the better ductility
and tensile modulus.
obtained
from
of temperature
higher in the homopolymer
than in the three copolymers. were
failed
of HV led to a significant
are sensitive
measurement
and
of the copolymer
applications.
expect
brittle
in this group of polymers
balance of strength
of possible
MPa) and the
but it was very brittle
(Tab/e 3). The copolymer
(2.5%) quite
HV the homo-
(36
with HV content of about 1 1% may have the
that copolymers
22%
reduced
MPa),
Unfortunately,
the tensile strength
As the crystallinity for
(2500
the HV content
about by the presence (Table2)
that the Tg
of the molecular
similar to PHB. A significant
were
PGA
strength
also
The copolymer
its crystallinity
of PHB was observed
Biomaterials
tensile
7%
or
with 3-HV were
with
tensile
Since the measurements
taken above the Tg of the four polymers,
of PHB was
degradation
to the copolymers
had the highest
and failed at low elongation with
for
moulding
of the copolymer.
As compared
13.5%.
T,,, were
similar
composition
lower
suited
to 2°C. This result indicated
these
of H, after
such as injection
was rather independent
of
by the standard
is typical
range -5
In
acid”.
Although position
the
behaviour
in a narrow
of these copolymers
highest
to be better
The Tg of PH B and its copolymers
extrusion.
polymer
HV have substantially
therefore
techniques
in ductility and toughness
of HV. The copolymer amorphous.
melt processing
only when
spite
with
points and appear
between
value
of copolymers
The copolymers
melting
variation
in this study for
are not
HV.
highest films
of
for about 1 h56. Thus, the melting temperature
that there may be a
batch
with
of the
samples
valid observations
employed
was almost completely
300
polymer.
included
degree of crystallinity
point
same
to form
procedure
distribution
weights
stability.
of a in the
be a reflection
molecular
to
samples
material
weight
we assume
had the second
the polymers
ICI
high.
presence
In polymer
batch
generally
too crystalline
casting
extremely
of all
could possibly
PHB and its copolymers PHB
was
of
were
molecular
In addition,
degree
high-
molecular weights, weight distribution
of the
of the polymer.
this kind, conventional very meaningful.
indication
bimodal
melt at 190°C
of PHB at 17 1 “C is very close to the upper limit of its thermal
of the
of low-molecular-weight
The
Time (min)
Figure 4 GPC of the copolymer of hydrowybutyric acid with 7 mol% of HV. Both retention time (lower axis) and the relative weight average molecular weight (X 1000) based on polystyrene standards (upper axis) are given. The chromatogram shows a bimodal molecular weight distribution with a high molecular weight fraction and about 36% (by weight) of low molecular weight material. Similar molecular weight distributions were found for all ICI samples tested (homopolymer of hydroxybotyric acid, Lot No. BX G04; copolymer of hydroxybutyric acid and 7 mol% of HV. Lot No. BX PSM4: copolymer of hydroxybutyric acid and 13.5 mol% of HV. Lot No. BX PO 10: copolymer of hydroxybutyric acid and 22 mol% of HK Lot No. BX PSM5/1).
amount
(Table1) and in all samples
lower
polydispersity
relatively
Retention
Z’O
material
by the supplier
weights
the calculated (Table
18
is
essentially
and a significant
of low-molecular-weight
4). The molecular found
14
of PHB and its copolymers
that all four samples
of a high-molecular-weight
were
HV
of the ICI samples
with HV revealed
li
7
the by
(PHB) were
no sharp transitions
curves.
Instead,
became
softer
E’ decreased with
increasing
Physico-mechanical
PC1
compression 0.5
PCL exhibited
several
the other tested exceptionally Another 235 typical
property
all other
The sample weight
to reason
increase weight
than
aliphatic
modulus
us
with
increasing
the mechanical than
leaves
PLA.
Although
nearly
90°C
that
observation
most of the other
polyesters
were
a
in the glassy
be
state at PLA and
state at room
were
unable
solvent,
the
phenomenon polymer. films.
to
Overall,
films
polymer
layer
was
to the other
a
broke
aliphatic
by
our
evaporation
into
pieces.
were
tested.
opaque,
unusual
flexible
polymer
as
suitable
Contrary
Despite
films
molecular
weight.
are therefore
was
The
second
more
stable
The
although weight
at the time
the tests
some
qualitative were
crystalline
conclusions.
HDA
anhydride)
was a pliable break.
analogue
of about
material Because
bonate)
and
of
and formed
clear, was
the
testing,
135°C
for
position
temperature
with
the
Transparent
the
known
ease.
between
the two
stability
against
of
a Td of
indicate
a decom-
for poly( BPA-carbonate).
obtained
from
FT-IR spectro-
samples,
we assume
of poly(BPA-iminocarbonate)
of
the
at temperatures group
indis-
amorphous
exceptional
degraded
tendency
to
virtually
of the thermal
evidence
is due
of
iminocarbonate about
and an organic
‘$”
--(=&k-o+
130-l
cyanate
bond 50°C
(Scheme
to
into
a
1).
nrQ-O-C=N + HO+
Z
thermal
to derive
similar
and
at yield
strength
polyanhydrides
and tested
applications.
properties
of aliphatic
was
found
to be an extremely
films
were
readily
obtained
by
relatively
present
low
imposes
thermal
of
property
Polycarbonates,
on the imino
to
and an
stability
to be an inherent
lacking
group,
cannot
and are consequently
by
melt
stability
more
spectroscopic
by TGA
polymer
sample
avoided
under
during carefully
or injection temperatures,
occurred
analysrs
thermal
compression controlled moulding, would
only as
about
some 100°C.
80°C
for
of
the
degradation moulding
conditions. which
clearly
at about
that
as low
requires
the
of
Although
revealed
at temperatures
moulding,
poly(BPA-
techniques.
poly(BPA-iminocarbonate)
compression
of
on the processibility
fabrication
as measured occurred
thermal limitations
IR
degradation Since
limited
in general.
loss
135”C, It
is known group
stable.
material
weight
modulus.
type
iminocarbonate) this
Poly(SA-
the
of reaction
this
reaction
to a hydroxyl
appears
undergo The
Glass transitrons to 50°C.
Thus,
atom
thermally
small
had
function.
leading
the extra hydrogen
polyanhydrides with
dissociation
in polyurethanes,
higher
58, PTMC
with
of over 350°C
of partially
a
particularly
completely
Our TGA results
decomposition
hydroxyl
was
poly(BPA-iminocarbonate)
analysis
as
poly(BPA-iminocar-
differences
reduction
that the thermal
some
In correspondence
be
materials
regarded
were
films
of polyiminocarbonates
PTMC polycarbonates5’,
brittle
be
polymer
were
poly(BPA-iminocarbonate).
occur
molecular
materials
mechanical
two
this
striking
a sharp
isocyanate
two
to drug-delivery
however,
used poly(BPA-carbonate),
appearance,
strong
One of the most
poly(BPA-iminocarbonate)
it is possible
-50
can
polymers
weight
large elongations
low
in too
for adhesion
may, with
poly(BPA-carbonate) Both
formed
strength
that exhibited of their
or blends
physical
polymer residual
45-50°C.
tensile
reactivity,
to be limited
material.
By
molecular
polyanhydrides
in the range from had low
hydrolytic
two
by
PTMC
to
performed. The
except
material
of the widely
characterization
extrusion
soft
copolymers
for
be mechanically
application, The
weight
obtarned
after
anhydride)
for mechanical their
amorphous
The
temperatures
not observed
The
in mind,
may
of
poly(CPP-SA-IS0
initial
used
molecular values
volume
PTMC
surgery.
A similar
about
limitations
fractions.
of
unable
poly(SA-HDA
were
predominantly
were
high
were
not certain
tested
appear
which
sample amorphous.
of well-defined
for
anhydride).
the absolute
poly(BPA-iminocarbonate)
degradable
poly-
atmosphere
we were
results
tested
after
together
without
immediately
anhydride)
and had a higher
films
these
tested
polyanhydride,
we were
With
melting
Since
dissociate
and
polymers,
an inert
precautions,
test
recoil
Polyiminocarbonates
incomplete.
poly(CPP-SA-IS0
transparent
test specimens
were
these
most
for medical
processibility
to all other
of poly(CPP-SA-IS0
anhydride)
their
cast under
and specimens
preparation.
investigated
to prepare solvent
compositions
are the hydrolytically being
limited
were
dry nitrogen
than
molecular
property
degradation.
anhydrides
obtain
currently
This
it difficult
massive
different
The
of
as PLA or PHB.
polymers
polyesters:
The polyanhydrides
polymers
applications. made
that
Based on preliminary with
other.
load deformed
to stick
and the relative
any practical after
in block
scopic
polyanhydrides
unstable
to find
of the
Polyanhydrides Two
each
low
no elastic
hydrodynamic
It seems
tinguishable.
This
crystallinity
yielded
rather
PCL
upon
due to the high moulding
PCL
of
procedure;
is probably
Compression
compared
obtain
solvent-casting
virtually
had a tendency
between
a large
interesting.
We of the
such
the
temperature. standardized
a
up and
PCL is can
its Tg whilst
above
to
prevention useful
weight,
in part by the fact that PCL is in the rubbery
room temperature,
using
and J. Kohn
of low-molecular-
little doubt
This
point
weak it
of PCL would
molecular
properties
= 50 000)
weaker
strength
I. Engelberg
be rolled
was
by Amgen
chloroform.
little tensile
(Table 3).
There against
as determined
polyesters.
and exhibited
= 44 000)
tensile
pressed
40°C
could
and the films
The large difference
is more
(IV,
at
films
breaking.
polymers:
PTMC had a low Tg (- 14°C) and was completely
had Td between which
The
deformation when
stability.
to us was of intermediate
L-PLA (M,
explained
T,,, of 57°C.
low
polyesters
that the mechanical
with
significantly
its
of degradable
moulding
tonnes.
without
are its
of PCL available
somewhat
comparison
aliphatic
esters)
and a low
stands
and
among
noteworthy
PCL had a Td of 350°C
of poly(ortho
molecular
not found
Most
of PCL is its high thermal
tested
255”C,
strength
properties
polyesters.
Tg of -62°C
low
and
unusual
aliphatic
unusual
Whereas
and
propelties
could
only be
Processing
would
require
be impossible
by even
without
degradation. The replacement
isosteric
imino
strength
of
the
group
of the carbonyl had
polymer:
little
effect
group
by the nearly
on the
mechanical
poly(BPA-iminocarbonate)
Biomaterials
199 1, Vol
12 April
was
301
Physico-mechanical properties of degradable polymers: I. Engelberg and J. Kohn
60
SUMMARY
1
AND CONCLUSIONS in Tab/es
The data collected to compare widely
2 and 3 represent
the engineering
investigated,
large number necessity
properties
degradable
of available
be
biomaterials.
polymers,
incomplete.
In
addition,
that cannot
example,
the need to keep the experimental
is in direct
L
012345678
parameters
specimens
that exhibit
be readily
we
Figure 5 Stress-strain curves for two different polyiminocarbonates. Curve A was obtained from a sample of poly(BPA-iminocarbonate) with a mol wt of 105 000. The curve indicates that this polymer is brittle with mostly elastic deformation. Curve B was obtained from poly/Dat-Tyr-Heximinocarbonate) with a mol wt of 10 1 000. Poly(Dat-Tyr-Hex-iminocarbonate) is less brittle and exhibits both elastic and plastic deformation. (First published in reference 53 and reproduced here with permission).
and
as strong
as poly(BPA-carbonate)
poly(BPA-carbonate) its
tensile
bonate),
against
most striking two
of similar
modulus,
was
modulus
under
stress-strain
the
sharply
failed
at 4%
bonds
optimized
that the possible
thereby
reducing
relative
to
general conclusions: poly(L-lactic
showed
the highest
was
found
mechanical
properties
in the
were
unchanged,
poly( BPA-imino-
somewhat
processing
and
Consequently,
reminiscent
required
the
onset
compression-moulded extrusion
Under
Although represents
a very of
modification polymer excellent
the
film
strength
increased
Biomaterials
BPA
change
the
effect
by
de-
properties
Dat-Tyr-Hex chemical
this
structural
of the resulting
polymers
are
the polymers
very
range and
available
polymers
cover a wide
95°C).
making
degradable
soft
to obtain virtually
futile search
Thus,
biomedical
screws,
applications
or scaffolds
of strength
(from -62”
to
any desirable
Tg
it has been implied
as
for
of severed
bone nerve
of material and biological
in many cases the biological
Although
instance
drug-releasing
for the reconnection
fibres require unique combinations
nontoxic
goal. Unfortunately,
such
grafts,
high-
or blends of the
for just one more
advanced
vascular
as
materials
the Tg of the tested
Likewise,
is no longer a valid research stents,
shades
range of over 160°C
material’5s intraarterial
such
from copolymers
or blending.
polymers properties.
and weak
all intermediate
materials.
by copolymerization
the poly(ortho
and its copolymers,
materials
to very
it possible
that ‘the generally
available,
strong
are accessible
currently
in our study,
aliphatic polyesters.
of physico-mechanical
L-PLA virtually
usually
all fall into this category.
the currently brittle
challenge.
related to aliphatic
PLA, PGA, poly( hydroxybutyrate)
From
the
mech-
is that the most widely
polymers
included
medical
devices)
with improved
research
of all
in terms of
inert polymers
bone fixation
observation
would
and tensile
strongest
were clearly inferior
esters),
molecular-weight
high-molecular-
strength
Among
Finally,
the
poly(BPA-iminocarbonate)
Thus, for many possible
degradable
PCL and PTMC
with
that had been fabricated
of tensile
or are structurally
properties.
in the of
properties
properties
(toxicity, degradation profile in viva, cell adhesion, blood compatibility, etc.) have not been determined, it appears to be unlikely
that the currently
be able to cover all desirable
available
materials
will indeed
combinations.
Poly(Dat-Tyr-Hex-imino-
amorphous
and
exhibited
retained
the
by poly(BPA-
ACKNOWLEDGEMENT
to poly(BPA-iminocarbon-
and tensile were
noticeable
be
as well.
of
In comparison
could
conditions,
limited.
completely
forming
ate), the tensile was
decomposition.
controlled
fundamental polymer,
thermal
may be possible
replacement
Tyr-Hex-iminocarbonate)
to
margin of
for
without
carefully
surprisingly
was
iminocarbonate).
ductility
in a wider
thermal
on the mechanical
was
carbonate)
of
moulding
the
of
of 14°C
relative
temperature
at 65-70°C
or injection
structure
in Ts resulted
the
investigated
and
strength
in agreement
to some of the advanced
general
such as PTMC
a derivative
7 7). led to a reduction
values
is an important
Another
and ductility
poly(Dat-Tyr-Hex-iminocarbonate)
composition.
polymer
and
of degradable
anical strength
preparation
conditions.
polymers
cover a very wide
(Table 2). Since the Td remained
the reduction
between
is we
sample
(such as degradable
development
controlled
even the mechanically
strength
applications
mobility,
but brittle
of BPA by Dat-Tyr-Hex,
(Structure
poly( BPA-iminocarbonate)
302
Overall,
mechanical
polyesters
hydrogen
chain
Tg of poly( Dat-Tyr-Hex-iminocarbonate)
safety
5,
L-PLA.
The replacement dipeptide
of
of poly(BPA-iminocarbonate)
to be a strong
of high-molecular-weight tyrosine
limit
degradable
individual
strictly
in mind it is possible to draw a few
acid)
However,
for tensile
between
Of all tested polymers,
weight
by the
point,
of interchain
could
poly( BPA-carbonate).
carbonate)
of the
poly(BPA-carbonate)
presence
the ductility
The
(Figure
At this
processing
test
in this study to
our data on tensile
not necessarily
such as PEEK or Kevlar”.
poly(BPA-imino-
100%47-4g.
in polyiminocarbonates
whose
shown
data. Whilst
elongation,
to fail only at about
of
brittleness
is clearly
curve of poly( BPA-iminocarbonate)
carbonate) speculate
in terms
properties
increased
This
A) and the elongation
known
weight
are
With this limitation
tested
poly(BPA-iminocar-
in the mechanical
poly(BPA-iminocarbonate). Curve
for
surpassed
M Pa for poly(BPA-carbonate)).
difference
polymers
molecular
(21 50 MPa
2000
and
during
Consequently,
as the
obtain
values
and
values obtained from test specimens
modulus. about
conditions
measurement.
For
and measureto
the best possible
uniform
certain
to optimize
polymer
Since we attempted
had to use
are
conditions
the desire
each
to the
addressed.
preparation
collect data that would allow comparisons experimental
(%I
with of
and modulus.
polymers
Elongation
conflict
processing strength
OIr
there
limitations
uniform as possible during sample
group of
Owing
any such study will by
practical
ment
a first attempt
of a diverse
modulus
slightly
(Figure
199 1. Vol 12 April
5, Curve
reduced, B).
of poly(Dat-
The authors
whilst
Incorporated,
its
Research
would
like to express
Nova
their gratitude
Pharmaceutical
and Development
Corporation,
to Amgen
Corporation,
Stolle
SRI International,
Phvsico-mechanical
and to Dr S. Lrn for contributing study.
In addition,
acknowledge
the
the helpful
manuscript
free polymer
authors
comments
samples
like
to
for this
and corrections
and J. Heller
Huang
(University
(MIT),
Dr
D.
(Stolle
Incorporated).
the viscosity
their help with the preparation 8902468,
by National
by Zimmer Focused
Development and Dr C.G.
measurements
and
and Mr S. Pulapura
for
of the manuscript.
Science
Foundation
Incorporated
23
24
This work and
A.M.
20,
and
25
D.K.,
26
of
Patent
3.839.297.
Miller,
R.A.,
Brady,
Biodegradable
changes
Mater.
Res.
1977.
11,
Lews,
D.H..
polymers,
Systems,
(Eds
NY,
1990,
release
and
pp. l-4
synthesis 1987,
2:
an In
catalystin
octoate
copolymer
sutures,
Degradation
rates
US
and
of oral
polyglycolates):
copolymer
ratios.
Rate
J. Biomed.
of
bloactlve
agents
Polymers
M. Chasln).
from
as
Marcel
lactlde/
Drug
De//very
Dekker,
New
York,
by mwoorgamsms:
technology
and
5. 246-250
Williams, (PHB)
D.F..
On
the
homopolymer
copolymers,
I”
use
1
Tibtech and
D.E..
in Biodegradable
E.A. and Senior,
polymers
for
copolymers.
71 1-719
R. Langer
hydroxyvalerate Dawes,
Part 1,
494-498
I” PLA/PGA
Polymer
N.D.
1, 22,
C.C.. Use ofstannous
(polylactates
Controlled
glycollde
Miller.
198
polymers
home-and
and CutrIght.
with
hydroxybutyrate
27
J.M.
modlftcatlon
D..
J. Kohn
1975
Implants
Byrom,
and
and copolymers:
L(-)-/act/de-glycolide
resorbable
USA,
homo-
acid)
Polymer
manufacture
I Engelberg
464
D. and Versfelt,
economics.
Giving Grant.
REFERENCES
acid)
1459-l
Gilding.
degradation,
Wasserman. the
grant DMR-
and by a Johnson
1979,
Reed,
vitro 22
R. Langer
Incorporated),
and to Mr D. Eshelman
was supported
S.J.
polymers-
surgery-poly(glycolic)/poly(lactlc
Special thanks are due to Mr Kofi
for performing
GPC analyses
and
21
Drs
Professor
Professor
Research
Dr S. Lin (Zimmer
Pitt (Amgen A. Antwi
International),
of Connecticut),
Lewis
Corporation),
(SRI
Polymer
of the
of degradable
surgery-polyglycollc/poly(lactlc
gratefully
made by Dr A. Domb (Nova Pharmaceutical),
D. Friend
Johnson
would
propemes
blodegradatlon
and
B/omatena/s
P.J
1987,
8.
The role and regulation
mwoorganwns.
Adv
of
poly-fi-
poly-8-hydroxybutyrate-
Muob.
129-l
37
of energy
Phys/o/.
reserve
1973,
10,
135-266 1
2
Kronenthal,
R.L.,
Polym.
Sci.
Techno/.
Heller.
J.. Controlled
bloerodlble 3
Langer,
Biodegradable 1974, release
polymers, R.S.
and
blomaterlals
polymers
8,
1 19-l
in controlled
drug
active
1980,
N.A..
and
surgery.
28
compounds
from
1, 5 l-57
Present delwery
and
29
future
systems,
applications
Biomaterials
of
1 98 1,
5
Zaikov.
G E., Quantltatwe Rev.
Shah.
C.,
T.
erodlble
1984,5, 6
Chem.
Htguchl,
catalysed
aspects
Macromol.
of polymer
Phys.
and
degradation
1985,
C25(4),
Hlmmelsteln,
polymeric
matncles
K.J.,
I” the living
Drug
of poly(ortho
R.A.,
delivery
from
ester)s, Biomaterials
8
Christel.
A wew
J.W..
of vascular
Winter,
for
D.F.
stems.
Internal
Circulation
1989,
79(2),
Wtley
1988
fixation,
7980,
Et Sons,
polyglycollde
(Eds
New
G.D.
York.
NY,
F.R.
and
Plates
and
33
Mater.
Heller.
Sparer,
J.,
Bwdegradable
Rozema. VI.
Biomaferials
1987,
in vitro,
rods
for
35
of absorbable,
fracture
Res.
J.A.,
Cheginl,
179-l
Zentner.
as drug
Marcel
N.
and
Masterson,
absorbable
22,
and
Polymers
Chasln).
of
1988,
R.V.
Heller. delivery
B.J.. devices.
and
of therapeutic
delivery
Poly(ortho
systems,
New
York,
esters),
(Eds
NY,
in
R. Langer
USA,
37
1990,
use of poly(ortho
agents,
J. Bioact.
esters)
J , Penhale.
Heller,
release
Controlled
Release
Mansdort),
Marcel
Compat.
Polym.
D.W.H., of
Helwing,
R.F.
norethlndrone
Delivery
Systems,
Dekker
Inc.,
and
from (Eds
New
Fritzlnger.
38
York,
T.J. NY,
Roseman USA,
B.K..
esters),
in
and
1983,
J., Frltzlnger, release
polymers.
J. Control.
J.. Ng.
Bruns,
R.A., Gaynon,
the
Cho,
B.K., Ng,S.Y.
M.G.
and
Heller,
polyorthoesters
and
39
-
Heller,
J
Penhale,
poly(ortho
20
Pitt.
C.G..
B.K.,
in
40
I. Linear
Sanders,
S.S., Use of poly(ortho
5-fluorouracll
and
a
LH-RH
devices
L.M.,
polyonhocarbonates,
esters)
D.W.H. by the
Sci. (PO/m.
Lett.
Frazza,
E.J. and Schmitt,
Mater.
Res.
GIldIng.
D.K.
Symp. and
1971, Reed,
Sci.
US
Patent
and
reaction Ed.)
Helwng,
R.F.,
of diketene
1980,
E.E., A new
18,
acetals
4.079.038.
and
43
of
Res.
polyols,
44
J. Biomed.
45
1, 43-58 A.M.,
Biodegradable
for
use
Biochem.
J
Sot.
in Biodegradable
R. Langer
and M. Chasln).
pp. 71-l
Cohen,
S.,
Plast.
20
Biodegradable
plastic
Eng.
1975.
Kllmas.
26,
Tech. Pap.
D.M.
and Schlndler.
of poly(e-caprolactone)
I”
3779-3787 A. and
of allphatlc
Woodward.
polyesters,
S.C.,
J.
The
Control
Rel
dlffuston
F.. Schlndler.
A. and
of poly(c-caprolactone).
Bao, Y.T. and Samuel,
I” polymers,
Applications,
Pitt,
1 Biomed.
ACS
Good),
N.K.P.,
Estimation
rn Controlled-Release Symposwm
American
Series,
Chemical
of
Technology,
Society.
Vol. 348,
WashIngton.
pp 49-70
J.V.,
Blends
contalnlng
Academic
Press,
New
Gratzl,
M.M..
C.G.,
Moatamed,
19,437-444
in Polymer
Pitt.
P.S.,
A.L..
polymers,
Feng.
Blends,
polyesters
poly(c-caprolactone)
(Eds
York,
NY,
D.R.
1978,
G.L..
related Vol 2,
pp 369-389
Surles.
degradation
copolymers
and
Paul and S. Newman)
USA,
Kimmel.
II. The
and thetr
X.D.,
Song,
in
46
C.X.
block
J. and
Schlndler.
of poly(o,L-lactide).
I” vwo,
SC.,
(Polym.
HIII.
and
Carothers,
J.W.
A.,
poly(c-
Biomaterials
1981,
2.
Chen,
W.Y.,
Lett.
Synthesis
and
of