Cardiovascular Vincent

Dousset,

MD

a

Felix

W. Wehrli,

Popliteal Artery MR Imaging-US

terms:

Arteries,

Arteries,

popliteal,

namics,

924.12984,

MR

studies,

ies,

924.12984

parative

extremities,

924.12984 924.1299

924.1299 a

a

a

Blood

Magnetic

924.91 Blood, flow Blood vessels,

a

vessels,

resonance

US (MR),

a

dystudcom-

studies

Radiology

1991;

a

Adehine

Louie,

MD

John

a

Listerud,

179:437-441

T

HE

feasibility

of measuring

hemo-

without

dynamic parameters by means magnetic resonance (MR) imaging has been demonstrated for a variety

of time-of-flight 14) methods.

(1-9) However,

none

of which

planar nique,

Fourier which,

stantial

was

is based

accuracy (5,7,15),

on the

flow-encoding in our view,

advantages

niques. The

objective

to assess

oven of the

the

of

and phase (10only a few

studies have assessed the any one of these methods

against

bi-

other

an accepted

standard,

em realized

in the

form

sional embodiments spin-echo detection

repro-

duplex was and

lat-

of one-dirnen-

by means by Feinberg

of and

Mark (17) and a gradient-echo implementation by Hennig et al (13). A second objective of the study was to the

technique’s

potential

detailed information velocity distribution,

not obtainable with Doppler the purpose of this validation, normal popliteal artery was

US. For the chosen.

The rationale for this choice was as follows: (a) The popliteal artery is readily accessible to duplex scanning. (b) It provides a large mange of both forward and reverse velocities covering the entire spectrum of physiobogic flow velocities. (c) It provides a straight vessel segment where flow is

expected tumbed. required parallel

to be relatively (d) No angular because the to the z axis.

encoding

undiscorrection is vessel is nearby

first

Market

St.

Ste

420,

Philadelphia,

PA

19104.

RSNA,

1991

METHODS

Subjects Popliteal artery flow dynamics were studied in 11 male and female volunteers (age range, 21-53 years; mean, 37 years)

(13),

where

the

in a plane

parallel

of the

velocity-encoding

moment

vent

velocity

olution

aliasing.

of 2 cm/sec

locity-encoding

spins

With per

steps

a velocity

pixel for

and

each

res-

64 ye-

velocity

di-

rection (128 total), a range of velocities from 0 to 128 cm/sec is covered. Temporal resolution was achieved by electrocardiographically triggering the pulse sequence. The intersequence delay (temporal resolution) chosen was 43 msec. Further, the pulse sequence, which is illustrated

pensated

in

on the

Figure

2, is velocity

com-

section-selection/readout

axis.

The protocol used consisted lowing steps: A series of axial bocalizer images were obtained

by means

Re-

ceived October 30, 1990; revision requested December 28; revision received January 17, 1991; accepted January 22. Address reprint requests to F.W.W. C

AND

vascular

gradient is governed by its amplitude and the spacing between the two lobes of opposite polarity. Spacing and step size thus determine the velocity resolution, which needs to be appropriately selected to pre-

of a transmit/receive

to determine

velocity-encoding

MATERIALS

peripheral

to the flow direction and subsequently velocity-sensitized by means of stepping a bipolar gradient in the direction of flow (Fig 1). Velocity is encoded by stepping a bipolam balanced gradient (18), which takes the place of the spatial phase-encoding gradient and imparts a first moment on the spins, as originally suggested by Moran (16), followed by the collection of a gradient echo. In contrast to Moran’s approach, however, velocity is encoded in one spatial direction only to achieve clinically practical imaging times. The

coil Hospi3440

known

Imaging

are excited

study

and

MR

velocity

tech-

present

accuracy

any

abnormalities, recruited in the Department of Radiology in the investigators’ institution.

All MR data were obtained on a 1 .5-T clinical imager (Signa; GE Medical Systems, Milwaukee) with use of biplanar

techhas sub-

ultrasound (US). The method first suggested by Moran (16)

for providing on intmaluminal

the Department of Radiology, University of Pennsylvania,

PhD

of

ducibility of the Fourier flow-encoding technique by calibrating it

highlight

I From tal of the

MD,

Hemodynamics: Correlation’

Temporally resolved velocity measurements in the pophiteal arteries of 11 healthy subjects were obtained by means of magnetic resonance (MR) imaging with use of the Fourier flow-encoding technique. Excellent agreement with corresponding Doppler ultrasonography (US) data (r .97, slope 0.99, intercept -1.5 cm/sec) was demonstrated over the entire velocity range from 50 to -20 cm/sec. The method was rapid and its implementation straightforward. Further, MR imaging was shown to provide the intraluminal velocity distribution relevant for the determination of true flow rates, not obtainable with Doppler US. Index

PhD

Radiology

the

level

image

of the folspin-echo bilaterally

extremity at which

would

be

the ob-

tamed. The site of measurement selected was an arbitrary but easily reproducible landmark-the intercondylar space-to ensure that the site of analysis was the same in all subjects. The parameters of the velocity-encoding pulse sequence were as follows: repetition time, 43 msec (mini-

437

mum

intersequence

delay);

echo

time,

30

msec; flip angle, 30#{176}; field of view, 16 cm; section thickness, 5 mm; number of signals averaged, one; velocity resolution, 20 mm/sec per pixel; and number of velocity encodings, 128. Under these conditions, 15-25

temporal

commodated, heart rate. velocity

increments

could

placement. Peak forward and revelocities of the three wave compoderived from the two modalities

were

then

identity

correlated, was

encoding

was

the

Doppler

of

2 mm-

duplex

subjected to The Doppon an ATL 9

instrument

(Advanced

Technology Laboratories, Bothell, Wash) with use of a 5-MHz transducer. The cursor was placed at the location of interest (ie, at the center of the vessel) with the US beam at an angle of 45#{176}-60#{176} (19-21).

The velocity

measured

regres-

Figure 3 shows a typical set of MR data, consisting of a transaxial bocalizem image (Fig 3a) with which the velocity analysis was performed, as

(Fig 3b-3d). phase-dependent velocity

From

waveform

meimages

an analysis flow images, was

with multiple flow (A, B, and

(Fig

4c). This

to result of the

from pressure

behavthe

opgradi-

ent built up along the wave propagation front following ventricular contraction, and vascular compliance (20,21). The correlation of the maximum central velocity data from the two modalities (peak forward and reverse velocities for the peaks labeled A, B, and C) is illustrated in Figure 5. A correlation coefficient of .97, a slope of 0.99, and an intercept of -1.5 cm! sec (suggesting that MR overesti-

RESULTS

of temporally flow-encoding

(19,22)

ion is known posing effects

sion.

well as a series solved Fourier

US

Mark

by linear

oscillation and reverse

C peaks)

ac-

equivalent

The same volunteers were Doppler (duplex) US analysis. ler waveforms were obtained

damped forward

and the line of

determined

depending on the subject’s The total imaging time for the

128 heartbeats or approximately utes per extremity.

Ultra

be

cursor verse nents

of the the

determined

(Fig 4). Note the close correspondence of the MR-derived waveform with the Doppler velocity waveform, which exhibits the characteristic RF

was the maxi-

-411)

A V

mum (central) velocity, representing the fastest-moving red blood cells, which is also the quantity measured with MR imaging.

TE

-5

____

a-

Vessels

Data

Analysis

The Fourier tamed

maximum central velocity in the flow-encoding images was obby measuring the maximum excur-

sion

of the

in

each

resolved was

velocity data, ated no more tion

C F E

bolus

temporally correction

of the

direction). Doppler

SIIca

Imaging

made

of the

images. to the

No angular tllpate

MR-derived

as the vessel typically devithan 10% from the omienta-

velocity-encoding

gradient

(z

US velocities were read off the velocity waveform by means of

Figure 1. Principle ing technique. Spins

of flow-encoding imagare excited in a plane

perpendicular

velocity

signal

is read

to

the

out in the same

vector,

and

3.

(a) Transaxial

spin-echo

localizer

tion

and

readout

compensated.

the

direction.

echo

time,

gradients

RF

GFE

are

first-moment

radio frequency, flow-encoded gradient. =

TE

=

C.

b.

a. Figure

Figure 2. Flow-encoding pulse sequence. Instead of spatial phase encoding, velocity is encoded by means of a bipolar gradient (G) coincident with the section-selection orientation. This gradient is stepped to encode velocity. Note further that the section-selec-

15-25

image,

showing the flow bolus (white arrow) from popliteal set of temporally resolved velocity-encoding images tion of the velocity during cardiac cycle. Annotations R wave (in percent).

together

with

flow-encoding

artery (black arrow). (b-d) in normal subject showing pertain to the phase delay

image, Typical evolufrom the

d.

438

a

Radiology

May

1991

0

10

20

30

40

50

60

70

80

9(

% PHASE b.

C.

Figure

4. (a) Velocity waveform, derived from the temporally resolved MR images of Figure 3b-3d, plotted as a percentage of the delay from the R wave. Note the tripolar waveform characteristic of the damped oscillation resulting from the vascular resistance to the pressure wave and the vessel’s compliance. Peaks labeled A, B, and C are the first forward, first reverse, and second reverse maximum velocities, mespectively. LPA = left popliteal artery. (b) Duplex localizer image with cursor indicating site of velocity measurement. (C) Doppler US waveform.

ple the entire RR interval. Finally, the temporal resolution achievable with MR imaging may be inferior that achievable with US, although

MR is capable

of correctly

delineat-

ing the complex bidirectional form. On the positive side,

MR imaging

as well as the Pouncebot index (also denoted “resistive index”), defined as (VA VB)/ VA, where VA and VB represent the velocities pertaining to peaks A and B, respectively. The latter is a measure of vascular impedance (20,23). The remarkable degree of mutual agreement observed is indicative of the high accuracy of MR imaging in hemodynamic analysis. :60.40.20

0 Maximum

20 Velocity

40

60

80

1001201s0

us

Figure 5. Correlation MR imaging velocity

DISCUSSION

of Doppler US and data in 11 subjects. Ve-

locities compared are the maximum central velocities. Note the three clusters of data points pertaining to the A, B, and C peaks, respectively (see Fig 4a). The correlation coefficient was .97 with a slope of 0.99 and an intercept of -1.5 cm/sec. indicating the excellent accuracy of the MR measurements. The high-velocity A peak (arrow) found in one of the test subjects with both modalities is abnormal.

Hemodynamic niphenal vascular flow encoding

mates

flow

found.

velocity

by

less

than

5%)

The data is summarized in the Table, which lists means and standard deviations for the three peaks as denived by means of the two modalities,

Volume

179

a

Number

2

distinguishes

ages

itself

in

several respects from Doppler US. It has, of course, the disadvantage of not being real time and therefore of being sensitive to variations in heart rate. A further limitation (which can be circumvented by using retrospective gating) (24) is the failure to sam-

information

of Figure

6, which

were

ac-

quired to illustrate the significant changes in velocity profile occurring during different phases of the cardiac cycle. Whereas the profile is Poiseuibhan during most of the cardiac cycle, this is not the case during the occurnence of strong acceleration or decel-

flow of the pewith MR means of

assessing the hemodynamic panametens of the peripheral arterial system. The method is found to be accurate oven a wide range of velocities and its implementation straightforward. It thus lends itself as an add-on to a morphologic imaging examination, with only insignificantly prolonged total examination time.

MR imaging

were

analysis system is a viable

wavehowever,

not obtainable with Doppler US in that it allows the complete distribution of velocities across the vessel lumen. This becomes evident in the higherresolution Fourier flow-encoding im-

eration, when vail, causing

(cm/sec)

provides

to

behavior

the flow

inertial forces to be pluglike.

is present,

preThis

for exam-

ple, during the phases preceding and following directional inversion of flow (ie, where the slope is maximal) (Fig 6e, 6j, 6n). Subsequently, the flow becomes predominantly laminan, as indicated by a more parabolic profile (Fig 6f-6h, 6k, 6o). During meversal of the flow direction, flow near the vessel wall exceeds the flow centrally (Fig 6i, 6m) (25,26), that is, opposite to the behavior found duning laminar flow. The multiple reversal of the flow direction is a consequence of the yessel’s elastic properties (vascular compliance) causing damped oscillation (22), a property characteristic of a penipherab arterial circuit (19). It is well known impedance

that,

during is lowered

stress, vascular (through an

Radiology

a

439

a.

b.

C.

d.

e.

f.

8

h.

i.

j-

k.

1.

q.

r.

m.

n.

Figure bution field

of view

selected

Fourier flow-encoding lumen during complete

was

10 cm,

corresponding

except during phases of strong acceleration flow near the vessel wall exceeds the central particularly head in

well

seen

in

p.

0.

6. High-resolution across the vascular

f

(straight

arrow).

images cardiac

of popliteal artery cycle. The images

subject showing temporal changes in velocity distriwere obtained with the protocol outlined earlier, except that the pixel size of 0.39 mm. During most of the cardiac cycle flow is laminar (f, h, k, o), where it is more pluglike (e, j, n). Further, during reversal of flow direction,

to a spatial

and deceleration, velocity (i, m). Note The

popliteal

vein

in normal

also the superposition

is displayed

of a small

as a signal

of constant

adapting process to satisfy the increased demand for oxygen and thus increased blood flow to the muscles), which causes virtual disappearance of the reversal in flow direction

flow

(F) on the

rate

of the popliteal

highly

laminar

artery, flow

(arrow-

assumption CONCLUSIONS

of radial

symmetry:

F

27

=

i:

rv1(r)dr,

(26,27).

The practical implications knowledge of the intmaluminal

of the veboc-

where

ity

permit-

beats

distributions

the

are

calculation

in milliliters

per

in their

of true second,

flow

rates

which

is

difficult to achieve with Doppler US because of the limitations imposed on sample gate size (28). Flow rates are more significant than flow velocities since the former obviously function of the vessel’s radius vessel’s diameter, which varies

among individuals) and and pathologic changes. Since we locity profile

crement

branch and

C).

the

ting

collateral amplitude

and,

further,

since

the

yein-

yes-

can fairly readily be from the axial spin-echo

Of course, the rates derived in ically depend on precision of the ments.

For

surements further

that

the

following

nc.

440

a

Radiology

example,

the

gives

for each requirement

is difficult

vessels.

which

number

of

accuracy of flow this manner will cnitthe accuracy and geometric measure-

nor

bution

the

cardiac is that

neither

intraluminal

be fully

diam-

phase. A the pixel

the luminal a condition

allow which

quantitation cannot

fied phy.

of true flow rates, accurately be quanti-

by means The short

of Doppler examination

for

small

well

vessel

study of the hemodynamics tracranial vasculature.

velocity

radially

distri-

sonogratime

(2-3

minutes) and the excellent neproducibibity (no dependence on transducer placement) are further benefits. Also, unlike in US, a single data set can provide velocity data for more than one vessel. Another potential application in which US fails to perform

the

to satisfy

Finally, may

vessel

the cardiac cycle compliance, separate mea-

size be much less than diameter of the vessel,

bocahizem image, we can, in principle, determine the flow mate (in milliliters per minute) by numerically solving equation,

the

n the number of temporal phases, r0 the vessel radius, and v1(r) the velocity at radial position r for the ith temporal increment.

eter changes during as a result of vascular which would require

physiologic

directly observe the for every temporal

sel’s diameter determined

are a (ie, the

N represents per minute,

The present study indicates that measuring blood flow with MR imaging is likely to create a role for this modality in vascular examinations, as it is capable of providing both functional and morphologic information. Further, MR imaging is shown to have the potential for providing detailed hemodynamic information, which includes the intrabuminal yebocity distribution, and thus should

due

On

symmetnot

the

real

to lack

other time

of a bone

hand, and

thus

window of the

MR

imaging

the

quality

May

is in-

is of

1991

the data is sensitive to variations in RR interval. Other potential problems relate to the possible noncoincidence of the flow-encoding gradient and vessel orientation, which requimes application of an angle correction or, alternatively, oblique orientation of the flow-encoding gradient, both of which can be achieved by prior acquisition of an angiographic localizer image. Finally, the data analysis currently is labor-intensive, requiring measurement of the bolus excursion

by

means

of a cross-hair

6.

7.

8.

9.

10.

cursor on each of the images. Howevem, this may be a temporary deficiency, since opportunities exist for automating the data analysis procedune. U

11.

12.

References 1.

2.

3.

Axel L. Blood flow effects in magnetic resonance imaging. AJR 1984; 143:11571166. Axel L, Shimakawa A, MacFall JR. A time-of-flight method of measuring flow velocity by magnetic resonance imaging. Magn Reson Imaging 1986; 4:199-205. Wehrli

FW,

Shimakawa

A, MacFall

JR.

sist

5.

Tomogr

1985;

K, Matsudav

Visualization tive analysis MR imaging. 199.

T, Sakurai

15.

16.

T, et al.

of moving fluid: quantitaof blood flow velocity using Radiology 1986; 159:195-

17.

18.

Volume

179

a

Number

2

Wehrli

FW.

Time-of-flight

effects

in

MR

imaging of flow. Magn Reson Med 1990; 14:187-193. Edelman RR, Mattle HP, Kleefield HP, Silver MS. Quantification of blood flow with dynamic MR imaging and presaturation bolus tracking. Radiology 1990; 171:551-556. Van Dijk P. Direct cardiac NMR imaging of heart wall and blood flow velocity. Comput Assist Tomogr 1984; 8:429-436. Nayler G, Firmin DN, Longmore DBJ. Blood flow imaging by cine magnetic resonance. J Comput Assist Tomogr 1986; 10:715-722. Walker MF, Souza SP, Dumoulin CL. Quantitative flow measurement in phase contrast MR angiography. J Comput Assist Tomogr 1988; 12:304-313. Hennig J, Mueri M, Brunner P. Friedburg H. Fast and exact flow measurements with the fast Fourier flow technique. Magn Reson Imaging 1988; 6:369-372. Merboldt KD, H#{227}nicke W, Frahm J. Rapid Fourier

9:537-545.

Wehrli FW, Shimakawa A, Gullberg GT, MacFall JR. Time-of-flight MR flow imaging: selective saturation recovery with gradient refocusing. Radiology 1986; 160:781-785. Shimizu

14.

et

al. MR imaging of venous and arterial flow by a selective saturation-recovery spin-echo (SSRSE) method. J Comput As4.

13.

Hennig J, Mueri M, Brunner P. Friedburg H. MR imaging of flow using the steadystate selective saturation method. J Comput Assist Tomogr 1987; 11:872-877. Matsuda T, Shimizu K, Sakurai T, et al. Measurement of aortic blood flow with MR imaging: comparative study with Doppler US. Radiology 1987; 162:857-861.

flow

imaging

using

FLASH

se-

quences. Magn Reson Med Biol 1988; 1:137-145. Maier SE, Meier D, Boesiger P. Moser UT, Vieli A. Human abdominal aorta: cornparative measurements of blood flow with MR imaging and Doppler US. Radiology 1990; 171:487-492. Moran PR. A flow velocity zeugmatographic interlace for NMR imaging in hu. mans. Magn Reson Imaging 1982; 1:197203. Feinberg

A, Mark

AS.

Human

brain

mo-

tion and cerebrospinal fluid circulation demonstrated with MR velocity imaging. Radiology 1987; 163:793-799. O’Donnell M. NMR blood flow imaging using multi-echo phase contrast sequences. Med Phys 1985; 12:59-64.

19.

Standness DEJ. niques in vascular EF, ed. Noninvasive in vascular

20.

21.

22.

24.

25.

26.

27.

St Louis:

Mosby,

1985;

waveforms:

problems

and

solu-

tions. In: Bernstein EF, ed. Noninvasive diagnostic techniques in vascular disease. St Louis: Mosby, 1985; 40-57. Pourcelot L. Applications cliniques de l’examen Doppler transcutan#{233}. In: Peroneau L, ed. Velocimetrie ultrasonore Doppler. Paris: INSERM, 1974. Glover GH, Pelc NJ. A rapid-gated cine MRI technique. Magn Reson Annu 1988; 299-234. Milnor RM. Pulsatile pressure and flow. In: Milnor WR, ed. Hemodynamics. Baltimore: Williams & Wilkins, 1982; 97-134. Edelman RR, Rubin JB, Buxton RB. Physics and instrumentation, In: Edelman RR, Hesselink JR. eds. Clinical magnetic resonance imaging. Philadelphia: Saunders, 1990; 109-182. Caro CG. The mechanics of the circulation.

28.

disease.

11-18. Nelson TR, Pretorius DH. The Doppler signal: where does it come from and what does it mean? AJR 1988; 151:439-447. Taylor KJW, Holland S. Doppler US. I. Basic principles, instrumentation, and pitfalls. Radiology 1990; 174:297-307. Johnston KW, Kassam MS. Processing Doppler signals and analysis of peripheral arterial

23.

Doppler ultrasonic techdisease. In: Bernstein diagnostic techniques

New

1978. Fronek ments ity.

In:

agnostic Louis:

York:

Oxford

University

A. Quantitative in arterial disease Bernstein

EF,

techniques Mosby,

1985;

ed.

velocity of lower

Press,

measureextrem-

Noninvasive

in vascular 554-562.

di-

disease.

St

Popliteal artery hemodynamics: MR imaging-US correlation.

Temporally resolved velocity measurements in the popliteal arteries of 11 healthy subjects were obtained by means of magnetic resonance (MR) imaging w...
921KB Sizes 0 Downloads 0 Views