Radiographic Mammographic Donald
Exposure Dose
R. Jacobson,
PhD
Calculator Calculator’
A set of dose signed
calculators
to facilitate
culation dose)
has been
quick
of surface
and
phy
de-
easy
exposure
for routine
Users Guide (3), recdosage in mammogra-
Mammography-A
ommend
and
specfied
dose.
by
On
aver-
age glandutar dose in mammographic studies. Acceptance by technologists has been good, and in two inspections by the Joint Commission for Accreditalion of Health Care Organizations, the calculators were deemed adequate to satisfy the diagnostic radiology standards that doses be monitored.
employing a grid should deliver no than 8 mGy (0.8 rad) for a twoview study. Recently published federal guidelines governing Medicare reim-
bursement
for screening
zations
creditation (JCAHO)
breast
screen-film
mammogram
grid (4). We present
tic
radiology
ance
Commission of Health
standard
individual
specifies
Our
that
“monitors
a
doses
from diagnostic radiology procedures” (1). An analysis (2) of the new standards states that compliance with this standard requires that exposure values be determined for all routine procedures for
each
individual
calculation on
x-ray
system.
dose
must
of x-ray
calculators of x-ray
The
made from the at a variety of Tabulated standose values are not acceptable. standard, in its strictest interpreta-
This tion, means that the value delivered in a radiographic should be available for the quested. JCAHO
standard
specifies “with respect ards from equipment, ence
to
the
currently thority
radiation
dose
procedure patient if re-
DR.2.2.5 (2) further to radiation haz(there
be)
adher-
of any
recommendations
recognized on
of the
and hazards,
reliable such
auas
the
National
Council on Radiation Proteclion and Measurements (NCRP), and any requirements of appropriate licensing agencies or other governmental bodies.” The NCRP, in report no. 85,
From the Department of Radiology, Medical Physics Section, Medical College of Wisconsin, 8700 W Wisconsin Aye, Milwaukee, WI 53226. From the 1989 RSNA scientific assembly. Received July 9, 1990; revision requested August 8; revision received September 19, 1991; accepted September 23. Address reprint requests to the author. ( RSNA, 1992 I
578
#{149} Radiology
a
performing in compli-
NCRP
standards.
and Methods is based
device
of a circular
slide
rule.
on exposure-to-dose assuming
backscatter calibrated
be based
with
for dose,
and
(0.3
on the An
ex-
ample of a device that is used to calculate surface exposure is shown in Figure 1. With an appropriate scale factor, and
equipment in question kilovolt peak settings. dard
JCAHO
calculating
principle
based
measurements
actual
with
Materials
for diagnos-
(DR.2.2.4.2)
qualified
for AcOrgani-
Care
mammograms
specify a maximum dose of 3 mGy rad) per view for a 4.5-cm compressed
1992; 182:578-580
HE 1992 Joint
glan-
more
measurements
T
average
of an analysis publication also
and
Index terms: Breast radiography, radiation dose, 00.11 a Dosimetry #{149} Radiations, exposure to patient and personnel
Radiology
using
the basis ratio, this
dose,
a typical
distance
exposure (or skin peak employed for with the source-to-
(SOD)
used
for
amination. Next, the value pere seconds used is located
outer
scale,
provides
for the
factor, the device could be to indicate skin dose. To ar-
rive at the surface dose), the kilovolt the study is aligned
object
conversion value
the
result.
ex-
for milliamon the
and the adjacent desired
the
inner
scale
Measure-
ments of radiation output (free in air) performed over a range of kilovolt peak settings and made on the basis of system usage, for example, at every 5 or 10 kVp between 60 and 120 kVp, are used to calibrate each calculating device for a specific x-ray tube. In the event of a tube replacement, or if routine quality control measurements indicate a change in tube output, a new dose calculator can be quickly and easily established for that tube. The error in surface exposure for a mis-setting of kilovolt peak is typically 3%-4% per kilovolt peak between 60 and 120 kVp. The error due to ignoring
backscatter and
is dependent
on field
size
beam energy, but a typical value for the increase in surface exposure due to backscatter is 30% ± 10 (5). could
be
included
in
the
bration of the device. In mammography, the recommended dose specification is average glandular
is generally
on to Some exand theoretical work (6,7) average glandular dose to
cali-
to patient
has related
exposure
agreed
risk
in air. These
(3).
data are pre-
sented in the literature in graphic (3,6) and tabular (8) form, and these sources generally agree with each other to within 5%. The examination parameters required to calculate average glandular dose are the output (mR [2.58 x 10 mCi/kg]/mAs) and half-value layer (HVL) at the kilovolt peak used, milli-
ampere seconds, and compressed breast thickness. The calculation procedure is tedious and time consuming and is simplified by the use of a mammographic dose
calculator. The operation of this (Fig 2) also is based on the principle of a circular slide rule. Examination input parameters are kilovolt peak, milliampere seconds, and compressed
device
breast
thickness.
In use,
the wheel
is rotated
until
the
relevant kilovolt peak and milliampere second values are aligned. The average glandular dose is then read adjacent to the compressed thickness. A correction for fatty or dense breasts, which varies between 5% and 10%, depending on compressed thickness (8), is also incorporated into the device. The calculator must be individually calibrated for a particular tube on the basis of measurements of radiation output and HVL at a variety of kVp settings. Average glandutar dose (Dg) is defined by the simple relationship D = Dg X Xa, where Dg is the average glandular dose in centigrays (rads) per roentgen exposure (without backscatter) and X, is the mcident exposure that would be measured free-in-air for the kilovolt peak and mitliampere second values used in the examination (3). The graphic data in reference 3 relating Dg to HVL and compressed breast thickness for a motybdenum target tube was found empirically to bedescribed by the following equation: Dg = HVL x lIT x 2.2, where T is the compressed breast thickness. This equation describes the graphic
tween pressed 5 cm.
x-ray
This factor
which
correspond penmentat
of risk-benefit specifies maximum recommended dose values; for example, mammography performed with a screen-film system
(or skin
radiography
that
be
dular
cat-
and
data 0.3 and
breast Therefore,
within
2%
for
HVL
be-
0.4 mm
Al and comthickness between 3 and with
the
equation
X,,
=
mAs x mR/mAs for the relevant kilovolt peak, the average glandular dose for a molybdenum tubetarget can be computed as follows: Dg = mAs x mR/ mAs x HVL x lIT x 2.2. The numerical constant divided by the compressed breast thickness is in-
February
1992
1.
2.
Figures
1, 2.
Mean
(1) Radiographic
Glandular
kVp
mR/mAs
;i. Zk_ L
Dose
x HVL
i .:.ll:
---
9L
-
Figure 3. Back calculator.
corporated calculator ing
to the
=
X
calculator
exposure
1/x
Calculator
z
100
=
MARK
E .3’
3/
iL
.2Z
22.
ifl
LZ
side
of mammographic
into
.
dose
of the dose factor correspondthickness. The
the design
as a single appropriate
of measured radiation output (mR/mAs) and HVL at a given kilovolt peak are combined into a single parameter that is dependent on kilovolt peak and is unique for each mammographic x-ray system. The back of the mammographic dose calculator (Fig 3) provides a worksheet for recording of these measurements and calibration of the device. The reciprocal of the product of output product
and
HVL,
multiplied is
Volume
They are fixed for each radiographic x-ray tube. The devices are easy to use, and acceptance by technologists has been good. Radiation dose is not presently being determined for each examination, but the calculators are in place if room-to-room comparisons are desired or a patient requests the dose for a particular study. Both hospitals underwent JCAHO inspections in 1989, and in each case, the inspector was satisfied with the calculator as a method of monitoring diagnostic doses. In the mammography imaging area, the reaction of the technologists to the ments
of two
near
the console
dose
182
exposure calculators have in the radiology departNumber
#{149}
calculator
To evaluate ments made
hospitals.
has
calculator,
have
been compared
puted
values
also been
positive.
the validity of measurewith the mammographic
dose
results obtained
obtained
with
with
it
hand-com-
by using
pub-
data (3,8) for several patient studies covering a wide range of breast sizes and tissue compositions. Agreement between the two methods was within a few percentages in all cases. lished
Discussion
In view of the general consensus that average glandular dose is the preferred method to specify patient dose in mammography, and because of the NCRP recommended dose limits, compliance with the JCAHO standards should be accomplished in mammography by measuring average glandular dose and assuring
Results The surface been installed
dose calculator.
by 100 (designated
the value that must be transferred to the wheel of the calculator. With the wheel in the calibration position, the MARK value for each kilovolt peak is located on the numerical scale of milliampere seconds, and an arrow is drawn by hand on the wheel and labeled with the appropriate kilovolt peak.
“MARK”)
and (2) mammographic
2
that
it is within
the
recom-
mended maximum limits. Currently, in diagnostic radiology other than mammography, there is no consensus as to
how tissue dose should be specified to reflect patient risk. Consequently, surface entrance exposure (or skin dose) was and
chosen as the index to be measured monitored. The JCAHO will accept
a methodology that is state-of-the-art (personal communication, 1989). Skin dose or surface exposure will be acceptable until a consensus is reached within the radiology community as to how diagnostic doses should be measured and monitored.
Alternatively, a computer program can be used to obtain surface exposure or skin dose data from radiologic examinations. If a personal computer or terminal is available in the x-ray department, a program could be written that would give the desired result on the basis of input of room number, kilovolt peak, milliampere seconds, and SOD. The drawback of a single terminal would be inconvenience and possible multiuser conflicts. Alternatively, a spreadsheet-type
output
could
be gen-
erated for each x-ray tube. To cover the range of 60-120 kVp and 1-500 mAs for a 50-100-cm SOD, with an incremental accuracy of 10%, between 10 and 20 pages
of numbers
would
be required
per tube. A larger increment between entries would reduce the required number of pages, but manual interpolation between
values
In summary, calculators
cept
would
then
be required. of these dose on the conrule, and the cat-
the design was
patterned
of a circular slide culators are easily calibrated for a particular x-ray system. Surface exposure is determined for general diagnostic procedures, and average glandular dose Radiology
579
#{149}
can be calculated for mammographic examinations. The calculators are inexpensive and easy to use, and have been found by JCAHO inspectors to satisfy the requirement that diagnostic doses be monitored. U
References 1.
Accreditation brook Terrace,
manual
Ill: Joint
for hospitals. Commission
Verification Identification
Oakon Ac-
identify
Index Spine,
terms: anatomy
Radiology
lumbosacral
of L-5 and of S-i).
nine
182:580-581
transitional
vertebrae
quent
are common congenital a reported prevalence (1-6). Since magnetic
structural
anomalies
have
often
been overlooked. of disk disease
Incorrect localization due to lumbosacral transitional vertebrae may account for inconsistency between clinical presentation and MR image analysis. We attempted to ensure verification of disk disease location and presence of a transitional body by obtaining cervicotho-
From the University of Nebraska College Medicine, Omaha (P.Y.H.) and the Department of Radiology, University of Nebraska Medical Center, 600 5 42nd St, Omaha, NE 68198-1045 I
0.15,
F.J.H.). Received July 15, 1991; revision requested August 9; revision received Septemben 4; accepted September 17. Address reprint requests to F.J.H. V RSNA, 1992
580
Radiology
#{149}
along
lumbar and by counting
with MR verte-
of
hundred consecutive patients for evaluation of the tumbosa-
was
applied
of view.
with Other
a large image
(48-cm)
enhancing
techniques such as flow compensation, saturation, and no phase wrap can also be used. Each patient was positioned with the isocenter at the iliac crests. Six sagittal images (scouts) of the cervicothoracic spine were obtained with a section thickness of 5 mm. These images included C-2 through T-12. Acquisition of the six images required approximately 2 of additional imaging time. The image that best demonstrated the odontoid process of C-2 was chosen for vertebrae counting purposes. By means of the cross-hair cursor on the monitor and beginning at C-2, the cervical and minutes
scout
thoracic vertebrae were counted caudad to T-11 or T-12. T-11 or T-12 was then marked with the cursor, and the distance between this point and the isocenter appeared at the bottom of the monitor arrow). an “5” denoting
L, Villafana
T, Day JL, Lightfoot DA. in mammography. Radiology 1984; 150:577-584. Hammerstein CR, Miller DW, White DR. Masterson ME, Woodard HQ, Laughlin JS. Absorbed radiation dose in mammography. Radiology 1979; 130:485-491. Handbook of glandular tissue doses in mammography. Publication no. FDA 858239. Rockville, MD: Department of Health and Human Services, 1985.
Dosage evolution
7.
8.
screen (case 1, Fig la, curved The distance was indicated by (superior) followed by a number
the distance
superior
to the isocenter.
The (S 95 case 1) was recalled on a routine lumbar spine image by means cursor (Fig lb, curved arrow). was localized, lumbar vertebrae the T-12
indicating
level
number mm in sagittal of the After T-12 could
be easily counted in a caudal direction. For patients who move between ob-
and Methods
in this study. Imwith a 1.5-T superconducting magnet (Signa; GE Medical Systems, Milwaukee) and 4.6 version software. Imaging parameters for the cervicothoracic sagittat scout image included a multiple-echo, multiplanar (MEMP) pulse sequence with repetition time (TR) of 400 msec, echo time (TE) of 10 msec (400/10), matrix of 256 x 192, and two signal averages. An Oxford magnet with a G2 body coil (Oxford Superconducting Technology, Carteret, field
anomalies, of 3% to 21% resonance (MR) imaging has become the diagnostic study of choice for evaluating disk disease of the lumbar spine, these fre-
with
images
in routine
cral spine participated aging was performed
NJ) UMBOSACRAL
scout
angle.
Two referred
Stanton
Vertebrae’
obtained
Materials
6.
on MR Images:
imaging studies brae caudad from C-2 rather than cephalad from the presumed lumbosacral
Spine, abnormalities, 331.131 a Spine, MR. 331.1214
1992;
Segments
those
standard images used in magnetic resonance imaging studies of the lumbar spine, and vertebrae were counted down from C-2 rather than up from L-5. In 200 patients, these techniques reveated 24 transitional vertebrae (15 of sacratization of tumbarization
5.
racic sagittat
transitional vertebrae and disease localion, cervicothoracic sagittat scout images were obtained in additon to the
cases cases
4.
of Transitional
Y. Hahn, BA J. Strobel, MD Francis J. Hahn, MD
accurately
3.
ofLumbosacral
Paul John
To
2.
creditation of Health Care Organizations, 1992;10. Tolbert DD, Hubbar LB. Bushong SC, Khan FM, Payne JT. Quality control issues governing the radiation physicist. Admin Radiol 1989; 8(9):46-52. Mammography-a user’s guide. Report no. 85. Bethesda, Md: National Council on Radiation Protection and Measurements, 1986. 42 CFR 494 (d). Federal Register. December 31, 1990; 251:53525. Harnisor RM. Backscatter factors for diagnostic radiology (1-4 mm Al HVL). Phys Med Biol 1982; 27:1465-1474.
in miffimeters
tainment of a routine lumbar image and a cervicothoracic
spine MR sagittal scout, another thoracolumbar sagittal scout (case 2, Fig 2a) can be obtained by use of the license plate (5 x 11-cm) surface coil and imaging parameters of 400/10, 48-cm field of view, 256 x 192 matrix, and two signal averages. The position of the diaphragm, liver, retroperitoneum, and aorta can also be used as supporting criteria for accurate localization of lower thoracic vertebrae.
Representative Case woman
Cases
1.-A 32-year-old presented with
right-handed occasional back-
ache (during the previous 6 months) radiating to the right buttock area and extending into the back of the thigh. The pain was associated with numbness, tingling, and weakness, and was aggravated by sitting or standing upright for long periods. The pain was relieved by non-weight-bearing rest. The cervicothoracic sagittal scout and vertebrae count from C-2 revealed six lumbar vertebrae, indicating lumbarization of 5-1 (Fig 1). Mild spinal stenosis at L5-L6 (arrow) was also evident. Case 2.-A 36-year-old man presented with recent symptoms of a backache radiating down the lateral thigh to the right foot. The backache typically improved with rest. The clinical presentation pointed to a diagnosis of L-5 radiculopathy. Routine lumbar MR sagittat images (2,000/30) showed a bulging disk at L5-S1 when the level of pain was determined by the conventional method of counting cephalad from the presumed lumbosacral angle. Cervico-
February
1992