2271

583 SEPARATION OF ESTROGEN CONJUGATES BY HIGH PRESSURE LIQUID CHROMATOGRAPHY Musey, P. 1.1, D. C. Collins, and J. R. K. Preedy, Departments of Medicine and Biochemistry, Emory University School of Medicine, Atlanta, Georgia 30303.

Received:

l-19-78 ABSTRACT

High pressure liquid chromatography using a prepacked commercial strong anion exchanger column (u Partisil 10 SAX, 25 cm x 4.6 mm) was used to separate a mixture of eight estrogen conjugates. Chromatographic conditions using aO.OlM potassium phosphate or 0.1 M NaCl as solvent in the isocratic mode are described for the separation of estrone glucosiduronate, 17B-estradiol-3_glucosiduronate, 17&estradiol17_glucosiduronate, estriol-3_glucosiduronate, estriol-16a-glucosiduronate, estriol-17_glucosiduronate, estrone sulfate and 17B-estradiol3-sulfate. This system gives high resolution of the estrogen conjugates in small eluent volumes in less than 30 min. The advantages of this high pressure liquid chromatographic system over other methods of separation are discussed. INTRODUCTION There are three main classes of estrogen conjugates found in biological fluids: glucosiduronates,

sulfates and sulfoglucosiduronates.

Analysis of complex mixtures of estrogen conjugates requires elaborate preparation and

multiple chromatographies

(2-4).

In 1969, we improved

the original (5) technique of DEAE-Sephadex chromatography of estrogen conjugates using a linear gradient of NaCl (6).

However, dextran or

cellulose based ion exchangers have the disadvantage of being soft easily collapsable gels in which the bed volume changes continuously with changes in pH, ionic strength and pressure.

In this situation,

there is no fixed relationship between elution volume and retention of the estrogen conjugate.

This problem was partly solved by development

of an isocratic mode of separation by Musey -et al (7) which allows calculation of the elution volume of a conjugate from the molarity of the eluent. Though high pressure ion exchange chromatography has been used extensively

Voibne

31,

in the analyses of nucleotides

Nwnber

4

S

(8). very little work has

wnnoltmm

April,

2978

been done with steroid conjugates (g-11). have lately been used with some success

Cellulose ion exchangers

in high speed systems to sepa-

rate organic acids (12) and estrogen conjugates

(13).

With the recent

introduction of rigid permanently bonded supports and the speed advantage provided by high pressure liquid chromatography

(HPLC), we have

explored the possibility of separating complex mixtures of estrogen conjugates normally found in physiological

fluids using a high pressure

system. EXPERIMENTAL Materials Estrogen conjugates used in this study included estrone-3-glucosiduronate (ElG) (14); 17B-estradiol-3-glucosiduronate (E23G); 17Bestradiol-17-glucosiduronate (E217G); estriol-3-glucosiduronate (E33G); estriol-16-glucosiduronate (E316G); estriol-17-glucosiduronate (E317G); estrone sulfate (ElS) and 17B-estradiol-3-sulfate (E2S). All the steroid conjugates were obtained from Sigma Chemical Co., St. Louis, MO. The purity of each steroid conjugate was checked by DEAE-Septadex chromatography (7) followed by sulfuric -acid fluorescence (15).. The only impurity found in some of the samples was free steroid which was removed by ether extraction. Estrogen sulfoglucosiduronate was not available to us at the time of this study. Sodium chloride and potassium dihydrogen phosphate solutions were prepared from analytical grade reagents and deionized distilled.water. The pH of the solutions was adjusted with 1 M acetic or phosphoric acid. The eluents were all filtered throu h a 5 um Millipore Solvinert filter (Millipore Corp., Bedford, Mass. 3 before use. Apparatus A relatively simple liquid chromatograph was assembled from comIt consisted of a mercial modules and regular laboratory equipment. Milton Roy (Philadelphia, Pa.) Simplex high pressure pump with a pulse dampener(Lab Data Control, Riviera Beach, Fl.), an in-line 2 micron stainless steel filter and a Valco six-port rotary injection valve (Glenco Scientific, Houston, TX.). The detection system was an inexpensive dual wave'length (280/254 nm) UV Spectrophotometer with a 18 vl flow cell (Glenco Scientific) connected to a 10 inch flat bed potentiometric recorder (Beckman Instrument, Inc.). When radioactive samples were involved in the chromatography, the effluent was collected directly into scintillation vials on a fraction collector. The columns used in this study were prepacked commercial strong anion exchanger, u Partisil 10 SAX, 25 cm x 4.6 mn (Whatman, Inc., Clifton, N. J.). Two of these columns were often operated in series

S

585

T'mmOIDI

to improveresolution. METHODS AND RESULTS All experiments were carried out in the isocratic mode using potassjum phosphate buffer or sodium chloride solutions.. Phosphate buffer concentration of 0.02 M was found adequate for elution of the estrogen conjugates. Standard solutions of authentic estrogen conjugates were prepared in 20% methanol in distilled water in approximate concentrations of 1 ug/pl. Aliquots of up to 10 pl per sample were injected Znto the column. The elution of some estrogen conjugates with phosphate buffer are shown in Figure 1.

Potassium phosphate concentratims aslow as 0.008

or 0.01 M were enough to elute even EJS.

However, the elution of the

Ez3G COLUMN: y Partisil SAX t 4mm x 25cml MODE: ISOCRATlC ELUENT.: 0.01 M KH,PO, pH4.2

FLOW RATE : 0.4mi /min SAMPLE:

0 Figure I.

5 TIME ~~N*~

STANDARDCONJUGATES

15

High pressure liquid ch~atogr~p~ of 52176, E23G, Elf;and ElS on anion exchange micropartisil column using a phosphate buffer as eluent; inlet pressure = 500 psi.

S

586

TPBEOSDl

conjugates in phosphate buffers tion of some of the conjugates.

too fast to allow adequate resolu-

was

Preliminary runs with sodium chloride

solutions indicated that better separation could be achieved with this solvent system. this eluent.

All subsequent chromatographies

Sodium chloride concentration

were conducted with

of 0.1 M was found to give

maximum resolution of the conjugates. Figure 2 shnws the separation of a mixture of four estrogen conjugates using 0.1 M NaC.1 as the eluent.

This column was operated

at 500 psi with a solvent delivery of 0.4 ml/min. system showed a separation of ElG, ElS, E23G and

This chromatographic

E217G that

was

COLUMN : y Par tisil SAX (4mm x 25crn)

m

:,G

MODE : ISOCRATIC ELUENT FLOW

W

RATE

SAMPLE

: O.IM

N&l

pH 5

: Q4ml/min ; STANDARD

CONJUGATES

.

o-5 Figure 2.

IO TIME Separation of E217G, E23G, ElG and E S by HPLC on anion exchange micropartisil column using aaC1 as eluent. Inlet pressure = 500 psi.

S

TDEOXDI

587

superior to the potassium phosphate system shown in Figure 1. Resolution of the estrogen conjugates could be improved by chromatography on two SAX ion exchange columns in series (Figure 3).

In this

system the back pressure was raised to 1200 psi. This gave a flow rate of approximately 0.8 ml/min. Comparison of Figures 2 and 3 indicates that the retention time of each of the estrogen conjugates was increased significantly in the dual column system leading to a much better separation of the individual compounds. The retention times of the conjugates were found to be consistent and reproducible with standard conjugates. In order to test the effect of urinary residue on the separation of estrogen conjugates in the HPLC system, 50 ml of urine was processed on Amberlite XAD-2 resin as described elsewhere (16). Methanolic solutions of standard estroaen

COLUMN :

Ez17G EIS

Ea E316C E,l7G

MODE:

i

r

Partisil SAX 4mm x 25cm)x2

IS OCRATIC

ELUENT : O.IM Ndl pH 4.8 08 ml/min FLOW RATE: SAMPLE : STANDARD CONJUGATES

s % ol

a 0

‘I‘

0

20

25

TIME t MN -1 Figure 3.

Separation of a test mixture of estrogen conjugates on two micropartisil anion exchange columns connected in series.

ST

588

DEOXDI

conjugates were added to the XAD extract and evaporated to dryness. The residue was redissolved in 2 ml of 20% aqueous methanol and aliquots were injected into the chromatograph.

Table I shows the

retention times of five repetitive injections of the urinary extract. There was insignificant variation in the retention times of the conjugates over several applications.

However, the elution volumes of the

estrogen conjugates were significantly

increased after several months

of repeated use or on applications of heavy extracts from large volumes of urine or plasma. (Whatman

Inc.,

This was obviated by introduction of a pre-column

Clifton, N. J.).

A pre-column

is a short guard column

which protects an analytical column when dirty samples are to be run routinely or when the contaminating

components of the sample may include potentially

compounds.

It is usually packed with pellicular equiva-

lent of the analytical column. We have used HPLC extensively to analyze radioactive metabolites Table I.

Retention times (minutes) of estrogen conjugates added to urine extract. Chromatographic conditions were as described in Figure 3. 1

2

3

4

5

Average

E316G/E317G

8.6

8.6

8.9

9.4

9.6

9.0

E33G

10.1

10.2

10.5

10.6

11.0

10.5

E217G

14.1

14.0

14.4

14.8

15.3

14.5

E23G

14.6

14.6

14.9

15.3

15.1

15.1

ElG

17.5

17.5

17.9

18.1

18.8

18.0

E25

20.0

20.2

20.7

21.0

21.6

20.7

El5

22.4

22.6

23.1

23.8

24.4

23.3

Run

Conjugates

S of estrone in human plasma.

TmPEOfDII

Figure 4 shows a representative chromato-

gram of such analyses using chromatographic described in Figure 3. intervals.

589

conditions similar to those

Radioactive fractions were collected at 20 sec.

The retention times of the radioacitve estrogen conjugates

were very similar to those obtained for the standard compounds in Figure 3.

EIG

15

E217G

12

9

6

3

0 Figure 4.

5

IO

15 20 25 30 TIME (MIN.)

35

Separation of radioactive metabolites in plasma following iv administrationof 3H-El into human subject. Chromatographic conditions were similar to Figure 3 but with a solvent flow rate of 0.6 ml/min at 1000 psi

DISCUSSION The high pressure chromatographic

system that we have described

for the separation of estrogen conjugates is more convenient and faster than conventional methods currently in use.

The separation of

is complete in about 30 min usingHPLC,whereas

these conjugates

tional ion exchange cellulose or Sephadex chromatography

conven-

takes over 18

hrs and requires about 1 liter of eluent (6,7). The sequence of elution of the conjugates on conventional anion exchange Sephadex or high pressure anion exchange cellulose is generally in order of their relative polarities.

However, on the bonded

silica gel columns used for HPLC investigation, reverse order.

they tend to elute in

This reverse sequence was maintained even when we ran

the samples with NaCl solutions without adjusting the pH. therefore that reverse phase chromatography

It seems

on bonded silica gel

columns would be ideal for weak acids like estrogen conjugates. The selectivity and resolution of the estrogen conjugates in this study is superior to ion exchange Sephadex or cellulose chromatographic methods. E317G.

There is, however, insufficient separation of E316G and Other authors (13) have found this to be the case for chroma-

tography on ion exchange cellulose. little consequence

Since E317G is thought to be of

in normal extracts of estrogen conjugates (17)

inability to separate it from E316G is not a major problem. significant to note that none of the chromatographic

It is

procedures

currently in use allow for simultaneous separation of estriol and estradiol-estrone

mixed conjugates, be it either celite partition,

paper, Sephadex LH-20, TLC or ion exchange chromatography. We have already demonstrated that estrogen conjugates can be

,ed rather successfully in isocratic mode on DEAE-Sephadex and ;ellulose systems (7). The excellent resolution we have achieved with isocratic elution in this HPLC is very significant and will reduce the substantial initial investment in HPLC system at least for estrogen analyses. A gradient elution in HPLC requires a complicated solvent delivery system consisting of two separate high pressure pumps interfaced with a sophisticated solvent prograannerand a pressurized mixing chamber.

It also requires a detector with a flow through reference

cell to correct for baseline shifts. The use of NaCl solutions does however have a disadvantage because of the long term corrosive effect of NaCl on steel columns and tubing. van Del Wal -et al (13) have reported at length on the factors and mechanics of separation of estrogen glucosiduronates on a high pressure cellulose system. Our experiences with cellulose ion exchangers using a similar system suggest that sulfates and glucosiduronates are poorly resolved. Therefore, it seems that bonded silica gels are the preferred supports for HPLC separation of steroid conjugates based on selectivity, speed, resolution and resistance to high pressure. The advantages of the present high pressure liquid chromatographic system over other methods currently in use for separating estrogen conjugates are the convenience, speed of analyses and the small volumes of eluent required (approximately 10 ml).

Other advantages of the system

are the use of a rather low concentration of counter ion, it does not require a gradient and it can be used to separate mixed conjugates of estriol, 178-estradiol and estrone simultaneously.

592

S

r?lEIEOfDll AC~O~~LEDGE~&NT

This research was supported by NIH Grants ROl-HL-16394 and h, AM-13468. Dr. Collins is a recipient of U.S. Public Health Service Career Research Development Award l-K04-AM-70381 from the National Institute of Arthritis, DigestSve and Metabolic Diseases.

1. 2. 3. 4. f: 7. 8. 9. 10. ::: 2::

15. if:

Mail reprint requests to: Dr. P. I. Musey, Department of Medicine, Emory University School of Medicine, 69 Butler St., S. E., Atlanta, Georgia 30303. Paul, W., C. Stitt, W. 3. Dignam and S, Kushinsky, 3 CHRO~~TOG 45:381 (1969). Taul, w., c. stftt, W.J. Dignam and S. Kushinsky, J CHROMATOG 45:31j2(1962). Jirku, ii.and M: Levitz, 3 CLIN Er~DOCR~NO~METAEI=:615 (1969). Hahnel, R., ANAL BIOCHEM &184 (j965). tfo+o~rk~_Ri'P. I. Musey.and M. Nilsen, STEROIDS H:191 (1969). 0. C. Collins and J. R. K. Preedy, STEROIDS g:657 (19773. l ' Peterson, E. A+ jn +nratnrv Tpy_hniams __.___. .__...- _ ___ in . -Eliachemistrv ___..-...-___ and -. L ~~~~~ Bioloav. Vo~1 2, T. S. Work and E. Work, eds., Elsevier, i970, p: 228. Id, A. G., 8. A. Lodge and N. J. Pound, J CHROMATGG SC1 Butterfii' 11:401 (1973). Henry, R. A., J. A. Schmit and J. F. Die&man, 3 CHROMATOG SC1 9_z513 (1971). Beyer, W. and W. Korozowich* ANN NEW YORK ACAD SC1 E393 (1968). Mororowich, W. 3 CHROMATOG SC1 12:453 (1974). van Del Wal, S. and J. F. K. HaEr, J CHROMATOG E:353 (1974). The following trivial names and abbreviations for steroid conjugates have been used in this paper: ~strone-3-glucosiduronate (ElG~=l7-oxoestra-l,3,5(lO)-trien-3-y1B-D-glucop>ranosiduronate; 17%estradiol-3-glucosiduronate (E23G)=178-hydroxyestra-1,3,5(10t&en-3-yl-B-D-giucopyranosiduronate; 17B-estradiol-17-glucosiduronate(E217G)= 3- hydroxyestra-1,3,5 ~lO)~trien-178-yl-8-D-glucopyranosiduronate; estriol-3-glucosiduronate.(E33G)=l~,l78-dihydroxyestra-l,3,5~10~trien-3-yl-B-D-glucopyranosiduronafe; estriol-16 .-glucosiduronate(E3166)=3,178-dihydroxyestra-1,3,5 (IO)-trien-16a-yl-B-D-glucopyranosiduronate; estriol-17-glucosiduronate (E3176)=3,16a-dihydroxyestra-1,3,5(10)trien-17$-yl-8-D-~lucopyranosiduronate; 17B-estradiol-3-sulfate (E23S)=17f3-hydroxyestra=l,3,5(lO)-trien3-yl-sulfate; estrone-3-sulfate (E~3S)=17-oxoestra-l,3,5(lO)-trien-3-yl-sulfate. Preedy, 3. R. K. and E. H. Atken, J 610 CHEM 23:1300 (1961). Bradlow, H. t., STEROIDS 1?:265 (1968). Smith, E. R. and A. E. Keme, B~OCHEM J 105:1047 (1967).

Separation of estrogen conjugates by high pressure liquid chromatography.

2271 583 SEPARATION OF ESTROGEN CONJUGATES BY HIGH PRESSURE LIQUID CHROMATOGRAPHY Musey, P. 1.1, D. C. Collins, and J. R. K. Preedy, Departments of M...
527KB Sizes 0 Downloads 0 Views