Harrell
G. Chotas,
Small
MS
#{149} Richard
Object
L. Van
Lung,
Mediastinum, phy, digital, 60.1215 Radiology
radiography,
C
radiography
HEST
plays
an essen-
of exposure
From
States.
a chest
contrast
detail
image
that
has
in all regions.
an
image
x-ray
exposure
tech-
obtained
high
penetrate
with
enough the
subdiaphragm
good With
radiographic
niques,
and
Unfortunately,
differences in x-ray that exist between chest it extremely difficult to
acquire
an
to ade-
mediastinum
areas
results in overexposure hyperlucent lung areas.
normally of the more The use of
wide-latitude films to better span the exposure range reduces this problem somewhat, but object contrast regions.
at the in one
Equalization the that
expense or more
radiography
dynamic reaches
of small image
enables
range of x-ray exposure the image recorder to be
reduced by altering the x-ray beam profile incident on the chest so that the mediastinal and subdiaphragmatic
areas
more The
receive
radiation
ization (Optical
comparatively
than
advanced
the
radiography Industries
Netherlands)
search
Building,
Department
is the
first
AMBER
images
have
Duke
C.E.R.) and Eastman
University
Kodak
Medical
Company,
Thoracic
Imaging
Center,
Health
on image
Box 3302,
Sciences
Division,
Durham,
Division,
with
the by expoobject image
ity) and may be further affected by the presence of cross-scattered radiation between chest regions. With this subject contrast information, a film
with
the appropriate
latitude
and
characteristic curve shape can be selected to achieve the desired level of display contrast in each image region.
The acterize
goal of this study was to charthe effects of exposure equalon small object contrast in two (a) by using a linear exposure
ization ways: detector to quantify is available to the (subject contrast),
selected
the contrast that image recorder and (b) by using a
screen-film
the nonlinear of the image trast).
sys-
studies
of dramatic of ana(1,2).
of the effects
Rm 139, Bryan
NC 27710
Rochester,
to be used
been comparof a screen-
recorder. Subject contrast may be related to patient size, body habitus, and clinical condition (eg, lung clar-
AMBER
Several
in visualization in the mediastinum
Research
system
have Selection
image
tionally
to observe effects final
the
(radiographic
Contrast
compared
system
transformation recorder on
con-
in AMBER
with
that
acquired
images
found
was
in conven-
images.
commercially
noted
evaluations
film
equal-
equalization
chest radiography. performance
Quantitative
of Radiology,
equalization
however, few (3,4).
region.
(AMBER) system Oldelft, Delft, The
scanning
tem for observer
lung
multiple-beam
improvements tomic detail
the
MD
AMBER system can be facilitated understanding the effects that sure equalization has on small contrast that is available to the
United
MATERIALS AMBER The sure raphy
(H.G.C.,
NY (R.L.V.M.).
Re-
G.A.J.,
Received
January 14, 1991; revision requested March 8; revision received April 1; accepted May 9. Address reprint requests to H.G.C. Dr Richard Van Metter is an employee of the Eastman Kodak Company. Eastman Kodak manufactures the radiographic intensifying screens and the ifims used in this study. Eastman Kodak has joint marketing agreements with Delft Instruments Medical Imaging, which manufactures the AMBER system used in this study. C RSNA, 1991
AND
System AMBER
equalization and has
METHODS
Description
system
is a dedicated
system for been described
expo-
chest radiogin detail
elsewhere (3). Briefly, the AMBER system (Fig 1) projects a horizontal fan beam that scans
I
E. Ravin,
in the
quately
180:853-859
#{149} Carl
content, atively
conventional
60.1215
PhD
tial role in the diagnosis of thoracic disease and is the most frequently performed x-ray examination the very large transmittance regions makes
radiography, 67.1215 . Radiogra60.1215 #{149} Thorax, radiography,
1991;
Johnson,
Radiography’
available terms:
Allan
in AMBER
Chest
The ability of a commercially available scanning equalization system for chest radiography to render small object contrast in the lung-, mediastinum-, and subdiaphragm-equivalent regions of an acrylic chest phantom was quantitatively evaluated. Images from nine chest phantoms that represented a wide range of patient sizes and dynamic ranges of x-ray transmittance were analyzed. Subject contrast was measured with a photostimulable phosphor detector, and images were acquired in both equalized and nonequalized (conventional) imaging modes. Available subject contrast in the lung-equivalent region was 8%15% lower in the equalized images compared with the nonequalized finages in all phantoms (patient types); contrast in the mediastinum-, retrocardiac-, and subdiaphragm-equivalent regions was 11%-63% higher in the equalized images, with the degree of improvement increasing as patient size and dynamic range increased. Images of each phantom were also acquired with the screenfilm systems currently in use at the authors’ institution, permitting an assessment of the relative performance (in terms of radiographic contrast) of these imagers with and without use of equalization. Index
#{149} G.
PhD
Contrast
Conventional
and
Melter,
the
patient
in a vertical
direction,
exposing the entire chest in 0.8 seconds. The chest image is recorded on an area detector such as film or a photostimulable phosphor (PSP) imaging plate. The fan
Abbreviations: beam equalization
stimulable
AMBER = advanced radiography, PSP
=
multiple photo-
phosphor.
853
Film cassette
Lead strip
Patient (top)
Aft sit with detector Lead strip (bottom)
array
Projected beam
Computer
Beam
Beam
modulators
Fore
(Electronic
feedback
fan
intensity
modulators
control)
slit X-ray
tube
2. Figures
1, 2.
(1) Illustration
of AMBER
system
components.
A horizontal and film tracks
x-ray
fan beam
scans
the chest
vertically,
exposing
an area
detector
(eg, screen-film cassette). A detector array between patient the moving fan beam, and associated exposure modulators dynamically alter the x-ray intensity of each of 20 beam segments during the scan. A 12:1 antiscatter grid (not shown) and narrow lead strips above and below the detector array help reduce scattered radiation in the final image. (2) Illustration of exposure modulators shows the AMBER modulator positions and resultant beam intensity profile when the x-ray fan beam simultaneously crosses the lung and mediastinum.
beam
consists
ments,
each
of 20 adjacent of which
beam
can
be
dently controlled by an intensity tor (located in front of the x-ray
a corresponding
exposure
seg-
tamed with conventional including the distribution
indepenmodulatube) and
detector
diation
back
centimeter) array, and
mechanism
modulators
is measured an electronic
adjusts
each
to partially
and
with feedthe
are
subdiaphragm, fully open (100%)
effects of a small, medium, and large dynamic range of x-ray transmittance on image content. Each phantom was made of acrylic plates (approximately 2.2 cm thick)
x-ray
local
cent lators
regions such as the lungs, the moduare closed in relation to the amount
exposure
flux;
in the
reaching the open), thereby
in these
source-to-image inches. All images
areas
more
plates.
acquired
(35
fixed with
in imaging
ter grids, an
beam
evaluation
equalization ages, the
geometry,
energy, of the alone.
degree
etc)
the
effects
of exposure
the ISO scribed
400 speed in Image
ventional nism was obtained intensity
region.
with
system For the
the equalization and radiographs
a scanning
beam
integrated over the Previous investigations
onstrated used with radiographs
854
images, disabled,
screen-film Acquisition.
decon-
mechawere of uniform
entire chest have dem-
that the AMBER system, when equalization disabled, produces comparable with those ob-
#{149} Radiology
Full chest-equivalent
x 43 cm)
torn
the manufacturer’s “normal” setting, and the system speed (relative base density) was set to 1.26. This system setup provides excellent clinical images when used with
and 1 .2-cm-thick
were
and
acrylic
acrylic
interleaved
plates
with
medi-
subdiaphragm-equivalent
lent plates were made large enough to model both the mediastinum and the heart, and subsequent references to the mediastinal area also pertain to the retrocardiac area. Scatter fractions in the phan-
to permit
For the AMBER imof equalization was set
the medifull chest (in-
plates to provide a balanced distribution of scattering material at all depths within the phantoms. The mediastinum-equiva-
antiscat-
and
0.6-cm-
astinum-
at 72
AMBER system, including both equalized (AMBER) and nonequalized (conventional) images. This was done to eliminate unnecessary experimental variables (eg, differences
with
sary
(minifilm
(Fig 2). The was
shapes to represent subdiaphragm, and
cluding lung) regions. When fully assembled, each phantom resembled the object shown in Figure 3. Fine adjustments in phantom thickness were made as neces-
hyperlu-
detectors reducing
distance were
of varying astinum,
the to maxi-
mize
of radiation mum, 10%
x-ray
ra-
(4).
Nine geometric chest phantoms representing small, medium, and large patient sizes (three phantoms in each size) were constructed to independently assess the
exposure that reaches the image recorder from each beam segment. In the relatively opaque regions of the chest such as the mediastinum modulators
image
Phantoms
of the beam
equalize
recorded
(be-
tween the patient and the image recorder). As the fan beam scans the patient, the local exposure that is transmitted by the patient and the antiscatter grid (12:1, 36 lines per the detector
in the
radiography, of scattered
to
images
were
measured
with
beam stop technique described and Vogelstein (5), and it was that
the
scatter
fractions
the
lead
the
by Plewes
confirmed in images of the
acrylic phantoms (24%-36% in the lungequivalent areas and 48%-81% in the mediastinumand subdiaphragm-equivalent areas) sured ages
were
consistent
with
in conventional obtained with
an
scatter grid Thicknesses
by Niklason
each phantom thicknesses
are
were thicknesses
acrylic transmission those
of acrylic
of the
shown
those
human equivalent
in Table
properties
lung
thickness
phantoms
to
medium,
was
and
range
values,
tected
exposure
held The
computed
in the
20:1
desired
for
the
small,
that they differences population,
consistent
with
dynamic ratio
were
of de-
5:1,
medium,
respectively. were
small,
regions
of equalization,
ranges
all
(ie,
lung-equivalent
use
ranges,
namic
for
size
as the
pa-
hospital.
constant
patient
large).
largest
at our
subdiaphragm-equivalent
dynamic
tient in
and
clinically
of a given
believe mission
the x-ray regions
Lung
and of
average-size,
examined
without
1. Lung
equivalent
smallest,
tients
and
imanti-
et al (6). in each region
set by establishing that had gross
unobstructed
mea-
chest
Figure 3. Acrylic chest phantom. Nine geometric chest phantoms were constructed of 2.2-cm-thick acrylic plates, with full-area “lung” plates interleaved with smaller plates representing the mediastinal and subdiaphragm regions of the chest. Small aluminum cylinders were included in each phantorn to create test signals for quantitative analysis.
and These
selected
because
10:1, large dywe
span the range of transobserved in our paand those
because
they
are
obtained
by
Lem-
September
1991
Table
sure latitude of the screen-film was the case in this investigation. Four images were obtained
1 Phantom
Acrylic
Thicknesses
Phantom Size
Dynamic
Lung
Mediastinum
Small
6.7
14.5
Medium
6.7
16.2
Medium
Large
6.7
18.0
Small
11.1 11.1 11.1 15.3 15.3 15.3
Medium Large Large
Small
Medium Large The dynamic spectively.
range values
*
Table
were
exiting
2 Settings
Exposure
E#{231}posure Setfings(mA)
Conven-
Phantom
Dynamic
Size
Range
Small Medium Large
All All All
use
AMBER
of the
same
Adjustments within each altering the
tional
65
21
130
52 130
330
et al (7) in a clinical
mers
population
measurement
with
by attaching a 3 x 4 0.6-cm-thick alu-
were
positioned
projected region (three
lung-equivalent
so
into the in each
lung), three in the mediastinum-equivalent region, and three in the subcliaphragm-equivalent region. This “signals plate” was positioned midway between the front and back surfaces of each phantom. Measurements sion and AMBER
region
of each
of local system
phantom
x-ray transmisresponse in each
were
made
by
means of the x-ray detection and readout capability of the AMBER system. The AMBER system contains an on-board com-
puter system two-dimensional sure)
and
232
serial
that measures and stores arrays of detector (expo-
modulator
data used
during
(feedback
the procedure.
communications
line
control)
An RSand
com-
puter terminal permits access to this information in tabular form. The AMBER system’s
detector
values
23.7
23.4
26.0 28.6
26.8 30.3 and
chest
each
large
region
tional
medium
phantoms,
after
Image
re-
conven-
as two
used
Conn)
screen-ifim
as image
has
been
are linear
systems
detectors.
The PSP
described
in detail
else-
When
the
imaging
to x rays, the incident
stored
as a latent
readout
system
image.
plate
radiation
is
An automated
uses a scanning
laser beam
to stimulate the stored energy in the plate and to extract the image information, resuiting in a digital image that is immedi-
ately
available
for display monitors.
for quantitative
analysis
or
on laser-printed film or video In this study, only the digital
data from the PSP system were anathe laser-printed ifims were not
image
lyzed; used.
Unlike
film, the sensitivity
of the PSP
imaging plate to x rays is linear (with respect to the logarithm of exposure) over a very
wide
[2.58
x
been
shown
exposure
106_2.58
range x
i0_2
to be well
(0.01-100 mCfkgJ)
suited
mR
and
for quantita-
tive radiographic exposure measurement (10). PSP images in this investigation were processed with the PSP system at a fixed sensitivity of 200, fixed latitude cades), and fixed time between
Volume
180
the #{149} Number
detected 3
exposure
sure
intensity.
When
used
(4.0 deexposure
in this
way,
the
PSP system is superior to film as a quantitative image detector in measuring x-ray that
may
fall outside
Kodak),
C
an ISO
screen-film because
currently
they
in use in
ting)
was
current
adjusted
(milliampere
set-
independently
for each
phantom, and these settings are listed in Table 2. Exposure settings were selected
scanning
so that the modu25% open
the lung-equivalent by
the
“modulator
region data”
ar-
and lighter
than
the average
lung
optical density; the estimated average optical density was used in this study to represent the middle optical density of a wellexposed lung and to account for the variety of optical densities in that region.
Image
Analysis
Contrast was measured the PSP and film images test
signal
locations
separately in at each of the 12
in each
image,
and
the
mean contrast in each chest-equivalent region of each image was computed. In the PSP images, the subject contrast detected in the image plane was computed as (E0 , where E,, was the ob-
has
exposures
observation
recorders
The generator
darker
is
set by examining
direct
x-ray
exposed
(Eastman
a
ray of the AMBER system). Images obtamed at all exposure settings had average optical densities in the lung-equivalent regions comparable with those of clinical images in each film type (1.33 ± .07 for AMBER and 0.99 ± .03 for conventional radiographs). Every clinical chest image, of course, contains lung regions that are
as well
(8,9). Briefly, the PSP system is on a reusable storage phosphor imaging plate that serves as an area detector and resides in a cassette that is similar to (and the same size as) a traditional screenfilm cassette.
Ortho Lanex
our hospital for clinical images obtained with the AMBER system and in our dedicated chest examination rooms, respectively. Film was developed in an RP X-OMAT processor (Eastman Kodak), and periodic sensitometry was performed throughout the procedure to verify processing stability. Film was processed 5 minutes after x-ray exposure. All images were obtained at 140 kV with 4.3-mm aluminum-equivalent filtration.
(as indicated
(PCR 901; Philips
Shelton,
traditional
were
with
system. The two were selected
are the image
when
Systems,
and two
one was obtained
for each AMBER image lators were approximately
system
of each
images
and the other with gradient, 1.8) with
screens
250-speed combinations
Acquisition
Medical
images,
detector film (average PSP
and readout (20 minutes) to permit the conversion of digital pixel values to expo-
permitting
to detected
25.6
exposure, of relative x-ray flux exiting each location in the chest phantom. The desired dynamic range of x-ray transmittance for each phantom was with
respect
data
23.1
where based
rectangular)
plates. The cylinders that six signals were
19.3 22.3
tional exposures and then adjusting acrylic thicknesses in the chest phantoms. Similarly, the degree to which each modulator was closed or left open at each scanning position was determined by examin-
system
to one of the full chestphantom
cylinders (large
19.6 21.2
A PSP imaging
technique.
mediastinumand subdiaphragm-equivalent areas. Image test signals used for contrast
minum equivalent
14.2 17.2 20.6
ing the modulator arrays after an equalized exposure was made. This information was used to establish the appropriate x-ray exposure level for each phantom, as described below.
in dynamic range values phantom size were made by acrylic thicknesses only in the
analysis were created array of 0.6-cm-diameter,
Subdiaphragm
5:1, 10:1, and 20:1 for the small, medium,
two AMBER
as
conventional images. Of the AMBER images, one was obtained with a PSP detector and the other with T-MAT L film (average gradient, 2.2) with Lanex regular screens (Eastman Kodak, Rochester, NY), an ISO 400-speed system. Of the conven-
f
Tr
Range*
Small
phantom:
Thidmess (cm)
.
system,
the expo-
served mean x-ray exposure in the circular shadow of the aluminum test signal and was the mean exposure in the area immediately surrounding that shadow. Exposure measurements (in milliroentgens) used in this equation were cornputed from mean PSP pixel values in each small
area
the mean
as 121.29*
PSP pixel
where
P
is
value.
This equation is of the form F = a*10P which has been reported previously (10), with coefficients a and 13 derived by fitting a curve to data from a calibrated exposure series and the
Radiology
#{149} 855
resultant
PSP
perirnental
pixel
Subject
contrast
ages
values
conditions
can
be
formation
measured
necessarily
the
as the
is available
captured
ex-
earlier.
in the
interpreted
that
image
under
described
PSP
im-
contrast
in-
to (though
by)
the
C
a
a
0
not
C 0
C.)
screen-film
a 0
recorder.
In the film images,
contrast
was
sured as the difference densities inside and
in observed surrounding
signal
quantity
location.
This
meaoptical test
s/s
is referred
s/U
S/L
Phantom
each
U/S
N/N
M/L
(Size/Dynamic
L/S
L/N
L/L
the was
outside mean used
tical
each
value in the
densities
test
of these difference
were
Pa),
with
It should
be
area,
±0.01
e C 0
a
C.)
0
Op-
with
(Tobias
a TA-
Associates,
s/s
Ivy-
s/s
that
S/N
S/L
Phantom
precision.
noted
N/S
U/N
Mu
(Size/Dynamic
L/S
L/U
Lu
reason, in the
used
and and
the
image
computed
plane
detected
was
exposure
AMBER Dynamic
aphragm-equivalent with age,
lung-
C 0
range for
C.)
ratio
and
of
subdi-
as measured
pixel values to exposure
in the PSP values.
im-
RESULTS Figure trast
4 illustrates
measured
the
in the
subject
PSP
con-
images
s/s
S/M
S/L
Phantom
M/S
tAIL
M/M
(Size/Dynamic
1/S
L/M
L/L
in the
AMBER
and
Figure 4. Bar graphs illustrate the subject contrast of test signals measured in the image plane with the PSP image recorder. Separate plots for the observed contrast in lung-, mediastinum-, and subdiaphragm-equivalent regions are shown as a function of phantom size and dynamic range. Black bars = AMBER images, striped bars = conventional images. S = small, M = medium, L = large.
with
conven-
size
thickness)
(lung
increasing
num
and
Subject
lent
by
regions on
in the
of the
average,
AMBER
conventional
same
phantom
C 0
the mediastinumphragm-equivalent subject
was
images
tional
images
with
increasing
increasing
that
of the
8%-15%).
higher
than
In
and
in the
in the
conven-
decreased
phantom
dynamic
sure equalization contrast relative
increased subject to the conventional
11%-63%).
In the
regions,
generally centage centages
856
of 31%
increased by with equalization. are summarized
#{149} Radiology
Medastthum
Phantom
6.
Bar graphs
show
of
2(
14
Subdiaphrgm
Region
5. Bar graph shows ratios of subject contrast in AMBER and conventional images for each region of interest (lung-, mediastinum-, and subdiaphragm-equivalent regions) and for each phantom (size and dynamic range indicated). Ratios greater than 1.0 indicate higher subject contrast in the AMBER image; ratios less than 1.0 indicate AMBER contrast is lower than in the conventional image. S = small, M = medium, L = large.
contrast
Figure
me-
expo-
(range,
subdiaphragm-
the
1/1
-
s/s
MIS 1/S 5/M SlIMIJM
S/L
M/t,
IA
Phantom
Figure 7. Bar graph illustrates the dynamic range (DR) of each phantom in an unequalized exposure as determined by means of two independent measurement techniques. Differences in measured values for a given phantom may be attributable to scattered radiation between regions and systematic differences in the measurement techniques. S = small, M = medium, L = large.
and
In the
regions,
equivalent
L/U
both
size
range.
by an average
Liip
Figure
than
diastinum-equivalent
image
0
and the subdiaregions, however,
contrast
AMBER
1/S
Range)
1/
z
images
images
U/N
U/S
(Size/Dynamic
0 0
in mediasti-
lower
(range,
comparison
S/I
radiographic contrast measured in AMBER and conventional images. Separate plots for the observed contrast in the lung-, mediastinum-, and subdiaphragm-equivalent regions are shown as a function of phantom size and dynamic range. Black bars = AMBER images, striped bars = conventional images. S = small, M = medium, L = large.
.
lung-equiva-
11%
Figure
S/N
Phantom
a
thickness).
in the
U/I
s/s
rela-
variations range (ie,
subdiaphragm contrast
was,
was
I JUL.,.
2/
phantom
but
lively unaffected phantom dynamic
___________________
:2:
in
tional images is shown. In the lungequivalent region, subject contrast decreased
J
Range)
each region of each chest phantom. Data from each of the three regions are plotted separately, and the subject contrast
Ill.
1./N
a
detector arrange in the
areas
the mean transformed
f
between For this
as the
in the
L/S
Range)
of
of dynamic were also made
comparison with ray measurements.
N/L
_._Subdiaphragm
O.4O’j
measurements
the distance the detector.
measurements image plane
S/PA S/L M/S N/N Phantom (Size/Dynamic
Range)
x-ray transmission through the chest can be strongly influenced by the detection technique the phantom
Range)
0
and
measurements calculation.
measured
TBX densitometer land,
signal
(Size/Dynamic
to
as radiographic contrast and is a function of both subject contrast and film gamma. Four optical density measurements were
made
Phantom
Range)
was
a larger perThese perin Figure 5.
6 shows
the
radiographic
contrast measured in each region of each phantom. In every case, radiographic contrast in the AMBER im-
exceeded conventional ages
phantom. gions, the AMBER
sured
that image
measured in the of the same
In the lung-equivalent reradiographic contrast in the images exceeded that meain the conventional images by
an average num-equivalent
exceeded
of 57%
.
In the
regions,
that
measured
mediastithe
contrast
in the
ventional images by an average 382% . In the subdiaphragm-equiva-
lent
region,
this
contrast
increase
even greater. In two cases, the intensity of the subdiaphragm-equivalent regions was not visually able in the conventional images
September
conof
was signal detectbut
1991
Log X-Ray
Transmittance
Log X-Ray
(Relative)
Transmittance
(Relative)
b.
a. Figure
trate
8. Graphs illustrate the effective small-area contrast)
was
still
on the
measurable
AMBER
Dynamic
and
response and
of two film (b) film contrast
and
visualized
images.
range
measurements
of
within 10% of the target ratios 10:1, and 20:1) in each dynamic
(5:1, range
group. Dynamic range values computed from the PSP exposure data, however, were consistently smaller than the AMBER-derived values in the medium and large dynamic range phantom groups, and within both of those groups, the discrepancy creased as the phantom size thickness) increased.
in(lung
DISCUSSION Contrast
The primary purpose of the AMBER system is to modify the x-ray exposure intensities entering the chest
to substantially
posure aphragm
increase
of the mediastinum while maintaining
the
ex-
and subdiproper
x-ray fluence through the lungs for optimum film exposure. One expected result of exposure equalization is a change
in
the
distribution
of
tered
radiation
which,
in turn,
ences ages,
the scatter measured
fractions in the as the ratio of
scat-
influim-
scattered and total (primary plus scattered) radiation. This effect has been measured in both an acrylic step wedge and in an anthropomorphic chest phantom by Chotas et al (4). In both cases, exposure equalization increased
the
scatter
fraction
thinner, lung-equivalent alive to the nonequalized Volume
180
#{149} Number
in the
regions (relexposure) 3
in this
significantly
fractions
each phantom without use of equalization are shown in Figure 7. Dynamic range values computed from the AMBER detector array data were
Subject
types used (gamma).
study
in regard
decreased
in the
to (a) film
the
mediastinum-
scatter and
subdiaphragm-equivalent regions. The changes in image scatter fractions were attributed to changes in relative amounts of cross-scatter between chest regions. In this study, we demonstrate that AMBER exposure equalization decreases the subject contrast of small test signals in the lung field and increases the subject contrast in the mediastinum and subdiaphragm. This finding is consistent with the altered scatter fractions reported by Chotas et al, but cannot be compared quantitatively because of the difference in the phantoms used in the two studies. It is interesting to note that the magnitudes of the observed changes in subject contrast (AMBER vs conventional) were very similar in all phantoms and in all simulated chest regions (Fig 4), with differences ranging only from 0.03 to 0.06. This is rather surprising given the large variations that existed in phantom dynamic range and “lung” thickness. This
fields sents trast
contrast
reduction
of the AMBER
in the
images
lung
repre-
only a small fraction of the conthat is available in the conven-
tional image (ie, causes minor subject contrast degradation), but in the more opaque chest regions, this same contrast
change
is large
compared
that in the conventional significantly increases subject contrast).
Radiographic the
with
image (ie, mediastinal
Contrast
Increased radiographic test signal areas was
contrast observed
of in
optical
density
(dotted
line
segments
illus-
all regions of all the AMBER images when compared with that in the conventional film radiographs. These observed changes in radiographic contrast
arise
from
several
factors,
as
follows: 1. Equalization compresses the dynamic range of transmitted exposure, which permits a higher contrast film to be used. In this investigation, Ortho
C film,
which
is routinely
used
for
conventional upright chest examinations, has been replaced by T-MAT L film. With film optical density between 1.0 and 1.5, this change in film causes
a contrast
increase
of approxi-
mately
35% at any particular density. 2. The average optical density of the lungs, mediastinum, and subdiaphragm regions is higher on the AMBER images than on the conventional images. Because film contrast increases with optical density up to 2.0 in these
selected
films,
the
AMBER
images have higher small object contrast in all chest regions. 3. As shown in Figure 4, AMBER changes the subject contrast, with contrast reduced in the lungs and increased in the mediastinum and subdiaphragm. Predictions of radiographic contrast on the basis of subject contrast (from the quantitative PSP image data) and the sensitometry curves for the two selected films agree well with the observed radiographic contrast values. This agreement increases our confidence in the use of the PSP system as a tool for quantitative exposure measurements. Figure 8 shows predictions of film optical density and film contrast for the two film types used in this study. Radiology
#{149} 857
The two represent
thickest lines the response
in each of the
plot two film
types as they were used in this investigation, and the thinnest line represents the predicted response of the
higher-contrast film (Kodak T-MAT without AMBER exposure equalization. In the optical density plot (Fig 8a),
the
inflection
points
in the
L)
higher
contrast film curve show the range over which AMBER exposure equalization occurred. Within this range, local film response (ie, in areas smaller than the AMBER system equalizes) is indicated by the dashed line segments.
Predicted
local
film
contrast
is
presented in Figure 8b. This figure illustrates that the higher contrast film, when used with AMBER, should yield a higher small area contrast at all image locations and span a larger effective x-ray transmittance range relative
to the
wider
latitude
film
the conventional ment. This agrees our study.
imaging environwith the results
Dynamic
Measurements
Range
A comparison values
by
of dynamic
means
ment
techniques
array
vs PSP
of the
b.
in
of
range
two
(AMBER exposure
a.
measure-
detector
data)
revealed
significant differences in the dynamic range of each phantom in the medium and large dynamic range phantom groups (Fig 7). This difference may
result
from
exposure during whereas
a point
plate
measures
sure
accumulated
procedure radiation regions).
the
fact
that
the
detector array measures at a distinct moment the imaging procedure,
AMBER
on
the
the
PSP
integrated
x-ray in time
expo-
throughout
the
(ie, including both primary and scatter from adjacent The
AMBER-based
Observations
The potential clinical findings is illustrated ventional and AMBER our
graphs
in Figure
importance of in the conchest radio-
9. In the
postero-
anterior AMBER images, dramatic improvement in visualization of mediastinal, retrocardiac, and subdiaphragmatic areas is due to the increased radiographic contrast in these areas. Normal anatomic landmarks
such as the tracheal carina, azygoesophageal recess, and posterior costophrenic gutters are routinely and easily visualized. In the lateral 858
#{149} Radiology
Figure
9.
Conventional
and
AMBER
chest
radiographs.
image obtained with Ortho C film with Lanex image obtained with T-MAT L film with Lanex obtained with Ortho C film with Lanex medium with T-MAT L film with Lanex regular screens.
medium regular screens.
(a) Conventional
posteroanterior
screens. (b) AMBER posteroanterior screens. (c) Conventional lateral image (d) AMBER lateral image obtained
exposure
data, therefore, are likely to be less contaminated by scattered radiation and may represent the true dynamic range of the phantom better. Clinical
d.
C.
imaging
AMBER
sualization
image, there is improved viof the apical areas of the
lung. Accordingly, abnormalities in the mediastinum, retrocardiac, subdiaphragmatic, and apical lung areas are potentially more easily recog-
nized.
without
might of the artifacts
Considerable given to the
attention
choice
has
been
of film used
with
the AMBER system in our hospital. In choosing a film, we attempted to satisfy two basic requirements. First, the
film must that spans
have an exposure latitude the dynamic range of
transmitted examinations. ily met with
x-ray
and difficult requirement film must provide “good” contrast detail in all image
films
compression provides.
exposures in clinical chest This requirement is easa variety of commercial
due
to the
that the The second,
dynamic AMBER more
range system subjective
is that the anatomic regions excessive image contrast that decrease the diagnostic utility image by amplifying system or
rendering
an
image
with
so much contrast that the normal “gestalt” of the image is lost. Plewes et al (11) have shown that the maximum film contrast that can be used effectively in scanning equalization systems varies inversely with the area of the beam. However, some equalization artifacts such as edge enhancement at the interface between regions of large transmission differences (eg, at the diaphragm or the heart-lung border) can become distracting as film contrast increases. Other AMBER arti-
September
1991
facts (such as vertical stripes in the image that can result when the modulators are driven to their limit) are made more apparent with higher film contrast. Extreme contrast enhancement of normal anatomy can also be problematic in a diagnostic task, particularly when AMBER images are viewed in an environment that includes other (conventional) chest radiographs. Since the clinical evaluation of radiographic images requires both pattern recognition prominence
and estimations of the of various structures,
large trast
differences in radiographic conamong chest images in a radiology department can be disadvantageous in the detection and diagnosis of disease. In preliminary studies, we have determined
that
a satisfactory
balance
of these requirements can be achieved by using a film (T-MAT L) that provides approximately 20% greater contrast than the film used in our hospital for conventional chest radiography. The reduction in subject contrast in the lungs that results from exposure equalization (due to increased scatter fractions) with AMBER
is slightly more than offset by this increase in ifim contrast. Higher contrast films were not acceptable because they were judged to yield images that were too different from the conventional chest films and that were more likely to highlight image artifacts. It is important to note, however, that in our institution, the standard film for conventional chest radiography is a wide latitude film. In other institutions in which a higher contrast film is used for conventional chest radiography (and is therefore already familiar to the radiologists), an alternative film selection might prove acceptable for use with An important advantage of AMBER is that more film types can become eligible for use in chest radiography (due to the reduced dynamic range requirements), and, therefore, the radiologist has more options when selecting the screen-film system that is most visually acceptable. #{149}
180
#{149} Number
3
Conf. Madison: 3.
4.
5.
Schultze
H, et aL
Kool
U, Busscher
Advanced
radiography simulated nodule ogy 1988; 169:35-39. Ravin CE. Advanced equalization radiography ization
DLT,
Vlasbloem
multiple-beam equalin chest radiology: a detection study. Radiol-
Chest Imaging
Proc
Medical
Physics,
6.
7.
1987;
60-
H, Schultze multiple-beam
Kool U. AMBER: a equalization sysradiography. Radiology 1988;
169:29-34. Chotas HG, Floyd CE, Dobbins JT, Lo JY, Ravin CE. Scatter fractions in AMBER imaging. Radiology 1990; 177:879-880. Plewes DB, Vogelstein E. A scanning sys-
tem for chest radiography exposure control: practical tion. Med Phys 1983; Niklason LT, Sorenson
Scattered Med Phys Lemmers Elburg H, transmission population:
imaging
with regional implementa-
10:655-663. JA, Nelson
JA.
radiation in chest radiography. 1981; 8:677-681. HESAJ, Schultze Kool U, Van Van Metier R. Low frequency of the chest in an out-patient implications Proc
system.
SPIE
for the AMBER 1990; 1231:437-
441.
CRB, Matthews CC, Scheinhorn D, Balter S. Digital imaging of the chest. J Thorac Imaging 1985; 1:1-13.
8.
Merritt
9.
Sonoda
M, Takano
Computed laser
stimulated
J, Kato
M, Miyahara
radiography
utilizing
luminiscence.
H.
scanning
Radiology
1983; 148:833-838. 10.
Floyd
CE, Chotas
HG,
Dobbins
JT, Ravin
CE. Quantitative radiographic imaging using a photostimulable phosphor system. Med Phys 1990; 17:454-459.
References 1.
63. Vlasbloem scanning
tem for chest
AMBER.
2.
Volume
clinical experience.
11.
Plewes
DB, McFaulJ,
Ivanovic
M.
Maxi-
mi.zing film contrast for scanning equalization radiography. Med Phys 1990; 17:357361.
multiple beam (AMBER): early
Radiology
#{149} 859