Steve H. Parker, MD Kenneth D. Hopper,
#{149} Jeffrey
MD
D. Lovin, MD #{149} William E. Jobe, MD F. Yakes, MD #{149} Brian J Burke, MD
Stereotactic
Breast
Biopsy
G
terms: 00.1299
Breast
#{149}
Radiology
Biopsies,
technology
Breast
neoplasms,
#{149}
neoplasms,
#{149} Breast, diagnosis,
localization,
ing
lesions
are
referred
calization
followed
sy.
a relatively
While
From
tinely (1,2),
safe
the
Radiology Army
Associates,
Medical
Center,
our
in such
with
inability
biopsies
to sample
the
accuracy. available provided
to apply
biopsy
fidently
our
gun lesion.
conducted mammography
with
fine-needle
Cob
with
breast
and
con-
a mammognaphically
tions tactic
Englewood,
The advent stereotactic us with the
have
Recent with
aspiration been
Cob
(S.H.P.,
(S.H.P.,
J.D.L.,
investiga-
use of stereoin conjunction
W.E.J.); J.M.L.,
biopsy
successful
in en-
diagnosis numbers
but of
Department K.D.H.,
of Radi-
W.F.Y.,
B.J.B.);
of Radiology, Pennsylvania State University, Milton S. Hershey Medical Cen(K.D.H.). From the 1989 RSNA scientific assembly. Received January 9, 1990;
revision
March
requested
14; revision
#{149}
received
and
accepted
May
8. Address
reprint
requests
to
8200 E Belleview, Suite 124, Englewood, CO 80011. herein are the private views of the authors and are not to the views of the Department of the Army or the Depart-
of Defense. RSNA, 1990 See also the articles by Hopper et al (pp 671-676), Poster et al (pp 677-679) and the editorial by Bernardino (pp 615-616) in this issue.
Gun’
to those
concomitant consecutive opsy
of the
corresponding,
surgical patients
biopsies referred
in 103 for bi-
of mammographically
nonpalpable
sugges-
lesions.
MATERIALS
AND
METHODS
During a 13-month period, under a protocol approved by the clinical investigation/human use committee and after informed consent was obtained, 103 stereotactic needle core breast biopsies were performed. After the needle core biopsy and localization wire placement, the patient underwent traditional surgical excibiopsy.
The
surgical
biopsies
were
performed by one of three surgical residents, all under direct staff supervision. All needle core biopsies by one of four radiologists
ologists with
lesions
experience
to the
target
suggestive
breast
sies
sional
and the Department ter, Hershey, Penn
S.H.P., Radiology Imaging Associates, The opinions or assertions contained be construed as official or as reflecting
MD
false-negative results and cases with insufficient tissue sampling (4-6). We therefore decided to compare results of stereotactic needle core gun biop-
tive
by radiologists enjoyed improved
with pinpoint of commercially mammography
bi-
Aurora,
biop-
procedure,
the use of an automated biopsy device (3). We were frustrated in our initial attempts to utilize the instrument for breast biopsies, however, because of
00.125
Imaging
bo-
performed and we have
success
1990; 176:741-747
Fitzsimons
needle
by surgical
abling breast cancer have had significant
1
for
there is frequently some delay in anranging the biopsy, the cost is significant, and there is some potential for complications and disfigurement. Accurate, dependable needle biopsies would eliminate or considerably reduce these drawbacks. Percutaneous biopsy of various sites other than the breast are rou-
(FNAB)
ology,
M. Luethke,
a Biopsy
REATER
means opsy, 00.3
with
utilization of breast screening in asymptomatic women is leading to the discovery of an even-increasing number of nonpalpable lesions suggestive of cancer. Patients with more ominous-appear-
One hundred three patients underwent stereotactic breast biopsy with an 18-, 16-, or 14-gauge cutting needle and a biopsy gun. After biopsy, a localization wire was placed and surgical biopsy performed. There was agreement of the histologic results in 89 cases (87%) including 14 of 16 cancers (87%) (x 0.806). The gun biopsy yielded the correct diagnosis in four cases involving a lesion (including one cancer) that was missed at the surgical biopsy. Nine cases in which the lesion was missed at gun biopsy can be related to insufficient needle size, the greater difficulty in using one of the two stereotactic devices, and early inexperience with the technique. A 14-gauge needle was used in the last 29 biopsies, the results of which agreed with the surgical pathologic findings in 28 cases (97%). With greater experience, stereotacticguided large-gauge automated percutaneous biopsy may prove to be an acceptable alternative to surgical biopsy in women with breast masses suspected at mammography. Index
#{149} James
#{149} Wayne
and the
two
Biopty
Urobogicab, by Radiplast, junction with
were (two
radiology gun
Covington, Uppsala, 18-, 16-,
performed staff radi-
residents)
(distributed
by
Bard
Ga; manufactured Sweden) or 14-gauge
in conBiopty-
cut needles. Sixty-five patients underwent biopsy with use of 18-gauge needies, nine patients with 16-gauge needles, and 29 patients with 14-gauge needles. One hundred one patients underwent biopsy with the “long-throw” Biopty gun, which has a 2.3-cm needle excursion. Two patients underwent biopsy with the “short-throw” Biopty gun, which has a 1.15-cm needle excursion. For the first 30 biopsies, stereotactic guidance was provided by the Senographe Mammographic System 600T (GE Medical Systems, Milwaukee) coupled with the Steneotix computerized stereotactic needle localization device (GE Medical Systems). The biopsy method used with this device is described elsewhere (7). Because of the logistical problems nebated to the use of the Biopty gun with the Senographe unit, we switched to the Mammotest Stereotactic System (Fischer
ment
725-727),
and
Elvin
et al (pp
Abbreviation:
FNAB
fine-needle
aspiration
biopsy.
741
Figure
1. Mammotest lies comfortably
Patient
stereotactic system. on the table with
her breast placed through an opening in the table. All hardware is located beneath the table, with the compression plate (solid anrow), stereotactic apparatus (open straight arrow),
biopsy
gun,
curved
arrow)
out
gun
is firmly
mounted bility
on of
locked depth
unwanted
and of
housing
patient’s
into slide, gun/needle
a.
(open sight.
Biopsy
the housing reducing
the
possi-
displace-
ment.
Imaging, Denver) after the first 30 patients. The remaining 73 patients underwent biopsy with this system. The Mammotest unit allows the patient to lie prone on a table with her breast protruding through an opening at the front of the table. The compression plate, x-ray tube, stereotactic apparatus, and the biopsy device are all located underneath the table, out of sight of the patient and free from encumbrances (Fig 1). The manufacturer designed a dedicated housing for the Biopty gun that attaches to the depth slide on the stereotactic apparatus. In addition, software was incorporated that provides an audible and visual warning on the digitizer if the 2.3-cm excursion of the needle would result in the needle striking the Bucky plate. The first step in the Mammotest biopsy sequence is to obtain a mammographic bocalizing view that includes visualization of the whole breast to ensure that the besion is indeed within the aperture of the compression plate. After confirmation, the breast within the aperture is cleansed with povidone-iodmne (Betadine; Purdue Frederick, Norwalk, Conn). Stereo views are then obtained by swinging the x-ray tube 15#{176} off center in each direction. The two stereo images are exposed adjacent to each other on the same film (Fig 2a). Computer-generated coordinates derived from the stereo images are dialed into the stereotactic unit, and the Biopty gun with needle already loaded is aligned automaticably on the proper trajectory. The needle is then advanced (by moving the gun housing) through a small skin nick to the calculated depth, and stereo views are obtamed again to confirm that the needle is poised over the lesion (Fig 2b). The needle is then backed up at least 2 mm to ensure proper postfine positioning of the sample notch within the lesion (3). The gun is fired, and a final set of stereo views ensures that the needle has tra742
#{149} Radiology
b.
C.
Figure
2.
Aligning
are obtained, aligning
and
the
gun
the
gun
and
the coordinates and
needle
document
that
the
needle
and
generated
onto
the skin nick to the computed needle tip is over the lesion.
needle the
proper
crements All biopsy
to posteriorly
from the center passes traverse
these
trajectory.
needle
views (b)
The
position.
are dialed needle
depth, and stereo views are obtained (c) Final set of stereo views is obtained has
traversed
the
versed the lesion (Fig 2c). The needle is removed, and the core tissue obtained is placed in fonmalin. Three to four passes are made through the lesion to canvass it anteriorly
confirming
from
in
measured
in-
of the lesion. the same skin
nick.
After the stereotactic biopsy, a conventionai needle localization hook-wire was placed by using the same coordinates as
(a) Stereo
into
is advanced
again after
views
the main
unit
through
to confirm that the gun is fired
the to
lesion.
used for the needle core biopsy. A conventional mammogram was obtained to ensure adequate wire placement. The patient then underwent outpatient surgical excisional biopsy under local anesthetic. The local anesthetic occasionally had to be supplemented by intravenously administered sedatives. The histopathobogic specimens from each biopsy were then compared by the same pathologist to evaluate for agreement.
September
1990
There were 20 with agreement from the needle specimens in 41 nomas diagnosed missed
fibroadenomas, between findings core and surgical cases. Four fibroadeat surgery were
at stereotactic
gun
biopsy.
One of the fibroadenomas was missed at gun biopsy because the stereotactic machine was out of calibration. Another fibroadenoma was missed
a. Figure
b.
3. Comparison of gun needle core and surgical histopathologic cancer. (a) Intraductal carcinoma from the gun core specimen (arrows). carcinoma from surgical excisional biopsy specimen (arrows).
specimens of breast (b) Same intraductal
because
of patient
movement.
A third was most likely missed as result of displacement of the lesion away from the advancing needle The fourth fibroadenoma missed gun biopsy was probably also due lesion movement. Two fibroadenomas
diagnosed
missed
at gun
at surgery
biopsy
(Fig
a tip. at to
were
4).
In the 55 cases of fibrocystic change, the findings from the needle cone and surgical specimens were in full
agreement
in 37 and
partial
agreement in 14 cases. Full agreements were tabulated when three or more of the elements of that panticulan area
of fibrocystic
change
were
correlated (Fig 5). If only one or two elements were correlated, then the case was tabulated as a partial agreement. There were four cases in which the gun biopsy specimen failed to contain any elements of fibrocystic change and contained only normal breast tissue. These may be a reflection of the random distribution of fiRESULTS
biopsy
The results from 102 of the 103 patients who underwent steneotactic gun biopsy were available for companison with results from surgical excisional biopsy. (Needle biopsy specimens from one patient were lost.) Samples sufficient for histobogic analysis were obtained by means of stereotactic needle core biopsy in 101 of the 102 patients (99%). The case in which insufficient tissue was obtained was one of the two biopsies performed with the “short-throw” Biopty gun and an 18-gauge needle. In 12 other cases there was disagreement between the histopathobogic diagnoses reached with the two biopsy methods. In eight of those cases, adequate tissue was obtained with the gun biopsy, but the diagnosis made was different from that made from the surgical specimen. In the other four cases, the results at gun biopsy corresponded to the mammographic findings, but these findings were not seen at surgery. For the 102 patients, ic = 0.806 reflecting strong agreement between the stereotactic needle core and surgical excisional Volume
176
#{149} Number
3
ing
results
brocystic breast
(Table).
There were 16 cancers ductal, two intraductal,
mucinous
carcinoma)
(13 infiltratand one
with
agree-
ment between findings from the needle and surgical biopsy specimens in 14 cases (Fig 3). The gun biopsy core specimen did not include a small focus of intraductal carcinoma found in a subsequent permanent surgical specimen. This was probably due to improper calibration of the stereotactic machine. In the other case of cancer without correlation, the stereotactic gun biopsy yielded a diagnosis of mucinous carcinoma that was missed at the ensuing surgical excisional biopsy.
The
presence
of cancer
was
bat-
en substantiated by means of histologic examination of the surgical specimen after modified radical mastectomy. In another case, the gun biopsy core did not include a focus of lobular carcinoma in situ found in the surgical specimen that was obtamed in a region remote from the area of fat necrosis that was causing the mammognaphic abnormality. The fat necrosis was successfully diagnosed from both the gun core and the frozen surgical material.
change tissue.
within
normal
There were 1 1 other diagnoses in the series including six cases in which only normal breast tissue was found in both the stereotactic and surgical specimens. There were two cases of intramammary lymph nodes with good correlation of results between biopsy techniques. In fact, in one of the two cases, results correlated so well that the pathologist believed he could identify the defect from the gun cone on the surgical specimen (Fig 6). In one case with a radial scan and one with an area of fat necrosis,
results
Finally, a cicatnix tactic biopsy was Subjectively,
also
correlated
well.
diagnosed at stereomissed at surgery. it was
noted
that
breast tissue (unlike prostate, liver, or kidney) was frequently fatty and friable and tended to fragment and disintegrate when placed in formalin. In consultation gists, it became
clean
with our pathobothat the histo-
logic with
correlations were not perfect the 18- or even 16-gauge nee-
dles,
especially
in cases
of fibrocystic
change. It was decided to use larger-gauge needle. Therefore, 74 patients underwent biopsy Radiology
a still after with #{149} 743
b.
a. Figure
4. Fibroadenoma missed at surgical biopsy. (a) Stereo views demonstrating over cluster of microcalcifications (arrows). (b) Gun core sample showing
poised
tissue (straight arrow) and Radiograph of surgical cifications. At histopathobogic
microcalcification specimen showing examination,
(c)
sue without
evidence
of the sclerotic
(curved arrow) of a sclerotic end of localization hook-wire the specimen contained only
needle the fibrous
fibroadenoma. but no microcalnormal breast tis-
fibroadenoma.
the 18- and 16-gauge needles, the remaining 29 underwent biopsy with 14-gauge needles. With the 14-gauge needle there was only one case of fi-
bleeding occurred, the 14-gauge needle.
brocystic of results.
tis occurred, and all were successfulby treated with antibiotics. It is not possible to determine with certainty
change without correlation Evaluation of the 14-gauge
needle vealed more
cores by our pathologists respecimens that were much consistent and uniform than
those
obtained
with
the
gauge needles. The one documented broadenoma missed
was controlled pressure. Three
whether from
case of a fiat gun biopsy re-
sulted from breast movement with use of the Senographe Stereotix unit. No unwanted patient on gun/needle
movement was observed with use of the Mammotest unit. In the two cases of documented lesions (one fibroadenoma and one cancer) that were
tion
of surgical
and
we have
As
mammography
No
ening
None
of the
103
patients
suffered
immediate significant complications. Two vasovagal reactions occurred during use of the Senographe Stereotix; one caused patient and breast movement resulting in a gun biopsy miss.
No
reactions
Mammotest 744
Radiology
#{149}
unit.
occurred
No significant
with
the
biopsy, in nearana-
more
commonplace lesions
are
for an increasing
the
to detect
would
screening
compound
method ogists noses
form opsies,
more the
number
of result bow-
for breast biopsy cancers (9), and this the increase in biCurrently, the
opsy performance. most common means is surgical excision.
sive, able,
and
will most likely some advocate
threshold earlier
be-
detected,
time-consuming, and occasionally
of breast This is an
biopsy expen-
uncomfortunreliable
of diagnosis (8,10). As radiolwhose robe it is to make diagand, more specifically, to per-
image-guided it should
percutaneous be natural
that
this
biwe
adapt our techniques to the diagnosis of breast disease. Most investigators have focused on FNAB in an attempt
goal
(4-6,11-13).
mammognaphic the theoretical
cy of skinny
case
DISCUSSION
need
en-
to accomplish
Steneotactic has improved
needle
guidance accura-
placement,
but
FNAB results in a substantial number of cases in which insufficient tissue obtained (6%-47%) and in a significant number of false-negative findings
breast biopsies (8). In addition,
were
needle
complication biopsies of other
solved by checking the calibration with a phantom prior to each biopsy. problems
excisional
of an infectious ly 1,000 gun tomic sites.
suspicious
with the Senognaphe SterThe time required for the has decreased with expenithat the average time for (excluding placement hook-wire) is now ap20-30 minutes.
resulted
biopsy,
yet to log a single
comes
calibration
cases
core
C.
or surgical biopsy. Howis a known complica-
missed at gun biopsy, calibration problems occurred with the Mammotest unit. These problems were
countered eotix unit. procedure ence such the procedure of localizing proximately
of
with manual of local celluli-
three
needle
localization, ever, cellulitis
smaller-
easily cases
these the
even with use Minor oozing
(1%-31%)
(14).
In addition
with
FNAB, the occasional false-positive results (1% or less), the poor ability make definitive benign diagnoses, and the inability to distinguish between in situ and invasive carcinoma have made surgeons and others reluctant
to rely
on
is
to
percutaneous
breast biopsy findings to make definitive treatment decisions (14-16). Compounding the above problems is the fact that cytopathobogy is a difficult art and science that is not available in its highest form in many institutions.
Since
many
believe
that
aspiration biopsy will not achieve a high degree of success without the constant availability of a highly skilled
and
trained
cytopathobogist,
attempts to perform breast FNAB at many institutions are likely to be unsuccessful (4,14,17). Many pathologists and surgeons believe cone tissue is superior to matenial obtained from needle aspiraSeptember
1990
that
the
14-gauge
sistently
a. Figure
b. 5.
Histopathobogic
ent elements.
gun
core
specimen
of fibrocystic
changes
showing
several
differ-
(a) Gun
core specimen showing multiple ectatic ducts (straight arrows) brosis (curved arrow). (b) Another portion of the same core shows ductal hyperplasia straight arrows) and papilbomatosis (open curved arrow). These findings are difficult, impossible,
to
detect
at aspiration
and fi(open if not
biopsy.
needles
provided
the
most highest
conquali-
ty, intact cores. We also thought that the larger cores would further decrease the chances of sampling error. We continued to use the Biopty gun rather than the conventional 14gauge needle because of the ease of use of the gun, the decreased patient discomfort and increased specimen quality and integrity due to the splitsecond sampling, and the ability of the gun to pierce both hand and mobile breast lesions before they have a chance to move out of the path of the needle. Despite the use of steneotactic guidance, with its theoretical accuracy of 1 mm in needle core gun biopsies,
there
were
still
nine
cases
in
which failed
a. Figure 6. (a) Surgical of resultant
b. Comparison
of histopathobogic
specimen showing gun core specimen.
specimens
gun core defect
tion biopsy, since it provides histologic information permitting definitive benign diagnoses and more complete characterization of mabignant lesions (1,2,15,16). When a definitive benign diagnosis (eg, fibroadenoma) can be made, open biopsy can be confidently avoided. The rebiabbe, complete, preoperative characterization of the type of malignant lesion present allows for the appropniate tailoring of the patient’s treatment without repetition of biopsy when and if surgery is performed. One of the reasons that radiologists might not have routinely used largegauge core needles in breast biopsy in the past is the theoretic increase in morbidity that is seen when largegauge needles are used elsewhere in the body (1,2). In adapting FNAB techniques to breast biopsy, investigators naturally used the needles and the techniques that they had honed during their experience in biopsy of other areas of the body, apparently without considering the fact that the breast is quite forgiving and virtually devoid of structures that could result Volume
the stereotactic gun biopsy to yield the diagnosis that was successfully made at surgical biopsy. However, we believe there were several possible explanations for these misses. First, our definite inexpenience with two different stereotactic machines used during our study had to be overcome. In addition, our pathobogists had to gain experience and confidence in handling and interpreting the cones. Second, because the 18-gauge needles used in the first 64 patients did not always obtain full
176
#{149} Number
3
of an
intramammary
in the center
lymph
of the lymph
node.
intact
node.
(b) Portion
cores,
lesions
in significant morbidity if tnansgressed by a core biopsy needle. Also, there have been conflicting reports regarding the relative efficacy of aspiration needles versus core biopsy needles in the breast (18-21). The studies purporting superior resuits with FNAB, however, compared multiple skinny needle passes to one core needle pass and were performed without an automated, rapid-fire cone
conjunction with it did not reliably tissue (22). Third,
biopsy
nomas
instrument
or pinpoint
imag-
ing guidance (neither stereotactic ultrasound [US] guidance). They failed to address the poor ability FNAB
results
to enable
nor also of
a definitive
benign diagnosis, an important consideration in breast biopsy as the majonity of lesions sampled are indeed benign. Without definitive benign diagnoses, it is impossible to differentiate between a potential false-negative result due to a sampling error and a true-negative result such as fibnoadenoma, sclerosing adenosis, or fibrocystic
Because
change.
the
breast
sists
of extraordinarily
fatty
tissue,
our
frequently friable
pathologists
conand
found
may
not
have
been as thoroughly canvassed as we believed from the stereo images. Also, the short-throw gun was used on two occasions until we realized, in another study, that obtain high-quality we experienced
significant calibration problems on at least two separate occasions. Fourth, patient movement causing subsequent breast movement after the mitial
stereo
calculations
were
made
ne-
suited in inaccurate coordinates in two cases. Fifth, we realized that at least one and possibly two fibroadewere
missed
because
of lesion
movement within the breast. This seems to be a frequent characteristic of fibroadenomas, confirmed under direct visualization during our USguided biopsies, since they do not appear to be firmly anchored in the breast. Therefore, as a needle is manually advanced toward the lesion, it can rotate or slip out of the path of the needle. Presently, if we suspect that a lesion is moving away from our needle tip, we back the tip up as much as 1.5 cm from the lesion. Then, when the gun is fired, the split-second the needle sion before
rapid-fire projection of allows it to pierce the leit can slip out of the way.
Radiology
#{149} 745
Finally, ity
it is apparent
plays
a role
that
in breast
serendipdiagnosis.
In
other words, there are occasions when the surgeons unwittingly make a diagnosis with their generous sungical sample that may have nothing whatsoever to do with the mammographic finding. We believe the sungical diagnosis of lobular carcinoma in situ in our series was just such an example; the mammognaphic finding in that case was fat necrosis. There is nothing that can be done about lesions missed at stereotactic gun biopsy that surgical
were
serendipitously
biopsy;
however,
seen the
nique is actually a further extension of the “minimal volume excision” approach advocated by Gallagher et al (27); however, it is much more expeditious and eliminates surgery entinely for benign lesions. Thus, the potential for psychological stress and physical trauma is lessened considerably.
A beneficial by-product of needle cone breast biopsy is the nearly immediate feedback the radiologist neceives regarding the actual histopathologic makeup of the mammographic finding will naturally
at
other
in question. improve future
This mam-
causative factors in our misses can be addressed and if these factors are eliminated, the success rate should improve further. In determining whether stereotactic gun biopsies might be performed as an alternative to surgical excisional biopsies, one should also take a close look at surgical biopsies, which have been referred to in some instances as “ritualistic rather than realistic” (15). Although surgery is considered a standard of reference, it should not be mistaken for perfection. On the contrary, needle localization breast biopsy is a semiblind operation with a technical difficulty that should not be underestimated even by the most experienced surgeons (23). The lesion miss rate is as high as one in five procedures with local anesthesia (10). Considerable bleeding can occur during a surgical biopsy that can obscure the operative field and preclude confident excision of deep lesions (23). Also, if the wine inadventantly dislodges, migrates, on is transected, the surgeon can become disoriented and excise the wrong tissue (24,25). Finally, lack of communication between the radiologist who has placed the localizing wire and the surgeon can result in a surgical
mographers’ interpretation abilities, and a less experienced mammographer will become highly competent much more quickly. As noted by Hall, “no book or lecture can substi-
miss (26). Surgical time-consuming,
vorable cost-benefit analysis that occurs when benign to malignant ratios rise with the inclusion of such be-
biopsies potentially
are
also nerve-
racking, and require a great deal of resources (8). The delays frequently encountered impose a psychological stress on the patient awaiting the procedure. Combining the time in the radiology and surgery departments,
the
needle
localization
surgical
tissue
with
only
746
#{149} Radiology
(8). and
surgery
may
have
been
performed
in
the process of the biopsy. Now such patients might still go straight to sungery; however, this sequence may change as the trend toward more conservative
ment tinues.
and
of primary Acquisition
nonsungical
breast
treat-
cancer
con-
of a definitive
histologic needle core biopsy specimen may be desirable prior to consideration of any type of treatment, surgical on otherwise. Currently there is some debate about the advisability of short-interval follow-up of less suggestive or “probably benign” lesions versus recommendation of immediate surgical biopsy (28). The rationale for short-interval
for
follow-up
surgical
biopsy.
is the
With
approach
a
to the
definitive
tion
suits,
and
and
the
amount
of sun-
References 1.
Charboneau
JW,
Reading
CC,
CT and sonographically opsy: current techniques tions. 2.
AJR
1990;
1989; Parker
rected
TJ.
154:1-10.
153:929-935. SH, Hopper
KD.
percutanelesions. AIR
Yakes
WF,
JL, Carter
TE. biopsies
percutaneous
sy gun.
Welch
guided needle biand new innova-
Gazelle GS, Haaga JR. Guided ous biopsy of intraabdominal
MD, Ownbey Radiology
1989;
Gibson
Image-diwith a biop-
171:663-669.
4.
Dowlatshahi K, Gent HJ, Schmidt R, Jokich PM, Bibbo M, Sprenger E. Nonpalpable breast tumors: diagnosis with stereotaxic localization and fine-needle aspi-
5.
Ciatto
ration.
6.
Radiology
5, Del
7.
lesion:
breast
Hall tial
core
Med
F.
1986;
Radiology C.
in 2,594
Stereotaxic
mammogralesions.
SH, Jobe WE, Luethke Stereotactic percutanebiopsy:
technical
experience.
Screening
problems
Nonfine-
non-palpable
tion and initial 1990; 3:135-143. 8.
P.
1:1033-1036.
Lovin JD, Parker JM, Hopper KD. ous
G, Auer
biopsy 1989;
Bravetti stereotaxic
cytology.
detected
Lancet
170:427-433.
MR.
needle aspiration 1989; 173:57-59. Azavedo E, Svane phically
bi-
1989;
Turco
breast
fine-needle
whether or not it is performed with administration of general anesthesia and whether it is performed on an in- or outpatient basis. The cost of a stereotactic gun biopsy is approxi-
utiliza-
screening
reduce
palpable
on
increasing
required for therapy of the cancers diagnosed (29). Stereotactic mammography allows pinpoint needle placement, and the large cones obtained from 14-gauge needles furthen decrease the possibility of sampling error. The large cores also ebimmate the need for an experienced cytopathobogist and allow for histologic evaluation permitting definitive benign diagnoses and complete charactenization of malignant lesions. Utilization of a mechanized biopsy device eliminates the need for mastery of difficult aspiration techniques with the push of a button. It is therefore our opinion that steneotactic breast gun biopsy overcomes many of the problems associated with FNAB and may prove to be a cost-effective, expedient, and dependable alternative to surgical biopsy. U
histo-
opsy in Denver varies between $1,217 and $2,673 depending
With
gery
the
pathologic diagnosis of mammographic lesions could be forthcoming. The cost of a surgical excisional
$675.
of mammographic
the resultant increase in performance of breast biopsy, the national cost of surgical biopsy could rise considerably. It is therefore important that cost-effective means of breast cancer diagnosis are identified and utilized. Stereotactic gun biopsy of the breast may help fill this need. In conclusion, to justify a percutaneous breast biopsy triage system for suggestive nonpalpabie mammographic lesions, the system must be extremely accurate, reduce on eliminate surgical biopsy with benign re-
3.
unfa-
lower cost, discomfort, and psychological trauma associated with stereotactic gun biopsies, a more aggressive
and
skin nick, leaving no scar. It can be performed quickly and easily after the original mammogram is obtamed, even the same day. This tech-
of teaching” puzzling
sometimes highly suggestive mammographic pseudolesions that can appear after surgical biopsy should no longer be a dilemma (19). Previously, in patients with highly suggestive lesions or obvious mammognaphic cancers, the definitive
sions
surgical excisional biopsy may requine up to 3 hours. Postoperative hematoma, cellulitis, and poor cosmesis can result. By contrast, a stereotactic gun biopsy obtains
tute for this type Also, the frequently
mately
adapta-
Breast
Dis
mammography: on
the
horizon.
potenN Engl
314:53-55.
September
1990
9.
10.
1 1.
Choucair
RJ, Holcomb
Hughes
TG.
breast 456.
lesions.
Norton
LW,
ization
breast
17.
Am J Sung 1988; 156:453Pearlman
NW.
biopsy:
aspiration
sis of primary 103:178-183.
Kahky
Needle
accuracy
cytology breast
localcost.
versus
Needle
breast
JM.
evaluation
Surgery
tion sis
1987;
biopsy
of 20.
BS, Wang
by means
cytology.
DL, Cruz AB,
Am J Sung 1988;
Nonpalpable
HH,
breast
of fine-needle
Radiology
Fein
V,
JK. Mayo
15.
21.
Clin
Proc
1986;
Abramson
DJ.
Milit
Med
1987;
176
#{149} Number
Volume
breast
152:594-595.
3
Vorherr
H.
tissue
Shabot
tablishing picious
22.
biopsy. 23.
fine
needle
pre-operative
of the
biopsy
26. with
in the diagnosis in breast
YH.
IM, needle
diagnosis masses.
Schick
Aspiration
to Tru-Cut
the
27.
28.
P. Nie-
cytology biopsy Surg
is sus-
1982;
196:122-126. Hopper KD, Baird DE, Reddy W, et al. Efficacy of automated biopsy guns versus conventional biopsy needles in the pygmy pig. Radiology 1990; 176:671-676. Norton LW, Zeligman BE, Pearlman NW. Accuracy and cost of needle localization breast biopsy. Arch Sung 1988; 123:947950.
of the
1987;
localiza-
150:787-788.
breast.
Sung
Gynecol
164:399-403.
Gallagher
WJ,
Cardenosa
McCarthy
KA,
Kopans
29.
AJR
G,
DB.
of nonpalpable 1989;
of the Invest
Rubens
interpreted
RaJR.
Minimal-vol-
breast
le-
153:957-961.
Brenner RJ, Sickles EA. Acceptability periodic follow-up as an alternative opsy for mammographicably detected sions
in es-
of clinically Ann
lesions
ume excision
1986; 22:1045-1052.
Goldberg
1988;
March
Yankaskas BC, Knelson MH, Abernathy ML, Cuttino JT, Clark RL. Needle local-
sions.
Eur
AJR
SA,
biopsy
ization biopsy of occult lesions breast: experience in 199 cases. diol 1988; 23:730-733.
of
cancer.
RJ, Feig
of breast
J, Gunderson SB, Gunderson AL, Cogbill TH, Travelli R, Strutt P. Needie localization and biopsy of nonpalpaObstet
aspiration
metastases
breast
25.
tion wire. Landercasper
ble
Histopa-
Clin Oncol
PS, Wechsler
Migration
diagno-
breast.
biopsy
Davis
DE.
2:239-254.
Breast
MM,
24.
aspira-
multihole needles for histologic and cytologic examination. Am J Obstet Gynecol 1985; 151:70-76. Pederson L, Guldhammer B, Kamby C, Aasted M, Rose C. Fine needle aspiration
superior
61:377-381.
Outpatient
1978;
berg R, Pilch
475. 16.
of carcinoma
J Cancer
GS, Goellner JR, Welch JS, Martin Fine-needle aspiration of the breast.
and
in the
thology
soft
aspi-
1989; 171:373-
Sheikh FA, Tinkoff GH, Kline TS, Neal HS. Final diagnosis by fine-needle aspiration biopsy for definitive operation in breast cancer. Am J Sung 1987; 154:470-
needle
cytology
and Tnt-Cut
lesions:
376.
Grant
Sundaram M, Wolverson MK, Heiberg E, Pilla T, Vas WG, Shields JB. Utility of CTguided abdominal aspiration procedures. AJR 1982; 139:1111-1115. Elston CW, Cotton RE, Davies CJ, Blamey RW. A comparison of the use of the “Tnt-Cut”
Finein the diagno-
aspiration
masses.
156:450-452. Hann L, Ducatman
Mclntire
14.
18.
19.
HV.
palpable
ration
cancer.
MP, Rone R, Duncan
Gaskill
13.
R,
of nonpalpable
Am J Sung 1988; 156(2):13B-l5B. Smith C, Butler J, Cobb C, State D.
needle
12.
MB, Mathews
Biopsy
as probably
of to bibe-
benign.
Ra-
diology 1989; 171:645-646. Kopans DB. Fine-needle aspiration of clinically occult breast lesions. Radiology 1989; 170:313-314.
Radiology
#{149} 747