Transfemoral Balloon Aortic Occlusion during Open Cardiopulmonary Resuscitation Improves Myocardial and Cerebral Blood Flow’ PAUL A. SPENCE, M.D., FRCS(C), HIROSHI Departments

IIDA, M.D.,

ROBERT M. LUST, PH.D., W. RANDOLPH CHITWOOD, JR., M.D., You Su SUN, M.D., AND ERLE H. AUSTIN III, M.D., FACS

of Cardiac Surgery and Physiology, East Carolina University

School of Medicine,

FACS,

Greenville, North Carolina 27858-4354

Submitted for publication November 20, 1989

These experiments were designed to determine whether the limited cardiac output during open cardiac massage could be preferentially directed to the coronary and cerebral vessels by balloon occlusion of the descending thoracic aorta. Sixteen dogs were instrumented to monitor cardiac output and left atrial, right atrial, right ventricular, left ventricular, and arterial blood pressures. Measurements of myocardial and cerebral blood flow distribution during massage were made using the radioactive microsphere technique. Each animal underwent two episodes of fibrillation and resuscitation. In one episode the arrest was managed by open massage alone, and in the other, open massage was accompanied by balloon occlusion, with the order randomized. When compared to control, open cardiac massage was associated with a significant decrease in mean arterial pressure; however, the addition of balloon occlusion produced a 130% increase in the mean arterial pressure that was obtained during open CPR (control, 93 + 5 mm Hg; massage alone, 35 + 2 mm Hg; massage + balloon, 76 + 2 mm Hg, P < 0.01). In a similar fashion, although the absolute blood flow was reduced by 50% when compared to control, the blood flow (ml/min/g) to the brain and heart during massage was 100% better when balloon occlusion was employed (brain: control, 0.41+ 0.03; massage only, 0.05 f 0.01; massage + balloon, 0.25 k 0.02, P < 0.01; heart: control, 1.46 k 0.11; massage alone, 0.35 + 0.05; massage + balloon, 0.71 + 0.05, P < 0.01). These results suggest that aortic occlusion significantly increased myocardial and cerebral perfusion patterns during ventricular fibrillation and open cardiac massage. Percutaneous transfemoral balloon aortic occlusion during CPR may be a useful adjunct to standard therapy. o 1990 Academic

Press.

Inc.

INTRODUCTION

The outcome of patients who undergo cardiopulmonary resuscitation (CPR) depends largely on the successful ’ Presented at the Annual Meeting of the Association for Academic Surgery, Louisville, KY, November 15-l&1989.

restoration of cardiac and cerebral function [ 11. Flow to both of these organs is markedly reduced during resuscitation efforts [2-61. For many years surgeons have been trained to cross-clamp the aorta during open cardiac massage when an emergency thoracotomy is performed during cardiac arrest. Increasing experience with femoral cannulation and the widespread availability of cardiac catheterization, percutaneous balloon angioplasty, and intraaortic balloon counterpulsation suggests that transfemoral occlusion of the aorta, producing a hemodynamic equivalent to the open cross-clamp technique, might be feasible during CPR. These experiments were designed to determine whether the limited cardiac output during cardiopulmonary resuscitation could be preferentially directed to the coronary and cerebral vessels by balloon occlusion of the descending thoracic aorta just past the origin of the subclavian artery. METHODS

Sixteen anesthetized (pentobarbital, 30 mg/kg iv bolus, supplemented pm) adult mongrel dogs, weighing between 19 and 24 kg, were intubated and ventilated with room air. Adjustments to the ventilator were made to maintain arterial POz above 90 mm Hg, arterial oxygen saturation above 93%) and arterial PC02 between 38 and 42 mm Hg. Blood gases were sampled routinely and bicarbonate was administered as needed to maintain arterial pH between 7.37 and 7.43. A left thoracotomy was performed and both internal mammary arteries were cannulated: one with a fluid-filled catheter for arterial blood sample withdrawal, and the other with a Millar catheter-tipped pressure transducer (Millar Instruments, Houston, Texas) to monitor arterial blood pressure. The pericardium was widely excised. The pulmonary artery was carefully dissected free from any connective tissue attachments and encircled with an electromagnetic flow probe (Zepeda Instruments, Seattle, WA) which was used to monitor cardiac output. Purse-string sutures were placed in the right atria1 appendage, and a fluid-filled catheter was inserted to measure right atria1 pressure. A second Millar catheter was inserted, advanced into the right ventricle and used

217 All

oozz-4804/90 $1.50 Copyright 0 1990 by Academic Press, Inc. rights of reproduction in any form reserved.

218

JOURNAL

OF SURGICAL

RESEARCH:

to monitor right ventricular pressure. In a similar fashion purse-string sutures were also placed in the left atria1 appendage. Through one set a fluid-filled catheter was placed and used to monitor left atria1 pressure and for injection of radioactive microspheres. Through the other pursestring a third Millar catheter was inserted, advanced into the left ventricle, and used to monitor left ventricular pressure. All catheters, probes, and transducers were connected to a polygraph (Hewlett-Packard Instruments, Los Angeles, CA) and monitored continuously for the duration of the experiment. A femoral cut down was performed, and an intraaortic balloon catheter (Datascope, Paramus, New Jersey) was advanced to the level of the origin of the subclavian artery. The position of the balloon was confirmed manually through the thoracotomy. The balloon was left deflated at this time, and the presence of the balloon did not significantly influence resting hemodynamics. Baseline measurements of cardiac output and right atrial, left atrial, arterial, right ventricular, and left ventricular pressures were obtained. The first microsphere injection, used to establish control perfusion patterns, was made at this time. The hearts were then fibrillated with a short burst of altering current. Open chest manual cardiac resuscitation was then performed at a rate of 100 compressions per minute. Ninety seconds after the onset of fibrillation and the initiation of cardiac massage, samples of all hemodynamic measurements were obtained, and the second microsphere injection was made. After 3 min of fibrillation and open cardiac massage the hearts were defibrillated and the animals were allowed to recover for 30 min. Due to the short period of CPR in these experiments, no systemic heparinization was used. In all cases, return to baseline hemodynamic parameters for a period of not less than 15 min was confirmed. This included the stabilization of all arterial blood gas parameters. The hearts were then fibrillated with an alternating current for a second time and the episode was repeated. A third injection of microspheres was made and data sampling again occurred 90 set after the initiation of open cardiopulmonary resuscitation. At the end of 3 min, the animals were again successfully defibrillated and allowed to return to stable baseline values. In one of the two episodes, the intraaortic balloon was inflated, but the episode in which the balloon was inflated was selected at random. At the end of the experiment, the animals were euthanized by lethal overdose. The heart and brain were removed and placed in fixative for later sectioning. Blood flow distribution was determined using lo-pm radioactive microspheres (New England Nuclear, Boston, MA) according to the reference withdrawal method of Rudolph and Heymann [7]. The injection techniques as well as the withdrawal and tissue sectioning procedures have been detailed previously [8]. Briefly, reference arterial withdrawal at a known, fixed rate was begun 15 set before microsphere injection and continued for at least 2 min after the injection. One milliliter (approximately 3 X lo6

VOL. 49, NO. 3, SEPTEMBER

1990

spheres) of agitated ultrasonic solution labeled with scandium-46, niobium-95, or ruthenium-103 were selected at random and injected into the left atrium. Streaming or nonuniform distribution of the microspheres in these experiments was checked by comparing flows to the right and the left cortex. Lateral differences in cortical flows of more than 15% were criteria for disregarding ail data obtained from that injection. In the present study no data were excluded on this criteria. Samples of the left ventricle, right ventricle, interventricular septum, left and right cerebral cortex, cerebellum, and medullary brain stem were weighed and analyzed for gamma activity (Packard Instrument Co., Sterling, VA). By computer analysis, counts were then translated to flow values and normalized for tissue weight to yield blood flow determinations in milliliters/minute/gram. Analysis of variance for repeated measures with appropriate post-hoc tests were used to determine statistical significance between all blood flow and hemodynamic determinations. Significance in all cases were determined at the P < 0.01 level. In conducting these experiments we adhered strictly to the Guiding Principles for the Care and Use of Laboratory Animals developed by the American Physiological Society. RESULTS

Cardiac output fell during open cardiac massage from a prefibrillation value of 2.28 f .08 liters/min to 1.10 + .09 and 1.19 * .08 in the massage alone and massage + balloon groups, respectively. While the decrease in both CPR groups was significantly different from control, neither of the CPR groups was different from each other. Similarly, increases in right and left atria1 pressures from control values of 4 and 5 mm Hg, respectively, to 10 and 11 mm Hg were observed during CPR, but again no significant differences were detected between the CPR groups. Mean arterial pressure fell immediately during fibrillation and could be maintained at only 38% of control values by open cardiac massage alone. Addition of the balloon occlusion to the massage procedures produced an increase in mean arterial pressure of 114% above open massage alone and restored mean arterial pressure to 81% of control values (Fig. 1). Figure 2 demonstrates that a significant gradient in peak systolic pressure developed between the left ventricle and the aorta during open cardiac massage alone, suggesting a significant loss in the transfer of developed pressure. In contrast, peak arterial pressures were not significantly different from peak left ventricular pressures when open massage was accompanied by balloon occlusion. There was better maintenance of peripheral resistance with balloon occlusion, as evidenced by the increased diastolic arterial pressure that was maintained in this group (Fig. 3). A summary of the differences in pressure patterns observed in the two CPR groups can be seen from the sample tracings shown in Fig. 4. The coronary perfusion gradient, calculated as the difference between arterial diastolic pressure and left ventricular end-dia-

SPENCE

n q Control q CPR + balloon l

H 70E 60.E. $3 502 $ 40a 30-

* pc.01 “.s control * pg.01 vs no balloon

ET AL.: BALLOON

OCCLUSION

DURING

219

CPR

q

Control @ CPR + balloon

CPR - no balloon ** * T g

60

L

50

If

40

a'u

30

n

CPR no balloon

** I

* pc.01 vs control ** pc.01 vs no balloon

20

20 10

10

0

0

Ao

FIG. 1. Changes in mean arterial pressure with fibrillation and open cardiac massage, with or without accompanying balloon occlusion.

stolic pressure, decreased to 15% of control during fibrillation and massage. Addition of balloon occlusion to the open cardiac massage increased the coronary perfusion gradient more than 300% from massage alone and restored the coronary perfusion gradient to approximately 70% of control (Fig. 5). Calculated blood flow to the heart and brain averaged between 12 and 20% of control values during open CPR alone, but were in excess of 50% of control values when open CPR was accompanied by balloon occlusion (Figs. 6 and 7). DISCUSSION

The concept of aortic occlusion during open cardiac resuscitation is not new. For many years, surgeons have been trained to clamp the aorta when open cardiac massage was performed in the setting of an emergency thoracotomy, although we could find no experimental literature to explain this practice. Preliminary experiments using manual clamping of the descending aorta had suggested that such an occlusion would produce beneficial results. Employing the intravascular balloon was decided upon, in part, as a prelude to an eventual closed-chest

fii! Control n CPR - no balloon q CPR + balloon

140 120 H E a, 2 %

* p

Transfemoral balloon aortic occlusion during open cardiopulmonary resuscitation improves myocardial and cerebral blood flow.

These experiments were designed to determine whether the limited cardiac output during open cardiac massage could be preferentially directed to the co...