Advances in Heart Failure Valvular Heart Disease in Patients Supported With Left Ventricular Assist Devices Teresa S. Wang, BS; Adrian F. Hernandez, MD, MHS; G. Michael Felker, MD, MHS; Carmelo A. Milano, MD; Joseph G. Rogers, MD; Chetan B. Patel, MD

H

eart failure (HF) has become increasingly prevalent with the progressive development of HF risk factors in our aging population.1 Although valvular heart disease (VHD) is not similarly considered as a public health concern (overall United States prevalence of 2.5%),2 many patients with advanced systolic HF have concurrent VHD. Because of the limited number of suitable donor organs for cardiac transplantation, advanced HF patients who have refractory or ­end-stage HF are often treated with left ventricular assist devices (LVADs).3 The impact of concomitant valvular lesions in those undergoing LVAD implantation has not been fully characterized and the management of these lesions is a subject of intense interest, particularly for patients for whom subsequent heart transplantation is not a viable option. First-generation LVADs exhibited a pulsatile-flow mechanism, using positive displacement to move blood to the periphery.4 These regular intervals help mimic the ­systole-diastole phases of native cardiac physiology, which differs substantially from the mechanism of newer continuous-flow devices in which a nonpulsatile stream of blood provides systemic circulation. Second- and third-generation continuous-flow LVADs provide smaller size and enhanced durability, allowing for long-term support with improved quality of life.5 However, there are unique implications of contemporary LVAD support on cardiac circulation and valvular mechanics. Chronic unloading of the left ventricle (LV) via continuous, nonpulsatile flow alters blood flow dynamics through the left-sided cardiac chambers, with implications for the aortic valve (AV). Alteration of cardiac mechanics through continuous-flow can also impact the unsupported right ventricle (RV), leading to tricuspid regurgitation (TR). In this review, we aim to provide a background on the most common aortic, tricuspid, and mitral valve pathologies in patients supported with LVADs and highlight the outcomes and management of patients with VHD who undergo LVAD insertion.

AV Pathology AV pathology is uncommon in advanced HF patients before LVAD implantation.6 Pal et al7 report in a single-center study (n=281) that only 4% of patients with advanced HF had moderate-to-severe aortic insufficiency (AI) at the

time of referral for heart transplantation or LVAD. However, many studies have described progression of AI after ­continuous-flow LVAD support.8,9

Aortic Blood Flow Dynamics Continuous-flow LVADs draw blood from the LV and direct flow into the ascending aorta, thus increasing cardiac output and aortic root pressure while simultaneously decreasing LV pressure (Figure 1). This creates a continuous transvalvular pressure (aortic pressure–LV pressure) gradient across the AV (Figure 2).6 The AV opens when transvalvular pressure is less than zero and remains closed when greater than zero. Because many LVAD devices are set to near full support where the LV is fully decompressed, the transvalvular pressure is greater than zero, and the AV remains closed throughout the cardiac cycle (ie, LVAD provides flow in series). In contrast, when the transvalvular pressure is periodically allowed to decrease, blood is ejected through the AV and the heart operates in parallel with the LVAD, with circulation supported via both native LV function and the LVAD (partial support).6

LVAD-Induced Commissural Fusion Based on the physiology described above, continuous-flow LVADs may lead to fusion or deterioration of the AV (Figure 1). Fusion develops when fibrous tissue deposits along the commissures of the leaflets, resulting in leaflet adherence that reduces the opening area of the valve.6 The closed state maximally stretches the leaflets, preventing nutrient flow and stimulating collagen production and remodeling.6,10 Closed valves can also degenerate from disuse, especially in the setting of trauma from high-velocity supravalvular blood flow.8 Finally, a closed AV coupled with limited antegrade flow allows for stasis of blood on the ventricular surface of the valve favoring thrombus formation and organization, which further contributes to leaflet fusion.6,8 This leads to retraction of the leaflet tips and generation of a central orifice that becomes fixed in the absence of intermittent AV opening. The transvalvular pressure generated from LVAD support can then result in regurgitant flow from the aorta to the LV through the central orifice and development of AI.11

From the School of Medicine (T.S.W.), Division of Cardiology (A.F.H., G.M.F., J.G.R., C.B.P.), and Division of Cardiothoracic Surgery (C.A.M.), Duke University Medical Center, Durham, NC. Correspondence to Chetan B. Patel, MD, Duke University Medical Center Box 3034, Durham, NC 27710. E-mail [email protected] (Circ Heart Fail. 2014;7:215-222.) © 2014 American Heart Association, Inc. Circ Heart Fail is available at http://circheartfailure.ahajournals.org

DOI: 10.1161/CIRCHEARTFAILURE.113.000473

Downloaded from http://circheartfailure.ahajournals.org/ 215at NORTH DAKOTA STATE UNIV on June 25, 2015

216  Circ Heart Fail  January 2014

Figure 1. Altered aortic blood flow dynamics. Left ventricular assist devices (LVADs) lead to 2 major alterations: (1) distorted stress sharing at the aortic root and (2) aortic valve (AV) closure secondary to increased transvalvular pressure. Closure of the AV contributes to thrombus formation, commissural fusion of the AV leaflets, and de novo aortic insufficiency (AI)/aortic stenosis (AS) or progression of preexisting AI.

Aortic Insufficiency Preexisting Aortic Insufficiency Initiation of LVAD support may exacerbate preexisting AI via the mechanisms highlighted above.9,12 An incompetent AV can lead to a closed loop with blood flowing from the LV through the LVAD into the aorta and back into the LV down a continuous pressure gradient. This recirculating blood volume may result in recurrence of HF symptoms because of increased LV pressure and volume as well as worsening AI.6 Importantly, although LVAD implantation is associated with an increased rate of development and progression of AI,8,9,13 no increase in mortality has been observed in patients who develop ­moderate-to-severe AI.13 Given the potential for AI to progress after LVAD insertion, many surgeons will correct moderate or greater degrees of AI at the time of implantation. Replacement with a bioprosthetic valve was previously the optimal method for AI correction but

has fallen out of favor given the need for greater duration of cardioplegic arrest, potential for thromboembolic complications,14 and reports of bioprosthetic AV fusion and stenosis.12,15 Additionally, concomitant aortic procedures are associated with increased operative risk and an ≈5-fold increase in 30-day mortality.7 Use of a mechanical prosthesis is not recommended because of presumed high risk of thrombus formation and the potential for subsequent embolization.16 LV outflow tract closure is a definitive strategy for treating preexisting AI while circumventing challenges associated with an AV prosthesis, and multiple methods have been described.14,17–19 The Park stitch is a simple coaptation stitch placed at the central portion of the 3 aortic cusps. Bryant et al14 sutured the commissures either centrally with a single buttressed stitch or along the full length of noncoaptation. Adamson et al18 describe an alternative method using felt strips along each leaflet that are anchored to the aortic wall. Suture repair of the AV is simple, requires a shorter period of cardioplegic arrest, and may be associated with reduced thrombosis and embolism compared with AV replacement. However, potential complications include AI recurrence, sutures tearing or eroding through the AV, and lack of ability to maintain hemodynamic stability with pump failure. Moreover, this method complicates the process of weaning from LVAD support should the LV recover function.20 De Novo Aortic Insufficiency Post-LVAD Implantation Multiple studies have found that de novo development of AI is common and can occur after a relatively short period of LVAD support.8 Rajagopal et al13 described that AI development was more pronounced in recipients of continuous-flow LVADs compared with pulsatile-flow LVADs but found no association between preoperative AI grade and development of moderate or greater postoperative AI. In other studies, progression of AI was correlated with female sex, smaller body surface area,

Figure 2. Creation of transvalvular pressure gradient with left ventricular assist device (LVAD) implantation. A normally functioning heart is able to create enough left ventricle (LV) pressure (120 mm Hg) to eject blood through the aortic valve (AV); transvalvular pressure is ≤0 mm Hg, and the AV can open (A). LVADs decompress the LV (5–10 mm Hg) while also increasing aortic root pressure to create transvalvular pressure >0 mm Hg, which closes the AV (flow in series, B).

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

Wang et al   Valvular Heart Disease and LVADs   217 HeartMate II support, increasing aortic sinus diameter, and an AV that remained predominantly closed.8 For patients who develop de novo AI, medical therapy includes vasodilators to lower systemic afterload and diuretics to lower ventricular preload. In cases where refractory HF symptoms or cardiogenic shock develop,14 inotropic support may be required until a surgical intervention can be implemented. There have been several reports of minimally invasive and percutaneous strategies for treatment of patients with previously implanted LVADs experiencing HF secondary to progressive AI.21–23 Santini et al21 described transcatheter aortic valve implantation with CoreValve (Medtronic, Milwaukee, WI) as a life-saving intervention in a patient who developed de novo AI after LVAD insertion. D’Ancona et al22 described successful transcatheter aortic valve implantation with a SAPIEN valve (Edwards Lifesciences, Irvine, CA) through a transapical approach in a patient who had an otherwise normal AV with absence of aortic calcifications. However, neither the SAPIEN nor CoreValve device has been Food and Drug Administration approved for this particular use. Parikh et al23 described percutaneous transcatheter AV closure in 5 patients using the Amplatzer device (AGA Medical, Plymouth, MN), which improved cardiac hemodynamics. We have yet to understand the long-term outcomes of these innovative techniques, particularly any susceptibility to development of AI similar to native valves. However, these reports provide promising evidence for novel techniques in the treatment of AI in patients with LVADs.

Aortic Stenosis Because the LVAD is not dependent on flow through the AV when functioning in series, preexisting aortic stenosis (AS) does not adversely affect the normal functioning of an LVAD.6,24 On the rare occasion that the LVAD could be explanted, AV replacement may be considered to permit successful weaning of LVAD support.6

Current Approach and Future Direction Recent recommendations from the International Society of Heart Lung Transplantation suggest that moderate or greater AI be considered for surgical correction at the time of LVAD implantation (Table 1).16 We favor suture closure of the AV leaflets as the treatment of choice given its low risk of complications and durable results.6,14,17 Transcatheter aortic valve implantation and percutaneous transcatheter AV closure could become reasonable options in the future with modifications in the design of the prosthesis or implantation technique. Preexisting AS is not a significant concern unless partial or parallel support is being implemented to maintain native pulsatility. Finally, given the mechanism of AV commissural fusion, we think that achieving a range of optimal pump speed that allows for adequate AV opening is another method of prevention against de novo pathology.

Tricuspid Valve Pathology Patients with advanced HF frequently progress to develop symptoms of biventricular failure with associated dilation of both the LV and RV. Dilation of the RV leads to secondary tricuspid annular dilation, apical leaflet displacement, and

restriction with resultant incomplete coaptation.25,26 Between 30% and 64% of patients with advanced HF will have associated TR,26 which further contributes to RV failure through the development of right-sided volume overload and reduced RV ejection. Because development of worsening RV failure after LVAD implantation is associated with reduced survival,26,27 tricuspid valve (TV) pathology is a major concern at the time of surgery.

Effect of LVAD Implantation on Tricuspid Regurgitation LVAD therapy favorably reduces LV end-diastolic pressure and pulmonary venous pressures.6 Reduced pulmonary pressures can decrease RV afterload and help normalize ­right-sided hemodynamics with resolution of TR. Furthermore, LVAD therapy significantly reduces central venous pressure and RV end-diastolic dimensions, thus improving RV function and contractility.28 However, in the immediate postoperative period, flow can be limited across the pulmonary vasculature, thus delaying reduction of RV afterload. Furthermore, patients often undergo volume resuscitation in the perioperative period, which exacerbates RV dilation and TR.25 Finally, unloading of the LV by the LVAD results in a leftward shift of the interventricular septum that can restrict the TV leaflets, thereby worsening TR26 (Figure 3). Holman et al29 found that TR acutely worsened, and that RV enlargement persisted 30 days postimplant, presumably because of the mechanisms noted above. In contrast, Piacentino et al26 found that at late follow-up (mean 156±272 days), TR decreased significantly (32% with moderate-severe TR versus 49% preoperatively), suggesting that RV remodeling is an incomplete and slow process.

Concomitant Tricuspid Valve Repair With LVAD Implantation Because RV failure can lead to prolonged hospitalizations and increased mortality, there is interest in concomitant tricuspid valve repair (TVR) during LVAD implantation in select patients.30 Potential benefits of TVR include reduced right atrial pressure and reduced sequelae of venous congestion, leading to improvement in renal and hepatic function.25 Downsides might include longer operative time, increased bleeding risks, and potential for development of tricuspid stenosis. Furthermore, there are no studies examining the impact of concomitant TVR on long-term outcomes, such as survival free of transplant, quality of life, and readmission rates.30 Several single-center studies have examined the short-term impact of this procedure for moderate-to-severe TR25,30–32; however, to our knowledge, there have been no such multicenter studies. Saeed et al30 reported on 42 patients with m ­ oderate-to-severe TR undergoing LVAD implantation, with 8 patients receiving concomitant TVR. TVR prolonged the operative time and showed similar outcomes to LVAD implantation alone, including incidence of RV failure, end-organ function, and survival. This absence of benefit led the authors to suggest that TVR may not be necessary at the time of LVAD implantation. Similarly, Krishan et al25 reported on 51 LVAD patients (37 concomitant TVR, 14 control) and found no difference in outcomes including postoperative hemodynamics, rates

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

218  Circ Heart Fail  January 2014 Table 1.  Options in the Management of Preoperative or Progressive Aortic Insufficiency in Patients With LVAD Support Management

Pros

Mechanical valve

-

Cons

Comments

High risk of thrombus formation Potential for subsequent embolization

Not recommended

Medical therapy (ie, vasodilators, diuretics)

Lower systemic afterload and ventricular preload

Provides temporary relief and symptom management

May develop medically refractory heart failure or cardiogenic shock, requiring surgical intervention

Bioprosthetic valve

Left ventricle can eject through artificial valve if device fails

Time-consuming operation Prone to thrombosis/embolism

Previously thought to be best method, has fallen out of favor

TAVI (not FDA approved for AI)

Immediate symptom relief Minimally invasive

Paravalvular regurgitation Lack of long-term data

Two case studies with good results Novel

Percutaneous transcatheter AV closure

Effective; reduces pulmonary capillary wedge pressure

Lack of long-term data

Few case studies Novel

 Felt strips anchored to aortic wall18

Safe, well tolerated

Device failure may lead to hemodynamic instability

Definitive strategy

 Suture closure of AV commissures14,17

Simple, quick ≥1 yr durability in case reports Lower risk of complications

Device failure may lead to hemodynamic instability

Quick technique with evidence of durability

LV outflow tract closure

AI indicates aortic insufficiency; AV, aortic valve; FDA, Food and Drug Administration; LV, left ventricle; LVAD, left ventricular assist device; and TAVI, transcatheter aortic valve implantation.

of bleeding, duration of mechanical ventilation and inotropic support, and hospital mortality. However, given the lack of obvious increase in perioperative risk, they advocated for a liberal approach, supporting the use of TVR at LVAD implantation. Piacentino et al31 performed 2 studies on concomitant TVR.31,32 The first involved 115 patients with ­moderate-to-severe TR, 34 of whom underwent concomitant TVR. A related report looked at 200 consecutive patients with continuous-flow LVAD, 61 of whom had ­moderate-to-severe TR (33 underwent concomitant TVR).32 Concomitant TVR reduced postprocedure RV failure and postoperative TR compared with LVAD alone. Additionally, there was a trend toward improved survival and early clinical outcomes, including decreased postoperative inotrope use and decreased incidence of renal dysfunction.

Current Approach and Future Direction These single-center studies are provocative but do not provide definitive guidelines about effectiveness of concomitant TVR. However, with some evidence showing improved early clinical outcomes with TV intervention, it is reasonable to suggest that concomitant TVR provides salient benefits in patients with moderate-to-severe TR receiving long-term LVAD support.16 Whether this intervention is repair versus replacement will also need to be evaluated given comparable outcomes in the Mayo experience.33 Alternatively, in the event that the LVAD is used as partial support, lower speeds could help prevent leftward shift of the interventricular septum, thus allowing closure of the tricuspid leaflets and obviating the need for repair. There remains a need for large prospective, randomized multicenter trials investigating clinical use and benefit of concomitant TVR. This would be a unique undertaking given the heterogeneity in TV pathology and degree of RV failure in this population. Standardized measures of TR severity and RV function (using core laboratories in a gatekeeper capacity)

would be required as entry criteria for the study. LVAD speed management as well as adjuvant therapies may have to be protocoled using echocardiographic and possibly hemodynamic guidelines (ie, degree of permissible pulmonary hypertension) to minimize variability in practice between centers. Furthermore, end points measured should extend beyond perioperative data and RV function and focus on index length of stay, readmission rates, functional status, and quality of life.

Pulmonic Valve Pathology Pulmonic Regurgitation and Stenosis There are no studies comprehensively investigating the incidence and clinical impact of pulmonic stenosis or regurgitation in patients supported with LVADs. In theory, either lesion might limit net flow across the pulmonary vasculature, thus delaying reduction of RV afterload and contributing to RV failure. This may be less clinically relevant given the ability of LVAD therapy to lower pulmonary pressures,34 thereby generating a pressure gradient favoring antegrade flow across the pulmonic valve. Therefore, pulmonic valve pathology may have minimal untoward consequences and is unlikely to be a frequent site of intervention. However, intrinsic pulmonary arterial hypertension, disease secondary to long-standing valvular disease, or previous chronic pulmonary emboli may not always respond to reduction of left-sided filling pressures with LVAD therapy.

Mitral Valve Pathology There is a paucity of information about mitral valve pathology in patients supported with LVADs, and surgical correction of mitral valve disease (particularly for mitral regurgitation, MR) has not been used frequently in patients with LVAD.35 Nonetheless, hemodynamically significant mitral valve disease is common among patients with advanced HF and remains the most prevalent VHD in the United States.36

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

Wang et al   Valvular Heart Disease and LVADs   219 risk. Further investigation is necessary to define the correlation between MR and AI in the setting of varying pump speeds, especially given the implications for AV opening and AI progression (Figure 4). Mitral valve repair might be indicated if lower speeds are necessary to preserve AV integrity or maintain pulsatility within the peripheral circulation.

Current Approach MR and mitral stenosis are not contraindications to LVAD implantation. Currently, severe MR does not require surgical repair unless there is expectation of recovery or LVAD explant.16 Mitral stenosis with a mean pressure gradient >10 mm Hg should be considered for concomitant mitral valve repair.

Previously Implanted Prosthetic Valves Aortic Valve For patients with preexisting bioprosthetic AV, a functioning valve does not require removal or replacement.16 For patients with previously implanted mechanical prostheses undergoing LVAD implantation, 2 options exist: (1) replacement with a bioprosthetic valve or (2) exclusion of the valve from circulation. For example, the Cohn sandwich plug technique uses a felt circular patch for outflow tract closure and excludes the metallic valve leaflets from affecting blood flow by creating a barrier on both the ventricular and aortic side of the prosthesis.20

Mitral Valve Figure 3. Transesophageal echocardiogram demonstrating tricuspid regurgitation and septal shift. (A) Transesophageal echocardiographic image obtained in the operating room showing tricuspid regurgitation after implantation of a Heartmate II left ventricular assist device. Septum is bowed toward the left ventricle, and there is severe tricuspid regurgitation. (B) Further imaging, again showing decompression of the left ventricle but reduced tricuspid regurgitation after tricuspid valve repair with annuloplasty.

Mitral Regurgitation Most MR associated with cardiomyopathy is functional and results from LV dilation. With LVAD decompression, LV dimensions decrease and allow the mitral leaflets to coapt, making MR insignificant during device support.9,35 Morgan et al37 found that continuous-flow LVAD implantation significantly decreased the severity of MR (moderate-severe) from 76.0% preoperatively to 8.0% at 1 and 6 months postoperatively. However, there is still potential risk for residual MR when device support cannot be optimized to provide low left-sided filling pressures, such as when there is suction of the interventricular septum over the inflow cannula at higher speeds or the need for intermittent AV opening.30,35 Some have described the use of an edge-to-edge technique in the treatment of moderate-to-severe MR, which fixes the anterior leaflet to the posterior leaflet at a central location and may be effective for a central regurgitant jet.38 It is important to note that exposure of the valve via the left atrium usually requires cardioplegic arrest and possibly increased operative

The presence of prosthetic or mechanical mitral valves is not a contraindication for LVAD placement.39 Prosthetic valves are exposed to normal blood flow after LVAD implantation and should not be at risk for thrombus formation. Thus, patients with previous prosthetic mitral valves typically do not require additional mitral valve surgery, but may require a higher level of anticoagulation postoperatively.24

Valvular Heart Disease and LVAD Speed Modification Ideally, continuous-flow LVADs would operate with total systemic blood flow consisting of contributions from both the LVAD and the LV. The pump speed would therefore be low enough for the AV to intermittently open yet high enough to maintain adequate systemic perfusion and unloading of the LV with reduction in MR grade.40 Intermittent opening of the AV via optimized pump speed could be advantageous in delaying and preventing de novo pathology, as well as allowing myocardial recovery and remodeling through maintenance of native contractility.40,41 It has also been suggested that opening of the AV will more closely mimic physiological pulsatility, which may be helpful in delaying development of acquired von Willebrand disease and pathological gastrointestinal bleeding.42 Finally, incomplete unloading of the LV may help maintain a normal interventricular septal position, creating less distortion of the tricuspid apparatus and reduced TR. Although this provides an optimal scenario, the combination of fully unloading the LV and intermittent AV opening may not be achievable in many patients without recurrence of HF symptoms.

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

220  Circ Heart Fail  January 2014 Table 2.  Current Approaches for Valvular Heart Disease in Patients Supported With LVADs Pathology

Current Approach

Aortic insufficiency

Correct moderate or greater AI at time of LVAD implantation (if patient deemed appropriate to receive LVAD)* Consider surgical intervention for mild AI Suture closure is preferred method

Aortic valve prostheses

Prefer bioprosthetic valve if replacement undertaken* Cohn sandwich plug technique for previously implanted mechanical prostheses

Mitral insufficiency

Not contraindication to LVAD implantation*† May require surgical repair if anticipation of ventricular recovery†

Mitral stenosis

Mean pressure gradient ≥10 mm Hg should be corrected at time of implantation

Mitral valve prostheses

May require higher level of anticoagulation* No need for routine replacement of a properly functioning mechanical valve†

Tricuspid regurgitation

Moderate or greater lesion should prompt consideration for concomitant repair†

AI indicates aortic insufficiency; and LVAD, left ventricular assist device. *In accordance with Recommendations for the Use of Mechanical Circulatory Support from American Heart Association.24 †In accordance with The 2013 International Society of Heart Lung Transplantation Guidelines for Mechanical Circulatory Support: Executive Summary (Level C recommendations).16

Summary Figure 4. Transthoracic echocardiographic findings with varying left ventricular assist device pump speed. Echocardiogram (parasternal long-axis view) reveals (A) mild continuous aortic regurgitation (arrow) and moderate mitral regurgitation (triangle) at 10 000 rpm and (B) moderate continuous aortic regurgitation (arrow) and mild-moderate mitral regurgitation (triangle) at 11 600 rpm.

Currently, clinical management of pump speed is assessed through clinical history and serial echocardiograms, which assess LV unloading, opening of the AV, interventricular septum position, and degree of MR.41 Topilsky et al9 described use of echocardiography to adjust pump speed perioperatively and found that variables such as tricuspid lateral annulus velocity, estimated left atrial pressure, and mitral deceleration index may be useful for future ramp optimization studies. Identifying the ideal pump speed, individualized based on severity of HF and degree of MR, will be one of the most important factors in management of this unique population. A promising protocol by Columbia University provides a standardized echocardiographically guided assessment for speed optimization and diagnosis of device malfunction to be used at discharge. This elegant study made speed adjustments (mean±424 rpm) in 61% of subjects and provided an understanding of device settings to preserve LV geometry, reduce MR severity, and allow intermittent AV opening.41 To our knowledge, the Columbia Ramp Study is one of the first standardized clinical ramp tests that both optimizes pump settings and diagnoses device malfunction.

Patients with advanced HF and coexisting VHD who undergo LVAD implantation constitute a growing population and pose unique challenges in clinical and surgical management. Contemporary continuous-flow LVADs alter blood flow dynamics and cardiovascular physiology and can induce progression of preexisting VHD or lead to de novo VHD. Although we have defined our recommendations for treatment options (Table 2), continued changes to device technology, patient selection, and surgical techniques will undoubtedly lead to further changes in practice. Finally, the rapid growth in the field of mechanical circulatory support necessitates collaborative multicenter approaches to study the impact of these practice changes in the form of prospective randomized trials.

Acknowledgments We thank Dr Priyesh Patel for his assistance in the artistic development of Figure 2.

Disclosures Dr Hernandez reports research funding from Amylin, Johnson & Johnson, and Portola Pharmaceuticals; consulting for AstraZeneca, Johnson & Johnson, Ortho-McNeil-Janssen Pharmaceuticals, and Corthera. Dr Felker reports research funding from Amgen, BG Medicine, Cytokinetics, Johnson & Johnson, Roche Diagnostic Corp, and Otsuka; consulting for Amgen, Cytokinetics, Roche, Otsuka, and Novartis. Dr Rogers reports consulting from Bayer AG, Boston Scientific Corporation, Medtronic, Inc, and Thoratec Corporation. Dr Patel reports consulting for Thoratec Corp and Heartware Inc. Dr Milano reports consulting for Thoratec Corp and Heartware Inc. Ms Wang has no conflicts of interest.

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

Wang et al   Valvular Heart Disease and LVADs   221

References 1. Vigen R, Maddox TM, Allen LA. Aging of the United States population: impact on heart failure. Curr Heart Fail Rep. 2012;9:369–374. 2. Iung B, Vahanian A. Epidemiology of valvular heart disease in the adult. Nat Rev. 2011;8:162–172. 3. Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, Long JW, Ascheim DD, Tierney AR, Levitan RG, Watson JT, Meier P, Ronan NS, Shapiro PA, Lazar RM, Miller LW, Gupta L, Frazier OH, Desvigne-Nickens P, Oz MC, Poirier VL; Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) Study Group. Long-term use of a left ventricular assist device for end-stage heart failure. N Engl. 2001;345:1435–1443. 4. Loor G, Gonzalez-Stawinski G. Pulsatile vs. continuous flow in ventricular assist device therapy. Best Pract Res Clin Anaesthesiol. 2012;26:105–115. 5. Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM 3rd, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH; HeartMate II Investigators. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl. 2009;361:2241–2251. 6. John R, Mantz K, Eckman P, Rose A, May-Newman K. Aortic valve pathophysiology during left ventricular assist device support. J Heart. 2010;29:1321–1329. 7. Pal JD, Klodell CT, John R, Pagani FD, Rogers JG, Farrar DJ, Milano CA; HeartMate II Clinical Investigators. Low operative mortality with implantation of a continuous-flow left ventricular assist device and impact of concurrent cardiac procedures. Circulation. 2009;120(11 Suppl):S215–S219. 8. Cowger J, Pagani FD, Haft JW, Romano MA, Aaronson KD, Kolias TJ. The development of aortic insufficiency in left ventricular assist ­device-supported patients. Circ Heart Fail. 2010;3:668–674. 9. Topilsky Y, Oh JK, Atchison FW, Shah DK, Bichara VM, Schirger JA, Kushwaha SS, Pereira NL, Park SJ. Echocardiographic findings in stable outpatients with properly functioning HeartMate II left ventricular assist devices. J Am. 2011;24:157–169. 10. Balachandran K, Sucosky P, Jo H, Yoganathan AP. Elevated cyclic stretch alters matrix remodeling in aortic valve cusps: implications for degenerative aortic valve disease. Am J. 2009;296:H756–H764. 11. Connelly JH, Abrams J, Klima T, Vaughn WK, Frazier OH. Acquired commissural fusion of aortic valves in patients with left ventricular assist devices. J Heart. 2003;22:1291–1295. 12. Feldman CM, Silver MA, Sobieski MA, Slaughter MS. Management of aortic insufficiency with continuous flow left ventricular assist devices: bioprosthetic valve replacement. J Heart. 2006;25:1410–1412. 13. Rajagopal K, Daneshmand MA, Patel CB, Ganapathi AM, Schechter MA, Rogers JG, Milano CA. Natural history and clinical effect of aortic valve regurgitation after left ventricular assist device implantation. J Thorac. 2013;145:1373–1379. 14. Bryant AS, Holman WL, Nanda NC, Vengala S, Blood MS, Pamboukian SV, Kirklin JK. Native aortic valve insufficiency in patients with left ventricular assist devices. Ann Thorac. 2006;81:e6–e8. 15. Barbone A, Rao V, Oz MC, Naka Y. LVAD support in patients with bioprosthetic valves. Ann Thorac. 2002;74:232–234. 16. Feldman D, Pamboukian SV, Teuteberg JJ, Birks E, Lietz K, Moore SA, Morgan JA, Arabia F, Bauman ME, Buchholz HW, Deng M, Dickstein ML, El-Banayosy A, Elliot T, Goldstein DJ, Grady KL, Jones K, Hryniewicz K, John R, Kaan A, Kusne S, Loebe M, Massicotte MP, Moazami N, Mohacsi P, Mooney M, Nelson T, Pagani F, Perry W, Potapov EV, Eduardo Rame J, Russell SD, Sorensen EN, Sun B, Strueber M, Mangi AA, Petty MG, Rogers J; International Society for Heart and Lung Transplantation. The 2013 International Society for Heart and Lung Transplantation Guidelines for mechanical circulatory support: executive summary. J Heart. 2013;32:157–187. 17. Park SJ, Liao KK, Segurola R, Madhu KP, Miller LW. Management of aortic insufficiency in patients with left ventricular assist devices: a simple coaptation stitch method (Park’s stitch). J Thorac. 2004;127:264–266. 18. Adamson RM, Dembitsky WP, Baradarian S, Chammas J, May-Newman K, Chillcott S, Stahovich M, McCalmont V, Ortiz K, Hoagland P, Jaski B. Aortic valve closure associated with HeartMate left ventricular device support: technical considerations and long-term results. J Heart. 2011;30:576–582. 19. Cohn WE, Frazier OH. The sandwich plug technique: simple, effective, and rapid closure of a mechanical aortic valve prosthesis at left ventricular assist device implantation. J Thorac. 2011;142:455–457. 20. Cohn WE, Demirozu ZT, Frazier OH. Surgical closure of left ventricular outflow tract after left ventricular assist device implantation in patients with aortic valve pathology. J Heart. 2011;30:59–63.

21. Santini F, Forni A, Dandale R, Ribichini F, Rossi A, Franchi G, Onorati F, Vassanelli C, Mazzucco A, Faggian G. First successful management of aortic valve insufficiency associated with HeartMate II left ventricular assist device support by transfemoral CoreValve implantation: the Columbus’s egg? J Am. 2012;5:114–115. 22. D’Ancona G, Pasic M, Buz S, Drews T, Dreysse S, Hetzer R, Unbehaun A. TAVI for pure aortic valve insufficiency in a patient with a left ventricular assist device. Ann Thorac. 2012;93:e89–e91. 23. Parikh KS, Mehrotra AK, Russo MJ, Lang RM, Anderson A, Jeevanandam V, Freed BH, Paul JD, Karol J, Nathan S, Shah AP. Percutaneous transcatheter aortic valve closure successfully treats left ventricular assist ­device-associated aortic insufficiency and improves cardiac hemodynamics. J Am. 2013;6:84–89. 24. Peura JL, Colvin-Adams M, Francis GS, Grady KL, Hoffman TM, Jessup M, John R, Kiernan MS, Mitchell JE, O’Connell JB, Pagani FD, Petty M, Ravichandran P, Rogers JG, Semigran MJ, Toole JM; American Heart Association Heart Failure and Transplantation Committee of the Council on Clinical Cardiology; Council on Cardiopulmonary, Critical Care, Perioperative and Resuscitation; Council on Cardiovascular Disease in the Young; Council on Cardiovascular Nursing; Council on Cardiovascular Radiology and Intervention, and Council on Cardiovascular Surgery and Anesthesia. Recommendations for the use of mechanical circulatory support: device strategies and patient selection: a scientific statement from the American Heart Association. Circulation. 2012;126:2648–2667. 25. Krishan K, Nair A, Pinney S, Adams DH, Anyanwu AC. Liberal use of tricuspid-valve annuloplasty during left-ventricular assist device implantation. Eur J. 2012;41:213–217. 26. Piacentino V 3rd, Williams ML, Depp T, Garcia-Huerta K, Blue L, Lodge AJ, Mackensen GB, Swaminathan M, Rogers JG, Milano CA. Impact of tricuspid valve regurgitation in patients treated with implantable left ventricular assist devices. Ann Thorac. 2011;91:1342–1346. 27. Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, Massey T, Milano CA, Moazami N, Sundareswaran KS, Farrar DJ; HeartMate II Clinical Investigators. Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac. 2010;139:1316–1324. 28. Morgan JA, Paone G, Nemeh HW, Murthy R, Williams CT, Lanfear DE, Tita C, Brewer RJ. Impact of continuous-flow left ventricular assist device support on right ventricular function. J Heart. 2013;32:398–403. 29. Holman WL, Bourge RC, Fan P, Kirklin JK, Pacifico AD, Nanda NC. Influence of longer term left ventricular assist device support on valvular regurgitation. ASAIO J. 1994;40:M454–M459. 30. Saeed D, Kidambi T, Shalli S, Lapin B, Malaisrie SC, Lee R, Cotts WG, McGee EC Jr. Tricuspid valve repair with left ventricular assist device implantation: is it warranted? J Heart. 2011;30:530–535. 31. Piacentino V 3rd, Ganapathi AM, Stafford-Smith M, Hsieh MK, Patel CB, Simeone AA, Rogers JG, Milano CA. Utility of concomitant tricuspid valve procedures for patients undergoing implantation of a ­continuous-flow left ventricular device. J Thorac. 2012;144:1217–1221. 32. Piacentino V 3rd, Troupes CD, Ganapathi AM, Blue LJ, Mackensen GB, Swaminathan M, Felker GM, Stafford-Smith M, Lodge AJ, Rogers JG, Milano CA. Clinical impact of concomitant tricuspid valve procedures during left ventricular assist device implantation. Ann Thorac. 2011;92:1414–1418. 33. Deo SV, Hasin T, Altarabsheh SE, McKellar SH, Shah IK, Durham L 3rd, Stulak JM, Daly RC, Park SJ, Joyce LD. Concomitant tricuspid valve repair or replacement during left ventricular assist device implant demonstrates comparable outcomes in the long term. J Card. 2012;27:760–766. 34. Salzberg SP, Lachat ML, von Harbou K, Zünd G, Turina MI. Normalization of high pulmonary vascular resistance with LVAD support in heart transplantation candidates. Eur J. 2005;27:222–225. 35. Aaronson KD, Patel H, Pagani FD. Patient selection for left ventricular assist device therapy. Ann Thorac. 2003;75(6 Suppl):S29–S35. 36. Patel JB, Borgeson DD, Barnes ME, Rihal CS, Daly RC, Redfield MM. Mitral regurgitation in patients with advanced systolic heart failure. J Card. 2004;10:285–291. 37. Morgan JA, Brewer RJ, Nemeh HW, Murthy R, Williams CT, Lanfear DE, Tita C, Paone G. Left ventricular reverse remodeling with a continuous flow left ventricular assist device measured by left ventricular end-diastolic dimensions and severity of mitral regurgitation. ASAIO J. 2012;58:574–577.

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

222  Circ Heart Fail  January 2014 38. Maisano F, Torracca L, Oppizzi M, Stefano PL, D’Addario G, La Canna G, Zogno M, Alfieri O. The edge-to-edge technique: a simplified method to correct mitral insufficiency. Eur J. 1998;13:240–245. 39. Rao V, Slater JP, Edwards NM, Naka Y, Oz MC. Surgical management of valvular disease in patients requiring left ventricular assist device support. Ann Thorac. 2001;71:1448–1453. 40. Bishop CJ, Mason NO, Kfoury AG, Lux R, Stoker S, Horton K, Clayson SE, Rasmusson B, Reid BB. A novel non-invasive method to assess aortic valve opening in HeartMate II left ventricular assist device patients using a modified Karhunen-Loève transformation. J Heart. 2010;29:27–31. 41. Uriel N, Morrison KA, Garan AR, Kato TS, Yuzefpolskaya M, Latif F, Restaino SW, Mancini DM, Flannery M, Takayama H, John R, Colombo

PC, Naka Y, Jorde UP. Development of a novel echocardiography ramp test for speed optimization and diagnosis of device thrombosis in ­continuous-flow left ventricular assist devices: the Columbia ramp study. J Am. 2012;60:1764–1775. 42. Uriel N, Pak SW, Jorde UP, Jude B, Susen S, Vincentelli A, Ennezat PV, Cappleman S, Naka Y, Mancini D. Acquired von Willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation. J Am. 2010;56:1207–1213. Key Words: advanced heart failure ◼ mechanical circulatory support ◼ valvular heart disease

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

Valvular Heart Disease in Patients Supported With Left Ventricular Assist Devices Teresa S. Wang, Adrian F. Hernandez, G. Michael Felker, Carmelo A. Milano, Joseph G. Rogers and Chetan B. Patel Circ Heart Fail. 2014;7:215-222 doi: 10.1161/CIRCHEARTFAILURE.113.000473 Circulation: Heart Failure is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231 Copyright © 2014 American Heart Association, Inc. All rights reserved. Print ISSN: 1941-3289. Online ISSN: 1941-3297

The online version of this article, along with updated information and services, is located on the World Wide Web at: http://circheartfailure.ahajournals.org/content/7/1/215

Permissions: Requests for permissions to reproduce figures, tables, or portions of articles originally published in Circulation: Heart Failure can be obtained via RightsLink, a service of the Copyright Clearance Center, not the Editorial Office. Once the online version of the published article for which permission is being requested is located, click Request Permissions in the middle column of the Web page under Services. Further information about this process is available in the Permissions and Rights Question and Answer document. Reprints: Information about reprints can be found online at: http://www.lww.com/reprints Subscriptions: Information about subscribing to Circulation: Heart Failure is online at: http://circheartfailure.ahajournals.org//subscriptions/

Downloaded from http://circheartfailure.ahajournals.org/ at NORTH DAKOTA STATE UNIV on June 25, 2015

Valvular heart disease in patients supported with left ventricular assist devices.

Patients with advanced HF and coexisting VHD who undergo LVAD implantation constitute a growing population and pose unique challenges in clinical and ...
674KB Sizes 0 Downloads 0 Views