Journal of the American College of Cardiology  2014 by the American College of Cardiology Foundation Published by Elsevier Inc.

Vol. 63, No. 14, 2014 ISSN 0735-1097/$36.00 http://dx.doi.org/10.1016/j.jacc.2014.01.019

Mechanisms, Pathophysiology, and Clinical Aspects of Incomplete Stent Apposition Guilherme F. Attizzani, MD,*yz Davide Capodanno, MD, PHD,*x Yohei Ohno, MD,* Corrado Tamburino, MD, PHD*x Catania, Italy; Jundiaí, Brazil; and Cleveland, Ohio Incomplete stent apposition (ISA) is characterized by the lack of contact of at least 1 stent strut with the vessel wall in a segment not overlying a side branch; it is more commonly found in drug-eluting stents than bare-metal stents. The accurate diagnosis of ISA, initially only possible with intravascular ultrasound, can currently be performed with higher accuracy by optical coherence tomography, which also enables strut-level assessment due to its higher axial resolution. Different circumstances related both to the index procedure and to vascular healing might influence ISA occurrence. Although several histopathology and clinical studies linked ISA to stent thrombosis, potential selection bias precluded definitive conclusions. Initial studies usually performed single time point assessments comparing overall ISA percentage and magnitude in different groups (i.e., stent type), thus hampering a comprehensive understanding of its relationship with vascular healing. Serial intravascular imaging studies that evaluated vascular response heterogeneity recently helped fill this gap. Some particular clinical scenarios such as acute coronary syndromes, bifurcations, tapered vessels, overlapping stents, and chronic total occlusions might predispose to ISA. Interventional cardiologists should be committed to optimal stent choices and techniques of implantation and use intravascular imaging guidance when appropriate to aim at minimizing acute ISA. In addition, the active search for new stent platforms that could accommodate vessel remodeling over time (i.e., self-expandable stents) and for new polymers and/or eluting drugs that could induce less inflammation (hence, less positive remodeling) could ultimately reduce the occurrence of ISA and its potentially harmful consequences. (J Am Coll Cardiol 2014;63:1355–67) ª 2014 by the American College of Cardiology Foundation

Incomplete stent apposition (ISA), also described as stent malapposition, is a morphological characteristic defined by the lack of contact between at least 1 stent strut and the underlying intimal surface of the vessel wall in a segment not overlying a side branch. Due to the lack of sufficient resolution and the inability of providing cross-sectional images, angiography does not allow the diagnosis of ISA, whereas high-resolution intravascular imaging modalities, such as intravascular ultrasound (IVUS) (axial resolution of approximately 150 mm) and optical coherence tomography (OCT) (axial resolution of approximately 10 to 15 mm) do allow diagnosis (1). According to the time that ISA is diagnosed relative to the index procedure, ISA is labeled as acute (i.e., diagnosed post-procedure), persistent (i.e., diagnosed post-procedure and persistent at follow-up assessment), and

From the *Division of Cardiology, Ferrarotto Hospital, University of Catania, Catania, Italy; yDivision of Interventional Cardiology, Pitangueiras Hospital, Jundiaí, SP, Brazil; zHarrington Heart and Vascular Institute, University Hospitals, Case Medical Center, Cleveland, Ohio; and the xExcellence Through Newest Advances (ETNA) Foundation, Catania, Italy. Dr. Attizzani has received consultant fees from St. Jude Medical, Inc. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose. Manuscript received October 21, 2013; revised manuscript received December 12, 2013, accepted January 7, 2014.

acquired (i.e., not present post-procedure, but identified at follow-up assessment). In the absence of baseline intravascular imaging assessment, it is described simply as late ISA (2,3). Several factors are responsible for this pathological phenomenon and its different presentations, as follows: 1) inadequate stent implantationdin this setting, there are basically 2 possibilities: a) marked mismatch between stent size selection and luminal dimensions (i.e., stent diameter smaller than reference lumen diameter), a situation in which, regardless of optimal stent expansion, ISA will occur (4); or b) stent underexpansion despite an adequate stent–artery ratio due to several different factors, such as inadequate pressure of implantation and/or plaque-related factors (Fig 1); 2) chronic stent recoil (5); 3) thrombus dissolution after primary percutaneous coronary intervention (PCI) (6); 4) positive vessel remodeling (6–11); and 5) inadequate (i.e., insufficient and/or delayed) neointimal hyperplasia (12–15). Incomplete Stent Apposition and Potential Relationship With Adverse Clinical Outcomes A clear relationship between ISA and adverse cardiovascular events is still controversial. Although some studies have

1356

Attizzani et al. Incomplete Stent Apposition

failed to demonstrate such connections (7,16–18), others have done so (12,19–24) (Table 1). BMS = bare-metal stent(s) Lee et al. (25) demonstrated that DES = drug-eluting stent(s) among 30 patients with very late EES = everolimus-eluting stent thrombosis (ST), 73.9% of stent(s) the drug-eluting stent (DES) group ISA = incomplete stent exhibited ISA (among whom apposition 64.7% had late acquired ISA), IVUS = intravascular whereas none of the bare-metal ultrasound stent (BMS) group had ISA. OCT = optical coherence Although the results showed a tomography higher incidence of ISA in paOLP = overlapping tients with DESs compared with PCI = percutaneous coronary BMSs, confirming data from intervention previous reports (20), they also PES = paclitaxel-eluting highlighted that ISA is not an stent(s) isolated factor responsible for ST = stent thrombosis DES thrombosis (i.e., 26.1% of SES = sirolimus-eluting DES patients had stent thromstent(s) bosis without evident ISA) (25). ZES = zotarolimus-eluting Guagliumi et al. (26) demonstent(s) strated by combined IVUS and OCT assessments that the presence and magnitude (i.e., area and distance) of ISA were significantly higher in patients with DES thrombosis compared with those without DES thrombosis. In addition, these investigators implicated 1 mechanism that led to ISA (i.e., positive remodeling) and another mechanism that could be a consequence of ISA (i.e., uncovered stent struts) as independent predictors of DES thrombosis (Fig. 2). Eosinophilic-rich inflammatory infiltrates associated with positive remodeling were also Abbreviations and Acronyms

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

demonstrated (27). Alfonso et al. (27) also identified ISA in patients with ST using IVUS and OCT (ISA was identified in 40% and 47% of patients, respectively). Kang et al. (28) performed OCT imaging in 33 patients with very late ST and found that in the DES group, 52% and 64% had ISA and thrombi, respectively, whereas no patients presented with ISA in the BMS group. ISA was also more frequently found in 124 patients with very late definite DES versus BMS thrombosis (52% vs. 16%, respectively, p ¼ 0.005) in a multicenter registry. Notably, greater ISA area was also demonstrated in the DES group compared with the BMS group. Moreover, among the DES group, a higher prevalence of ISA was identified in sirolimus-eluting stents (SES) compared with paclitaxel-eluting stents (PES) (58% vs. 37%, respectively, p ¼ 0.02) (29). More recently, higher rates of ISA were revealed in 34 patients with DES (56%) and BMS (11%) thrombosis in Italy (30). Although extremely insightful, cautious interpretation of these results is crucial due to potential selection bias. Understanding the underlying mechanisms involved in ISA is of paramount importance to the interventional cardiologist because it enables the identification of which features are potentially modifiable and which resources should be utilized during the index intervention to optimize implantation quality and minimize the likelihood of having ISA and its potentially harmful consequences (31–33). Although physiological vascular healing after stent implantation leads to progressive reduction of ISA over time (12–14,34,35), recent serial OCT assessments elucidated this complex mechanism more clearly, showing that the greater the acute ISA (i.e., after the index procedure), the higher the possibility of its persistence at follow-up (14,15). Furthermore, these studies demonstrated that acute ISA affects the overall vascular response negatively, ultimately leading to delayed stent strut coverage (12–15) and even to ST (12). The comprehension of this complex scenario, moreover, is key for investigators and biomedical engineers so that they can actively search for constant improvements in mechanical (i.e., stent platforms) (36) and biological (i.e., antirestenotic drugs, polymers) (37) components of coronary artery scaffolds. Histopathology Investigations

Figure 1

ISA in Vessels With Similar Dimensions

(A) A perfectly rounded vessel (red circle) with 3-mm diameter is depicted. (B) The gray-dashed lines correspond to a representation of a 2.5-mm stent implanted with high pressure, reaching 2.7 mm. In this situation, although the stent is very well and evenly expanded, incomplete stent apposition (ISA) occurred because there is a clear stent–vessel mismatch. (C) Conversely, a 3.0-mm stent also implanted with high pressure is seen, in which the horizontal diameter reaches 3.1 mm, but the vertical diameter is only 2.5 mm as the consequence of calcified plaques (yellow with red dots) that impair proper expansion, therefore, leading to ISA in some regions of the stent. Figure is by Craig Skaggs.

Histopathology studies have been instrumental in enabling a better understanding of mechanisms that lead to ISA. The pivotal demonstration by Virmani et al. (37) that Cypher (Cordis, Johnson & Johnson, New Brunswick, New Jersey) SES thrombosis was linked to hypersensitivity reactions and to a complex proinflammatory milieu raised the initial concerns regarding the safety issues of DESs (37). The investigators identified the presence of ISA with thick layers of fibrin thrombus separating stent struts from the vessel wall. In a larger series of patients, Joner et al. (38) ratified the complexity of the multifactorial scenario (i.e., procedural and clinical related features) associated with DES failure, describing the presence of ISA coupled with delayed arterial

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

Table 1

Impact of ISA on Cardiovascular Events

N

Clinical Presentation

Type of Stent

Assessed by

Follow-Up (Months)

Positive Association Between ISA and Cardiovascular Events

Design

Assessment of Malapposition

Guagliumi et al. (22)

PP, SC

21

Stable/ACS

ZES

OCT

6

Malapposed struts per stent: 2.0  4.2% (OLP), 3.8  7.1% (non-OLP)

No

Kubo et al. (23)

PP, SC

45

Stable/ACS

SES

OCT

9

Malapposed struts per stent: 0.2  1.2% (stable), 4.0  6.7% (ACS)

No

Guagliumi et al. (24)

PP, SC

77

Stable/ACS

SES/PES/ZES/BMS

OCT

6

Malapposed struts per stent: 2.9  7.0% (SES; OLP), 1.9  4.2% (SES; non-OLP), malapposed struts per stent: 5.5  15.6% (PES; OLP), 0.7  2.1% (PES; non-OLP), malapposed struts per stent: 0.1  0.2% (ZES; OLP), 0.1  0.1% (ZES; non-OLP), malapposed struts per stent: 0.1  0.1% (BMS; OLP), 0.2  0.9% (BMS; non-OLP)

No

Guagliumi et al. (22)

PP, SC

42

Stable/ACS

EES CoCr/EES PtCr

OCT

6

Malapposed struts per stent: 0.5%, malapposition area 0.02 mm2 (EES CoCr), malapposed struts per stent: 0.0%, malapposition area 0.01 mm2 (EES CoCr)

No

Hong et al. (16)

RP, SC

557

Stable/ACS

SES/PES

IVUS

6

LSM CSA: 3.0  1.9 mm2 (SES), 3.2  1.9 mm2 (PES) (LSM occurred in 12.1% of lesions)

No

Steinberg et al. (17)

PP, MC

1,580

Stable/ACS

PES/BMS

IVUS

9

Late-acquired ISA was identified in 2.7% of BMS, 3.1% of PES-SR, and 15.4% of PES-MR

No

Tanabe et al. (18)

PP, MC

469

Stable/ACS

PES/BMS

IVUS

6

Mean ISA area: 3.6  1.2 mm2 (PES-SR), 2.1  1.4 mm2 (PES-MR), 3.0  2.1 mm2 (BMS)

No

Cook et al. (19)

PP, MC

194

Stable/ACS

SES/PES

IVUS

8

Max ISA CSA: 4.5  5.0 mm2 (SES), 5.0  5.1 mm2 (PES), total ISA length: 2.3  1.4 mm (SES), 1.0  0.9 mm (PES)

Yes (the presence of ISA after DES implantation was associated with a higher rate of MI and VLST)

Cook et al. (21)

PP, SC

188

Stable/ACS

SES/PES

IVUS

8

Max ISA CSA: 8.3  7.5 mm2 (VLST), 4.0  3.8 mm2 (control), max ISA length: 6.3  6.3 mm2 (VLST), 1.5  1.0 mm2 (control)

Yes (ISA is highly prevalent in patients with VLST after DES implantation)

ACS ¼ acute coronary syndrome(s); BMS ¼ bare-metal stent(s); CoCr ¼ cobalt-chromium; CSA ¼ cross-sectional area; EES ¼ everolimus-eluting stent(s); ISA ¼ incomplete stent apposition; IVUS ¼ intravascular ultrasound; LSM ¼ late stent malapposition; MACE ¼ major adverse cardiac event(s); MC ¼ multicenter; MI ¼ myocardial infarction; MR ¼ moderate-release; OCT ¼ optical coherence tomography; OLP ¼ overlapping; PCI ¼ percutaneous coronary intervention; PES ¼ paclitaxel-eluting stent; PP ¼ prospective; PtCr ¼ platinum-chromium; RP ¼ retrospective; SC ¼ single center; SES ¼ sirolimus-eluting stent(s); SR ¼ slow-release; VLST ¼ very late stent thrombosis; ZES ¼ zotarolimus-eluting stent(s).

Attizzani et al. Incomplete Stent Apposition

First Author (Ref. #)

1357

1358

Figure 2

Attizzani et al. Incomplete Stent Apposition

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

Late ISA, Thrombosis, and Neoatherosclerosis

This 60-year-old patient had multiple drug-eluting stents implanted in overlapping fashion in the right coronary artery (RCA) 6 years before the current hospitalization due to non–ST-segment elevation myocardial infarction. (A) Angiography demonstrates “haziness” in the mid-RCA (white arrow). The white dashed lines in the longitudinal optical coherence tomographic (OCT) image (bottom) correspond to the respective locations of the cross sections identified by the same letter. (B) A normal pattern of thin neointimal proliferation is depicted without incomplete stent apposition (ISA) or uncovered struts. (C, D, E). However, ISA with some attached thrombus is depicted (white arrows). (F) A huge intraluminal thrombus burden is depicted (white arrow), whereas in (G) another normal pattern (thin, bright tissue covering the struts) of vascular response is depicted. (H) A transition to an abnormal pattern of stent coverage is shown as low-signal intensity tissue, identified as peri-struts (i.e., acellular material, such as fibrin or proteoglycan). (I) Although normal tissue covers the struts from 1 to 7 o’clock, the white dashed square in J demonstrates lipidic tissue (marked light attenuation with poorly defined borders, white asterisk) covering the stent (i.e., neoatherosclerosis), which almost “hides” the stent struts (white arrows), and a thin cap overlying the lipid plaque (white arrowheads). This case has important teaching points, as follows: 1) In a long stented segment, there is minimal ISA, but it is linked to stent thrombosis; therefore, it is important to perform lesion-level assessments rather than display only an overall percentage of the ISA and the assessment of ISA clusters in clinical trials; 2) Although a clear relationship between the acellular tissue peri-strut (i.e., potentially secondary to inflammation drug and/or polymer toxicity) identified in H and the ISA (i.e., eventually secondary to positive remodeling due to drug and/or polymer toxicity) cannot be clearly derived from this case, one can speculate that these pathological phenomena could be linked by a proinflammatory milieu (35); and 3) Similar rationale of item 2 can be derived from the lipidic neointimal tissue covered by thin fibrous cap identified more proximally (I and J).

healing and the presence of fibrin surrounding the stent struts as 1 of the mechanisms associated with ST (35) (Fig. 2). More recently, Cook et al. (39) correlated the histopathology of thrombus aspirates with IVUS findings in patients with very late DES thrombosis. Interestingly, they were able to establish a correlation between the amount of eosinophil infiltrates with the extent of positive vessel remodeling, suggesting hypersensitivity reactions of all 3 layers of the vessel wall in patients with ISA; in patients without positive remodeling, however, only a few or no eosinophils were detected. Despite the outstanding contribution of histopathology research toward a better comprehension of the mechanisms that lead to ISA and its potential consequences, the data presented by these studies should be carefully interpreted because they represent a highly selected population (i.e., patients who had a DES implanted and died). Therefore, although ISA is suggested

as 1 of the potential mechanisms linked with adverse cardiovascular events in these studies, a linear cause and effect relationship cannot be derived from these findings. Another significant role played by histopathology in this scenario is the validation of intravascular imaging findings, which is of utmost importance because they enable accurate in vivo qualitative (rather than merely quantitative) assessments of vascular responses in a prospective fashion, which could ultimately facilitate the understanding of the pathophysiology of ISA. The feasibility and accuracy of tissue characterization utilizing OCT has been previously demonstrated (40–42). This could be particularly important to help elucidate, for example, the underlying mechanisms that lead to ISA in different scaffolds (Fig. 3) (43). Radu et al. (44) suggested the term “pseudoapposed” to define stent struts that were “apposed” to low-signal intensity regions, which were likely composed of acellular material such as fibrin or

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

proteoglycan, and similar to what has been described by histopathology (38). Further investigations are required to assess the relevance of these findings. Fundamental Differences Between IVUS and OCT in ISA Interpretation Due to their high resolution, intravascular imaging modalities are essential to properly diagnose ISA in vivo. Although a complete description of IVUS and OCT imaging is beyond the scope of this paper, it is important to be cognizant of the fundamental differences and potential additive values of these imaging systems in the diagnosis of ISA, hence, facilitating data interpretation. IVUS is a sound-based imaging system with an axial resolution of approximately 150 mm and a penetration of 4 to 8 mm. It enables detailed assessment of vessel size because the external elastic membrane is clearly visualized, therefore allowing accurate quantification of positive vascular remodeling (45,46). The diagnosis of ISA by IVUS is determined by the presence of blood speckle (or also by color Doppler imaging in the case of lower frequency IVUS) behind the stent struts not overlying a side branch. Data are provided in cross-sectional (i.e., area of ISA) and longitudinal (i.e., volume of ISA) levels (6). Conversely, OCT is a nearinfrared, light-based system with an approximately 10-fold higher resolution (approximately 10- to 15-mm axial resolution) compared with IVUS, but with less tissue penetration (1 to 3 mm). This lack of deep tissue penetration precludes accurate quantification of positive vessel remodeling, whereas the blood-free environment obtained during OCT image acquisition, coupled with its higher axial resolution, provides a sharper delineation of the stent–lumen interface compared with IVUS, thus allowing easier image interpretation (1,47) (Fig. 4). Several authors demonstrated that OCT is superior to IVUS in detecting ISA, including in the left main coronary artery (1,12,27,48). Another important piece of information added by OCT on top of cross-sectional and longitudinal levels of assessment is the strut-level analysis, which enables the diagnosis and quantification of ISA in each individual strut. It is also important to highlight that the initial diagnosis of ISA by OCT, including that performed in the catheterization laboratory in our daily practice, is qualitative, whereas the final determination of whether or not 1 strut exhibits ISA is quantitative (i.e., considering blooming, strut, and polymer thicknesses) (Fig. 5) (49,50). Taken together, these characteristics indicate that IVUS and OCT complement each other in the diagnosis and elucidation of ISA mechanisms (26,27). Study Designs and Endpoints Several studies have evaluated ISA by single time point (i.e., follow-up) assessments (51–55). Although the information provided by these studies is extremely valuable in the determination of ISA rates using different stents in vivo, it is

Attizzani et al. Incomplete Stent Apposition

1359

impossible to completely elucidate this morphological feature in a single “snapshot.” Healing after stent implantation is a continuum influenced by several different factors, related to both baseline characteristics (i.e., clinical and procedure related) and to alterations that occur along this complex biological process (i.e., inflammation, delayed healing, and positive remodeling); therefore, serial intravascular imaging studies using IVUS and/or OCT (11,12,14,15,35,56) have helped to fill the gap left by single time point assessments. Despite the importance of serial invasive imaging assessments to comprehensively characterize the natural history of ISA, increased costs, coupled with ethical issues of submitting patients to the inherent risks of repeated invasive interventions, x-ray exposure, and contrast media injections, have hampered their more extensive utilization (57,58). Another consideration is that data presentation across studies evaluating ISA lacks standardization. The majority of trials provide overall percentages, areas, and volumes of ISA associated with the groups (i.e., different stents) being evaluated (59,60), rather than identifying and quantifying segments in which the abnormality is present (i.e., lesionlevel assessment) (61), therefore restricting the assessment of intra- and interindividual heterogeneity of vascular responses, and consequently, of ISA. Räber et al. (62) demonstrated that, despite an overall equivalent percentage of ISA late after SES and PES implantation, significant clustering of ISA was depicted in the former, suggesting different patterns of vascular responses between the scaffolds being evaluated. Gutiérrez-Chico et al. (15) described heterogeneous morphological healing patterns in patients with acute ISA using serial OCT imaging assessments. Another tool to assess whether heterogeneity exists in vascular responses is the coefficient of variation, which is obtained by dividing each lesion into subsegments (i.e., dividing the standard deviation by the mean percentage of ISA) (35,63–65). These analytical approaches could ultimately expand our knowledge of assessing pathological vascular reactions to stents, particularly ISA, and avoiding, for instance, the “diluting” of the results of clustered ISAs that may occur in localized segments of certain individuals, and in the entire population of the study. Whether these findings significantly affect clinical outcomes remains to be demonstrated. Incomplete Stent Apposition in Different Clinical Scenarios Acute coronary syndromes. Nakazawa et al. (66) used histopathology to demonstrate that vascular healing was delayed and ST rates were increased in patients who underwent DES implantation for myocardial infarction compared with stable angina; culprit sites in the former had greater fibrin deposition and inflammation, together with less neointimal proliferation. In addition, higher rates of struts penetrating the necrotic core and greater external elastic membrane areas were found. Hong et al. (67) showed in vivo

1360

Figure 3

Attizzani et al. Incomplete Stent Apposition

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

Representative Images of Late Stent Thrombosis in First-Generation SES and ISA in PES

(A) Histological sections from sirolimus-eluting stents (SES). A 40-year-old woman who received 2 SES in the left anterior descending artery (LAD) and right coronary artery (RCA) 17 months ante mortem died suddenly 4 days after surgical removal of a melanoma (wide excision). Antiplatelet therapy (aspirin and clopidogrel) was discontinued 5 days before the surgery. Histological sections of the SES in the LAD showed total thrombotic occlusion and diffuse inflammation (a). Numerous inflammatory cells were observed within the neointimal area (b). Inflammatory reaction predominantly consists of T lymphocytes (c) (CD45RO) and eosinophils (d) (Luna stain). Note that the same reaction was observed in the SES in the RCA (e), and severe inflammation resulted in incomplete stent apposition (ISA) of stent struts (f). (B) Histological sections from a PES showing ISA. A 69-year-old man who received a paclitaxel-eluting stent (PES) in a saphenous vein graft died suddenly 3 months after stent placement. Histological sections showed thrombotic occlusion in the PES (a, b); note the ISA was secondary to severe fibrin deposition (c). A 48-year-old man with a PES implant in the proximal LAD died suddenly at 40 months. Histological sections showed thrombotic occlusion of the PES (d). Most struts show ISA with fibrin deposition underneath the stent struts (e, f). Thr ¼ thrombus. Adapted and reprinted with permission from Nakazawa et al. (43).

that BMS implantation during primary PCI was an independent predictor of late ISA, but it was not linked to increased rates of adverse events at 3-year follow-up. In a serial assessment by IVUS, BMS and PES rates of acute ISA were equivalent, whereas PES were more frequently associated with late acquired ISA at follow-up, mostly because of thrombus dissolution and positive remodeling. The investigators also found that an area of acute ISA >1.2 mm2 determined resolved ISA from persistent ISA (6). Likewise, no differences were observed between SESs and BMSs

regarding acute ISA rates after primary PCI (38.5% and 33.8%, p ¼ 0.51), whereas late acquired ISA was more often identified in SES, mainly secondary to positive vessel remodeling (84% of the cases) (68). Data obtained from the HORIZONS-AMI (Harmonizing Outcomes With Revascularization and Stents in Acute Myocardial Infarction) study did not link early stent thrombosis with the presence of acute ISA (69); in addition, the 13-month follow-up of the HORIZONS-OCT study included 188 stents (PES and BMS) and showed higher rates of late ISA with PES

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

Figure 4

Attizzani et al. Incomplete Stent Apposition

1361

ISA by IVUS and OCT

In A-I, an intravascular ultrasound (IVUS) cross-sectional image shows stent struts (red arrows) with ISA (yellow asterisks). The same co-registered region of interest is demonstrated in an OCT image (A-II), in which one can also identify the stent struts (red arrows) with ISA (yellow asterisks). In B-I, however, the IVUS image does not enable a clear diagnosis of ISA, whereas in the same co-registered region (B-II) imaged by OCT, the presence of stent struts (red arrow) with ISA (yellow asterisks) is clearly depicted. The blue lines in the longitudinal views located in the bottom of each cross section demonstrate the region of the pullback from which the cross sections were obtained. Abbreviations as in Figures 2 and 3.

compared with BMS (70). Gonzalo et al. (71) used OCT imaging to demonstrate, in vivo, that DESs implanted during primary PCI compared with stable and/or unstable angina more often had ISA at 9-month follow-up (75% vs. 25.8%, respectively, p ¼ 0.001). The investigators also identified DES implantation during primary PCI as the only independent predictor of ISA (odds ratio: 9.8, 95% confidence interval: 2.4 to 40.4; p ¼ 0.002). Davlouros et al. (72) showed in a small study that ISA was correlated with a lack of coverage 6 months after PES implantation for acute coronary syndromes. Kubo et al. (23) compared unstable versus stable angina patients who underwent DES implantation and

Figure 5

demonstrated significantly higher rates of both acute (67% vs. 32%; p ¼ 0.038) and late ISA (33% vs. 4%, respectively; p ¼ 0.012) in the former setting; no adverse events were reported in the 55 patients in this study at 9-month followup. Adrenergic vasoconstriction, which leads to underestimation of the reference vessel size, which then leads to an equivocal stent choice (i.e., smaller than vessel dimensions), is another potential cause of ISA after acute myocardial infarction (73). The European Guidelines on ST-segment elevation myocardial infarction recommend intracoronary administration of nitrates for more accurate stent sizing during primary PCI (74).

Quantification of ISA by OCT

In daily clinical practice, the initial assessment of ISA by OCT is qualitative, as demonstrated in A; however, the accurate determination of ISA is ultimately quantitative, because several factors such as polymer and strut thicknesses (different for each stent) and one-half of the blooming should be taken into account for the determination of the final malapposed distance. ISA quantification of some struts are demonstrated by the purple squares in B, whereas the region highlighted by the white square in C demonstrates a strut (blue arrow), which although seemed to be ISA by qualitative assessment, was actually properly apposed (i.e., in contact with the lumen boundary, blue dashed line), whereas another strut was demonstrated to be malapposed (green square, malapposed distance calculation on the right side of the figure). Abbreviations as in Figure 2.

1362

Attizzani et al. Incomplete Stent Apposition

Overlapping stents. In overlapping (OLP) stent regions, a greater amount of metal and increased concentrations of eluting drugs or polymers might increase the risk of flow disruption, inflammation, fibrin deposition, and thrombotic events. Finn et al. (75) performed histopathological evaluation of OLP stents in animal models and showed that although both first-generation SES and PES promoted delayed arterial healing and inflammation at OLP sites compared with BMS, PES had greater fibrin deposition and medial cell loss, as well as a greater external elastic membrane area compared with both SES and BMS. In concordance, immature neointima was found by Shinke et al. (76) in OLP PES regions; furthermore, these investigators used angioscopy to identify more thrombus in PES compared with BMS OLP regions. Recently, an in vitro assessment linked BMS deployed in ISA or OLP fashions with higher thrombogenicity compared with well-apposed, length-matched controls (1.58-fold, p < 0.001; and 2.32-fold, p < 0.001). The investigators also demonstrated that thin struts were less prone to thrombosis compared with thick-strut stents (77). The concerns raised by histopathology and in vitro data were amplified by in vivo OCT evaluation performed by Tanigawa et al. (78), who showed significantly higher rates of acute ISA in OLP versus non-OLP regions despite aggressive stent optimization, and by Guagliumi et al. (24), who demonstrated higher rates of late ISA in OLP segments of SES and PES compared with zotarolimus-eluting stents (ZES) and BMS. In the same study, numerically higher, although not statistically significant, rates of ISA were found in OLP compared with non-OLP PES, whereas no differences were shown in the SES, ZES, and BMS groups (24). Aoki et al. (79) did not highlight concerns regarding ISA in OLP PES segments at 9-month follow-up. Meanwhile, Otake et al. (80) demonstrated that despite the vessel size increase after OLP PES, no significant ISA was demonstrated; likewise, no significant ISA was demonstrated in the everolimus-eluting stent (EES) group in the same retrospective IVUS study at 9-month follow-up. In addition, Dawkins et al. (81) did not show increased ISA in OLP segments of PES. An interesting insight in this setting was recently demonstrated by Guagliumi et al. (22); similar stent platforms and antirestenotic drugs (i.e., zotarolimus), but different polymers and release kinetics, led to different rates of ISA in OLP regions (i.e., slow-release ZES showed higher rates of ISA compared with fast-release ZES). Different stent platforms (i.e., cobaltchromium vs. platinum-chromium), but same drug-release kinetics, in EES did not reveal differences in terms of ISA (82). The importance of the underlying plaque in the results of OLP stenting was investigated by Tahara et al. (83), who demonstrated that increased rates of ISA and uncovered struts were present when OLP stenting was performed in normal-appearing arterial segments, whereas restenosis occurred exclusively in segments with previous highgrade stenosis. Chronic total occlusions. Several different mechanisms might be implicated in ISA after chronic total occlusion

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

PCI. Stent–vessel mismatch, which occurs because the occluded artery can appear smaller than its actual size due to chronic hypoperfusion, could be responsible for acute ISA, whereas injury to the adventitial layer caused by the guidewire, creation of a false lumen, or subintimal stenting could contribute to late acquired or persistent ISA. Hong et al. (16) identified chronic total occlusion PCI as an independent predictor of DES ISA. In the same study, the incidence of ISA in this scenario was 27.5%. Bifurcations. Coronary artery bifurcations are complex structures in which stent–vessel interactions are dependent on several variables, such as stent strut location (i.e., flow divider or lateral wall) (84), side branch angulation, plaque distribution, or the technique used for stent implantation (i.e., 1 or 2 stents) or post-dilation (i.e., kissing-balloon or separate main vessel and side branch dilations). The difficulty of achieving adequate stent apposition utilizing 2 stent techniques in coronary artery bifurcations was demonstrated in vitro by Murasato et al. (85) and Ormiston et al. (86), and their results were consistent in an in vivo IVUS study that revealed that more than 60% of the ISA was proximal to the carina in lesions treated with the “crush” technique (87). In line with these results, Foin et al. (88) showed, in vitro, that “crush” stenting led to higher rates of ISA compared with the Culotte and T with protrusion technique. Tyczynski et al. (89) used OCT imaging to demonstrate higher rates of ISA in a side branch compared with non-side branch segments after bifurcation stenting. Interestingly, no differences between simple and complex techniques were revealed (89). Several interventions have been proposed aimed at optimizing stent implantation in bifurcations. Distal cell guidewire re-crossing to the side branch led to a reduction of ISA both in vitro and in vivo (90,91). In the latter, OCT-oriented guidewire re-crossing to the side branch significantly reduced rates of ISA compared with angiography-oriented procedures (9.5% vs. 42.3%, respectively; p < 0.0001). Foin et al. (92) compared the use of a final kissing balloon in vitro with a 2-step (main branch and side branch) final balloon dilation technique in bifurcations. Although the investigators demonstrated equivalent rates of ISA in the side branch ostium region for both techniques, they depicted higher rates of ISA at the proximal stent edge with the kissing balloon (30.7  26.4% vs. 2.8  9.6% for 2 step, p ¼ 0.0016) (92), which could be reduced by performing an additional balloon dilation in the proximal stent segment (33.4% vs. 0.6%, respectively, for nonproximal post-dilation and proximal post-dilation; p ¼ 0.02) (93). More recently, Rahman et al. (94) demonstrated that although kissing-balloon inflations led to some amount of proximal main branch stent asymmetry, the improved stent expansion secondary to that maneuver was sufficient to adequately appose stent struts in the majority of cases. Di Mario et al. (95) proposed that stent optimization in bifurcations could be maximized by high-resolution intravascular imaging guidance. A large area of ISA, which was not suspected after angiography-guided stent optimization with a kissing balloon, was seen by

Attizzani et al. Incomplete Stent Apposition

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

3-dimensional OCT imaging in a case of left main coronary artery distal bifurcation intervention recently described by Fujino et al (96). Among 403 patients with IVUS-guided SES implantation in the left main coronary artery (including distal bifurcation), 28 presented with acute ISA; however, no correlations between this pathological phenomenon and adverse cardiovascular events at 2-year clinical follow-up were established (97). So far, however, no clinical benefits have been found with the use of routine intravascular imaging guidance during bifurcation stenting; therefore, further research in this field is mandatory (98). Vessel tapering. Vessel tapering is frequently observed in long coronary lesions, which might lead to ISA due to marked differences in vessel geometry in the same region of interest. Although multiple post-dilations with different balloon sizes is the usual strategy used with the aim of obtaining adequate apposition when stenting distinct vessel dimensions, aggressive post-dilation with balloons disproportionately larger than the nominal stent diameter can lead to stent strut deformation (99) and eventual polymer damage (100,101). This could negatively affect vascular responses. Carrozza et al. (102) described increased stent recoil with larger acute expansion diameters in which excessive DES overstretching left important gaps between the strut cells, which could induce suboptimal antirestenotic drug delivery (103). Importantly, Russo et al. (104) used an animal to demonstrate that vessel overexpansion might be counterproductive because it leads to more intense neointimal proliferation. Therefore, reaching equipoise between adequate (but not excessive) stent expansion and absence of ISA while

Figure 6

1363

avoiding vessel injury is a difficult goal to accomplish in tapered coronary arteries. Potential Alternatives to Reduce ISA in Daily Clinical Practice As previously described, ISA is multifactorial; features directly related to the index intervention and others associated with vascular response to stents are implicated in its pathophysiology. Therefore, it is likely that combined, rather than isolated approaches, are necessary to successfully reduce ISA rates. Although a clear relationship between ISA and adverse cardiovascular events is still controversial, interventional cardiologists must be committed to minimizing acute ISA by optimizing stent choice and techniques of implantation. Some novel devices could be helpful in addition to conventional strategies in this setting. Verheye et al. (105) demonstrated promising results with the use of self-expandable nitinol stents (STENTYS, STENTYS SA, Paris, France) in coronary bifurcations. Interestingly, these devices (either BMS or PES) have small interconnections that enable easy provisional stenting and access to the side branch. Progressive stent expansion was demonstrated by increasing stent area over time, which could potentially reduce rates of persistent and late acquired ISA because the stent conforms to vascular remodeling. Minimal ISA was revealed at 6-month IVUS follow-up. This same device was tested during primary PCI and demonstrated an early reduction of ISA compared with balloon-expandable stents (106). Self-expandable stents could also help reduce ISA in tapered lesions because the same stent could properly

STEMI in a Tapered Vessel Treated With Self-Expandable Stents

(A) An ostial occlusion in the left anterior descending artery (LAD) during ST-segment myocardial infarction (STEMI). After flow restoration (B), marked vessel tapering is observed in the mid LAD (white arrows). Final result after implantation of 2 STENTYS nitinol self-expandable stents in the mid and proximal LAD in overlapping fashion is shown in (C). White arrows identified by letters have their OCT assessments represented with the corresponding letters in E and F; a marked discrepancy in stent areas is shown, whereas distal and proximal references are shown in D and G.

1364

Attizzani et al. Incomplete Stent Apposition

accommodate different vessel dimensions along the lesion (Fig. 6), thereby avoiding complications (e.g., dissections and vessel perforation) in smaller segments while allowing optimal stent apposition in larger segments. Further investigations are necessary to establish whether self-expandable stents could reduce adverse clinical outcomes compared with balloonexpandable stents in all of these clinical settings. More recently, everolimus-eluting bioresorbable vascular scaffolds (Absorb, Abbott Vascular, Santa Clara, California) emerged as a new tool in the armamentarium against coronary artery disease. These novel devices demonstrated promising results in randomized studies (107) and are being progressively incorporated into routine clinical practice in Europe, the Middle East, and Asia (108–110). Bioresorbable vascular scaffolds are more conformable than metallic stents (111), and their struts, including those associated with ISA, are completely resorbed 2 to 3 years after implantation. This feature could potentially eliminate concerns related to very long-term metallic exposure to the coronary circulation in cases of delayed healing (i.e., ISA and noncoverage) (112,113). Whether the implantation of these novel devices can reduce adverse events related to late ISA compared with metallic stents in real-world populations warrants further investigation. Another critical step in reducing late ISA, which is not directly procedure related, is the active search for polymers and antirestenotic drugs that continue to inhibit neointimal hyperplasia and induce less toxicity and inflammatory reactions in the vascular wall, resulting in less positive vascular remodeling. This “perfect equilibrium” between drug and/or polymer combinations has been pursued thoroughly by investigators over recent years in both pre-clinical and clinical trials with promising results in terms of reducing ISA (32,37,114–116). Table 2

Summary of Conclusions

ISA is a multifactorial, morphological feature that is affected by procedure-related elements and by stent-vessel interactions over time. ISA is more commonly identified in DES than in BMS. OCT is superior compared with IVUS in diagnosing ISA. Although an association between ISA and adverse clinical events is strongly suggested by histopathology and in vivo studies, highly selected populations may introduce bias to these findings; therefore, ISA is the most likely 1 feature among several others that might predispose to ST. The magnitude of acute ISA may predict its resolution and impact on vascular healing. Standardization in data reporting and lesion-level assessment with the evaluation of clustering and heterogeneity could ultimately lead to a better understanding of ISA and its impact on clinical outcomes. Particular clinical scenarios such as acute coronary syndromes, OLP stents, bifurcations, vessel tapering, and CTO are more prone to exhibit ISA. Adequate stent sizing, optimal techniques of implantation, and intravascular imaging guidance can effectively reduce ISA rates. Novel scaffolds such as self-expandable stents could play a role in reducing ISA because they can accommodate eventual positive vessel remodeling over time. Finding antirestenotic drugs and/or polymers that cause less inflammation, and therefore, less positive remodeling, is another important aspect aimed at reducing ISA. CTO ¼ chronic total occlusion, DES ¼ drug-eluting stent(s); other abbreviations as in Table 1.

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

Conclusions The main highlights of this comprehensive review are summarized in Table 2. ISA is a multifactorial, pathological feature, and although a clear cause-effect relationship with ST is still unclear, a body of data suggests that it might be one of the contributors in the prothrombotic vascular milieu. Optimal stent sizing and techniques of implantation, together with the use of new technologies, could minimize the occurrence of ISA. Acknowledgments

The authors thank the following experts for their valuable contribution to our paper: Gaku Nakazawa, MD, and Tsutomu Saiki, BMET, from the Tokai University School of Medicine, Kanagawa, Japan, and Yusuke Fujino, MD, from the New Tokyo Hospital, Chiba, Japan. Reprints and correspondence: Dr. Davide Capodanno, Department of Cardiology, Ferrarotto Hospital, University of Catania, Via Citelli 1, 95100 Catania, Italy. E-mail: [email protected].

REFERENCES

1. Bezerra HG, Attizzani GF, Sirbu V, et al. Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention. J Am Coll Cardiol Intv 2013;6:228–36. 2. Takano M, Inami S, Jang IK, et al. Evaluation by optical coherence tomography of neointimal coverage of sirolimus-eluting stent three months after implantation. Am J Cardiol 2007;99:1033–8. 3. Fujino Y, Attizzani GF, Nakamura S, Costa MA, Bezerra HG. Coronary artery aneurysms after sirolimus-eluting stent implantation: multimodality imaging evaluation. J Am Coll Cardiol Intv 2013;6: 423–4. 4. Kume T, Waseda K, Ako J, et al. Intravascular ultrasound assessment of postprocedural incomplete stent apposition. J Invasive Cardiol 2012;24:13–6. 5. Fujino Y, Attizzani GF, Nakamura S, Costa MA, Bezerra HG. Frequency-domain optical coherence tomography assessment of stent constriction 9 months after sirolimus-eluting stent implantation in a highly calcified plaque. J Am Coll Cardiol Intv 2013;6: 204–5. 6. Guo N, Maehara A, Mintz GS, et al. Incidence, mechanisms, predictors, and clinical impact of acute and late stent malapposition after primary intervention in patients with acute myocardial infarction: an intravascular ultrasound substudy of the Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) trial. Circulation 2010;122:1077–84. 7. Ako J, Morino Y, Honda Y, et al. Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis. J Am Coll Cardiol 2005;46:1002–5. 8. Shah VM, Mintz GS, Apple S, Weissman NJ. Background incidence of late malapposition after bare-metal stent implantation. Circulation 2002;106:1753–5. 9. Feres F, Costa JR Jr., Abizaid A. Very late thrombosis after drugeluting stents. Catheter Cardiovasc Interv 2006;68:83–8. 10. Mintz GS, Shah VM, Weissman NJ. Regional remodeling as the cause of late stent malapposition. Circulation 2003;107:2660–3. 11. Siqueira DA, Abizaid AA, Costa J de R, et al. Late incomplete apposition after drug-eluting stent implantation: incidence and potential for adverse clinical outcomes. Eur Heart J 2007;28:1304–9. 12. Ozaki Y, Okumura M, Ismail TF, et al. The fate of incomplete stent apposition with drug-eluting stents: an optical coherence tomographybased natural history study. Eur Heart J 2010;31:1470–6.

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67 13. Gutiérrez-Chico JL, Regar E, Nüesch E, et al. Delayed coverage in malapposed and side-branch struts with respect to well-apposed struts in drug-eluting stents: in vivo assessment with optical coherence tomography. Circulation 2011;124:612–23. 14. Fujino Y, Attizzani GF, Bezerra HG, et al. Serial assessment of vessel interactions after drug-eluting stent implantation in unprotected distal left main coronary artery disease using frequency-domain optical coherence tomography. J Am Coll Cardiol Intv 2013;6:1035–45. 15. Gutiérrez-Chico JL, Wykrzykowska J, Nüesch E, et al. Vascular tissue reaction to acute malapposition in human coronary arteries: sequential assessment with optical coherence tomography. Circ Cardiovasc Interv 2012;5:20–9. 16. Hong MK, Mintz GS, Lee CW, et al. Late stent malapposition after drug-eluting stent implantation: an intravascular ultrasound analysis with long-term follow-up. Circulation 2006;113:414–9. 17. Steinberg DH, Mintz GS, Mandinov L, et al. Long-term impact of routinely detected early and late incomplete stent apposition: an integrated intravascular ultrasound analysis of the TAXUS IV, V, and VI and TAXUS ATLAS workhorse, long lesion, and direct stent studies. J Am Coll Cardiol Intv 2010;3:486–94. 18. Tanabe K, Serruys PW, Degertekin M, et al. Incomplete stent apposition after implantation of paclitaxel-eluting stents or bare metal stents: insights from the randomized TAXUS II trial. Circulation 2005;111:900–5. 19. Cook S, Eshtehardi P, Kalesan B, et al. Impact of incomplete stent apposition on long-term clinical outcome after drug-eluting stent implantation. Eur Heart J 2012;33:1334–43. 20. Hassan AK, Bergheanu SC, Stijnen T, et al. Late stent malapposition risk is higher after drug-eluting stent compared with bare-metal stent implantation and associates with late stent thrombosis. Eur Heart J 2010;31:1172–80. 21. Cook S, Wenaweser P, Togni M, et al. Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation 2007;115:2426–34. 22. Guagliumi G, Ikejima H, Sirbu V, et al. Impact of drug release kinetics on vascular response to different zotarolimus-eluting stents implanted in patients with long coronary stenoses: the LongOCT study (Optical Coherence Tomography in Long Lesions). J Am Coll Cardiol Intv 2011;4:778–85. 23. Kubo T, Imanishi T, Kitabata H, et al. Comparison of vascular response after sirolimus-eluting stent implantation between patients with unstable and stable angina pectoris: a serial optical coherence tomography study. J Am Coll Cardiol Img 2008;1:475–84. 24. Guagliumi G, Musumeci G, Sirbu V, et al. Optical coherence tomography assessment of in vivo vascular response after implantation of overlapping bare-metal and drug-eluting stents. J Am Coll Cardiol Intv 2010;3:531–9. 25. Lee CW, Kang SJ, Park DW, et al. Intravascular ultrasound findings in patients with very late stent thrombosis after either drug-eluting or bare-metal stent implantation. J Am Coll Cardiol 2010;55:1936–42. 26. Guagliumi G, Sirbu V, Musumeci G, et al. Examination of the in vivo mechanisms of late drug-eluting stent thrombosis: findings from optical coherence tomography and intravascular ultrasound imaging. J Am Coll Cardiol Intv 2012;5:12–20. 27. Alfonso F, Dutary J, Paulo M, et al. Combined use of optical coherence tomography and intravascular ultrasound imaging in patients undergoing coronary interventions for stent thrombosis. Heart 2012;98:1213–20. 28. Kang SJ, Lee CW, Song H, et al. OCT analysis in patients with very late stent thrombosis. J Am Coll Cardiol Img 2013;6:695–703. 29. Kosonen P, Vikman S, Jensen LO, et al. Intravascular ultrasound assessed incomplete stent apposition and stent fracture in stent thrombosis after bare metal versus drug-eluting stent treatment: the Nordic Intravascular Ultrasound Study (NIVUS). Int J Cardiol 2013; 168:1010–6. 30. Pesarini G, Dandale R, Rigamonti A, et al. Late and very late coronary stent thrombosis: intravascular ultrasound findings and associations with antiplatelet therapy. Catheter Cardiovasc Interv 2013;82:1056–65. 31. Radu M, Jørgensen E, Kelbæk H, Helqvist S, Skovgaard L, Saunamäki K. Optical coherence tomography at follow-up after percutaneous coronary intervention: relationship between procedural dissections, stent strut malapposition and stent healing. EuroIntervention 2011;7:353–61.

Attizzani et al. Incomplete Stent Apposition

1365

32. Uren NG, Schwarzacher SP, Metz JA, et al. Predictors and outcomes of stent thrombosis: an intravascular ultrasound registry. Eur Heart J 2002;23:124–32. 33. Alfonso F, Suárez A, Angiolillo DJ, et al. Findings of intravascular ultrasound during acute stent thrombosis. Heart 2004;90:1455–9. 34. Gutiérrez-Chico JL, Jüni P, García-García HM, et al. Long term tissue coverage of a biodegradable polylactide polymer-coated biolimus-eluting stent: comparative sequential assessment with optical coherence tomography until complete resorption of the polymer. Am Heart J 2011;162:922–31. 35. Attizzani GF, Bezerra HG, Ormiston J, et al. Serial assessment by optical coherence tomography of early and late vascular responses after implantation of an absorbable-coating sirolimus-eluting stent (from the first-in-human DESSOLVE I trial). Am J Cardiol 2013;112: 1557–6. 36. Tanigawa J, Barlis P, Dimopoulos K, Dalby M, Moore P, Di Mario C. The influence of strut thickness and cell design on immediate apposition of drug-eluting stents assessed by optical coherence tomography. Int J Cardiol 2009;134:180–8. 37. Virmani R, Guagliumi G, Farb A, et al. Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious? Circulation 2004;109:701–5. 38. Joner M, Finn AV, Farb A, et al. Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk. J Am Coll Cardiol 2006;48:193–202. 39. Cook S, Ladich E, Nakazawa G, et al. Correlation of intravascular ultrasound findings with histopathological analysis of thrombus aspirates in patients with very late drug-eluting stent thrombosis. Circulation 2009;120:391–9. 40. Attizzani GF, Bezerra HG, Chamié D, et al. Serial evaluation of vascular response after implantation of a new sirolimus-eluting stent with bioabsorbable polymer (MISTENT): an optical coherence tomography and histopathological study. J Invasive Cardiol 2012;24:560–8. 41. Templin C, Meyer M, Müller MF, et al. Coronary optical frequency domain imaging (OFDI) for in vivo evaluation of stent healing: comparison with light and electron microscopy. Eur Heart J 2010;31: 1792–801. 42. Malle C, Tada T, Steigerwald K, et al. Tissue characterization after drug-eluting stent implantation using optical coherence tomography. Arterioscler Thromb Vasc Biol 2013;33:1376–83. 43. Nakazawa G, Finn AV, Vorpahl M, Ladich ER, Kolodgie FD, Virmani R. Coronary responses and differential mechanisms of late stent thrombosis attributed to first-generation sirolimus- and paclitaxel-eluting stents. J Am Coll Cardiol 2011;57:390–8. 44. Radu M, Jørgensen E, Kelbaek H, Helqvist S, Skovgaard L, Saunamäki K. Strut apposition after coronary stent implantation visualised with optical coherence tomography. EuroIntervention 2010; 6:86–93. 45. Maehara A, Mintz G, Weissman N. Advances in intravascular imaging. Circ Cardiovasc Interv 2009;2:482–90. 46. Mintz GS. What to do about late incomplete stent apposition? Circulation 2007;115:2379–81. 47. Bezerra HG, Costa MA, Guagliumi G, Rollins AM, Simon DI. Intracoronary optical coherence tomography: a comprehensive review. J Am Coll Cardiol Intv 2009;2:1035–46. 48. Fujino Y, Bezerra HG, Attizzani GF, et al. Frequency-domain optical coherence tomography assessment of unprotected left main coronary artery disease–a comparison with intravascular ultrasound. Catheter Cardiovasc Interv 2013;82:E173–83. 49. Mehanna EA, Attizzani GF, Kyono H, Hake M, Bezerra HG. Assessment of coronary stent by optical coherence tomography, methodology and definitions. Int J Cardiovasc Imaging 2011;27:259–69. 50. Prati F, Guagliumi G, Mintz GS, et al. Expert review document part 2: methodology, terminology and clinical applications of optical coherence tomography for the assessment of interventional procedures. Eur Heart J 2012;33:2513–20. 51. Moore P, Barlis P, Spiro J, et al. A randomized optical coherence tomography study of coronary stent strut coverage and luminal protrusion with rapamycin-eluting stents. J Am Coll Cardiol Intv 2009;2: 437–44. 52. Chen BX, Ma FY, Luo W, et al. Neointimal coverage of bare-metal and sirolimus-eluting stents evaluated with optical coherence tomography. Heart 2008;94:566–70.

1366

Attizzani et al. Incomplete Stent Apposition

53. Kim JS, Jang IK, Kim JS, et al. Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents. Heart 2009;95:1907–12. 54. Matsumoto D, Shite J, Shinke T, et al. Neointimal coverage of sirolimus-eluting stents at 6-month follow-up: evaluated by optical coherence tomography. Eur Heart J 2007;28:961–7. 55. Kim BK, Shin DH, Kim JS, et al. Optical coherence tomographybased evaluation of malapposed strut coverage after drug-eluting stent implantation. Int J Cardiovasc Imaging 2012;28:1887–94. 56. Kawamori H, Shite J, Shinke T, et al. Natural consequence of postintervention stent malapposition, thrombus, tissue prolapse, and dissection assessed by optical coherence tomography at mid-term follow-up. Eur Heart J Cardiovasc Imaging 2013;14:865–75. 57. Durán A, Hian SK, Miller DL, Le Heron J, Padovani R, Vano E. A summary of recommendations for occupational radiation protection in interventional cardiology. Catheter Cardiovasc Interv 2013;81:562–7. 58. ACT Investigators. Acetylcysteine for prevention of renal outcomes in patients undergoing coronary and peripheral vascular angiography: main results from the randomized Acetylcysteine for Contrastinduced nephropathy Trial (ACT). Circulation 2011;124:1250–9. 59. Kim JS, Jang IK, Fan C, et al. Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial. J Am Coll Cardiol Intv 2009;2:1240–7. 60. Li S, Wang Y, Gai L, et al. Evaluation of neointimal coverage and apposition with various drug-eluting stents over 12 months after implantation by optical coherence tomography. Int J Cardiol 2013; 162:166–71. 61. Finn AV, Nakazawa G, Joner M, et al. Vascular responses to drug eluting stents: importance of delayed healing. Arterioscler Thromb Vasc Biol 2007;27:1500–10. 62. Räber L, Baumgartner S, Garcia-Garcia HM, et al. Long-term vascular healing in response to sirolimus- and paclitaxel-eluting stents: an optical coherence tomography study. J Am Coll Cardiol Intv 2012; 5:946–57. 63. Attizzani GF, Bezerra HG. Contemporary assessment of stent strut coverage by OCT. Int J Cardiovasc Imaging 2013;29:23–7. 64. Morino Y, Limpijankit T, Honda Y, et al. Late vascular response to repeat stenting for in-stent restenosis with and without radiation: an intravascular ultrasound volumetric analysis. Circulation 2002;105: 2465–8. 65. Guagliumi G, Bezerra HG, Sirbu V, et al. Serial assessment of coronary artery response to paclitaxel-eluting stents using optical coherence tomography. Circ Cardiovasc Interv 2012;5:30–8. 66. Nakazawa G, Finn AV, Joner M, et al. Delayed arterial healing and increased late stent thrombosis at culprit sites after drug-eluting stent placement for acute myocardial infarction patients: an autopsy study. Circulation 2008;118:1138–45. 67. Hong MK, Mintz GS, Lee CW, et al. Incidence, mechanism, predictors, and long-term prognosis of late stent malapposition after bare-metal stent implantation. Circulation 2004;109:881–6. 68. van der Hoeven BL, Liem SS, Dijkstra J, et al. Stent malapposition after sirolimus-eluting and bare-metal stent implantation in patients with ST-segment elevation myocardial infarction: acute and 9-month intravascular ultrasound results of the MISSION! Intervention study. J Am Coll Cardiol Intv 2008;1:192–201. 69. Choi SY, Witzenbichler B, Maehara A, et al. Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy. Circ Cardiovasc Interv 2011;4:239–47. 70. Guagliumi G, Costa MA, Sirbu V, et al. Strut coverage and late malapposition with paclitaxel-eluting stents compared with bare metal stents in acute myocardial infarction: optical coherence tomography substudy of the Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) trial. Circulation 2011;123:274–81. 71. Gonzalo N, Barlis P, Serruys PW, et al. Incomplete stent apposition and delayed tissue coverage are more frequent in drug-eluting stents implanted during primary percutaneous coronary intervention for ST-segment elevation myocardial infarction than in drugeluting stents implanted for stable/unstable angina: insights from optical coherence tomography. J Am Coll Cardiol Intv 2009;2: 445–52.

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67 72. Davlouros PA, Nikokiris G, Karantalis V, et al. Neointimal coverage and stent strut apposition six months after implantation of a paclitaxel eluting stent in acute coronary syndromes: an optical coherence tomography study. Int J Cardiol 2011;151:155–9. 73. van Werkum JW, Heestermans AA, Zomer AC, et al. Predictors of coronary stent thrombosis: the Dutch Stent Thrombosis Registry. J Am Coll Cardiol 2009;53:1399–409. 74. Steg P, James SK, Atar D, et al. ESC Guidelines for the management of acute myocardial infarction in patients presenting with ST-segment elevation. Eur Heart J 2012;33:2569–619. 75. Finn AV, Kolodgie FD, Harnek J, et al. Differential response of delayed healing and persistent inflammation at sites of overlapping sirolimus- or paclitaxel-eluting stents. Circulation 2005;112: 270–8. 76. Shinke T, Li J, Chen JP, et al. High incidence of intramural thrombus after overlapping paclitaxel-eluting stent implantation: angioscopic and histopathologic analysis in porcine coronary arteries. Circ Cardiovasc Interv 2008;1:28–35. 77. Kolandaivelu K, Swaminathan R, Gibson WJ, et al. Stent thrombogenicity early in high-risk interventional settings is driven by stent design and deployment and protected by polymer-drug coatings. Circulation 2011;123:1400–9. 78. Tanigawa J, Barlis P, Dimopoulos K, Di Mario C. Optical coherence tomography to assess malapposition in overlapping drug-eluting stents. EuroIntervention 2008;3:580–3. 79. Aoki J, Mintz GS, Weissman NJ, et al. Chronic arterial responses to overlapping paclitaxel-eluting stents: insights from serial intravascular ultrasound analyses in the TAXUS-V and -VI trials. J Am Coll Cardiol Intv 2008;1:161–7. 80. Otake H, Honda Y, Yamasaki M, et al. Vascular response to overlapping everolimus-eluting stents: comparison with paclitaxel-eluting stents. Circ J 2010;74:1023–5. 81. Dawkins KD, Grube E, Guagliumi G, et al. Safety and efficacy of multiple, overlapping polymer-based paclitaxel-eluting stents. EuroIntervention 2007;3:213–21. 82. Guagliumi G, Capodanno D, Ikejima H, et al. Impact of different stent alloys on human vascular response to everolimus-eluting stent: an optical coherence tomography study: the OCTEVEREST. Catheter Cardiovasc Interv 2013;81:510–8. 83. Tahara S, Bezerra HG, Sirbu V, et al. Angiographic, IVUS and OCT evaluation of the long-term impact of coronary disease severity at the site of overlapping drug-eluting and bare metal stents: a substudy of the ODESSA trial. Heart 2010;96:1574–8. 84. Nakazawa G, Yazdani SK, Finn AV, Vorpahl M, Kolodgie FD, Virmani R. Pathological findings at bifurcation lesions: the impact of flow distribution on atherosclerosis and arterial healing after stent implantation. J Am Coll Cardiol 2010;55:1679–87. 85. Murasato Y, Horiuchi M, Otsuji Y. Three-dimensional modeling of double-stent techniques at the left main coronary artery bifurcation using micro-focus X-ray computed tomography. Catheter Cardiovasc Interv 2007;70:211–20. 86. Ormiston JA, Currie E, Webster MW, et al. Drug-eluting stents for coronary bifurcations: insights into the crush technique. Catheter Cardiovasc Interv 2004;63:332–6. 87. Costa RA, Mintz GS, Carlier SG, et al. Bifurcation coronary lesions treated with the “crush” technique: an intravascular ultrasound analysis. J Am Coll Cardiol 2005;46:599–605. 88. Foin N, Alegria-Barrero E, Torii R, et al. Crush, culotte, T and protrusion: which 2-stent technique for treatment of true bifurcation lesions? Insights from in vitro experiments and micro-computed tomography. Circ J 2013;77:73–80. 89. Tyczynski P, Ferrante G, Moreno-Ambroj C, et al. Simple versus complex approaches to treating coronary bifurcation lesions: direct assessment of stent strut apposition by optical coherence tomography. Rev Esp Cardiol 2010;63:904–14. 90. Foin N, Torii R, Alegria E, et al. Location of side branch access critically affects results in bifurcation stenting: insights from bench modeling and computational flow simulation. Int J Cardiol 2013;168: 3623–8. 91. Alegría-Barrero E, Foin N, Chan PH, et al. Optical coherence tomography for guidance of distal cell recrossing in bifurcation stenting: choosing the right cell matters. EuroIntervention 2012;8:205–13. 92. Foin N, Torii R, Mortier P, et al. Kissing balloon or sequential dilation of the side branch and main vessel for provisional stenting of

Attizzani et al. Incomplete Stent Apposition

JACC Vol. 63, No. 14, 2014 April 15, 2014:1355–67

93. 94.

95. 96.

97.

98.

99.

100.

101. 102. 103.

104. 105. 106.

bifurcations: lessons from micro-computed tomography and computational simulations. J Am Coll Cardiol Intv 2012;5:47–56. Foin N, Secco GG, Ghilencea L, Krams R, Di Mario C. Final proximal post-dilatation is necessary after kissing balloon in bifurcation stenting. EuroIntervention 2011;7:597–604. Rahman S, Leesar T, Cilingiroglu M, et al. Impact of kissing balloon inflation on the main vessel stent volume, area, and symmetry after side-branch dilation in patients with coronary bifurcation lesions: a serial volumetric intravascular ultrasound study. J Am Coll Cardiol Intv 2013;6:923–31. Di Mario C, Iakovou I, van der Giessen WJ, et al. Optical coherence tomography for guidance in bifurcation lesion treatment. EuroIntervention 2010;6 Suppl J:J99–106. Fujino Y, Attizzani GF, Tahara S, et al. Frequency-domain optical coherence tomography assessment of unfavorable kissing-balloon result in unprotected left main intervention. J Am Coll Cardiol Intv 2013;6: 1324–5. Kang SJ, Ahn JM, Song H, et al. Comprehensive intravascular ultrasound assessment of stent area and its impact on restenosis and adverse cardiac events in 403 patients with unprotected left main disease. Circ Cardiovasc Interv 2011;4:562–9. Biondi-Zoccai G, Sheiban I, Romagnoli E, et al. Is intravascular ultrasound beneficial for percutaneous coronary intervention of bifurcation lesions? Evidence from a 4,314-patient registry. Clin Res Cardiol 2011;100:1021–8. Basalus MW, van Houwelingen KG, Ankone MJ, Feijen J, von Birgelen C. Micro-computed tomographic assessment following extremely oversized partial postdilatation of drug-eluting stents. EuroIntervention 2010;6:141–8. Watanabe T, Fujita M, Awata M, et al. Integrity of stent polymer layer after drug-eluting stent implantation: in vivo comparison of sirolimus-, paclitaxel-, zotarolimus- and everolimus-eluting stents. Cardiovasc Interv Ther 2014;29:4–10. Guérin P, Pilet P, Finet G, et al. Drug-eluting stents in bifurcations: bench study of strut deformation and coating lesions. Circ Cardiovasc Interv 2010;3:120–6. Carrozza JP Jr., Hosley SE, Cohen DJ, Baim DS. In vivo assessment of stent expansion and recoil in normal porcine coronary arteries: differential outcome by stent design. Circulation 1999;100:756–60. Foin N, Sen S, Allegria E, et al. Maximal expansion capacity with current DES platforms: a critical factor for stent selection in the treatment of left main bifurcations? EuroIntervention 2013;8: 1315–25. Russo RJ, Silva PD, Yeager M. Coronary artery overexpansion increases neointimal hyperplasia after stent placement in a porcine model. Heart 2007;93:1609–15. Verheye S, Ramcharitar S, Grube E, et al. Six-month clinical and angiographic results of the STENTYS self-apposing stent in bifurcation lesions. EuroIntervention 2011;7:580–7. van Geuns RJ, Tamburino C, Fajadet J, et al. Self-expanding versus balloon-expandable stents in acute myocardial infarction: results from the APPOSITION II study: self-expanding stents in ST-segment

107. 108.

109.

110.

111.

112.

113.

114.

115.

116.

1367

elevation myocardial infarction. J Am Coll Cardiol Intv 2012;5: 1209–19. Serruys P, Ormiston JA, Onuma Y, et al. A bioabsorbable everolimuseluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods. Lancet 2009;373:897–910. Manna AL, Ohno Y, Attizzani GF, Tamburino C. Successful retrograde recanalization of chronic total coronary occlusion with multiple bioresorbable vascular scaffolds (‘full polymer jacket’): initial experience and rationale. Eur Heart J 2013;34:2925. Grasso C, Attizzani GF, Patané M, Ohno Y, Capodanno D, Tamburino C. First-in-human description of everolimus-eluting bioabsorbable vascular scaffold implantation for the treatment of drug-eluting stent failure: insights from optical coherence tomography. Int J Cardiol 2013;168:4490–1. Kajiya T, Liang M, Sharma RK, et al. Everolimus-eluting bioresorbable vascular scaffold (BVS) implantation in patients with STsegment elevation myocardial infarction (STEMI). EuroIntervention 2013;9:501–4. Gomez-Lara J, Garcia-Garcia HM, Onuma Y, et al. A comparison of the conformability of everolimus-eluting bioresorbable vascular scaffolds to metal platform coronary stents. J Am Coll Cardiol Intv 2010; 3:1190–8. Serruys PW, Onuma Y, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes. J Am Coll Cardiol 2011;58:1578–88. Onuma Y, Serruys PW, Perkins LE, et al. Intracoronary optical coherence tomography and histology at 1 month and 2, 3, and 4 years after implantation of everolimus-eluting bioresorbable vascular scaffolds in a porcine coronary artery model: an attempt to decipher the human optical coherence tomography images in the ABSORB trial. Circulation 2010;122:2288–300. Tada N, Virmani R, Grant G, et al. Polymer-free biolimus a9-coated stent demonstrates more sustained intimal inhibition, improved healing, and reduced inflammation compared with a polymer-coated sirolimus-eluting Cypher stent in a porcine model. Circ Cardiovasc Interv 2010;3:174–83. Barlis P, Regar E, Serruys PW, et al. An optical coherence tomography study of a biodegradable vs. durable polymer-coated limuseluting stent: a LEADERS trial sub-study. Eur Heart J 2010;31: 165–76. Guagliumi G, Sirbu V, Musumeci G, et al. Strut coverage and vessel wall response to a new-generation paclitaxel-eluting stent with an ultrathin biodegradable abluminal polymer: Optical Coherence Tomography Drug-Eluting Stent Investigation (OCTDESI). Circ Cardiovasc Interv 2010;3:367–75.

Key Words: drug-eluting stent(s) - incomplete stent apposition intravascular ultrasound - optical coherence tomography - percutaneous coronary intervention - stent malapposition.

Mechanisms, pathophysiology, and clinical aspects of incomplete stent apposition.

Incomplete stent apposition (ISA) is characterized by the lack of contact of at least 1 stent strut with the vessel wall in a segment not overlying a ...
2MB Sizes 2 Downloads 3 Views