Method for rapid MRI quantification of global cerebral metabolic rate of oxygen Suliman Barhoum, Michael C Langham, Jeremy F Magland, Zachary B Rodgers, Cheng Li, Chamith S Rajapakse and Felix W Wehrli A recently reported quantitative magnetic resonance imaging (MRI) method denoted OxFlow has been shown to be able to quantify whole-brain cerebral metabolic rate of oxygen (CMRO2) by simultaneously measuring oxygen saturation (SvO2) in the superior sagittal sinus and cerebral blood flow (CBF) in the arteries feeding the brain in 30 seconds, which is adequate for measurement at baseline but not necessarily in response to neuronal activation. Here, we present an accelerated version of the method (referred to as F-OxFlow) that quantifies CMRO2 in 8 seconds scan time under full retention of the parent method’s capabilities and compared it with its predecessor at baseline in 10 healthy subjects. Results indicate excellent agreement between both sequences, with mean bias of 2.2% (P = 0.18, two-tailed t-test), 3.4% (P = 0.08, two-tailed t-test), and 2.0% (P = 0.56, two-tailed t-test) for SvO2, CBF, and CMRO2, respectively. F-OxFlow’s potential to monitor dynamic changes in SvO2, CBF, and CMRO2 is illustrated in a paradigm of volitional apnea applied to five of the study subjects. The sequence captured an average increase in SvO2, CBF, and CMRO2 of 10.1 ± 2.5%, 43.2 ± 9.2%, and 7.1 ± 2.2%, respectively, in good agreement with literature values. The method may therefore be suited for monitoring alterations in CBF and SvO2 in response to neurovascular stimuli. Journal of Cerebral Blood Flow & Metabolism advance online publication, 13 May 2015; doi:10.1038/jcbfm.2015.96 Keywords: cerebral blood flow measurement; energy metabolism; magnetic resonance; MRI

INTRODUCTION The brain relies almost entirely on aerobic metabolism. Alterations in cerebral metabolic rate of oxygen (CMRO2) may represent an important index of metabolic dysregulation in common brain diseases including Alzheimer’s, Parkinson’s, multiple sclerosis, and Huntington’s disease.1–4 So far, most studies involving measurement of CMRO2 have relied on positron emission tomography.5,6 Another method for CMRO2 quantification is based on measurement of brain oxygen saturation (SvO2) via jugular vein catheterization,7 while the cerebral flow is measured by either optical coherence tomography or Doppler ultrasound. In recent years, some noninvasive magnetic resonance imaging (MRI) approaches have emerged for measuring CMRO2 based on quantifying SvO2 via measurement of blood transverse relaxation time (T2).8–10 The most commonly used approach, T2-relaxation under spin tagging (TRUST), is combined with phase-contrast MRI to quantify the total cerebral blood flow (CBF). TRUST in its various embodiments8,11–14 has been found to be reliable and reproducible and has showed its usefulness in a number of translational patient studies including multiple sclerosis.1 However, its relatively long acquisition time of about 4 minutes limits its utility for assessing dynamic CMRO2 changes in response to physiologic challenges. In recent work by some of the authors, global CMRO2 was measured at 30-second temporal resolution,15 here referred to as OxFlow (as the method quantifies both CBF and oxygen saturation in a single step). OxFlow was designed to quantify superior sagittal sinus (SSS) SvO2 by means of MR susceptometry-based oximetry16,17

and simultaneously CBF in the major arteries feeding the brain, by means of four interleaved gradient-recalled echoes alternating between brain and neck locations, yielding SvO2 and CBF. The 30-second temporal resolution OxFlow was implemented to detect changes in CMRO2 in response to a hypercapnic challenge.18 Measurement of cerebrovascular reactivity can be achieved by means of volitional apnea in the form of a breathhold test (BHT).19 Previous reports showed that BHT is useful to obtain functional information in patients with carotid occlusive disease.20,21 It also simulates the chronic intermittent hypoxia that occurs in obstructive sleep apnea, and thus the BHT may provide a particularly relevant metric of cerebrovascular reactivity in these patients.22 However, OxFlow’s acquisition time of 30 seconds limits the technique’s use for obtaining time-resolved CMRO2 measurements in response to the BHT. More recently, a very fast (3-second temporal resolution) MRI method for whole-brain CMRO2 measurement that quantifies both SvO2 and CBF in the SSS was introduced.23 However, the method requires an initial calibration step to quantify the SSS blood flow (SSS-BF) to total CBF (tCBF) ratio, allowing subsequent derivation of tCBF from continuously-measured SSS-BF. There are several motivations for measuring neurometabolic parameters with high temporal resolution. Among these are blood oxygen level-dependent-based methods to ascertain whether the gas mixture breathing stimuli (hypercapnia and hyperoxia)24 or breath-hold25 are isometabolic and over what time frame. This is achievable by applying the technique via global stimuli. Apnea, as in obstructive sleep apnea is another important area of investigation. Again, this is associated with global variations in oxygen

Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA. Correspondence: Dr FW Wehrli, Laboratory for Structural NMR Imaging, Department of Radiology, University of Pennsylvania, 1 Founders Pavilion, 3400 Spruce Street, Philadelphia, Pennsylvania 19104, USA. E-mail: [email protected] This work was supported by NIH grants R21-HD069390, R01-HL122754, K25-HL111422, and R01-HL109545. Received 23 February 2015; revised 8 April 2015; accepted 13 April 2015

Fast MRI quantification of CMRO2 S Barhoum et al

2

Figure 1. (A) Eight-second temporal resolution OxFlow (F-OxFlow) sequence in which data are collected by alternating between two anatomic locations; (B) sagittal maximum intensity projection indicating the slice locations of SvO2 (orange) and cerebral blood flow (CBF) (blue) measurement; (C) velocity map showing internal carotid arteries (long arrow) and vertebral arteries (short arrow); (D) phase difference image at the level of the superior sagittal sinus (SSS) with region of interest (ROI) indicated for background tissue phase measurement.

metabolism and is therefore assessable via global CMRO2 measurement. The purpose of this work was to design and implement a substantially enhanced OxFlow sequence (henceforth referred to as fast OxFlow or F-OxFlow) that quantifies CMRO2 with fourfold increased temporal resolution (8 seconds) compared with the original implementation,15 and to evaluate its performance at baseline and during a volitional apnea paradigm.

Cerebral metabolic rate of oxygen is quantified by combining venous and arterial oxygenation and CBF using Fick’s law:26 ð1Þ

where Ca is the blood’s oxygen carrying capacity expressed in μmol O2 per 100 mL blood. Ca was calculated for each subject based on their hemoglobin level assuming 1.39 mL of O2 per gram of hemoglobin as the O2 carrying capacity of hemoglobin.27 Global CBF represents the total in-flow through the feeding left and right internal carotid and vertebral arteries. SaO2 and SvO2 are the arterial and SSS venous oxygen saturation, respectively. For healthy subjects SaO2 was assumed to be 98% at baseline but during the volitional apneic paradigm SaO2 was monitored with a digital pulse oximeter clipped to the right middle finger. The MR susceptometry-based oximetry was used to quantify SvO2 in the SSS instead of internal jugular veins to avoid severe static field inhomogeneity due to the proximity of oral cavity and trachea. Previous studies8,28 have shown good agreement between the SvO2 measured in the SSS and internal jugular veins SvO2. Susceptometry-based oximetry relies on measurement of the magnetic susceptibility difference Δχ = χdo Hct (1 − SvO2/100) between the intravascular blood and surrounding tissue, with χdo = 4π (0.27) p.p.m. representing the susceptibility difference (in SI units) between fully deoxygenated and fully oxygenated blood,29 and Hct the hematocrit. By modeling the vessel Journal of Cerebral Blood Flow & Metabolism (2015), 1 – 7

 1  1 ð2Þ ΔB  ΔχBo 3 cos 2 y - 1 ¼ χ do Hctð1 - Sv O2 =100ÞBo 3 cos 2 y - 1 6 6 where θ is the vessel tilt angle relative to the main field Bo. Cerebral blood flow was quantified using a nongated phase-contrast technique as described in Jain et al.15

Pulse Sequence for Cerebral Metabolic Rate of Oxygen Quantification

MATERIALS AND METHODS Principle of Magnetic Resonance Imaging-Based Cerebral Metabolic Rate of Oxygen Quantification

CMRO2 ¼ C a CBFðSa O2 - Sv O2 Þ

as a long paramagnetic cylinder,16,17 the induced magnetic field (ΔB) relative to the surrounding tissue can be approximated as:

The parent OxFlow pulse sequence15 was designed to simultaneously quantify SvO2 in the SSS and average blood flow velocity in internal carotid and vertebral arteries with four interleaves of RF-spoiled gradient-recalled echoes alternating between the two anatomic locations. To increase the temporal resolution of CMRO2 quantification, the OxFlow sequence was modified to three interleaves with a different acquisition order (Figure 1A). The order of interleaves and corresponding slice locations are shown in Figures 1A and 1B. Henceforth, the modified sequence will be referred to as fast OxFlow or F-OxFlow. Pulse sequences were coded in SequenceTree, an open-source integrated software environment, and exported to MRI scanner with a custom-designed export module.30 All but the second interleave were flow compensated. The first and second interleaves were acquired above the carotid bifurcation to quantify CBF by computing the phase difference image (Figure 1C). In the first interleave, the first gradient moment (m1) of the slice-selection gradient (Gz) is nulled whereas in the second interleave the third lobe of Gz is shifted to yield m140. The third interleave acquires three echoes at the level of the SSS. However, the phase difference was computed between the equalpolarity echoes to quantify SSS SvO2. The second RF excitation pulse is applied at the level of the neck 12 ms after the first RF excitation, the third at the level of the SSS 12 ms after the second RF excitation pulse. The scheme of Figure 1A is repeated again 16 ms later. In this manner, spins at the level of the neck are excited with effective repetition times (TRs) of 12 and 28 ms, while at the level of the SSS TR = 40 ms. Experiments were conducted with the following scan parameters: FOV = 208 × 208 mm2, slice thickness = 5 mm, matrix size = 208 × 208. Scan parameters for the first and © 2015 ISCBFM

Fast MRI quantification of CMRO2 S Barhoum et al

3 Table 1. CBF (mL/100 g/min) quantified from pairs of velocity-encoded images (referred to as CBF1 and CBF2) and the relative difference between them for a range of TR values for the third interleave in four subjects obtained with the three-interleave phase-contrast sequence 1

2

3

4

Mean ± s.d.

P-value (t-test)

TR3 = 12 ms CBF1 (mL/100 g/min) CBF2 (mL/100 g/min) Difference (%)

54.7 ± 4.8 54.2 ± 5.3 0.9

44.3 ± 5.1 44.9 ± 5.3 − 1.3

47.6 ± 1.0 47.1 ± 1.0 1.1

38.2 ± 5.2 38.5 ± 1.0 − 0.8

46.2 ± 6.9 46.2 ± 6.5 0.0 ± 1.2

0.97

TR3 = 16 ms CBF1 (mL/100 g/min) CBF2 (mL100 g/min) Difference (%)

54.3 ± 6.6 54.6 ± 6.4 − 0.6

44.2 ± 3.5 44.0 ± 4.0 0.5

46.5 ± 1.9 47.1 ± 1.7 − 1.3

38.5 ± 1.1 39.2 ± 1.1 − 1.8

45.9 ± 6.5 46.2 ± 6.5 − 0.8 ± 1.0

0.21

TR3 = 25 ms CBF1 (mL/100 g/min) CBF2 (mL/100 g/min) Difference (%)

54.0 ± 4.3 54.9 ± 3.9 − 1.7

42.4 ± 3.6 43.7 ± 3.4 −3

46.9 ± 2.6 45.8 ± 2.3 2.4

39.6 ± 2.4 38.6 ± 3.0 2.6

45.7 ± 6.3 45.8 ± 6.8 0.0 ± 2.9

0.96

TR3 = 40 ms CBF1 (mL/100 g/min) CBF2 (mL/100 g/min) Difference (%)

53.1 ± 3.7 54.0 ± 4.7 − 1.7

44.8 ± 3.4 43.2 ± 3.7 3.6

47.5 ± 2.9 46.1 ± 2.2 3

39.0 ± 2.5 39.6 ± 2.7 − 1.5

46.1 ± 5.9 45.7 ± 6.1 0.9 ± 2.8

Subject #

0.59

Abbreviations: CBF, cerebral blood flow; TR, repetition time.

second interleaves: bandwidth = 240 Hz/pixel, VENC = 60 cm/s at baseline and 80 cm/s during the apneic paradigm, flip angle = 15°; for the third interleave, bandwidth = 320 Hz/pixel, echo spacing (ΔTE) between equalpolarity echoes = 7 ms, flip angle = 16°. The difference in the TR between interleave one and two is a potential cause of signal modulation from the inflow effect, which could lead to a systematic error in CBF quantification. Another source of error results from the slight difference in the T1 steady-state characteristics between the acquired signals used for CBF quantifications at interleaves one and two. However, the error associated with the variations in the T1 steady-state characteristics is negligible given the average longitudinal relaxation time of the arterial blood (1,700 ms),31 which is two orders of magnitude longer than difference in TR between interleaves one and two. The potential ramifications of these problems were investigated in four healthy subjects by using a three-interleave phase-contrast sequence. The first and third interleaves were flow encoded (VENC = 60 cm/s) while the second interleave was flow compensated. Velocity maps were obtained from interleaves one and two, and two and three. To investigate potential bias caused by computing velocity maps with different degrees of T1 steady-state mismatch, a range of TR values were chosen for the third interleave (TR3 = 12, 16, 25, and 40 ms), while TR = 12 ms was used for first and second interleaves.

Institutional Review Board of the University of Pennsylvania. Analysis of patient data was performed in accordance with the Declaration of Helsinki. Written consent was obtained from all subjects. Ten healthy human subjects (mean age = 33 ± 6 years, two females and eight males) participated in the study of baseline CMRO2, which consisted of measurement of SSS SvO2 and CBF at the level of the neck using 8second and 30-second temporal resolution OxFlow sequences in the same session with five successive measurements being made for each method. Hct was measured for each subject via a capillary blood sample obtained from the fingertip (Hb 201+, Hemocue, Angelholm, Sweden). The angiogram of the head and neck vessels was generated with a 2D time-of-flight sequence to determine the proper location of the axial plane transecting the SSS at approximately right angle, as well as the slice distal to the carotid bifurcation, and the tilt angle of the SSS with respect to Bo.15 The location of both slices is indicated in the maximum intensity projection sagittal image (Figure 1B). Further, to calculate CMRO2 per 100 gram of brain mass, a T1-weighted 3D MPRAGE (magnetization-prepared rapid gradient-echo) image of 1 mm isotropic spatial resolution was acquired to estimate brain volume for each subject. A brain density of 1.05 g/mL was used to calculate the brain mass.33

Image Processing and Analysis

The apneic paradigm involved three repeated blocks of 32-second normal breathing baseline, 32-second breath hold apnea, and 88-second normal breathing recovery period. The time-course data of the apneic paradigm were averaged over the three repeated blocks to improve SNR. The subjects performed breath-hold at normal end expiration after being instructed to breathe in and breathe out. Arterial oxygen saturation (SaO2) was continuously measured during the apneic paradigm using a digital pulse oximeter. To exclude artifacts from inspiration and expiration during the last 6 seconds of the baseline period,23 the first three data points in the timecourse data of the apneic paradigm were averaged to quantify baseline oximetric parameters value of SvO2, SaO2, and CBF, from which CMRO2 was derived. The average CMRO2 in response to the breath-hold challenge was quantified using the last two data points of the breath-hold challenge and the first point in the recovery period from the time-course data for each subject. In this manner, errors from residual signal modulations during successive inspiration and expiration before the start of the breath-hold period are minimized.23 The temporal delay in the measured SaO2 using the pulse oximeter was corrected by shifting backward in time the measured SaO2 curve for each subject such that the oxygen resaturation in the arterial blood takes place approximately 8 seconds after the end of the breath-hold period.23 This corresponds to the circulatory transport delay time between the lungs and brain.34

For SvO2 quantification, a phase difference image (Figure 1D) was computed from  echoes 1 and 3 in the third interleave as Δϕmap ¼ arg Z 1 Z 3 ¼ γΔB ´ ΔTE, where Z1 and Z3 are the complex pixel values of the two echoes and the asterisk indicates the complex conjugate; γ is the proton gyromagnetic ratio. The static background field inhomogeneity is characterized by a low spatial-frequency modulation across the FOV so its effects can be approximated and reduced with a second-order polynomial.32 After subtracting the polynomial fit from the raw phase difference image Δϕmap, the phase difference (Δϕ) between intravascular blood and surrounding tissue was computed to quantify SvO2 as:   2jΔϕj ´ 100 ð3Þ S v O2 ¼ 1 γχ do ΔTEBo ð cos 2 y - 1=3ÞHct For CBF quantification, a phase difference image (Figure 1C) was computed from the first and second interleaves.

Magnetic Resonance Imaging Experiments All imaging was performed at 3T on a Siemens Tim Trio system (Siemens Medical Solutions, Erlangen, Germany) with a 12-channel head and neck coil combination. Human subject studies were approved by the © 2015 ISCBFM

Quantification of Cerebral Metabolic Rate of Oxygen in Response to An Apneic Challenge

Journal of Cerebral Blood Flow & Metabolism (2015), 1 – 7

Fast MRI quantification of CMRO2 S Barhoum et al

4 Table 2. Oximetric parameters (means ± s.d.) at baseline: SvO2 (%), CBF (mL/100 g/min), CMRO2 (μmol/100 g/min) for 10 subjects obtained with F-OxFlow and OxFlow Subject #

1 2 3 4 5 6 7 8 9 10 Means ± s.d.

F-OxFlow (8-second temporal resolution)

Oxflow (30-second temporal resolution)

SvO2

CBF

CMRO2

SvO2

CBF

CMRO2

71.8 ± 3.9 61.7 ± 0.8 64.6 ± 2.7 67.6 ± 2.4 71.7 ± 0.7 63.2 ± 1.0 69.0 ± 2.1 64.9 ± 3.0 68.0 ± 0.3 68.5 ± 2.9 67.1 ± 3.4

54.3 ± 5.5 38.3 ± 1.5 45.0 ± 3.7 47.7 ± 1.8 53.7 ± 2.0 40.9 ± 1.8 51.4 ± 1.8 45.2 ± 5.0 40.0 ± 1.0 44.4 ± 3.4 46.1 ± 5.6

122 ± 10 135 ± 7 128 ± 8 132 ± 8 130 ± 5 130 ± 6 131 ± 8 112 ± 7 90 ± 1 119 ± 4 123 ± 14

73.1 ± 2.6 61.9 ± 0.5 65.3 ± 2.1 66.2 ± 1.6 70.7 ± 1.4 64.1 ± 2.5 67.6 ± 2.2 61.4 ± 3.4 63.7 ± 2.2 68.5 ± 1.3 66.3 ± 3.8

55.4 ± 3.0 38.2 ± 2.1 45.5 ± 1.5 45.8 ± 2.8 51.8 ± 1.5 42.1 ± 2.4 50.2 ± 3.6 41.5 ± 1.4 37.6 ± 1.3 42.9 ± 3.3 45.1 ± 5.9

121 ± 5 134 ± 6 128 ± 5 133 ± 8 130 ± 6 130 ± 5 133 ± 6 115 ± 9 97 ± 5 114 ± 5 124 ± 12

Abbreviations: CBF, cerebral blood flow; CMRO2, cerebral metabolic rate of oxygen. Paired two-tailed t-test P-values are 0.18, 0.08, and 0.56 for SvO2, CBF, and CMRO2, respectively.

RESULTS Figure 1B shows an example of a sagittal localizer venogram indicating the location of SvO2 and CBF measurements. Representative phase difference images at two anatomic locations for velocity and SvO2 quantification are shown in Figures 1C and 1D, respectively. The CBF values (i.e., CBF1 and CBF2) from both velocity maps quantified for four subjects using the three-interleave phase contrast sequence are listed in Table 1. The average fractional differences between CBF1 and CBF2 for TR3 ranging from 12 to 40 ms were 1.03%, 1.05%, 2.4%, and 2.5 %, i.e., the bias resulting from unequal TR is negligible. Baseline SvO2, CBF, and CMRO2 values quantified with OxFlow and F-OxFlow are listed in Table 2. Average SvO2, CBF, and CMRO2 values across all 10 subjects for F-OxFlow and OxFlow were 67.1 ± 3.4% and 66.3 ± 3.8% (P = 0.18, paired two-tailed t-test), 46.1 ± 5.6 and 45.1 ± 5.9 mL/100 g/min (P = 0.08, paired two-tailed t-test), and 123 ± 14 and 124 ± 12 (μmol/100 g/min) (P = 0.56, paired two-tailed t-test). These values are in good agreement with those reported recently in similar cohorts of young subjects.15,18 Further, a recent study by some of the authors of the current paper showed excellent agreement between susceptometry-based oximetry and TRUST in quantifying venous oxygenation (SvO2).35 It thus seems appropriate to compare F-OxFlow with its parent version. Figures 2A to 2C show correlations of the quantified parameters between the data obtained with the two methods. All three functional parameters were strongly correlated among the two sequences: SvO2 (R2 = 0.76, P o 0.001), CBF (R2 = 0.92, P o 0.0001), and CMRO2 (R2 = 0.96, P o 0.0001). The sensitivity of the sequence in capturing changes in the oximetric parameters in response to a breath-hold challenge is illustrated in Figure 3, which represents the group-averaged timecourse plots over five subjects of the measured parameters SvO2, SaO2, CBF, and CMRO2. The data show the expected increases in CBF and SvO2 along with a slight reduction in SaO2. Average increases in SvO2, CBF, and CMRO2 were 10.1 ± 2.5% (P = 0.0008, two-tailed t-test), 43.2 ± 9.2% (P = 0.0005, two-tailed t-test), and 7.1 ± 2.2% (P = 0.002, two-tailed t-test), respectively, while the average decrease in SaO2 was − 3.4 ± 0.9% (P = 0.0011, two-tailed ttest). The average baseline values, average maximum values, and average change (i.e., increase or decrease) in SvO2, SaO2, CBF, and CMRO2 for the five subjects evaluated are listed in Table 3. Journal of Cerebral Blood Flow & Metabolism (2015), 1 – 7

DISCUSSION An MRI-based method denoted OxFlow has recently been reported for quantifying whole-brain CMRO2 in 30 seconds scan time.15 Although adequate for assessing baseline CMRO2 and in response to hypercapnia,18 the metabolic response to the stimulus often takes place on a shorter time-scale. Here, we designed and implemented a substantially faster version of the parent OxFlow pulse sequence15 that allows for simultaneous quantification of SSS SvO2 and CBF in 8 seconds. The shortened scan time makes the technique potentially suited for monitoring dynamic changes. Comparison of the performance of the two sequences yielded strong correlations for SvO2 (R2 = 0.76, Po 0.001), CBF (R2 = 0.92, Po 0.0001), and CMRO2 (R2 = 0.96, Po 0.0001) (Figure 2). Further, the data indicate excellent agreement between both sequences with mean absolute differences of 2.2%, 3.4%, and 2.0% for SvO2, CBF, and CMRO2, respectively, none of the means being significantly different from each other (P = 0.18, P = 0.08, and P = 0.56 for SvO2, CBF, and CMRO2, respectively). Each measurement was repeated five times consecutively in each session resulting in an average SD in the quantified SvO2, CBF, and CMRO2 of 2.0%, 2.8%, and 6.4%, respectively (Table 2). The data suggest that at baseline the quantified SvO2, CBF, and CMRO2 to be stable physiologic parameters. The data in Figure 3 suggest that F- OxFlow is able to monitor the hemodynamic and metabolic changes occurring during an apnea paradigm. Although Doppler ultrasound36 and nearinfrared spectroscopy37 have provided insight into the physiologic response to apneic events in subjects with obstructive sleep apnea, the latter two techniques detect the relative dynamic changes in either CBF or cerebral tissue oxygenation, rather than CMRO2 in absolute physiologic units. Our results are in good quantitative agreement with those reported for a nearly identical paradigm using a rapid MRI-based technique in which both CBF and SvO2 were measured in the SSS.23 The observed increase in CMRO2 likely arises because breath-hold (i.e., volitional apnea) is a mixed hypercapnic/hypoxic stimulus. In a previous study by some of the authors of the present paper based on a similar susceptometric approach showed that hypercapnia is isometabolic.18 However, T2-based oximetry (i.e., TRUST) showed 5 ± 2% increase of CMRO2 in response to mild hypoxia.10 Therefore, the increase in CMRO2 in response to volitional apnea is plausible, if seen as a combination of isometabolic (hypercapnia) and pro-metabolic (hypoxia) stimuli, as observed in recent work by Rodgers et al.23 © 2015 ISCBFM

Fast MRI quantification of CMRO2 S Barhoum et al

5

Figure 2. Correlations comparing SvO2 (A), cerebral blood flow (CBF) (mL/100 g/min) (B), and calculated cerebral metabolic rate of oxygen (CMRO2) (μmol/100 g/min) (C) between OxFlow (30 seconds temporal resolution) and F-OxFlow (8 seconds temporal resolution) in 10 subjects. Error bars represent standard deviations across five repeats in each scanning session. The dashed lines denote the 95% confidence interval for the linear fit, which encompass the line of identity in each case. Table 3. Summary of the average values of oximetric parameters (means ± s.d.): SvO2 (%), SaO2 (%), AVO2 (%), CBF (mL/100 g/min), and CMRO2 (μmol/100 g/min) at rest state and in response to breath-hold challenge for five subjects obtained with F-OxFlow (P = 0.0008, 0.0005, and 0.002 for SvO2, CBF, and CMRO2, respectively) Subject # SvO2 (%) Average baseline Maximum Increase (%)

Figure 3. Group-averaged time-course SvO2 (%), SaO2 (%), cerebral blood flow (CBF) (mL/100 g/min), and cerebral metabolic rate of oxygen (CMRO2) (μmol/100 g/min) over five subjects during an apneic paradigm. The shaded area represents the volitional apnea period. Average values of the oximetric parameters at baseline were calculated from the first three data points (black-filled squares corresponding to the bracketed section ‘Baseline’) from the timecourse data for each subject. The average CMRO2 in response to volitional apnea is calculated from the three CMRO2 data points (red-filled squares corresponding to the bracketed section ‘EndApnea’) from the time-course data for each subject. Maximum SvO2 and CBF and the minimum SaO2 values were calculated using data points denoted by the yellow-filled squares from the average timecourse data of the five subjects.

To achieve accurate nontriggered PC velocity quantification, it is necessary to minimize the inflow-related blood signal modulation between successive pulse cycles due to variation in blood-flow velocity during the cardiac cycle.38 Of course, as long as TR is long enough for complete spin replacement between successive excitations, no amplitude modulation can occur. This is easily achieved in OxFlow where the slice thickness and effective TR are 5 mm and 70 ms, respectively, i.e., blood water protons will be refreshed for average blood flow velocity as low as 7.2 cm/s. However, in F-OxFlow the duration of the pulse sequence cycle at the level of the flow measurement is much shorter, alternating between 12 and 28 ms. Nevertheless, our data comparing F-OxFlow to the parent pulse sequence (Table 1) indicate nonsignificant bias in the extracted data. These findings are in accordance with results by Bakker et al,38 who showed that for monophasic flow waveforms (such as carotid and vertebral arteries) signal modulation for small flip angles ⩽ 15° is negligible as a source of error in nontriggered phase-contrast velocity measurements. © 2015 ISCBFM

SaO2 (%) Average baseline Minimum Decrease (%) AVO2 (%) Average baseline Minimum Decrease (%)

1

2

3

4

5

Means ± s.d.

65.5

62.3

67.6

64.9

73.4

66.7 ± 4.2

74.8 13.3

70.4 12.2

74 9

71 9

78.9 7.2

73.8 ± 3.4 10.1 ± 2.5

98

99

98

98.5

98

98.3 ± 0.4

94 − 4.2

97 −2

95 − 3.1

95 − 3.6

94 − 4.2

95.0 ± 1.2 − 3.4 ± 0.9

32.5

36.7

30.4

33.6

24.6

31.6 ± 4.5

19.2 27.6 21 25.3 − 51.5 − 28.3 − 36.6 − 28.2

CBF (mL/100 g/min) Average 41.1 baseline Maximum 73.8 Increase (%) 56.9

15.4 21.7 ± 4.9 − 46 − 38.1 ± 10.5

37

44.4

43.5

49.8

43.2 ± 4.7

51.5 32.8

69 43.4

63.4 37.2

79.2 45.6

67.4 ± 10.6 43.2 ± 9.2

CMRO2 (μmol/100 g/min) Average 114 124 baseline Average 119 131 maximum Increase (%) 4.3 5.5

132

128

107

121 ± 10

143

141

116

130 ± 12

8

9.7

8.1

7.1 ±2.2

Abbreviations: CBF, cerebral blood flow; CMRO2, cerebral metabolic rate of oxygen.

In recent work by some of the present authors, a three-second temporal resolution sequence was reported for quantification of CMRO2 by measuring both SvO2 and CBF at the SSS (SSS-BF) instead of in the neck arteries.23 This sequence, although faster by a factor of 2.5, requires running a 12-second phase-contrast reference scan with two-slice interleaves alternated between the SSS and neck arteries at baseline. The reference sequence is required to estimate tCBF from the SSS-BF:CBF ratio at baseline. In Journal of Cerebral Blood Flow & Metabolism (2015), 1 – 7

Fast MRI quantification of CMRO2 S Barhoum et al

6 dynamic applications such as the study of the response to breathhold, it is assumed that SSS-BF:CBF ratio remains constant during the paradigm. This ratio was shown to remain constant for healthy subjects in response to a mixed hypercapnic-hypoxic paradigm.23 However, it is conceivable that this ratio could vary during physiologic challenges in certain disease populations, for example, those with disorders affecting venous hemodynamics. Recently, a simplified version of the OxFlow technique in which, similarly, both SvO2 and tCBF were measured in the SSS, has been applied in a translational study in neonates.39 However, rather than measuring the SSS-BF:CBF ratio it was assumed to be 50%. One potential problem of all methods designed to quantify temporal changes in AVO2 (such as induced apnea) is pulse oximetry used for measuring SaO2. Errors resulting from poor perfusion at the measurement site (e.g. finger) may provide false readings.40 In principle, SaO2 can be quantified in the internal carotids and vertebral arteries using MR susceptometry. However, the presence of the trachea and air spaces in the neck cause severe static field inhomogeneities, making it difficult to measure SaO2 in the neck arteries. Further, it is challenging to detect small desaturation effects since the phase difference of the signal with respect to tissue is small. This is an area of future investigation, especially with the better background field removal methods that now exist. Finally, all image reconstruction and analysis is currently performed offline from the raw k-space data. The logical next step will be to more tightly integrate the processing into the scanner's pipeline and operator console to facilitate its use in a clinical setting. In conclusion, we have introduced an accelerated version of an MRI-based method that allows estimation of global brain CMRO2 at 8-second temporal resolution by simultaneously quantifying CBF and SvO2. The method may be suited for monitoring the response to nonsteady-state stimuli such as apnea. AUTHOR CONTRIBUTIONS MCL, ZBR, and FWW conceived the study and edited the manuscript. MCL, ZBR, JFM, and CL contributed to the development of the MRI pulse sequences. SB analyzed and drafted the manuscript. CSR contributed to data interpretation. All authors reviewed the manuscript critically for intellectual content, and read and approved the final manuscript.

DISCLOSURE/CONFLICT OF INTEREST The authors declare no conflict of interest.

REFERENCES 1 Ge Y, Zhang Z, Lu H, Tang L, Jaggi H, Herbert J et al. Characterizing brain oxygen metabolism in patients with multiple sclerosis with T2-relaxation-under-spintagging MRI. J Cereb Blood Flow Metab 2012; 32: 403–412. 2 Ishii K, Kitagaki H, Kono M, Mori E. Decreased medial temporal oxygen metabolism in Alzheimer's disease shown by PET. J Nucl Med 1996; 37: 1159–1165. 3 Tanaka M, Kondo S, Okamoto K, Hirai S. Cerebral perfusion and oxygen metabolism in Parkinson's disease: positron emission tomographic study using oxygen-15-labeled CO2 and O2. Nihon Rinsho 1997; 55: 218–221. 4 Leenders KL, Frackowiak RS, Quinn N, Marsden CD. Brain energy metabolism and dopaminergic function in Huntington's disease measured in vivo using positron emission tomography. Mov Disord 1986; 1: 69–77. 5 Mintun MA, Raichle ME, Martin WR, Herscovitch P. Brain oxygen utilization measured with O-15 radiotracers and positron emission tomography. J Nucl Med 1984; 25: 177–187. 6 Ito H, Ibaraki M, Kanno I, Fukuda H, Miura S. Changes in cerebral blood flow and cerebral oxygen metabolism during neural activation measured by positron emission tomography: comparison with blood oxygenation level-dependent contrast measured by functional magnetic resonance imaging. J Cereb Blood Flow Metab 2005; 25: 371–377. 7 Mayberg TS, Lam AM. Jugular bulb oximetry for the monitoring of cerebral blood flow and metabolism. Neurosurg Clin N Am 1996; 7: 755–765. 8 Xu F, Ge Y, Lu H. Noninvasive quantification of whole-brain cerebral metabolic rate of oxygen (CMRO2) by MRI. Magn Reson Med 2009; 62: 141–148.

Journal of Cerebral Blood Flow & Metabolism (2015), 1 – 7

9 Chen JJ, Pike GB. Global cerebral oxidative metabolism during hypercapnia and hypocapnia in humans: implications for BOLD fMRI. J Cereb Blood Flow Metab 2010; 30: 1094–1099. 10 Xu F, Liu P, Pascual JM, Xiao G, Lu H. Effect of hypoxia and hyperoxia on cerebral blood flow, blood oxygenation, and oxidative metabolism. J Cereb Blood Flow Metab 2012; 32: 1909–1918. 11 Liu P, Xu F, Lu H. Test-retest reproducibility of a rapid method to measure brain oxygen metabolism. Magn Reson Med 2013; 69: 675–681. 12 Xu F, Uh J, Brier MR, Hart J, Jr., Yezhuvath US, Gu H et al. The influence of carbon dioxide on brain activity and metabolism in conscious humans. J Cereb Blood Flow Metab 2011; 31: 58–67. 13 Lu H, Ge Y. Quantitative evaluation of oxygenation in venous vessels using T2-Relaxation-Under-Spin-Tagging MRI. Magn Reson Med 2008; 60: 357–363. 14 Rodgers ZB, Englund EK, Langham MC, Magland JF, Wehrli FW. Rapid T2- and susceptometry-based CMRO2 quantification with interleaved TRUST (iTRUST). Neuroimage 2015; 106: 441–450. 15 Jain V, Langham MC, Wehrli FW. MRI estimation of global brain oxygen consumption rate. J Cereb Blood Flow Metab 2010; 30: 1598–1607. 16 Haacke EM, Lai S, Reichenbach JR, Kuppusamy K, Hoogenraad FGC, Takeichi H et al. In vivo measurement of blood oxygen saturation using magnetic resonance imaging: a direct validation of the blood oxygen level-dependent concept in functional brain imaging. Hum Brain Mapp 1997; 5: 341–346. 17 Fernandez-Seara MA, Techawiboonwong A, Detre JA, Wehrli FW. MR susceptometry for measuring global brain oxygen extraction. Magn Reson Med 2006; 55: 967–973. 18 Jain V, Langham MC, Floyd TF, Jain G, Magland JF, Wehrli FW. Rapid magnetic resonance measurement of global cerebral metabolic rate of oxygen consumption in humans during rest and hypercapnia. J Cereb Blood Flow Metab 2011; 31: 1504–1512. 19 Ratnatunga C, Adiseshiah M. Increase in middle cerebral artery velocity on breath holding: a simplified test of cerebral perfusion reserve. Eur J Vasc Surg 1990; 4: 519–523. 20 Muller M, Voges M, Piepgras U, Schimrigk K. Assessment of cerebral vasomotor reactivity by transcranial Doppler ultrasound and breath-holding - a comparison with acetazolamide as vasodilatory stimulus. Stroke 1995; 26: 96–100. 21 Silvestrini M, Troisi E, Matteis M, Cupini LM, Caltagirone C. Transcranial Doppler assessment of cerebrovascular reactivity in symptomatic and asymptomatic severe carotid stenosis. Stroke 1996; 27: 1970–1973. 22 Somers VK, White DP, Amin R, Abraham WT, Costa F, Culebras A et al. Sleep apnea and cardiovascular disease. Circulation 2008; 118: 1080–1111. 23 Rodgers ZB, Jain V, Englund EK, Langham MC, Wehrli FW. High temporal resolution MRI quantification of global cerebral metabolic rate of oxygen consumption in response to apneic challenge. J Cereb Blood Flow Metab 2013; 33: 1514–1522. 24 Yablonskiy DA. Cerebral metabolic rate in hypercapnia: controversy continues. J Cereb Blood Flow Metab 2011; 31: 1502–1503. 25 Thomason ME, Foland LC, Glover GH. Calibration of BOLD fMRI using breath holding reduces group variance during a cognitive task. Hum Brain Mapp 2007; 28: 59–68. 26 Kety SS, Schmidt CF. The effects of altered arterial tensions of carbon dioxide and oxygen on cerebral blood flow and cerebral oxygen consumption of normal young men. J Clin Invest 1948; 27: 484–492. 27 West JB. Pulmonary physiology and pathophysiology: an integrated, case-based approach. 2nd ed. Philadelphia, PA: Lippincott Williams & Wilkins, 2007. 28 Jain V, Magland J, Langham M, Wehrli FW. High temporal resolution in vivo blood oximetry via projection-based T(2) measurement. Magn Reson Med 2012; 70: 785–790. 29 Spees WM, Yablonskiy DA, Oswood MC, Ackerman JJ. Water proton MR properties of human blood at 1.5 Tesla: magnetic susceptibility, T(1), T(2), T*(2), and non-Lorentzian signal behavior. Magn Reson Med 2001; 45: 533–542. 30 Magland JF, Li C, Langham MC, Wehrli FW. Pulse sequence programming in a dynamic visual environment: sequencetree. Magn Reson Med 2015 (in press). 31 Lu H, Clingman C, Golay X, van Zijl PC. Determining the longitudinal relaxation time (T1) of blood at 3.0 Tesla. Magn Reson Med 2004; 52: 679–682. 32 Langham MC, Magland JF, Floyd TF, Wehrlil FW. Retrospective correction for induced magnetic field inhomogeneity in measurements of large-vessel hemoglobin oxygen saturation by MR susceptometry. Magn Reson Med 2009; 61: 626–633. 33 Kretschmann HJ, Kammradt G, Krauthausen I, Sauer B, Wingert F. Brain growth in man. Bibl Anat 1986; 28: 1–26. 34 Batzel JJ, Kappel F, Schneditz D, Tran HT. Cardiovascular and respiratory systems: modeling, analysis, and control. Society for Industrial and Applied MathematicsPhiladelphia, PA2007.

© 2015 ISCBFM

Fast MRI quantification of CMRO2 S Barhoum et al

7 35 Barhoum S, Rodgers ZB, Langham M, Magland JF, Li C, Wehrli FW. Comparison of MRI methods for measuring whole-brain venous oxygen saturation. Magn Reson Med 2014 e-pub ahead of print 27 June 20142010.1002/mrm.25336; e-pub ahead of print. 36 Furtner M, Staudacher M, Frauscher B, Brandauer E, Esnaola y Rojas MM, Gschliesser V et al. Cerebral vasoreactivity decreases overnight in severe obstructive sleep apnea syndrome: a study of cerebral hemodynamics. Sleep Med 2009; 10: 875–881. 37 Valipour A, McGown AD, Makker H, O'Sullivan C, Spiro SG. Some factors affecting cerebral tissue saturation during obstructive sleep apnoea. Eur Respir J 2002; 20: 444–450.

© 2015 ISCBFM

38 Bakker CJ, Kouwenhoven M, Hartkamp MJ, Hoogeveen RM, Mali WP. Accuracy and precision of time-averaged flow as measured by nontriggered 2D phasecontrast MR angiography, a phantom evaluation. Magn Reson Imaging 1995; 13: 959–965. 39 Jain V, Buckley EM, Licht DJ, Lynch JM, Schwab PJ, Naim MY et al. Cerebral oxygen metabolism in neonates with congenital heart disease quantified by MRI and optics. J Cereb Blood Flow Metab 2014; 34: 380–388. 40 DeMeulenaere S. Pulse oximetry: uses and limitations. Nurse Pract 2007; 3: 312–317.

Journal of Cerebral Blood Flow & Metabolism (2015), 1 – 7

Method for rapid MRI quantification of global cerebral metabolic rate of oxygen.

A recently reported quantitative magnetic resonance imaging (MRI) method denoted OxFlow has been shown to be able to quantify whole-brain cerebral met...