Review Article

MRI-based methods for quantification of the cerebral metabolic rate of oxygen

Journal of Cerebral Blood Flow & Metabolism 2016, Vol. 36(7) 1165–1185 ! Author(s) 2016 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav DOI: 10.1177/0271678X16643090 jcbfm.sagepub.com

Zachary B Rodgers1,2, John A Detre1,3 and Felix W Wehrli1

Abstract The brain depends almost entirely on oxidative metabolism to meet its significant energy requirements. As such, the cerebral metabolic rate of oxygen (CMRO2) represents a key measure of brain function. Quantification of CMRO2 has helped elucidate brain functional physiology and holds potential as a clinical tool for evaluating neurological disorders including stroke, brain tumors, Alzheimer’s disease, and obstructive sleep apnea. In recent years, a variety of magnetic resonance imaging (MRI)-based CMRO2 quantification methods have emerged. Unlike positron emission tomography – the current ‘‘gold standard’’ for measurement and mapping of CMRO2 – MRI is non-invasive, relatively inexpensive, and ubiquitously available in modern medical centers. All MRI-based CMRO2 methods are based on modeling the effect of paramagnetic deoxyhemoglobin on the magnetic resonance signal. The various methods can be classified in terms of the MRI contrast mechanism used to quantify CMRO2: T2*, T20 , T2, or magnetic susceptibility. This review article provides an overview of MRI-based CMRO2 quantification techniques. After a brief historical discussion motivating the need for improved CMRO2 methodology, current state-of-the-art MRI-based methods are critically appraised in terms of their respective tradeoffs between spatial resolution, temporal resolution, and robustness, all of critical importance given the spatially heterogeneous and temporally dynamic nature of brain energy requirements.

Keywords Magnetic resonance, cerebral hemodynamics, functional MRI, neurophysiology, positron emission tomography, arterial spin labeling Received 25 November 2015; Revised 16 February 2016; Accepted 22 February 2016

Introduction Brain energy requirements Although comprising only 2% of total body mass, the brain accounts for approximately 20% of the body’s total oxygen consumption.1 The brain is almost entirely dependent on oxidative metabolism of glucose to meet its significant energy requirements and is therefore especially susceptible to hypoxia and ischemia. If circulation to the brain is stopped, such as following cardiac arrest, loss of consciousness occurs in seconds, and brain tissue is irreversibly damaged in as little as three minutes.2 Because of the brain’s dependence on oxygen to meet its energy needs, the cerebral metabolic rate of oxygen (CMRO2) provides a direct, quantitative measure of the brain’s energy utilization and represents an important marker of brain tissue viability and function.

The linkage between regional brain activity and changes in cerebral blood flow (CBF), cerebral glucose metabolism (CMRGlu), and CMRO2 was initially studied using evoked activity, which tends to cause highly localized changes in metabolism. While this provides insight into the structural and functional organization

1

University of Pennsylvania Medical Center, Philadelphia, PA, USA Laboratory for Structural, Physiologic, and Functional Imaging, Department of Radiology, Philadelphia, PA, USA 3 Center for Functional Neuroimaging, Department of Neurology, Philadelphia, PA, USA 2

Corresponding author: Felix W Wehrli, Laboratory for Structural, Physiologic, and Functional Imaging, Department of Radiology, 1 Founders Building, 3400 Spruce Street, Philadelphia, PA 19104, USA. Email: [email protected]

1166 of the brain, measurement of such local metabolic changes is challenging, requiring spatial imaging rather than whole-organ measurement. However, the majority of brain blood-flow and metabolism is not associated with dynamic responses to the environment, but rather represents more sustained ‘‘state’’ or ‘‘trait’’ effects. This so-called ‘‘resting-state’’ brain function is organized into distinct distributed networks of brain regions such as the ‘‘default mode’’ network,3 which was originally identified by its elevated brain metabolism at rest compared to task-activated conditions. While the precise cellular mechanisms relating neuronal signaling and cerebral metabolism are still an area of active investigation, it is known that approximately 80% of resting CMRO2 is devoted to active signaling processes, i.e. the propagation of action potentials and the restoration of those potentials after neurotransmission.4 Because CMRO2 is remarkably well conserved in normal physiology, there is much interest in understanding physiologic and pathologic states in which CMRO2 is affected. Although alterations in CMRO2 have been suggested in several common neurologic disorders, these changes appear to be relatively small given the critical importance of oxygen in maintaining tissue viability. Pathological changes in CMRO2 have been studied most intensively in patients with cerebrovascular disease in whom declining CMRO2 despite increased oxygen extraction in the brain is thought to represent the final stage of autoregulatory decompensation.5

Journal of Cerebral Blood Flow & Metabolism 36(7) CBF changes are large enough to be observed by the naked eye, as demonstrated in 1881 by Italian physiologist Angelo Mosso,9 who observed increased regional pulsations in the brains of patients performing mental tasks while undergoing neurosurgery. In 1890, Charles Roy and Charles Sherrington hypothesized that such CBF responses were related to increases in metabolic demand.10 However, quantitative methods for testing such hypotheses did not yet exist. A major breakthrough came in 1945, when Seymour Kety and Carl Schmidt introduced the first method for quantifying CBF and CMRO2 in humans based on the Fick Principle.11–13 In the 1950s through 1970s, Kety and his student, Louis Sokoloff, led the development of autoradiographic methods to create quantitative images of CBF and CMRO2 in animals.14 The advent of positron emission tomography (PET) allowed extension of this autoradiographic work to humans, where it was used to detect regional activations associated with specific mental functions, for example, language processing.15 Seminal work by Peter Fox and Marcus Raichle16,17 demonstrated that in response to mental and motor tasks, local CBF increases in excess of CMRO2 (Figure 1). This mismatch between changes in blood flow and oxygen metabolism during task activation provides the physiological foundation for the blood-oxygen-leveldependent (BOLD) functional magnetic resonance imaging (fMRI) method, which allows observation of brain activity based on MRI-measured changes in local blood oxygenation. The BOLD fMRI effect was first observed by Seiji Ogawa in hypoxic rats in 199018 and

Measuring brain activity: A historical perspective Developing methods for quantification and imaging of brain activity and metabolism has long been a major goal of basic and clinical neuroscience research. One of the major challenges in studying brain function is that the brain is relatively inaccessible to direct measurement. The electrical signals produced by the discharge of neuronal action potentials provide perhaps the most direct measure of brain activity. These signals were first observed by Richard Caton in 1875,6 using galvanometric measurements in animals during chewing and visual stimulation. This work was expanded to humans with the invention of electroencephalography (EEG) by Hans Berger in 1929,7 which allowed noninvasive measurement of brain electrical activity through the scalp. Unfortunately, EEG and magnetoencephalography8 – a complementary technique based on measurement of magnetic fields – have limited spatial resolution and depth penetration. Despite producing relatively small changes in brain oxygen metabolism, motor and mental processes are associated with larger changes in CBF. In fact, these

Figure 1. Early demonstration of uncoupling between blood flow and CMRO2 during visual stimulation using multimodal PET. Despite a large increase in occipital blood flow, corresponding CMRO2 changes are below the level of detection. This mismatch between flow and metabolism provides the basis for the BOLD fMRI technique. Reproduced from Fox et al.,16 with permission.

Rodgers et al.

1167

Figure 2. Early demonstration of the BOLD effect in humans. (a) T1-weighted axial image with several regions of interest (ROIs) indicated by square boxes; (b) T2*-weighted BOLD image at the same slice location; (c) Pseudo-color map of BOLD image intensity changes in response to visual stimulation and (d) corresponding time-course plots of signal intensities from ROIs in (a), indicating restriction of BOLD effect to ROIs (1 and 2) in the visual cortex. Reproduced from Ogawa et al.,19 with permission.

subsequently demonstrated in humans by Ogawa19 (Figure 2) and several other groups.20–22 Increased BOLD contrast with task activation reflects a regional net decrease in venous deoxyhemoglobin due to the proportionally larger changes in CBF as compared to CMRO2 during regional brain activation. Unlike PET-based metabolic imaging, BOLD fMRI does not require injection of radioactive tracers, can be performed on standard clinical imaging systems, and provides higher spatial and temporal resolution. In the past 20 years, use of BOLD fMRI has grown exponentially and provided enormous insights into the spatiotemporal functional or organization of the brain.

Why quantify CMRO2? Though the contribution of BOLD fMRI to basic neuroscience cannot be overstated, the technique has fundamental limitations. The method’s relative simplicity belies the enormous complexity of interpreting the underlying physiologic meaning of BOLD data. BOLD signal changes are predominantly vascular in origin, driven by changes in CBF and cerebral blood volume (CBV) more so than metabolism. The relationship between the BOLD signal and its vascular and metabolic determinants is a complex one, modified by a number of physiologic and external factors. Therefore, the BOLD signal must always be quantified

in terms of changes relative to a baseline state, and it cannot be used to quantify baseline cerebral metabolism. These relative BOLD signal changes can vary significantly within23 and across24 subjects, reducing statistical power in studies aimed at detecting group differences. Group comparisons can also be confounded by vascular effects unrelated to metabolism, for instance, in studies of aging.25,26 Although simultaneous fMRI and intracortical EEG in animals have shown that BOLD signal change correlates with underlying electrical activity,27 BOLD signal onset latency is an order of magnitude greater compared to electrophysiological changes.28 Furthermore, the BOLD signal primarily originates from the post-capillary venous vasculature, which may be far from the site of activation.29 In contrast to BOLD signal, CMRO2 is a direct gauge of brain activity, and can be measured in absolute physiologic units. It is hypothesized that CMRO2 may therefore provide a more sensitive measure of the brain’s neurometabolic state, and improve sensitivity to detect group differences and longitudinal changes. Animal studies have suggested that the CMRO2 response to stimuli exhibits closer spatiotemporal correlation with neuronal activation compared to BOLD signal.28 Accordingly, CMRO2 imaging could provide a more precise and accurate tool for studying the temporal dynamics of neurologic function and spatial

1168

Journal of Cerebral Blood Flow & Metabolism 36(7)

organization of brain functional systems. Developing improved methods for the quantification and mapping of CMRO2 is of great interest to both basic scientists and clinicians. Pre-clinical applications of brain oxygen consumption imaging have been demonstrated in several of the most common neuropathologies, including stroke,30,31 brain tumors,32 and Alzheimer’s disease (AD).33

CMRO2 quantification theory Brain oxygen delivery and consumption In aerobic metabolism, oxygen acts as the final electron acceptor of the electron transport chain, driving the formation of adenosine triphosphate (ATP), the main energy substrate in the body 6O2 þ C6 H12 O6 þ 36ADP þ 36Pi ! 6CO2 þ 6H2 O þ 36ATP

ð1Þ

Oxidative phosphorylation of glucose (and additionally ketone bodies during starvation) is the principal mechanism through which the brain can produce ATP for any appreciable amount of time. While nonoxidative glucose consumption (glycolysis) plays an important role in providing rapid energy during functional activation,16,34 it is considerably less efficient than aerobic metabolism, producing only 2 moles of ATP per mole of glucose compared to 36 moles of ATP. Thus, there is significant evidence that increased energy demand even during transient brain activation is met largely through oxidative metabolism.35 Oxygen is not produced endogenously and must be continuously transported to the brain via the blood. This transport is achieved by passive diffusion of oxygen across the alveolar and capillary membranes of the lungs, where it binds to hemoglobin in erythrocytes, and, to a much lesser extent, is dissolved in blood plasma. The degree of binding of oxygen to hemoglobin is governed by the partial pressure of oxygen, represented by the hemoglobin dissociation curve (Figure 3). In normal physiologic and atmospheric conditions, the oxygen partial pressure (pO2) in the arterial blood is 80–100 mmHg, resulting in an arterial oxygen saturation (Ya) of approximately 98% hemoglobin oxygen saturation (%HbO2). Because pO2 in the brain (and other end-organs) is much lower, oxygen is released in these tissues from hemoglobin and diffuses along a decreasing pO2 gradient: across the capillary membrane, through the cellular membrane of neurons and glial brain cells, and finally into the mitochondria where aerobic metabolism takes place. The oxygen depleted venous blood, with about 1/3 of its hemoglobin desaturated, is returned to the heart and lungs. The different

Figure 3. Hemoglobin (Hb) dissociation curve illustrating the relationship between oxygen partial pressure (pO2) and percent hemoglobin oxygen saturation (%HbO2). Normal values for arterial blood and brain tissue are indicated. The sigmoidal shape of the curve results from cooperative binding of oxygen at the four heme sites of the Hb tetramer, and facilitates unloading of oxygen from blood to brain tissue along the pO2 gradient.

magnetic properties of oxygenated and deoxygenated hemoglobin provide the foundation for MR-based quantification of blood oxygen saturation and CMRO2.

Fick principle for CMRO2 quantification Assuming all oxygen extracted from the blood is used for ATP production, CMRO2 can be quantified via the Fick principle11–13 CMRO2 ¼ Ca  CBF  ðYa  Yv Þ

ð2Þ

where Ya and Yv are the arterial and venous oxygen saturation in %HbO2, CBF is the CBF in mmol per minute per 100 g of brain tissue, Ca is the arterial oxygen content of fully saturated arterial blood (i.e. blood with Ya¼100 %HbO2) in mmol of O2 per mL blood, giving CMRO2 in mmol per minute per 100 g of brain tissue. Ca is a Hb-dependent constant Ca ¼ 0:620559  Hb

ð3Þ

where Hb is in g/dL and the scaling factor is calculated based on a hemoglobin molar mass of 64,458 g/mol.36 Ya can be measured continuously with a digital pulse oximeter, leaving Yv and CBF to be quantified from the MR imaging experiment. YaYv is often called the arteriovenous oxygen difference (AVO2D), and brain

Rodgers et al. oxygen extraction is frequently reported in terms of the OEF, which is equal to AVO2D/Ya. OEF and AVO2D values are very similar for Ya in the normal physiologic range and are sometimes used interchangeably. Kety and Schmidt12,13 were the first to quantify CMRO2 in humans using the Fick Principle. Ya and Yv were measured directly via co-oximetry of arterial and venous blood. To quantify CBF, they integrated the differential concentration of nitrous oxide (N2O) in the arterial and venous blood during continuous inhalation of N2O gas. Though accurate and well validated, the technique is highly invasive, requiring catheterization of the femoral artery and jugular vein. It also provides only a single steady-state global measurement of Yv, CBF, and CMRO2. However, this landmark work laid the foundation for the vast array of CBF, oxygenation, and metabolism measurement and mapping techniques introduced in the subsequent 60 years.

Non-MRI-based CMRO2 quantification methods Optical methods Several optical methods exist for quantification of CBF and oxygen saturation. These methods take advantage of the different wavelengths at which oxyhemoglobin (oHb) and deoxyhemoglobin (dHb) absorb light. Jugular bulb oximetry involves intravenous insertion of a fiber optic probe and determination of Yv from the absorption spectrum of different wavelengths of light. The method can be combined with transcranial Doppler37 measurement of CBF for quantification of CMRO2. While the approach has the advantage of allowing continuous bedside monitoring, the catheter insertion is invasive and prone to possible serious complications such as carotid artery puncture,38 and the technique is prone to errors due to poor catheter tip placement and calibration.39 Optical methods can also be applied to regional measurement of blood flow and oxygenation. Nearinfrared spectroscopy (NIRS)40 involves application of light with wavelength 650–1100 nm, which can penetrate the scalp, skull, and brain to a depth of a few centimeters. The measured absorption spectra are used to estimate relative concentrations of oHb and dHb, which are then used to derive Yv. A complementary optical technique for CBF measurement, diffuse correlation spectroscopy (DCS),41 measures the scattering of near-infrared light in tissue. NIRS and DCS can be combined to quantify local CMRO242; however, the low penetration depth of light inherently limits the approach to measurement of cortical regions near the skull surface. Thus, the technique is best suited to studies involving neonates,43 whose skulls are thinner,

1169 or in animals, where smaller brain sizes and use of cranial windows improve light penetration.

PET PET involves the injection or inhalation of exogenous, positron-emitting radioactive tracers with specific chemical properties. The emitted positrons annihilate with surrounding electrons to produce equal energy 511 keV photons that are emitted in opposite directions and detected by the PET scanner. The position of these detection events can be used to tomographically compute the spatial distribution of annihilation events, and, thus, the spatial distribution of the tracer. PET-based quantification of CMRO244–46 uses intravenous injection of H215O water to quantify CBF and separate inhalation of radioactive 15O2 gas to quantify OEF, which together can be used to quantify CMRO2. Although the method is considered the gold standard for CMRO2 mapping, it has a number of significant limitations. It is invasive, requiring both arterial and venous punctures as well as significant exposure to radiation. Due to the complexity and cost of the protocol, and the need for an on-site cyclotron to produce the 15 O tracers, only a handful of sites around the world are equipped to conduct PET-based CMRO2 studies. While the generated CMRO2 maps are of good quality, the spatial resolution is somewhat coarse (&5 mm3). Finally, the acquisition time for PET-based CMRO2 is on the order of minutes, requiring that measurements be made in steady-state and thus limiting its ability to resolve time-course responses to dynamic stimuli.

MRI-based CMRO2 quantification methods MRI contrast and CMRO2 Unlike PET, MRI is inherently non-invasive and relatively ubiquitous in modern medical centers. In MRI imaging, radiofrequency (RF) pulses are applied to induce precession of nuclear spin magnetic moments in the tissue or object of interest. Electromagnetic induction produces a signal, which decays with a time constant called the effective transverse relaxation rate (R2*). The portion of the signal decay due to static magnetic field inhomogeneity can be removed by spin-echo refocusing, thus isolating the transverse relaxation rate (R2). These two quantities define a third relaxation rate, R20 ¼ R2*  R2, which represents the rate of signal decay due to static magnetic field inhomogeneities. Of note, R20 is only an exponential decay constant when the distribution of the inhomogeneous field is Lorentzian. These relaxation rates are often defined in terms of relaxation times: T2* ¼ 1/R2*,

1170

Journal of Cerebral Blood Flow & Metabolism 36(7)

T2 ¼ 1/R2, and T20 ¼ 1/R20 . The varying chemical and structural properties of tissues have characteristic effects on the time evolution of the MR signal, allowing generation of images with widely varying contrast. The tunability and variability of MRI contrast have made the modality an enormously powerful tool for clinical diagnosis and scientific discovery. A fundamental limitation of MRI is that the degree of polarization of nuclear magnetic moments is quite small, only a few parts per million, and the exponential time constant for repolarization (the T1) is quite long, on the order of seconds for water protons in many tissues of interest. For this reason, tradeoffs between signal-to-noise-ratio (SNR), spatial resolution, and temporal resolution must be considered in almost all MRI applications, including CMRO2 measurement. In principle, MRI is possible with any nuclei possessing non-zero spin magnetic moment. Direct detection of oxygen with 17O MRI is one possible approach to MR-based CMRO2 quantification. However, enriched 17O is enormously expensive, and the detection sensitivity is low. Because the human body is mostly composed of hydrogen-containing water molecules, 1H, the principal isotope of hydrogen (99% abundance), is the nucleus of choice for most human MRI imaging, including CMRO2 techniques. The various 1H-MRI CMRO2 quantification methods are all based on application of the Fick Principle. CBF is typically measured non-invasively using either phase-contrast MRI (PC-MRI)47 in large cerebral vessels or mapped on a voxel-wise basis with arterial spin labeling (ASL).48,49 Ya can be measured with pulse

oximetry or assumed to be near 98 %HbO2 in normal conditions. This leaves quantification of Yv, the crux of most MR-based CMRO2 quantification techniques. Yv is itself of interest in certain applications, for instance, in stroke, where it may provide a marker for potentially salvageable tissue in the ischemic penumbra region.30,31 MR-based Yv quantification takes advantage of the unique magnetic properties of the metalloprotein hemoglobin, first demonstrated by magnetic mass balance experiments conducted by Pauling and Coryell in 1936.50 In the deoxygenated state, the Fe2þ heme iron’s six electrons in the five 3d orbitals are distributed across the eg and t2g orbitals, resulting in four unpaired electrons and a spin S ¼ 2. When the heme iron becomes oxygenated, the ligand field separating the t2g and eg orbitals is increased, making the configuration in which all electrons occupy the three t2g orbitals more energetically favorable, and resulting in spin pairing with a net electron spin S ¼ 0. Thus, only dHb is paramagnetic, whereas oHb is diamagnetic. dHb paramagnetism causes the magnetic susceptibility of blood to scale linearly with dHb concentration, and also has various effects on the relaxation rates of blood and surrounding tissue. The paramagnetism of dHb is exploited in a variety of MR-based techniques to quantify Yv and CMRO2, as summarized in Table 1. These methods can be categorized based on the tissue compartment in which the effects of dHb are modeled (extravascular vs. intravascular) and the MR contrast method used to quantify these effects (T2*, T20 , T2, or susceptibility), as outlined in the sections that follow. Furthermore, the methods

Table 1. Summary of MR-based Yv/CMRO2 quantification methods and their respective features. Signal Origin

Contrast

Spatial Res.

Method

Yv Temp. Res.

Simul. CBF?

Extravascular

T2* T20 T2

Voxel-wise Voxel-wise Global Regional

Calibrated BOLD51 qBOLD52 TRUST53 TRU-PC54 Projection-based T255 QUIXOTIC56 VSEAN57 OxFlow58 Quantitative Venography131 Zhang et al.59

0:03 8:30 0:24 2:50 0:15 27:30 6:18 0:28 15:42 60:00

YES NO NO NO NO NO NO YES NO NO

Intravascular

Voxel-wise Susceptibility

Global Regional Voxel-wise

Note: ‘‘Signal origin’’ is the tissue compartment in which signal used for Yv quantification is modeled. ‘‘Contrast’’ is the MRI contrast mechanism used in the Yv quantification model. ‘‘Spatial Res.’’ is the spatial resolution for Yv quantification in minutes:seconds. ‘‘Method’’ is the name/acronym or authors associated with the published method and the most relevant citation. ‘‘Yv Temp. Res.’’ is the reported approximate temporal resolution for a single Yv measurement (ignoring any requisite planning or calibration scans). ‘‘Simul. CBF?’’ denotes whether the method/pulse sequence measures CBF simultaneously with Yv. Of note, all of the above techniques allow quantification of absolute Yv and CMRO2, whereas calibrated BOLD techniques provide only relative changes in CMRO2 in response to a stimulus. This limitation is removed by some recent BOLD calibration techniques employing multiple gas-mixture breathing calibration,60,61 as discussed in the text.

Rodgers et al.

1171

are distinguished by whether the measurements are made on a global, regional, or voxel-wise basis, and the resultant tradeoff between their spatial and temporal resolutions.

Extravascular T2*-based methods (Calibrated BOLD) Because CMRO2 measurement provides greater physiological insights compared to BOLD contrast alone, the development and application of BOLD fMRI have been paralleled by attempts to resolve CMRO2 from the BOLD signal. This requires modeling the various physiologic factors that contribute to the BOLD signal, schematically illustrated in Figure 4. During neural activation, there is a local increase in both CMRO2 and CBF; however, the increase in CBF is several times greater than what is required to meet the additional CMRO2 demand,17 resulting in a counterintuitive decrease in the OEF and thus [dHb], the local concentration of deoxyhemoglobin in a voxel. CBF also independently increases the venous cerebral blood volume (CBVv) fraction, which acting alone would increase [dHb]. Overall, the CBF washout effect dominates, resulting in a reduction in [dHb] and the characteristic increase in signal intensity in R2*-weighted BOLD fMRI images. The BOLD signal is a simple exponential function of R2* and echo time (TE). Because fractional BOLD signal changes are only a few percent, the exponential function can be linearized BOLD ffi TE  R2 BOLD0

ð4Þ

where the subscript 0 denotes the baseline state and  denotes the change from baseline to activation. R2* can be expressed as R2 ¼ A  ðCBVv  ½dHbv  CBVv0  ½dHbv0 Þ

ð5Þ

where the subscript v denotes the venous blood compartment, and A is a scaling factor that incorporates effects due to vessel geometry, magnetic field strength, and the susceptibility difference between blood and tissue.62,63 The supralinear (i.e.  > 1) effect of [dHb] in equation (5) results from the combined effects of: (1) the linear exponential decay of the signal for spins in the vicinity of large venous vessels as predicted by the static dephasing model of Yablonskiy and Haacke64 and (2) the quadratic exponential decay of the signal for spins in the vicinity of small vessels (capillaries) due to diffusion of water molecules in large local field gradients, somewhat analogous to exchange between two chemical shift sites as predicted by the Luz and

Figure 4. Schematic diagram illustrating the various physiologic contributions relating neural activity to BOLD signal: During neural activation, both CMRO2 and CBF increase; however, the latter far exceeds the additional CMRO2 requirements, thereby resulting in a decrease in OEF (and thus local dHb concentration). Further, CBF independently increases the venous cerebral blood volume (CBVv) fraction, which acting alone would increase [dHb]. Overall, the CBF effect dominates, causing a reduction in [dHb] and concomitant increase in BOLD signal intensity.

Meiboom65 model for fast exchange. Monte Carlo simulations based on the distribution of vessel sizes in the brain have suggested a  value of 1.5 at 1.5 T.63 However, the majority of functional imaging studies in recent years have been conducted at 3.0 T field strength, where  is predicted to be closer to 1.3.66 The traditional physical interpretation of  is somewhat oversimplified. For example, it does not account for intravascular BOLD signal, which is especially significant at lower field strengths. However, at very high field, the intravascular contribution to the BOLD signal is negligible. Recently, Shu et al.67 determined b experimentally in the rat brain at 9.4 T field strength by

1172

Journal of Cerebral Blood Flow & Metabolism 36(7)

systematically varying the blood magnetic susceptibility via administration of a superparamagnetic contrast agent. The authors reported values on the order of 0.8 in the neocortex of rat brains, significantly lower than values used at 3.0 T. Furthermore, while  was found to be quite uniform across different neocortical regions, lower values were measured in sub-cortical regions. This suggests caution must be exercised in treating  as a spatially uniform constant across the entire brain. Equation (5) requires mapping of venous CBV. Although non-invasive techniques for direct quantification of CBV exist, most notably vascular-space-occupancy (VASO) MRI,68 they suffer from low sensitivity, and cannot easily distinguish between arterial and venous CBV. Furthermore, the VASO method requires imaging at a specific blood inversion null point, making whole-brain coverage difficult. Thus, CBV is generally derived from ASL-measured CBF using the Grub power relationship69 

CBV CBF ¼ CBV0 CBF0

 ð6Þ

where  is the Grubb constant. While the original Grubb constant value (0.38) accounts for total CBV changes, because only dHb-containing CBVv, rather than total CBV, determines the BOLD signal, recent work has suggested a lower value for a of 0.1870 or 0.23.71 Like ,  is typically treated as a constant, but may in fact vary between subjects and brain regions.72–74 Combining equations (4) to (6) and invoking the Fick Principle (with Ya approximated to equal 100 %HbO2) gives an expression – known as the Davis model51 – relating BOLD signal changes to CMRO2 and CBF       BOLD CMRO2  CBF  ¼ M  1 BOLD0 CBF0 CMRO2 j0 ð7Þ Here M is the BOLD calibration constant, defined as M ¼ TE  A  CBVvo  ½dHBvo

ð8Þ

Appraisal of equation (8) demonstrates that M is the maximum possible BOLD signal change that would result from total washout of all dHb from a voxel.51,75 This can be appreciated by assuming that in such a maximal BOLD response, the increase in CBF would dominate any changes in CMRO2, thus reducing the entire term in parentheses in equation (7) to unity.

Despite the aforementioned limitations of treating  and  as constants, simulations based on a more complete BOLD signal model suggest that the general form of equation (7) is remarkably valid if the traditional physical interpretations of  and  are relaxed and they are instead treated as fitting constants.76 This heuristic approach suggests lower values for  (0.14) and  (0.91) are optimal. With knowledge of M, subsequent measurement of CBF and BOLD during a functional paradigm allows quantification of fractional changes in CMRO2 by solving equation (7). However, M varies both across subjects and brain regions,77,78 and must therefore be ‘‘calibrated’’. Davis et al. first demonstrated an approach to BOLD calibration via hypercapnic gas-mixture breathing.51 Assuming that hypercapnia does not result in changes in CMRO2 allows further simplification of equation (7)     BOLD CBF  ¼ M  1 BOLD0 CBF0

ð9Þ

M can be solved from equation (9) via simultaneous measurement of BOLD and CBF signal during both baseline and hypercapnic states. Application of hypercapnia calibrated BOLD to functional tasks may provide improved intra- and inter-subject reproducibility compared to traditional BOLD signal methods.79 However, the method has several limitations. Breathing hypercapnic gas can induce breathlessness, which may be especially problematic in patients who are distressed or infirm. Furthermore, the assumption of isometabolism to hypercapnia has been challenged in several recent papers employing direct CMRO2 quantification based on T2.80,81 Finally, due to the large exponent on the relative CBF changes in equation (9), the technique is highly sensitive to noisy ASL-derived CBF data. Recently, an alternative approach to BOLD calibration based on hyperoxic gas-mixture breathing was proposed.82 Hyperoxia is assumed to cause minimal changes in blood flow.83 Following the derivation of the deoxyhemoglobin dilution model75 and assuming minimal hyperoxic CBF changes results in an alternative calibration equation     BOLD ½dHbv  ¼ M  1 BOLD0 ½dHbv0

ð10Þ

where [dHb]v is assumed to vary uniformly across the brain and is quantified via capnographic measurement of end-tidal O2 (EtO2) before and during hyperoxia. This approach avoids the patient discomfort associated with hypercapnia as well as the sensitivity of the model

Rodgers et al. to errors in ASL-derived CBF (as CBF is assumed to remain constant). Unlike hypercapnia calibration (equation (9)), equation (10) can be derived from equation (5) without invoking the Fick Principle, and therefore does not require assuming a particular value for Ya. However, it has the major disadvantage of requiring an assumed baseline Yv value in order to derive [dHb]v from EtO2 measurements. While Yv is relatively uniform across the brain,17 it varies considerably even between healthy subjects.53,58 Finally, hyperoxia may induce a modest reduction in CBF.83 Although this can be incorporated into the model, it is difficult to measure for individual subjects due to the low sensitivity of ASL to small hyperoxic flow changes. ASLbased CBF measurement during hyperoxia is further complicated by the T1 shortening of blood due to dissolved O2. Recently, several groups60,61 have proposed combining multiple gas-mixture breathing paradigms in order to improve the precision of calibrated BOLD, as well as to allow extension of calibrated BOLD to the quantification of not just fractional CMRO2 changes but also resting-state CMRO2 in absolute physiologic units. Example images obtained from such a dual gas approach (hypercapnia and hyperoxia) are shown in Figure 5. In addition to providing M maps for calibration of subsequent task-based fMRI studies, these dual gas methods can also provide resting-state OEF, CBF, and CMRO2 maps. Of course, the need for multiple gas manipulations adds significant complexity to the experimental protocol, and also increases the physiologic assumptions inherent in the calibration process.

1173 As pointed out earlier, the calibration constant M essentially represents the maximal BOLD effect achievable by replacement of all dHb by oHb. Recently, Blockley et al.85 suggested an alternative calibration experiment not relying on gas mixture breathing. Since the majority of the BOLD effect results from R20 -based signal attenuation, the idea is to capture this portion of the signal modulation by voxel-wise measurement of R20 , for instance, by using an asymmetric spin-echo sequence. However, since a small portion of the BOLD signal is due to irreversible spin dephasing, a scaling parameter is required to correct the calibration constants derived in this manner. Furthermore, since R20 is also affected by magnetic field perturbations unrelated to dHb, the accuracy of the method will hinge on how well such extraneous field inhomogeneities can be eliminated. While the calibration constant M has been shown to depend on the anatomic location in the brain, recent work suggests that it may also depend on the resting brain state. Using an R20 -based approach to M mapping in anesthetized rats, Shu et al.86 found M to depend on both the level and method of anesthesia. Calibrated BOLD is the single current CMRO2 quantification technique with both voxel-wise spatial resolution and sufficient temporal resolution for application to functional experiments. However, it has numerous challenges, including a complex experimental setup and the many physiologic assumptions inherent in the signal model and calibration procedures. Thus, significant spatial averaging is often required to reduce noise and recover physiologically plausible CMRO2

Figure 5. Group average (N ¼ 7) maps of resting-state parameters obtained from dual gas (hyperoxia and hypercapnia) calibrated BOLD. Unlike traditional BOLD calibration using a single gas-mixture challenge, dual gas mixture paradigms allow generation of absolute resting-state CBF, OEF, and CMRO2 maps in addition to M maps. Images of M, OEF, CBF, and CMRO2 exhibit plausible patterns of anatomic contrast. Gray–white matter contrast is absent in the map of resting OEF, consistent with previous PET studies. Reproduced from Gauthier and Hoge 84 with permission.

1174

Journal of Cerebral Blood Flow & Metabolism 36(7)

values. BOLD calibration techniques will benefit from higher SNR BOLD/ASL acquisition strategies. Improved results may also be achieved by replacing physiologic assumptions (i.e. isometabolism to hypercapnia) with additional direct measurements of these physiologic parameters. Advances in BOLD calibration techniques are discussed extensively in a recent review.87

Extravascular T20 -based methods Like T2*-based calibrated BOLD, T20 -based methods involve modeling the effects of intravascular dHb on the temporal evolution of extravascular signal. Rather than invoking the semi-empirical model of equation (5), which requires calibration and corresponding physiologic assumptions, the MRI signal behavior is modeled explicitly in terms of known or measurable physical quantities by invoking the so-called static dephasing regime (SDR). In the presence of magnetic field inhomogeneities, spins accumulate phase at different rates determined by the local magnetic field strength, causing phase incoherence and signal decay. Diffusion of spins between sites of different field strengths can also cause signal decay. In the SDR, it is assumed that static dephasing leads to complete signal decay before diffusion dephasing has an appreciable effect. The time behavior of extravascular MR signal in the SDR was first described by Yablonskiy and Haacke.64 By modeling small blood vessels as a network of randomly oriented cylinders, Yv can be expressed in terms of the spin-echo reversible decay rate, R20 4 R02 ¼ CBVv       do  HCT  ð1  Yv Þ  B0 3 ð11Þ where do is the susceptibility difference between fully oxygenated and fully deoxygenated erythrocytes, Hct is the blood hematocrit, and B0 is the main magnetic field strength. It is noted that equation (11) has the same functional form as equation (5); however, the SDR does not consider diffusion effects and thus  ¼ 1. The SDR model also predicts that in a free induction decay (FID) experiment, the time-dependent MR signal, S(t), exhibits different behavior in the short and long time scales related to the characteristic time, tc "

#  2 4 t SðtÞ ¼ Sð0Þ  exp   CBVv  R2  t 3 tc for

t 5 1:5  tc

ð12Þ

  SðtÞ ¼ Sð0Þ  exp½CBVv  exp R2  t for t 4 1:5  tc

ð13Þ

CBVv can be solved by fitting the FID MR signal to equations (12) and (13), leaving measurement of R20 to determine Yv via equation (11). Several spin-echo-based pulse sequences exist for R20 mapping,88–90 all of which involve measuring the MR signal evolution in the vicinity of a spin echo. The method was first applied to human studies by An and Lin,91 and later improved upon by He and Yablonskiy,52 who additionally considered the effects of static field inhomogeneities and signal contributions from cerebrospinal fluid (CSF) and intravascular blood, dubbing the method ‘quantitative BOLD’ (qBOLD). qBOLD-derived resting-state OEF maps are displayed in Figure 6. In subsequent studies, the qBOLD technique was validated relative to pulse oximetry in rats92 and has shown promise in defining the ischemic penumbra region in acute stroke and post-stroke recovery in humans.93 A recent iteration of the method94 suggests that a simulationbased ‘‘fingerprinting’’ approach95 may allow better fitting of acquired data to the qBOLD model. For more detailed discussion, the reader is directed to a recent review.96 Although qBOLD achieves quantitative mapping of Yv without gas calibration, successful implementation is challenged by the need to fit data to a multi-parameter model with many underlying assumptions. While the original signal model defined by equations (11) to (13) does not account for contributions from diffusion, inclusion of diffusive processes as signal modulators has been addressed in more recent work.97 In addition to theoretically improving the accuracy of qBOLDderived Yv, diffusive BOLD models can also provide information on mean vessel size. This approach has shown potential for imaging of tumors,98 which in addition to affecting local metabolism, also tend to be highly vascularized relative to surrounding tissue. It is important to note that because R20 mapping techniques are based on spin-echo sampling, they have relatively long acquisition times compared to echo-planar imaging (EPI)-based T2* mapping (calibrated BOLD). Thus, unlike calibrated BOLD, qBOLD is inherently limited to measurement of steady-state Yv and CMRO2. However, it has the advantage of not requiring gas-mixture breathing calibration.

Intravascular T2-based methods T2-based CMRO2 quantification methods are based on isolation of venous blood T2, which can be related to Yv through theoretical and empirical models. As blood water protons diffuse through the magnetic field inhomogeneities created by dHb-containing erythrocytes,

Rodgers et al.

1175

Figure 6. Representative maps of estimated brain parameters obtained with a high-resolution GESSE (gradient-echo sampling of spin echo)88 pulse sequence. The top leftmost image is a high-resolution anatomic image. The rest of the maps are deoxygenated blood fraction (%), OEF (%), R2 of brain tissue (R2t, s1), interstitial fluid and cerebrospinal fluid (ISF/CSF) volume fraction, ISF/CSF frequency shift (Hz), R20 of brain tissue (s1), and brain deoxyhemoglobin concentration (M). Reproduced from He et al.,52 with permission.

spin-echo irreversible (T2) dephasing occurs in the intravascular space. This decay can be estimated by a multi-spin-echo Carr-Purcell-Meiboom-Gill (CPMG) pulse sequence, containing 180-degree pulses with spacing tCPMG. The Luz and Meiboom65 model for two-site fast exchange can be modified and applied for diffusion of spins between intra- and extra- erythrocyte compartments99–101 to quantify R2 of venous blood  R2, blood ¼ R2, plasma þ Hct  R2, v þ ð1  HctÞ  ð!v Þ2 t  2 CPMG   1   tanh tCPMG 2 ð14Þ where R2,plasma is the relaxation rate of blood plasma, R2,v and !v are the relaxation rate and susceptibility-shift differences, respectively, between the erythrocytes and plasma for exchanging water in the blood,  is the exchange time between frequency-shifted sites, and tCPMG is the time spacing between consecutive 180degree pulses in the CPMG echo train. This complex model can be simplified by combining multiple physical quantities into calibration constants, resulting in a second-order polynomial relating R2 of blood and Yv102 R2,blood ¼ A þ B  ð1  Yv Þ þ C  ð1  Yv Þ2

ð15Þ

where A, B, and C are Hct- and CPMG spacing (tCPMG)-dependent constants which can be determined empirically from blood samples. Such an approach to Yv quantification was first demonstrated by Wright et al.101 in the major thoracic vessels and later demonstrated in cerebral veins in response to visual stimulation.99,103 Figure 7 shows an example calibration curve obtained from ex vivo blood samples. Determining the values of the calibration constants in equation (15) is a major challenge in implementing these T2-based methods. The constants lack specific physical meaning and must be determined empirically from measurements in blood samples using a range of precisely controlled oxygenation and Hct levels. Furthermore, the constants must be derived separately for different field strengths and tCPMG times, ideally using a sequence with a CPMG echo train identical to that used for in vivo T2 mapping. Care must be taken in the calibration experiment to maintain normal blood chemistry (e.g. temperature and pH) and prevent blood settling. Another challenge for T2-based methods is the need to isolate pure blood signal, as partial volume mixing of tissue or CSF will significantly bias measured T2 and Yv. Even in the largest vessels, such as the superior sagittal sinus (SSS), blood signal isolation is non-trivial due to the relatively large voxel sizes required by the fast imaging readouts used in CPMG-based

1176 T2-mapping sequences. Several different approaches have been proposed, as discussed below. Spin-tagging venous blood isolation (TRUST). T2-RelaxationUnder-Spin-Tagging (TRUST)53 isolates venous blood signal via application of spin tagging, similar in principle to ASL-based CBF quantification. Subtraction of images acquired with and without venous blood inversion isolates pure blood signal. Application of various amounts of CPMG T2 weighting prior to imaging

Figure 7. Calibration curve relating blood T2 to Yv and Hct for tCPMG ¼ 10 ms. Each red symbol represents a blood sample with known Hct and Yv obtained from gold-standard blood co-oximetry, with blood T2 measured from MR. The data are used to derive the fitting constants in equation (15), allowing subsequent determination of Yv via measurement of blood T2. Reproduced from Lu et al.,102 with permission.

Journal of Cerebral Blood Flow & Metabolism 36(7) allows for quantification of blood T2. This method has been applied in both the SSS and internal jugular veins, with very comparable values obtained,104 suggesting that the SSS can be used as a surrogate for global Yv. Combining TRUST with PC-MRI CBF quantification allows determination of CMRO2.104 Figure 8 provides a schematic illustration of the TRUST method. Since its introduction, TRUST has been improved in terms of speed and reliability106 and extensively validated.102,105,107 It has been applied widely in physiologic investigations of the effects of hypercapnia,81 hypoxia and hyperoxia,108 caffeine,109 exercise,110 and cognitive training,111 in studies of neonatal development112 and normal aging,113 and in diseases including multiple sclerosis114 and mild cognitive impairment.115 Although providing a robust and reliable approach to global Yv quantification, TRUST temporal resolution for CMRO2 quantification is on the order of minutes, due to the temporally inefficient nature of CPMGbased T2-mapping and the need for a separate measurement of CBF with PC-MRI. This limits the method’s application to steady-state measurement of CMRO2. Phase-contrast venous blood isolation (TRU-PC, projectionbased T2). As an alternative to spin tagging, blood signal can also be isolated through complex difference subtraction of images acquired with different first gradient moments,54,55 in a similar manner to PC-MRI blood flow quantification. An advantage of this approach over TRUST is that it does not require spin tagging in a drainage territory corresponding to a particular vein and can thus be applied to smaller cortical

Figure 8. Illustration of the positioning of T2-relaxation-under-spin-tagging (TRUST) MRI and representative results. (a) Imaging slice (red) and labeling slab (blue) of TRUST MRI scan. Imaging slice was positioned to be parallel to anterior-commissure–posteriorcommissure line and 20 mm above sinus confluence. (b) Typical data of TRUST MRI. Subtraction of control and label images yields pure venous blood signal, which is subject to increasing T2 weightings. (c) Monoexponential fitting of blood signal in superior sagittal sinus as a function of effective echo time (eTE) results in T2 estimation. T2 value can then be converted to venous oxygenation (Yv) via a calibration plot.102 Reproduced from Liu et al.,105 with permission.

Rodgers et al. vessels with arbitrary vessel geometry, as demonstrated using the T2-Relaxation-Under-PhaseContrast (TRU-PC) method.54 A similar technique using a projection-readout was demonstrated to quantify Yv in as little as 15 s.55 However, because the complex difference signal is also dependent on blood flow, multiple averages are required to ensure blood flow effects are removed, limiting application of these phase-contrast-based techniques to situations of steady-state blood flow, and partially negating the advantage of high temporal resolution. While these techniques can provide Yv in regional vessels, the utility of regional Yv quantification based on intravascular (as opposed to voxel-wise) measurement is fundamentally limited. Unlike arterial perfusion territories, venous drainage territories corresponding to particular vessels are poorly defined. Thus, determining the Yv of a given tissue region from the Yv measured within nearby veins is a major challenge. A recently proposed method for quantitative imaging of venous drainage territories aims to address this challenge.116 However, the method’s spatial and temporal resolution are significantly limited by the need to separately spatially encode both the venous drainage vessels as well as their corresponding upstream tissue regions. Velocity selective excitation venous blood isolation (QUIXOTIC, VSEAN). Clever application of velocity selective excitation pulses117 in combination with blood tagging and T2-preparation similar to TRUST can be used to specifically isolate signal from venous blood in the postcapillary venous compartment. Two similar techniques, QUantitative Imaging of eXtraction of Oxygen and TIssue Consumption (QUIXOTIC)56 and Velocity Selective Excitation with Arterial Nulling (VSEAN),57 use this approach to obtain voxel-wise quantification of Yv. However, because post-capillary blood comprises only a few percent of total parenchymal volume, these techniques suffer from low sensitivity and SNR. As such, they require multiple averages and thus have long acquisition times even when being limited to a single acquisition slice. CSF signal contamination is also a concern, and may result in overestimation of Yv values.56

Intravascular susceptibility-based methods While the aforementioned techniques are based on modeling the effect of paramagnetic dHb on transverse relaxation rates, Yv can also be more directly quantified by measurement of blood susceptibility itself. This technique exploits the relative paramagnetism of dHb versus oHb, which causes the susceptibility of whole

1177 blood relative to surrounding tissue, , to be linearly related to Yv118  ¼ Hctðdo ð1  Yv Þ þ oxy Þ

ð16Þ

where Hct is venipuncture-derived hematocrit, and do and oxy are the experimentally determined volume susceptibility differences between fully oxygenated and deoxygenated erythrocytes and fully oxygenated erythrocytes and water, respectively. Values of 4p  0.273 p.p.m and 4p  0.008 p.p.m. (SI units) are used for do and oxy, respectively, based on theoretical calculations119,120 as well as ex vivo calibration experiments.119,121 Of note, the model in equation (16) assumes that the susceptibilities of water, plasma, and tissue are the same to within experimental noise, and also assumes a single value for the mean corpuscular hemoglobin content (MCHC), the concentration of hemoglobin per volume of packed red blood cells. MCHC is known to vary in certain diseases, particularly anemia. Thus, a more complete model is obtained by substituting MCHCnorm/Hb in place of Hct in equation (16), where Hb is the individual subject’s measured Hb, and MCHCnorm is the MCHC value assumed in the derivation of the do and oxy constants. A MCHCnorm value of 33.3 g/dL was used in deriving the above values of do and oxy.121 Although blood susceptibility cannot be measured directly, it induces a local field offset, B, which can be measured with an MRI multi-echo gradient-recalled echo (GRE) field mapping sequence as B ¼  =TE

ð17Þ

where  is the difference in phase accrual between echoes spaced apart by TE in the blood versus surrounding reference tissue, and  is the proton gyromagnetic ratio. Thus, solving for Yv hinges on determining  from the measured B. The relationship between the two quantities can be described in terms of a convolution in the image domain B ¼ d 

ð18Þ

where d ¼ cos2yr/4pr3 is the dipole kernel. Equation (18) can also be expressed as a point-wise multiplication in the Fourier domain FðBÞ ¼ D  X

ð19Þ

where D ¼ 1/3 – kz2/k2 is the dipole kernel in the Fourier domain. Because D contains zeros on a pair of conical surfaces at 54.6 degrees relative to the z-direction,

1178

Journal of Cerebral Blood Flow & Metabolism 36(7)

inversion of the dipole kernel is ill-posed.122 Regularization or conditioning is necessary in order to find a unique solution for  given B. This inversion problem is a major focus of the growing field of quantitative susceptibility mapping (QSM) (reviewed in Wang et al.123). Although inversion of equation (18) is ill-posed in the general case, it can be solved for certain simple susceptibility distributions, including ellipsoids and specifically cylinders, as the latter approximate the geometry of blood vessels. By modeling a vein of interest as a pseudo-infinite cylinder (i.e., with length  diameter) and accounting for field cancellation due to the Lorentz sphere contribution, the relationship between B and  is given by equation (20)124,125 1 B ¼ B0 ð3 cos2  1Þ 6

ð20Þ

where is the vessel angle with respect to B0, and the reference region outside the vessel is chosen to be sufficiently far from the vessel axis such that the induced field is negligible. Combining equations (16), (17), and (20) allows for determination of Yv by measurement of  Yv ¼ 1 

oxy 6ð =TEÞ þ 2 B0 do Hctð3 cos  1Þ do

ð21Þ

Study of equation (21) reveals that for a given Yv, larger vessel angles will result in increasingly smaller phase shifts between the intra and extravascular compartments, thus reducing the precision of Yv quantification. In fact, approaching the magic angle (54.6 degrees), equation (21) becomes indeterminate. For this reason, the model is ideally suited for smaller vessel angles, and theoretical modeling suggests it is

quite precise for vessels with angles less than 30 degrees.126 The model is schematically illustrated in Figure 9. In addition to providing a surrogate for global Yv,55,104 the SSS is long (in terms of length:diameter ratio), relatively straight, and nearly parallel to the B0 field when the subject is lying supine in an MRI scanner, making it an excellent candidate for application of the infinite cylinder model. Application of this susceptometry-based oximetry (SBO) approach to Yv quantification in the SSS has been validated both theoretically126 and with anatomical phantom models.58 Because the method is based on a simple 2D GRE phase mapping sequence, SBO is very fast compared to T2-based methods such as TRUST. Furthermore, it can be naturally interleaved with GRE-based PC-MRI for additional simultaneous quantification of CBF and, therefore, CMRO2. This combined blood Oxygenation and Flow (dubbed ‘OxFlow’) pulse sequence has been used to quantify CMRO2 with 30-s temporal resolution.58 The method was later applied to study the CMRO2 response to hypercapnia, both in healthy subjects127 and neonates with congenital heart disease.128 More recent iterations of the OxFlow method have achieved temporal resolutions as low as two seconds with the aid of view sharing.129 This accelerated OxFlow sequence was applied to study the CMRO2 response to breath-hold apnea in healthy subjects (Figure 10),129 as well as in patients with obstructive sleep apnea (OSA).130 Compared to control subjects, OSA patients were found to have both reduced baseline CMRO2, as well as a reduced ability to maintain CMRO2 during volitional breath holds. Such blunting of the normal CMRO2 response to apnea may in part explain the significant neurologic comorbidities associated with OSA.131

Figure 9. Schematic illustration of a blood vessel approximated by the infinite cylinder model. Due to paramagnetic dHb, MR signal in the blood has additional phase (fblood) compared to surrounding reference tissue (fref). The phase difference (f) and vessel angle (y) are the two imaging-derived parameters used to compute Yv in equation (21).

Rodgers et al. Compared to TRUST and other T2-based Yv techniques, SBO provides a simplified model relating measured signal parameters to Yv and does not require sequence-specific calibration. As mentioned, a major limitation of the technique is the requirement for vessels with appropriate geometry, such as the SSS. Another limitation is that static magnetic field inhomogeneity from non-dHb sources (e.g. air–tissue interfaces) must be removed. This can be accomplished with polynomial fitting,132 but this approach will perform poorly in situations of severe background field variation (e.g. near metal orthodontics).

1179 The second subset of susceptibility-based Yv quantification techniques involves 3D phase mapping of the brain and dipole inversion by means of QSM techniques. This has been approached by: (1) focusing on only the intravascular signal to determine Yv of large and medium sized (MRI-resolvable) veins,133,134 and (2) modeling the susceptibility effects of the small venous blood compartment of tissue to determine voxel-wise Yv.59 The first approach has been used to produce venograms of Yv values in the venous vascular tree (Figure 11),133 whereas the second gives voxel-wise Yv maps similar to the qBOLD technique. In both

Figure 10. High temporal resolution CMRO2 quantification applied to breath-hold apnea in 10 young healthy subjects. (a) Magnitude image with a superior sagittal sinus (SSS) ROI outlined and corresponding velocity and phase-difference (f) maps illustrating an increase in blood flow velocity and decrease in bloodtissue phase difference in response to apnea. Time-course plot of cohortaveraged Ya, Yv, and tCBF (b) and AVO2D, tCBF, and CMRO2 (c) absolute parameter values in response to apnea, with the 30 second apneic period illustrated by the grey bar. Error bars indicate 1 s.d. Across subjects, a small but statistically significant (6.0 3.6%, P ¼ 0.00044) increase in CMRO2 in response to apnea was observed between the baseline (‘‘Base’’) and the end-apnea (‘‘EA’’) periods. Reproduced from Rodgers et al.,129 with permission.

1180

Journal of Cerebral Blood Flow & Metabolism 36(7)

Figure 11. Quantitative oxygenation venograms produced from whole-brain 3D phase mapping and QSM. Color bar indicates intravascular Yv. Venograms are shown from three healthy volunteers. In the first subject, major veins in the brain are labeled, including the superior sagittal sinus (SSS), inferior sagittal sinus (ISS), straight sinus (SS), transverse sinus (TS), and superior anastomic vein (SAV). Reproduced from Fan et al.,133 with permission.

methods, ASL can be used to measure CBF. Acquisition times for the 3D field maps are several minutes, thus, both techniques are limited to measurement of baseline physiology or steady-state stimuli. As with T2-based techniques for small vessel Yv quantification, the venogram approach to Yv-determination requires relating vessel-specific Yv values to surrounding voxels, for which there is currently no feasible approach. As with SBO, vessel angles near the magic angle will have an increasingly small effect on the induced blood-tissue phase difference, which may explain the pronounced correlation between vessel angle and Yv observed with the technique.133 Although the voxel-wise QSM approach avoids these problems by focusing on the microvascular component of tissue, due to the small blood volume fraction of the microvessels, the induced changes in susceptibility for a

given Yv are approximately two orders of magnitude smaller than the corresponding intravascular susceptibility shifts. In fact, the intravascular blood susceptibility model (equation (16)) ignores the effects of such shifts in tissue susceptibility entirely. Furthermore, the voxel-wise method requires imaging during both baseline and an assumed isometabolic stimulus (with caffeine used in Zhang et al.59) in order to separate blood and non-blood susceptibility effects.

Conclusions CMRO2 is a key physiological parameter with the potential to yield important insights into both normal brain function and brain pathophysiology in disease states. Pathophysiological alterations in CMRO2 have been studied in cerebrovascular disease, where

Rodgers et al. reductions in CBF limit the delivery of sufficient oxygen to meet metabolic needs, ultimately leading to cerebral infarction. Alterations in CMRO2 may also occur due to mitochondrial dysfunction, which has been associated with numerous neurodegenerative disorders,135 or may be due to alterations in energy regulation such as in the Warburg effect136 described in cancer. A number of magnetic resonance techniques for measurement and mapping of CMRO2 have emerged during the past two decades. Whole-brain CMRO2 quantification techniques, in particular TRUST and SBO, have been successfully applied in a variety of studies of normal human physiology and disease, suggesting the potential of global CMRO2 as a clinical biomarker. However, even in physiologic and pathologic states, alterations in CMRO2 are small, demanding that techniques have high levels of precision, accuracy, and reproducibility. This is illustrated, for instance, by the disparate results obtained regarding the CMRO2 response to hypercapnia with TRUST81 and SBO.127 As CMRO2 becomes more widely adopted as a clinical biomarker, clinicians need to understand the potential biases and relative robustness of the various approaches, and recent work has begun to explore the relative agreement between methods.80,137,138 In contrast to global CMRO2 methods, voxel-wise mapping techniques – both T2*- and T20 -based – are at an earlier stage of technical development, and there is much space for improvement in terms of both acquisition strategies and theoretical models. Because they do not rely on gas breathing calibration, T20 -based methods may ultimately emerge as the preferred approach to steady-state CMRO2 mapping in clinical populations, for instance, in patients presenting with ischemic stroke. In contrast, T2*-based BOLD calibration methods – though complicated by the need for gas calibration – are unique among all MR-based CMRO2 quantification techniques in that they can achieve both high temporal resolution and voxel-wise spatial resolution. Calibrated BOLD could greatly improve the power of fMRI in longitudinal and group comparison studies by removing confounding by vascular effects unrelated to metabolic differences. Significant progress continues to be made in developing and refining MR-based methods for CMRO2 quantification. Much work remains in addressing critical challenges and defining new avenues of discovery. Funding This work was supported by the following grants from the National Institutes of Health: R21 HD069390, R01 HL109545, R01 HL122754.

1181 Declaration of conflicting interests The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Authors’ contribution Dr. Rodgers drafted this work, Dr. Wehrli advised Dr. Rodgers and edited this work, Dr. Detre reviewed this work.

References 1. Rolfe DF and Brown GC. Cellular energy utilization and molecular origin of standard metabolic rate in mammals. Physiol Rev 1997; 77: 731–758. 2. Safar P. Resuscitation from clinical death: pathophysiologic limits and therapeutic potentials. Crit Care Med 1988; 16: 923–941. 3. Raichle ME, MacLeod AM, Snyder AZ, Powers WJ, Gusnard DA and Shulman GL. A default mode of brain function. Proc Natl Acad Sci USA 2001; 98(2): 676–82. 4. Attwell D and Laughlin SB. An energy budget for signaling in the grey matter of the brain. J Cereb Blood Flow Metabol 2001; 21: 1133–1145. 5. Fisher M and Bastan B. Identifying and utilizing the ischemic penumbra. Neurology 2012; 79(13 Suppl 1): S79–S85. 6. Canton R. The electric currents of the brain. Br Med J 1875; 2: 278. 7. Berger H. U¨ber das Elektrenkephalogramm des Menschen. Archiv fu¨r Psychiatrie und Nervenkrankheiten 1929; 87: 527–570. 8. Cohen D. Magnetoencephalography: evidence of magnetic fields produced by alpha-rhythm currents. Science 1968; 161: 784–786. 9. Mosso A. Ueber den Kreislauf des Blutes im Menschlichen Gehirn: Von Veit. Leipzig: Veit, 1881. 10. Roy CS and Sherrington CS. On the regulation of the blood-supply of the brain. J Physiol 1890; 11: 85–108. 11. Adolf F. Ueber die Messung des Blutquantums in den Herzventrikeln. Verh Phys Med Ges Wurzburg 1870; 2: 16–28. 12. Kety SS and Schmidt CF. The determination of cerebral blood flow in man by the use of nitrous oxide in low concentrations. Am J Physiol 1945; 143: 55–63. 13. Kety SS and Schmidt CF. The nitrous oxide method for the quantitative determination of cerebral blood flow in man: theory, procedure and normal values. J Clin Invest 1948; 27: 476–483. 14. Sokoloff L, Reivich M, Kennedy C, et al. The [14C]deoxyglucose method for the measurement of local cerebral glucose utilization: theory, procedure, and normal values in the conscious and anesthetized albino rat. J Neurochem 1977; 28: 897–916. 15. Petersen SE, Fox PT, Posner MI, et al. Positron emission tomographic studies of the cortical anatomy of singleword processing. Nature 1988; 331: 585–589. 16. Fox PT, Raichle ME, Mintun MA, et al. Nonoxidative glucose consumption during focal physiologic neural activity. Science 1988; 241: 462–464.

1182 17. Fox PT and Raichle ME. Focal physiological uncoupling of cerebral blood flow and oxidative metabolism during somatosensory stimulation in human subjects. Proc Natl Acad Sci U S A 1986; 83: 1140–1144. 18. Ogawa S, Lee TM, Kay AR, et al. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc Natl Acad Sci U S A 1990; 87: 9868–9872. 19. Ogawa S, Tank DW, Menon R, et al. Intrinsic signal changes accompanying sensory stimulation: functional brain mapping with magnetic resonance imaging. Proc Natl Acad Sci U S A 1992; 89: 5951–5955. 20. Frahm J, Bruhn H, Merboldt KD, et al. Dynamic MR imaging of human brain oxygenation during rest and photic stimulation. J Magn Reson Imaging 1992; 2: 501–505. 21. Bandettini PA, Wong EC, Hinks RS, et al. Time course EPI of human brain function during task activation. Magn Reson Med 1992; 25: 390–397. 22. Kwong KK, Belliveau JW, Chesler DA, et al. Dynamic magnetic resonance imaging of human brain activity during primary sensory stimulation. Proc Natl Acad Sci U S A 1992; 89: 5675–5679. 23. Aguirre GK, Zarahn E and D’Esposito M. The variability of human, BOLD hemodynamic responses. Neuroimage 1998; 8: 360–369. 24. Handwerker DA, Ollinger JM and D’Esposito M. Variation of BOLD hemodynamic responses across subjects and brain regions and their effects on statistical analyses. Neuroimage 2004; 21: 1639–1651. 25. Ances BM, Liang CL, Leontiev O, et al. Effects of aging on cerebral blood flow, oxygen metabolism, and blood oxygenation level dependent responses to visual stimulation. Hum Brain Mapp 2009; 30: 1120–1132. 26. D’Esposito M, Zarahn E, Aguirre GK, et al. The effect of normal aging on the coupling of neural activity to the bold hemodynamic response. Neuroimage 1999; 10: 6–14. 27. Logothetis NK, Pauls J, Augath M, et al. Neurophysiological investigation of the basis of the fMRI signal. Nature 2001; 412: 150–157. 28. Malonek D and Grinvald A. Interactions between electrical activity and cortical microcirculation revealed by imaging spectroscopy: implications for functional brain mapping. Science 1996; 272: 551–554. 29. Menon RS, Ogawa S, Hu X, et al. BOLD based functional MRI at 4 Tesla includes a capillary bed contribution: echo-planar imaging correlates with previous optical imaging using intrinsic signals. Magn Reson Med 1995; 33: 453–459. 30. Baron JC, Bousser MG, Rey A, et al. Reversal of focal ‘‘misery-perfusion syndrome’’ by extra-intracranial arterial bypass in hemodynamic cerebral ischemia. A case study with 15O positron emission tomography. Stroke 1981; 12: 454–459. 31. Sobesky J, Zaro WO, Lehnhardt FG, et al. Does the mismatch match the penumbra? Magnetic resonance imaging and positron emission tomography in early ischemic stroke. Stroke 2005; 36: 980–985. 32. Miles KA and Williams RE. Warburg revisited: imaging tumour blood flow and metabolism. Cancer Imaging 2008; 8: 81–86.

Journal of Cerebral Blood Flow & Metabolism 36(7) 33. Ishii K, Kitagaki H, Kono M, et al. Decreased medial temporal oxygen metabolism in Alzheimer’s disease shown by PET. J Nucl Med 1996; 37: 1159–1165. 34. Vaishnavi SN, Vlassenko AG, Rundle MM, et al. Regional aerobic glycolysis in the human brain. Proc Natl Acad Sci U S A 2010; 107: 17757–17762. 35. Hoge RD, Atkinson J, Gill B, et al. Linear coupling between cerebral blood flow and oxygen consumption in activated human cortex. Proc Natl Acad Sci U S A 1999; 96: 9403–9408. 36. Toffaletti J and Zijlstra WG. Misconceptions in reporting oxygen saturation. Anesth Analg 2007; 105(6 Suppl): S5–S9. 37. Aaslid R, Markwalder TM and Nornes H. Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral arteries. J Neurosurg 1982; 57: 769–774. 38. Coplin WM, O’Keefe GE, Grady MS, et al. Thrombotic, infectious, and procedural complications of the jugular bulb catheter in the intensive care unit. Neurosurgery 1997; 41: 101–107; discussion 107–109. 39. Sheinberg M, Kanter MJ, Robertson CS, et al. Continuous monitoring of jugular venous oxygen saturation in head-injured patients. J Neurosurg 1992; 76: 212–217. 40. Jobsis FF. Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 1977; 198: 1264–1267. 41. Boas DA, Campbell LE and Yodh AG. Scattering and Imaging with diffusing temporal field correlations. Phys Rev Lett 1995; 75: 1855–1858. 42. Verdecchia K, Diop M, Lee TY, et al. Quantifying the cerebral metabolic rate of oxygen by combining diffuse correlation spectroscopy and time-resolved near-infrared spectroscopy. J Biomed Opt 2013; 18: 27007. 43. Durduran T, Zhou C, Buckley EM, et al. Optical measurement of cerebral hemodynamics and oxygen metabolism in neonates with congenital heart defects. J Biomed Opt 2010; 15: 037004. 44. Frackowiak RS, Lenzi GL, Jones T, et al. Quantitative measurement of regional cerebral blood flow and oxygen metabolism in man using 15O and positron emission tomography: theory, procedure, and normal values. J Comput Assist Tomogr 1980; 4: 727–736. 45. Mintun MA, Raichle ME, Martin WR, et al. Brain oxygen utilization measured with O-15 radiotracers and positron emission tomography. J Nucl Med 1984; 25: 177–187. 46. Ito H, Ibaraki M, Kanno I, et al. Changes in cerebral blood flow and cerebral oxygen metabolism during neural activation measured by positron emission tomography: comparison with blood oxygenation leveldependent contrast measured by functional magnetic resonance imaging. J Cereb Blood Flow Metabol 2005; 25: 371–377. 47. Moran PR. A flow velocity zeugmatographic interlace for NMR imaging in humans. Magn Reson Imaging 1982; 1: 197–203.

Rodgers et al. 48. Detre JA, Leigh JS, Williams DS, et al. Perfusion imaging. Magn Reson Med 1992; 23: 37–45. 49. Williams DS, Detre JA, Leigh JS, et al. Magnetic resonance imaging of perfusion using spin inversion of arterial water. Proc Natl Acad Sci U S A 1992; 89: 212–216. 50. Pauling L and Coryell CD. The magnetic properties and structure of hemoglobin, oxyhemoglobin and carbonmonoxyhemoglobin. Proc Natl Acad Sci U S A 1936; 22: 210–216. 51. Davis TL, Kwong KK, Weisskoff RM, et al. Calibrated functional MRI: Mapping the dynamics of oxidative metabolism. Proc Natl Acad Sci 1998; 95: 1834–1839. 52. He X and Yablonskiy DA. Quantitative BOLD: mapping of human cerebral deoxygenated blood volume and oxygen extraction fraction: default state. Magn Reson Med 2007; 57: 115–126. 53. Lu H and Ge Y. Quantitative evaluation of oxygenation in venous vessels using T2-relaxation-under-spin-tagging MRI. Magn Reson Med 2008; 60: 357–363. 54. Krishnamurthy LC, Liu P, Ge Y, et al. Vessel-specific quantification of blood oxygenation with T2-relaxationunder-phase-contrast MRI. Magn Reson Med 2014; 71: 978–989. 55. Jain V, Magland J, Langham M, et al. High temporal resolution in vivo blood oximetry via projection-based T2 measurement. Magn Reson Med 2013; 70: 785–790. 56. Bolar DS, Rosen BR, Sorensen AG, et al. QUantitative Imaging of eXtraction of oxygen and TIssue consumption (QUIXOTIC) using venular-targeted velocityselective spin labeling. Magn Reson Med 2011; 66(6): 1550–1562. 57. Guo J and Wong EC. Venous oxygenation mapping using velocity-selective excitation and arterial nulling. Magn Reson Med 2012; 68: 1458–1471. 58. Jain V, Langham MC and Wehrli FW. MRI estimation of global brain oxygen consumption rate. J Cereb Blood Flow Metabol 2010; 30: 1598–607. 59. Zhang J, Liu T, Gupta A, et al. Quantitative mapping of cerebral metabolic rate of oxygen (CMRO) using quantitative susceptibility mapping (QSM). Magn Reson Med 2015; 74: 945–952. 60. Bulte DP, Kelly M, Germuska M, et al. Quantitative measurement of cerebral physiology using respiratorycalibrated MRI. Neuroimage 2012; 60: 582–591. 61. Gauthier CJ and Hoge RD. A generalized procedure for calibrated MRI incorporating hyperoxia and hypercapnia. Hum Brain Mapp 2013; 34: 1053–1069. 62. Ogawa S, Menon RS, Tank DW, et al. Functional brain mapping by blood oxygenation level-dependent contrast magnetic resonance imaging. A comparison of signal characteristics with a biophysical model. Biophys J 1993; 64: 803–812. 63. Boxerman JL, Hamberg LM, Rosen BR, et al. MR contrast due to intravascular magnetic susceptibility perturbations. Magn Reson Med 1995; 34: 555–566. 64. Yablonskiy DA and Haacke EM. Theory of NMR signal behavior in magnetically inhomogeneous tissues: the static dephasing regime. Magn Reson Med 1994; 32: 749–763.

1183 65. Luz Z and Meiboom S. Nuclear magnetic resonance study of the protolysis of trimethylammonium ion in aqueous solution. J Chem Phys 1963; 39: 366–370. 66. Mark CI, Fisher JA and Pike GB. Improved fMRI calibration: precisely controlled hyperoxic versus hypercapnic stimuli. Neuroimage 2011; 54: 1102–1111. 67. Shu CY, Sanganahalli BG, Coman D, et al. Quantitative beta mapping for calibrated fMRI. Neuroimage 2015; 126: 219–228. 68. Lu H, Golay X, Pekar JJ, et al. Functional magnetic resonance imaging based on changes in vascular space occupancy. Magn Reson Med 2003; 50: 263–274. 69. Grubb RL Jr, Raichle ME, Eichling JO, et al. The effects of changes in PaCO2 on cerebral blood volume, blood flow, and vascular mean transit time. Stroke 1974; 5: 630–639. 70. Chen JJ and Pike GB. MRI measurement of the BOLDspecific flow-volume relationship during hypercapnia and hypocapnia in humans. Neuroimage 2010; 53: 383–391. 71. Chen JJ and Pike GB. BOLD-specific cerebral blood volume and blood flow changes during neuronal activation in humans. NMR Biomed 2009; 22: 1054–1062. 72. Rostrup E, Knudsen GM, Law I, et al. The relationship between cerebral blood flow and volume in humans. Neuroimage 2005; 24: 1–11. 73. Ciris PA, Qiu M and Constable RT. Noninvasive MRI measurement of the absolute cerebral blood volume-cerebral blood flow relationship during visual stimulation in healthy humans. Magn Reson Med 2014; 72: 864–875. 74. Mark CI and Pike GB. Indication of BOLD-specific venous flow-volume changes from precisely controlled hyperoxic vs. hypercapnic calibration. J Cereb Blood Flow Metabol 2012; 32: 709–719. 75. Hoge RD, Atkinson J, Gill B, et al. Investigation of BOLD signal dependence on cerebral blood flow and oxygen consumption: the deoxyhemoglobin dilution model. Magn Reson Med 1999; 42: 849–863. 76. Griffeth VE and Buxton RB. A theoretical framework for estimating cerebral oxygen metabolism changes using the calibrated-BOLD method: modeling the effects of blood volume distribution, hematocrit, oxygen extraction fraction, and tissue signal properties on the BOLD signal. Neuroimage 2011; 58: 198–212. 77. Ances BM, Leontiev O, Perthen JE, et al. Regional differences in the coupling of cerebral blood flow and oxygen metabolism changes in response to activation: implications for BOLD-fMRI. Neuroimage 2008; 39: 1510–1521. 78. Chiarelli PA, Bulte DP, Piechnik S, et al. Sources of systematic bias in hypercapnia-calibrated functional MRI estimation of oxygen metabolism. Neuroimage 2007; 34: 35–43. 79. Leontiev O and Buxton RB. Reproducibility of BOLD, perfusion, and CMRO2 measurements with calibratedBOLD fMRI. Neuroimage 2007; 35: 175–184. 80. Rodgers ZB, Englund EK, Langham MC, et al. Rapid T2- and susceptometry-based CMRO2 quantification with interleaved TRUST (iTRUST). Neuroimage 2015; 106: 441–450.

1184 81. Xu F, Uh J, Brier MR, et al. The influence of carbon dioxide on brain activity and metabolism in conscious humans. J Cereb Blood Flow Metabol 2011; 31: 58–67. 82. Chiarelli PA, Bulte DP, Wise R, et al. A calibration method for quantitative BOLD fMRI based on hyperoxia. Neuroimage 2007; 37: 808–820. 83. Bulte DP, Chiarelli PA, Wise RG, et al. Cerebral perfusion response to hyperoxia. J Cereb Blood Flow Metabol 2007; 27: 69–75. 84. Gauthier CJ and Hoge RD. Magnetic resonance imaging of resting OEF and CMRO(2) using a generalized calibration model for hypercapnia and hyperoxia. Neuroimage 2012; 60: 1212–1225. 85. Blockley NP, Griffeth VE, Simon AB, et al. Calibrating the BOLD response without administering gases: comparison of hypercapnia calibration with calibration using an asymmetric spin echo. Neuroimage 2015; 104: 423–429. 86. Shu CY, Herman P, Coman D, et al. Brain region and activity-dependent properties of M for calibrated fMRI. Neuroimage 2016; 125: 848–856. 87. Blockley NP, Griffeth VE, Simon AB, et al. A review of calibrated blood oxygenation level-dependent (BOLD) methods for the measurement of task-induced changes in brain oxygen metabolism. NMR Biomed 2013; 26: 987–1003. 88. Yablonskiy DA and Haacke EM. An MRI method for measuring T2 in the presence of static and RF magnetic field inhomogeneities. Magn Reson Med 1997; 37: 872–876. 89. Stables LA, Kennan RP and Gore JC. Asymmetric spinecho imaging of magnetically inhomogeneous systems: theory, experiment, and numerical studies. Magn Reson Med 1998; 40: 432–442. 90. Ma J and Wehrli FW. Method for image-based measurement of the reversible and irreversible contribution to the transverse-relaxation rate. J Magn Reson B 1996; 111: 61–69. 91. An HY and Lin WL. Quantitative measurements of cerebral blood oxygen saturation using magnetic resonance imaging. J Cereb Blood Flow Metab 2000; 20: 1225–1236. 92. He X, Zhu M and Yablonskiy DA. Validation of oxygen extraction fraction measurement by qBOLD technique. Magn Reson Med 2008; 60: 882–888. 93. Lee JM, Vo KD, An H, et al. Magnetic resonance cerebral metabolic rate of oxygen utilization in hyperacute stroke patients. Ann Neurol 2003; 53: 227–232. 94. Christen T, Pannetier NA, Ni WW, et al. MR vascular fingerprinting: a new approach to compute cerebral blood volume, mean vessel radius, and oxygenation maps in the human brain. Neuroimage 2014; 89: 262–270. 95. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature 2013; 495: 187–192. 96. Yablonskiy DA, Sukstanskii AL and He X. Blood oxygenation level-dependent (BOLD)-based techniques for the quantification of brain hemodynamic and metabolic properties – theoretical models and experimental approaches. NMR Biomed 2013; 26: 963–986.

Journal of Cerebral Blood Flow & Metabolism 36(7) 97. Dickson JD, Ash TWJ, Williams GB, et al. Quantitative phenomenological model of the BOLD contrast mechanism. J Magn Reson 2011; 212: 17–25. 98. Lemasson B, Christen T, Serduc R, et al. Evaluation of the relationship between MR estimates of blood oxygen saturation and hypoxia: effect of an antiangiogenic treatment on a gliosarcoma model. Radiology 2012; 265: 743–752. 99. Oja JM, Gillen JS, Kauppinen RA, et al. Determination of oxygen extraction ratios by magnetic resonance imaging. J Cereb Blood Flow Metab 1999; 19: 1289–1295. 100. van Zijl PC, Eleff SM, Ulatowski JA, et al. Quantitative assessment of blood flow, blood volume and blood oxygenation effects in functional magnetic resonance imaging. Nat Med 1998; 4: 159–167. 101. Wright GA, Hu BS and Macovski A. Estimating oxygen saturation of blood in vivo with MR imaging at 1.5 T. J Magn Reson Imaging 1991; 1: 275–283. 102. Lu H, Xu F, Grgac K, et al. Calibration and validation of TRUST MRI for the estimation of cerebral blood oxygenation. Magn Reson Med 2012; 67: 42–49. 103. Golay X, Silvennoinen MJ, Zhou J, et al. Measurement of tissue oxygen extraction ratios from venous blood T(2): increased precision and validation of principle. Magn Reson Med 2001; 46: 282–291. 104. Xu F, Ge Y and Lu H. Noninvasive quantification of whole-brain cerebral metabolic rate of oxygen (CMRO2) by MRI. Magn Reson Med 2009; 62: 141–148. 105. Liu P, Dimitrov I, Andrews T, et al. Multisite evaluations of a T2-relaxation-under-spin-tagging (TRUST) MRI technique to measure brain oxygenation. Magn Reson Med 2015; 75: 680–687. 106. Xu F, Uh J, Liu P, et al. On improving the speed and reliability of T2-relaxation-under-spin-tagging (TRUST) MRI. Magn Reson Med 2012; 68: 198–204. 107. Liu P, Xu F and Lu H. Test-retest reproducibility of a rapid method to measure brain oxygen metabolism. Magn Reson Med 2013; 69: 675–681. 108. Xu F, Liu P, Pascual JM, et al. Effect of hypoxia and hyperoxia on cerebral blood flow, blood oxygenation, and oxidative metabolism. J Cereb Blood Flow Metab 2012; 32: 1909–1918. 109. Xu F, Liu P, Pekar JJ, et al. Does acute caffeine ingestion alter brain metabolism in young adults? Neuroimage 2015; 110: 39–47. 110. Chapman SB, Aslan S, Spence JS, et al. Shorter term aerobic exercise improves brain, cognition, and cardiovascular fitness in aging. Front Aging Neurosci 2013; 5: 75. 111. Chapman SB, Aslan S, Spence JS, et al. Neural mechanisms of brain plasticity with complex cognitive training in healthy seniors. Cereb Cortex 2015; 25: 396–405. 112. Liu P, Huang H, Rollins N, et al. Quantitative assessment of global cerebral metabolic rate of oxygen (CMRO2) in neonates using MRI. NMR Biomed 2014; 27: 332–340. 113. Peng SL, Dumas JA, Park DC, et al. Age-related increase of resting metabolic rate in the human brain. Neuroimage 2014; 98: 176–183.

Rodgers et al. 114. Ge Y, Zhang Z, Lu H, et al. Characterizing brain oxygen metabolism in patients with multiple sclerosis with T2-relaxation-under-spin-tagging MRI. J Cereb Blood Flow Metab 2012; 32: 403–412. 115. Liu J, Zhu YS, Khan MA, et al. Global brain hypoperfusion and oxygenation in amnestic mild cognitive impairment. Alzheimer’s Dement 2014; 10: 162–170. 116. Wong E and Guo J. VENTI: venous territory imaging using remote sensing. In: Proceedings of the 21st Annual ISMRM Scientific Meeting, Salt Lake City, Utah, ISMRM, 2013. 117. Wong EC, Cronin M, Wu WC, et al. Velocity-selective arterial spin labeling. Magn Reson Med 2006; 55: 1334–1341. 118. Cerdonio M, Morante S and Vitale S. Magnetic susceptibility of hemoglobins. Methods Enzymol 1981; 76: 354–371. 119. Spees WM, Yablonskiy DA, Oswood MC, et al. Water proton MR properties of human blood at 1.5 Tesla: magnetic susceptibility, T(1), T(2), T*(2), and nonLorentzian signal behavior. Magn Reson Med 2001; 45: 533–542. 120. Yablonskiy DA. Cerebral metabolic rate in hypercapnia: controversy continues. J Cereb Blood Flow Metab 2011; 31: 1502–1503. 121. Jain V, Abdulmalik O, Propert KJ, et al. Investigating the magnetic susceptibility properties of fresh human blood for noninvasive oxygen saturation quantification. Mag Reson Med 2012; 68: 863–867. 122. Li L and Leigh JS. Quantifying arbitrary magnetic susceptibility distributions with MR. Mag Reson Med 2004; 51: 1077–1082. 123. Wang Y and Liu T. Quantitative susceptibility mapping (QSM): Decoding MRI data for a tissue magnetic biomarker. Mag Reson Med 2014; 73: 82–101. 124. Ferna´ndez-Seara MA, Techawiboonwong A, Detre JA, et al. MR susceptometry for measuring global brain oxygen extraction. Mag Reson Med 2006; 55: 967–973. 125. Haacke EM, Lai S, Reichenbach JR, et al. In vivo measurement of blood oxygen saturation using magnetic resonance imaging: a direct validation of the blood oxygen level-dependent concept in functional brain imaging. Hum Brain Mapp 1997; 5: 341–346. 126. Li C, Langham MC, Epstein CL, et al. Accuracy of the cylinder approximation for susceptometric measurement of intravascular oxygen saturation. Mag Reson Med 2012; 67: 808–813.

1185 127. Jain V, Langham MC, Floyd TF, et al. Rapid magnetic resonance measurement of global cerebral metabolic rate of oxygen consumption in humans during rest and hypercapnia. J Cereb Blood Flow Metab 2011; 31: 1504–1512. 128. Jain V, Buckley EM, Licht DJ, et al. Cerebral oxygen metabolism in neonates with congenital heart disease quantified by MRI and optics. J Cereb Blood Flow Metab 2014; 34: 380–388. 129. Rodgers ZB, Jain V, Englund EK, et al. High temporal resolution MRI quantification of global cerebral metabolic rate of oxygen consumption in response to apneic challenge. J Cereb Blood Flow Metab 2013; 33: 1514–1522. 130. Rodgers ZB, Leinwand SL, Keenan BT, et al. Cerebral metabolic rate of oxygen in obstructive sleep apnea at rest and in response to breath-hold challenge. J Cereb Blood Flow 2015; 36: 755–767. 131. Kushida CA, Nichols DA, Quan SF, et al. The apnea positive pressure long-term efficacy study (APPLES): rationale, design, methods, and procedures. J ClinSleep Med 2006; 2: 288–300. 132. Langham MC, Magland JF, Floyd TF, et al. Retrospective correction for induced magnetic field inhomogeneity in measurements of large-vessel hemoglobin oxygen saturation by MR susceptometry. Magn Reson Med 2009; 61: 626–633. 133. Fan AP, Bilgic B, Gagnon L, et al. Quantitative oxygenation venography from MRI phase. Magn Reson Med 2014; 72: 149–159. 134. Haacke EM, Tang J, Neelavalli J, et al. Susceptibility mapping as a means to visualize veins and quantify oxygen saturation. J Magn Reson Imaging 2010; 32: 663–676. 135. Lin MT and Beal MF. Mitochondrial dysfunction and oxidative stress in neurodegenerative diseases. Nature 2006; 443: 787–795. 136. Warburg O. On the origin of cancer cells. Science 1956; 123: 309–314. 137. Barhoum S, Rodgers ZB, Langham M, et al. Comparison of MRI methods for measuring wholebrain venous oxygen saturation. Magn Reson Med 2015; 73: 2122–2128. 138. Fan AP, Scha¨fer A, Huber L, et al. Baseline oxygenation in the brain: correlation between respiratorycalibration and susceptibility methods. Neuroimage 2015; 125: 920–931.

MRI-based methods for quantification of the cerebral metabolic rate of oxygen.

The brain depends almost entirely on oxidative metabolism to meet its significant energy requirements. As such, the cerebral metabolic rate of oxygen ...
2MB Sizes 0 Downloads 14 Views