William Emmanuib
J.
Drasler, PhD I. Protonotarios,
Rheolytic Removal
#{149} Mark
MS
L. Jenson, #{149} Robert
Catheter for of Thrombus’
The authors present a percutaneous thrombectomy system (rheolytic thrombectomy catheter [RTCJ) in which high-velocity jets of saline solution are used to lyse and remove thrombus. The catheters (4-6 F) direct a 10,000-15,000-psi (0.7-1.05 x 10kPa) jet of saline solution onto an exhaust port from orifices at the end of the catheter. The jet entrains clot and resulting fragments and brings them into the high-velocity region for lysis and removal. Whole blood clots (10-15 cm) placed in 6-9-mm-diameter tubing were completely dissolved and removed with the RTC in less than 1 minute. In vivo use in a canine model resulted in lysis and removal of clots from a femoral artery, without vessel damage. The small caliber, flexibility, and effective lysis of this system suggest its potential usefulness in large central vessels that are difficult to access surgically and in small-diameter vessels that require more rapid removal of thrombus than can be achieved with thrombolytic therapy. Index terms: Catheters technology #{149} Thrombolysis, bosis, experimental Radiology
1992;
and
catheterization, 9*12992
#{149} Throm-
182:263-267
R
J.
BS #{149} Gregory G. Dutcher, MS
Wilson,
MD, FRCPC #{149} Joseph C. Possis, BMechE
#{149} Zinon
Percutaneous
EMOVAL
veins,
of thrombus and vascular
creasingly common acute ischemia and
venous outflow five to balloon bytic therapy tomy, alone angioplasty,
from arteries, grafts is in-
for treatment reestablishment
of of
and is often adjuncangioplasty. Thrombo-
on mechanical or in combination anticoagubation,
thrombecwith or other
procedures, is becoming more cepted as a treatment for acute nary being
accoro-
and peripheral ischemia examined as treatment
vein thrombosis and bolism. Thrombobytic streptokinase,
and is for deep
pulmonary therapy
urokinase,
emwith
or another
plasminogen activator is the most common form of such treatment but the risk of hemorrhagic complications, along with the length of treatment time, limits the use of thrombobytic
drugs
(1).
Various
mechanical
thrombectomy methods are also used, including the time-honored Foganty balloon embolectomy procedure and other newer catheter systems. Several mechanical systems ablate thnombus but and
do not others
bus
with
remove attempt suction
the debris to remove
due
to size
We developed small,
(2-4), throm-
or mechanical
(5-8). Some of these devices ited in the channel size that necanabized or are otherwise some
or stiffness
a method
high-velocity
jets
means
are limcan be cumben-
of saline
same From neapolis,
Possis Medical, 8325 10th MN 55427 (W.J.D., M.L.J.,
R.G.D.,
Z.C.P.),
and Toronto
General
Ave N, MmJ.M.T., E.I.P.,
Hospital
and Hospital for Sick Children, Toronto (G.J.W.). Received January 14, 1991; revision requested February 25; final revision received July 22; accepted August 12. Address reprint requests to W.J.D. W.J.D. is vice president of research and development for Possis Medical, the company that developed and will manufacture the thrombectomy device described in this article. 2 9* indicates generalized vein and artery involvement. ,,
RSNA,
1992
percutaneous
catheter.
The
sobu-
apart. The fragments of thrombus rccirculate in this region and repeatedly pass near the jets until they are bro-
into
smaller
and
smaller
remnant
pass
into
particles.
thrombus an
particles
exhaust
lumen
then
near
the
catheter tip and are removed through the catheter. An additional jet impinges on the exhaust lumen and provides a barge stagnation pressure to aid in the evacuation of the thrombus debris and to further break up any fragments that may be too large to pass
into
the
lumen.
The jets are oriented such that direct close-range impingement onto the vessel wall is avoided; thrombus removed from the vessel wall primariby by
the
fluid
tion.
A balloon
helps
to center
mixing
and
at the
catheter
the
tip
ments.
The
source
pressure
(2.1
x i0
device
tip
to keep
can
the
be used
of at least
kPa)
and
can
is
necircula-
away from the vessel walls, and lates the region, thus minimizing potential for escape of thrombus
jets
isothe fragat a
30,000
cut
psi,
a variety
of materials at the higher pressures. Because high pressures arc not necessary to break apart thrombus and because
they
could
be dangerous
to the
vessel wall, the RTC is used at up to approximately 15,000 psi (1.05 x 10 kPa) for thrombectomy. This RTC system has shown promisc in initial
in vitro
iments
may
and oven
present testing of the
other
report of the possible
and
offer
in vivo
clinical
methods.
exper-
advanThe
describes the initial RTC and suggests some applications of this
system.
mul-
tiple small-diameter jets at the tip of a rheobytic thrombectomy catheter (RTC) are directed into a thrombus. These high-velocity jets have sufficient energy to cut into the thrombus and create a region of intense fluid mixing in the region of the thrombus, which helps to break the thrombus
ken
The
tages
(5,7).
in which
tion from a catheter tip are used to break up thrombus and allow withdrawal of the resulting debris via the I
M. Thielen
MATERIALS Device
AND
METHODS
Construction were
fabricated
of 4-6-F
2.0-mm
outer
diameter)
dual-lumen
ing
1). The
RTCs (Fig
the inflation
Abbreviations: rafluonoethylene, tomy catheter.
smaller
lumen
ePTFE RTC
lumen
for a latex
= =
expanded rheolytic
(1.3tub-
was
used
balloon
as
(to
polytetthrombec-
263
RHEOLYTIC THROMBECTOMY
SYSTEM
Figure 1. RTC is 2 mm in diameter and contains seven jet orifices at the tip. Six equally spaced jets form a conical pattern around the catheter and provide the necessary circulalion patterns for clot lysis, and one jet impinges on the exhaust lumen to aid in debris withdrawal. The guide wire and exhaust lumen (the same lumen) can be seen; no centening balloon is shown.
Rheolytic center
the
also
catheter
coaxially
contained
stainless-steel larger lumen debris lumen.
tip in the vessel) high-pressure
tubing provided
to supply the jets. A for the exhaust of
and was also The catheter
eight
jet orifices
(0.001-0.002
that
were
inches)
impinged debris
used as the tip contained
guide-wire four
25-50
evacuation;
the
to
One
lumen
RTC
system
remaining
is shown
supplied
to the
catheter
jets
were
in Figure
pressurized and
was
saline activated
was
with
minor
capable
of providing
saline solution to 30,000 psi in the present
of the
to the
per
passed through of exhaust woubd
sion. Although exhaust lumen exhaust
vent
and
sufficient
flow
flow,
exhaust
a roller equal
pump that
of
catheter at up The catheters used 50 mL
port so the of saline
the
pump
flow
was
through
used
atrate infu-
to pre-
the exhaust
lumen.
In Vitro
Testing
eter,
264
(ePTFE)
6 mm)
vascular
and
#{149} Radiology
to obtain
at one the pores
better
was advanced functioning
was
grafts
allowed
clot,
in the The
inflated
so they
some each
would
pressure to) tube or graft,
ten was The
used
catheter
procedures
same materials embolectomy
four farther
advanced
In cathe-
times.
used
through
the whole length. The was opened longitudinally
approximately
two
more
through
the it was times
tubing or and pho-
tographed.
which and guidepump
debris;
Small arteries,
were
distal
(inner
diam-
sd branches bosis in the
to clot.
The
were
used
placed
to the
graft,
grafts femoral anThe grafts in diameter.
on the and
femorab
nearby
yes-
were ligated to ensure thromgraft and vessels. No additives to aid
clamps
was
inserted
the femorab
artery
artery
and
the
in thrombus
formation.
were either
or graft into
the
or fem-
oral artery by advancement with a guide catheter under fluoroscopic guidance. Aften the thrombus was reached, the RTC was advanced through the vessel with the jets functioning; the thrombus was lysed and removed by using the RTC at 10,00015,000 psi (0.7-1.05 x io kPa). Heparin was not used in the saline solution supplied to the RTC, but was administered intravenously thrombosis
after
ing balloon
was
to avoid any thrombobysis.
inflated
artifactual The
during
ation, and cane was taken inflation, so that the chance
from
bnis
passage
saline
center-
RTC
open-
to avoid ovenof vessel
the centering
balloon
be minimized. The centering served to isolate the debris
would
balloon and block
also de-
the vessel; the infused mixed with the debris and
along
solution
carried it out the exhaust lumen. A balloon embolectomy-thrombectomy
was
performed
in the
crab vessel as a control; artery on the graft was
Testing
clamps
distal
RTC
a carotid
procedure
In Vivo
the an
into
damage
each
passed through or graft, and
then graft
apply
material walls. embolectomy
approximately
it could be length of tubing
used as were
(and
was
until
whole
touch
the the
per-
were balloons
days,
and
directly
through the and the centering
embolectomy
formed a control.
3-7
After removed,
adhesion.
Bilateral ePTFE intcrpositionab were placed in the superficial teries of six 28-38-kg canines. were 3-5 cm long and 5 mm
Fresh animal blood was placed in woyen polyester tubing (inner diameter, 8 mm) and expanded polytetrafluoroethylene
RTC jets
time,
was
clamped through
the clot was allowed to age for 1-7 the RTC was used to lyse the clot.
Balloon
the jet impingement on the was sufficient to drive the
excessive
used to limit the flow of exhausted thrombus the catheter to a disposable collection bag.
balloon inflated, and the exhausted debris was collected. The clamps were left in place to ensure that the catheter did not simply push the clot.
a
The steady
minute at 10,000x i0 kPa). Tubing
catheter
days, The with
grafts were was forced
material
After
solution
variation
to supply the (2.1 x 10 kPa). study typically
of saline solution 15,000 psi (0.7-1.05
tached
cyclic
and also drives the roller pump thrombus debris is carried from
tubing and end; blood
2. A
with
footswitch by the catheter operator. pump provided an approximately pressure
A bag of sterile saline solution supplies a disposable pump, and supplies the RTC. The centering balloon inflation port shown. A nondisposable unit drives the disposable high-pressure
jet
standard saline solution bag was used for the supply, and a similar empty bag was used to coblect the exhausted thrombus debris. A disposable positive-displacement pump
RTC system. the solution access are
to aid in
oriented at a retrograde angle to avoid direct contact with the vessel wall while providing the necessary mixing and recircubation for effective clot lysis. The
Figure 2. pressurizes wine lumen
im
in diameter.
on the exhaust
Thrombectomy Catheter
and
Radiographic suits were
and
After
explanted,
the
femoral access.
procedures
procedure,
stained
collected during was examined
animal,
for
contrabat-
of the ne-
control
and examined with electron microscopy.
one
the
documentation
of study
obtained.
either used
RTC
with
was
the vessels Evans
blue,
bight and scanning Distal effluent was use
in
two
cases
and
for particulate matter. In the RTC was passed at 10,000
January
1992
dispensed, thrombus
diffuse residual mural was apparent on radio-
graphs
obtained
after
the
balloon
cedure; back of fluonoscopic and evaluation during the may
have
contributed
thrombus.
Figure
graphic
a.
b.
Figure 3. Radiographs documenting thrombectomy performed with the (a) Left femonal stump indicates occluded femoral artery and ePTFE graft; seen distally verify that an adequate amount of contrast medium has been diate results show patent artery and graft. Graft anastomoses can be seen.
All
experiments
animal
care
RTC in a dog. small branches infused. (b) Imme-
complied
and
use
with
guidelines
(9).
in the polyester grafts resulted
in
bysis of the clot in each case. A typical 10-cm-long clot was lysed and removed at 10,000-13,000 psi (0.7Figure
4.
Femoral
artery and graft thrombectomy
planted 2 weeks after the RTC. The specimen Evans
blue
damage.
stain
was treated
to examine
The proximal
for
artery
cxwith
0.91
x i0
RTC
left very
with
bus, less than that left with the Fogarty procedure; the walls of even the 8-mm-diameter tubing were cleaned effectively with the 2-mm-diameter
endothelial
is at the left.
The 5-mm-diameter, 3-cm-long ePTFE graft indicates the size of the specimen. The nonendothelialized surface of the graft absorbed
the stain
and
served
as a positive
psi
(0.7
io
x
kPa)
without for graft
through
a manner
similar
bolysis, and immediately. was allowed
damage catheter
to that
the vessels In another to recover
used
ing.
Volume
182
#{149} Number
The reto
resulted from was passed in for
throm-
were explanted case, the animal for 2 weeks before
radiography was performed were explanted and stained nations of patency and the
1
minute.
The
remaining
throm-
and vessels for determiextent of heal-
Testing RTC
was
used
to byse and
rc-
move occlusive thrombi from eight vessels. The average age of the thrombus was 4 days (range, 3-7 days). The
average
pressure
used
was
12,000
psi
(0.84 x i0 kPa). The RTC was successfully used to remove thrombus and restore patency in all cases. The
RTC for
jets 1-2
were
typically
minutes
functioning
in each
loon
thrombectomy
used
to open
stain, whereas tery absorbed
four
vessel.
procedures vessels.
the
native
femonab
an-
of thrombus. attributable marks
seen
were
near
the
graft.
The
pattern
Bal-
were
of damage
that
was
consistent shearing Histologic
with the stretching and forces imposed by a balloon. and scanning electron mi-
croscopy
examination
strate
significant
in the vessels used. Figure vessel use.
did
damage in which 5 shows
not
intact
intima
the RTC was a small focal fi-
in an otherwise
with
demon-
to the
elastica
normal after
RTC
For most of the procedures, the RTC was contained in a guide catheten that helped to position the device and allowed for contrast medium delivery. In a few cases, the RTC was manipulated over a 0.014-inch guide wire, which demonstrated that the device can be used over a wire if neccssary. Debris
lumen clumps cells.
collected
from
the
exhaust
of the catheter consisted of of platelets and red blood Effluent
downstream ing
A randomly
initial
little stain and was free No vessel trauma was to RTC use, but clamp
brim deposit
RTC.
undisturbed
the manipulations placement on clotting,
determine whether the procedure. The
in one
little
control.
In Vivo
arteries, quired
kPa)
radio-
of the
occlusion and the immediate results of successful thrombectomy with the RTC; the graft anastomoses can be seen in Figure 3b. After the control balloon thrombectomy procedure was performed in the contrabateral vessels of several animals, the procedure was abandoned because of excessive bleeding, swelling, and other complications, which were possibly due to the procedure being performed in the area without sufficient time being allowed for healing after graft implantation. Explanted vessels after thrombectomy with the RTC showed little damage; the explanted graft and vessel obtamed from an animal 2 weeks after thnombectomy are shown in Figure 4. The nonendothebiabized surface of the ePTFE graft absorbed the Evans blue
stress
Testing
Use of the RTC tubes and cPTFE
residual
the
RTC was capable of removing 1-weekold thrombus from both the ePTFE graft and the native vessel. The vessels in which balloon thrombectomy was used showed a
standard
RESULTS In Vitro
to this
3 shows
demonstration
pro-
guidance procedure
the
blood
from RTC
procedure
that
the
was
collected
thrombus consisted
dunpri-
many of individual red blood cells with occasional platelet and red cell clumping (Fig 6). Radiology
#{149} 265
Figure
5.
Histologic
tery shows inner focal
elastic fibnin
wise
normal
of the
section
of femoral
ar-
intact
vessel wall and undamaged lamina after RTC use. A small deposit can be seen in the other-
figure
vessel.
The lumen
(original
is at the top
magnification,
x200). Figure 6. Light photomicnognaphs of exhausted debris RTC use. (a) Debris collected from the catheter exhaust seen, which indicates the scale (original magnification, downstream from the thrombus during thrombectomy blood cells (original magnification, x500).
DISCUSSION The potential for an effective
clinical applications thrombectomy system
are many. The ability of the RTC to lyse thrombi that were at least one
week ness
old in vivo in a variety
ability
to use
ously
offers
indicates
its useful-
of situations.
the
device
reduced
The
pencutane-
time
and
cost
of
treatment and less surgical trauma. The RTC may result in less vessel trauma than do Foganty-type balloon embolectomy procedures, which can produce various vessel injuries (10,1 1), and percutaneous access through an easily accessible periphcrab vessel combined with fluonoscopic
guidance
may
allow
the
treat-
ment of thrombi that are not easily treated with balloon embobectomy. Removal of thrombotic debris rather than fragmentation of the thrombus and creation of multiple small emboli seems
desirable.
Although
controlled
studies of distal embobization to verify the benefit of debris evacuation with the RTC have not been performed, the exhaust of debris from the catheten tip combined with the breaking of thrombus into small fragments seems preferable to balloon embobectomy procedures. Further testing is under way to abten the
angle
of the
retrograde
jets
that generate the necincubation pattern. Jets that direct the flow radially without an axial component tend to restrict any axial particulate embolization reflux proximal to the catheter tip. The centering balloon on the catheten shaft aids in restricting proximal and distal embolization. Additional research to evaluate the need for an 266
#{149} Radiology
and effluent blood collected during lumen. Single red blood cells can be x200). (b) Effluent blood collected consisted primarily of individual red
isolation catheter through which the RTC would be delivered is currently under way. This isolation catheter would contain an inflatable balloon affixed to its tip and could be positioned to protect side branches from possible embobization and to isolate the vessel from arterial blood flow during the rheolytic procedure. Although heparin anticoagulation would likely be used perioperativeby, the occurrence of bleeding complications
may
be less
lytic agents, due the RTC. Indeed, useful in patients lytic
drug
avoided risk
than
treatment
because
with
thrombo-
to the local action the RTC may be in whom thromboshould
be
of its associated
of hemorrhagic
of
high
complications.
In
addition, whereas thrombobytic drugs often require many hours to induce complete thrombolysis (1,12), treatment with the RTC may provide rapid symptomatic relief in a matter of a few minutes. The combination of thrombolytic drugs and the RTC is theoretically appealing. The RTC can serve to quickly
break
up
and
remove
barge
thrombi, and thrombolytic drugs would be effective on any remaining smaller thrombi, including debris that may not have been removed through the catheter and thrombi in distal yessels
that
are
too
small
for
access
with
the RTC. The addition of thrombolytic on anticoagulant drugs to the saline solution used with RTC may be a convenient means of delivering the drugs
and achieving tions at the
Other
appear
may
system. vessels
diameter is more
local concentrasite.
mechanical
devices
that
high treatment
thrombectomy
to have
shortcomings
be addressed The much
may useful
with
ability larger
of the than
indicate than
the
RTC
RTC to clear the catheter
that other
the
system
devices
that
produce channels of a size only simibar to that of the catheter (2,3,5-7). Treatment of an entire lesion with a single insertion of the RTC may be possible
due
to the
breaking
up
and
removal of thrombus through the catheter, and would thus avoid the repeated withdrawal and reinsertion required with some aspiration devices (6,7). Preliminary research with rheobytic catheters of similar size and flexibility has
demonstrated
catheters vessels possible
moval arteries follow
to gain in animals,
the
ability
access which
application
of these
to coronary suggests
of the
RTC
the in re-
of thrombus from coronary (13). The ability of the RTC a guide wire provides access
distal vascubatune and in peripheral applications Early in vitro testing
to to
side
branches as webb. demonstrated
that when the jets were spaced closer together, fewer passes with the device were needed to effectively clean the tubing walls. The earlier four-jet prototype
rheobytic
catheters
were
there-
fore replaced with catheters with up to seven jets. The present study did not attempt to quantify any differJanuary
1992
ences between the prototypes either in vitro or in vivo. The exhaust ability of the catheters is affected by the cxhaust lumen size, but other factors such
as jet
size
and
position
are
also
important. Smaller catheter tubing allows the use of smaller guide cathetens and sheaths; a barge guide catheten was used with all RTC prototypes in this study, but this technique could be optimized with smaller guide catheters used with smaller RTC prototypes in future studies. The variation of catheter tubing diameter was more influenced by the availability of suitable tubing than by choice; the flexibibity, handling ease, and exhaust ca-
Acute pulmonary emboli that require treatment present a variety of undesirable choices. Open-vessel surgicab embolectomy is associated with great operative risk (14), and thrombolytic drug therapy is time-consuming and involves the risk of bleeding complications. Rapid treatment of barge pulmonary emboli by a percutaneous device has been advocated (5,14), and the RTC may offer fast, effective treatment and thus reduce
2.
mortality
6.
and
morbidity.
In addition,
may follow (13). The RTC may be useful for treatment of vascular graft thrombosis, including saphenous vein
immediate lysis of any large thrombi in the deep veins by using the RTC may be desirable, to reduce the likeihood of recurrent pulmonary emboli. The ability of the RTC to follow a guide wire and gain access to coronary vessels suggests its possible use in acute myocardial infarction, after sufficient preclinical and clinical testing has been performed. In conclusion, the RTC system is effective for percutaneous bysis and removal of thrombus. The ability of the RTC system to lyse thrombus up to 1 week old has been demonstrated both in vitro and in vivo in canine arteries and in vascular grafts. Initial vessel examinations from short-term in vivo studies were favorable, and follow-up in one animal demonstnated patent vessels and ePTFE vascuban graft 2 weeks after thrombectomy with this technique. The size and flexibility of the catheter and its ability to follow a guide wine suggest possible application to the treatment of acute coronary thrombosis. #{149}
or synthetic bypass venous hemodialysis
References
pacity
as a function
of catheter
were not quantified the prototypes used cessfully
exhausted
tubing
in this study, but in this study sucdebris in each
case. Infusion of contrast material through the exhaust lumen of the RTC allows assessment of the vessel without the use of a separate catheter. Use of fluoroscopy with the RTC may provide for timely treatment of the underlying disease. In the case of an arterial atherosclerotic stenosis, the thrombus may be rapidly and safely removed with the RTC, providing initial relief and aiding in nadiobogic examination of the stenosis, and standard dilation balloon angioplasty or revasculanization or mechanical atherectomy
tion
or rheobytic
to treatment
atherectomy
grafts and anteniografts, in addi-
of the
native
artery.
1.
Eisenbud
Treatment
DE, Brener
of acute
BJ, Shoenfeld
vascular
with intra-anterial urokinase. 1990; 160:160-165.
Volume
182
#{149} Number
1
3.
Kensey KR, Nash JE, Abrahams C, Zarins CK. Recanalization of obstructed arteries with a flexible, rotating tip catheter. Radiology 1987; 165:387-389. Bildsoe MC, Moradian GP, Hunter DW,
Castaneda-Zuniga WR, Amplatz K. Mechanical clot dissolution: new concept. Ra4.
5.
diology 1989; 171:231-233. Hong AS, Chae J, Dubin SB, Lee 5, Fishbein MC, Siegel RJ. Ultrasonic clot disruption: an in vitro study. Am Heantj 1990; 120: 418-422. Hawkins IF, Helms R, Spencer C, Hawkins MC. Mechanical spiral embolectomy
catheter.
7.
nipseed
8.
Semin
mt
Radiol
WD, Acher
CW, BurgessJH.
Per-
cutaneous aspiration thromboembolectomy. Radiology 1985; 156:61-66. Guenther RW, Vorwerk D. Aspiration
catheter for percutaneous clinical results. Radiology 9.
1985; 2:414-418.
Sniderman KW, Bodner L, Saddekni 5, Srur M, Sos TA. Percutaneous embolectomy by transcatheten aspiration. Radiology 1984; 150:357-361. Stark EE, McDermottJC, Crummy AB, Tur-
273. Guide
animals. thesda,
for the
care
and
NIH publication Md;
National
thrombectomy: 1990; 175:271use
of laboratory
no. 86-23. BeInstitutes
of Health,
1985. 10.
11.
Dobrin PB. Mechanisms and prevention of arterial injuries caused by balloon embolectomy. Surgery 1988; 106:457-466. Bowles CR, Olcott C, Pakter RL, Lombard C, Mehigan JT, Walter JF. Diffuse arterial narrowing as a result of intimal pnoliferation: a delayed complication of embolec-
tomy with the Foganty balloon Vase Sung 1988; 7:487-494. 12.
13.
14.
catheter.
Bang NU, Wilhelm OG, Clayman MD. Aften coronary thrombolysis and repenfusion, what next? J Am Coil Cardiol 1989; 14:837849. Drasler WJ,Jenson ML, Wilson GJ, et al. A rheolytic system for percutaneous coronary and peripheral plaque removal. Angiology 1991; 42:90-98. Greenfield U. Vena caval interruption and pulmonary embolectomy. Clin Chest Med 1984; 5:495-505.
R, et al.
occlusions Am J Sung
Radiology
#{149} 267